PHODS User S Manual 02 20
PHODS User S Manual 02 20
User’s Manual
Rev. 02/20
2 PHODS User’s Manual
COPYRIGHT NOTICE
This user’s manual is copyrighted, all rights reserved. It may not, in whole or in part, be copied,
photocopied, reproduced, translated, or reduced to any electronic medium or machine readable form
without prior consent in writing from Aqua Lung.
NOTICE OF NON-LIABILITY
Before the (PHODS) Portable Helicopter Oxygen Delivery System is put to use, it is the responsibility of
any user who will use this device to become familiar with the operation and safety aspects of this device.
Using the system improperly could cause a failure and lead to possible damage and/or personal injury.
Aqua Lung assumes no responsibility for damage, accidents, injury or death that may result from
the misuse of this device or equipment. This includes any use of this device or equipment outside
the scope of common sense, the instruction manual, inserts and other related documentation.
(Portions of this manual are excerpts from the MH OPC-M2 instruction manual, used by permission
of Mountain High Equipment & Supply Company®, Redmond, OR.)
CONTENTS
WARNING: Oxygen is a highly oxidizing gas and can vigorously accelerate combustion. It
can provide a catalyst for spontaneous combustion resulting in personal injury or death
if not used properly and with caution. DO NOT use any type of oil or grease on any of the
fittings, valves or cylinders. DO NOT smoke while in use. DO NOT operate near an open flame.
WARNING: This device is classified as, and is only suitable for use as, a supplementary
breathing apparatus (SBA) for aviation use up to 18,000 MSL. It is intended to supply
the needed amount of oxygen for persons during excursions at flight altitudes where
supplemental oxygen is needed. This device is not suitable for any type of medical
operations. This device is not suitable for SCBA (Self Contained Breathing Apparatus) or
SCUBA (Self Contained Underwater Breathing Apparatus) operations.
WARNING: When operating the PHODS around high power emitters such as friendly or
threat systems, PHODS may stop providing oxygen, or may start pumping oxygen when
not expected or required. If the unit is not operational, turning the OPC power OFF then
ON will be required. An operational check must therefore be performed immediately prior
to use at high altitudes.
WARNING: Before using the PHODS, it is important to receive hypoxia awareness and
aeromedical training. Use of the PHODS without proper training is dangerous and can
result in serious injury, or death.
WARNING: DO NOT apply any type of petroleum-based lubricant or grease to any part
of the PHODS. Hydrocarbon contamination, including petroleum-based lubricants, can
cause spontaneous combustion when exposed to compressed oxygen.
6 PHODS User’s Manual
WARNING: It is important to fill the cylinder only with Oxygen 90% or greater. If filling the
PHODS from any other source than a Deployable Oxygen Refill Station (DORS) ensure
that the moisture content does not exceed 0.005 milligrams of water vapor per liter of
gas at 760 millimeters of mercury (mmHg) at a temperature of 70ºF / 21.1ºC. Excess water
vapor in the breathing gas can cause ice to form inside the PHODS and interfere with the
operation of the system at colder temperatures.
WARNING: Do not attempt to loosen or remove the regulator CPC fitting or safety burst plug
on the regulator under any circumstances. This should only be done by a qualified technician.
WARNING: The oxygen cylinder may burst from impact of small arms fire, resulting in
overpressure shock, shrapnel injury and/or flash fire.
WARNING: DO NOT modify or alter the length of any hoses or tubing. This should only
be done by a qualified technician, using original manufacturer hose assemblies.
CAUTION: DO NOT apply any type of aerosol spray to the PHODS. Doing so may cause
permanent damage to certain plastic components, including the delivery tubing.
CAUTION: PHODS operation has not been verified below -25ºF / -32ºC and performance
may be degraded at lower temperatures.
NOTE: DO NOT store the OPC connected to the cylinder while the inlet is under pressure. If
the lines are disconnected the ends must be covered to prevent the entry of debris, dust or dirt
into clean hoses.
NOTE: Factory prescribed service for the PHODS must be performed every 12 months by a
factory trained service technician.
NOTE: When instructed to “OPEN” the handwheel, turn the handwheel counter-clockwise. the red
indicator ring will not be visible in the handwheel window, this indicates the system is in the “ON”
position. When instructed to “CLOSE” the handwheel, turn the handwheel clockwise. the red indicator
ring will be visible in the handwheel window, this indicates the system is in the“OFF” position.
