Identify Patient-Ventilator Asynchrony
Identify Patient-Ventilator Asynchrony
Identify Patient-Ventilator Asynchrony
Introduction
Trigger Asynchrony (Phase 1)
Trigger Asynchrony Can Occur in Any Ventilation Mode
Flow Asynchrony (Phase 2)
Volume Ventilation With a Fixed Flow Pattern
Pressure Ventilation With Variable Flow
Termination Asynchrony (Phase 3)
Delayed Termination
Premature Termination
Expiratory Asynchrony (Phase 4)
Summary
Patient-ventilator interaction can be described as the relationship between 2 respiratory pumps: (1) the
patient’s pulmonary system, which is controlled by the neuromuscular system and influenced by the
mechanical characteristics of the lungs and thorax, and (2) the ventilator, which is controlled by the
ventilator settings and the function of the flow valve. When the 2 pumps function in synchrony, every
phase of the breath is perfectly matched. Anything that upsets the harmony between the 2 pumps results
in asynchrony and causes patient discomfort and unnecessarily increases work of breathing. This article
discusses asynchrony relative to the 4 phases of a breath and illustrates how asynchrony can be iden-
tified with the 3 standard ventilator waveforms: pressure, flow, and volume. The 4 phases of a breath
are: (1) The trigger mechanism (ie, initiation of the inspiration), which is influenced by the trigger-
sensitivity setting, patient effort, and valve responsiveness. (2) The inspiratory-flow phase. During both
volume-controlled and pressure-controlled ventilation the patient’s flow demand should be carefully
evaluated, using the pressure and flow waveforms. (3) Breath termination (ie, the end of the inspiration).
Ideally, the ventilator terminates inspiratory flow in synchrony with the patient’s neural timing, but
frequently the ventilator terminates inspiration either early or late, relative to the patient’s neural
timing. During volume-controlled ventilation we can adjust variables that affect inspiratory time (eg,
peak flow, tidal volume). During pressure-controlled or pressure-support ventilation we can adjust
variables that affect when the inspiration terminates (eg, inspiratory time, expiratory sensitivity). (4)
Expiratory phase. Patients with obstructive lung disease are particularly prone to developing intrinsic
positive end-expiratory pressure (auto-PEEP) and therefore have difficulty triggering the ventilator.
Bedside evaluation for the presence of auto-PEEP should be routinely performed and corrective ad-
justments made when appropriate. Key words: ventilator graphics, waveforms, asynchrony, patient-venti-
lator interface. [Respir Care 2005;50(2):202–232. © 2005 Daedalus Enterprises]
Introduction cade. Both direct and anecdotal evidence (see below), how-
ever, suggests that bedside use of ventilators’ graphics
Ventilator graphics are available on almost all current capabilities is widely underutilized, and standard ap-
mechanical ventilators and have been available for evalu- proaches or guidelines for graphics interpretation are often
ating the patient-ventilator interface for more than a de- lacking.
Ventilator Factors
Trigger variables: esophageal pressure, flow, or shape signal
Sensitivity setting
Rise-time capability
Design, mode, and settings of the flow delivery system
Flow pattern selected
Design of the exhalation valve
How positive end-expiratory pressure is generated by the software
Extraneous flow (eg, from a nebulizer or added oxygen)
Patient Factors
Sedation level: pain, splinting
Inspiratory effort/respiratory drive; neural timing
Pathology of the respiratory system or abdomen; secretions
Intrinsic positive end-expiratory pressure
Fig. 1. The primary causes of ventilator-related deaths, 1995–2003. Size and type of airway
(Adapted from Reference 1, with permission.) Presence of leaks
Fig. 2. Airway pressure (top), flow (middle), and volume (bottom) waveforms from a normal subject during synchronized intermittent
mandatory ventilation with pressure support. The 4 phases of the breath are numbered. Phase 1 is the initiation of patient effort, which
indicates achievement of the trigger threshold (2 cm H2O) that opens the inspiratory valve. Phase 2 represents the relationship between flow
delivery, as determined by the ventilator’s flow algorithm, and the patient effort (the first and third breaths are pressure support breaths,
in which flow is partially dependent on patient effort, and the second breath is the mandatory breath, which has a constant-flow pattern).
Notice the scooped-out appearance of the pressure waveform during the mandatory breath, which indicates that the inspiratory flow was
inadequate. Phase 3 is the breath-termination point, which varies based on the type of breath; for the middle breath the inspiratory time
is set on the ventilator, but the inspiratory time for the pressure support breath is based on the termination criterion, which in this case is
5% of the peak flow. Phase 4 is the expiration portion of the breath. During this phase the breath should be inspected for evidence of
intrinsic positive end-expiratory pressure (auto-PEEP). This expiratory flow waveform returns to zero prior to the next breath, which
indicates the absence of auto-PEEP.
The use of guidelines, standards, and protocols for as- “Although triggering composes only a small part of the entire
sessing and treating disease states improves patient out- inspiratory cycle, inappropriate setting or design may increase
comes.3 Similarly, a standard approach to analysis of ven- the patient’s effort and inspiratory muscle work.”4,5
tilator waveforms should improve patient comfort, reduce “In a demand-flow system (pressure-trigger), the trig-
work of breathing (WOB), and perhaps improve outcomes. ger variable is a set pressure that must be attained at the
Evaluation of patient-ventilator synchrony can be broken onset of inspiration for the ventilator to deliver fresh
down into 4 phases (Fig. 2): evaluation of triggering; eval- gas into the inspiratory circuit.”4 Most microprocessor-
uation of adequate flow delivery; evaluation of breath ter- based ventilators use pressure-triggering to initiate both
mination; and evaluation of intrinsic positive end-expira- the mandatory breaths (assist-control and synchronized
tory pressure (auto-PEEP), which is the primary clinical intermittent mandatory ventilation) and spontaneous
complication associated with the expiratory phase. For or- breaths (continuous positive airway pressure, synchro-
ganization purposes and as a methodical approach for cli- nized intermittent mandatory ventilation, pressure sup-
nicians to use at the bedside, the present article is orga- port ventilation).4
nized according to these 4 phases.
Definition: Trigger Asynchrony. This term has been
Trigger Asynchrony (Phase 1) defined as “muscular effort without ventilator trigger.”6
Though this definition describes the problem when patient
Definition: Trigger. “The trigger variable is defined as effort fails to trigger the ventilator, we will also discuss
the variable that is manipulated to deliver inspiratory flow.”4 several additional triggering problems: double-triggering,
Fig. 3. Schematic of a typical breath. P-T ⫽ trigger pressure. D-T ⫽ inspiratory trigger time. D-B ⫽ time to return airway pressure to zero.
