UNIVERSAL TEMPERATURE TRANSMITTER WITH DISPLAY

SEM710

SIMPLE CONFIGURATION VIA USB PORT

ISOLATED PT100, THERMOCOUPLE INPUT

DISPLAY TEMPERATURE in °C /°F or OUTPUT DRIVE in mA

PUSH BUTTON USER TRIM

(4 to 20) mA TWO WIRE OUTPUT

HOUSING ABS PLASTIC OR STAINLESS STEEL

INTRODUCTION

The SEM710 is the first of a new generation head mounted temperature transmitter with display feature, from Status Instruments. It
has been designed to accept most common temperature sensor inputs and provide the user with a standard two wire (4 to 20) mA
output signal. Isolation is provided between input and output and all temperature ranges are linear to temperature. The addition of
a display provides the user with instant information of the loop condition at the point of measurement. It comes in either our SCH4
ABS plastic connector head or our new SCH15 Stainless steel connector head.

Designed for ease of use, our latest USB interface is fitted for quick and easy configuration. Just connect a standard USB cable
between the SEM710 and your PC. Using our free configuration software, your PC will automatically upload the existing configuration
data and guide you through any changes you wish to make. To further help save time, the SEM710 does not need to be wired to a
power supply during the configuration process, it is powered via the USB interface from your PC. The following parameters are
configurable :-

INPUT TYPE LOW RANGE HIGH RANGE UNITS BURNOUT USER TRIM DISPLAY

PT100
Up/Scale
Or Input @ 4 mA Input @ 20 mA °F, °C On , Off °C, °F, mA
Down/Scale
T/C K, J, E, N, T, R, S

The SEM710 is also provided with user push button trim, allowing trim adjustments at both 4 mA and 20 mA. The user trim function
can be locked during configuration if not required. The display will flag error message for sensor failure, also the display will flash
under or over range alternatively with actual Input reading, when the temperature exceeds the set range.

100.0
+

3 - +
PT100 Vs
+ Loop
2 Load
-
1
-

SEM710
Input 2 Wire (4-20) mA Output
USB

OneTemp Pty Ltd


Ph: 1300 768 887 Adelaide l Melbourne l Sydney l Brisbane www.onetemp.com.au
UNIVERSAL DIN RAIL TRANSMITTER

SPECIFICATION @ 20 °C

INPUTS

INPUT RANGE ACCURACY STABILITY O/C CJ Sensor IMPEDANCE


(Note 1) (Note 6) (Note 3) excitation
(Note 4)
K (-200 to 1370) °C 0.1 % of FSR ±0.5 °C ± 0.01 % of FSR / °C Yes Yes - 1 MΩ
J (-100 to 1200) °C (type T 0.2 % FSR. ± 0.5 °C) (Note 5)
E (-100 to 1000) °C
N (-180 to 1300) °C
T (-100 to 400) °C
R (-10 to 1760) °C ± 0.5 °C ±0.1 % of FSR
(Note 2)
S (-10 to 1760) °C ± 0.5 °C ±0.1 % of FSR
(Note 2)

P (-200 to 850) °C ± 0.1 °C / ±0.05 % of rdg ± 0.005 % of FSR / °C No <450 uA

Key Rdg = Reading ; FSR = Full Scale Range ; O/C = programmable open circuit sensor detect; CJ = Cold junction error

Notes 1. Accuracy for PT100 and T/C do not include sensor and cold junction errors.
2. Only over the range (800 to 1600) °C
3. Cold junction range (-20 to 70) °C, Accuracy ± 0.5 °C , Tracking ± 0.05 °C
4. PT100 input Maximum lead resistance 20 R, Lead effect 0.015 °C / Ω.
5. Impedance – not including 0.2 uA open circuit detect bias current effect.
6. Ambient (-10 to 50) °C

OUTPUT

Type Two wire current sink; signal range (4 to 20) mA; full range (3.8 to 24) mA
Supply (15 to 30) V dc , 24 V nominal giving Max loop load of 450 R @ 24 V
Response time < 500 ms to reach 95 % of final value ; Start up time < 3 s
Calibration Accuracy ± 5 uA
Loop Effects Loop ripple 0.03 % of FSR; Supply sensitivity 0.05 uA / °C ; supply ripple rejection < ± 5 uA error @ 1 V
rms 50 Hz ripple
Protection Reverse connection and over-voltage protection. Max over voltage current 100 mA.
User Trim Raise and lower buttons, active for offset when output is between (3.8 to 6) mA, Span between (18 to 22)
mA. Trim lock option.
Stability ± 5 uA / °C.

GENERAL

Isolation Input to output tested at 500 V dc.


Ambient operating (-40 to 85) °C (10 to 95) % RH non condensing. Storage (-40 to 85) °C
Approvals CE tested to BS EN 61326
Protection IP67 when used with SCH4 head
Display 4 Digit seven segment Red LED Height 7.3 mm
Display Resolution 0.1 °C , 0.1°F, or 0.00 mA
Display Errors Err (sensor error) ; Flash OUER (over range) Flash UNDR ( under range)

C REFER TO
F
I INSTRUCTION MANUAL BEFORE USE
DM640

ORDER CODE : SEM710


Terminals Screw terminal
Ø66.25
ASSOCIATED PRODUCTS
Cable 2.5 mm Max ORDER CODES
21.50
USB CABLE A/M TO MINI B/M 48-200-0001-01
Colour Grey USB Link Software F0C @ www.status.co.uk Downloads

SCH4 ABS Plastic Enclosure


Ø74.80 SCH15 Stainless Steel Enclosure

Status Instruments Ltd Tel: +44 (0)1684 296818


Green Lane Business Park Fax: +44 (0)1684 293746
Green Lane, Tewkesbury Email: [email protected]
Gloucestershire, UK Website: www.status.co.uk
GL20 8DE D2466_01_04 CN4808 SEM710 data sheet