Hi 98194

Download as pdf or txt
Download as pdf or txt
You are on page 1of 37

Instruction Manual

HI 98194, HI 98195,
HI 98196
Multiparameter Meters

w w w . h a n n a i n s t . c o m
1

Downloaded from www.Manualslib.com manuals search engine


TABLE OF CONTENTS
Dear Customer, CHAPTER 1- INTRODUCTION ....................................................................... 6
Thank you for choosing a HANNA Instruments® product. 1.1 Preliminary Examination ................................................................................. 6
Please read this instruction manual carefully before using the instrument. 1.2 Model Identification ....................................................................................... 6
It will provide you with the necessary information for correct use of the instrument, 1.3 General Description....................................................................................... 6
as well as it’s versatility. 1.4 Display and Keypad Description ...................................................................... 8
If you need additional technical information, do not hesitate to e-mail us at CHAPTER 2 - QUICK START ........................................................................... 9
[email protected] or visit our website www.hannainst.com for our
2.1 Sensor and Probe Installation .......................................................................... 9
worldwide contact list.
2.2 Basic Operation ........................................................................................... 11
2.3 Help Function .............................................................................................. 11
CHAPTER 3 - SPECIFICATIONS ...................................................................... 12
3.1 System Specifications ..................................................................................... 12
3.2 Probe Specifications ...................................................................................... 16
3.3 Sensor Specifications ..................................................................................... 17
CHAPTER 4 - PROBE INSTALLATION .............................................................. 18
4.1 Sensor Descriptions ....................................................................................... 18
4.2 Sensor Preparation/Activation ........................................................................ 19
4.2.1 pH Preparation ..................................................................................................... 19
4.2.2 ORP Activation ..................................................................................................... 19
4.2.3 D.O. Sensor Activation ......................................................................................... 20
4.2.4 EC Sensor Preparation .......................................................................................... 20
4.3 Sensor Installation ........................................................................................ 20
CHAPTER 5 - INITIALIZATION AND MEASUREMENT ....................................... 23
5.1 Battery Installation ........................................................................................ 23
5.2 Meter Initialization ........................................................................................ 24
5.3 Measurement Mode ...................................................................................... 25
5.4 Setup Menu Structure .................................................................................... 26
CHAPTER 6 - PARAMETER SETUP MENU ........................................................ 27
6.1 Select Parameters .......................................................................................... 27
6.2 Parameter Units ............................................................................................ 27
6.2.1 Temperature Unit .................................................................................................. 27
6.2.2 TDS Unit (HI 98194, HI 98195 only) ..................................................................... 27
6.2.3 DO Concentration Unit (HI 98194, HI 98196 only) ................................................ 27
HANNA instruments® reserves the right to modify the design,
6.2.4 Pressure Unit ........................................................................................................ 27
construction or appearance of its products without advance notice.
2 3

Downloaded from www.Manualslib.com manuals search engine


6.2.5 Resistivity Unit (HI 98194, HI 98195 only) ............................................................. 28 8.1.9 Meter Password .................................................................................................... 46
6.2.6 Seawater Sigma Unit (HI 98194, HI 98195 only) .................................................... 28 8.1.10 Meter ID ............................................................................................................ 46
6.2.7 EC Resolution Unit (HI 98194, HI 98195 only) ....................................................... 28 8.1.11 Language ........................................................................................................... 47
6.2.8 Absolute EC Resolution Unit (HI 98194, HI 98195 only) .......................................... 28 8.1.12 Restore Factory Settings ....................................................................................... 47
6.2.9 TDS Resolution Unit (HI 98194, HI 98195 only) ..................................................... 28 8.2 Probe Setup ................................................................................................. 47
6.3 Parameter Coefficients ................................................................................... 29 8.2.1 Probe ID .............................................................................................................. 47
6.3.1 EC Reference Temperature (HI 98194, HI 98195 only) ............................................ 29 CHAPTER 9 - STATUS .................................................................................... 48
6.3.2 EC Temperature Coefficient (HI 98194, HI 98195 only) .......................................... 29 9.1 Meter Status ................................................................................................. 48
6.3.3 TDS Factor (HI 98194, HI 98195 only) .................................................................. 29 9.2 Probe Status ................................................................................................ 48
6.4 Averaging ................................................................................................... 29 9.3 GLP Data .................................................................................................... 49
CHAPTER 7 - CALIBRATION MODE ................................................................ 30 CHAPTER 10 - LOGGING MODE ................................................................... 52
7.1 Quick Calibration ........................................................................................ 31 10.1 Logging Menu Structure ............................................................................... 53
7.2 pH Calibration ............................................................................................. 32 10.2 Logging On Meter ...................................................................................... 53
7.2.1 Preparation .......................................................................................................... 32 10.2.1 One Sample On Meter ....................................................................................... 54
7.2.2 Procedure ............................................................................................................ 33 10.2.2 Continuous Meter Log ........................................................................................ 54
7.2.3 pH Calibration Error Messages .............................................................................. 34 10.3 Log Recall ................................................................................................. 55
7.3 Relative mV Calibration ................................................................................. 35 10.4 Log Notes .................................................................................................. 56
7.3.1 Preparation .......................................................................................................... 36 10.4.1 Remarks ............................................................................................................ 56
7.3.2 Procedure ............................................................................................................ 36 10.4.2 Delete All Remarks .............................................................................................. 56
7.4 Dissolved Oxygen Calibration (HI 98194, HI 98196 only) .............................. 36 CHAPTER 11 - PC CONNECTION .................................................................. 57
7.5 Conductivity Calibration (HI 98194, HI 98195 only) ...................................... 38 11.1 Software Installation .................................................................................... 57
7.6 Temperature Calibration ................................................................................ 42 11.2 Meter to PC Connection .............................................................................. 57
7.7 Atmospheric Pressure Calibration ................................................................... 43 CHAPTER 12 - TROUBLESHOOTING / ERROR MESSAGES ............................... 59
CHAPTER 8 - SYSTEM SETUP ......................................................................... 44 APPENDIX
8.1 Meter Setup ................................................................................................. 44 A - PROBE MAINTENANCE ................................................................................ 61
8.1.1 Time ................................................................................................................... 44 B - PROBE DEPLOYMENT .................................................................................. 63
8.1.2 Date .................................................................................................................... 44 C - ACCESSORIES ............................................................................................. 65
8.1.3 Auto Poweroff ...................................................................................................... 45 D - WARRANTY ................................................................................................. 70
8.1.4 Key Beep ............................................................................................................. 45
8.1.5 Error Beep ........................................................................................................... 45
8.1.6 Decimal Separator ................................................................................................ 45
8.1.7 LCD Contrast ....................................................................................................... 45
8.1.8 LCD Backlight Intensity .......................................................................................... 46

4 5

Downloaded from www.Manualslib.com manuals search engine


Chapter 1 - INTRODUCTION Main features of the HI 9819X systems:
• Rugged meter and probe
1.1 PRELIMINARY EXAMINATION • Easy to use
Remove the instrument from the packing material and examine it carefully to • Measure up to 14 parameters and display of up to 12 parameters
make sure that no damage has occurred during shipping. If there is any damage,
• Waterproof protection (IP67 for the meter and IP68 for the probe)
notify your Dealer or the nearest HANNA Customer Service Center immediately.
• Graphic LCD with backlight
Note Save all packing materials until you are sure that the instrument • Built-in barometer for D.O. concentration compensation (HI 98194, HI 98196
functions correctly. Any damaged or defective items must be returned only)
in their original packing material with the supplied accessories.
• Quick calibration feature
• Measurement check to eliminate any erroneous readings
1.2 MODEL IDENTIFICATION
HI 98194 pH/EC/DO Multiparameter meter with HI 7698194 probe. • Autorecognition of probe and sensors
HI 98195 pH/ORP/TDS/NaCl/Resistivity meter with HI 7698195 probe. • Log-on-demand and automatic logging (up to 45,000 samples) on meter for
all parameters
HI 98196 pH/ORP/Dissolved Oxygen meter with HI 7698196 probe.
• Graphical display of logged data
1.3 GENERAL DESCRIPTION • USB interface for PC communication
HI 9819X is a portable logging multiparameter system (instrument, probe) that • Auto-ranging for EC readings (HI 98194, HI 98195 only)
monitors up to 14 different water quality parameters (7 measured, 7 calculated). • Good Laboratory Practice feature, the last 5 calibrations are automatically
The microprocessor-based intelligent multisensor probes allows measurement of stored
many water quality parameters such as pH, ORP, dissolved oxygen, conductivity • Field-replaceable sensors with color coded caps
and temperature with data logging. The system is easy to setup and easy to use. • Meter is powered with alkaline batteries
The HI 9819X features a graphic, backlit display that automatically sizes the
digits to fit the screen with on-screen graphing capability. Each parameter is fully
configurable.
HI 9819X was designed to withstand harsh environments and is the ideal solution
for field measurements of lakes, rivers and sea.
The meter meets IP67 standards (30 minute immersion at a depth of 1 m) and
the multisensor probe meets IP68 standards (continuous immersion in water).

6 7

Downloaded from www.Manualslib.com manuals search engine


1.4 DISPLAY AND KEYBOARD DESCRIPTION Chapter 2 - QUICK START

2.1 SENSOR AND PROBE INSTALLATION


• Sensor o-rings must be lubricated with the supplied grease prior to installation.
• HI 769819X probes have 2 or 3 sensor connectors sockets identified with
color-coded triangles:
• Connector 1 (red): For either pH/ORP, pH sensor
• Connector 2 (white): For dissolved oxygen sensor (HI 98194, HI 98196
only)
• Connector 3 (blue): For EC sensor (HI 98194, HI 98195 only)
• Position the connector key towards the center of the probe, make sure the
connector is seated correctly (the sensor will no longer move freely) before
tightening the locking threads.
• To protect the sensors, screw the protective shield onto the probe body.
• With the meter off, connect the probe to the instrument input on the top of the
meter. Align the pins and key then push the plug into the socket and tighten
the thread.
• Turn the meter on by pressing the On/Off key. The meter will automatically
recognize the probe and the installed sensors and identify them on the probe
status screen.
• Press Measure to view the measurement screen.