D-I ⫽ inspiratory delay time. Area 1 ⫽ trigger pressure-time product. D-E ⫽ expiratory time delay. P-E ⫽ supra-plateau expiratory pressure
change. Area E ⫽ expiratory pressure area. (From Reference 7, with permission.)
auto-triggering, and insensitive trigger (triggering that re- the breath until after the inspiratory delay time has passed.
quires excessive patient effort). If we account for the pressure-drop across the endotra-
Trigger asynchrony is only one type of problem asso- cheal tube and the potential presence of auto-PEEP, the
ciated with a patient fighting the ventilator. Though bed- patient may not receive any positive inflation support until
side clinicians are inclined to think of trigger problems as some time after the inspiratory delay time has passed.
being associated primarily with the sensitivity setting on Currently, there are only 2 common types of trigger mech-
the ventilator, the definition has been expanded to address anism available on commercial ventilators in the United States:
other variables that are influenced by the patient’s inspira- pressure trigger and flow trigger. Though initial clinical stud-
tory effort or respiratory drive, and the rate at which the ies indicated that flow-triggering offered some advantage in
ventilator supplies gas to the circuit (Fig. 3).7 These vari- reducing trigger asynchrony, recent advances in the develop-
ables include (1) the traditional “trigger pressure” or “valve ment of pressure transducers have resulted in nearly equiva-
sensitivity,” which can be adjusted by the clinician; (2) the lent or comparable results.8 –10 A third type of trigger mech-
pressure maximum, which is the most negative pressure or anism—the shape-signal or “shadow” trigger—is available in
largest downward deflection in the airway pressure wave- the European market and is now available in the United States
form—this value may be more negative than the trigger as “AutoTrak” on the Respironics BiPAP ventilator.11 This
pressure if the patient has a strong respiratory drive; (3) shape-signal mechanism uses a mathematical model derived
the inspiratory trigger time, which is the time elapsed be- from the pressure and flow signals;12,13 it has some promise
tween the initial patient effort and the point at which the for reducing trigger asynchrony, and it may be better toler-
airway pressure reaches the maximum baseline pressure— ated.14 –19 Initial clinical studies have looked at various ap-
for patients with low respiratory drive, it takes longer for plications of this derived signal, such as inclusion in propor-
the airway pressure to reach the trigger pressure; (4) time tional-assist algorithms, but this triggering mechanism still
to return trigger pressure to zero or baseline—this time is has some problems with repetitive auto-cycling (runaway).15,20
affected by how rapidly the ventilator is able to supply
flow to pressurize the circuit, and is influenced by the
Trigger Asynchrony Can Occur in Any Ventilation
slope setting; and (5) the inspiratory delay time, which is
Mode
the total time delay from the initial patient effort until the
pressure waveform returns to baseline—this is the sum of
the inspiratory trigger time and the time to return trigger to Clinical studies indicate that ventilator-dependent pa-
baseline. The patient does not receive any assistance with tients experience trigger asynchrony in all of the common
Fig. 9. Flow, airway pressure (Paw), and esophageal pressure (Pes) in a patient with severe chronic obstructive pulmonary disease and
ventilated with pressure support. The dotted lines indicate the beginning of inspiratory efforts that triggered the ventilator. The thin, black
arrows indicate nontriggering inspiratory efforts. Notice the time delay between the beginning of inspiratory effort and ventilator triggering.
Ineffective (nontriggering) efforts occurred during both mechanical inspiration and expiration. Those ineffective efforts can easily be
identified on the flow waveform; ineffective efforts during mechanical inspiration abruptly increase inspiratory flow, whereas during expi-
ration they result in an abrupt decrease in expiratory flow (open arrows in the flow waveform). The set respiratory frequency is 12
breaths/min, but the patient is making 33 inspiratory efforts per minute. (From Reference 2, with permission.)
Fig. 10. The upper panel shows waveforms of esophageal pressure (Pes in cm H2O), pressure at the airway opening (Pao in cm H2O), and
flow (in L/min) from a tracheostomized patient with trigger asynchrony during flow-controlled, volume-cycled (assist/control) ventilation. The
patient’s inspiratory efforts are identified by the negative Pes swings. The positive end-expiratory pressure (PEEP) is set at zero. Pao
appropriately drops to zero during expiration, indicating little circuit or valve resistance. Trigger asynchrony is evident; there is one triggered
breath (white arrows) for every 3– 4 inspiratory efforts (black arrows point to nontriggering efforts). Prolonged expiratory flow is due to airflow
limitation. Pes swings have little effect on retarding the expiratory flow and even less effect on Pao, depending on the phase of expiration.
In the lower panel, PEEP was increased to 10 cm H2O, so Pao during expiration is 10 cm H2O. There is persistent flow at end-expiration,
which indicates auto-PEEP. Trigger asynchrony has improved; there is one triggered breath for every 2–3 inspiratory efforts. There is less
limitation of expiratory flow, and the Pes swings are more effective in retarding the persistent expiratory flow. Peak inspiratory pressure and
Pes have slightly increased (compared to the waveforms in the upper panel), which probably indicates a higher end-expiratory lung volume
and total PEEP level. (From Reference 6, with permission.)
Fig. 13. Double-triggering seen in flow and volume waveforms from volume-controlled ventilation. Continued subject effort during the
second breath causes the airway pressure to drop below the trigger threshold, which initiates an additional “stacked” breath. Note the large
increase in peak airway pressure caused by the stacked breath and the high peak expiratory flow following the stacked breath.
Fig. 14. Inadequate flow and volume result in double-triggering during volume-controlled ventilation. During the first breath, slow valve-opening
(#1, as seen early in the flow waveform) and inadequate peak flow cause the dished-out appearance of the pressure waveform (#2). In the second
breath, again the valve-opening is too slow (#2) and inadequate peak flow results in additional gas being pulled through the demand valve (#3),
as seen by the appearance of the flow bump near the end of the flow waveform and the additional tidal volume beyond the set volume (in the
volume waveform). Continued inspiratory effort by the subject results in a second trigger and a stacked breath (#4). The subject was breathing
through a mouthpiece and filter, and was disconnected following the stacked breath, as seen in the volume waveform.
Fig. 15. Auto-triggering created by small circuit leak (tubing connection). The waveform sequence is from a Bear 1000 ventilator, set on
intermittent mandatory ventilation plus 5 cm H2O of pressure support, connected to a Michigan Instruments test lung. The small leak caused
the circuit pressure to drop below the set positive end-expiratory pressure and trigger the ventilator.
Fig. 16. Representative flow, airway pressure (Paw), esophageal pressure (Pes), and arterial blood pressure (BP) waveforms from a patient
who underwent mitral-valve replacement and tricuspid annuloplasty for mitral stenosis, tricuspid regurgitation, and aortic regurgitation. With
triggering sensitivity set at 1 L/min (left), pressure support ventilation was activated between 2 synchronized intermittent mandatory
ventilation breaths. When trigger sensitivity was changed to 4 L/min (right), pressure support breaths disappeared and there was marked
oscillation in flow, Paw, and Pes. Cardiogenic oscillation was evaluated as the peak inspiratory-flow fluctuation (A), amplitude in the flow
oscillation (B), amplitude in airway pressure (C), and amplitude in esophageal pressure (D). Also note that the baseline of esophageal
pressure was elevated when auto-triggering occurred, suggesting hyperinflation of the lungs. (From Reference 26, with permission.)
fort the patient contributed to the delivery of the breath. form, we can determine that the ventilator has sensed the
Figure 19 illustrates the pressure pattern associated with a beginning of patient effort and correctly initiated the breath.
constant-flow mandatory breath during passive breathing. In this instance, however, as patient effort increases, the
Figures 20 and 21 illustrate breathing sequences during peak flow set on the ventilator no longer meets the pa-
which the first breath is passive and during the second tient’s flow demand and the airway pressure waveform
breath the patient took a more active role; note the dished becomes progressively dished out.
