Discovery Ct750 HD Lightspeed VCT: Installation Manual
Discovery Ct750 HD Lightspeed VCT: Installation Manual
Discovery CT750 HD
LightSpeed VCT
Installation Manual
(Book 2 of 2)
5308551-1EN
Rev. 5.0
GE HEALTHCARE
DIRECTION 5308551-1EN, REV. 5.0 LIGHTSPEED VCT/DISCOVERY CT750 HD INSTALLATION MANUAL
Page 200
GE HEALTHCARE
DIRECTION 5308551-1EN, REV. 5.0 LIGHTSPEED VCT/DISCOVERY CT750 HD INSTALLATION MANUAL
Chapter 1
Position Subsystems .............................................................................................29
Section 1.0
Installer/FE Notices ........................................................................................... 29
1.1 Shipping, Warehouse and Transportation Warning..........................................................29
1.2 International Shipments ....................................................................................................29
1.3 On Site Warning ...............................................................................................................29
1.4 Service Actions .................................................................................................................29
Section 2.0
Introduction ....................................................................................................... 30
2.1 Skill Set Required .............................................................................................................30
2.2 All Installers ......................................................................................................................31
2.3 Floor and Room Preparation ............................................................................................31
2.4 Overview...........................................................................................................................32
2.5 Pre-Installation Template..................................................................................................32
2.6 Required Common Tools and Supplies ............................................................................32
Section 3.0
Delivery Procedure............................................................................................ 35
3.1 System Transportation - Temperature Extremes..............................................................35
3.2 Stored Systems ................................................................................................................35
3.3 Construction Site Storage.................................................................................................35
3.4 Construction Site Installations ..........................................................................................35
3.5 Working with the Mover ....................................................................................................37
3.6 Damage In Transportation ................................................................................................38
3.7 A1 Breaker........................................................................................................................39
3.8 Installation Support Kits ....................................................................................................39
3.9 Installation Conditions.......................................................................................................40
Section 4.0
Layout the Floor Template ............................................................................... 41
4.1 Time & Personnel .............................................................................................................41
4.2 Tools and Test Equipment................................................................................................41
Section 12.0
Removing Table Shipping Dollies ................................................................... 78
12.1 Time and Personnel..........................................................................................................78
12.2 Tools and Test Equipment................................................................................................78
12.3 Preparation .......................................................................................................................78
12.4 Procedure .........................................................................................................................78
Section 13.0
Rear Entry Cable Box........................................................................................ 80
13.1 Rear Entry with Surface Floor Duct ..................................................................................80
Section 14.0
Install Table Footswitch Assembly.................................................................. 81
14.1 Time and Personnel..........................................................................................................81
14.2 Tools and Test Equipment................................................................................................81
14.3 Procedure .........................................................................................................................81
Section 15.0
Remove Gantry Tilt Bracket ............................................................................. 85
15.1 Time and Personnel..........................................................................................................85
15.2 Tools and Test Equipment................................................................................................85
15.3 Procedure .........................................................................................................................85
Section 16.0
Position the Power Distribution Unit............................................................... 86
16.1 Time and Personnel..........................................................................................................86
16.2 Tools and Test Equipment................................................................................................86
16.3 Procedure .........................................................................................................................86
Section 17.0
Install Operator Console (GOC6) ..................................................................... 89
17.1 Time and Personnel..........................................................................................................89
17.2 Tools and Test Equipment................................................................................................89
17.3 Procedures .......................................................................................................................89
Section 18.0
Seismic Mounting.............................................................................................. 94
18.1 Time and Personnel..........................................................................................................94
18.2 Tools and Test Equipment................................................................................................94
18.3 Procedures .......................................................................................................................94
18.4 Uninterruptible Power Supply (UPS) ................................................................................95
Chapter 2
Power, Ground & Interconnect Cables .................................................................97
Section 1.0
Introduction ....................................................................................................... 97
1.1 System Component Identification .....................................................................................98
1.2 Cable Color Identifiers ......................................................................................................99
Section 2.0
System Interconnect Diagram ........................................................................ 102
Section 3.0
Console Connections...................................................................................... 103
3.1 SCIM, Keyboard, Trackball and Mouse Installation
.............................................................. (On Lean Cart in the desktop component tray)104
3.2 Connecting the Media Tower
.......................................................... (On Lean Cart in the User Tower & Accessories)107
3.3 Optional MOD Drive ....................................................................................................... 107
3.4 Connecting the LCD Monitor.......................................................................................... 108
3.5 Power Panel Connections.............................................................................................. 110
3.6 Console Connections ..................................................................................................... 110
3.7 Install Drives................................................................................................................... 112
3.8 USB Barcode Reader Option ......................................................................................... 114
Section 4.0
Install Options.................................................................................................. 115
4.1 Install Optional Remote Monitor (on Lean Cart)............................................................. 115
4.2 Install Respiratory Gating Option (on Lean Cart) ........................................................... 115
4.3 Install Cardiac Gating IVY Monitor and Stand Option IVY 3150B (HD) ......................... 115
4.4 Install Injector Option (on Lean Cart) VCT /HD.............................................................. 115
4.5 Install IVY Monitor and Stand IVY A (on Lean Cart) VCT .............................................. 115
4.6 Customer Accessories (Head Holders and Extender) (on Lean Cart) ........................... 115
4.7 UPS Installation (on skid)............................................................................................... 115
4.8 Flat Tabletop used with Respiratory Gating ................................................................... 115
Section 5.0
Gantry Cable Connections.............................................................................. 116
5.1 Gantry Option Board CT750 HD Only............................................................................ 118
5.2 CT750 HD IPC Cable Connections................................................................................ 119
