SEM710
SEM710
SEM710
INTRODUCTION
The SEM710 is the first of a new generation head mounted temperature transmitter with display feature, from Status Instruments. It
has been designed to accept most common temperature sensor inputs and provide the user with a standard two wire (4 to 20) mA
output signal. Isolation is provided between input and output and all temperature ranges are linear to temperature. The addition of
a display provides the user with instant information of the loop condition at the point of measurement. It comes in either our SCH4
ABS plastic connector head or our new SCH15 Stainless steel connector head.
Designed for ease of use, our latest USB interface is fitted for quick and easy configuration. Just connect a standard USB cable
between the SEM710 and your PC. Using our free configuration software, your PC will automatically upload the existing configuration
data and guide you through any changes you wish to make. To further help save time, the SEM710 does not need to be wired to a
power supply during the configuration process, it is powered via the USB interface from your PC. The following parameters are
configurable :-
INPUT TYPE LOW RANGE HIGH RANGE UNITS BURNOUT USER TRIM DISPLAY
PT100
Up/Scale
Or Input @ 4 mA Input @ 20 mA °F, °C On , Off °C, °F, mA
Down/Scale
T/C K, J, E, N, T, R, S
The SEM710 is also provided with user push button trim, allowing trim adjustments at both 4 mA and 20 mA. The user trim function
can be locked during configuration if not required. The display will flag error message for sensor failure, also the display will flash
under or over range alternatively with actual Input reading, when the temperature exceeds the set range.
100.0
+
3 - +
PT100 Vs
+ Loop
2 Load
-
1
-
SEM710
Input 2 Wire (4-20) mA Output
USB
SPECIFICATION @ 20 °C
INPUTS
Key Rdg = Reading ; FSR = Full Scale Range ; O/C = programmable open circuit sensor detect; CJ = Cold junction error
Notes 1. Accuracy for PT100 and T/C do not include sensor and cold junction errors.
2. Only over the range (800 to 1600) °C
3. Cold junction range (-20 to 70) °C, Accuracy ± 0.5 °C , Tracking ± 0.05 °C
4. PT100 input Maximum lead resistance 20 R, Lead effect 0.015 °C / Ω.
5. Impedance – not including 0.2 uA open circuit detect bias current effect.
6. Ambient (-10 to 50) °C
OUTPUT
Type Two wire current sink; signal range (4 to 20) mA; full range (3.8 to 24) mA
Supply (15 to 30) V dc , 24 V nominal giving Max loop load of 450 R @ 24 V
Response time < 500 ms to reach 95 % of final value ; Start up time < 3 s
Calibration Accuracy ± 5 uA
Loop Effects Loop ripple 0.03 % of FSR; Supply sensitivity 0.05 uA / °C ; supply ripple rejection < ± 5 uA error @ 1 V
rms 50 Hz ripple
Protection Reverse connection and over-voltage protection. Max over voltage current 100 mA.
User Trim Raise and lower buttons, active for offset when output is between (3.8 to 6) mA, Span between (18 to 22)
mA. Trim lock option.
Stability ± 5 uA / °C.
GENERAL
C REFER TO
F
I INSTRUCTION MANUAL BEFORE USE
DM640