out appearance of the second breath. Figure 22 represents The reason for evaluating flow asynchrony is to adjust
a similar sequence, with increasing amounts of patient the ventilator to match the patient’s flow demand. This can
effort. generally be accomplished by adjusting the peak flow set-
Flow asynchrony is sometimes mistaken for trigger asyn- ting on the ventilator until the pressure waveform pattern
chrony, and in particular trigger problems associated with most closely represents the condition observed during pas-
incorrect sensitivity settings. The difference can be deter- sive breathing. However, the patient’s flow demand may
mined by carefully evaluating the vertical alignment of the vary considerably, based on the neural drive to breathe.
pressure and flow waveforms. Figure 23 illustrates a se- The flow variability is analogous to the difference in flow
quence in which patient effort progressively increased. By demand between the resting breathing pattern and that
adding vertical lines to the leading edge of the flow wave- associated with increasing athletic activity: the higher the
Fig. 17. Evaluating trigger sensitivity with the airway pressure and flow waveforms. The black arrows indicate initiation of mandatory
constant-flow breaths. The sensitivity setting is – 4 cm H2O. During the first inspiratory effort the trigger threshold is not reached and the
breath is time-triggered, but the patient effort is sufficient to pull gas through the demand valve (seen as the flow that precedes the
constant-flow pattern). The next patient effort occurs before the synchronized intermittent mandatory ventilation window is open and the
patient again pulls gas through the demand valve. The second mandatory breath is time-triggered (constant flow occurs during patient
exhalation). Preceding the third mandatory breath, patient effort reaches the trigger threshold, but again a small amount of gas is pulled
through the demand valve (the leading edge of the flow curve is slightly rounded).
Fig. 19. These waveforms represent relaxed, passive breathing by a subject connected to a Bear 100 ventilator during intermittent
mandatory ventilation plus 2 cm H2O continuous positive airway pressure (PEEP). The circled breath is a mandatory volume breath with a
constant-flow pattern.
Fig. 20. Two-breath sequence during volume-controlled ventilation. The first breath is passive. During the second breath the subject exerted
additional inspiratory effort, which scooped out the airway pressure waveform.
Fig. 21. Two-breath sequence during volume-controlled ventilation, created in a laboratory setting, using a Servo 300A ventilator and a
Michigan Instruments test lung. Effort during the second breath was created by manually lifting the test lung. The shaded area represents
the pressure-time product associated with the additional simulated inspiratory effort.
Fig. 22. Three-breath sequence obtained with a Bear 1000 ventilator connected to Michigan Instruments test lung during volume-controlled
ventilation. Progressive increases in patient effort during breaths 2 and 3 were created by manually lifting the test lung. The dished-out
appearance of the airway pressure waveform illustrates the changes from the passive breath when flow does not meet patient demand.
Fig. 23. Evaluation of trigger threshold and patient effort. Progressive increases in patient effort are evidenced by the increasingly scooped
appearance of the airway pressure waveform (downward arrows). Trigger synchrony is evaluated by interpreting the beginning of the flow
waveform, and by the time relationship to the initiation of patient effort in the airway pressure waveform (upward arrows at the beginning
of the constant-flow pattern). Even though inspiratory effort progressively increased, the valve correctly opens after a small drop in airway
pressure. Inadequate peak flow, however, causes the dished-out appearance of the airway pressure waveform, as the patient exerts more
effort.
Fig. 24. Comparison of peak flow during a constant-flow mandatory breath (circled) and during pressure support breaths during synchro-
nized intermittent mandatory ventilation. The peak flow of the mandatory breath is substantially less than the peak flow associated with the
patient-oriented pressure support breath. The inadequacy of the ventilator peak flow is evidenced by the scooped-out appearance of the
airway pressure waveform (arrow) and the additional flow that the patient pulls through the demand valve in excess of the constant-flow
setting (bump in the middle of the constant-flow waveform).
Fig. 25. These waveforms illustrate inadequacy of ventilator flow while using the descending-ramp flow pattern. Upper panel: As flow decreases
during the latter portion of the breath, the patient demand for flow results in scooping of the airway pressure waveform (arrow). Lower panel: The
ventilator peak-flow setting was increased from 60 L/min to 120 L/min, which eliminated the scooped appearance of the airway pressure
waveform. P ⫽ airway pressure. (Courtesy of Kenneth D Hargett RRT, The Methodist Hospital, Texas Medical Center, Houston, Texas)
Fig. 27. Airway pressure (Paw), flow, and esophageal pressure (Pes) waveforms from a patient with chronic obstructive pulmonary disease,
ventilated with volume assist-control ventilation, with 2 inspiratory flow rates: 30 L/min and 90 L/min. With both flow rates, tidal volume was
kept constant (0.55 L). Ineffective (nontriggering) inspiratory efforts are indicated by arrows. Increasing the expiratory time (by increasing
inspiratory flow at constant tidal volume) decreased dynamic hyperinflation, which reduced the number of ineffective inspiratory efforts,
which increased the respiratory rate. (From Reference 2, with permission.)
the prolonged inspiratory time reduced the time alotted for the peak flow setting was increased (lower panel of Fig.
expiration, and resulted in air-trapping. Auto-PEEP can be 27), inspiratory time decreased, expiratory time increased,
positively identified in a flow waveform where expiratory and auto-PEEP decreased, which resulted in fewer missed
flow does not return to zero prior to the onset of the next trigger attempts. With modern ventilators the incidence of
breath. In addition, the peak airway pressure increases— ineffective triggering does not seem to differ between flow-
another indication of air-trapping. triggering and pressure-triggering systems,33 but shape-
In addition to flow patterns, VT and peak flow settings signal triggering shows promise for improving synchrony
also influence inspiratory and expiratory timing. This is for many patients with COPD.34
particularly important with patients who have severe air- Patients with asthma (Fig. 28) can suffer deleterious
flow limitation. In the literature there are examples of effects from large VT and flow-time relationships that re-
patients who were on fixed flow patterns and developed sult in hyperinflation and missed trigger attempts.32 Dur-
auto-PEEP.2,32 The upper panel of Figure 27 illustrates a ing one of the breaths (vertical lines in Fig. 28) much of
patient who had COPD and a low peak-flow setting that the neural inspiratory time is spent in pulling through the
resulted in air-trapping and missed trigger attempts. After auto-PEEP, such that the patient effort is nearly over be-
Fig. 31. Over-distention “beak” on a pressure-volume loop. In this example, volume in excess of 350 mL abruptly increases pressure,
indicating over-distention. VT ⫽ tidal volume. Pcirc ⫽ circuit pressure. (Courtesy of Kenneth D Hargett RRT, The Methodist Hospital, Texas
Medical Center, Houston, Texas)
Pressure Ventilation With Variable Flow sure-controlled ventilation, however, the peak flow is no
longer fixed, but is variable and depends on various fac-
Evaluation of flow asynchrony during pressure- tors, including; (1) set target pressure, (2) patient effort,
controlled ventilation also requires viewing the pressure- and (3) respiratory-system compliance and resistance. Of
time waveform to determine the adequacy of flow deliv- those, only the pressure can be set by the clinician. The
ery. As with volume-controlled ventilation, the flow pressure should be set based on the lowest pressure setting
delivery is assessed by evaluating the presence or lack of that achieves the desired VT. The main flow parameter that
concavity in the pressure-time waveform. During pres- can be adjusted during pressure-controlled ventilation is
Fig. 32. These waveforms illustrate pressure support breaths using different rise time (pressure slope, or rate of valve-opening) settings,
ranging from very slow (1) to very fast (5) valve opening, performed with a nearly constant or stable inspiratory effort (flow demand). The
airway pressure waveform appears scooped-out initially and then progresses to a more constant-pressure pattern as the valve setting is
changed to open faster. Breath 4 has the best valve-opening setting for these conditions. In breath 5 the valve opens a little too fast and
causes a slight pressure-overshoot early in the airway pressure waveform.