5.3 CT750 HD Cardiac Monitor Setup (IVY B)..................................................................... 120
5.4 VCT GOB and Option Panel Installation ........................................................................ 122
5.5 VCT Cardiac Monitor Setup and Installation (IVY Monitors) .......................................... 123
5.6 TGPU Connections ........................................................................................................ 124
Section 6.0
Table Connections........................................................................................... 125
Section 7.0
PDU Cable Connections & Configuration .....................................................126
7.1 Introduction to NGPDU .................................................................................................. 126
7.2 Panel - 380 - 480VAC Mains “TS1” Input Power Connection
completed by the customer electrician.127
7.3 Panel - Circuit Breakers ................................................................................................. 128
7.4 HVDC Connection.......................................................................................................... 129
7.5 440V Connection............................................................................................................ 129
7.6 Gantry & Console Power Connections (HD/GOC6)....................................................... 130
7.7 Console Power Cable Plug Removal ............................................................................. 131
7.8 Console Power Connection............................................................................................ 131
7.9 PDU Control Cable......................................................................................................... 132
7.10 System Ground Connection ........................................................................................... 132
Chapter 3
System Continuity & Ground Checks.................................................................137
Section 1.0
System Continuity (Mechanical Contractor)................................................ 137
1.1 Time and Personnel........................................................................................................137
1.2 Tools and Test Equipment..............................................................................................137
1.3 Procedure .......................................................................................................................137
Section 2.0
Site Ground Continuity Check ....................................................................... 140
Section 3.0
Shim Installations............................................................................................ 140
3.1 Time and Personnel........................................................................................................140
3.2 Tools and Test Equipment..............................................................................................141
3.3 Preparation .....................................................................................................................141
3.4 Procedure .......................................................................................................................142
3.5 Finalization......................................................................................................................143
Chapter 4
System Covers: Installation & Alignment ..........................................................145
Section 1.0
Process Overview ........................................................................................... 145
1.1 Gantry Front Cover Installation.......................................................................................146
1.2 Gantry Rear Cover Installation ......................................................................................148
1.3 Gantry Top Covers .........................................................................................................148
1.4 Gantry Side Covers ........................................................................................................149
1.5 Scan Windows ................................................................................................................149
1.6 Align Gantry Covers........................................................................................................150
1.7 Install Gantry Base Covers .............................................................................................152
Section 2.0
Install Console Covers.................................................................................... 155
2.1 Time and Personnel........................................................................................................155
2.2 Tools and Test Equipment..............................................................................................155
2.3 Preparation .....................................................................................................................155
2.4 Procedures .....................................................................................................................155
Section 3.0
.......................................................................................................................... 155
3.1 Install Panels ..................................................................................................................156
3.2 Re-install Side Panel ......................................................................................................156
Appendix A
Gantry Cover Removal and Dolly Setup ............................................................ 167
Section 1.0
Gantry Cover Removal .................................................................................... 167
1.1 Time & Personnel........................................................................................................... 167
1.2 Tools and Test Equipment ............................................................................................. 167
1.3 Procedures..................................................................................................................... 167
1.4 Gantry Rear Cover Removal.......................................................................................... 174
Section 2.0
CT UMI Tilting Dolly......................................................................................... 175
2.1 Installation Procedure .................................................................................................... 175
2.2 Tilting the Dolly............................................................................................................... 175
Section 3.0
Gantry Auxiliary (Mini) Dolly Installation....................................................... 176
3.1 Time & Personnel........................................................................................................... 176
3.2 Tools and Test Equipment ............................................................................................. 176
3.3 Procedures..................................................................................................................... 176
Appendix B
Pictorial Representation of Required Tools ...................................................... 181
Appendix C
Regulatory Clearance Quick Reference Guide.................................................. 185
Section 1.0
Regulatory Code Description ......................................................................... 185
Section 2.0
Terms and Definitions..................................................................................... 186
Section 3.0
Minimum Regulatory Working Clearance by Major Subsystem ................. 187
Section 4.0
Room Dimensions .......................................................................................... 189
4.1 Suggested Room Size ....................................................................................................189
4.2 Typical Room Size..........................................................................................................189
4.3 Minimum Room Size (Limited Access) ...........................................................................189
4.4 Regulatory Caution .........................................................................................................190
4.5 Operational Caution........................................................................................................190
Section 5.0
Egress ............................................................................................................. 190
Section 6.0
How to Measure............................................................................................... 191
Section 7.0
Minimum Room Layouts................................................................................. 192
Section 8.0
Suggested and Typical Room Layouts ......................................................... 194
Chapter 5
Electrical Introduction..........................................................................................213
Section 1.0
Installer/FE Notices ......................................................................................... 213
Section 2.0
Introduction ..................................................................................................... 213
Section 3.0
Review Mechanical Hand Off Material........................................................... 213
Section 4.0
Calibration Training Requirements ............................................................... 213
Section 5.0
Required FE Common Tools and Supplies................................................... 213
5.1 Cleanliness .....................................................................................................................214
Section 6.0
Requirements/Assumptions........................................................................... 214
Section 7.0
FE Workflow..................................................................................................... 214
Section 8.0
Overview for Completing Installation............................................................ 215
8.1 System-Level Tasks .......................................................................................................215
Chapter 6
Electrical Integration and Safety Verifications .................................................. 219
Section 1.0
Computer Integration and Configuration ...................................................... 219