NOTE: When instructed to disconnect or reconnect the OPC hose to the CPC fitting, use the
following method:
Disconnect: Depress the button on the CPC fitting to remove the OPC hose.
Reconnect: Insert the OPC hose into the CPC fitting, making sure the button “Pops Up”. Check that
the OPC hose is firmly attached prior to pressurizing the system.
If you have any questions or do not understand the information in these Warnings, Cautions, and
Notes, please contact an Aqua Lung Technical Advisor before proceeding.
7
PRODUCT DESCRIPTION
(PHODS) Portable Helicopter Oxygen Delivery System is an oxygen delivery system that can be attached
directly to a crew member’s vest and helmet to support missions at high altitudes.
The (PHODS) Portable Helicopter Oxygen Delivery System is made up of the following components:
• Regulator and oxygen cylinder.
• Automatic oxygen pulse controller with inlet and outlet hoses.
• Flexible nasal cannula w/ QD coiled hose assembly.
Cannula
Assembly
Nosepiece
First Stage
Oxygen
Cylinder
OPC Pocket
Oxygen Pulse
Controller
Figure 1
8 PHODS User’s Manual
OXYGEN
Figure 4
NOTE: Due to the small size of the dial indicator and the working pressure of 2100 psi (145
BAR), it is possible that the needle will only go just to the green or next to it. This is acceptable
provided the unit is charged on a calibrated filling system.
NOTE: While the dial indicator indicates a full cylinder on the ground, the colder temperatures at
higher altitudes will reduce the pressure in the cylinder. This is a normal occurrence. At higher
altitudes, aircrew members must check the cylinder pressure in the user worn cylinders more often.
9
Figure 5
4. Ensure the OPC pocket is installed correctly on the vest using the MOLLE system straps. Install the OPC into
the pocket. Insert the 90° QD fitting into the CPC connector on the regulator until it “clicks” into place (Figs. 6 & 7).
Figure 6 Figure 7
INSERTING
Push in the tubing
5. Connect outlet tube into OPC. To Insert Tubing:
Push the tubing into the connector until resistance is felt, REMOVING
1. Push in the connector
then push a little further, about 1/8 inch / 3mm (Fig. 8). collar.
2. Pull the tubing straight
Gently tug on the tubing to make sure it is captured. To back while holding the
Remove Tubing: Push the tubing in slightly, then push collar in.
in the connector collar while pulling gently on the tubing.
Figure 8
10 PHODS User’s Manual
CAUTION: DO NOT pinch the outlet hose when inserting it into the “out” connector.
CAUTION: When removing the tubing, DO NOT pull without first pushing in on the collar,
as this will likely damage the connector.
Figure 9
Figure 10
11
NOTE: If the nasal cannula mount is not installed in the helmet, it must be installed prior to the
use of PHODS.
Internal Mount
Figure 11
Figure 12
Figure 13
Figure 14
12 PHODS User’s Manual
External Mount
Figure 15
Figure 16
Figure 17
Figure 18
13
Notes:
• The OPC oxygen delivery schedule is calibrated to provide the oxygen normally required by a
healthy average-size person using a cannula at a given altitude. Individual needs may be different.
• In all modes, the OPC delivers more oxygen as altitude increases (altitude compensating).
• See OXYGEN FLOW RATES (Fig. 23) for a comparison of oxygen flow rates for the different OPC modes.
• Mode setting changes are indicated by a brief flash of the Battery LED.
Notes:
• Below 8,000 ft pressure altitude, no oxygen is dispensed, but the OPC continuously monitors breathing.
• When the barometric pressure is low, operation will begin at a slightly lower altitude than when the
barometric pressure is high.
• ON mode is the most efficient delivery setting and will provide the greatest oxygen duration.
14 PHODS User’s Manual
Notes:
• (¹) An extremely enriched but fixed amount of oxygen is delivered at all altitudes up to ~ 32,000 ft PA.
Above that, the amount of oxygen delivered increases with additional altitude (altitude compensating)
and is actually the same amount that would otherwise be delivered in ON mode in that altitude
range. See OXYGEN FLOW RATES (Fig. 23).
• At lower altitudes, much more oxygen will be used than with ON mode, and oxygen duration will be
significantly decreased. R/M is by far the least efficient delivery setting in the lower altitude range.
