DrägerSensor AC e
DrägerSensor AC e
Gas permeable
membrane Measurement Electrode
Teflon housing
However, the sensor can only withstand a low differential pressure. The pressure difference between the gas space on the membrane and
the back of the sensor must not generally exceed ± 30 mbar. Some sensors only allow ± 5 mbar.
Sensor for CO
Electrochemical sensors:
Interference
• Cross-sensitivity to other gases can lead to a positive or
negative signal change
The Dräger lower detection limit is the concentration display, which with high probability is not due to the zero-point unsteadiness but to a gas measurement. The
detection limit only applies within moderate variations in ambient temperature and humidity conditions
Rule of thumb: zero unrest times two detection limits; detection limit times two smallest alarm threshold; Pre-alarm times two main alarms
occurance
zeo LDL pre alarm
mean 2ppm 4ppm 8ppm
zero
95%
50%
zero noise
sensor
distribution
signal
DrägerSensor AC | Oertel, Marcus | 24.07.2023
Basics of electrochemical sensors
T90 time:
• The t90 time is independent from the gas
high concentration
concentration and defines the time till 90%
concentration
of the final reading has been reached. mid concentration
• T63 is equal to the time constant tau in the linear time-invariant first-order system.
• A system is referred to as a linear time-invariant system if its behavior has both the
property of linearity and is independent of time shifts. This independence from time
shifts is called time invariance. This describes a mathematical model for physical
systems whose behavior suddenly changes exponentially.
t0 time
• Most gas detectors have an exponential response when gas is supplied. t63 t99
1 x τ (tau) = 63,2%; 2 x τ (tau) = 86,5%; 3 x τ (tau) = 95,0%; 4 x τ (tau) = 98,2%; 5 x τ (tau) = 99,3%
In the instructions for use for the electrochemical sensors you can find the alarm
response time for gas exposure:
t0 time
t63 t99
The factors for the alarm thresholds come from the standards for the performance requirements for measuring the concentration of
chemical substances in workplace atmospheres.
− An air-filled reserve volume is provided in the sensor for this purpose. Above 95% r.h. the reserve volume is no longer sufficient in the
long run and internal overpressure occurs, which leads to the destruction of the sensor.
− High humidity (> 95% r.h.) can lead to condensation on the membrane, which impedes gas access.
air
water
sulfuric acid
reversibler Prozess
air
KOH
Electrolyte
Measurement electrode
Pre-reaction:
6HF → 6F- + 6H+ HF gas
6H+ + 5Br- + BrO3- → 3Br2 ↑ + 3H2O
Cathode reaction: Wick
Displayed ppm HF
Open design
9
Sensor response is a function of flow
6
0.5 l/m = 0.6 m/s
3
Measured ppm HF
Constant flow
10
Linear sensitivity e.g., 0.5 L/min in
5
pumped operation.
1
1 5 10 ppm exposed
DrägerSensor AC | Oertel, Marcus | 24.07.2023
DrägerSensor AC
Working principle – „open sensor“:
During commissioning, the sensor must be filled with electrolyte, which
must also be replaced if the "calibration" provides insufficient signals, or the
electrode is contaminated with dirt or chemicals. Normally, the zero signal is
stable and has almost no drift. After high doses of gas, the span must be
checked
Excessive exposure to HF leads to deposition of white crystals, which are LiF from
reaction with the electrolyte.
Described in SI 31:
To remove the salt crystals, use a weak water jet and a smooth brush, e.g., a soft
toothbrush. Please use applicable gloves to avoid skin contact of the salt crystals.