1.1 Introduction and Flowchart............................................................................................. 219
1.2 Inventory System Software and Restore System State ................................................. 220
1.3 Determine System Configuration ................................................................................... 223
1.4 Configure Site Specific Set Up ................................................................................... 225
1.5 Set Time and Date ......................................................................................................... 237
1.6 Install Customer Options................................................................................................ 237
1.7 Network and Service Integration and Checkout............................................................. 241
1.8 Camera .......................................................................................................................... 242
1.9 Save System State and Start Up Applications ............................................................... 243
Section 2.0
Table Gantry Integration .................................................................................244
2.1 Introduction .................................................................................................................... 244
2.2 Check Alignment Lights ................................................................................................. 245
2.3 Autovoice/Intercom Checks ........................................................................................... 246
2.4 CT System X-Ray ON Indicators, Cautions & Warning Labels...................................... 248
2.5 Check Warning Labels ................................................................................................... 249
2.6 Process Product Locator Cards ..................................................................................... 249
2.7 Install Service Cabinet ................................................................................................... 250
2.8 Check X-Ray Lights ....................................................................................................... 251
2.9 Table Height Characterization........................................................................................ 251
2.10 Interference Test ............................................................................................................ 252
2.11 Gantry Tilt Verification Test............................................................................................ 253
2.12 Leakage Current Test .................................................................................................... 257
Chapter 7
Image Quality........................................................................................................ 259
Section 1.0
Image Quality Process Overview Flowchart ................................................. 259
Section 2.0
Reference Procedure - Scanning w/Service Protocols ................................ 260
Section 3.0
Tube Warm Up and Fast Cal........................................................................... 260
Section 4.0
Table/Gantry Alignment Procedure ............................................................... 260
4.1 Time & Personnel ...........................................................................................................260
4.2 Tools and Test Equipment..............................................................................................261
4.3 Preparation .....................................................................................................................261
4.4 Procedure .......................................................................................................................261
4.5 Tilt Alignment Tool Procedure ........................................................................................262
Section 5.0
Prepare the QA Phantom................................................................................ 268
Section 6.0
Center Phantom............................................................................................... 268
6.1 Required Tools ...............................................................................................................268
6.2 Procedure .......................................................................................................................268
Section 7.0
Tomographic Plane Indication ....................................................................... 269
Section 8.0
Image Quality Test .......................................................................................... 271
8.1 Preparation .....................................................................................................................271
8.2 Procedures .....................................................................................................................271
Section 9.0
System Functional Test .................................................................................. 271
9.1 Preparation .....................................................................................................................271
9.2 Procedure .......................................................................................................................272
9.3 Cardiac Functional Test..................................................................................................273
9.4 Connect Pro Functional Test ..........................................................................................274
9.5 Finalization......................................................................................................................274
Section 10.0
Save System State .......................................................................................... 274
Chapter 8
Customer Options Installation & Verification ....................................................275
Section 1.0
DICOM Network ............................................................................................... 275
1.1 Introduction .....................................................................................................................275
1.2 Preparation .....................................................................................................................276
1.3 Procedures .....................................................................................................................276
1.4 Declaring the Scanner on Advantage NET Protocol Devices/Systems ..........................283
1.5 Declaring the Scanner on DICOM Protocol Devices/Systems .......................................283
Section 2.0
DICOM HIS/RIS Setup ..................................................................................... 285
2.1 Prerequisites...................................................................................................................285
Chapter 9
System-Level Safety Tests .................................................................................. 293
Section 1.0
Patient Touch Leakage Test ........................................................................... 293
1.1 Time and Personnel ....................................................................................................... 293
1.2 Tools and Test Equipment ............................................................................................. 293
1.3 Safety ............................................................................................................................. 293
1.4 Preparation..................................................................................................................... 293
1.5 Procedures..................................................................................................................... 294
1.6 Ground Resistance Checks Procedure (done during mechanical install) ...................... 297
1.7 Finalization ..................................................................................................................... 297
Section 2.0
CT System Chassis Leakage Test.................................................................. 298
2.1 Time & Personnel........................................................................................................... 298
2.2 Tools and Test Equipment ............................................................................................. 298
2.3 Safety ............................................................................................................................. 298
2.4 Procedures..................................................................................................................... 298
2.5 Finalization ..................................................................................................................... 300
Appendix D
System Configuration Data Sheets .................................................................... 301
Section 1.0
Requirements................................................................................................... 301
Section 2.0
Manual Film Composer Options..................................................................... 301
Section 3.0
System Network Configuration ...................................................................... 302
Section 4.0
Network Application (Image Transfer) Configuration .................................. 303
Section 5.0
Camera Application Configuration ................................................................ 303
Appendix E
Informational Support Procedures ..................................................................... 305
Section 1.0
Symbols........................................................................................................... 305
Section 2.0
ESD Grounding Points................................................................................... 307
Section 3.0
Manually Mount/Unmount a CD .................................................................... 310
3.1 Introduction .....................................................................................................................310
3.2 Procedures .....................................................................................................................310
Chapter 5
Electrical Introduction
TOOL COMMENT
Standard FE Tool Kit See new tool list
Fluke 87 DVM or equivalent must be calibrated yearly.
AC Clamp-on amp meter must be calibrated yearly.
Dale 600 or 601 must be calibrated yearly.
