743 Rancimat: Manual
743 Rancimat: Manual
743 Rancimat: Manual
Manual
8.743.8003EN / 2019-09-25
Metrohm AG
CH-9100 Herisau
Switzerland
Phone +41 71 353 85 85
Fax +41 71 353 89 01
[email protected]
www.metrohm.com
743 Rancimat
Manual
8.743.8003EN / 2019-09-25
Technical Communication
Metrohm AG
CH-9100 Herisau
[email protected]
Table of contents
1 Introduction 1
1.1 Instrument description ......................................................... 1
1.2 Rancimat method ................................................................. 2
1.3 About the documentation ................................................... 3
1.3.1 Symbols and conventions ........................................................ 3
1.4 Safety instructions ................................................................ 4
1.4.1 General notes on safety ........................................................... 4
1.4.2 Electrical safety ........................................................................ 5
1.4.3 Flammable solvents and chemicals ........................................... 6
1.4.4 Recycling and disposal ............................................................. 6
4 Operation 25
4.1 Fundamentals of operation ............................................... 25
4.1.1 Starting and exiting the program ........................................... 25
4.1.2 Terms .................................................................................... 26
4.1.3 Control window .................................................................... 26
4.1.4 Results window ..................................................................... 28
4.1.5 File types ............................................................................... 31
4.1.6 Context-sensitive menus ........................................................ 32
6 Troubleshooting 163
6.1 Problems ........................................................................... 163
8 Accessories 173
Index 174
Table of figures
Figure 1 Measuring arrangement (schematic representation) ........................... 3
Figure 2 Front 743 Rancimat ........................................................................... 7
Figure 3 Rear 743 Rancimat ............................................................................ 8
Figure 4 Mounting accessories for the air supply (rear of the instrument) ...... 11
Figure 5 Equipping the reaction and measuring vessels ................................. 14
Figure 6 Equipping the reaction vessel for determining Delta T ...................... 52
Figure 7 Accessories for GLP test Temperature ............................................ 146
1 Introduction
stating whether the GLP tests have been fulfilled. Metrohm offers a GLP
test set to carry out these tests .
Measuring
vessel
Conductivity
measuring cell
Measuring
solution
Reaction vessel
Sample
Heating block
CAUTION
WARNING
This instrument has left the factory in a flawless state in terms of technical
safety. To maintain this state and ensure non-hazardous operation of the
instrument, the following instructions must be observed carefully.
WARNING
Flammable substances
WARNING
The heating block of the 743 Rancimat can be heated to 229.9 °C.
Flammable substances may ignite at such temperatures.
Adjust the oven's maximum heating temperature to the sample that is
to be analyzed.
WARNING
WARNING
WARNING
Mains voltage
WARNING
WARNING
WARNING
743 Rancimat
1
2
Type:
Nr.: f:
U: S:
From To
flask flask
Air out
Fuse
100-115V: 4A(TH) RS 232 Pt 100
220-240V: 2A(TH)
Air/N 2in
WARNING - Fire Hazard -
For continued protection replace only
with the same type and rating of fuse
4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
Figure 3 Rear 743 Rancimat
3 Installation
3.1.2 Checks
Immediately after receipt, check whether the shipment has arrived com-
plete and without damage by comparing it with the delivery note.
3.1.3 Location
The instrument has been developed for operation indoors and may not be
used in explosive environments.
Place the instrument in a location of the laboratory suitable for operation
and free of vibrations, if possible protected from corrosive atmospheres
and contamination by chemicals.
The instrument should be protected against excessive temperature fluctua-
tions and direct sunlight.
NOTICE
1
From To 9
flask flask
2
Air out
10
3
11
Air/N2 in
12
4 5 6 7
Figure 4 Mounting accessories for the air supply (rear of the instru-
ment)
NOTICE
The dust filter serves for filtering the air sucked through the air
pump and must be replaced at periodic intervals (see chapter 5.2,
page 161).
CAUTION
Do not fill the hot molecular sieve directly into the drying flask
after regeneration, as otherwise the plastic filter on the filter tube
will melt.
Wait until the molecular sieve has cooled down before filling.
NOTICE
■ Screw the other end of the FEP tubing onto the Air/N2 in con-
nection (4-12).
NOTICE
With an external gas supply, the gas flow cannot be regulated in the PC
program. The gas flow must be set manually using the reducing valve
and the gas flow display (see chapter 4.2.4, page 40).
2
11
12
13
3
14
4
5 15
6
16
7
8 17
18
9
19
10
20
WARNING
The foam barrier can melt if it projects too deeply into the heating
block.
Ensure that the foam barrier is at least 7 cm above the base of the
reaction vessel.
NOTICE
WARNING
WARNING
Ensure that the instrument is never put into operation with fuses of
another type, otherwise there is a risk of fire!
3 Replace fuses
■ Carefully take the defective fuses out of the fuse holder and
replace them with two new fuses suitable for the set mains volt-
age of the type TH (slow-acting, with high switching capacity):
– 115 V 4 A (TH) Metrohm No.: U.600.0022
– 230 V 2 A (TH) Metrohm No. U.600.0107
WARNING
NOTICE
3.4 Connecting a PC
3.4.1 Connecting the 743 Rancimat and the PC
CAUTION
Always switch the Rancimat and PC off before you connect the two
devices with the RS-232 cable 6.2134.100.
CAUTION
1 ■ Exit 743 Rancimat if you have started the software after the instal-
lation.
■ Start the installation file by double-clicking and follow the instruc-
tions of the installation program.
When starting the program for the first time, the Administrator password
must be set. Proceed as follows:
2 Start program
■ In the Windows Start menu under Pro-
grams ▶ Metrohm ▶ Rancimat, click on the menu item Ranci-
mat.
The login window for the Administrator opens:
4 Log in as Administrator
■ Enter the Administrator password set before in the Password
field and confirm with [OK].
■ Confirm the message "There are no units configured yet!"
with [OK].
4 Operation
This section describes the most important points for operation of the 743
Rancimat.
You can find more detailed information in the Online help of the software,
which you can use to rapidly and easily call up the required information
everywhere, via [F1].
NOTICE
If you are starting the program for the first time, you first have to set an
Administrator password (see chapter 3.4.3, page 22).
1 Start program
■ In the Windows Start menu under Pro-
grams ▶ Metrohm ▶ Rancimat, click on the menu item Ranci-
mat.
The following dialog window appears:
2 Log in
■ Select the required User.
■ Enter the Password and confirm with [OK].
3 Close program
■ Click on the menu item File ▶ Exit or the symbol in the upper
right corner of the dialog window.
For opening and closing the Results window, see page 29.
4.1.2 Terms
Control window The main window 743 Rancimat Control is
called the control window. It contains all the
functions for monitoring and controlling the
Rancimats connected to the PC.
Method A method comprises all parameters for carrying
out and evaluating a determination.
Determination Determination refers to the automatic determi-
nation of the induction time and/or stability time
for a sample. In order to carry out a determina-
tion, a method suitable for the samples must be
selected.
Results The results of a determination are automatically
saved in the database Repos.mrd and can be
viewed in the Results window.
Reevaluation Reevaluation refers to the subsequent revision of
a determination, in particular the manual specifi-
cation of the induction time using tangents.
Recalculation The recalculation can be used to subsequently
recalculate results, formulas and standards.
Extrapolation Extrapolation can be used to convert the results
measured at various temperatures to a required
target temperature. In addition, the factor for
converting the induction time to the standard
time can be determined.
Structure
The elements of the Control window are the menu bar, the toolbar, the
control panel and the status bar.
