Two-Wire TRANSDUCER: Series
Two-Wire TRANSDUCER: Series
Two-Wire TRANSDUCER: Series
Two-wire TRANSDUCER
+ +
+ +
https://www.shinkawa.co.jp/sst/
Published in June.2021
WKN-142K - - / NB1
The WKN Series transmitter combines the signal conditioning circuit with the probe driver into one
package for easy installation. The two wire current loop drives the transducer and transmits signals at System Cable Length Output Range Mounting Plate Non-incendive
-
the same time, saving wiring cost. 1 5m 1 0 to 100 μm pk-pk 1 35 mm DIN Rail Mounting Plate
CSA:
2 7m 2 0 to 125 μm pk-pk 2 Screw Mounting Plate
ClassⅠDiv.2 Gr.A,B,C,D
Ideal for monitoring of general purpose rotating 5 9m 3 0 to 200 μm pk-pk 1
ATEX:
equipment and compressors. 4 0 to 250 μm pk-pk
Ex nAⅡT4 Gc
5 0 to 400 μm pk-pk
Compact, space-saving (50% smaller, 60% lighter
than the conventional) Transmitter (for thrust)
Current output allows for long wiring distances.
WKN-142T - - / NB1
Direct connection to PLCs and DCS control systems.
No external signal conditioner required.
System Cable Length Output Range Mounting Plate Non-incendive
Phoenix terminal type allows for quick and efficient wiring
1 5m 1 -0.6 to 0 to +0.6 mm 1 35 mm DIN Rail mounting plate CSA:
2 7m -0.635 to 0 to +0.635 mm 2 Screw mounting plate ClassⅠDiv.2 Gr.A,B,C,D
RoHS and CE compliance 2 1
5 9m (-25 to 0 to +25 mils) ATEX:
Ex nAⅡT4 Gc
System configuration Sensor
Jam nut
WKN transmitter Thread
Coaxial connector
Rated voltage power/current output Flexible armor
Sensor tip
PWR +12 TO
0-125 μ
vibration :
7m System
WKN-142K2-2
WL sensor
2-Wire
Remove label for ZERO SPAN adjust
vibration
4-20 mA
Transmitter
+35 VDC
Do not connect it to GAP OUT.
can cause false alarm.
Grounded AC powered equipment
GAP OUT
GAP OUT
+18.5* to 35 VDC
WW extension cable
7.87 mV/μ
GAP
(787 mV/100 m) 1 to 5 VDC
Receiver
System cable length : 5 m, 7 m, or 9 m 250 Ω
Oscilloscope etc.
* Assumes the receiver 250 Ω, other 30 Ω
WL-142K05 - / NB1
Specification
Armor Thread Size Unthreaded Length* (L1) Case Length* (L2) Cable Length** (L3) Non-incendive
Transmitter Outline Drawing (by mount type)
For WK series users:
Holes of screw mounting plates are the same as WK series.
Unit: mm You can replace your WK series with WKN series without making new holes. With M1 M8 X 1 10 mm step, 0 to 230 mm 10 mm step, 20 to 250 mm 1 0.5 m CSA:
35 mm DIN rail mounting plate type A (Without Fluoro M2 M10 X 1 e.g.) 06=60 mm, L1≤L2-20 mm e.g.) 25=250 mm 2 1.0 m ClassⅠDiv.2
Specifications WKN-142K WKN-142T resin coating) Gr.A,B,C,D
U1 1/4-28 UNF-2A 0.1 in step, 0 to 9.2 in 0.1in step, 0.8 to 9.9 in 3 5.0 m 1
Name plate GAP output connector With U2 3/8-24 UNF-2A e.g.) 04=0.4 in, L1≤L2-0.7 in e.g.) 35=3.5 in 4 7.0 m ATEX:
4 to 20 mA Output Range 0 to 100 μm pk-pk -0.6 to 0 to +0.6 mm or
Sensor input connector 0 to 125 μm pk-pk -0.635 to 0 to +0.635 mm Ex nAⅡT4 Gc
5 9.0 m
Current Output
0 to 200 μm pk-pk (-25 to 0 to +25 mils) T (With Fluoro * Specify length in mm for M thread, in inch for UNF-2A thread
0 to 250 μm pk-pk resin coating) ** For sensors of cable length 5.0, 7.0, 9.0 meters, extension cable cannot be
0 to 400 μm pk-pk used. Connect the sensor cable directly to the transmitter.
L Without
Logo, model code and 4 to 20 mA Conversion ±1.5 % of full scale range
serial number Accuracy (Input to test pin and current output)
Extension Cable
Terminal block Maxi. Load Resistance 43.5 × (Vps-12)Ω (Vps = power supply voltage)
Coaxial connector Coaxial connector
Calibration Target JIS SCM440 flat surface Flexible armor
DIN rail mounting plate Scale Factor Error* 5 m, 7 m System: 7.87 mV/μm ±6.5 % typ.
9 m System 7.87 mV/μm ±10 % typ.
(Values include interchangeability errors) to sensor to transmitter