QA Phantom and phantom holder
Lockout/Tagout kit, or equivalent
Safety Glasses
Table 5-1 FE Tools and Supplies
5.1 Cleanliness
The system must not be located in, near, or around construction. The room should be clean and
clear of construction dust and installation materials. Do not power on the system if requirement in
Book 1, Chapter 1, Section 3.13
Complete the tasks listed and check the appropriate box on the GE e-4879 form (explained in
Section 8.0 on page 215).
8.1.1 General
HVAC system is operational and environmental data reported on the GE e-4879 form.
System realignments completed, if required.
Broadband installed and operational.
Power and ground audit completed.
8.4 Options
Check the appropriate box on the GE e-4879 form to verify the installation and proper functionality
of all customer-ordered options.
Injector installed and operational.
Advantage Windows Workstation installed and functional tests completed.
Advantage 4D installed and functional tests completed.
Filming/Camera/DASM installed and operational.
Modem installed and functional tests completed.
UPS installed and functional tests completed.
Network items installed and functional tests completed.
GE e-4879 completed; see Section 9.0. (Required for installations in ALL countries.)
FDA 2579 completed; see Section 9.0. (Required ONLY for U.S. installations.)
Any PQRs or PSRs encountered have been reported.
All FMIs for system completed, if necessary.
All dispatching activities (03-04-10 codes) completed.
Customer acceptance checks completed.
System transfer completed and appropriate GE Healthcare personnel notified.
All outstanding customer installation issues have been addressed.
Chapter 6
Electrical Integration and Safety Verifications
This section describes the reconfiguration, system state restore, options installation, and monitor
adjustment procedures.
Start
Restore Camera
System State
Determine Save
System Configuration System State
Reconfigure
Operator’s Console Adjust
Monitor
Set
Time & Date
End
Install Customer
Options
Locate the box with the system software and option DVD disks, system order sheets, product
locator cards, and system reconfig DVD. This tray is in the top shelf on the Lean cart labeled
“Software. If not Lean packed, locate these items.”
You should find the following software CD documents:
• System Operating Software CD set
• Learning and Reference Guide
• Tip Simulator
• Advanced Applications
• Service Information
Note: There may be other items in addition to those above.
1.2.1.3 Preparation
Your system should have a system state DVD, located in the software box on the Lean cart.
The system state DVD contains:
• Characterization
• Calibrations
• Gen Cal
• Other Data
If you cannot locate the shipped system state DVD and your console data is not present, you must
do a complete recalibration of your system. If the system data is present and your Save State disk
is missing, complete a Save State now.
1.2.1.4 Procedures
The installation process uses all the system state files. At this time, use the system state DVD to
restore all files.
1.) If you are not on the Service Desktop, click the SERVICE DESKTOP Icon.
2.) Click the UTILITIES icon.
3.) Select SYSTEM STATE.
4.) Insert the DVD into the DVD drive.
5.) Select CHARACTERIZATION AND CAL.
7.) Select RESTORE to restore the system characterization and phantom calibration files to the
system.
Restore State can take as long as ten minutes, although typical times average about three
minutes. When Restore State completes, dismiss the tool, and proceed to the next section.
If any error should occur during the restore process, see the Software Load Procedure manual
(Load From Cold) for information regarding possible error messages and their recovery.
8.) Click NO for the Reset Scan Hardware popup message.
9.) Click DISMISS.
1.3.1 Preparation
For convenient removal and use during installation, System Configuration Data Sheets are located
in Appendix D on page 301.
Below is a summary of key information, some of which is required from the customer, to complete
system configuration. When gathering this information, refer to Configure Site Specific Set Up on
page 225.
System File Information:
• Hospital name (Ask the customer for ALL related fields.)_____________________________
• Service ID________________________________________________________________
Patient Info:
• Next MOD#_______________________________________________________________
• Exam #, Diagnostic # 50000 default____________________________________________
• Click YES to regenerate database.
• Click NO for Mobile System.
• HIPAA ___________________________________________________________________
Preference File Information:
• Doctor’s title_______________________________________________________________
• Date Format_______________________________________________________________
• Time Format______________________________________________________________
• Language type_____________________________________________________________
• Selected Fast Cal KV’s – default - ALL unless instructed otherwise_____________________
• Dose Information – default - Unless instructed otherwise_____________________________
• Dicom – default - Unless instructed otherwise_____________________________________
Hardware File Information:
• Select table type GT 2000 or GT 1700__________________________________________
• Default for all others
• Network printer – default_____________________________________________________
Network file Information:
• Suite Name – (from FE or hospital)_____________________________________________
• Host Name – (from FE or hospital)______________________________________________
• IP Address________________________________________________________________
• Net Mask _________________________________________________________________
• Broadcast Address__________________________________________________________
• Default Gateway___________________________________________________________
• Advanced options – default - Unless instructed otherwise by the FE____________________
____________________________________________________________________________
Note: The document collector box that arrived with your system contains the Software Installation
Procedures manual, which documents the reconfiguration procedure in more detail.
1.4.1 Preparation
On the following screens, you should make the changes necessary, pressing the corresponding
button at the top of the screen to move from screen to screen. When you are done, you can either
press the ACCEPT button to start the reconfiguration process, or press the QUIT button to exit
without changing the system configuration.
While the reconfiguration is going on, messages are displayed in a shell window that closes when
reconfiguration is complete. Should you later want to review the reconfiguration output, it is logged
in:
/var/adm/install.log.YYYYMMDDWWWHHMMSS
Where
YYYYMMDDWWWHHMMSS is the Date/Time that the reconfiguration was started.
To view the file, type: more /var/adm/install.log.YYYYMMDDWWWHHMMSS
It is possible to abort the reconfiguration while entering information on the reconfiguration screens.