• R/M may also be considered “Recovery/Medical” mode as the flow provided is more on the order
of an emergency ventilation protocol. R/M may be used with discretion to “recover” from exertion or
to otherwise mitigate the effects of sudden hypoxia.
Notes:
• F20 setting augments the amount of oxygen otherwise provided at a given pressure altitude in order
to compensate for the additional plenum volume associated with a facemask.
• More oxygen will be used than with ON mode, and oxygen duration will be decreased. See OXYGEN
FLOW RATES (Fig. 23).
• Pilots making night approaches and landings may use the R/M or F20 modes to provide additional
oxygen in order to enhance night vision.
15
Figure 23
ON Mode
• Oxygen is only delivered above the 8,000 ft pressure altitude (PA) threshold.
• Oxygen delivery is suspended when the altitude drops below 8,000 ft PA.
• Above 8,000 ft PA, more oxygen is delivered as altitude increases (altitude compensating).
F20 Mode
• Oxygen is delivered at all altitudes.
• A fixed amount of oxygen is delivered at all altitudes below 8,000 ft PA.
• Above 8,000 ft PA, more oxygen is delivered as altitude increases (altitude compensating).
• The amount of oxygen delivered at any altitude is greater than what would otherwise be provided in
the ON mode (enriched flow).
R/M Mode
• Oxygen is delivered at all altitudes.
• A fixed amount of oxygen is delivered at all altitudes up to ~ 32,000 ft PA.
• The fixed amount of oxygen delivered in the lower altitude range is significantly greater than what
would otherwise be provided in either ON or F20 mode (extremely enriched flow).
• The amount of oxygen delivered above ~32,000 ft PA is the same as provided in ON mode, and
likewise, more oxygen is delivered as altitude increases (altitude compensating).
Notes:
• At any altitude, ON mode will always provide the best efficiency and greatest oxygen duration.
• F20 mode will always provide more oxygen than ON mode.
• At lower altitudes, R/M mode will provide the most oxygen (by far).
• The OPC will continue to operate above 32,000 ft PA, although at such high altitudes physiological
factors begin to compromise the effectiveness of an SBA (Supplemental Breathing Apparatus)
system such as the OPC.
16 PHODS User’s Manual
Figure 24
LEGEND
Figure 25
The Oxygen Flow Indicator LED (O2) displays conditions related to oxygen flow, while the Battery
Warning LED displays battery-related conditions. An initial oxygen pulse is delivered as part of the
Power-Up Test sequence. Oxygen pulses are subsequently delivered with each inhalation during
normal operation.
Notes:
• Status LEDs flash green for Night Vision Goggle (NVG) compatibility, and are therefore difficult to
see through the NVG. LEDs should be viewed directly (not through the NVG) and the OPC should
be located in such a manner as to make this convenient.
• OPC status should be checked periodically to verify proper operation.
17
The oxygen pulse will be evident from the sound of the valve opening, or can be felt in the nostrils if the
cannula has been donned. See PHODS PRE-FLIGHT CHECKLIST.
Power-Up Test: Fail (Bad Battery Alarm) (Fig. 27) 1 Sec 2 Sec 3 Sec (repeats)
Figure 27
Power-Up Test: Fail (Internal Error Alarm) (Fig. 28) 1 Sec 2 Sec 3 Sec (repeats)
An INTERNAL ERROR ALARM indicates that the OPC unit will not function properly and must be serviced.
18 PHODS User’s Manual
Figure 29
Exceptions:
• In ON mode but below 8,000 ft pressure altitude, an oxygen pulse will not be delivered and the Oxygen
LED will not flash.
• If the detected breathing rate exceeds 30 bpm (breaths-per-minute), the OPC delivers oxygen only on
every other breath.
Apnea Alarm / Abandoned Alert (Fig. 30) 1 Sec 2 Sec 3 Sec 4 Sec (repeats / 2 sec)
Figure 30
An APNEA ALARM is generated if the time since the last inhalation event exceeds an interval that is
dependent on the pressure altitude. This interval is ~30 seconds at lower altitudes, but much less at
higher altitudes where the onset of hypoxia is more rapid and its effects are more pronounced. The
APNEA ALARM does not respond below 8,000 ft pressure altitude when operating in ON mode.
Low Battery-1 Alert (First Warning) (Fig. 32) 1 Sec 2 Sec 3 Sec 4 Sec (repeats / 2 sec)
Battery Warning LED displays continuous single flashes.