Menus
The Control window contains the following main menus:
File Create new methods, open existing methods,
manage methods, open Results window, print,
new login, close program.
View Activate/Deactivate toolbar and status bar, dis-
play instrument information, change live param-
eters, display status overview, display event log-
ging, select instrument.
Tools Manage cell constants, perform GLP tests, set
timer, set gas flow control, start all channels,
determine Delta T, record temperature, perform
service diagnosis.
Symbols
New method Create a new method (see chapter 4.5.1, page 56).
Open method Open an existing method (see chapter 4.5.1, page 56).
Method Manager Open, rename and delete methods (see chapter 4.5.1, page 56).
Results Open Results window (see "Opening and closing the Results window", page
29).
Live parameters Display live parameters which can be changed for a current determination (see
chapter 4.6.6, page 85).
Status overview Display status overview for the instruments connected (see chapter 4.3.2, page
45).
Instrument informa-
tion Instrument information (see chapter 4.3.1, page 43).
GLP status Display GLP status (see chapter 4.8.3, page 145).
Structure
The elements of the Results window are the menu bar, the toolbar, the
subwindows (determination overview, live, single and multiple graphs)
and the status bar.
Menus
The Results window contains the following main menus:
File Open database, print, close dialog window.
Edit Copy, select, delete content of the filter.
View View selection: determination overview, determi-
nation and method data, GLP.
Determination Search, sort, filter, display graphs, perform
extrapolation, perform recalculation, export
determinations, delete determinations.
Format Format determination overview.
Options Carry out general program settings.
Window Arrange subwindows.
Help Display program-specific Online help.
Symbols
The toolbar of the Results window contains the following symbols:
Print Print results, curves and overview lists (see chapter 4.7.6, page 129).
Display determina-
tion overview Display determination overview (see chapter 4.7.1, page 88).
Ascending Sort column in ascending order (see chapter 4.7.1, page 88).
Descending Sort column in descending order (see chapter 4.7.1, page 88).
Delete filter Delete content of the filter (see chapter 4.7.1, page 88).
Filter selection Only display selected determinations (see chapter 4.7.1, page 88).
Selection not in filter Hide selected determinations (see chapter 4.7.1, page 88).
Single graph Display single graph (see chapter 4.7.3, page 113).
Multiple graph Display multiple graph (see chapter 4.7.3, page 113).
Export to Word Export Determination overview to Word (see chapter 4.7.6, page 129).
Export to Excel Export Determination overview to Excel (see chapter 4.7.6, page 129).
Export determination
Export determination and method data to a TXT file (see chapter 4.7.6, page
and method data
129).
Tile vertically Arrange subwindows on top of each other (see chapter 4.7.8, page 142).
Tile horizontally Arrange subwindows beside each other (see chapter 4.7.8, page 142).
Zooming
Proceed as follows:
4.1.8 Help
You can call up help for the current topic anywhere with the symbol ,
the menu item Help ▶ Help topics or the key [F1].
Green text can be clicked on. This takes you to another help
topic.
Violet text marks menu items, parameters or buttons in the
program.
Blue text marks titles and important information.
2 Select interface
■ Under COM port, select the serial interface of the PC to which
the instrument is connected.
3 Restart program
■ Close the program with the menu item File ▶ Exit.
■ In the Windows Start menu under Pro-
grams ▶ Metrohm ▶ Rancimat, click on the menu item Ranci-
mat.
■ Enter the Administrator password and confirm with [OK].
The program is opened. The message Instr.#: Ready now appears
in the status bar for all connected instruments.
NOTICE
User lists are best created and passwords are best entered immediately
after the first program start.
The setup for the users and their access rights is carried out in the follow-
ing order:
1. Add a new group
2. Adjust rights for this group
3. Create users in this group
After having entered the access rights, the Login window for selecting the
user and entering the password appears each time the program is started.
All methods, determinations and reports are marked with the user name.
It is possible to change the user at any time with the menu item
File ▶ New login....
Deleting a user
2 Delete user
■ Right click on the user to be deleted to activate the menu item
Delete user.
■ Confirm the message Do you really want to delete user "user
name"? with [Yes].
Deleting a group
2 Delete group
■ Right click on the group to be deleted in order to select the menu
item Delete group.
■ Confirm the message Do you really want to delete group
"group name"? with [Yes].
4.2.3 Timer
The function Timer can be used to start the two heating blocks A and B,
independently of one another, automatically and at a time specified by
you. The Rancimat has to be switched on and the software must be run-
ning.
Two different examples will explain how to program the timer.
Proceed as follows:
1 Switch on
■ Click on the menu item Tools ▶ Gas flow control ▶ Gas flow
on.
2 Switch off
■ Click on the menu item Tools ▶ Gas flow control ▶ Gas flow
off.
Proceed as follows:
NOTICE
Proceed as follows:
1 Quit program
■ Close the program Rancimat (see chapter 4.1.1, page 25).
2 Optimize program
■ In the windows Start menu, click on the menu item Pro-
grams ▶ Metrohm ▶ Rancimat ▶ Optimize ▶ Program.
1 Export database
■ Export determinations not currently required from the database
Repos.mrd into a newly created database (see "Exporting deter-
minations to another database", page 134).
■ Close the Results window and the program Rancimat (see chap-
ter 4.1.1, page 25).
2 Optimize database
■ In the Windows Start menu, click on the menu item Pro-
grams ▶ Metrohm ▶ Rancimat ▶ Optimize ▶ Database.
Serial number
Serial number of the selected instrument.
Program number
Number of the EEPROM program of the selected instrument.
Last maintenance
Date of the last maintenance with signature of the service technician who
has carried out the service work.
Last calibration
Date of the last instrument adjustment with signature and code of the
responsible body.
NOTICE
Unit
Instrument number (1…4).
Block
Instrument block (A or B).
Channel
Channel (1…4).
Method
Name of the method loaded.
Sample ID1
Sample identification 1.
Status
Status of the channel.
ready: No active measurement. The channel is ready for starting a deter-
mination.
running: Instrument is measuring.
finished: Determination finished. The channel is ready for starting a new
determination.
error: Communication error between instrument and PC.
Stab. time
Stability time determined.
Induc. time
Induction time determined.
Set temp.
Setpoint temperature (defined in the method).
Current temp.
Current block temperature.
Proceed as follows to display and save all events of the Rancimats con-
nected to the PC:
NOTICE
You can find the meaning of the symbols and columns in the event
overview at "Meaning of the symbols and columns in the event
overview", page 48.
2 Update view
■ Click on the symbol or the menu item View ▶ Refresh to
view the current status of the log file.
2 Delete entries
■ Enter the maximum age in days, which the entries are allowed to
have in the list. All older entries are deleted.
■ Confirm the entry with [OK].
Type
- Information on correctly executed events.
Date
Date of the event. The format depends on the settings defined in Win-
dows in Control Panel ▶ Country settings ▶ Date.
Time
Time of the event. The format depends on the settings defined in Win-
dows in Control Panel ▶ Country settings ▶ Time.
User
Name of the user logged in at the time of the event.
Unit
Instrument number (1…4).
Event
Description of the event.
Temperature Conductivity
18 °C 127 µS/cm
19 °C 130 µS/cm
20 °C 133 µS/cm
21 °C 136 µS/cm
22 °C 138 µS/cm
23 °C 141 µS/cm
24 °C 144 µS/cm
25 °C 147 µS/cm
2
3
4
5
6
10
Figure 6 shows in detail how the accessories for determining Delta T are
mounted. Proceed as follows:
NOTICE
Create a unique method for both blocks, so that you can sepa-
rately enter the temperature correction for both blocks.