Fig. 33. Effect of minimum (mn), medium (md), and maximum (mx) rise times on trigger pressure-time product (T-PTP) during pressure
support ventilation (white bars) and pressure assist/control modes (black bars) with 5 ventilator brands. T-PTP decreased as rise time
increased, with all the ventilators, in both pressure support and pressure assist/control mode. (From Reference 7.)
Fig. 35. Representative flow, volume, airway pressure (Paw), esophageal pressure (Pes), and end-tidal carbon dioxide pressure (PETCO2)
waveforms from a patient with spontaneous ventilation (SV), and one undergoing pressure support ventilation (pressure support of 15 cm
H2O). The rise time was modulated so that the plateau pressure was reached after a time ranging from 0.1 second (T 0.1) to 1.5 seconds
(T 1.5). Each reduction of the pressure ramp slope was associated with an apparently dose-dependent progressive increase in Pes swings
(arrows). Tidal volume appeared to remain constant. (From Reference 43, with permission.)
Fig. 36. Waveform recorded by the freestanding respiratory mon- Fig. 37. These waveforms are from the same subject as the wave-
itor on postoperative day 9. The rise-time setting is 10%. Note the forms in Figure 36, in which the rise time was 10% (slow valve-
scooped-out appearance of the airway pressure (Paw) waveform opening). In the above figure, the rise time is 1% (rapid valve-
(arrows), which indicates inadequate flow. When the initial flow opening), which causes a much steeper slope in the beginning of
output is lower than patient demand, the rise in airway pressure the inspiratory flow waveform. The pressure support level is
may be delayed or prolonged, or in some case may prohibit at- achieved earlier in the inspiratory phase (arrow), and there is a
taining the set pressure support level. V̇ ⫽ flow. VT ⫽ tidal volume. pressure plateau. V̇ ⫽ flow. VT ⫽ tidal volume. Pes ⫽ esophageal
Pes ⫽ esophageal pressure. (Adapted from Reference 39.) pressure. (Adapted from Reference 39.)
the initial ventilator rise-time setting did not provide ade- valve-opening setting may be preferable.40 – 42 The pres-
quate flow for the patient’s flow demand (slow valve- sure-overshoot occurs when there is a mismatch be-
opening). The rise-time setting was increased to meet the tween the patient’s flow demand and the rate of flow
patient’s demand (Fig. 37), which improved the pressure- delivery determined by the rise-time setting (see Fig.
time waveform (pressure rises more rapidly to the target 38). Rapid rise time can also cause flow asynchrony,
pressure) and decreased the esophageal pressure swing. patient discomfort, and premature termination of inspi-
In contrast, there are both patient and ventilator con- ration. Premature termination can be caused by either of
ditions in which rapid valve-opening causes pressure- 2 mechanisms: (1) early in flow delivery, the initial
overshoot early in the breath, in which case a slower pressure-overshoot may exceed the target pressure and
Fig. 38. Rapid valve-opening (fast rise time) results in pressure-overshoot in the airway pressure waveform (arrows). In addition, the initial
flow in excess of patient demand creates a high peak flow and a steeper flow decrease, which can cause premature termination of
inspiration. (Courtesy of Kenneth D Hargett RRT, The Methodist Hospital, Texas Medical Center, Houston, Texas)
terminate inspiration as part of the secondary breath- of rise time, even during volume-controlled ventilation
termination criteria (pressure termination), or (2) late in (Fig. 40). In the example in Figure 40, the rise-time
the flow delivery, the flow-termination point may occur setting (inspiratory rise-time percent) was set at its low-
sooner (percentage of peak flow is now a higher flow- est point (ie, most rapid valve-opening setting, which
termination value because of the higher initial peak flow) can be accomplished only by depressing the knob and
(Fig. 39). The Servo 300A ventilator allows adjustment turning it to the zero position, which the manufacturer
Fig. 39. Airway pressure (above) and flow (bottom) waveforms during 2 different rise time settings. The faster rise time (second breath)
results in a higher peak flow and pressure-overshoot (arrow). In addition, the flow-termination criterion (percent of peak flow) is reached
sooner because of the higher peak flow, and the breath terminates sooner. I ⫽ inspiratory. (Courtesy of Kenneth D Hargett RRT, The
Methodist Hospital, Texas Medical Center, Houston, Texas)
Fig. 40. With these waveforms the rise-time (on the Servo 300A ventilator) during volume-controlled ventilation was set to its fastest position
(shortest rise time), which causes pressure oscillations early in the pressure waveform and also on the pressure-volume loop. On the Servo
300 ventilator the rise time can be set in any mode. Accessing the fastest rise-time setting requires depressing the setting knob and turning
it to the zero position, which the manufacturer does not recommend under normal conditions.
does not recommend under normal circumstances), which ciated with the inspiratory cycle. The remaining types
causes pressure oscillations early in the pressure wave- of asynchrony are associated with expiratory events.
form. These include premature or delayed breath termination,
Though adjustment of the rise time is considered fine- and problems associated with overlap of expiratory mus-
tuning of the flow parameter, it is fairly easy to accomplish cle activity into the inspiratory cycle and simultaneous
by observing the flow and pressure waveforms simulta- contraction. While the latter are difficult to evaluate
neously and adjusting the rise time to maximize the ap- without neural recordings, evaluation of breath-termi-
pearance of the waveform (almost square appearance of nation is possible with the use of the standard pressure
the pressure waveform, no concavity, and no overshoot; and flow waveforms, and it has also been documented
see asterisk in Fig. 32). In comparison to volume-con- with neural recordings.47,48 Figure 41 illustrates the dif-
trolled ventilation, the ability to more nearly match the ference between premature termination (ventilator cy-
patient flow waveform during pressure-controlled ventila- cles off before the end of expiratory-muscle neural ac-
tion, and the ventilator’s ability to increase or decrease tivity) and delayed termination (ventilator does not cycle
flow in response to patient demand are desirable traits that off until after expiratory-muscle activity is initiated).
ultimately improve flow synchrony.