1. Graphic LCD
2. Battery level indicator
3. Softkeys
4. On/Off key: turn the meter on and off
5. Lamp key: turn the backlight on and off
6. Alphanumeric keyboard: insert alphanumeric codes
7. / Arrow keys: scroll the displayed options/message
8. HELP key: obtain information about the displayed screen
9. ESC key: return to the previous screen
10. Softkey functions defined on display

8 9

Downloaded from www.Manualslib.com manuals search engine


2.2 BASIC OPERATION
The main operating modes for HI 9819X are measurement, logging and setup.
The measurement screen can be configured to display a single measurement or
up to 12 simultaneous measurements by using the numbers 1-7 on the keypad.
Use the arrow keys to scroll through the measurements not being displayed.
See section 5.3 for more details.
The measurement units will blink if the system has not been calibrated and the
measurement number will blink when the reading is out of range.
Press Log to display the logging menu. You can either log a single sample on the
meter or start an interval log on the meter. See chapter 10 for more details.
Press Menu to enter setup mode. You can configure which parameters you
want to measure, calibrate the sensors, change system settings and view the
meter and probe status.
2.3 HELP FUNCTION
HI 9819X features context sensitive HELP, which provides useful information
regarding the displayed screen.
Simply press the HELP key to access this function, then use the arrow keys to
scroll through the message.
To escape from the HELP window, press the HELP key again or ESC.

10 11

Downloaded from www.Manualslib.com manuals search engine


Chapter 3 - SPECIFICATIONS CONDUCTIVITY (HI 98194, HI 98195 only)
Range 0 to 200 mS/cm
3.1 SYSTEM SPECIFICATIONS (absolute EC up to 400 mS/cm)
TEMPERATURE Resolution
Manual 1 µS/cm; 0.001 mS/cm; 0.01 mS/cm; 0.1 mS/cm; 1 mS/cm
Range -5.00 to 55.00 °C;
Automatic 1 µS/cm from 0 to 9999 µS/cm
23.00 to 131.00 °F;
0.01 mS/cm from 10.00 to 99.99 mS/cm
268.15 to 328.15 K
0.1 mS/cm from 100.0 to 400.0 mS/cm
Resolution 0.01 °C; 0.01 °F; 0.01 K Automatic (mS/cm) 0.001 mS/cm from 0.000 to 9.999 mS/cm
Accuracy ± 0.15 °C; ± 0.27 °F; ±0.15 K 0.01 mS/cm from 10.00 to 99.99 mS/cm
Calibration Automatic at 1 custom point 0.1 mS/cm from 100.0 to 400.0 mS/cm
Accuracy ±1 % of reading or ±1 µS/cm whichever is greater
pH/mV
Calibration Automatic single point, with 6 standard solutions
Range 0.00 to 14.00 pH; ± 600.0 mV
(84 µS/cm, 1413 µS/cm, 5.00 mS/cm, 12.88 mS/cm,
Resolution 0.01 pH; 0.1 mV 80.0 mS/cm, 111.8 mS/cm) or custom point
Accuracy ± 0.02 pH; ± 0.5 mV
Calibration Automatic 1, 2 or 3 points with automatic recognition RESISTIVITY (HI 98194, HI 98195 only)
of 5 standard buffers (pH 4.01, 6.86, 7.01, 9.18, 10.01)
and 1 custom buffer Range 0 to 999999 Ω·cm;
(depending on measurement setup) 0 to 1000.0 kΩ·cm;
ORP 0 to 1.0000 MΩ·cm
Range ± 2000.0 mV Resolution Depending on resistivity reading
Resolution 0.1 mV Calibration Based on conductivity or salinity calibration
Accuracy ± 1.0 mV
Calibration Automatic at 1 custom point (relative mV) TDS (Total Dissolved Solids) (HI 98194, HI 98195 only)
DISSOLVED OXYGEN (HI 98194, HI 98196 only) Range 0 to 400000 ppm (mg/L);
(the maximum value depends on the TDS factor)
Range 0.0 to 500.0 %
0.00 to 50.00 ppm (mg/L) Resolution
Manual 1 ppm (mg/L); 0.001 ppt (g/L);
Resolution 0.1 %
0.01 ppt (g/L); 0.1 ppt (g/L); 1 ppt (g/L)
0.01 ppm (mg/L)
Automatic 1 ppm (mg/L) from 0 to 9999 ppm (mg/L)
Accuracy 0.0 to 300.0 %: ± 1.5 % of reading 0.01 ppt (g/L) from 10.00 to 99.99 ppt (g/L)
or ± 1.0 % whichever is greater; 0.1 ppt (g/L) from 100.0 to 400.0 ppt (g/L)
300.0 to 500.0 %: ± 3 % of reading Automatic ppt (g/L) 0.001 ppt (g/L) from 0.000 to 9.999 ppt (g/L)
0.00 to 30.00 ppm (mg/L): ± 1.5 % of reading 0.01 ppt (g/L) from 10.00 to 99.99 ppt (g/L)
or ±0.10 ppm (mg/L) whichever is greater; 0.1 ppt (g/L) from 100.0 to 400.0 ppt (g/L)
30.00 ppm (mg/L) to 50.00 ppm (mg/L): ± 3 % of reading
Accuracy ±1 % of reading or ±1 ppm (mg/L) whichever is greater
Calibration Automatic 1 or 2 points at 0, 100 % or 1 custom point
Calibration Based on conductivity or salinity calibration
12 13

Downloaded from www.Manualslib.com manuals search engine


SALINITY (HI 98194, HI 98195 only) METER BATTERY LIFE
Range 0.00 to 70.00 PSU The power consumption of the HI 9819X multiparameter systems are dependent
Resolution 0.01 PSU on two things:
Accuracy ±2% of reading or ±0.01 PSU whichever is greater 1. The measurement system configuration (sensor configuration)
Calibration Based on conductivity calibration 2. The meter configuration (logging interval and backlight use)
The following table estimates the meter’s battery life connected to a HI 769819X
SEAWATER SIGMA (HI 98194, HI 98195 only) probe with backlight off. (Note: backlighting use consume the most power).
Range 0.0 to 50.0 σt, σ0, σ15 The table variables, battery selection and parameter selection.
Resolution 0.1 σt, σ0, σ15
Accuracy ± 1σt, σ0, σ15 pH/ ORP, DO, EC
Backlight OFF, 1 s log 280 hours
Calibration Based on conductivity or salinity calibration
Backlight OFF, 4 min log 360 hours
Backlight OFF, 10 min log 400 hours
ATMOSPHERIC PRESSURE
Backlight ON, 4 min log 50 minutes
Range 450 to 850 mm Hg; 17.72 to 33.46 in Hg; Backlight ON, 10 min log 50 minutes
600.0 to 1133.2 mbar; 8.702 to 16.436 psi;
0.5921 to 1.1184 atm; 60.00 to 113.32 kPa
Resolution 0.1 mm Hg; 0.01 in Hg; 0.1 mbar
0.001 psi; 0.0001 atm; 0.01 kPa
Accuracy ±3 mm Hg within ±15°C from calibration temperature
Calibration Automatic at 1 custom point

METER SPECIFICATIONS
Temperature Compensation Automatic from -5 to 55 °C (23 to 131 °F)
Logging Memory 45,000 records
(continuous logging or log-on-demand of all parameters)
Logging Interval 1 second to 3 hours
PC Interface USB (with HI 9298194 software)
Waterproof Protection IP67
Environment 0 to 50 °C (32 to 122 °F); RH 100 %
Battery Type 4 x 1.5 V, AA alkaline batteries
Battery Life 360 hours without backlight / 50 hours with backlight
Dimensions/Weight
221 x 115 x 55 mm (8.7 x 4.5 x 2.2”) / 750 g (26.5 oz.)

14 15

Downloaded from www.Manualslib.com manuals search engine


3.2 PROBE SPECIFICATIONS 3.3 SENSOR SPECIFICATIONS
HI 769828-0 HI 769828-1 HI 769828-2 HI 769828-3
Sensor Inputs 3 for HI 7698194 Description pH pH/ORP Dissolved Oxygen EC
2 for HI 7698195 and HI 7698196 Measure Type
Primary Unit pH, mV (pH) pH, mV (pH/ORP) D.O. (% sat. & conc.) EC
Sample Environment Fresh, brackish, seawater
Measure Range 0.00 to 13.00 pH 0.00 to 13.00 pH 0.0 to 500.0 % 0.0 to 200.0 mS/cm
Waterproof protection IP68 ±600.0 mV ±600.0 mV 0.00 to 50.00 mg/L 0.0 to 400 mS/cm
±2000.0 mV (absolute)
Operating Temperature -5 to 55° C *
Temperature Range -5 to 55°C -5 to 55°C -5 to 55°C -5 to 55°C
Storage Temperature -20 to 70° C
Color Code Red Red White Blue
Maximum Depth 20 m (66 ft.) * Materials Tip: glass (pH) Tip: glass (pH); Pt (ORP) Cat/An: Ag/Zn Stainless steel
Dimensions 342mm (13.5”), Junction: ceramic Junction: ceramic Membrane: HDPE electrodes AISI 316
Body: PEI Body: PEI Body: white top ABS Body: ABS/EPOXY
(without cable) dia=46 mm (1.8”) Electrolyte: gel Electrolyte: gel CAP
Reference: double Reference: double
Weight 570g (20.1 oz.)
Maintenance HI 70300 HI 70300 HI 7042S none
(with batteries and sensors)
Solution (storage solution) (storage solution) (D.O. electrolyte)
Cable Specification Multistrand-multiconductor shielded cable with internal strength Dimensions 118 x 15 mm 118 x 15 mm 99 x 17 mm 111 x 17 mm
member rated for 68 kg (150 lb) intermittent use
Depth 20 m (65’) 20 m (65’) 20 m (65’) 20 m (65’)
Wetted Materials Body: ABS
Threads: Nylon
Shield: ABS/ 316 SS
Temp probe: 316 SS
O-rings: EPDM