From the time a LOW BATTERY-1 ALERT is first issued,
O
the OPC should continue to operate properly for about 2
another eight (8) hours (@ 77°F/25°C, and depending on Figure 32
battery quality).
LOW
BATTERY-1
Low Battery-2 Alarm (Second Warning) (Fig. 33) 1 Sec 2 Sec 3 Sec 4 Sec (repeats / 2 sec)
OPC BATTERY
Alerts (Fig. 34)
PRE-MISSION IN-MISSION
OPC
POWER-UP (Nominal) (Nominal)
TEST
Figure 34
Notes:
• Replace the batteries immediately if any of the LOW BATTERY Alerts are indicated during the
PHODS PRE-FLIGHT CHECKLIST. DO NOT embark on a mission with weak batteries. A set of
fresh AA alkaline batteries should always be included as part of a “spares” kit.
• If a mission is begun with a LOW BATTERY-1 Alert, the OPC unit may have no more than 2 hours
of operation remaining.
• If a mission is begun with a LOW BATTERY-2 Alarm, the OPC unit may be on the verge of imminent
shut down.
• If a LOW BATTERY-2 Alarm is received during a mission, the batteries should be replaced at the
first opportunity.
• If operation is continued with a LOW BATTERY-2 Alarm and subsequent FLOW-FAULT Alarms are
received, this likely indicates that the oxygen-dispensing valve has ceased operating, and the OPC
unit is no longer dispensing oxygen (see BATTERY LIFE AND DEPLETION).
• It is particularly important to heed BATTERY ALERTS in the course of missions conducted at higher altitudes.
• The times for remaining operation (“Time-to-Failure”) are approximations only, based on nominal
conditions (see BATTERY LIFE AND DEPLETION).
21
The OPC uses two (2) AA alkaline batteries. Remove the battery door
by gently pressing down on the battery cover tab, then slide the door
out and away from the unit. Insert the batteries with proper polarity as
shown on the label inside the battery compartment. Replace the door by
sliding it back in until it snaps into place.
Notes:
• Batteries will be a tight fit. Take care not to damage the batteries
and/or connectors when inserting or removing the batteries. Do not
short the battery terminals.
• Install the batteries with the proper polarity as shown.
−
• Use good quality alkaline batteries only (DURACELL ULTRA,
ENERGIZER MAX or equivalent). + +
• DO NOT USE LITHIUM BATTERIES! No damage will result, but
the LOW BATTERY Alerts WILL NOT function correctly as they are −
calibrated for alkaline batteries. Figure 35
• DO NOT mix old and new batteries! Replace all batteries at the same time.
• Remove the batteries during long periods of non-use. Battery leakage
and corrosion can damage the OPC.
• Fresh batteries should be installed prior to each mission, and
removed at the completion of each mission.
• Dispose of batteries properly. Do not burn. Use collection points
when available.
• See OPC-M2 SPECIFICATIONS for battery-life information.
22 PHODS User’s Manual
STORING PHODS
PHODS can be stored under a fairly wide temperature range 185°F (+85°C) to -79.6°F (-62°C). To protect
the system from excessive humidity/rain, etc. store the unit in the supplied case.
In practical use the following precautions should be observed:
1. The OPC unit, oxygen hoses, nasal cannula’s, etc., should be disconnected from the oxygen supply
and stored in a secure manner to ensure that dirt and debris do not enter the inlet and outlet tubes. The
supplied protective case is recommended.
2. Remove the batteries from the OPC unit when it is not in use.
3. Always store the cylinder with the handwheel in the closed position (red indicator ring visible). The
unit should not be stored with oxygen in the cylinder.
4. PHODS must be tested before use in accordance with the PHODS PRE-FLIGHT CHECKLIST.
TROUBLESHOOTING GUIDE
General Checks
WARNING - If after consulting the troubleshooting guide section of this manual the situation is
not resolved, immediately return the PHODS system to a qualified technician for inspection or
contact an Aqua Lung Technical Advisor.
OPC AUTO-COMPENSATE
Breathing Sensor and Altitude Compensation
As absolute atmospheric pressure decreases with altitude, breathing efforts consequently assert less
pressure on breathing sensors to the point where inhalation may not be properly detected at higher
altitudes. Additionally, breathing effort tends to diminish as the partial pressure of CO2 also decreases
with altitude. The OPC must compensate for these physical and physiological effects when ascending
to higher altitudes.
