2016 VM CARACTERISTICAS DE DISEÑO Clin Chest Med
2016 VM CARACTERISTICAS DE DISEÑO Clin Chest Med
2016 VM CARACTERISTICAS DE DISEÑO Clin Chest Med
Modern Mechanical
Ventilators
Neil R. MacIntyre, MD
KEYWORDS
Mechanical ventilation Modes of ventilation Patient–ventilator interactions
Ventilator-induced lung injury Closed-loop control
KEY POINTS
Mechanical ventilator design features can be challenging to understand because of imprecise and
confusing terminology.
Mechanical ventilator design features have evolved, reflecting new awareness of ventilator-induced
lung injury and patient–ventilator synchrony.
As newer designs are introduced, clinical adoption will require outcome data to support their use.
meant parameters that the ventilator manipulated flow or a set inspiratory pressure. With flow
targeting, the ventilator adjusts pressure to main- ventilatory support. The mode controller is an elec-
tain a clinician-determined flow magnitude and tronic, pneumatic, or microprocessor-based sys-
pattern (sine, square, accelerating, decelerating); tem designed to provide the proper combination
with pressure targeting, the ventilator adjusts of breaths according to set algorithms and feed-
flow to maintain a clinician-determined inspiratory back data (conditional variables). The 5 most com-
pressure. Modern systems also usually allow mon modes are volume assist control (VACV),
adjustment of the rate of pressure rise in pressure pressure assist control (PACV), volume synchro-
targeting. Cycle variables are generally a set vol- nized intermittent mandatory ventilation (V-SIMV),
ume, a set inspiratory time, or a set reduction in pressure synchronized intermittent mandatory
inspiratory flow as the lung fills. This flow-cycling ventilation (P-SIMV), and stand-alone pressure
criterion is manufacturer-specific (eg, 25%–35% support ventilation (PSV) (Table 1).4 Examples of
of peak flow), or it can be clinician-adjusted on proprietary names for these basic modes are given
many newer machines. A secondary cycling in Table 2. Depending upon the set control breath
mechanism may be present on some devices if rate, VACV and PACV can range from totally ma-
inspiratory time exceeds a certain fraction chine controlled to totally patient assisted. V-
(eg, 80%) of a set total cycle time. Breaths can SIMV and P-SIMV can provide VA and VC or PA
also be cycled off if pressure limits are exceeded. and PC breaths respectively interspersed with
With this approach, breath delivery algorithms either unsupported or pressure supported (PS)
from modern mechanical ventilators can be breaths. Data from international surveys5 indicate
broken into 5 basic breaths: volume control (VC), that the most commonly used mode worldwide is
volume assist (VA), pressure control (PC), pressure volume assist control, with pressure assist control
assist (PA), and pressure support (PS) (Fig. 1).4 a distant second. IMV modes have been steadily
decreasing in use, while stand-alone PSV modes
have been increasing in use.
Basic Modes of Ventilatory Support
Choice of mode depends upon the clinical goals
The availability and delivery logic of different and an understanding of ventilator breath design
breath types define the mode of mechanical features.4 Mandatory breath rates are set
Fig. 1. The 5 basic breaths defined by trigger, target, and cycle variables. Depicted are airway pressure, flow, and
volume tracings over time. Solid lines reflect set changes; dotted lines reflect variable changes from effort or me-
chanics changes. The five basic breaths. Volume control is machine triggered, flow targeted, and volume cycled.
Volume assist is patient triggered, flow targeted, and volume cycled. Pressure control is machine triggered, pres-
sure targeted, and time cycled. Pressure assist is patient triggered, pressure targeted, and time cycled. Pressure
support is patient triggered, pressure targeted, and flow cycled. (From MacIntyre NR. Principles of mechanical
ventilation. In: Broaddus VC, editor. Murray and Nadel’s textbook of respiratory medicine, 6th edition.
New York: Elsevier; 2016. p. 1762; with permission.)
Downloaded for Anonymous User (n/a) at Cooper University Hospital-Rowan University from ClinicalKey.com by Elsevier on March 02, 2019.
For personal use only. No other uses without permission. Copyright ©2019. Elsevier Inc. All rights reserved.
