Disassembling X-Y Unit

Disassembling X-Y Unit

PAL Service Training | 2009 | TE | Version 1.0 V E R T R A U L I C H / For Internal Use Only Page 1
Disassembling X-Y Unit

Index

Removing the electronic PCBs


Removing the covers
Removing ribbon cables
X-Y axis position sensors
Installing the ribbon cables
Assembling X-Y axis
Limitation of repair
Testing X-Y axis

PAL Service Training | 2009 | TE | Version 1.0 V E R T R A U L I C H / For Internal Use Only Page 2
Disassembling X-Y Unit

Removing the electronic PCBs

Remove the screws from the covers at the back side of the instrument.
Remove the MOTIO or Control board and disconnect the ribbon cables.
Remove the CPU board if applicable.
Disconnect the cable from the Gas Flush tube if applicable (use a gripping plier).

PAL Service Training | 2009 | TE | Version 1.0 V E R T R A U L I C H / For Internal Use Only Page 3
Disassembling X-Y Unit

Removing the Covers


1
Remove both side covers from the X axis.
Remove the front cover from the Y axis.
Remove the rear cover from the Y axis.

Note: Check if the Ribbon Cable is deformed or even damaged.

2 3

PAL Service Training | 2009 | TE | Version 1.0 V E R T R A U L I C H / For Internal Use Only Page 4
Disassembling X-Y Unit

Removing the ribbon cables

Pay special attention to the X-Y axis ribbon cable .


Disengage the X/Y carriage from the X axis and disengage as well the Y axis Profile.
Remove the X-Y and the X-Z ribbon cables.

PAL Service Training | 2009 | TE | Version 1.0 V E R T R A U L I C H / For Internal Use Only Page 5
Disassembling X-Y Unit

Exchanging X-Y position sensors

Exchange the X axis position


sensor PCB (Observe the
orientation of the cable connector).

Press the PCB away from the


metallic X/Y-carriage during
tightening the two screws..

The reason for this pre-caution is to give


enough space for the sensor component and
the blind shutter built-in the X-Profile.

PAL Service Training | 2009 | TE | Version 1.0 V E R T R A U L I C H / For Internal Use Only Page 6
Disassembling X-Y Unit

Exchanging X-Y position sensors

Exchange the y axis position sensor


PCB (Observe the orientation of the
cable connector).

Ensure the contact of the


Grounding Conductor with the PCB
and the carriage.

Note: Leave the grounding point at the motor


untouched.

PAL Service Training | 2009 | TE | Version 1.0 V E R T R A U L I C H / For Internal Use Only Page 7
Disassembling X-Y Unit

Exchanging X-Y position sensors

Mount the Y-Axis Position Sensor


PCB. Press the board from the side
against the X/Y-Carriage.

The reason for this pre-caution is to


avoid a contact from the senor
component against the Toothed
Rack built-in the Y-Axis Profile.
Check this point carefully when the
Y-Profile is engaged.

Note: Leave the grounding point at the motor


untouched.

PAL Service Training | 2009 | TE | Version 1.0 V E R T R A U L I C H / For Internal Use Only Page 8
Disassembling X-Y Unit

Installing the ribbon cables

Connect the X-Y and X-Z ribbon cables to the carriage.


Install the Polymer Guiding Shield for the Y-Axis as shown in the picture on the right
side.

PAL Service Training | 2009 | TE | Version 1.0 V E R T R A U L I C H / For Internal Use Only Page 9
Disassembling X-Y Unit

Assembling X-Y Axis

Engage the X/Y-Carriage into the profile of the


X-Axis.

Engage the Y-Axis profile with the rollers


from the X/Y-Carriage.

PAL Service Training | 2009 | TE | Version 1.0 V E R T R A U L I C H / For Internal Use Only Page 10
Disassembling X-Y Unit

Assembling X-Y Axis

Form a natural bow with the Ribbon Cables


including the FlexBand.

Insert the clip in the cut-out - press


firm to hold the cables in position.

PAL Service Training | 2009 | TE | Version 1.0 V E R T R A U L I C H / For Internal Use Only Page 11
Disassembling X-Y Unit

Assembling X-Y Axis

Move the Y/Z Assembly to the left and right side and check from side of the X-
Axis if the Ribbon Cables are centered and fully lined up with each other and the
Flexband.

PAL Service Training | 2009 | TE | Version 1.0 V E R T R A U L I C H / For Internal Use Only Page 12
Disassembling X-Y Unit

Assembling X-Y Axis

Connect the ribbon cables to the MOTIO or Control PCB and reinstall it.
Reinstall the CPU board if applicanle.

PAL Service Training | 2009 | TE | Version 1.0 V E R T R A U L I C H / For Internal Use Only Page 13
Disassembling X-Y Unit

Limitation of Repair

When it comes down to the rollers there is a strict limitation. Do NOT touch the
fixation.
The rollers are matched with the side distance against the Guide Rods in the X-Y-
Assembly.
The Unit can take damage within a few months if the alignment is too hard.
The alignment can only be done in the factory using a dedicated setting gauge.

PAL Service Training | 2009 | TE | Version 1.0 V E R T R A U L I C H / For Internal Use Only Page 14
Disassembling X-Y Unit

Limitation of Repair

Toothed Rack should not moved in position neither for the X-Axis nor for the Y-Axis
Assembly.
The play of the Pinions from the X- and Y- Motors is matched to the Toothed Rack.
Uncontrolled changes can lead to destruction of the Assembly.
In any case where an adjustment seems necessary is it advisable to contact CTC
Analytics.

A single ribbon cable or a position sensor PCB are individually available.

PAL Service Training | 2009 | TE | Version 1.0 V E R T R A U L I C H / For Internal Use Only Page 15
Disassembling X-Y Unit

Testing X-Y Axis

Install the Injection Unit to the X-Y-Assembly.


Power up the PAL System.
Test Sensors.
Test Motors.
Run an Endurance Test.

PAL Service Training | 2009 | TE | Version 1.0 V E R T R A U L I C H / For Internal Use Only Page 16
PAL Firmware Hands On

PAL Service Training | 2009 | TE | Version 1.0 V E R T R A U L I C H / For Internal Use Only Page 17