After you have prepared everything for determining Delta T (see "Prepar-
ing Delta T determination", page 53), you can now execute the determi-
nation of Delta T as follows:
NOTICE
You can find the calibration data (T1 to T3) for Metrohm sensors
on the certificate enclosed with the sensor or on the label fastened
on the sensor cable.
NOTICE
4.5 Methods
4.5.1 Managing methods
2 Adjust parameters
■ Adjust the required parameters.
■ Click on the tab Evaluation and also adjust the parameters.
■ Repeat for all remaining tabs.
NOTICE
You can find a detailed description of all tabs and their parameters
at 4.5.2 Parameter description.
3 Save method
■ Click on the [Save as...] button.
The dialog window Save method as... opens.
■ Enter the method name and confirm with [Save].
1 Open method
NOTICE
1 Open method
■ In the Control window, click on the symbol or the menu
item File ▶ Open method....
The dialog window Open method appears.
■ Click on the required method and confirm the selection with
[Open].
The dialog window "Method name" appears.
NOTICE
1 Open method
■ In the Control window, click on the symbol or the menu
item File ▶ Open method....
The dialog window Open method appears.
■ Click on the required method and confirm the selection with
[Open].
The dialog window "Method name" appears.
2 Print parameters
■ Click on the [Print...] button.
The dialog window Print Setup opens. All the printers installed
on your PC appear under Name.
■ Select the required printer and confirm with [OK].
Temperature
Setpoint temperature of the sample.
Input range 50 - 220 °C
Delta T
Temperature correction: indication of how much the block temperature
has to be increased so that the sample reaches the setpoint temperature.
This value is determined through measurement of the deviation by the
current temperature of the sample from the heating block temperature
using a calibrated, external temperature sensor (see chapter 4.4.2, page
51).
NOTICE
Gas flow
Gas flow through the sample. The gas flow can only be set for both
blocks jointly. The value of the method loaded for block A is always deci-
sive.
Input range 7 - 25 L/h
Start options
Start delay
Waiting time until starting the measurement.
Input range 0 - 1000 min
Start mode
Mode for starting and stopping the channels.
Selection per channel | per 2 channels | per block
per channel
All channels can be started and stopped individually.
per 2 channels
2 channels each can be started and stopped jointly. The [Start] button
in the control window changes automatically for this.
per block
The 4 channels of a block can be started and stopped jointly.
Stop criteria
Time
on | off (Default value: off)
Time until the automatic stop of the determination.
Input range 0.1 - 9999 h
Conductivity
on | off (Default value: off)
Conductivity limit value. If this value is reached, the determination is
ended automatically.
Input range 1 - 400 µS/cm
Endpoint(s)
on | off (Default value: off)
Automatic stop of the determination, as soon as every active channel has
reached all endpoints (induction time and stability time).
CAUTION
If the gas flow is automatically switched off, the reaction vessels must
be removed from the instrument within 30 to 60 min after the end of
the determination. If the vessels are not removed, a vacuum could
develop over the course of time, which would thus result in back suc-
tion of the measuring solution into the hot reaction vessel.
NOTICE
If you select this evaluation mode, you always need to calibrate the
conductivity measuring cell beforehand.
Delta Kappa
Conductivity change for the determination of the stability time.
Input range 1 - 400 µS/cm
Evaluation delay
Waiting time until starting the curve evaluation.
Evaluation suppression
Start
Time from the start of the determination to the start of the evaluation
suppression.
Input range 0 - 100 h
End
Time from the start of the determination to the end of the evaluation sup-
pression.
Input range 0 - 100 h
Evaluation sensitivity
This value defines the minimum value which the maximum of the curve of
the second derivative has to reach in order to be accepted as a result. The
result, i.e. the induction time, is the time until the first accepted maximum.
The default value "1" will lead to the correct result for most applications.
Smaller values will cause the acceptance of smaller maximums of the sec-
ond derivative. Higher values require a higher maximum.
Input range 0.1 - 9999
Induction time
The induction time is the time to the break point of the curve conduc-
tivity vs. time recorded with the Rancimat. The induction time is a parame-
ter for the oxidation stability of the sample examined and corresponds to
a large extent to the results from the complex AOM method.
For automatic determination of the induction time, the second derivative
of the measured curve is used, which exhibits a maximum at the break
point and which can be displayed in the window Live graph. To enable
detection of a break point, certain criteria must be fulfilled with regard to
the height and width of the peak in the second derivative. The break point
can also be determined manually as the intersection point of the two tan-
gents at the extended straight branches of the curve (see "Setting tan-
gents manually", page 117).
Evaluation suppression
Beginning End
Induction time
Time
With certain samples, it may occur that the conductivity rises in steps, to a
small degree, long before the actual induction time (e.g. through side
reactions or with volatile compounds). To prevent this rise being evaluated
as an endpoint, the user can delay the evaluation or suppress it for a cer-
tain range.
Stability time
The stability time is the time which is necessary in order to attain a speci-
fied conductivity change Delta Kappa.
Conductivity
Conductivity
Change
Stability time
Time
End value
End value of the time axis with manual scaling.
Input range 0.5 - 100 h
End value
End value of the conductivity axis with manual scaling.
Input range 10 - 400 µS/cm
AOCS Cd12b-92
Formula calculation according to standard AOCS Cd12b-92 with Tar-
get temperature = 110 °C.
AOM
Formula calculation according to standard AOM with Target temper-
ature = 97.8 °C.
free
Formula calculation with freely selectable Target temperature.
Formula parameters
Factor
Parameter for the calculation of the standard time, which indicates the
factor by which the time changes per 10 °C. This value can be determined
through the extrapolation (see chapter 4.7.4, page 118) of results, which
have been recorded at different temperatures.
Input range 0.001 - 100
Target temperature
Target temperature for the calculation of the standard time.
Input range 20 - 300 °C
Include in report...
Header
on | off (Default value: on)
Output of the header with company settings (see ""Company settings"
tab", page 138), date of the printout, sample identifications, instrument
information, determination date, user, cell constant and serial number.
Method parameters
on | off (Default value: on)
Output of the method parameters.
Curve(s)
on | off (Default value: on)
Output of the curves.
Method description
on | off (Default value: off)
Output of the method description entered on the tab Description.
Results
on | off (Default value: on)
Output of the selected results.
Selection Induction time | Stability time | Standard time |
Formula 1 | Formula 2 | Formula 3
Send to file
on | off (Default value: off)
Automatic output of determination and method data after finishing the
determination as TXT file for further processing in other programs. The
content of this report is independent of the report elements defined at
Include in report....
Browse…
Selection of the directory for saving this file.
ID 1 ID 2
Peanut oil DBG2307
Induction time 3.26 h
3.26
30
25
20
15
µS/cm
10
0
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5
h
Send to file
4.6 Determinations
Standard sequence Optional steps Further information
Section .4.6.2
Prepare instrument and accessories
Section 4.6.3
Select method
no
Temperature
reached?
yes
Determination finished
Section 4.6.5
Clean vessels and accessories
NOTICE
Use new reaction vessels and air tubes for each measurement.
Blow the reaction vessel out with nitrogen before use.
Oils and fats made from products containing oil and fat must be extracted
as follows with petroleum ether (low boiling point) protected from light:
NOTICE
2 Extraction
■ Prepare an Erlenmeyer flask 300 mL with normal grinding.
■ Weigh in between 50 and 100 g (depending on oil/fat content) of
the sample in powder form (e.g. milk powder, cocoa powder,
hazel nut powder) or the finely ground material.
