CCT 32KW, 42KW, 50KW

CT X-RAY GENERATOR
SPELLMAN HIGH VOLTAGE ELECTRONICS CORPORATION

PAGE 1 OF 2

• 32, 42 and 50kW all in the same footprint and volume


• Utilizes Spellman’s own SPI interface (compatible
with other Spellman CT generators)
• 16/32 slice with 0.5 second rotation
• Supports Spellman’s add on Grid Box Option
• Integrated anode drive (supporting ball
or liquid metal bearings)

Spellman High Voltage has been designing, manufacturing GRID SPECIFICATIONS


and providing CT generators for OEM customers for over
three decades. We have unrivaled experience being the Optional grid box for
Deflection or Cutoff Grid
first provider of a CT generator used in a commercially
available continuous rotating CT scanner in 1985. Since
then Spellman has enhanced performance and developed Deflection Grid:
new CT generators year after year, setting the standard for The deflection grid (also called the Flying Focal Spot) allows for
CT generator technology. dynamic movement of the focal spot providing double X sampling
Spellman’s CCT series of CT generators provide a unique which increases effective resolution while decreasing unwanted
combination of performance and price. Ideally suited for the artifacts.
economy 16 slice CT scanner, this generator has the ability to Type: Deflection grid, single box assembly
be scaled from 32kW to 50kW. Its compact size and unique Input Power: 24Vdc, ±10% @ 4.2A maximum, negative
bend makes it suitable for small footprint gantry designs. input ties to chassis ground internally
Designed to easily pair with many off the shelf CT tubes. Connectors:
High Voltage Input:
OPTIONS
3 pin 75kV Federal Standard (Cathode, Large, Small)
DG Deflection Grid
CG Cutoff Grid High Voltage Output:
4 pin 75kV CA7 (Cathode, Large, G1, G2)
SPECIFICATIONS Number of Grids: 2 grids, switching out of phase
Input Voltage: Bandwidth: DC to 5kHz
Main: 400Vac, three phase, +15%/-20%, 50/60Hz Output Voltage: 0V to -4KV, from grid to cathode. User
Auxiliary: 220/230Vac, single phase, +15%/-10%, 50/60Hz input will command the bi-level grids to
Output Voltage: switch between adjustable voltages,
Range: 60kV to 140kV (bipolar, 30kV to 70KV) VLOW and VHI.
Accuracy: ±0.5% of setting The two grids switch out of phase: If G1 is
Output Current: at VLOW, G2 will be at VHI and vice versa.
Range: 10mA to 420mA, 50kW maximum. The VLOW and VHI levels for G1 and G2
10mA to 350mA, 42kW maximum. are independently adjustable from
10mA to 275mA, 32kW maximum. 0 to -4KV but VLOW range is constrained
by: |VLOW | ≤ | VHI |
Filament Output:
Type: Two filament outputs, large and small Cutoff Grid:
Voltage/Current:15Vac (30kHz-40kHz) @ 0-6A A cutoff grid truncates the high voltage at the end of the X-Ray
maximum, referenced to cathode output exposure, reducing patient radiation dosing that does not contribute
Power: to the imaging process.
Maximum mA: 420mA, 50kW maximum
Type: Cutoff grid, single box assembly
Maximum kV: 140kV, 50kW maximum
Peak: 50kW Input Power: 24Vdc, ±10% @ 0.75A maximum, negative
input ties to chassis ground internally
Maximum mA: 350mA, 42kW maximum Connectors:
Maximum kV: 140kV, 42kW maximum
High Voltage Input:
Peak: 42kW
3 pin 75kV Federal Standard
Maximum mA: 275mA, 32kW maximum (Cathode, Large, Small)
Maximum kV: 140kV, 32kW maximum High Voltage Output:
Peak: 32kW 4 pin 75kV CA7 (Cathode, Large, G1, G2)
Gantry Rotation: 0.5 seconds Number of Grids:
Starter: One, switched between “Zero” (0V≥Vg-c ZERO
Type: Dual Speed Starter ≥ -10V) and “Cutoff” (Vg-c CUTOFF = -4kV±10%
Compatibility: All industry standard X-Ray tubes @ steady state)

Corporate Headquarters
Hauppauge, New York USA
www.spellmanhv.com 128148-001 REV. A

+1-631-630-3000 FAX: +1-631-435-1620 Spellman High Voltage is an ISO 9001 and ISO 14001 registered company
e-mail: [email protected] Copyright © 2020 Spellman High Voltage Electronics Corp.
CCT 32KW, 42KW, 50KW
CT X-RAY GENERATOR
SPELLMAN HIGH VOLTAGE ELECTRONICS CORPORATION

PAGE 2 OF 2

CCT X-RAY GENERATOR DEFLECTION / CUTOFF GRID


DIMENSIONS: in.[mm] DIMENSIONS: in.[mm]

TOP VIEW TOP VIEW

11.50 [292.1] TO CG
10.77 [273.55]
0.312 [19]
MOUNTING
HOLE, 6PL 9.68 [245.87]

VOLTAGE

DANGER
HIGH
HV 9.75 [247.65]
OUTPUT
17.71 [449.8]

HV
OUTPUT

1.68 [42.7] 3.23 [81.91] 3.23 [81.91]


9.04 [229.5]
TO CG 11.14 [120]

SIDE VIEW

Ø0.375 [Ø9.5] 1.25 [31.75]


MTG HOLES 12PL
AIR
FLOW

4.43 [112.52]
SIDE VIEW

7.05 [179.2]

FRONT VIEW
10.25 [260.35]
2.13 [54.0] TO CG
9.75 [247.65]

24V IN CUSTOMER I/O


0.312 [19]
ALTERNATE
135° MOUNTING
HOLE, 4PL
5.49 [139.44]

2.14 [54.35]
17.13 [435.1]
INPUT 2.22 [56.38]
0.99 [25.14]

24.07 [611.4]

2.99 [75.94] 4.27 [108.45]

How to Order:
32kW maximum: PART NO.: CCT70PN32
42kW maximum: PART NO.: CCT70PN42
50kW maximum: PART NO.: CCT70PN50
Consult sales for grid options

Corporate Headquarters
Hauppauge, New York USA
www.spellmanhv.com 128148-001 REV. A

+1-631-630-3000 FAX: +1-631-435-1620 Spellman High Voltage is an ISO 9001 and ISO 14001 registered company
e-mail: [email protected] Copyright © 2020 Spellman High Voltage Electronics Corp.