Press the QUIT button at the top of the screen. There is NO safe way to abort the reconfiguration
after pressing the ACCEPT button. If the entries made in the screens were incorrect, DO NOT try
to stop the reconfiguration, instead wait for it to complete, and rerun reconfig, entering the correct
parameters.
1.4.2 Procedure
1.) Shut down applications from the Service Desktop.
2.) In an xterm window, log in as root:
a.) Type: su - ENTER
b.) Type the root password; press ENTER
3.) Launch the Install utility:
Type: reconfig ENTER at the prompt.
The OC displays the Install Utility Window as shown in Figure 6-3.
Comment: The following pages show the screens that are used to change the configuration of the system.
These screens are the same as those used for the Software Configuration during Load From Cold.
The actual screens vary depending on the current configuration of your system.
4.) Click the CONFIG button.
The OC displays the System Configuration - System Settings Screen as shown in Figure 6-4.
ID Item Description
a Hospital Name Configures the name that appears on images produced by this
scanner.
Example: St Marys Hospital
b Service ID Issued by the service organization.
Example: 262785CT2 (no spaces)
c Time Zone Selects the time zone for this site.
d Next Patient Exam # Configures the next exam number the scan user interface uses.
e Next Diagnostic Exam # Customer-selected; configures the next exam number the scan
user interface uses.
f Mobile System Indicates to the software if this CT is in a mobile environment or
not.
Table 6-1 System Setting Screen
ID Item Description
g Regenerate Database Determines whether the scan database is regenerated during
reconfiguration.
h Recreate Scan Disk Determines whether the Scan Array is recreated during
Array reconfiguration. Used only after HSDA Assembly replacement or
multiple Hard Disk Drive failures
i Rebuild Scan Disk Array Used only if replacing a single hard Disk Drive in the High Speed
Disk Array (HSDA) of GOP6 operator consoles. When a new hard
drive is installed in HSDA and the reconfig utility is executed, a
new set of buttons are displayed to allow the inclusion of the new
hard Disk Drive into the array.
This is not displayed during normal operation; it is displayed only
after a new HDD is installed in the HSDA, and the HSDA
recognizes it.
Table 6-1 System Setting Screen
VCT Identification
ID Item Description
a Doctor’s Title Title of the doctor (e.g. radiologist)
b Units for Patient Weight Identifies to the software if this site uses pounds or kilograms.
c Language Selects the language to display on the Application screens.
d Autovoice Language Configures the language heard in the scan room.
e Keyboard Selection Configures the language specific keyboard character set.
f Date Format Configures the format to display the date on the images.
g Time Format Configures the format to display the time on the images.
h Modified in Room Start Be sure OFF is selected.
If this site is in Japan, select ON.
i HIPPA Present Be sure OFF is selected, unless told differently
j Dose Information Display Option for the site to use in monitoring calculated patient dose.
k Preferred Fast Cal kV Configures the preferred kV that the Fast Cal routine will
calibrate (80, 100, 120, 140 in the Selected Preferred Fast Cal
kV field) Default is ON for VCT systems.
l Target Noise Index Table Select Table 2.
Table 6-2 Preferences Settings VCT
HD example below:
HD Identification
ID Item Description
a Doctor’s Title Title of the doctor (e.g. radiologist)
b Units for Patient Weight Identifies to the software if this site uses pounds or kilograms.
c Language Selects the language to display on the Application screens.
d Autovoice Language Configures the language heard in the scan room.
e Keyboard Selection Configures the language specific keyboard character set.
f Date Format Configures the format in which the date will be displayed on the
images.
g Time Format Configures the format in which the time will be displayed on the
images.
h Modified In Room Start Be sure OFF is selected, unless the site is in Japan, in which
case, this feature should be ON.
i HIPAA Present Be sure OFF is selected unless told differently.
Table 6-3 Preferences Settings HD
HD Identification
ID Item Description
j Dose Information Display Option for the site to use in monitoring calculated patient dose.
Use the default selection unless told differently. Select ON (full
CTDiw Display); Select ON WITHOUT TOTAL DLP (no Dose
Length Product Display); Select OFF (no CTDiw Display or Dose
Report, series 999 created)
k Dose Record Configures support for DICOM Dose SR Record option for
saving dose information with study. Default is OFF. The dose
information is saved in a DICOM structured report. Select ON -
Saves the dose information. Select OFF- turn off the option.
Select FULL- Save the does information in a DICOM X-RAY
Radiation Dose SR SOP Class
l Preferred Fast CAl KV Configures the preferred kV that the Fast Cal Routine will
calibrate. Defaulted ON for HD systems.
m Target Noise Index Table Be sure Table 2 is selected.
n Gantry Layout Configures the preference for Patient loading. Choose correct
orientation depending on site specific Gantry layout. Default is
ON RIGHT.
o Flip and Rotate Configures the preference for allowing the Flip and Rotate
feature to be turned on in the User Interface on the (Left) SCAN
monitor. Default is OFF.
Table 6-3 Preferences Settings HD
Note: This preference shall not be enabled unless specifically requested by the Customer and
Evaluation of Dose SR Compatibility functional check procedure has been executed and
indicates that the other hospital systems support the Dose Report SOP classes!
l.) Select the Preferred kVs for Fast Cal.