In general, the pressure-controlled modes (pressure sup-
Delayed Termination
port and pressure control) are better capable of meeting
inspiratory demand than is volume-controlled ventila-
tion.44,45 In addition, dual-control modes, which combine Parthasarathy et al studied delayed termination in a group
the attributes of pressure limiting and volume control, might of healthy subjects in whom airflow limitation was simu-
improve patient-ventilator synchrony.11,46 lated with a Starling resister.47 As seen in Figure 42, an
increase in pressure support resulted in a greater number
Termination Asynchrony (Phase 3) of missed trigger attempts. In evaluating the data, the non-
triggered attempts were preceded by breaths with larger
The first 2 types of asynchrony (missed trigger at- peak flow, higher VT, and prolonged inspiratory time. Non-
tempts or trigger delay and flow asynchrony) are asso- triggering attempts involved weaker inspiratory efforts
Fig. 44. These waveforms illustrate a subject being ventilated with a pressure support of 5 cm H2O. The subject’s neural timing precedes
the end of mechanical inflation and results in a pressure spike (large arrow) on the airway pressure waveform. Also note the rapid decline
in the inspiratory flow waveform near the end of the breath (small arrows) as a result of the subject’s expiratory effort.
gram recordings) to evaluate expiratory-muscle activity. the target pressure and the end of mechanical inflation. In
Parthasarathy et al47 used electromyography to confirm pressure-controlled modes, however, the inspiratory time
similar observations in normal subjects. is set by the clinician. In this case, subsequent reduction in
Delayed termination generally results in dynamic hyper- inspiratory time removed the pressure spike and the zero-
inflation, which causes trigger-delay and increases the num- flow plateau. Figure 46 illustrates another case study in-
ber of missed trigger attempts. Though there may be differ- volving delayed termination in the pressure-control mode.
ences between patient groups, most patients seem to have Inspiratory time is prolonged excessively, and the pressure
rather weak compensatory responses to acute delays in breath- waveform illustrates repeated expiratory attempts by the
termination52,53 and generally do not make acute changes in patient.
their expiratory timing following a single inspiration. Data from diseased subjects, with a variety of disorders,
Even in normal healthy subjects the effects of delayed indicate that inspiratory times are often in the range of 1
termination can be assessed by carefully evaluating the second or less.54 Though there are clearly instances in
pressure and flow waveforms. Figure 44 illustrates pres- which variations occur, the less-than-1-second guideline
sure and flow waveforms during pressure support ventila- should be used as an approximate starting point for eval-
tion of a healthy respiratory care student. Those wave- uating and setting inspiratory time.47,54 During pressure-
forms demonstrate the same delayed breath-termination controlled ventilation the initial inspiratory time can be set
characteristics as in the COPD patient noted above. There with the less-than-1-second guideline, and then subsequent
is an obvious pressure spike near the end of the breath, adjustments should be based on the time for the inspiratory
which coincides with a rapid decline in flow, indicating flow waveform to decay to zero, but not beyond, and to
the subject’s attempt to exhale. A similar pattern is seen in prevent a pressure spike near the end of inspiration. Dur-
the waveform in Figure 45, which was obtained during ing pressure-support ventilation the expiratory sensitivity
pressure-regulated volume-control ventilation (on a Servo can be adjusted within this range, while fine tuning can
300A ventilator). Again, there is an obvious pressure spike occur in 2 directions: if the breath is too long, there will be
at the same time as the zero-flow plateau in the flow a pressure spike near the end of the breath; if the breath is
waveform. The zero-flow plateau indicates achievement of too short, the patient may continue to inhale, resulting in
Fig. 45. Waveforms from a normal subject ventilated with pressure-regulated volume control on a Servo 300A ventilator. The subject’s
expiratory effort begins just prior to the end of the mechanical inspiratory time, which causes pressure spikes (arrows). Also note the small
inspiratory zero flow plateau at the end of the inspiratory flow waveform, which indicates that flow into the lung has stopped just prior to
mechanical expiration. In pressure-controlled modes the inspiratory time is set on the ventilator, and in this instance the ventilator
inspiratory time setting needs to be reduced slightly.
Premature breath-termination also has deleterious ef- We have discussed termination asynchrony as a sepa-
fects and causes asynchrony in patients with acute respi- rate consideration to emphasize the importance of per-
ratory distress syndrome. Tokioka et al55 evaluated the forming a stepwise analysis of the breathing pattern. The
Fig. 47. Left Panel: Flow (V̇), volume (V), airway pressure (Paw), and esophageal pressure (Pes) waveforms with termination criteria (TC) of 5% and
35% of peak flow, during ventilation with pressure support of 10 cm H2O. With TC 5% the breathing pattern was regular. Tidal volume was 390
mL and respiratory frequency was 17 breaths/min. The negative deflection of Pes during inspiration was minimal. With TC 35%, tidal volume
decreased to 281 mL, and respiratory frequency increased to 23 breaths/min. The inspiratory flow terminated despite continuous negative
deflection of Pes. Work of breathing increased from 0.20 J/L with TC 5% to 0.32 J/L with TC 35%. The arrows indicate continued patient
inspiratory effort on the expiratory flow waveform (convex pattern) and the airway pressure waveform (concavity). Right panel: TC 5% versus TC
45% during ventilation with pressure support of 10 cm H2O, with case 2. With TC 5%, inspiratory flow terminated simultaneously with the
cessation of the patient’s inspiratory effort, estimated by Pes. In contrast, premature termination with double-breathing (circled) occurred with TC
45%. Work of breathing also increased, from 0.42 J/L with TC 5% to 0.64 J/L with TC 45%. Note the larger Pes deflection during TC 45%. Also
during TC 45% note the more pronounced changes (arrows) in the expiratory flow curve and the expiratory portion of the airway pressure
waveform, caused by continued patient inspiratory effort despite early termination of the mandatory breath. (From Reference 55, with permission.)
remaining portion of the breath to consider is the expira- Shortened expiratory time creates the potential for air
tory time and the potential clinical consequences of short- trapping and auto-PEEP, which can cause trigger asyn-
ened or prolonged expiratory time. chrony because of the additional effort required to pull
Shortened expiratory time has major clinical implica- through the auto-PEEP to reach the trigger thresh-
tions because of the risk of causing auto-PEEP. Prolonged old.21,47,59,60 Breaths preceding missed breaths typically
expiratory time is of little consequence, unless the expi- have longer inspiratory times and shorter expiratory times,
ratory time is so long as to cause hypoventilation. Expi- and are associated with auto-PEEP. In completing the breath
ratory asynchrony can occur in conditions in which there analysis it is perhaps circuitous to note that the final con-
is delay in the relaxation of the expiratory-muscle activity sideration of auto-PEEP has as one of its main impacts the
prior to the next mechanical inspiration47 or overlap be- creation of trigger asynchrony, and that each of the fol-
tween expiratory and inspiratory muscle activity,56 –58 but lowing portions of the breath (inspiratory flow and termi-
those conditions are verified with neural measurements nation) also have as a consequence of asynchrony the cre-
and are beyond the scope of our discussion about use of ation of auto-PEEP. In other words, flow patterns that
the standard ventilator waveforms to evaluate patient-ven- increase inspiratory time (eg, lower peak flow during con-
tilator synchrony. trol ventilation, switch from constant-flow to descending-
deceleration and a convex pattern in the expiratory flow 13. Younes M. Proportional assist ventilation. In: Tobin MJ, editor. Prin-
waveform, which indicates continued patient effort. ciples and practice of mechanical ventilation. New York: McGraw-
Hill; 1994.
14. Gay PC, Hess DR, Hill NS. Noninvasive proportional assist venti-
Expiratory Asynchrony (Phase 4) lation for acute respiratory insufficiency: comparison with pressure
support ventilation. Am J Respir Crit Care Med 2001;164(9):1606–
1611.
The major consideration during the expiratory phase is 15. Passam F, Hoing S, Prinianakis G, Siafakas N, Milic-EmiliJ, Geor-
the presence or creation of auto-PEEP, which can be caused gopoulos D. Effect of different levels of pressure support and pro-
by insufficient expiratory time or by asynchrony created portional assist ventilation on breathing pattern, work of breathing
during the flow or termination phases. 1. Routinely check and gas exchange in mechanically ventilated hypercapnic COPD
patients with acute respiratory failure. Respiration 2003;70(4):355–
for auto-PEEP by inspecting the end of the expiratory flow 361.
waveform and by measuring auto-PEEP (most commonly 16. Navalesi P, Hernandez P, Wongsa A, Laporta D, Goldberg P, Got-
performed at the bedside with the expiratory-hold tech- tfried SB. Proporational assist ventilation in acute respiratory failure:
nique). 2. If auto-PEEP is identified, adjust, if possible, effects on breathing pattern and inspiratory effort. Am J Respir Crit
trigger sensitivity, peak flow, flow pattern, pressure slope, Care Med 1996;154(5):1330–1338.