16 17

Downloaded from www.Manualslib.com manuals search engine


Chapter 4 - PROBE INSTALLATION 4.2 SENSOR PREPARATION / ACTIVATION

4.1 SENSOR DESCRIPTIONS 4.2.1 pH Preparation


HI 769828-0 Combination pH sensor features a glass pH Remove the shipping cap from the pH sensor. If the shipping cap does not contain
sensitive bulb and a silver/silver chloride double junction any liquid, pour HI 70300 into shipping cap, place it back on the sensor and
reference with gelled electrolyte. soak for at least 1/2 hour before use. If HI 70300 is not available, pH 4.01
buffer may be substituted.
HI 769828-1 Combination pH/ORP sensor features a glass
sensitive bulb for pH readings, a platinum sensor for redox 4.2.2 ORP Activation
measurements and a silver/silver chloride double junction For improved redox measurements, the surface of the sensor must be clean and
reference with gelled electrolyte. smooth. A pretreatment procedure should be performed to ensure quick response.
Note See section 4.2.1 for pH preparation. The pretreatment of the sensor is determined by the pH and the ORP potential
values of the sample. Use the table below to determine the treatment required.
See section 4.2.2 for ORP activation.
First locate the typical sample pH. If the corresponding ORP value (mV) is higher
HI 769828-2 Galvanic dissolved oxygen (D.O.) sensor. The than the values in the table below, an oxidizing pretreatment is necessary. If the
thin gas permeable membrane isolates the sensor elements value is lower, a reducing pretreatment is necessary.
from the testing solution but allows oxygen to pass through.
The oxygen that passes through the membrane is reduced
at the cathode and causes a current, from which the oxygen pH mV pH mV pH mV pH mV pH mV
concentration is determined. The D.O. sensor conforms to
Standard Methods 4500-AG, EPA 360.1. 0 990 1 920 2 860 3 800 4 740

Note The D.O. sensor needs to be activated before instal- 5 680 6 640 7 580 8 520 9 460
lation. See section 4.2.3 for details. 10 400 11 340 12 280 13 220 14 160
HI 769828-3 4 ring-electrode conductivity (EC/TDS/Resistivity/
For reducing pretreatment: immerse the electrode for at least five minutes in
Salinity) sensor. The sensor is immune to polarization or surface
HI 7091.
coatings.
For oxidizing pretreatment: immerse the electrode for at least five minutes in
HI 7092.

18 19

Downloaded from www.Manualslib.com manuals search engine


4.2.3 D.O. Sensor Activation The HI 7698195 support two different sensors:
The D.O. probe is shipped dry. To prepare the sensor for use: Connector 1: pH, pH/ORP
• Remove the black & red plastic cap. This cap is used for shipping purposes Connector 2: EC
only and can be thrown away.
• Insert the supplied O-ring in to the membrane cap.
• Rinse the membrane with some electrolyte solution. Refill with clean electrolyte.
Gently tap the membrane cap to dislodge air bubbles. To avoid damaging the
membrane, do not touch it with your fingers or directly tap the membrane.
• With the sensor facing down screw the membrane cap counterclockwise to the
end of the threads. Some electrolyte will overflow.
• Rinse outside of sensor with deionized water.
• Invert sensor and inspect. There should be no bubbles or debris between the
membrane and sensor body.
4.2.4 EC Sensor Preparation
The EC sensor does not need to be soaked or hydrated before use. Use the small
brush included in the probe maintenance kit to clean and loosen any debris
before using.
4.3 SENSOR INSTALLATION
The HI 7698194 can support three different
sensors: Connector 1: pH, pH/ORP
Connector 2: D.O.
Connector 3: EC.
To make installation easier, the sensors have color- The HI 7698196 support two different sensors:
coded caps and the sockets are identified with Connector 1: pH, pH/ORP
colored triangles, corresponding to the colors of Connector 2: DO
the sensors (pH - red; EC - blue; DO - white).

20 21

Downloaded from www.Manualslib.com manuals search engine


For a correct installation: Chapter 5 - INITIALIZATION AND MEASUREMENT
• Grease the sensor O-ring with the lubricant found in the probe maintenance 5.1 BATTERY INSTALLATION
kit. DO NOT SUBSTITUTE other grease/lubricants as it may cause the O-ring
to swell. HI 9819X are supplied with 4 alkaline, size AA batteries.
The battery symbol on the LCD indicates the remaining battery life. The meter
• Insert the sensor into the correctly color coded opening while positioning the has a low battery warning, and when the symbol starts blinking, batteries should
connector key toward the center of the probe. Make sure the connector is be replaced with new ones. When the batteries are discharged the meter will
seated correctly (the sensor will no longer move freely) before tightening the automatically shut off to avoid erroneous readings.
locking threads with your fingers.
Replace batteries in nonhazardous areas only.
• Continue to tighten the locking threads with the tool supplied in the maintenance Remove the 4 screws on the rear of the instrument and insert the batteries
kit until the sensor is secured tightly against the probe body. observing polarity.
• To protect the sensors, screw the protective shield onto the probe body. Note: Do not mix old and new alkaline batteries.

• With the meter off, connect the probe to the DIN socket on the bottom of the
meter. Align the pins and key then push the plug into the socket. Secure the
probe by attaching the carabinier of the probe to the meter piston.
• Turn on the meter by pressing the ON/OFF key. The meter should automatically
recognize the installed sensors and identify them on the probe status screen. If
you have an error message or the sensor is not recognized, reconnect the
sensor(s) or probe and try again.

22 23

Downloaded from www.Manualslib.com manuals search engine


5.2 METER INITIALIZATION Two active soft keys are found at the bottom of the status screen.
After connecting the desired sensors to the probe and connecting the probe to • Press Measure to access the measurement mode.
the meter (see previous chapter), turn the meter on by pressing ON/OFF. • Press Param to access the “Select Parameter” menu. (This screen can also be
After the initialization has been completed if the probe is connected, the meter accessed from the main menu, see Chapter 6 for a detailed description.).
displays the PROBE STATUS SCREEN. The probe status screen identifies the • Press the DOWN arrow to view additional information about the probe.
probe and attached sensors. If an incompatible probe was connected the
following message appears and the probe has to be replaced. The compatible 5.3 MEASUREMENT MODE
meter-probe pairs are: HI 98194 - HI 7698194; HI 98195 - HI 7698195;
HI 98196 - HI 7698196. Measurement mode is one of the three main operating modes of HI 9819X
(along with logging mode and setup mode).
If an incompatible sensor was connected to the probe, “Incompat.sensor!” will
be displayed according to the respective connector. If the sensor was placed in During measurement mode HI 9819X will simultaneously measure data for all
a wrong connector, “Wrong input” will be displayed for that connector. enabled parameters.
• Use the numbers on the keyboard to select the number of parameters that are
shown on the screen at one time. The display will automatically resize the font.

• Press the [up] and [down] arrows to scroll through the enabled parameters if
they do not fit on one screen.
Note A flashing measurement value indicates that the measurement is out of
range.
A flashing measurement unit indicates that the user calibration has not
been done and is needed for accurate readings.
• Press Log to enter the log menu. See Chapter 11 for details.
• Press Menu to enter the main setup menu. The main menu accesses the pa-
rameter setup, calibration, system setup and status options. See the follow-
ing chapters for details.

24 25

Downloaded from www.Manualslib.com manuals search engine


5.4 SETUP MENU STRUCTURE Chapter 6 - PARAMETER SETUP MENU
From the main menu, use the arrow keys to
highlight Parameter Setup and then press Select.
The following options will be displayed:

6.1 SELECT PARAMETERS


Use the ARROW keys to scroll through the menu.
Press the right softkey to enable or disable a single
parameter, or the left softkey to enable or disable
all parameters. A checked box means that the
parameter is enabled.
Only the available parameters are present in the
list.
Note If the password protection is enabled, you will be required to enter
the password before any parameters can be modified.

6.2 PARAMETER UNITS

6.2.1 Temperature Unit


The user can select the measurement unit: °C, °F
or K. The default value is °C.

6.2.2 TDS Unit (HI 98194, HI 98195 only)


The user can select ppm - ppt or mg/L - g/L
measurement unit. The default value is ppm - ppt.

6.2.3 DO Concentration Unit (HI 98194)


The user can select ppm or mg/L. Dissolved Oxygen
concentration is calculated using % saturation,
conductivity and atmospheric pressure. The default
value is ppm.

6.2.4 Pressure Unit


The user can select one the following measurement
units: psi, mmHg, inHg, mbar, atm, kPA. The
default value is psi.

26 27

Downloaded from www.Manualslib.com manuals search engine


6.2.5 Resistivity Unit (HI 98194, HI 98195 only) 6.3 PARAMETER COEFFICIENTS
The user can select resistivity from one of the 6.3.1 EC Reference Temperature (HI 98194,
following measurement units: Ω·cm, kΩ·cm or HI 98195 only)
MΩ·cm. Resistivity is calculated from the conductivity
This value is used for temperature compensated
measurement. The default unit is MΩ·cm.
conductivity. All EC measurements will be referenced
to the conductivity of a sample at this temperature.
6.2.6 Seawater Sigma Unit (HI 98194, HI 98195 only)
Press the softkey to select the desired option; 20 °C or at 25 °C. The default value
This parameter is used for seawater analysis. It is is 25 °C.
calculated from the conductivity measurement and
depends on water pressure, temperature and salinity. 6.3.2 EC Temperature Coefficient (HI 98194,
The default value is σt. HI 98195 only)
Users can select the reference temperature: σt, σ0 The temperature coefficient Beta (β) is defined by
and σ15 (i.e. current temperature, 0 °C or 15 °C). the following equation (using 25 °C as an example):
EC25=ECx/(1+β(Tx-25))
6.2.7 EC Resolution (HI 98194, HI 98195 only)
Beta is a function of the solution being measured. For freshwater samples Beta is
The user can configure the conductivity resolution
approximately 1.90%/°C. If the actual temperature coefficient of your sample is
with one of the following options:
known, press Modify to enter the value. To confirm press Accept. The value can
Auto: the meter automatically chooses the range be within 0.00 and 6.00%/°C. The default value
to optimize the measurement. Readings can be in is 1.90%/°C.
µS/cm or mS/cm.
Auto mS/cm: the meter automatically chooses 6.3.3 TDS Factor (HI 98194, HI 98195 only)
the range to optimize the measurement, readings will be in mS/cm only. TDS stands for total dissolved solids, and it is a
1µS/cm, 0.001 mS/cm, 0.01mS/cm, 0.1mS/cm or 1mS/cm: the meter will not calculated value based on the conductivity of the
autorange, the measurement will be displayed with the selected resolution. The solution (TDS = factor x EC25). The TDS conversion factor can be set from 0.00 to
default value is Auto. 1.00. A typical TDS factor for strong ionic solutions is 0.5, while for weak ionic
solutions (e.g. fertilizers) is 0.7. Press Modify to enter the value, press Accept to
6.2.8 Absolute EC Resolution (HI 98194, HI 98195 only) confirm. The default value is 0.50.
Absolute conductivity displays the conductivity without temperature compensation.
6.4 Averaging
See 6.2.8 EC resolution for resolution details.
Averaging is a software filter to minimize sensor
Note A small letter “A” added to the µS/cm or mS/cm unit refers to an absolute
noise and provide more stable readings.
conductivity value (i.e. a conductivity reading with no temperature
Averaging is particularly useful to get a
compensation).
representative reading of the “average” value from
6.2.9 TDS Resolution (HI 98194, HI 98195 only) flowing water. Averaging will affect all
The user can configure the TDS resolution with one of the following options: measurements. This value should be kept low if
Auto: the meter automatically chooses the range to optimize the measurement, you want a fast response. Press Modify to select
readings can be in ppt or ppm. the desired number of samples to average. This
Auto ppt: the meter automatically chooses the range to optimize the measurement, value can be set from 1 to 20 samples. The
readings will be in ppt only. default value is 1.
1 ppm, 0.001 ppt, 0.01 ppt, 0.1 ppt or 1 ppt: the meter will display the Note Each reading takes 1 second, so when logging the first sample will be
measurement with selected resolution. The default value is Auto. delayed by a few seconds if averaging is used.
28 29