Measurement accuracy
measurement uncertainly (of meas. value) or ≤ ± 20%
minimum (whichever is the greater value) ≤ ± 1 ppm
Alarm response time, on gas exposure with 5x alarm threshold, t0...20 ≤ 30 seconds
Alarm response time, on gas exposure with 1.6x alarm threshold, t0...63 ≤ 60 seconds
Displayed name Gas / Vapor Chemical formula CAS number calibration group Relative Sensitivity
HF Hydrogen fluoride HF 7664-39-3 A 1.0
HCl Hydrogen chloride HCl 7647-01-0 A 1.0
HBr Hydrogen bromide HBr 10035-10-6 A 1.0
BF3 Boron trifluoride BF3 7637-07-2 A 1.0
SiF4 Silicon tetrafluoride SiF4 7783-61-1 A 1.0
GeF4 Germanium tetrafluoride GeF4 7783-58-6 A 1.0
WF6 Tungsten hexafluoride WF6 7783-82-6 A 1.0
BCl3 Boron trichloride BCl3 10294-34-5 A 1.0
TeCS Silicon tetrachloride SiCl4 10026-04-7 A 1.0
DCS Dichlorosilane SiH2Cl2 4109-96-0 A 1.0
POC Phosphorous trichloride oxide POCl3 10026-87-3 A 1.0
PCl3 Phosphorous trichloride PCl3 7719-12-2 A 1.0
ClF3 Chlorine trifluoride ClF3 7790-91-2 B 3.0
Acid Antimony pentachloride SbCl5 7647-18-9 A 1.0
Acid Thionyl chloride SOCl2 7719-09-7 A 1.0
Acid Titanium tetrachloride TiCl4 7550-45-0 A 1.0
Acid Trichloro silane SiHCl3 10025-78-2 A 1.0
Acid Tin tetrachloride SnCl4 7646-78-8 A 1.0
Acid Hydrogen iodide HI 10034-85-2 A 1.0
Acid Acetic acid CH3COOH 64-19-7 A 1.0
Acid Formic acid HCOOH 64-18-6 A 1.0
Acid Boron tribromide BBr3 10294-33-4 A 1.0
Acid Chlorosulfonic acid HSO3Cl 7790-94-5 A 1.0
Acid Germanium tetrachloride GeCl4 10038-98-9 A 1.0
DrägerSensor AC | Oertel, Marcus | 24.07.2023
DrägerSensor AC
Polytron 8100 Menu:
Gas settings
This function sets the gas type, measuring range and units of measurement.
Not all sensors offer a selection of values.
1. Select Settings > Sensor > Gas setting and confirm.
The current measured gas is displayed.
2. Select a measured gas from the list and confirm.
The current unit of measurement is displayed.
Note: It is recommended that the pre-alarm should not be closer than twice the LDL and the main alarm 4 times LDL to avoid
false alarms.
Fluorocarbon Based
Hydrocarbon Based
Ethylene-Propylene
CAS
Gas
Stainless Steel
Buthyl Rubber
Nitrile Rubber
Number
Carbon Steel
Chloroprene
Aluminium
PCTFE
Copper
Silicon
PVDF
PTFE
Brass
FKM
PA
PP
Hydrogen Bromide 10035-10-6 4 4 4 2 2 1 1 1 4 0 4 4 4 1 4 1 4 4
Legend
No data 0
Satisfactory 1
Satisfactory but risk of corrosion in presence of moisture 2
Acceptable under certain conditions 3
23 m
Comparison of tubing length: Delay due to different 0.5 m
length 20 sec.
Material: FEP; øi 4.8 mm
HF step change
5.5 ppm
T50 = 45 s
T90 = 90 s
With calibration mask
Blue line:
Green line:
Result:
No falling asleep
No sensitivity loss
Use of the diagnostic dongle does not make sense with the DrägerSensor AC
DrägerSensor AC | Oertel, Marcus | 24.07.2023
DrägerSensor Self-Test
Sensor Self-Test
2. Evaluation:
• temperature compensation
Polytron 7000 will perform an initial sensor test during the warmup time when a sensor gets connected. It is checked whether all
electrodes are connected to the correspondingly or are conductive. The result of the initial SST is also stored in the sensor memory.
If the sensor is dry, meaning not filled with electrolyte, a sensor error is displayed:
The sensor must be disconnected, filled with electrolyte and reconnected to reset the error on the transmitter.
• Polytron 7000 Display “- - - -“ Warning Error #132 “Sensor test failure! Refill sensor”
• After refilling as described in the Start-up kit (6809381) manual you must wait approx. 40 minutes until the warning is canceled.
• If the sensor was electrically disconnected for the refill an additional warm-up time (ready for operation) must be accounted for.
• After electrolyte refill the sensor needs to be calibrated after Warm-up time (ready for calibration)!
• Despite the sensor self-test, the electrolyte must be checked and replaced regularly. If the sensor runs dry, the electrodes can be damaged. A sensor
with low or depleted electrolyte will not meet the performance specification!
Test Sequence
Initial SST failed before warm up time #134 #134 SST Fault saved in sensor memory
Periodical SST failed 20 minutes after 2 consecutive tests #132 SST Warning saved in sensor memory
Periodical SST failed 40 minutes after 4 consecutive tests #125 SST Fault saved in sensor memory
#132 Sensor electrolyte has evaporated. Measurements will soon no longer be possible. Replenish sensor electrolyte