17. Appendini L, Purro A, Gudjonsdottir M, Baderna P, Patessio A,
inspiratory time, breath-termination criteria, expiratory
Zanaboni S, et al. Physiologic response of ventilator-dependent pa-
time, or respiratory rate, depending on the ventilation mode. tients with chronic obstructive pulmonary disease to proportional
3. Trigger problems associated with auto-PEEP may im- assist ventilation and continuous positive airway pressure. Am J
prove with applied PEEP, approximately equal to the auto- Respir Crit Care Med 1999;159(5 Pt 1):1510–1517.
PEEP.61 18. Giannouli E, Webster K, Roberts D, Younes M. Response of venti-
lator-dependent patients to different levels of pressure support and
proporational assist. Am J Respir Crit Care Med 1999;159(6):1716–
1725.
REFERENCES
19. Wysocki M, Meshaka P, Richard JC, Similowski T. Proportional-
1. Joint Commission on Accreditation of Health Care Organizations. assist ventilation compared with pressure-support ventilation during
Preventing ventilator-related deaths and injuries. Sentinel Event Alert. exercise in volunteers with external thoracic restriction. Crit Care
Issue 25, Feb 26, 2002. Available at http://www.jcaho.org/about⫹us/ Med 2004;32(2):409–414.
news⫹letters/sentinel⫹event⫹alert/sea 25.htm. (Accessed Decem- 20. Du HL, Ohtsuji M, Shigeta M, Chao DC, Sasaki K, Usuda Y, Yamada
ber 10, 2004). Y. Expiratory asynchrony in proportional assist ventilation. Am J
2. Kondili E, Prinianakis G, Georgopoulos D. Patient-ventilator inter- Respir Crit Care Med 2002;165(7):972–977.
action. Br J Anaesth 2003;91(1):106–119. 21. Leung P, Jubran A, Tobin J. Comparison of assisted ventilator modes
3. Stoller JK. The effectiveness of respiratory care protocols. Respi on triggering, patient effort, and dyspnea. Am J Respir Crit Care
Care 2004;49(7):761–765. Med 1997;155(6):1940–1948.
4. Sassoon CSH. Mechanical ventilator design and function: the trigger 22. Fabry B, Guttmann J, Eberhard L, Bauer T, Haberthur C, Wolff G.
variable. Respir Care 1992;37(9):1056–1069. An analysis of desychronization between the spontaneously breath-
5. Marini JJ, Capps JS, Culver BH. The inspiratory work of breathing ing patient and ventilator during inspiratory pressure support. Chest
during assisted mechanical ventilation. Chest 1985;87(5):612–618. 1995;107(5):1387–1394.
6. Chao DC, Scheinhorn DJ, Stearn-Hassenpflug M. Patient-ventilator 23. Rossi A, Polese G, Brandi G, Conti G. Intrinsic positive end-expi-
trigger asynchrony in prolonged mechanical ventilation. Chest 1997; ratory pressure (PEEPi). Intensive Care Med 1995;21(6):522–536.
112(6):1592–1599. 24. Younes M. Interactions between patients and ventilator. In: Roussos
7. Chatmongkolchart S, Williams P, Hess DR, Kacmarek RM. Evalu- C, editor. The thorax, 2nd ed. New York: Marcel Dekker; 1995:
ation of inspiratory rise time and inspiration termination criteria in 2367–2420.
new-generation mechanical ventilators: a lung model study. Respir 25. Hill LL, Pearl RG. Flow triggering, pressure triggering, and auto-
Care 2001;46(7):666–677. triggering during mechanical ventilation (editorial). Crit Care Med
8. Calzia E, Lindner KH, Stahl W, Martin A, Radermacher P, Georgieff 2000;28(2):579–581.
M. Work of breathing, inspiratory flow response, and expiratory 26. Imanaka H, Nishimura M, Takeuchi M, Kimball WR, Yahagi N,
resistance during continuous positive airway pressure with the ven- Kumon K. Autotriggering caused by cardiogenic oscillation during
tilators EVITA-2, EVITA-4 and SV 300. Intensive Care Med 1998; flow-triggered mechanical ventilation. Crit Care Med 2000;28(2):
24(9):931–938. 402–407.
9. Richard JC, Carlucci A, Breton L, Langlais N, Jaber S, Maggiore S, 27. Kacmarek R, Hess D, Stoller J. Airway pressure, flow and volume
et al. Bench testing of pressure support ventilation with three differ- waveforms, and lung mechanics during mechanical ventilation. In:
ent generations of ventilators. Intensive Care Med 2002;28(8):1049– Monitoring in respiratory care. St Louis: Mosby-Year Book; 1993:
1057. Chapter 16.
10. Takeuchi M, Williams P, Hess D, Kacmarek RM. Continuous pos- 28. Tobin MJ. What should the clinician do when a patient “fights the
itive airway pressure in new-generation mechanical ventilators: a ventilator?” Respir Care 1991;36(5):395–406.
lung model study. Anesthesiology 2002;96(1):162–172. 29. Nilsestuen JO, Hargett K. Managing the patient-ventilator system
11. Branson RD, Johannignam JA. What is the evidence for the newer using graphic analysis: an overview and introduction to Graphics
ventilation modes? Respir Care 2004;49(7):742–760. Corner. Respir Care 1996;41(12):1105–1122.
12. Zarske R, Doring M. ATC and PPC breathing support with optimum 30. Yang SC, Yang SP. Effects of Inspiratory flow waveforms on lung
patient comfort. Draeger Medizintechnik Gmbh 1997 ISBN mechanics, gas exchange, and respiratory metabolism in COPD pa-
3–926762-27–6. tients during mechanical ventilation. Chest 2002;122(6):2096–2104.
31. Kallet R. The effects of flow patterns on pulmonary gas exchange, 47. Parthasarathy S, Jubran A, Tobin MJ. Cycling of inspiratory and
lung-thorax mechanics, and circulation. Respir Care 1996;41(7):611– expiratory muscle groups with the ventilator in airflow limitation.
624. Am J Respir Crit Care Med 1998;158(5 Pt 1):1471–1478. Erratum
32. Georgopoulos D, Kondili E, Prinianakis G. How to set the ventilator in: Am J Respir Crit Care Med 1999;159(3):1023.
in asthma. Monaldi Arch Chest Dis 2000;55(1):74–83. 48. Parthasarathy S, Jubran A, Tobin MJ. Assessment of neural inspira-
33. Sassoon CS, Foster GT. Patient-ventilator asynchrony. Curr Opin tory time in ventilator-supported patients. Am J Respir Crit Care
Crit Care 2001;7(1):28–33. Med 2000;162(2 Pt1):546–552.