■ Cover the sample material with a layer of approx. 1 cm of petro-
leum ether (low boiling point).
■ Extract the oil or fat while continuously stirring for around 12 h
and keeping it protected from light.
NOTICE
NOTICE
NOTICE
NOTICE
You can find further information on the treatment of oil and fat samples in
Rancidity in Foods, Allen J.C., Hamilton R.J., Applied Science Publishers,
London and New York, 1983.
You can use water-soluble fats as pure substances and prepare them as
follows:
NOTICE
NOTICE
Alternatively you can melt out the emulsion fats (e.g. butter, margarine) at
a temperature, exceeding the melting point of these fats by 10 °C (i.e. at
approx. +50 °C), centrifuge them and pipette off the resultant oil phase.
Check and prepare the instruments and vessels as follows (see also figure
5, page 14):
NOTICE
NOTICE
Use new reaction vessels and air tubes for each measure-
ment. Blow the reaction vessels out with nitrogen before use so
that any particles adhering are removed.
1 Select method
■ Under Unit x Block A, click on the symbol in order to select
the method for Block A.
■ If required, also select a method for Block B.
NOTICE
2 Start heating
■ Under Unit x Block A, click on the button [Heat to xx °C].
■ If required, also switch on the heating for Block B.
The color of the button frame changes to red and the temperature
display (2-3) on the instrument flashes. The current temperature is
displayed directly next to the button.
Heating up duration for 120 °C: approx. 45 min.
Heating up duration for 200 °C: approx. 60 min.
NOTICE
If you wish to turn off the heating, click on the button [Heat to
xx °C] again.
The entry for ID1 can be selected from the automatically created list
of already entered identifications. This list can be deleted again in the
Control window with the menu item Options ▶ Clear sample
ID1 history.
NOTICE
■ Seal the channels not used with the sealing plugs or empty reac-
tion vessels for protection against impurities.
■ Insert the prepared reaction vessels into the openings of the heat-
ing block.
■ Connect the white silicone tubings which are fastened to the
reaction vessel cover to the tubing adapter M8 / olive (5-11) of
the measuring vessel covers.
■ Screw the FEP tubings 250 mm (5-1) to the thread adapters (5-3)
of the reaction vessel covers and the air supply connections (3-3)
of the Rancimat.
1 Start determination
■ Start of an individual determination: Click on the button
[Start] of the corresponding channel.
■ Start of all channels of Block A: Click on the button .
■ Start of all channels of Block B: Click on the button .
After the determination has been started, the live curve is dis-
played for every channel:
NOTICE
NOTICE
Replace the reaction vessel covers if they are not tightly on the
reaction vessel anymore or the material is obdurate.
1 Stop determination
■ In the Control window, click on the button [Stop].
The message Do you want to save this determination? appears.
During a determination:
After a determination:
Determination finished.
The channel is ready for starting a
new determination.
4.7 Results
4.7.1 Determination overview
The determination overview displays selected information on all determi-
nations which are saved in the opened database, in tabular form. You can
adjust this overview, i.e. you can for example hide fields, adjust the col-
umn width or change the font type.
You can also filter and sort the content of the determination overview
according to specific criteria.
You will find step-by-step instructions on various actions, which you can
execute in the determination overview, below. You can also execute many
actions in addition to the described methods via the context menu.
Further below in this section you will find a detailed description of all
parameters of the dialog windows of the determination overview (see
"Parameter description", page 96).
2 Open database
■ In the Results window, click on the menu item File ▶ Open
database....
The dialog window Open database appears.
■ Click on the required database and confirm the selection with
[OK].
1 Select fields
■ Click on the symbol or the menu item Format ▶ Select
fields.
The following dialog window appears:
■ Position the cursor in the title line of the table between two col-
umns.
The cursor assumes the following form:
■ Drag the column width to the required size keeping the left
mouse button pressed down.
Selecting determinations
1 Select a line
■ Place the mouse button on the left edge of the line to be
selected.
The cursor assumes the following form:
■ Click on the line.
Sorting determinations
1 Select determinations
■ See "Selecting determinations", page 91.
With a special filter you can define complex conditions for filtering and
sorting data sets. In the following section, you will see a simple example
of a special filter. You will find a detailed description of all parameters
under "BestimmungsübersichtFilter", page 103.
NOTICE
The two fields are displayed in the lower part of the window under
Field:
NOTICE
Deleting determinations
NOTICE
Parameter description
Selection
Selection of all fields, which can still be selected for the determination
overview. The following fields of the determination overview are
available:
Send to file
Directory in which the determination file (TXT file) has been saved.
Evaluation sensitivity
Evaluation sensitivity used for the determation of the induction time.
User
Name of the user who was logged in when recording the determination.
Determination date
Date and time at which the determination was started.
Block
Letter of the block used for the determination.
Delta Kappa
Conductivity change for the determination of the stability time.
Gas flow
Gas flow through sample.
Unit
Number of the instrument used for the determination.
ID 1
Sample identification 1.
ID 2
Sample identification 2.
Channel
Number of the channel used for the determination.
Method name
Name of the method with which the determination has been carried out.
Standard time
Automatically calculated standard time.
Serial number
Serial number of the instrument used to record the determination.
Stability time
Automatically or, with the reevaluation, manually determined stability
time.
Temperature
Temperature as defined in the method.
Cell constant
Cell constant of the conductivity measuring cell used to record the deter-
mination.
Display
Fields which have been selected for the determination overview.
Selection Fields of the determination overview
Column width
Width of the column in characters.
Entry 'Number of characters'
Standard width
on | off (Default value: on)
Reset the column width to standard width (18.5 characters).
[Best Fit]
Adjust the column width to the longest field content (including title).
Gridlines Shown
Horizontal
on | off (Default value: on)
Display horizontal grid lines.
Vertical
on | off (Default value: on)
Display vertical grid lines.
Cell Effect
Selection Flat | Raised | Sunken
Flat
Flat view of the cells.
Raised
Raised view of the cells.
Sunken
Sunken view of the cells.
Gridline Color
Color of the horizontal and vertical grid lines.
Selection Color
Background color
Color of the cell background.
Selection Color
Preview
Display of the selected format for the cells.
Font
Selection of the font type from the fonts installed on the PC.
Selection Font type
Effects
Underline
on | off (Default value: off)
Choice of underlining for characters.
Color
Color for the selected font type.
Selection Color
Font style
Choice of font style for the selected font type.
Selection Standard | Italic | Bold | Bold italic
Sample
Display of the selected font type.
Size
Font size in points for the selected font type.
Selection Font size
Find What
Required search term.
Entry 'Search term'
The following characters can be used as wildcards:
*: any character sequence (e.g. *ab, *ab*, ab*)
?: single character (e.g. a?a, ??a, ab?)
#: single numeral (e.g. ##20, ab###, #23##)
Search
Search mode
Selection All | Down | Up
All
Search in the entire table.
Down
Search downwards as far as the table end.
Up
Search upwards as far as the table start.
Match
Search area where the entered text is to be searched for.
Selection Whole field | Start of field | Any part of field
Whole field
The search is only successful if the entire field matches the search term.
Start of field
The search is successful if the start of the field content matches the
search term.
Any part of field
The search is successful if the field content partly matches the search
term.
Match case
on | off (Default value: off)
If this option is activated, the search is only successful if the uppercase and
lowercase of the search term and field content match.
Field
Selection of a field (column) from the determination overview.
Selection Field from the determination overview
Sort
Sorting for the selected field.
Selection Ascending | Descending | (not sorted)
Criteria
Criteria for filtering data sets.
or
Further criteria for filtering data sets.