Downloaded from www.Manualslib.com manuals search engine


Chapter 7 - CALIBRATION MODE 7.1 QUICK CALIBRATION
HI 9819X’s calibration routines are accessed by highlighting “Calibration” The quick calibration method provides a quick single point calibration for pH,
and pressing Select from the main menu. Calibration is the process that conductivity and dissolved oxygen sensors. HI 9828-25 calibration solution is
standardizes the electrical signal from the sensors to reagent standards of used for both pH and conductivity.
known value. • Fill the calibration beaker 2/3 full with HI 9828-25 calibration solution.
Calibrations are intuitive and menu driven. All • Slowly place the sensors into the solution and
calibration data is stored in the non volatile probe dislodge bubbles that may adhere to the sensors.
memory, allowing probes to be connected to • Screw the calibration beaker completely on the
different meters without recalibration. probe body. Some solution may overflow.
There are two types of calibrations available: the • Wait a few minutes for the system to stabilize.
“Quick calibration”, which is used for a single • From the “Calibration” menu select “Quick
point calibration of pH, Conductivity, and/or Dissolved Oxygen and is handy for calibration”.
field work; and the “Single param. calibration” that allows each parameter
• A three item calibration menu will appear (pH,
to be calibrated individually. The user may also restore each parameter to a
Conductivity and Dissolved oxygen) and “pH”
factory default calibration.
will start to blink along with the “Not ready”
Note The password will be required if password protection is enabled. message.
To optimize measurements, it is advisable to establish the optimum calibration • When the pH signal is stable, the “Ready”
period required for the measurement environment. message appears. Press Confirm to store the
Calibration requirements vary with deployment conditions, for example very turbid calibration data.
biologically-active waters may require more frequent cleanings and calibrations • The “Storing” message will appear as the
than cleaner waters. calibration proceeds to the next sensor. A
General calibration guidelines are listed below: checkmark will appear in the box next to “pH”
• Set up a routine service schedule where measurement integrity is validated. to indicate a successful calibration.
This is especially important for new installation sites or long deployments. Note To bypass any of the calibrations press
• Inspect sensor connectors for corrosion and replace damaged sensors. Skip to move to the next sensor in the
• Inspect sensor o-rings for damage and if necessary replace and lubricate with quick calibration menu.
the grease found in the probe maintenance kit. If the pH sensor is not installed the
• Do not handle the sensing surfaces of the sensors. message “pH sensor not installed! Skip
• Avoid rough handling and abrasive environments that can scratch the reactive to conductivity calibration” will appear.
surfaces of the sensors.
• Following the pH calibration, “Conductivity” will start to blink along with the
• Avoid long-term exposure of sensors to bright sunlight. If possible, calibrate in
“Not ready” message.
a shaded area.
• When the measurement is stable, “Ready” appears. Press Confirm to store
• Discard standards after use. Do not return the used standards to the bottles of
the calibration data and the “Storing” message will appear.
“fresh” solution.
• For measurements across a temperature gradient (when water temperature is Note If EC calibration is not required, skip to
drastically different from the standards), permit the sensors to reach thermal the D.O. quick calibration by pressing the
equilibrium before conducting calibrations or making measurements. The heat Skip softkey.
capacity of the probe is much greater than the air and the small beakers of • The message “Empty the beaker.” will appear.
calibration standards.

30 31

Downloaded from www.Manualslib.com manuals search engine


• Unscrew the calibration beaker and empty the solution. 7.2.2 Procedure
• Shake any remaining liquid off the probe and beaker. No droplets should remain The measured pH value is displayed, along with the temperature and the buffer
on the D.O. sensor membrane. value on the second level.
Note Do not attempt to dry wipe the D.O. sensor as damage to the membrane If necessary, press the Cal point softkey and use the arrow keys to select the
may occur. correct buffer.
• Screw the empty calibration beaker on the probe body. The beaker should not • Immerse the sensors in the first buffer rinse
be dry. solution and stir gently.
• Press Accept to close the displayed message. • Immerse the pH sensor and temperature
• When the measurement is stable, “Ready” probe into the selected buffer and stir gently.
appears. Press Confirm to store the calibration The temperature, pH buffer value and the “Not
data and the “Storing” message will appear. ready” message are displayed.
• Press OK to return to “Calibration” menu.
• Once the reading has stabilized the countdown
Note To quit the quick calibration procedure, timer will count down until the display shows the
press ESC at any time. “Ready” message.
After every calibration the quick calibration window will show a check mark in • Press Confirm to accept the calibration point.
the box next to the calibrated parameter. • After the calibration point is confirmed, to avoid
cross-contamination immerse the sensors in the next calibration buffer rinse
7.2 pH CALIBRATION solution and stir gently.
To optimize the pH measurement follow the general guidelines mentioned in the • Press Cal Point to select the next buffer (if necessary), and repeat the calibration
Chapter 7 introduction. procedure outlined above with the second and third buffers.
From the “Calibration” menu select “Single param. Note The calibration procedure can be terminated after a single or 2 point
calibration” and then “pH calibration”. The display calibration by pressing ESC. The message “Storing” followed by “Calibration
shows two options: “Calibrate pH” and “Restore completed” will be displayed.
factory calib.”.
• Press OK to return to the Calibration menu.
If a new pH sensor has been installed use “Restore
factory calib.” before performing a user calibration • Press Measure to return to the measurement screen.
as some warning messages are based on changes from previous calibrations. Custom buffer calibration
If “Restore Factory Calib” is selected, all user calibration data will be deleted and
• The HI 9819X permits a single custom buffer to
the default calibration is restored. A user calibration should follow immediately.
be used for pH calibration. This can be used
If “Calibrate pH” is selected, the user can perform a new calibration using up to
along with standard buffers as part of a 2 or 3
3 buffers (pH 4.01, 6.86, 7.01, 9.18, 10.01 or one custom buffer).
point calibration or as a single point.
When a 3-point calibration is performed, all old data are overwritten, while with
• To select this option first press Cal. point and
a single or 2-point calibration the meter will also use information from the previous
then Custom while the meter is waiting for stable
calibration, if exists.
reading.
7.2.1 Preparation • A text box window will appear. Use the keypad to enter the value of the buffer at
Pour small quantities of the selected buffer solutions into clean beakers. To the current temperature. The valid range for custom a buffer is from 0.00 to
minimize cross contamination, use two beakers for each buffer solution: the first 14.00 pH.
one for rinsing the sensor and the second one for calibration.
32 33

Downloaded from www.Manualslib.com manuals search engine


7.2.3 pH Calibration Error Messages 7.3 Relative mV CALIBRATION
The HI 9819X displays a series of messages if an error has occurred during The “ORP calibration” allows the user to perform a single point custom calibration
calibration. (relative mV) or to restore the factory calibration.
If the meter does not accept a pH calibration point, a short message is displayed The Oxidation-Reduction Potential (ORP), displayed in mV, is the voltage that
to indicate the possible error source. The following screens are examples: results from the difference in potential between the platinum ORP sensor and the
silver/silver chloride reference electrode. ORP values are not temperature
compensated, although ORP values can change with temperature (e.g. reference
electrode potential changes, sample equilibrium changes). It is important to report
ORP values together with the reference electrode used and the temperature.
The inert platinum ORP surface provides an electron exchange site with the sample
(or standard) and its surface. The electron exchange is typically very fast in
well-poised solutions (standards for example), but may be more lengthy in natural
water samples.
Calibration is typically not required for a new ORP sensor, but the process does
establish a baseline that can be used as a comparison for future validations.
Calibration is used to compensate for changes due to contamination of the
platinum surface and drift in the reference electrode.
A relative mV calibration can also be made to remove the voltage attributable to
the Ag/AgCl reference electrode (to display the ORP versus a SHE (standard
These are the available messages:
hydrogen electrode). This is really an arithmetic correction and is correct only at
• “Input out of scale”: the pH value is out of range. The pH sensor may require the standard temperature. For example, HI 7022L reads 470 mV at 20 °C
replacement. versus the Ag/AgCl reference. The ORP mV versus a SHE would be 675 mV.
• “Check sensor”: the electrode may be broken, very dirty or the user has (add 205 mV to the observed value).
attempted to calibrate the same buffer value twice.
• “Wrong buffer”: the displayed pH reading is too far from the selected buffer
value. This is often seen immediately after a buffer calibration has been completed
but before the pH sensor has been moved to the next buffer. Check if the
correct calibration buffer has been selected.
• “Invalid temperature”: the buffer temperature is outside the acceptable range.
• “Wrong buffer” / “Contaminated buffer” / “Check electrode”: the buffer is
contaminated or the sensor is broken or very dirty.
• “Check sensor” / “Clean sensor”: the electrode is broken or very dirty.
• “Wrong” / “Clear old calibration”: erroneous slope condition. These messages
appear if the slope difference between the current and previous calibration
exceeds the slope window (80% to 110%). Press the Clear softkey to cancel
the old data and continue the calibration procedure, or press ESC to quit the
pH calibration mode.