34. Prinianakis G, Kondili E, Kostaki M, Georgopoulos D. Effect of 49. Bartoli A, Bystrzycka E, Guz A, Jain SK, Noble MI, Trenchard D.
signal method of triggering and cycling off on patient-ventilator Studies of the pulmonary vagal control of central respiratory rhythm in
interaction. Eur Respir J 2002;20:s603. the absence of breathing movements. J Physiol 1973;230(2):449–465.
35. Tuxen DV, Williams TJ, Scheinkestel CD, Czarny D, Bowes G. Use
50. Tobin MJ. Monitoring of respiratory mechanics in spontaneously
of a measurement of pulmonary hyperinflation to control the level of
breathing patients. In: Tobin MJ, editor. Principles and practice of
mechanical ventilation in patients with acute severe asthma. Am Rev
intensive care monitoring. New York: McGraw-Hill; 1998:617–653.
Respir Dis 1992;146(5 Pt 1):1136–1142.
51. Haluszka J, Chartrand DA, Grassino AE, Millic-Emili J. Intrinsic
36. Williams TJ, Tuxen DV, Scheinkestel CD, Czarny D, Bowes G. Risk
PEEP and arterial PCO2 in stable patients with chronic obstructive
factors for morbidity in mechnically ventilated patients with acute
severe asthma. Am Rev Respir Dis 1992;146(3):607–615. pulmonary disease. Am Rev Respir Dis 1990;141(5 Pt 1):1194–
37. Marini JJ, Capps JS, Culver BH. The inspiratory work of breathing 1197.
during assisted mechanical ventilation. Chest 1985;87(5):612–618. 52. Jubran A, van de Graaff WB, Tobin MJ. Variability of patient-
38. Marini JJ, Rodriguez RM, Lamb V. Bedside estimation of the in- ventilator interaction with pressure support ventilation in patients
spiratory work of breathing during mechanical ventilation. Chest with chronic obstructive pulmonary disease. Am J Respir Crit Care
1986;89(1):56–63. Med 1995;152(1):129–136.
39. Johnson MD, Lawson JJ, Campbell RS. Using waveforms to opti- 53. Younes M, Kun J, Webster K, Roberts D. Response of ventilator-
mize inspiratory rise time during pressure support ventilation. Respir dependent patients to delayed opening of exhalation valve. Am J
Care 1997;42(5):520–526. Respir Crit Care Med 2002;166(1):21–30.
40. Branson RD, Campbell RS, Davis K Jr, Johannigam JO, Johnson DJ, 54. Tobin MJ, Chadha TS, Jenouri G, Birch SJ, Gazeroglu HB, Sackner
Hurst JM. Altering flowrate during maximum pressure support ven- MA. Breathing patterns. 2. Diseased subjects. Chest 1983;84(3):
tilation (PSVmax): effects on cardiorespiratory function. Respir Care 286–294.
1990;35(11):1056–1064. 55. Tokioka H, Tanaka T, Ishizu T, Fukushima T, Iwaki T, Nakamura Y,
41. MacIntyre NR, Ho LI. Effects of initial flow rate and breath termi- Kosogabe Y. The effect of breath termination criterion on breathing
nation criteria on pressure support ventilation. Chest 1991;99(1): patterns and the work of breathing during pressure support ventila-
134–138. tion. Anesth Analg 2001;92(1):161–165.
42. Bonmarchand G, Chevron V, Chopin C, Jusserand D, Girault C, 56. Dodd DS, Brancatisano T, Engel LA. Chest wall mechanics during
Moritz F, et al. Increased initial flow rate reduces inspiratory work of exercise in patients with severe chronic air-flow obstruction. Am
breathing during pressure support ventilation in patients with exac- Rev Respir Dis 1984;129(1):33–38.
erbation of chronic obstructive pulmonary disease. Intensive Care 57. Abe T, Kusuhara N, Yoshimura N, Tomita T, Easton PA. Differen-
Med 1996;22(11):1147–1154.
tial respiratory activity of four abdominal muscles in humans. J Appl
43. Bonmarchand G, Chevron V, Menard JF, Girault C, Moritz-Berthe-
Physiol 1996;80(4):1379–1389.
lot F, Pasquis P, Leroy J. Effects of Pressure ramp slope values on
58. Martin JG, De Troyer A. The behavioUr of the abdominal muscles
the work of breathing during pressure support ventilation in restric-
during inspiratory mechanical loading. Respir Physiol 1982;50(1):63–
tive patients. Crit Care Med 1999;27(4):715–722. Erratum in: Crit
73.
Care Med 1999;27(7):1404.
44. MacIntyre NR, McConnell R, Chang KC, Sane A. Patient-ventilator 59. Smith TC, Marini JJ. Impact of PEEP on lung mechanics and work
flow dyssynchrony: flow-limited versus pressure-limited breaths. Crit of breathing in severe airflow obstruction. J Appl Physiol 1988;
Care Med 1997;25(10):1671–1677. 65(4):1488–1499.
45. Cinnella G, Conti G, Lofaso F, Lorino H, Harf A, Lemare F, Bro- 60. Petrof B, Legare M, Goldberg P, Milic-Emili J, Gottfried S. Continuous
chard L. Effects of assisted ventilation on the work of breathing: positive airway pressure reduces work of breathing and dyspnea during
volume-controlled versus pressure-controlled ventilation. Am J Re- weaning from mechanical ventilation in severe chronic obstructive pul-
spir Crit Care Med 1996;153(3):1025–1033. monary disease. Am Rev Respir Dis 1990;141():281–289.
46. MacIntyre NR, Gropper C, Westfall T. Combining pressure-limiting 61. Tobin MJ, Fahey PJ. Management of the patient who is “fighting the
and volume-cycling features in a patient-interactive mechanical ventilator.” In: Tobin MJ, editor. Principles and practice of mechan-
breath. Crit Care Med 1994;22(2):353–357. ical ventilation. New York: McGraw-Hill; 1994:1149–1162.
Discussion the ventilators I know of, there is no patient to breathe through the circuit in
way you can breathe through that valve. case the machine shut down.
Sanborn: I’m somewhat confused
about the concept of breathing through Nilsestuen: You are absolutely right. Benditt: This harkens back to Dean
the demand valve. In the past there was Modern ventilators have reduced the pa- Hess’s “eyeball test.” If the patient seems
an actual demand valve—a scuba-type tient’s ability to do that. The sample to be synchronous with the ventilator, if
valve. In today’s ventilators there is no waveform that this refers to is from the you look at the flow waveform, will you
equivalent of that. The valve either trig- Bear 1000. In this ventilator there was a pick up abnormalities that you would
gers or it doesn’t trigger, and in most of backup demand valve that allowed the not suspect with the eyeball test?
Nilsestuen: If you carefully evalu- physical examination I’m going to about the experienced practitioners he
ate the flow waveform? Is that what look at the flow waveform to make has had in his classes who have come
you’re asking? Or does the eyeball test sure that things are OK here.” Person- back for advanced training, and how
mean observing the patient to deter- ally, I would do as you say—just ex- confident and familiar they are with
mine if he or she is synchronous with amine the patient, and if there’s a prob- all of these things, but how poorly they
the ventilator? lem then investigate it with these tools. do on actual examinations that reflect
But if it’s the other way around, then their understanding of what they do.