Design Features of Modern Mechanical Ventilators 609
Table 1
that dedicated APRV modes on most devices
The 5 basic modes defined by the breaths have the set inspiratory pressure referenced to at-
available mospheric pressure rather than the set expiratory
pressure. Proponents of APRV argue that the
Breath Types Available long I:E ratio raises mean Paw without additional
set positive end expiratory pressure (PEEP) or tidal
Mode VC VA PC PA PS Sp
volume (VT) and that the spontaneous efforts dur-
Volume assist control X X – – – – ing the inflation phase enhance gas mixing and
Pressure assist control – – X X – – cardiac filling.7,8 Examples of proprietary names
Volume SIMV X X – – X X are given in Table 2.
Pressure SIMV – – X X X X
Positive End Expiratory Pressure
Pressure support – – – – X –
PEEP can be generated in 2 basic ways: applied or
The breaths are the 5 breaths depicted in Fig. 1 plus an un-
assisted spontaneous breath (Sp). Note that the clinician- intrinsic. Applied PEEP is clinician set and is usu-
set breath rate can result in VACV and PACV being totally ally provided by valving systems in the expiratory
controlled ventilation (high set rate), virtually totally assis- limb. Modern ventilators also can adjust circuit
ted ventilation (very low or absent set rate), or assist con- flow during exhalation to assure PEEP mainte-
trol ventilation (intermediate set rate).
Abbreviations: PA, pressure assist; PC, pressure control;
nance in the setting of circuit leaks. Intrinsic
PS, pressure support; VA, volume assist; VC, volume control. PEEP develops in the setting of high minute venti-
From MacIntyre NR. Principles of mechanical ventila- lation, short expiratory times, and high airway
tion. In: Broaddus VC, editor. Murray and Nadel’s text- resistance/high compliance lung units. Impor-
book of respiratory medicine, 6th edition. New York: tantly, applied PEEP distributes evenly throughout
Elsevier; 2016. p. 1762; with permission.
the lung while intrinsic PEEP is highest in high
resistance/high compliance lung units and lowest
depending upon the reliability of the patient’s effort in low compliance/low resistance units.9 Conven-
to supply an appropriate number of breaths. Pres- tional approaches to PEEP generally rely on set
sure versus flow/volume targeting balances the PEEP and avoidance of intrinsic PEEP. However,
synchrony enhancement of pressure targeting proponents of APRV argue for the use of intrinsic
against the volume guarantee of flow/volume tar- PEEP to maximize expiratory flow and minimize
geting.6 When using patient-triggered pressure- expiratory time.
targeted breaths, cycling on time (PA breaths)
versus flow (PS breaths) depends on patient com- FEEDBACK CONTROL FEATURES
fort/synchrony.
Airway pressure release ventilation (APRV) is As mechanical ventilators have evolved, so has the
often touted as a new mode but in fact is simply capability for microprocessor-based systems to
a variant of P-SIMV, in which the inspiratory time monitor conditional variables and use this informa-
is set longer than the expiratory time. Patient ef- tion to automatically adjust timing, flow, pressure,
forts thus occur during the inflation phase and and even FiO2 (feedback control). An early example
can produce additional unassisted or PS breaths. was the use of a patient effort sensor (conditional
A point of confusion exists in setting up APRV in variable) to adjust the number of mechanical
Table 2
Examples of proprietary names for the 5 basic modes and 2 feedback features
Downloaded for Anonymous User (n/a) at Cooper University Hospital-Rowan University from ClinicalKey.com by Elsevier on March 02, 2019.
For personal use only. No other uses without permission. Copyright ©2019. Elsevier Inc. All rights reserved.
610 MacIntyre
breaths provided during either assist control 2 breath types depending on the number of patient
modes or SIMV.10 A variation on this breath rate efforts. The maximum pressure change from
feedback mechanism was mandatory (or mini- breath to breath on most systems generally is
mum) minute ventilation, which used minute venti- limited to a few centimeters of water to prevent
lation to adjust the number of positive pressure large swings in pressure and volume.
breaths delivered.11 Currently available systems These modes have been assessed clinically in 2
to partially close the loop will be described.12,13 settings. First, in severe parenchymal lung injury
(eg, ARDS [acute respiratory distress syndrome]),
Inspiratory Pressure and Flow Adjustments PRVC has been used as a way to provide more syn-
Based on Artificial Airway Geometry chronous pressure-targeted breaths while assuring
that safe tidal volume delivery is maintained. One
The endotracheal tube (ETT) imposes a significant
study demonstrated that this was possible, although
inspiratory resistance on a spontaneously breath-
a minority of patients had significant periods of time
ing patient.14 This imposed load can have an
with excess VT.19 Second, VS has been touted as a
impact on flow synchrony during interactive assis-
means to automatically wean patients, the theory
ted/supported breaths and can make it difficult to
being that as patients recover, they will make stron-
assess potential for ventilator withdrawal during
ger inspiratory efforts, and VS will automatically
periods of unassisted/unsupported breathing.