In the lines Criteria and or and in all following lines, data sets can be fil-
tered using expressions. An expression is a combination of operators,
functions and field names for determining a condition for filtering data
sets. In the following only the most important elements and rules are
described and a few selected examples are given. For more extensive
information please refer to the documentation on Microsoft Access.
Operators, functions, Mathematical +, –, =, *, /
field names operators
Comparison opera- <, >, <>, <=, >=, Between…, And…
tors
Logical operators AND, OR, NOT
Wildcards * (any character string), ? (single character), #
(single numeral)
Current date Date()
Current time Now() (date and time)
Other fields ["Field name"]
Links with various If conditions are specified in various fields, these conditions can also be
fields linked with AND and OR. The following applies:
Criteria in the same line apply as AND conditions, i.e. only those data sets
are displayed which satisfy all conditions.
Criteria in various lines apply as OR conditions, i.e. all data sets which fulfill
one condition are displayed.
Parameter description
ID 1
Sample identification 1.
ID 2
Sample identification 2.
Determination date
Date and time at which the determination was started.
User
Name of the user logged in when recording the determination.
Serial number
Serial number of the instrument used to record the determination.
Cell constant
Cell constant of the conductivity measuring cell used to record the deter-
mination.
GLP test
GLP status at the time of recording the determination. The following dis-
plays are possible:
GLP test switched off: The GLP monitoring was not activated for either
temperature or conductivity or gas flow.
GLP test due: The next GLP test was due.
GLP test not passed: The last GLP test carried out was not passed.
GLP test passed: The last GLP test carried out was passed.
Stability time
Automatically or, with the reevaluation, manually determined stability
time.
Delta Kappa
Conductivity change for the determination of the stability time.
Standard time
Automatically calculated standard time.
Method name
Name of the method.
Creation date
Date when the method was last saved.
Creator
Name of the logged in user who last saved the method.
Method description
Description of the method.
Temperature
Setpoint temperature of the sample.
Delta T
Temperature correction: indication of how much the block temperature
has been increased so that the sample reaches the setpoint temperature.
Gas flow
Gas flow through sample.
Start options
Start delay
Waiting time until starting the measurement.
Start mode
Mode for starting the channels:
per channel: all channels were started individually.
per 2 channels: 2 channels each have been started jointly.
per block: the 4 channels of a block were started jointly.
Stop criteria
Time
Time until the automatic stop of the determination.
Conductivity
Conductivity limit value. The determination was automatically ended upon
reaching this value.
Endpoint
Automatic stop of the determination, as soon as every endpoint (induction
and stability time) has been reached.
Delta Kappa
Conductivity change for the determination of the stability time.
Evaluation delay
Waiting time until starting the curve evaluation.
Evaluation suppression
Start
Time from the start of the determination to the start of the evaluation
suppression.
End
Time from the start of the determination to the end of the evaluation sup-
pression.
Evaluation sensitivity
Evaluation sensitivity used for the determation of the induction time.
Formula parameters
Factor
Parameter for the calculation of the standard time. It indicates by which
factor the time changes per 10 °C.
Target temperature
Target temperature for the calculation of the standard time.
Autoscaling
Automatic definition of the start and end value of the time axis.
Manual scaling
Manual definition of the end value of the time axis.
End value
End value of the time axis with manual scaling.
Autoscaling
Automatic definition of the start and end value of the conductivity axis.
Manual scaling
Manual definition of the end value of the conductivity axis.
End value
End value of the conductivity axis with manual scaling.
Report elements
Header
Output of the header.
Curve
Output of the curves.
Method description
Output of the method description entered on the tab Description.
Results
Output of the results selected below.
Method parameters
Output of the method parameters.
Results
Output of the selected results: Induction time, Stability time, Stan-
dard time, Formula 1...3.
Send to file
Automatic output of determination and method data after finishing the
determination as a TXT file.
Directory
Directory in which the report has been saved.
Description
The following warning messages are possible for incorrectly executed
determinations:
Temperature not reached: setpoint temperature at start of the determi-
nation not reached.
Missing measuring points (start time): measuring points are missing
from the given time (in s) onwards.
Missing measuring points (end time): measuring points are present
again from the given time (in s) onwards.
Manual Stop: the determination was stopped manually with [Stop].
Stop after communication error: Automatic stop of the determination
because of interrupted connection between the Rancimat and the PC.
Value
Value for warning (e.g. temperature, time).
NOTICE
1 Zoom graph
■ See "Zooming", page 32.
Proceed as follows to copy a graph to the clipboard and then paste it into
another program:
1 Copy graph
■ Right click in the dialog window of a graph.
The context-sensitive menu for graphs appears.
■ Click on the menu item Copy.
2 Paste graph
■ Open program (e.g. Word, Excel, any graphics program, etc.)
■ Paste the graph using the key combination [Ctrl] + [V].
Printing graphs
3 Save changes
■ In the dialog window Live graph click on the button [Save].
The new induction time is accepted in the determination overview.
In order to set the stability time manually, you need to adjust the conduc-
tivity change Delta Kappa. After this, the stability time is recalculated
automatically. Proceed as follows:
3 Save changes
■ In the dialog window Live graph click on the button [Save].
The new stability time is accepted in the determination overview.
For determining the induction time, you can manually set the tangents.
The induction time is then defined as the intersection of the tangents.
Set the tangents as follows:
5 Save changes
■ In the dialog window Live graph click on the button [Save].
The new induction time is accepted in the determination overview.
1 Delete tangents
4.7.4 Extrapolation
Extrapolation can be used to extrapolate the results measured at various
temperatures to the required target temperature. In addition, the factor
for converting the induction time to the standard time can be deter-
mined.
A detailed description of the parameters can be found at "Parameter
description", page 121.
NOTICE
NOTICE
NOTICE
The correlation limit for the correlation coefficient r2 can only be set
by an Administrator.
2 Adjust limit
■ Under correlation limit (r2) enter a value between 0 and 1 and
confirm with [OK].
See "Copying graphs and pasting them into other programs", page 116.
Parameter description
In the graphic temperature vs. time the linear regression is used to rep-
resent the calculated straight line T = ln(t)/B – ln(A)/B. The measured val-
ues of the selected determination are shown in black, the regression line
in green and the time calculated for the target temperature in red.
Results
Formula
Exponential formula for calculating the time extrapolated from the mea-
sured values for the target temperature.
A
Calculated coefficient A of the exponential formula.
B
Calculated coefficient B of the exponential formula.
r2
Calculated correlation coefficient. If this value is below the limit value (see
"Setting the correlation limit", page 120) specified by the administrator a
warning appears when opening the extrapolation window.
Time
Result of the extrapolation for the target temperature in hours and years.
Target temperature
Target temperature for the extrapolation.
Number of determinations
Number of determinations used for the extrapolation.
Standard factor
Value determined from the regression straight line which indicates the fac-
tor by which the time changes per 10 °C. This value can be entered in the
formula for the determination of the standard time as a factor (see "Cal-
culation of standard time", page 110).
[Settings]
Open the dialog window Program settings. The extrapolation parame-
ters are to be entered in the tab Extrapolation (see ""Extrapolation"
tab", page 140).
[Update]
Update graphics and result display. This function has to be carried out if
the selection of determinations in the determination overview or the set-
tings for the extrapolation have been changed.
[Detail]
Display only the area with the results of the determinations in the graph.
NOTICE
NOTICE
Recalculating formulas
NOTICE
NOTICE
Parameter description
Calculation active
on | off (Default value: on)
If this option is activated, the parameters can be re-entered for the calcu-
lation of the induction time.