34 35

Downloaded from www.Manualslib.com manuals search engine


7.3.1 Preparation If the % D.O. saturation range is calibrated, the D.O. concentration range will
Appendix D – ACCESSORIES lists Hanna solutions used for ORP calibrations. also be calibrated, and vice versa.
The calibration should be conducted at temperatures between 20-26 °C. The Dissolved oxygen concentration values are based
sensor should be clean and oil free. on % D.O. saturation, temperature, salinity and
atmospheric pressure. A standard solution or a
7.3.2 Procedure
reference D.O. meter may be used to compare
• From the “Calibration” menu select “Single readings during calibration.
param. calibration” and then “ORP calibration”. The calibration of the D.O. concentration range
The display shows two options: “Custom ORP” can only be performed at a single custom point (4 to 50 mg/L). It is recommended
and “Restore factory calib.”. to calibrate the D.O. sensor close to the values that will be measured.
• For a user calibration select “Custom ORP”. Choose “DO calibration” from the “Calibration” menu, select the D.O. calibration
• Fill a beaker with an ORP test solution (see type using the arrow keys and press Select to confirm.
APPENDIX D “Accessories”). % D.O. saturation
• Using the keypad, insert the numerical ORP
The calibration of the % D.O. saturation range can be performed at a single or
value and then press Accept to confirm.
2 standard points (0 % and 100 %), or at a single custom point (50 % to 500 %).
• The stability counter will count down and the
Procedure:
message “Ready ” and Confirm will be
displayed. • To calibrate at 100 %, fill the calibration beaker
with approximately 4 mm (5/32") of water and
• Press Confirm to accept the calibration point.
screw it onto the probe. The membrane should
• After confirmation, the following messages are displayed: “Storing” and
not be wet. This condition corresponds to air
“Calibration completed”.
100 % saturated with oxygen and water vapor.
• Press OK to return to the Calibration menu. • The reading, temperature, calibration point and
• Press Measure to return to the measurement screen. the “Not ready” message are displayed.
• To restore the factory calibration data, select the corresponding option in the • Once the reading has stabilized the countdown
“ORP calibration” menu and then press Select. timer will count down until the display shows
the “Ready” message.
7.4 DISSOLVED OXYGEN CALIBRATION (HI 98194, HI 98196 only)
• Press Confirm to accept the calibration point.
The accuracy of dissolved oxygen measurements is directly related to membrane
After confirmation, put the D.O. and temperature sensors into HI 7040L zero
cleanliness and calibration technique. Oily coating and biological contaminants
oxygen solution and wait for stability to be reached. The stability timer will count
are the primary cause of calibration drift in dissolved oxygen sensors. Unfortunately,
down and Confirm will appear. Press Confirm to store the calibration.
brushes or other cleaning objects may damage the membrane. Replacing the
membrane cap and electrolyte is the best way to perform periodic maintenance. • The following messages will appear: “Storing” and “Calibration completed”.
• Press OK to return to the “Calibration” menu.
Although it may be easier to calibrate the D.O. sensor prior to deployment, it is
• Press ESC twice to return to the main menu.
advised to calibrate at the site of deployment. Errors in measurement may result
if altitude and barometric pressure differ between the calibration and measurement • Press Measure to return to the measurement screen.
site. This is very important for autonomously logging probes. Note The user can perform a single point calibration by pressing ESC after
Note Perform either the % D.O. Saturation or D.O. Concentration calibration. the first point is accepted.
Note If the D.O. input is not within the acceptable range, the message “Invalid
input” is displayed.

36 37

Downloaded from www.Manualslib.com manuals search engine


Single point Custom % saturation calibration Note For a correct conductivity calibration, the probe shield or the calibration
• For a calibration at another known value place sensor and temperature probe beaker must be used.
into the known solution and change the calibration value, press the Cal. point
The conductivity calibration menu includes
softkey and select the desired point.
3 different types of calibration: Conductivity,
• To insert a different calibration value, press Cal. point and then Custom.
Absolute conductivity and Salinity.
Insert the desired value using the keypad, then press Accept.
The “Conductivity” option allows a single point
• When the reading is stable, the “Ready” message is displayed. Press Confirm
calibration with a standard solution selectable by
to store the calibration point.
the user. This calibration is temperature
• The following messages will appear: “Storing” and “Calibration completed”.
compensated.
• Press OK to return to the “Calibration” menu.
• Press ESC twice to return to the main menu. The “Absolute conductivity” option allows a single point calibration with a
• Press Measure to return to the measurement screen. conductivity solution of known non-temperature compensated value at the current
temperature.
D.O. concentration (only HI 98194)
The “Salinity” option allows calibration with a standard salinity solution.
Verify the barometric pressure, conductivity and
The 3 calibrations are related, so that each one will calibrate all 3 measurements.
temperature reading are correct. Calibrate them
if necessary. To calibrate the D.O. concentration Note To improve accuracy, choose a calibration standard near the sample
range, a solution with known Dissolved Oxygen conductivity.
concentration value is needed. The solutions used Choose “Conductivity calibration” from the “Calibration” menu, select the
to calibrate with should be determined calibration type using the arrow keys and press Select to confirm.
independently (for instance by Winkler titration). Place the D.O. sensor with
temperature sensor into the known solution. Conductivity
• From the “DO calibration” menu, select the “DO concentration” option, insert • Select the “Conductivity” option and press Select to confirm.
the known concentration. Allow the sensors to reach thermal equilibrium with • Fill the calibration beaker with a conductivity standard (see APPENDIX D -
the solution. Stir or agitate if possible to keep fresh solution in front of the “Accessories” for choosing the proper HANNA standard solution).
membrane and press OK. • Pour additional standard into a second beaker to be used to rinse the sensor.
• When the reading is stable, the stability timer will count down and <Confirm> • Immerse the sensor into the rinse standard by raising and lowering the beaker
will appear. Press Confirm to accept the value. a few times to ensure that the EC sensor channels are filled with fresh standard.
• When the messages “Storing” and “Calibration completed” appear, the • Place the calibration beaker over the EC sensor and dislodge any trapped
calibration is completed. To return to the “Calibration” menu, press OK. bubbles. Screw the beaker into place. Wait for the reading to stabilize.
• To return to the main menu, press ESC twice.
7.5 CONDUCTIVITY CALIBRATION (HI 98194, HI 98195 only)
A conductivity calibration is used to adjust for variations in cell factors by using
a standard solution of known conductivity. Oily coating and biological
contaminants are the primary cause of calibration drift in conductivity sensors.
This type of fouling changes the apparent cell geometry, resulting in a shift in cell • The main display shows the actual reading, while the secondary level displays
constant. Before performing a conductivity calibration inspect the EC sensor for the current temperature and the standard value.
debris or blockages. The EC electrodes are situated inside the two small channels
• To change the standard value, press Cal. point and the list of available standard
found in the bottom of the conductivity sensor. Clean using the small brush from
values is displayed: 0 µS/cm, 84 µS/cm, 1413 µS/cm, 5.00 mS/cm, 12.88
the probe maintenance kit. Flush with water. A mild detergent may be used to
mS/cm, 80.0 mS/cm and 111.8 mS/cm.
remove oily coatings. Always flush with clean water after cleaning.
38 39

Downloaded from www.Manualslib.com manuals search engine


• The third level displays the status message. Salinity
• Press Custom to insert a custom value (temperature compensated value). The measurement of salinity is based on the Practical Salinity Scale which uses
Insert the desired value using the keypad, then press Accept. the EC measurement. If the user has a standard with known PSU value it may be
• When the reading becomes stable, the stability timer will count down and used to calibrate the conductivity sensor.
Confirm will appear. Press Confirm to save the calibration. • Select “Salinity” from the “Conductivity calibration”
• After confirmation, the following messages are displayed: “Storing” and menu.
“Calibration completed”. • Use the keypad to enter the known salinity value
• Press OK to return to the “Calibration” menu. of the calibration solution. Press Accept to con-
• Press ESC twice to return to main menu. firm.
• Press Measure to return to the measurement screen. • Fill the calibration beaker with salinity standard of known value.
• Pour additional standard into a second beaker to be used to rinse the sensor.
Absolute Conductivity
• Immerse the sensor into the rinse beaker and raise and lower the beaker to
• Select “Absolute conductivity ” from the
ensure that the EC sensor channels are filled with fresh standard.
“Conductivity calibration” menu.
• Use the keypad to enter the custom value • Place the calibration beaker with standard over the EC sensor and dislodge
with the desired resolution. Press Accept to any trapped gas bubbles. Screw the beaker into place.
confirm. • Wait for the reading to stabilize. The stability timer will count down and Confirm will
• Fill the calibration beaker with conductivity standard with known conductivity appear.
at the temperature of standardization. • Note the temperature and adjust the salinity value if needed.
• Pour additional standard into a second beaker to be used to rinse the sensor. • Press Confirm to save the calibration.
• Immerse the sensor into the rinse beaker and raise and lower the beaker to • After confirmation, the following messages are displayed: “Storing” and
ensure that the EC sensor channels are filled with fresh standard. “Calibration completed”.
• Place the calibration beaker over the EC sensor and dislodge any trapped • Press OK to return to the “Calibration” menu.
bubbles. Screw the beaker into place. • Press ESC twice to return to the main menu.
• Wait for the reading to stabilize. The stability timer will count down and • Press Measure to return to the measurement screen.
Confirm will appear.
• Note the temperature and adjust the conductivity value if needed. Notes These procedures calibrate the slope value. To calibrate the offset, set
• Press Confirm to save the calibration. the calibration point at 0 µS/cm and repeat the procedure.
• After confirmation, the following messages are displayed: “Storing” and If the temperature input is not within
“Calibration completed”. the acceptable range (0 to 50°C), the
message “Invalid temperature” is
• Press OK to return to the “Calibration” menu.
displayed.
• Press ESC twice to return to the main menu.
If the conductivity input is not within the
• Press Measure to return to the measurement screen.
acceptable range, the message “Wrong
standard” is displayed.

40 41

Downloaded from www.Manualslib.com manuals search engine


7.6 TEMPERATURE CALIBRATION 7.7 ATMOSPHERIC PRESSURE CALIBRATION
The probe is factory calibrated for temperature readings. The user can perform Place HI 9819X in a wind-free area and choose
a single point temperature calibration or restore factory calibration. This procedure “Custom pressure” to perform a user calibration
requires a reference temperature measuring instrument. or “Restore factory calib”.
• Select “Temperature” from the “Calibration” menu. Note “Custom pressure” procedure requires
• Select “Calibrate temperature”. a reference barometer.
• Insert the probe in an isothermal bath with reference instrument and allow the Select the “Atm. pressure” from the “Calibration” menu.
probe to come to thermal equilibrum.
• Select the “Custom pressure” option.
• Use the keypad to enter the known temperature and then press Accept to
• Using the keypad, insert the numeric value that
confirm.
agrees with the reference meter and then press
• The stability timer will count down and the Accept to confirm.
message “Ready ” and “Confirm” will be
• The stability counter will count down and the
displayed.
message “Ready ” and “Confirm” will be
• Press Confirm to store the calibration point. displayed. Press Confirm to store the calibration
• After confirmation, the following messages are point.
displayed “Storing” and “Calibration completed”. • After confirmation, the following messages are displayed: “Storing” and
• Press OK to return to the “Calibration” menu. “Calibration completed”.
• Press Measure to return to the measurement • Press Measure to return to the measurement screen.
screen. • Press OK to return to the “Calibration” menu.
• To restore the factory calibration, select the • To restore the factory calibration, select “Restore factory calib.” in the “Pressure
corresponding option in the “Temperature calibration” menu and press Select.
calib.” menu and then press Select.