reduce inspiratory pressure. Conversely, inspiratory
Low-level (eg, 5–8 cmH2O) PS has been pro-
pressure would increase if patient effort diminished
posed as a way of eliminating the ETT resistive
or respiratory system mechanics worsened.
load.15 However, the PS algorithm supplies a con-
Whether this approach is superior to routine sponta-
stant inspiratory pressure, which, because of the
neous breathing trials (SBTs) is unclear.
high fixed resistance of the ETT, tends to under-
One must also be cautious in using VS in this
compensate the load at the beginning of the
weaning setting, because if the clinician set volume
breath. Patient muscle unloading thus is uneven
is excessive for patient demand, a patient may not
and may be suboptimal.
attempt to take over the work of breathing for that
To better address this loading pattern, many
volume, and thus support reduction and weaning
ventilators have the capability to calculate the
may not progress. In addition, if the pressure level
ETT resistance properties based on clinician input
increases in an attempt to maintain an inappropri-
of ETT length and diameter. The ventilator incorpo-
ately high set tidal volume in the patient with airflow
rates this calculation with measurements of instan-
obstruction, intrinsic PEEP (PEEPi) may result. VS
taneous flow to apply pressure proportional
may also inappropriately lower inspiratory pressure
to resistance throughout the total respiratory cy-
in a patient with excessive flow demands induced
cle.16,17 It must be recognized that the ETT
by pain, anxiety, or acidosis.20
compensation strategy is based on the input ge-
ometry of the artificial airway and cannot account
for changes in tube characteristics induced by Enhanced Feedback Control of Combination
kinks or partial occlusions or the relationship of Pressure- and Flow-Targeted Breaths
the tube opening against the tracheal wall.
Airway occlusion pressure (P0.1),21 oxygen satura-
tion (SpO2 ),22 and end-tidal CO2 concentrations23,24
Feedback Control of Combination Pressure-
have been incorporated into the pressure flow/vol-
and Flow-Targeted Breaths
ume hybrid breaths described. The one system that
Over the last 2 decades, several engineering inno- is commercially available uses end-tidal CO2 and
vations have attempted to combine the flow syn- respiratory rate along with the tidal volume to adjust
chrony advantages of pressure-targeted breaths the applied inspiratory pressure (SmartCare,
with the volume guarantee features of flow/volume Dragerwerk AG, Lubeck, Germany).24 The system
targeted breaths. The most common approach attempts to find an inspiratory pressure that main-
uses standard pressure-targeted breaths with the tains the respiratory rate and tidal volume in a
ventilator adjusting the pressure target according clinician-set comfort zone. The end-tidal CO2
to a clinician-set VT.18 When these breaths are serves as a backup signal to assure adequate venti-
exclusively supplied with time cycling, the mode lation is occurring. Inspiratory pressure is reduced
is commonly referred to as pressure-regulated vol- to as low a level as possible within these bound-
ume control (PRVC) but has a number of proprie- aries; then the clinician should be alerted to perform
tary names (see Table 2). When these breaths an SBT when this pressure reaches 9 cm H2O.
are supplied exclusively with flow cycling, the Although clinical trials have failed to consistently
mode is commonly referred to as volume support show an advantage to this approach,25,26 an auto-
(VS). Some ventilators will switch between these mated system that is just as good as clinicians
Downloaded for Anonymous User (n/a) at Cooper University Hospital-Rowan University from ClinicalKey.com by Elsevier on March 02, 2019.
For personal use only. No other uses without permission. Copyright ©2019. Elsevier Inc. All rights reserved.
Design Features of Modern Mechanical Ventilators 611
could have applications in settings with rapidly MODES DRIVEN BY NOVEL SENSORS OF
recovering patients or low availability of clinicians PATIENT EFFORT
to make frequent assessments.