If this option is deactivated, the original result is deleted.
Evaluation delay
Waiting time until starting the curve evaluation.
Evaluation sensitivity
This value defines the minimum value which the maximum of the curve of
the second derivative has to reach in order to be accepted as a result. The
result, i.e. the induction time, is the time until the first accepted maximum.
The default value "1" will lead to the correct result for most applications.
Smaller values will cause the acceptance of smaller maximums of the sec-
ond derivative. Higher values require a higher maximum.
The software suggests a value with which the highest maximum of the
second derivative is accepted as the induction time. If there is no sugges-
tion then the highest maximum has already been accepted.
Calculation active
on | off (Default value: off)
If this option is activated, Delta Kappa can be re-entered for the calcu-
lation of the stability time.
If this option is deactivated, the original result is deleted.
Delta Kappa
Conductivity change for the determination of the stability time in µS/cm.
Calculation active
on | off (Default value: off)
If this option is activated, the parameters for formula 1 can be re-
entered.
If this option is deactivated, the original result is deleted.
Calculation active
on | off (Default value: off)
If this option is activated, the parameters for formula 2 can be re-
entered.
If this option is deactivated, the original result is deleted.
Calculation active
on | off (Default value: off)
If this option is activated, the parameters for formula 3 can be re-
entered.
If this option is deactivated, the original result is deleted.
Calculation active
on | off (Default value: off)
If this option is activated, the parameters for the calculation of the stan-
dard time can be re-entered.
If this option is deactivated, the original result is deleted.
AOCS Cd12b-92
Formula calculation according to standard AOCS Cd12b-92 with Tar-
get temperature = 110 °C.
AOM
Formula calculation according to standard AOM with Target temper-
ature = 97.8 °C.
free
Formula calculation with freely selectable Target temperature.
Formula parameters
Factor
Parameter for the calculation of the standard time, which indicates the
factor by which the time changes per 10 °C.
Target temperature
Target temperature for the calculation of the standard time.
2 Print results
■ If required, adjust the column width (see "Formatting the deter-
mination overview", page 88).
NOTICE
In the overview table, only those columns are printed which are
within the maximum print width (15 cm).
2 Select determination
■ Select the determination of which the data is to be printed (see
"Selecting determinations", page 91).
3 Select data
■ Click on the symbol or the menu item File ▶ Print....
The dialog window Printing of... appears.
■ Select the option All determination and method data and
confirm with [Print].
The dialog window Print appears.
4 Print data
■ Select the required printer and click on [Print].
The determination and method data will be printed.
Printing graphs
1 Open graph
■ See "Single graph, multiple graph or live graph", page 114.
2 Print graph
■ Click on the symbol or the menu item File ▶ Print....
The dialog window Print appears.
■ Select the required printer and click on [Print].
The graph will be printed.
2 Select determination
■ Select the determination of which the report is to be printed (see
"Selecting determinations", page 91).
3 Select data
■ Click on the symbol or the menu item File ▶ Print....
The dialog window Printing of... appears.
4 Print report
■ Select the required printer and click on [Print].
The report will be printed.
2 Select determination
■ Select the determination of which the report is to be printed (see
"Selecting determinations", page 91).
3 Select data
■ Click on the symbol or the menu item File ▶ Print....
The dialog window Printing of... appears.
■ Select the option Determination according to settings.
5 Print report
■ Click on the [Print] button.
The dialog window Print appears.
■ Select the required printer and click on [Print].
The report will be printed.
1 Select determinations
■ See "Selecting determinations", page 91.
If you do not select any determinations, the entire overview is
exported.
2 Select determination
■ Select the determination of which the data is to be exported (see
"Selecting determinations", page 91).
3 Export data
■ Click on the symbol or the menu item Determina-
tion ▶ Export ▶ Determination and method data.
The dialog window Save to export file appears.
■ Select memory location, enter file name and confirm with [Save].
The determination and method data will be exported.
2 Select determination
■ Select the determination to be exported (see "Selecting determi-
nations", page 91).
3 Select database
■ Click on the menu item Determination ▶ Export ▶ Export to
database.
The dialog window Export data to... appears.
■ Select an existing database.
or
Under File name enter the name for the new database.
■ Confirm the selection or entry with [Save].
The following message appears (only for the Administrator):
NOTICE
2 Select determination
■ Select the determination of which the data is to be exported (see
"Selecting determinations", page 91).
3 Export data
■ Click on the symbol or the menu item Determina-
tion ▶ Export ▶ List of measured values.
The dialog window Save to export file appears.
■ Select the memory location, enter the file name and confirm with
[Save].
The measured values list will be exported.
You can carry out settings for the representation and output of results as
follows:
Header
on | off (Default value: on)
Output of the header with company settings (if defined, see "Company
settings" tab), date of the printout, sample identifications, instrument
information, determination date, user, cell constant and serial number.
Results
on | off (Default value: on)
Output of the results selected in the Results tab (see page 139).
Curve
on | off (Default value: on)
Output of the curves.
Method parameters
on | off (Default value: on)
Output of the method parameters.
Method description
on | off (Default value: on)
Output of the method description entered in the tab Description.
with offset
on | off (Default value: off)
If this option is activated, the curves are arranged with an offset above
one another in the case of an output of multiple graphs. Instead of the
usual lettering of the conductivity axis, the difference of the entire axis is,
in this case, printed in µS/cm.
[Company-logo…]
Open the window Find company logo for selecting the graphics file
with the company logo. Only WMF or BMP files can be used. After click-
ing on [Open], the logo is displayed in the adjoining window.
[Remove logo]
Remove the embedded company logo.
Logo on the...
Selection left side | right side
left side
The company logo is printed out on the left edge of the header.
right side
The company logo is printed out on the right edge of the header.
Company text
The text entered here is printed out on the side of the header opposite the
company logo.
Entry Characters
"Results" tab
On the Results tab, you can carry out the settings for the result output.
Induction time
on | off (Default value: on)
Stability time
on | off (Default value: off)
Standard time
on | off (Default value: off)
"Scaling" tab
On the Scaling tab, you can carry out the settings for scaling the curve
axes.
Acc. to method
Axis scaling in accordance with the settings defined in the method under
Curve representation (see chapter 4.5.2.3, page 68).
Autoscaling
Automatic definition of the start and end values of the curve axes.
Manual scaling
Manual definition of the start and end values of the curve axes.
X Start
Start value of the time axis in hours.
X End
End value of the time axis in hours.
Y Start
Start value of the conductivity axis in µS/cm.
Y End
End value of the conductivity axis in µS/cm.
"Extrapolation" tab
On the Extrapolation tab, you can carry out the settings for the extrapo-
lation.
Induction time
The induction time is used for extrapolation. If the induction time has
been determined manually, this value will be used and not the automati-
cally determined induction time.
Stability time
The stability time is used for extrapolation.
Target temperature
Temperature, for which the selected result is to be converted. The conver-
sion is realized using the regression straight line (from the extrapolation).
Input range freely selectable - °C
1 Open windows
■ Open all required dialog windows.
The following dialog windows can be opened in the Results win-
dow:
– Determination overview (View ▶ Determination over-
view)
– Graphs (Determination ▶ Graphics ▶ Single/Multiple/
Live graph)
– GLP window (View ▶ GLP)
2 Arrange windows
■ Click on the menu item Window ▶ Tile vertically or Win-
dow ▶ Tile horizontally or Window ▶ Cascade.
The opened dialog windows are correspondingly arranged.
The dialog window GLP properties for unit with serial no. x
appears.