42 43

Downloaded from www.Manualslib.com manuals search engine


Chapter 8 - SYSTEM SETUP 8.1.3 Auto Poweroff
The Auto Poweroff function is used to save battery
From the main menu, select “System setup” and life. After the set time is elapsed, the meter will:
then “Meter setup” or “Probe setup”. 1. automatically switch off, if in normal measurement
mode. Press On/Off to switch on again.
8.1 METER SETUP
2. enter a sleeping mode, if the continuous
Note If the password protection is enabled, logging mode is selected with a logging interval of at least 30 seconds. The “Auto
you will be required to enter the password Poweroff” message and the Wake up softkey appear on the LCD; logging is not
before any settings can be modified. stopped. Press Wake up to reactivate the display.
8.1.1 Time
The meter uses a real time clock for logging. The
time and time format are set in this function.
Press Modify and set the time using the keypad.
Press Accept to save the time. When using the
12 hour format, press A or P on the keypad for Available options are: Not used (disabled), 5, 10, 15, 20, 30 or 60 minutes.
AM or PM after you set the time. Press Modify to select the desired time interval. The default value is “not used”.
Press Format to change between 12 and 24 hour 8.1.4 Key Beep
formats. The default format is 24 hours. If enabled, an acoustic signal sounds every time
a key is pressed. A checked box indicates this function
has been enabled. The default setting is disabled.
8.1.5 Error Beep
If enabled, an acoustic signal sounds every time
an incorrect key is pressed, or when an error occurs.
A checked box indicates this function has been
enabled. The default setting is disabled.
8.1.2 Date 8.1.6 Decimal Separator
The date and date format are set in this function. The user can select the type of decimal separator:
Press Modify and set the date using the keypad. Press “dot” or “comma”. Press the softkey to select the
Accept to save the date. desired option. The default setting is “dot”.
Press Format to change between the available date 8.1.7 LCD Contrast
formats: DD/MM/YYYY, MM/DD/YYYY, YYYY/ The LCD contrast can be adjusted with this function.
MM/DD, YYYY-MM-DD, MM-DD-YYYY, and Press Modify to enter this function. Use the
DD-MM-YYYY. The default format is YYYY/MM/DD. ARROW keys to change the contrast level and
press Accept to save the new value. The default
value is 8.

44 45

Downloaded from www.Manualslib.com manuals search engine


8.1.8 LCD Backlight Intensity 8.1.11 Language
The LCD backlight intensity can be adjusted with The language used in the meter user interface can
this function. Use the arrow keys to change the level be changed. The default language is English.
and press Accept to save the new value. The default Please contact your local Hanna office for currently
value is 7. available languages.
8.1.9 Meter Password 8.1.12 Restore Factory Settings
The Meter Password protects against unauthorized configuration changes and This function restores measurement settings
log data erasure. When implemented, many setting and functions cannot be to their original factory values. This includes
modified or viewed. measurement units, coefficients, other measurement
To enable the password proceed as follows: configurations and all logged data. The factory
• Highlight “Meter Password” and press Modify. calibration for the sensor channels is not affected.
• Enter the desired password in the text box and press Accept. • Select the “Restore factory settings” and press
Select.
• The meter will ask to confirm: press Yes to confirm or No to escape.
8.2 PROBE SETUP
8.2.1 Probe ID
The probe can be labeled with an identification
code: press Modify and a text box will be displayed.
Note While typing, the characters are masked with a “*” (star) symbol. Use the keypad to enter the desired alphanumeric
• The meter will require password confirmation. Retype the same password and code and then press Accept.
press Accept to confirm. A maximum of 14 characters can be used.
• The meter returns to the “Meter Setup” menu. The checkbox corresponding to
the meter password is checked.
To disable the password protection highlight “Meter Password” and press Modify,
enter the password and then press Disable. “No password” appears in the text
box. Press Accept to confirm.
8.1.10 Meter ID
The Meter ID may be used to uniquely identify a
meter/operator. Press Modify and a text box appears.
Use the keypad to insert the desired alphanumeric
ID and press Accept to store the identification. A
maximum of 14 characters can be used.

46 47

Downloaded from www.Manualslib.com manuals search engine


Chapter 9 - STATUS 9.3 GLP DATA
GLP (Good Laboratory Practice) is a set of functions that allows the user to store
Useful information regarding the meter, probe (if or recall data regarding the probe calibration. This feature also allows the user
connected) and GLP calibration data are available to associate readings with specific calibrations.
for viewing by selecting “Status” from the main To view GLP data select “GLP” from the “Status”
menu. menu. The complete list of available parameters
9.1 METER STATUS appears. Select the desired parameter to view the
stored GLP information.
Select “Meter Status” to display information related
to the battery, logging, internal temperature, Note If no calibration data is available for the
password, Meter ID, serial number and firmware selected parameter, the display shows the
version. Press [up] and [down] to scroll through message “No GLP data available for this
the status screens. Press ESC to return to the measurement”. Press OK to return to the
“Status” menu. previous screen.
Note GLP data is stored for the last 5 calibrations.
This calibration history allows the user to detect when readings start to
change and sensors may require cleaning or replacement.
pH
• From the “GLP” menu, select the “pH” option.
9.2 PROBE STATUS • Data regarding the last pH calibration will be
displayed: offset, acidic slope, basic slope,
Select “Probe Status” to display information related buffers used, time and date of the calibration.
to the probe type, connected sensors, Probe ID,
serial number and firmware version. • Use the arrow keys to scroll through the stored
data for the last 5 calibrations.
• Press [up] and [down] to scroll through the status
screens. • Press ESC to return to the “GLP” menu.
• Press ESC to return to the “Status” menu. Note A “C” label near the buffer value indicates
a custom point, while an “H” indicates a
HANNA standard buffer value.
If a quick calibration was performed, the buffer values are replaced
with the “Quick calibration” indication.
If no pH calibration has been performed
or if calibration was cleared using the the
Note The probe status screen will automatically be displayed when the probe “Restore factory calib.” option the offset
sensor status has changed. If this occurs, the “Measurement Screen” and slope values are set to default, and
and “Parameter Selection” softkeys are available (see Section 5.2). the message “Factory calibration” is dis-
played. Press ESC to return to the previous screen.

48 49

Downloaded from www.Manualslib.com manuals search engine


ORP • Use the arrow keys to scroll through the stored data for the last 5 calibrations.
• From the “GLP” menu select the “ORP” option.
• Data regarding the last ORP calibration will be
displayed: calibration point, time and date.
• Use the arrow keys to scroll through the stored
data for the last 5 calibrations.
• Press ESC to return to the “GLP” menu.
Notes A “C” letter near the conductivity calibration indicates a custom point,
Notes If no ORP calibration has been performed or if calibration was cleared while an “H” indicates a HANNA standard value.
using the “Restore factory calib.” option the offset and slope values are If no conductivity calibration has been performed or if calibration was
set to default, and the message ”Factory calibration” is displayed. Press cleared using the “Restore factory calib.” option the offset and slope
ESC to return to the previous screen. values are set to default, and the message ”Factory calibration” is
displayed. Press ESC to return to the previous screen.
Dissolved Oxygen
• From the “GLP” menu select the “Dissolved Temperature
oxygen” option. • From the “GLP” menu select the “Temperature”
• Data regarding the last D.O. calibration will be option.
displayed: calibration points, % saturation or • Data regarding the last temperature calibration
concentration, time and date. will be displayed: calibrated point, time and date.
• Use the arrow keys to scroll throught the stored • Use the arrow keys to scroll through the stored
data for the last 5 calibrations. data for the last 5 calibrations.
Notes A “C” label near the calibration point indicates a custom point, while Notes If no user temperature calibration has been performed or if calibration
an “H” indicates a HANNA standard value. was cleared using the “Restore factory calib.” option the offset value is
When the % D.O. range is calibrated, also the D.O. concentration range set to default, and the message ”Factory calibration” is displayed. Press
is calibrated, and vice versa. ESC to return to the previous screen.
If no D.O. calibration has been performed or if calibration was cleared Atmospheric Pressure
using the “Restore factory calib.” option the offset and slope values are
set to default, and the message ”Factory calibration” is displayed. Press • From the “GLP” menu select “Atm. pressure”.
ESC to return to the previous screen. • Data regarding the last atmospheric pressure
calibration will be displayed: custom calibration
Conductivity point, time and date.
• From the “GLP” menu select the “Conductivity” • Use the arrow keys to scroll through the stored
option. data for the last 5 calibrations.
• Data regarding the last conductivity calibration Notes If no atmospheric pressure calibration has been performed or if
will be displayed: calibration point, cell constant calibration was cleared using the “Restore factory calib.” option the
value, calibration type (conductivity, absolute offset value is set to default, and the message ”Factory calibration” is
conductivity or salinity), time and date of the displayed. Press ESC to return to the previous screen.
calibration.

50 51

Downloaded from www.Manualslib.com manuals search engine


Chapter 10 - LOGGING MODE 10.1 LOGGING MENU STRUCTURE

The HI 9819X and the corresponding probe offer many logging options that From measurement mode, press Log to access the log menu.
can be combined based on user needs. The following figures describe the available
logging options.

10.2 LOGGING ON METER


• The data logged on the meter are organized
by lots. Up to 45,000 complete records can
be stored in up to 100 lots. Each lot can store
log-on-demand records and/or continuous
records with different parameter configurations.

52 53

Downloaded from www.Manualslib.com manuals search engine


10.2.1 One Sample On Meter 10.3 LOG RECALL

Use this option to log one set of enabled measurement parameters to the meter • Select “Meter log recall” to view logs that are stored
memory. on the meter. The meter will show the number
of available lots. Select “Lots” to view or delete
• If there are no lots saved on the meter, press individual lots.
New to create a new lot. Use the keypad to enter
the desired lot name and press Accept to confirm. • Use the arrow keys to select the desired lot and
Press OK to log the sample in the selected lot. then press View.

