Glassware SOP
Glassware SOP
Glassware SOP
4. PRINCIPLE OF CALIBRATION:
The basic principal of gravimetric method is to weigh the amount of distilled water delivered in a
single operation of the volumetric instrument, thus giving volume of the water. A number of repeat
5. PREPARATION:
6. CALIBRATION METHOD :
6.1 DETAILED CALIBRATION METHOD FOR PISTON PIPETTE (As per ISO 8655-6):
1. Ensure water in the balance water container at least half full. If the water level is too low, add
distilled water. Do not overfill the container. Ensure water in the pipetting container to a depth at
least 3 mm.
2. Press set zero key and ensure set zero of balance display.
2. Fit the suitable tip to the pipette.
3. Fill the tip with distilled water and expel to waste five times to reach humidity equilibrium in the
dead air volume of the pipette.
4. Calibrate the pipette at least three volumes i.e. nominal volume, 50% of the nominal volume,
10% of the nominal volume & For fixed volume pipette at nominal volume.
6. Set a range on pipette with rotating the cap clockwise or anticlockwise direction. Hold the
pipette in hand vertically and follows the process following ways shown in fig.1,
7. Press the operating button to the first stop. immersing the tip 2 to 3 mm below the surface of
distilled water
8. Release the operating button slowly to the ready position. This action will fill the tip with the
distilled water. Remove the tip vertically and carefully from the surface of distilled water.
9. Open the top sliding cover of analytical balance. Take the tip near to weighing vessel vertically.
Press the operating button to the second stop, and completely empty the tip in weighing vessel.
10. Release the operating button to the ready position. Remove the tip vertically and carefully from
the top sliding cover of analytical balance and close the top sliding cover.
11. Repeat the above same procedure for 10 measurements at each pre-decided volume and
record these readings in calibration data sheet.
12. Affix the calibration sticker indicating its identification and calibration date at appropriate place
on the instrument and prepare the calibration Certificate having contents as per 5.10 of Quality
Manual.
6.2 DETAILED CALIBRATION METHOD FOR GLASSWARE (As per ISO 4787:2010):
Where,
m2 is the balance reading of vessel with water, in grams;
Where,
m2 is the balance reading of vessel with water, in grams;
m1 is the balance reading of empty vessel, in grams (zero in case the balance was tared with the
volumetric instrument or receiving vessel);
ρw is the density of water at tw °C, in gm /ml
ρa is the density of air, in grams per millilitre, obtained from equation below at the temperature and
atmospheric pressure of the test;
Pa is the pressure, expressed in hectopascals (hPa);
Ø is the relative humidity, expressed as a percentage;
ta is the air temperature, expressed in degrees Celsius;
7. General calculation:
Where,
(a0)+((a1*10^(-2))*L26)+((a2*10^(-3))*(L26^2))+((a3*10^(-5))*(L26^3))+((a4*10^(-7))*(L26^4))
Where,
with the constants (ITS-90 temperature scale):
a0 is equal to 999.85308 kg/m3;
a1 is equal to 6.32693x10–2 °C–1 kg/m3;
a2 is equal to 8.523829x10–3 °C–2 kg/m3;
a3 is equal to 6.943248x10–5 °C–3 kg/m3;
a4 is equal to 3.821216x10–7 °C–4 kg/m3.
w (999.85308) + ((6.32693*10^ (-2))* t w + ((-8.52382910^ (-3))* t w ^2)) + ((6.943248*10^ (-5)) * t w ^3)) + ((-3.821216 *10^ (-7))* t w ^4))
Where
t w = water temperature in degree Celsius;
ρb is either the actual density of the balance weights when these are adjusted to their
nominal mass, or the reference density for which the weights have been adjusted (see
the note below), in grams per millilitre, or, when using an electronic balance without
weights, the (reference) density of the weights with which it has been adjusted; (density
of the weights is taken as 8 g/ml),
ϒ is the coefficient of cubical thermal expansion of the material of which the volumetric
instrument tested is made, in reciprocal degrees Celsius,
6. Uncertainty of Measurements
TYPE A
Where,
σ is Standard deviation,
n is number of Observation &
UA Standard deviation of mean or the standard uncertainty.
Degrees of freedom
νi = n – 1
TYPE B
(i) U1 = Uncertainty due to Uncertainty of standard (Weighing Balance) – from calibration certificate.
(ii) U2 = Uncertainty due to Repeatability of weighing balance used for calibration,
(iii) U3 = Uncertainty due to water temperature (considering accuracy and uncertainty of glass
thermometer used for water temperature measurement),
(iv) U4 = Uncertainty due to air temperature (considering accuracy and uncertainty of thermo-
barometer used for air temperature measurement),
(v) U5 = Uncertainty due to air pressure (considering accuracy and uncertainty of thermo-barometer
thermometer used for air pressure measurement),
Sensitivity coefficients are calculated using following formula (as per ISO/TR 20461):
i) Sensitivity Coefficient for Water Temperature :
4
m
Ctw iatwi 1
w2 i 1
m (k 2ta 0 k 3) k1 pa
Cta
w 2 (ta ta 0) 2
iii) Sensitivity Coefficient for Air Pressure :
m k1
Cpa
w (ta ta 0)
2
(i) U1 = Uncertainty due to Uncertainty of standard (Weighing Balance) – from calibration certificate.
(ii) U2 = Uncertainty due readability of weighing balance
Table
Coverage factor k for various effective Degree of Freedom with coverage probability of 95 %
νeff 1 2 3 4 5 6 7 8 9 10 11 12
k 12.71 4.30 3.18 2.78 2.57 2.45 2.36 2.31 2.26 2.23 2.20 2.18
νeff 13 14 15 17 20 25 30 ∞
k 2.16 2.14 2.13 2.11 2.09 2.06 2.04 1.96
7. REF IS STANDARD:
8. RECORD REF:
Calibration Data of Volume – IF-40