2 Activate monitoring
■ In the Temperature tab, activate the option Enable GLP moni-
toring.
■ Carry out the remaining settings (see "Parameter description",
page 144) and confirm with [OK].
3 Deactivate monitoring
■ Open the dialog window GLP properties for unit with serial
no. x (see step 1).
■ In the required tab (Temperature, Conductivity, Gas flow),
deactivate the option Enable GLP monitoring and confirm with
[OK].
Parameter description
Tolerance
Indication of the tolerance value for the relevant GLP test. If the result is
within the range Setpoint ± Tolerance, the test is regarded as passed. If
you click on [Set default], the specified standard value is entered in this
field.
Parameter description
Test type
Type of GLP test (Temperature, Conductivity or Gas flow).
Due in
Number of days until the next GLP test.
5
Figure 7 Accessories for GLP test Temperature
5 Aluminum cylinder
Proceed as follows in order to prepare the GLP test set according to Figure
7:
1 Assemble accessories
■ Insert the temperature sensor from above into the opening of the
PTFE cylinder.
■ Place the PTFE cylinder on the aluminum cylinder in such a way
that the temperature sensor fits into this opening.
■ Push the temperature sensor into the aluminum cylinder as far as
the limit stop.
2 Screw on accessories
■ Use the hexagon key to lightly tighten the hexagon screw.
NOTICE
Before starting the GLP test, ensure that the temperature of the inserted
GLP test set is stable. If the GLP test set is used in a cold heating block,
this condition is fulfilled after the normal heating up time of the heating
block and the GLP test set. If the GLP test set, on the other hand, is
used in an already heated heating block, it is necessary to wait at least
30 min before starting the GLP test.
WARNING
The aluminum cylinder of the GLP test set is very hot. Do not
touch!
4.8.4.2 Conductivity
The GLP test for conductivity measurement can be carried out either
with a standard solution with a known conductivity or with the test
resistor 6.2109.030. The test resistor is part of the optionally avaiable
GLP test set . The test resistor must be connected during measurement at
the electrode connector (3-2) of the 743 Rancimat.
NOTICE
Before carrying out the GLP test with standard solution for the first
time, the cell constants of the conductivity measuring cells must be
determined (see chapter 4.4.1, page 49).
1 Select instrument
■ In the Control window, use the menu item View ▶ Unit x to
select the instrument for which the GLP test is to be carried out.
■ Carrying out the GLP test with the test resistor: continue with
step 2.
Carrying out the GLP test with the standard solution: continue
with step 3.
Temperature Conductivity
18 °C 127 µS/cm
19 °C 130 µS/cm
20 °C 133 µS/cm
21 °C 136 µS/cm
22 °C 138 µS/cm
23 °C 141 µS/cm
24 °C 144 µS/cm
25 °C 147 µS/cm
Carrying out the GLP test for the gas flow measurement
■ Click on the menu item Tools ▶ GLP test ▶ Gas flow... and
confirm with [Next >].
The following dialog window appears:
NOTICE
To be able to print a test certificate, you have to activate the GLP moni-
toring (see "Activating/Deactivating GLP monitoring", page 143).
For the selected instrument, you can print a GLP certificate, which con-
tains the following information:
Print date, serial number of the instrument, results of the last GLP tests
(with date, test person and test results).
NOTICE
If you want an overview of all GLP tests carried out with the connected
units, the database Repos.mrd must be open. If another database is
open, only those GLP tests are displayed which have also been exported
into this database during export of the determinations.
NOTICE
To be able to print out the GLP results, you have to activate the GLP
monitoring (see "Activating/Deactivating GLP monitoring", page 143).
2 Select data
■ In the Results window, click on the menu item File ▶ Print....
If the database contains GLP test results which have been attained
with different instruments, the dialog window Select unit
appears.
Parameter description
Printing options
Selection Last certificate | Certificate of date | All tem-
perature GLPs | All conductivity GLPs | All gas
flow GLPs
Last certificate
Printout of the last GLP certificate of the selected overview window.
The printout contains the Serial number of the instrument, the Date,
the Tester and the Test result.
Certificate of date
Printout of the GLP certificate for the entered date. The printout con-
tains the Serial number of the instrument, the Date, the Tester and
the Test result.
All temperature GLPs
Printout of the detailed results for all GLP tests for temperature mea-
surement.
All conductivity GLPs
Printout of the detailed results for all GLP tests for conductivity mea-
surement.
All gas flow GLPs
Printout of the detailed results for all GLP tests for gas flow measure-
ment.
CAUTION
NOTICE
The molecular sieve filled in the drying flask (3-11) serves for adsorption
of disruptive oxidizing gases as well as of the water from the sucked in air.
You can regenerate the molecular sieve in the drying oven at approx.
+140…+180 °C for 24 to 48 h. You can order a new molecular sieve
under the order number 6.2811.000.
CAUTION
Do not fill the hot molecular sieve directly into the drying flask after
regeneration, as otherwise the plastic filter on the filter tube will melt.
Wait until the molecular sieve has cooled down before filling.
6 Troubleshooting
6.1 Problems
Problem Cause Remedy
The pump is louder The air flow is blocked ■ Check dust filter (3-20) and, if necessary,
than normal. somewhere before or after replace.
the pump. ■ Check filter tube (4-4) on the drying flask
cover for blockages and, if necessary, free
with a gentle tap.
■ Examine FEP tubings (2-6), (3-13),(3-14)
and (3-17) for blockages and, if necessary,
replace.
No air flow can be The air supply is blocked. ■ Remove FEP tubing (5-1) from the thread
detected in the reac- adapter (5-3). A slight air flow must be dis-
tion vessel (it does cerned here. If this is not the case, please
not bubble), contact Metrohm Service.
although the pump ■ Check FEP tubing (2-6) for blockages. If
is running. necessary, clean or replace.
■ Check thread adapter (5-3) and air tube
(5-5) on the reaction vessel cover for
blockages. If necessary, clean or replace.
The FEP tubing for the air Check FEP tubing (2-6) for cracks, kinks etc. If
supply is defective. necessary, replace.
The FEP tubing for the air Tighten FEP tubing (2-6) on both sides.
supply is not connected
correctly.
The air tube (5-5) does not ■ Push reaction vessel cover (5-8) down as
immerse in the sample. far as the limit stop.
■ Use more sample.
The connection is leaking. Check silicone tubing (5-2) for leakages and, if
necessary, replace.
The reaction vessel cover ■ If the reaction vessel cover is oblique or not
(5-8) is not mounted cor- completely mounted, press down as far as
rectly or is too loose. the limit stop.
■ If the reaction vessel cover is loose on the
reaction vessel despite correct assembly,
the cover has to be replaced.
The connection is wrongly ■ Ensure that the PTFE tube (5-17) is moun-
connected. ted to the air supply at the opening In
(5-13) of the measuring vessel cover.
■ Ensure that the silicone tubing (5-2) is
mounted on the tubing adapter, which is
mounted in the opening In (5-13).
■ Ensure that the reaction vessel is connected
to the measuring vessel belonging to the
corresponding measuring place.
The conductivity is The conductivity of the Check the conductivity of the water used and,
displayed in the water used is too high. if necessary, use deionized water from another
upper part of the source.
Live window during
startup. The conductivity measuring Check cell constant of the measuring cell. The
cell is contaminated. cell constant should be between 1.0 and 1.2.
If this is not the case, the measuring cell must
be cleaned (see chapter 4.6.5, page 84).
The induction times The reaction vessels used ■ Free the reaction vessel of particles (dust,
are not reproducible are not clean. cardboard, glass fragments etc.) with com-
for multiple determi- pressed air before weighing in the sample.
nations. ■ Only use new, unused reaction vessels.