Two new modes have been introduced over the
Feedback Control of Ventilator Breath last 2 decades that use unique feedback control
Delivery Based on Respiratory System based on patient effort to control positive pressure
Mechanics breath delivery.13,33 The first is proportional assist
ventilation (PAV), an approach that applies a
A novel approach to automated feedback control of clinician set pressure and flow gain on patient-
ventilator support controls a pressure-targeted generated flow and volume.34 PAV uses intermit-
breath using a VT, frequency and inspiratory-to- tent controlled test breaths to calculate resistance
expiratory (I:E) ratio algorithm based on respiratory and compliance. It can then use measured flow
system mechanics. Known as adaptive lung venti- and volume to calculate both resistive and elastic
lation or adaptive support ventilation (ASV),27–30 work. The clinician is required to set a desired pro-
the system calculates respiratory system me- portion of the total work that should be performed
chanics using several controlled test breaths. It by the ventilator. The ventilator then measures the
then uses a minimal work calculation31 to set the patient flow and volume demand with each breath
frequency–tidal volume pattern that minimizes the and adds both pressure and flow to provide the
combined resistance and compliance components selected proportion of the breathing work. PAV
of work. The ASV algorithm then attempts to mini- has been compared with power steering on an
mize intrinsic PEEP by measuring the expiratory automobile, an analogy that has much truth. Like
time constants (RCe 5 resistance compliance) PAV, power steering reduces the work to turn the
and providing an expiratory time of at least 3 RCe. wheels but does not automatically steer the car;
With ASV, clinicians must set the desired minute the driver must control the car’s ultimate direction
ventilation and the proportion of that minute just as the patient ultimately must control the
ventilation that the machine is to supply. Ideal magnitude of the breath and the timing of the
body weight also can be used to calculate the breathing pattern.
desired minute ventilation based on metabolic de- Because PAV requires sensors in the ventilator
mands and predicted dead space. Clinicians also circuitry to measure patient effort, it is susceptible
must set the PEEP and FIO2 in the United States to the same sensor performance and intrinsic
(described later in this article). When spontaneous PEEP issues that affect breath triggering in other
efforts occur with ASV, the algorithm responds assisted modes.24 Also like conventional assisted
with fewer mandatory breaths and adjusts inspira- modes, the clinician must set PEEP and FiO2.
tory pressure according to the minimal work tidal Finally, breath termination (cycling) is much
volume considerations listed previously. ASV has like pressure support and is determined by a
been shown to perform as designed, although in clinician-adjustable percentage of maximal inspi-
healthier lungs, tidal volumes may exceed lung- ratory flow.
protective guidelines.30 Meaningful outcome PAV has been shown in multiple studies to
studies do not exist. perform as designed.35,36 However, whether PAV
improves meaningful clinical outcomes (eg, seda-
Feedback Systems Controlling Positive End-
tion needs, shorter needs for mechanical ventila-
Expiratory Pressure and FiO2
tion) remains to be determined.
On a mechanical ventilator, an FiO2 controller A second novel mode is neurally adjusted venti-
conceptually could be coupled to a feedback latory assistance (NAVA), which utilizes a dia-
controller of PEEP to meet oxygenation and me- phragmatic electromyogram (EMG) signal to
chanical goals (ie, PaO2 or SpO2 targets balanced trigger, govern flow, and cycle ventilatory assis-
against lung compliance or plateau pressure). tance.37 The EMG sensor is an array of electrodes
One system approved outside the United States mounted on an esophageal catheter that is posi-
incorporates the PEEP-FiO2 table used by the Na- tioned in the esophagus at the level of the dia-
tional Institutes of Health (NIH) ARDS Network phragm. Ventilator breath triggering is thus
study.32 With this algorithm, PEEP and FiO2 combi- virtually simultaneous with the onset of phrenic
nations are guided by a PaO2 target range of 55 to nerve excitation of the inspiratory muscles, and
80 mm Hg and a plateau pressure limit of 30 to breath cycling is tightly linked to the cessation of
35 cm H2O. Although this table proved safe and inspiratory muscle contraction. Flow delivery is
effective in ARDS Network trials, whether an auto- driven by the intensity of the EMG signal (electrical
mated system using it will improve outcomes has activity of the diaphragm or EADi), and the clinician
yet to be demonstrated. sets a mL/mV gain factor.
Downloaded for Anonymous User (n/a) at Cooper University Hospital-Rowan University from ClinicalKey.com by Elsevier on March 02, 2019.
For personal use only. No other uses without permission. Copyright ©2019. Elsevier Inc. All rights reserved.