Benditt: No. Say I go to the bedside I might have to change my practice It’s been my observation for a long
and I look at the patient, and the pa- and say that to maximize synchrony I time that clinicians— critical care
tient looks relaxed, isn’t using acces- should include the flow waveform as nurses for example—are often very fa-
sory muscles, the abdominal muscles part of my routine rounds. Sounds like miliar and comfortable with all the he-
aren’t contracting, and everything we’ve got a little difference of opin- modynamic data they deal with, but
looks good to me as a clinician. Then ion as to whether that’s part of the may have incomplete understanding
I turn on the flow waveform. Will I eyeball test. of what those data represent physio-
see these things popping out at me logically and what they actually mean.
that I didn’t suspect? Nilsestuen: My sense is that there Likewise, physicians who have been
are some very obvious things that you in practice for a while may have de-
Nilsestuen: My take on that is that can see the patient do. But, for in- veloped routines and familiarity and
sometimes, yes, you will see some stance, if the neural timing is off just comfort and fluency with the data that
things that you didn’t suspect, because a little bit, and there’s just a tiny little they deal with every day, but may not
the patient is not yet exhibiting a strong bump in the pressure waveform, you interpret that data in a manner consis-
enough apparent effort or obvious won’t see that just by looking at the tent with our current understanding of
muscular activity that counters what patient, because the patient will look what it ought to mean. So I think this
the ventilator is doing. The subtleties fairly relaxed, whereas in fact he’s just is a very complex issue that involves
start showing up in the waveforms be- starting to recruit muscular activity both the clinician as subjective observ-
fore it’s obvious at the bedside. right at the end. That’s fine-tuning, er—the eyeball test— but also knowl-
definitely, but, I think looking at the edge of the technical aspects of what
Hess: Regarding what Josh Benditt graphics makes asynchrony very ap- the waveforms are telling us. I think
just brought up, I teach the respiratory parent, and then you can document neither of them is 100% the right an-
therapists, fellows, residents—anyone that there is a subtlety there that may swer, and they may sometimes be in
who will listen to me—to look at the improve patient comfort. conflict with one another.
patient, listen to the ventilator, and un-
derstand the physiology. I teach them to Pierson:* I think this stuff is pretty Hess: I am reminded, David, of a
look at the patient and listen to the ven- hard. I’m hearing Dean say that the picture that you published in one of
tilator, and if you see that the patient is eyeball test is valuable, and that how Martin Tobin’s books,1 which shows
making inspiratory efforts and you hear the patient looks in general should be the respiratory therapist peering at the
no response from the ventilator, there’s your overall guide as to how things ventilator, and the nurse peering at the
auto-PEEP. Then look for chest recoil, are going. But I’m also hearing Jon monitor, and the residents all looking
and if you see that the thorax is still saying that you can see things on the at the laboratory results, and nobody’s
recoiling when the next breath starts, graphics displays that are early warn- looking at the patient!
there’s auto-PEEP—the same thing ings of important things you haven’t
you’ll see on the flow waveform. So picked up by your observation. Now,
first look at the patient and listen to the I suspect that Jon would agree that REFERENCE
ventilator, then look at the graphics and there are things that one might see on
do all these other assessments to try to the waveforms that are not clinically 1. Pierson DJ. Goals and indications for mon-
itoring. In: Tobin MJ, editor. Principles and
understand the underlying physiology important. On the other hand, though, practice of intensive care monitoring. New
that is producing the clinical problem. the eyeball test depends very much on York: McGraw-Hill; 1998:33–44. [Figure
So, personally, I like graphics a lot, but whose eyeball it is. At the very begin- 2–2 on page 40]
I still rely a lot on the eyeball test. ning of his presentation Jon told us
Campbell: I agree with what you
Benditt: I think the reason that’s just said. I think it’s very important.
important, I would say, is because I * David J Pierson MD FAARC, Division of The patient’s appearance may lead you
don’t routinely on rounds go to the Pulmonary and Critical Care Medicine, Univer- to look at the graphics to diagnose
bedside and say, “OK, as a part of my sity of Washington, Seattle, Washington. and maybe fine-tune, but certainly with
advanced settings (such as rise time Hess: Call me old-fashioned. I look not forget that the primary setting is
and expiratory cycle sensitivity or ter- at graphics a lot and I like graphics a the pressure control, and the setting of
mination criterion), there’s no way you lot, but I have many times set PEEP pressure control is going to affect the
can set those appropriately just by your with patients who have auto-PEEP by flow waveform, or whatever wave-
view of what the patient’s doing. With simply looking at the patient, looking form you decide to look at. So what
lots of the graphs that Jon showed, if for failed triggering efforts, and just we really need is some tool that would
you didn’t have the line to line it up. . . . turning up the PEEP until every in- tell us if we have the base setting cor-
So I would never think you could put spiratory effort triggers the ventilator. rect before we take that next leap to
your hand on the patient’s chest or the secondary settings. The volume-
diaphragm and put your other thumb Pierson: But that’s your eyeball, support slides that you showed were
on the graphic and say, “Ah-ha, I think Dean, and not necessarily any random classic; they were showing all the
we’re in synchrony.” I don’t think eyeball. missed triggers. Well, maybe they just
that’s a possibility, and I think you didn’t have a high enough volume set.
really need those graphics to perfectly Harris: This sort of reminds me of Maybe it wasn’t the cycle criterion
set it. Respiratory therapists ask what the physical examination versus, say, that was the problem.
to do with the ventilator’s expiratory echocardiography for the heart. If you
sensitivity setting and I tell them to have a very experienced person who Benditt: I have a comment about
take the default, and if you see a pres- can understand what’s going on or has
Dr Pierson’s question, “Whose eye-
sure spike, adjust it. That’s the prob- a lot of experience looking at a pa-
ball is it?” If you think about it, it is
lem; we don’t have the equivalent of a tient, then the physical examination is
much more common to teach the
PDR [Physicians Desk Reference] for very helpful. If you have somebody
physical examination. Maybe it’s eas-
the ventilator that would tell us, “if who is not that experienced, it might
ier to get people to understand what
this, then that,” and “this is the dose,” not be so helpful.
muscle effort is in a patient who is
and “this is how to do it.” Jon, you described asynchrony as 2
asynchronous with the ventilator than
pumps that are not working together,
Nilsestuen: Another example of not in harmony, and I’m thinking that it is to evaluate waveforms, set up the
something that is far better done using we have a lot of information about line to look at the waveform overlap,
the graphics than with just the visual one of the pumps, the ventilator, but and so forth. I think you need at least
technique of looking at the patient is we don’t have a lot of information as much, and probably more, exper-
setting the correct inspiratory time dur- about what’s driving the human pump. tise to look at the waveforms as to
ing pressure-controlled ventilation. Should we be using things that Jon look at the muscle function at the bed-
There’s almost no way you can do was talking about? Should we be look- side. So, unless you are going to have
that visually by just looking at the pa- ing at diaphragm function? Should we an echocardiogram or have the thera-
tient, but it’s very easy to do if you be looking at neural time? Is it over- pist or some expert person do a daily
understand the graphics and you just kill, or would it be helpful to have that analysis of these waveforms for you
look at the flow waveform. Because information to synchronize the 2 and then put it in a report and send
then you can see the plateau if it’s too pumps? it to you, I think it’s probably better
long, you can see it chop off if it’s to teach them the physical examina-
terminating too soon. That’s maybe Campbell: We’re spending an aw- tion.
one of the best examples where it is ful lot of time talking about secondary
very clear that the graphics have a huge settings, such as setting the inspira- Hess: Glad to know I’m not the only
advantage. tory time during pressure control. Let’s old curmudgeon in the room.