• If there are existing lots on the meter, the meter • The meter displays a summary of all data related
will suggest a lot to store the sample. Press OK to the selected lot: number of samples, memory
to use the selected lot or Options to select a space used, time and date of the first and last readings.
different lot. This will add the new sample data
to an existing lot. A new lot can also be created • Press View to display the sample details for each
by pressing New. Press OK to log the sample in point. Use the arrow keys to change the sample
the selected lot. number in the selected lot. The sample number is
shown on the bottom right corner of the display.
• On the “Remarks” window, select Yes to go to
the Remarks screen. Press No to skip this Note Details are available only for the enabled
option. If Yes is selected, select a remark from parameters.
the list, or press New to create a new remark.
• Press Info to see record information for the current
• To return to the measurement screen, press ESC. sample (time & date, remark or serial number (if
10.2.2 Continuous Meter Log available).)

• Select “Start meter log” to log the currently enabled • Press Data to return to the previous screen or
parameters at the set logging interval on the Jump to select a different sample in the same lot.
meter. When Jump is pressed, a text box appears to insert
the desired sample number.
• To set the logging interval, highlight “Start meter
log” and press Options. The log interval time • Press ESC to return to the menu.
can set from 1 second to 3 hours. Press Modify • Choose “Plot” and the meter will create a list with
and use the arrow keys and keypad to enter the all available parameters that can be plotted.
desired log interval. Press Accept to confirm.
• Use the arrow keys to select the desired parameter.
• Press Select to edit the lot, remark, or tag, see Press Select to view the graph.
section 11.4.
• To stop the meter log, enter the log menu and
select “Stop meter log”.

54 55

Downloaded from www.Manualslib.com manuals search engine


• Use the arrow keys to move the cursor in the graph Chapter 11 - PC CONNECTION MODE
and highlight a sample. The sample data are
displayed below the graph. The logged data from a probe or meter can be transferred to a PC using the
HI 9298194 Windows® compatible application software. HI 9298194 offers a
• Press ESC to return to the parameter list.
variety of features and on-line-help is available.
• Press ESC again to return to the menu. HI 9298194 allows data to be imported into most spreadsheet programs (e.g.
Note The number of lot samples that can be plotted Excel©, Lotus 1-2-3©). After the data has been imported into a spreadsheet, all
is limited by the display resolution. To view a features of the spreadsheet program can be used to analyze and graph the
complete graph download data to PC. data.
Delete all lots 11.1 SOFTWARE INSTALLATION
• From ‘’Meter log recall’’ choose “Delete all lots” • Insert the installation CD into the PC.
and the meter will display the message “Do you • The software menu window should start automatically (if it does not, navigate
want to perform the current operation?”. Press to the main CD folder and double-click “setup.exe”). Click “Install software”
Yes to delete or No to return to the previous and follow the instructions.
screen.
• To return to the “Log recall” menu, press ESC. 11.2 METER TO PC CONNECTION
• With the meter OFF, disconnect the probe.
10.4 LOG NOTES • Connect the USB cable to the meter and to a USB port on the PC.
10.4.1 Remarks • Turn the meter ON and the message “PC connected” will be displayed.
A remark can be associated with each sample. The • Run the HI 9298194 application software.
meter can store up to 20 remarks. • Press Setting button on the top of the screen
• To add a remark, select “Log notes” from the Log and select the measurement units you which
menu, and then select “Remarks”. your data to appear with.
• The display shows a list of stored remarks. • To access the meter data select the “Meter”
button on the toolbar at the top of the screen.
• Press New to create a new remark, and use the
The PC-Meter connection will be established
keypad to enter the new remark in the text box.
and a new window will be displayed with meter data: status information (soft-
• Press Delete to delete the selected remark from ware version and date, SN, ID, battery level and free memory info), as well as
the meter. If the deleted remark is used in an existing a summary of logged data lots. The lots logged on the meter can be saved to
lot, the information will be still available in the lot the PC by pressing the “Download lot” button after the desired lot is selected.
data. • Once the lot has been downloaded, all the logged samples can be viewed.
10.4.2 Delete All Remarks
• Select “Delete all remarks” to delete all remarks.
The display will show the message “Do you want
to perform the current operation?”. Press Yes to
delete or No to return to the previous screen.
Windows® is a registered Trademark of “Microsoft Co.”

56 57

Downloaded from www.Manualslib.com manuals search engine


Meter to PC data Chapter 12 - TROUBLESHOOTING / ERROR MESSAGES
a. Select parameter units
b. Select Meter from toolbar HI 9819X displays error messages to aid in troubleshooting. Warnings are
c. Select Lot displayed for most issues, while Errors are displayed for critical issues.
See the calibration chapter for messages that can occur during calibration. Other
messages are listed below.

• “Log space full” appears when the meter memory


is full and additional data cannot be logged.
Delete one or more lots from the meter (Log /
Meter Log).

• “Power fault. Check the probe cable”: this


message may appear when powering up the
meter with a probe connected. If the meter
detects a high load on the probe connection
this message is triggered. Check the probe
cable. If the problem persists, contact the
HANNA service center.

• “Language data not available”: this message


appears when powering up the meter if the
language file is not seen by the meter. Restart
the meter to verify this is a true meter error.
If the problem persists, contact the HANNA
service center.
• ‘’Dead meter battery!’’: This message appears
if the meter batteries are too low to power the
meter and it will automatically turn off. Connect
the charger if using rechargeable C batteries
or replace the alkaline batteries to continue.

• ‘’User data corrupted!’’: This message appears


when powering up the user data stored on meter
are corrupted. Restart the meter. If the problem
persists, contact the HANNA service center.

58 59

Downloaded from www.Manualslib.com manuals search engine


• ‘’Warning x’’: Any other warning that appears at power-on is identified using APPENDIX A - PROBE MAINTENANCE
a numeric code. Restart the meter. If the problem persists, contact the
HANNA service center. Some meter/probe features can be accessed but The HI 7698292 probe maintenance kit includes HI 7042S (electrolyte solution
with no guarantee. for D.O. sensor), spare membranes with o-rings for D.O. sensor, a small brush
for cleaning EC , o-rings for sensor connectors and a syringe with grease to
lubricate these o-rings.
General Maintenance
• Inspect all sensor connectors for corrosion and replace sensors if necessary.
• Inspect sensor o-rings for nicks or other damage and replace sensor if necessary.
Lubricate only with grease from kit.
• ‘’Errors x’’: Any critical errors that appear are identified using a numeric code,
Use only the supplied grease as some lubricants can cause the
and the meter is automatically switched off. Contact the HANNA service
o-rings to expand or affect the turbidity calibration standards.
center.
• After prolonged storage or cleaning, calibration of the sensors is required.
• After use rinse the probe with tap water and dry it. The pH electrode bulb must
be kept moist. Dry the D.O. and EC sensors.
• Check GLP data under “Status” to ensure the sensor is still functioning properly.
pH and pH/ORP Sensor Maintenance
• Remove the sensor protective cap. Do not be alarmed if any salt deposits are
present. This is normal with pH/ORP electrodes and they will disappear when
rinsed with water.
• Shake down the sensor as you would do with a clinical thermometer to eliminate
any air bubbles inside the glass bulb.
• If the bulb and/or junction are dry, soak the electrode in HI 70300 storage
solution for at least one hour.
• To ensure a quick response time, the glass bulb and the junction should be kept
moist and not allowed to dry. Store the sensor with a few drops of HI 70300
storage solution or pH 4.01 buffer in the protective cap. Tap water may also be
used for a very short period (few days).

Never use distilled or deionized water to store pH sensors

• Inspect the sensor for scratches or cracks. If any are present, replace the sensor.
• Cleaning procedure: clean the sensor frequently by soaking it for 1 minute in
HI 70670 or HI 70671 cleaning solution. After cleaning soak the sensor in
HI 70300 storage solution before taking measurements.

60 61

Downloaded from www.Manualslib.com manuals search engine


D.O. Sensor Maintenance (only for HI 98194 and HI 98196) APPENDIX B - PROBE DEPLOYMENT
For a top performance probe, it is recommended to replace the membrane every
The Hanna HI 769819X have been designed for a variety of water quality
2 months and the electrolyte monthly.
measurements both in situ or in active deployments in urban or natural waters.
Proceed as follows: The HI 9819X systems may be used for discrete spot sampling with a meter
• Unscrew the membrane by turning it counterclockwise. and the meter’s log on demand function or, unattended with continuous
• Rinse a spare membrane with some electrolyte while monitoring and logging from the meter. These data are then downloaded to a
shaking it gently. Refill with clean electrolyte. meter or PC and can be plotted with logging software to obtain the graphical
• Gently tap the cap over a surface to ensure that no air bubbles remain log needed for interpretation of the essential physical property of the aqueous
trapped. Avoid touching the membrane. body of water.
• With the sensor facing down, completely screw the cap clockwise. Some
electrolyte will overflow. In all of these deployment situations data quality is dependent upon the site
location, service intervals, amount of coatings, sedimentation and vegetation,
If any deposit scales the sensor, gently brush the sensor surface with the supplied and the actual installation. The probe may be installed in a horizontal bank
brush, while paying attention to not damage the plastic body. Do not use the brush (fixed installation) or a vertical suspension. The maximum depth rating of 20 m
on the membrane. (65’) for the probe should be adhered to. (Note: actual sensor specifications
may be less). The location must be accessible for the duration of the measurement
EC Sensor Maintenance (only for HI 98196 and HI 98195)
(consider seasonal flooding, freezing and other acts of nature) when selecting a
• After every series of measurements, rinse the probe with tap water. site. Many conditions may affect the quality of measurements. Select an
• If a more thorough cleaning is required, clean the sensor with the supplied installation site that is representative of the water body being monitored. Avoid
brush or a non-abrasive detergent. Ensure that the two cylindrical holes in the areas without adequate water circulation. To protect equipment it is best to avoid
sensor are free of foreign material. exposure to wind, foam, turbulence, air temperature gradients/sun, extended
periods of high flow, extended periods of high sediment and floating debris. The
standard operating procedures (SOP) for the data gathering must be upheld.
This typically includes pre and post deployment checks of the sensors to validate
data gathered between calibrations, upholding service intervals, and following
any other site-specific procedures. Grab samples for laboratory analysis or spot
sampling with another probe are addition ways to validate the measurements
taken by unattended continuous logs.
The probe is suitable for installation in confined locations such as air vaults, river
intakes, vertical wells, tanks, etc.. The streamline diameter of the probe permits
insertion into 2" pipelines. Unlike probes that require a cable support for active
deployments the probe can be manually lowered and raised by the cable due to
it’s superior strength member.
It is suitable for installation in open moving waters; rivers, streams, ditches
(farmland drainage), conveyance canals, etc.. In these cases protecting the probe
from debris is important. If the probe is suspended from a pier or bridge position
it behind a support and anchor the cable/probe to a pipe.
It is suitable for deployment in open waters; monitoring lakes, ponds, wetland
basin, infiltration basins, bays. Schedule regular service to remove aquatic weed
growth that may be interfering with representative water samples.
62 63