612 MacIntyre
Like PAV, NAVA depends exclusively on patient 12. MacIntyre NR, Branson RD. Feedback enhance-
effort for timing, intensity, and duration of the ments on conventional ventilator breaths. In:
breath. Thus, like PAV, clinicians must set appro- Tobin MJ, editor. Mechanical ventilation, principles
priate alarms and backup positive pressure venti- and practice. 3rd edition. Philadelphia: McGraw
lation, especially for patients with unreliable Hill; 2011. p. 403–14.
respiratory drives. Also like PAV, clinicians must 13. Lellouche F, Brochard L. Advanced closed loops
set PEEP and FiO2. during mechanical ventilation (PAV, NAVA, ASV,
NAVA has been shown to perform as SmartCare). Best Pract Res Clin Anaesthesiol
designed,13,37–39 and conceptually, NAVA should 2009;23:81–93.
provide excellent patient–ventilator synchrony. 14. Bersten AD, Rutten AJ, Vedig AE, et al. Additional
However, data demonstrating improved outcomes work of breathing imposed by endotracheal tubes,
(eg, duration of mechanical ventilation, sedation breathing circuits, and intensive care ventilators.
needs) are lacking. Another concern with NAVA Crit Care Med 1989;17:671–80.
is the expense associated with the EMG sensor. 15. Bersten AD, Rutten AJ, Vedig AE. Efficacy of pres-
sure support in compensating for apparatus work.
Anaesth Intensive Care 1993;21:67–71.
REFERENCES
16. Guttmann J, Bernhard H, Mols G, et al. Respiratory
1. MacIntyre NR. Evolution of the modern mechanical comfort of automatic tube compensation and inspi-
ventilator. In: Hill N, Levy M, editors. Ventilator man- ratory pressure support in conscious humans. Inten-
agement strategies for critical care. New York: sive Care Med 1997;23:1119–23.
Marcel Dekker; 2000. p. 37–52. 17. Fabry B, Haberthur G, Zappe D, et al. Breathing
2. Chatburn RL, El-Khatib M, Mireles-Cabodevila E. pattern and additional work of breathing in sponta-
A taxonomy for mechanical ventilation: 10 funda- neously breathing patients with different ventilatory
mental maxims. Respir Care 2014 Nov;59(11): demands during inspiratory pressure support and
1747–63. automatic tube compensation. Intensive Care Med
3. Branson RD, Chatburn RL. Technical description 1997;23:545–50.
and classification of modes of ventilator operation. 18. Branson RD, MacIntyre NR. Dual control modes
Respir Care 1992;37:1026–44. of mechanical ventilation. Respir Care 1996;41:
4. MacIntyre NR. Principles of mechanical ventilation. In: 294–305.
Broaddus VC, editor. Murray and Nadel’s textbook of 19. Kallet RH, Campbell AR, Dicker RA, et al. Work of
respiratory medicine. 6th edition. New York: Elsevier; breathing during lung-protective ventilation in pa-
2016. p. 1761–77. tients with acute lung injury and acute respiratory
5. Esteban A, Ferguson ND, Meade MO, et al. Evolu- distress syndrome: a comparison between volume
tion of mechanical ventilation in response to clinical and pressure-regulated breathing modes. Respir
research. Am J Respir Crit Care Med 2008;177: Care 2005;50:1623–31.
170–7. 20. Jaber S, Delay JM, Matecki S, et al. Volume-guaran-
6. MacIntyre NR, McConnell R, Cheng KC, et al. Pa- teed pressure-support ventilation facing acute
tient-ventilator flow dyssynchrony: flow-limited changes in ventilatory demand. Intensive Care
versus pressure-limited breaths. Crit Care Med Med 2005;31:1181–8.
1997;25:1671–7. 21. Iotti GA, Braschi A. Closed loop support of ventilator
7. Habashi NM. Other approaches to open-lung venti- workload: the P0.1 controller. Respir Care Clin N Am
lation: airway pressure release ventilation. Crit Care 2001;7:441–51.