These kVs should include all kVs that the site uses for patient scanning. The default
selections are 80, 100, 120, and 140. Use the default selection unless the customer
prefers others.
m.) Verify that Table 2 is selected for Target Noise Index
n.) Choose the correct orientation for Gantry Layout depending on the orientation of the
gantry table as viewed from the operator’s console. Default is ON RIGHT.
o.) Flip and Rotate: Configures the preference for allowing the Flip and Rotate
feature to be turned on in the User interface on the (Left) SCAN Moniotr. Default is OFF.
This preference allows the Customer to apply custom orientation changes based on
Exam Type and reconstructions methods on the DICOM images that will be transferred
to PACS and related systems.
Note: This preference shall not be enabled unless specifically requested by the Customer and
Evaluation of Image Flip and Rotate Compatibility functional check procedure has been
executed and all DICOM test images pass orientation check!
8.) Click the HARDWARE button to display the Hardware Settings Screen for VCT see Figure 6-
7 for CT750 HD see Figure 6-8
VCT
ID Item Description
a Gantry Type Indicates the type of gantry installed with this system.
b Tube Type Indicates the type of X-ray tube installed in this system.
c DAS Type Indicates the type of DAS installed in this system
d Table Type Select the table type:
VT 1700 (GT Long)
VT 2000 (GT Short)
e PDU Type Indicates the type of PDU installed in this system.
f Scan Recon Indicates the number of VeRBs installed.
Hardware
g Number of VeRBs Indicates the number of VeRBs installed in this system.
Table 6-4 Hardware Settings
HD Example Below.
CT750 HD
ID Item Description
a Gantry Type Indicates the type of gantry installed with this system.
b Tube Type Indicates the type of X-ray tube installed in this system.
c DAS Type Indicates the type of DAS installed in this system
Table 6-5 Hardware Settings
CT750 HD
ID Item Description
d Detector Type Select the detector type: Colorado64 for HD
e Table Type Select the table type:
VT 1700 (GT Long)
VT 2000 (GT Short)
f PDU Type Indicates the type of PDU installed in this system.
g Scan Recon Indicates console type.
Hardware
h Number of VeRBs Indicates the number of VeRBs installed in this system.
Table 6-5 Hardware Settings
ID Item Description
a Suite Name The name this site is using on the system to identify the CT suite.
b Station Name
c Host Name Identifies the network hostname and AE Title of the CT system to
the hospital’s network
d IP Address Hospital’s IP Address for the system.
e Net Mask Hospital-provided; used if the CT system is on a subnet
f Broadcast Address Same as the IP Address, except the last digit group is set to 1’s
or 0’s, depending on the network configuration.
g Default Gateway Hospital-provided.
h AW DirectConnect Enable if the option is provided with the system.
i NIS Domain Name: Customer-provided site domain name.
j IP Address The IP Address for the NIS Server, if used. Hospital-provided.
k Enable Network Time Hospital decision.
Protocol
l Change DARC Subnet Hospital decision.
Table 6-6 Configure Network Settings
14.) When the configuration changes are complete, the system displays a prompt to reboot. Click
on YES. (See Figure 6-11.)
15.) The system automatically logs in as ctuser after the reboot. Select OK on the Autostart
Disabled popup message.
Important: You must set the time and date on the Host Computer with Application Software down.
1.) Open a Unix Shell and log in as root:
a.) Type: su - ENTER
b.) Type the root password; press ENTER
2.) Set the date and time for your time zone by updating the fields in the setdate routine.
a.) Type the following: {root@hostname}# setdate
Note: Type q to quit at any time; press ENTER to proceed.
To be accurate, this tool prompts you to enter the Second. Watch your clock or PC carefully to enter
the proper value, and press ENTER at the right second to set the accurate time. Press ENTER to
proceed.
b.) Type the current Year (1980-2030).
c.) Type the current Month (1-12).
d.) Type the current Day (1-30).
e.) Type the current Hour (Military time, 0-23).
f.) Type the current Minute (0-59).
g.) Type the current Second (0-59).
\Updating the time on the OC and DARC, Please Wait…
Ping darc (172.16.0.2) 56(84) bytes of data.
3.) Boot system to application level.
1.6.3 Preparation
Note: • Your system has one or more DVDs that contain customer-purchased options. Standard
options are required for system operation. Install the Options DVD(s) at this time, following the
instructions in this section. A more complete listing of options are located on the Service CD-
ROM shipped with your system.
• The following tables are options clarification and identification for installation purposes only.
• Standard options are loaded before additional options. Load the standard options in the order
listed in Table 6-7.
• Customer software options are located in the Option and Documentation Tray on the lean cart.
Additional options must be loaded in the following order for proper operation.
Marketing Name Installed Options Comments
Neuro 3D Filter NeuroFilter
AutoFilter-and-Transfer AutoFilter-and-Transfer
HD option (64 slice) Patient-64-Slice
HD Hi-Power Option HD Hi-Power Option
Smart Score Pro SmartScore Pro Requires AutomA for EKG
Monitor
Required for Cardiac options
CardIQ Snapshot CardIQ SnapShot
.35 Second Rotation Sub-0.4-Second-Scan Requires CardIQ SnapShot
ECG Trace OC EKG Viewer Requires CardIQ SnapShot
Cardiac Enhancement Filter Noise Reduction Filter Requires CardIQ SnapShot
SnapShot Pulse CardIQSnapShot-Cine
DentaScan DentaScan
CTPerfusion3Neuro CTPerfusion3Neuro
Volume Viewer VolumeViewer Install before DMPR
CardIQPlus OR CardIQXpressPlus OR
Cardiac Pro on OC CardIQ2XpressPro Requires CardIQ SnapShot
Card EP CardEP
AdvVesselAnalysis AdvVesselAnalysis
Auto Bone AutoBone
CTCBase CTColonoBase load only one
CTCPlus CTColonoPlus
CTC PRO on OC CTColonoPro
SmartStep SmartStep
Volume Shuttle AxialShuttle
Table 6-8 Additional Options
1.6.4 Procedures
1.) Locate all of the option DVDs.