The OPC employs an active algorithm that constantly and automatically adjusts the sensitivity of the
breathing sensors based on pressure altitude and detected breathing effort. It also filters out false-
triggers due to pneumatic artifacts. This helps ensure that all breaths are correctly detected so that the
OPC can respond reliably and deliver the proper amount of oxygen. This function is entirely automatic
and involves no user settings.
OPTIONAL FACEMASK
The oxygen mask design includes the mounting brackets necessary to connect the mask to the HGU-56/P
helmet, with or without the maxillofacial shield installed. It may be used with the optional snap adapters
to connect the mask to the HGU-84/P helmet without the maxillofacial shield installed. This provides
versatility to air crew members allowing them to customize the system according to mission demands.
The system remains stable throughout the flight performance envelope of helicopters, is easy to don, and
has quick disconnect features. The mask assembly’s quick release snaps and adjustable fasteners allow
three unique configurations wherein the maxillofacial shield and oxygen mask can be used independently
or together for maximum protection.
The mask includes an integrated M-169A/AIC dynamic 5-ohm microphone to aid communications.
Use of a mask will supply an adequate oxygen mixture to users who are under mental or physical stress
and who, as a result, will breathe through their mouths instead of their noses. Oxygen enters from the
OPC regulator hose and travels through 90° oxygen inlet port of the mask to allow the oxygen to enter
the mask cavity. Exhaled gases exit the mask through the exhalation valve (Fig. 37). Ensure the GMM
Plug is inserted into the 90° elbow on the opposite side of the mask (Fig. 38).
WARNING - The GMM plug must be inserted into the 90° inlet port on the opposite side of the
mask in order to function properly.
WARNING - Oxygen should be delivered to the inlet port on the inhalation valve side of the mask
to ensure the proper dilution of oxygen with the ambient air in the mask cavity.
WARNING - Disinfect the facemask or nasal cannula & nose piece after each use to avoid bacteriological
contamination and for cleanliness. Cleaning can be accomplished with isopropyl alcohol.
GMM Plug
Inlet Port
Figure 37 Figure 38
29
Note: In ON mode, the APNEA ALARM does not respond below 8,000 ft
pressure altitude.
30 PHODS User’s Manual
Notes:
• Specifications and limits are characterized from test results, or derived from underlying specifications.
• Unit is not water-proof! Keep away from rain and spray.
• Nominal Battery Voltage/Current values measured in ON mode setting @ 15 bpm typical.
• Battery-life values assume fresh alkaline batteries and normal operating conditions.
• Fresh batteries should be installed prior to each mission, and removed at the completion of each mission.
• Use good quality alkaline batteries only. DO NOT use Lithium batteries.
• DO NOT mix old and new batteries! Replace all batteries at the same time..
• Remove batteries during long-term storage to prevent battery leakage and corrosion.
• The Battery Minimum ‘Start-Up’ voltage is the level at which the OPC unit will initiate the POWER-
UP TEST. If the battery level is too low, the OPC will immediately generate a BAD BATTERY
ALARM. Otherwise the OPC will proceed to issue the initial oxygen pulse. This not only exercises
the valve and verifies the integrity of the entire oxygen system, but also serves as a stress-test for
the batteries. If the battery level is too low following the initial pulse, the OPC will generate a BAD
BATTERY ALARM. Only when the OPC successfully passes all of these tests will it then commence
operating (with or without a LOW BATTERY Warning). This behavior should help the operator in
determining if the OPC unit is bad or if the batteries are just too low.
• If a BAD BATTERY ALARM is generated, the OPC unit will enter “lock-out” mode and will not function
other than to display the BAD BATTERY ALARM. Batteries must be replaced at this point for the OPC
unit to resume proper functioning. However, once the OPC is running, it will continue to operate as long
as possible until the batteries are completely exhausted (see BATTERY LIFE AND DEPLETION).
• The OPC POWER-UP TEST also verifies the integrity of internal components and sensors and
generates an INTERNAL ERROR ALARM (see POWER-UP TEST) if any of these checks fail. An
INTERNAL ERROR ALARM indicates that the OPC unit will not function properly and must be serviced.
31
MAINTENANCE NOTES
PHODS
Portable Helicopter Oxygen Delivery System