The reaction vessel cover ■ If the reaction vessel cover is oblique or not
(5-8) is not mounted cor- completely mounted, press down as far as
rectly or is too loose. the limit stop.
■ If the reaction vessel cover is loose on the
reaction vessel despite correct assembly,
the cover has to be replaced.
The connection to the Ensure that no air can escape through leaks
measuring vessel is not when transferring from the reaction vessel to
mounted correctly. the measuring vessel.
The stability times The cell constant was not ■ Determine cell constant (see chapter 4.4.1,
are not reproducible determined or does not page 49).
for multiple determi- correspond to the value ■ Ensure that the measuring cell is always
nations. entered. used in the same channel, so that the cell
constant determined once also corre-
sponds to the measuring cell.
■ Ensure that the measuring cell is not con-
taminated. If necessary, clean (see chapter
4.6.5, page 84).
The stability time is Delta Kappa is not defined Ensure that the value of Delta Kappa defined
longer/shorter than correctly. in the method is correct (see chapter 4.5.2.2,
expected. page 65).
The measurement The air supply for the mea- Loosen tubing adapter (5-11) on the measur-
curves are extremely suring solution is directed ing vessel cover, turn the PTFE tube (5-17) in
noisy. to the conductivity measur- such a way that the air is no longer directed to
ing cell. the measuring cell and fix in this position.
Gas bubbles adhere to the ■ Ensure that the measuring cell is clean and
conductivity measuring cell free of fat. If necessary, clean thoroughly
during measurement. (see chapter 4.6.5, page 84).
■ In some cases ultra pure water contains a
large proportion of dissolved air which
escapes during measurement. In this case,
degas the ultra pure water before the mea-
surement for 5 to 10 min in a vacuum.
The curve shows a Side reactions occur at the ■ In the method, use the option Evaluation
step which means start or during the mea- delay or Evaluation suppression (see
that the induction surement, which cause the chapter 4.5.2.2, page 65).
time is no longer conductivity in the measur- ■ Do not use Endpoint(s) as a stop criterion
determined cor- ing cell to rise. in the method, but instead a defined Con-
rectly. ductivity (e.g. 200 µS/cm) (see chapter
4.5.2.1, page 61). The curves can then also
be evaluated manually after finishing the
determination.
The curve shows a The reaction vessel cover ■ Thoroughly clean the reaction vessel cover
step at the start of (5-8) and/or the tubing and the silicone tubing (see chapter 4.6.5,
the measurement, (5-2) still contain residues page 84).
which has not occur- which are transported into ■ Replace the reaction vessel cover and the
red in previous the measuring vessel silicone tubing from time to time.
measurements. through the warm air flow
when a new measurement
is started.
The induction time The option Evaluate Activate the evaluation of the induction time in
is not evaluated induction time is deacti- the method (see chapter 4.5.2.2, page 65).
automatically, vated in the method.
although a signifi-
The option Evaluation Deactivate the corresponding option in the
cant break point can
delay or Evaluation sup- method (see chapter 4.5.2.2, page 65).
be noticed in the
pression is defined in the
curve.
method, which prevents
the evaluation of the curve
in the corresponding time
period.
The measurement The measurement has been Keep the measurement running until the end-
aborts without an stopped manually. point is automatically found.
endpoint being
found. A time or conductivity is Deactivate time or conductivity as a stop crite-
defined as a stop criterion rion in the method (see chapter 4.5.2.1, page
in the method, which was 61) or increase the corresponding value for the
reached before the end- time or conductivity.
point.
Error The heating is defective. Switch the instrument off and on again. If the
error occurs again, please contact Metrohm
Service.
The maximum temperature Switch the instrument off and on again. If the
(220 °C) has been error occurs again, please contact Metrohm
exceeded by 10 °C. Service.
Error Error during self test. Switch the instrument off and on again. If the
error occurs again, please contact Metrohm
Service.
7 Technical specifications
Test resistor 10 kΩ
7.9 Housing
Cover material Polyurethane hard foam (PUR) with flame retardation for fire class
UL94VO, CFC-free
Base material Steel, painted
Width 405 mm
Height 268 mm (without accessories)
353 mm (with accessories)
Depth 466 mm
Weight 20.2 kg (without accessories)
27.6 kg (with accessories)
8 Accessories
Up-to-date information on the scope of delivery and optional accessories
for your product can be found on the Internet. You can download this
information using the article number as follows:
2 Enter the article number (e.g. 743) into the search field.
The search result is displayed.
NOTICE
Index
A Determination .......................... 26 Filter
Access rights Delete ................................ 95 Determinations ................... 93
Adjust ................................ 36 Filter ................................... 93 Formulas .................................. 70
Manage ............................. 35 Prepare ............................... 82 Fuse
Accessories Recalculate ....................... 122 Replace .............................. 18
Mount ................................ 10 Search ................................ 92
Administrator password Select ................................. 91 G
Set ..................................... 22 Sequence ........................... 76 Gas flow
Air purification Sort .................................... 92 Method .............................. 63
Mount accessories .............. 10 Start ................................... 84 Range ............................... 170
Air supply external Stop manually .................... 86 Switch on / off .................... 40
Mount accessories .............. 13 Determination data GLP monitoring
Air supply internal Display ............................. 104 Activate / Deactivate ......... 143
Mount accessories .............. 10 Export .............................. 134 Display status ................... 145
Print ................................. 130 GLP test
C Determination overview Conductivity ..................... 151
Calibration data ........................ 54 Display ............................... 88 Gas flow ........................... 154
Cell constant Export .............................. 133 Results .............................. 156
Determine .......................... 49 Format ............................... 88 Temperature ..................... 146
Clean Print ................................. 130 Graph
Accessories ......................... 84 Device information ................... 43 Print ................................. 131
Measuring vessel ................ 84 Diagnosis ................................. 43 Group
Reaction vessel ................... 85 Drying flask Add .................................... 36
Tubings .............................. 85 Mount ................................ 12 Adjust access rights ............ 36
Computer Dust filter Delete ................................ 38
Connect ............................. 20 Mount ................................ 11
Conductivity measurement Replace ............................ 161 H
Measuring range .............. 170 Heating
Resolution ........................ 170 E Range ............................... 169
Connect Electrostatic charge .................... 6 Start automatically .............. 39
Power grid .......................... 19 Emulsion fat ............................. 77 Start manually .................... 82
Control window ....................... 26 Evaluation ................................ 65 Turn off .............................. 83
Menus ................................ 27 Event overview ......................... 46 Heating up duration ................. 83
Structure ............................ 26 Export Heating up time ..................... 170
Symbols .............................. 28 Determination and method
Curves ...................................... 68 data ................................. 134 I
Determination overview .... 133 Induction time .......................... 66
D Measured value list ........... 136 Live graph ........................ 115
Database To another database ......... 134 Multiple graph .................. 114
Open .................................. 88 External temperature .............. 170 Recalculate ....................... 122
Open in Microsoft Access 97 Extrapolation .................... 26, 118 Single graph ..................... 114
......................................... 135 Extrapolation results Instrument communication ....... 34
Optimize ............................ 42 Print ................................. 133 Instrument information ............. 43
Repos.mrd .......................... 29
Delta T F L
Determine .......................... 51 Fat Live graph
Method .............................. 62 Emulsion fats ...................... 77 Display ............................. 113
Derivative Non-liquid, pure fats ........... 77 Print ................................. 131
Second derivative ............. 115 File types .................................. 31 Log file ..................................... 46