Med 2005;33(Suppl 3):S228–40. 22. Strickland JH, Hasson JH. A computer-controlled
8. Myers T, MacIntyre NR. Does airway pressure ventilator weaning system: a clinical trial. Chest
release ventilation offer important new advantages 1993;103:1220–6.
in mechanical ventilatory support? Respir Care 23. Dojat M, Harf A, Touchard D, et al. Evaluation of a
2007;52:452–60. knowledge-based system providing ventilatory man-
9. Kacmarek RM, Kirmse M, Nishimura M, et al. The ef- agement and decision for extubation. Am J Respir
fects of applied vs auto-PEEP on local lung unit Crit Care Med 1996;153:997–1004.
pressure and volume in a four-unit lung model. 24. Dojat M, Harf A, Touchard D, et al. Clinical evalua-
Chest 1995;108:1073–9. tion of a computer-controlled pressure support
10. Weisman IM, Rinaldo JE, Rogers RM, et al. Intermit- mode. Am J Respir Crit Care Med 2000;161:
tent mandatory ventilation. Am Rev Respir Dis 1983; 1161–6.
127:641–7. 25. Lellouche F, Mancebo J, Jolliet P, et al. A multicenter
11. Hewlett AM, Platt AS, Terry VG. Mandatory minute randomized trial of computer-driven protocolized
volume: a new concept in weaning from mechanical weaning from mechanical ventilation. Am J Respir
ventilation. Anaesthesia 1977;32:163–9. Crit Care Med 2006;174:894–900.
Downloaded for Anonymous User (n/a) at Cooper University Hospital-Rowan University from ClinicalKey.com by Elsevier on March 02, 2019.
For personal use only. No other uses without permission. Copyright ©2019. Elsevier Inc. All rights reserved.
Design Features of Modern Mechanical Ventilators 613
26. Rose L, Presneill J, Johnston L, et al. 33. Sinderby C, Beck J. Proportional assist ventilation
A randomised, controlled trial of conventional and neurally adjusted ventilatory assist–better ap-
versus automated weaning from mechanical venti- proaches to patient ventilatory synchrony? Clin
lation using SmartCare/PS. Intensive Care Med Chest Med 2008;29(2):329–42.
2008;34:1788. 34. Younes M. Proportional assist ventilation, a new
27. Laubscher TP, Heinrichs W, Weiler N, et al. An adap- approach to ventilatory support. Am Rev Respir
tive lung ventilation controller. IEEE Trans Biomed Dis 1992;145:114–20.
Eng 1995;41:51–9. 35. Bosma K, Ferreyra G, Ambrogio C, et al. Patient-
28. Iotti GA, Polito A, Belliato M, et al. Adaptive support ventilator interaction and sleep in mechanically
versus conventional ventilation for total ventilator ventilated patients; pressure support versus propor-
support in acute respiratory failure. Intensive Care tional assist ventilation. Crit Care Med 2007;35(4):
Med 2010;36:1371–9. 1048–54.
29. Weiler N, Eberle B, Latorre F, et al. Adaptive lung 36. Grasso S, Puntillo F, Mascia L, et al. Compensa-
ventilation. Anaesthesist 1996;45:950–6. tion for increase in respiratory workload during
30. Dongelmans DA, Paulus F, Veelo DP, et al. mechanical ventilation. Pressure-support versus
Adaptive support ventilation may deliver unwanted proportional-assist ventilation. Am J Respir Crit
respiratory rate–tidal volume combinations in pa- Care Med 2000;161(3 Pt 1):819–26.
tients with acute lung injury ventilated according 37. Sinderby C. Neurally adjusted ventilatory assist
to an open lung strategy. Anesthesiology 2011; (NAVA). Minerva Anestesiol 2002;68:378–80.
114(5):1138–43. 38. Brander L, Leong-Poi H, Beck J, et al. Titration and
31. Otis AB. The work of breathing. Physiol Rev 1954; implementation of neurally adjusted ventilatory
34(3):449–58. assist in critically ill patients. Chest 2009;135(3):
32. NIH ARDS Network. Ventilation with lower tidal vol- 695–703.
umes as compared with traditional tidal volumes 39. Spahija J, de Marchie M, Albert M, et al. Patient-
for acute lung injury and the acute respiratory ventilator interaction during pressure support venti-
distress syndrome. N Engl J Med 2000;342: lation and neurally adjusted ventilatory assist. Crit
1301–8. Care Med 2010;38(2):518–26.
Downloaded for Anonymous User (n/a) at Cooper University Hospital-Rowan University from ClinicalKey.com by Elsevier on March 02, 2019.
For personal use only. No other uses without permission. Copyright ©2019. Elsevier Inc. All rights reserved.