Downloaded from www.Manualslib.com manuals search engine


The probe is suitable for measurements in a flow cell. Pumping water to a APPENDIX C - ACCESSORIES
flow-through monitoring station has obvious pros and cons. Typically a shelter is
required to secure a pump, and flow chamber. A power requirement, shelter, PROBES packed in carton box, without sensors or protective shield
pump maintenance and higher installation cost need to be considered. Freeze
protection, security, and convenience of calibration and possibility of adding HI 7698194/4 HI 7698194 probe with 4 meter (13.1’) cable
multiple measurement points and antifouling preconditioning systems are HI 7698194/10 HI 7698194 probe with 10 meter (33’) cable
advantages to this type of installation.
HI 7698194/20 HI 7698194 probe with 20 meter (65.6’) cable
HI 7698194/40 HI 7698194 probe with 40 meter (131.2’) cable
HI 7698195/4 HI 7698195 probe with 4 meter (13.1’) cable
HI 7698195/10 HI 7698195 probe with 10 meter (33’) cable
HI 7698195/20 HI 7698195 probe with 20 meter (65.6’) cable
HI 7698195/40 HI 7698195 probe with 40 meter (131.2’) cable
HI 7698196/4 HI 7698196 probe with 4 meter (13.1’) cable
HI 7698196/10 HI 7698196 probe with 10 meter (33’) cable
General Guidelines for fixed installation:
• Select a water-sampling site that will allow collection of representative water HI 7698196/20 HI 7698196 probe with 20 meter (65.6’) cable
samples. HI 7698196/40 HI 7698196 probe with 40 meter (131.2’) cable
• Position the probe so the sensor surfaces face toward the flow. This will minimize
air bubble or fluid cavitation. Limit flow rate to moderate. Note: • Probes with different cable length are available upon request.
• Mount Probe 0 to 45° angle from vertical to avoid sensors (pH, pH/ORP) from • Order protective shields separately.
becoming electrically discontinuous due to internal electrolytes flowing away
from their internal cells.
• Install meter and probe where they will be accessible for maintenance as
required.
• Regularly visit water sampling sites to: check for damage to sensors, the
installation mountings, and the meter battery power.
• Remove aquatic weed growth that may be interfering with water sample
collection.
• Set up devices and programs for water monitoring and sampling.
• If the probe is suspended from a pier or bridge ensure that it is protected from
debris by positioning behind a support and anchoring the cable /probe to a
pipe.
• Have access to spare sensors and proper range standard solutions or buffers.
• Strictly follow the established SOP’s.
• Flow cell installation; Avoid trapped air. Maintain constant flow rate.

64 65

Downloaded from www.Manualslib.com manuals search engine


METERS WITH PROBES (packaged together in a sturdy carrying case SENSORS
with probe maintenance kit, HI 9298194 PC Application Software,
HI 7698194-0 pH sensor
HI 7698290 calibration beaker, HI 9828-20 calibration solution (230
mL), USB cable, Manual, appropriate probe shield and specified HI 7698194-1 pH/ORP sensor
sensors) HI 7698194-2 Dissolved Oxygen sensor
HI 98194 HI 98194 meter, probe with 4 meter (13.1’) cable, with HI 7698194-3 EC sensor
pH/ORP, EC, D.O. sensors.
HI 98194/10 HI 98194 meter, probe with 10 meter (33’) cable, with CABLES, CONNECTORS, ACCESSORIES
pH/ORP, EC, D.O. sensors. HI 7698290 Short calibration beaker
HI 98194/20 HI 98194 meter, probe with 20 meter (65.6’) cable, with HI 7698295 Short protective shield
pH/ORP, EC, D.O. sensors.
HI 7698297 Long, quick release flow cell
HI 98194/40 HI 98194 meter, probe with 40 meter (131.2’) cable, with
HI 7698292 Probe maintenance kit with HI 7042S (electrolyte solution
pH/ORP, EC, D.O. sensors.
for D.O. sensor), small brush, small hex key, O-rings for
HI 98195 HI 98195 meter, probe with 4 meter (13.1’) cable, with D.O. sensor (5 pcs.), O-rings for probe (5 pcs.) and
pH/ORP, EC sensors. syringe with grease to lubricate the O-rings
HI 98195/10 HI 98195 meter, probe with 10 meter (33’) cable, with HI 76981952 Probe maitenance kit with small brush, small hex key,
pH/ORP, EC sensors. O-rings for probe (5 pcs.) and syringe with grase to
HI 98195/20 HI 98195 meter, probe with 20 meter (65.6’) cable, with lubricate the O-rings.
pH/ORP, EC sensors. HI 9298194 PC application software
HI 98195/40 HI 98195 meter, probe with 40 meter (131.2’) cable, with HI 920015 Micro USB cable, PC to meter
pH/ORP, EC sensors.
HI 98196 HI 98196 meter, probe with 4 meter (13.1’) cable, with
pH/ORP, D.O. sensors.
HI 98196/10 HI 98196 meter, probe with 10 meter (33’) cable, with
pH/ORP, D.O. sensors.
HI 98196/20 HI 98196 meter, probe with 20 meter (65.6’) cable, with
pH/ORP, D.O. sensors.
HI 98196/40 HI 98196 meter, probe with 40 meter (131.2’) cable, with
pH/ORP, D.O. sensors.

66 67

Downloaded from www.Manualslib.com manuals search engine


QUICK CALIBRATION SOLUTIONS pH/ORP MAINTENANCE SOLUTIONS
HI 9828-20 Quick calibration solution, 230 mL HI 70670L pH/ORP cleaning solution for salt deposits, 500 mL
HI 9828-25 Quick calibration solution, 500 mL HI 70671L pH/ORP cleaning and disinfecting solution for algae,
fungi and bacteria, 500 mL
HI 9828-27 Quick calibration solution, 1 gal.
HI 70300L pH/ORP electrode storage solution, 500 mL
pH BUFFERS
DO SOLUTIONS
HI 5004 pH 4.01 buffer solution, 500 mL
HI 7040L Zero oxygen solution, 500 mL
HI 5046 pH 4.63 buffer solution, 500 mL
HI 7042S Electrolyte solution for D.O. sensor, 30 mL
HI 5005 pH 5.00 buffer solution, 500 mL
HI 76409A/P Spare membrane with O-ring (5 pcs.)
HI 5006 pH 6.00 buffer solution, 500 mL
HI 5068 pH 6.86 buffer solution, 500 mL CONDUCTIVITY STANDARD SOLUTIONS
HI 5007 pH 7.01 buffer solution, 500 mL HI 7030L 12880 µS/cm calibration solution, 500 mL
HI 5074 pH 7.41 buffer solution, 500 mL HI 7031L 1413 µS/cm calibration solution, 500 mL
HI 5008 pH 8.00 buffer solution, 500 mL HI 7033L 84 µS/cm calibration solution, 500 mL
HI 5009 pH 9.00 buffer solution, 500 mL HI 7034L 80000 µS/cm calibration solution, 500 mL
HI 5091 pH 9.18 buffer solution, 500 mL HI 7035L 111800 µS/cm calibration solution, 500 mL
HI 5010 pH 10.01 buffer solution, 500 mL HI 7039L 5000 µS/cm calibration solution, 500 mL

ORP SOLUTIONS
HI 7020L ORP test solution, 200/275 mV @ 20 °C, 500 mL
HI 7021L ORP test solution, 240 mV @ 20 °C, 500 mL
HI 7022L ORP test solution, 470 mV @ 20 °C, 500 mL
HI 7091L Reducing pretreatment solution, 500 mL
HI 7092L Oxidizing pretreatment solution, 500 mL

68 69

Downloaded from www.Manualslib.com manuals search engine


APPENDIX D - WARRANTY
All HANNA Instruments® meters are guaranteed for two years (sensors,
electrodes and probes for six months) against defects in workmanship and
materials when used for their intended purpose and maintained according to
instructions.
This warranty is limited to repair or replacement free of charge. Damages due to
accidents, misuse, tampering or lack of prescribed maintenance are not covered.
If service is required, contact the dealer from whom you purchased the instrument.
If under warranty, report the model number, date of purchase, serial number
and the nature of the problem.
If the repair is not covered by the warranty, you will be notified of the charges
incurred.
If the instrument is to be returned to HANNA Instruments®, first obtain a
Returned Goods Authorization number from the Technical Service department
and then send it with shipping costs prepaid.
When shipping any instrument, make sure it is properly packed for complete
protection.

Recommendations for Users


Before using this product, make sure that it is entirely suitable for the environment in which it is used.
Operation of this instrument in residential areas could cause unacceptable interferences to radio and
TV equipment, requiring the operator to take all necessary steps to correct interferences. The glass bulb
at the end of the electrode is sensitive to electrostatic discharges. Avoid touching this glass bulb at all
time. To maintain the EMC performance of equipment, the recommended cables noted in the instruction
manual must be used. Any variation introduced by the user to the supplied equipment may degrade the
instruments' EMC performance. To avoid electrical shock, do not use this instrument when voltage at
the measurement surface exceed 24 Vac or 60 Vdc. To avoid damage or burns, do not perform any
measurement in microwave ovens.
70 71

Downloaded from www.Manualslib.com manuals search engine


Thank you for reading this data sheet.

For pricing or for further information, please contact us at our UK Office, using the details
below.

UK Office
Keison Products,
P.O. Box 2124, Chelmsford, Essex, CM1 3UP, England.
Tel: +44 (0)330 088 0560
Fax: +44 (0)1245 808399
Email: [email protected]

Please note - Product designs and specifications are subject to change without notice. The user is responsible for determining the
suitability of this product.

Downloaded from www.Manualslib.com manuals search engine

You might also like