2.) Ensure that the Options DVD(s) are NOT write-protected at this time. The initial install requires
that the DVD be write-enabled; subsequent installs can be done with the DVD write-protected.
3.) Inventory product locator cards. There should be approximately 28 cards depending on the
options ordered. Use the sales order sheet that comes with the system to verify that all sold
items are present.
4.) If you are not on the Service Desktop, click the SERVICE DESKTOP icon.
5.) Click the CONFIGURATION icon.
6.) Click INSTALL OPTIONS.
A blank Options window is displayed (Figure 6-12).
7. Click INSTALL. A Select Mechanism (Figure 6-13) prompt appears. Click the button for the
mechanism through which Option Keys are installed.
click OK anyway, wait for the abort pop up, then abort the process.) Available options on the
DVD appear on the Software Options window. (See Figure 6-15.)
11.) Select each option in the left-hand column in the order given in Table 6-7 to install the
corresponding software. The options must be loaded in the order given in Table 6-7.
Note: Some options require you to complete the data field(s). Refer to the inventory sheet for the options
with data requirements. Have this information available when completing this section.
• Check the FDO to see what options were ordered.
• Compare FDO options to those on the Options DVD.
• If different, contact your local sales representative.
12.) Select INSTALL. A box may be displayed while the options are loading. When an option is
displayed in the Installed Options list on the right side, then installation of that option is
complete. Note that some options take a fraction of a second to install, while options like 3D
may take a half- minute.
13.) When all options have been installed, check the permanent options on the right side of the
screen against those ordered. Make changes as required and close this screen.
14.) Run Verify Options to confirm that all options are loaded.
15.) After the options are installed, select QUIT then QUIT again.
16.) Remove the DVD and write-protect the side with options.
17.) When the system prompts to reboot, click YES, and reboot the system to complete the installation.
• If additional network connections are needed for this installation, complete as required.
Confirm network operation.
• If additional service integration is required to complete this installation, complete as required.
• Verify that the system information on the service home page is correct and that system service
information is present on the service desktop.
1.8 Camera
1.8.3 Preparation
If a DASM is required, notify the PMU that the DASM is not supported on systems with GOC6.
For details on camera configuration, refer to the Software Load Procedures manual.
For details on troubleshooting the camera, refer to the System Service manual.
1.8.4 Procedures
1.) Click on the SERVICE DESKTOP icon.
2.) Select CONFIGURATION icon.
3.) Select INSTALL CAMERA.
4.) Read WARNING message, and click OK.
5.) From the remote printer list select a camera, and select ADD for new install.
a.) ADD
b.) UPDATE
c.) DELETE
6.) Select DICOM or POSTSCRIPT*.
* Follow the manufacturers suggested setup instructions..
7.) Follow the procedures on the screen.
Note: Camera and film information is required. Review this information with the customer. Data
sheets are available in Service Information CD under Alignment, Setup and Calibrations.
8.) Return to the Home Page
9.) Click the SERVICE DESKTOP icon.
10.) Click SHUTDOWN
11.) Click REBOOT.
1.) Insert a new Save State DVD into the SCSI Tower DVD RAM drive.
2.) Click the SERVICE DESKTOP icon.
3.) If reloading software, click UTILITIES.
4.) Select SYSTEM STATE.
7.) If the following message appears, insert a DVD into the DVD drive and click YES.
Use these procedures to functionally check every part of the table/gantry subsystem.
Start
Check
X-Ray Lights
Check Alignment
Lights
Perform Mechanical
Characterization
Intercom Check
Volume Set
Interference
Test
Check
Warning Labels
Verify Table
Elevation
Process Product
Locator Cards
Leakage Current
Test
Install Service
Cabinet
End
Required Tool
• Multimeter
2.2.3 Procedures
CAUTION Verify all personnel have cleared the system. The gantry rotates during this check.
1.) Adjust the scan room lights to normal customer operating levels.
2.) Turn ON the AXIAL DRIVE ENABLE and HVDC ENABLE switches (located on the gantry
service switch panel.
3.) Turn on the alignment light switch on the gantry service panel. The gantry rotates and the
alignment lights turn ON.
Press to talk
2.3.3 Procedures
PATIENT SPEAKER
To adjust the volume of the patient speaker in the table:
1.) Adjust the left-most volume thumb wheel on the SCIM while speaking into the console
microphone. (Press the bar on the SCIM to talk; release the bar to listen.)
2.) The assistant should be able to clearly hear the operator.
AUTOVOICE VOLUME
1.) On the Scan Desktop, select PROTOCOL MANAGEMENT.
2.) Select AUTO VOICE RECORD.
The labels on the system and the system manuals must comply with the country law, as listed in
Direction 5221102-1EN (found in the keyboard collector kit shipped with the system) regardless of
the user interface (UI) language that is chosen. Compliance to the law must be completed prior to
releasing the system to the customer.
Important: Do not cover English labels already on the system.
2.5.1 On SCIM
1.) Make sure the X-Ray warning label appears in the correct location on the SCIM.
2.) Record this information on GE Form e4879 located on the Service CD.
2.5.2 On Gantry
1.) Check that all laser warning labels are present on the gantry near the laser opening.
2.) There should also be warning labels on the lower right side of the gantry front cover.
3.) Record this information on GE Form e4879 located on the Service CD.
2.5.3 On Laser
1.) Make sure all laser warning labels appear in the correct location on the outside of the gantry.
2.) Obtain and install replacements for any missing labels.
1.) Collect the product locator cards shipped with the system. There should be approximately 28
product locator cards with the average system.
2.) Update the online product locator web site with the required hospital information.
3.) Confirm that the serial numbers on the cards shipped with the system match those found on
the web site for that GON number. Update the information, as required.
4.) Place the cards in a plastic bag, then place them in the service cabinet.
2.7.3 Procedures
1.) Place the cabinet in the location shown on the site print.
2.) Verify and place all of the service materials shipped with the system in the service cabinet.
Perform several scans following the steps below. Verify that the x-ray ON lights are ON during the
scans. When done, complete the information on GE form e4869 on the Service CD
1.) Make sure the axial drive enable and HVDC enable switches are ON.
2.) If you are not on the Service Desktop, click on the Service Desktop icon.
3.) Select DIAGNOSTICS.
4.) Select DIAGNOSTIC DATA COLLECTION.
5.) Set the scan time to 2.00.
6.) Set the kV to 80.
7.) Set the mA to 40.
8.) Press ACCEPT RX.
9.) Press START SCAN button when flashing.
10.) Record the above information on GE Form e4879 located on the Service CD.
The relationship of table height to ISO center and internal-to-external landmarks must be
characterized for proper interference matrix functionality.
Important: Do NOT perform tilt characterization.
1.) Select the CHARACTERIZE TABLE HEIGHT button from the interface.
2.) Follow the on-screen instructions.
Note: If the table height is less than 21 mm or greater than 25 mm, relative to ISO, you must adjust the
table height using the table leveling pad and adjusters. Raise or lower all adjusters equally to
achieve desired results.
PREREQUISITES
• Be sure that the System State was restored from DVD per Section 1.2.1 on page 220.
• Reset the hardware to download the new characterization values before performing the table/
gantry interference tests in this section.
CONFIGURATION
The table must:
• Be flashed with latest software.
• Have elevation, cradle, and IMS characterized.
• Be mechanically aligned to gantry.
• Have table/gantry characterization completed.
• Have the table extender installed.
TEST OUTLINE
The following tests verify the proper tilt and table interference matrix on the gantry.
• Gantry Tilt Verification Test – Section 2.11
• Position Tilt, Move Table to Interference Limit – Section 2.11.1
• Position Table, Move Tilt to Interference Limit – Section 2.11.2
• Tilt Limits When Table Below Scan Plane Lower Limit – Section 2.11.3
REQUIREMENTS
The following requirements are tested in this series of tests:
1.) No motion causes the table to hit the gantry (or gantry to hit the table)
2.) Requirement #1 includes the use of the table extender.
3.) No tilt motion causes the gantry tilting frame to touch the stationary base covers for any tilt
angle.
LIMITATIONS
These requirements are only met when the table is NOT in service mode.
2.11.3 Tilt Limits When Table Below Scan Plane Lower Limit
The following tests verify the table and tilt interference limits when the table height is below the scan plane.
Note: • “I” means top of gantry tilts toward the table base
• “S” means top of gantry tilts away from the table base.
Important: For all tests, make sure there is 2.5 cm of clearance between the gantry and table.
Test Steps Expected Results
Set the gantry tilt to zero. Cradle position on the display should read 0.0.
Move the cradle (and IMS) to the home Gantry tilt on the display should read 0.0.
position. Table height should read 585 ± 3 mm. (This
1 Lower the table all the way. value is the offset from 560 mm by the same
Set the internal landmark. amount as the upper table limit varies from
25 mm in the Gantry Tilt Verification Test #2 in
Table 6-10.)
Tilt the gantry forward and backward. Gantry tilt on the display should read S0.5 ± 0.5º.
2 Verify the following tilt limits:S0.5 & I30. Gantry tilt on the display should read
I30.0 ± 0.5º.
Set the gantry tilt to 0. Gantry tilt on the display should read 0.0.
Using the gantry push-buttons, move Cradle should stop at 10.0 mm ± 6 mm from the
the cradle in toward the gantry until it home position. (This value is not valid if the
3 stops. table/gantry mechanical characterization is
not correct as described in the Gantry Tilt
Verification Test #2 in Table 6-10. Ensure that
the front of the table is at least 25 mm from
the gantry.)
Set the gantry tilt to zero. Cradle position on the display should read 0.0
4
Move the cradle to the home position. Gantry tilt on the display should read 0.0.
Raise the table to a height of 386 mm. Table height should read 386 mm.
Verify the following tilt limits: S11.0 & Gantry tilt on the display should read
5
I30.0. S11.0 ± 0.5º.
Gantry tilt on the display should read I30.0 ± 0.5º.
Set the gantry tilt to 0. Gantry tilt on the display should read 0.0.
Set the internal landmark. The cradle should stop at 141 mm ± 6 mm from
6 Using the gantry push-buttons, move the home position.
the cradle (and IMS) in toward the
gantry until it stops.
Set the gantry tilt to zero. Cradle position on the display should read 0.0.
7
Move the cradle to the home position. Gantry tilt on the display should read 0.0.
Raise the table to a height of 242 mm. Table height should read 242 mm.
Verify the following tilt limits: S24.0 & Gantry tilt on the display should read
8
I30.0. S24.0 ± 0.5º.
Gantry tilt on the display should read I30.0 ± 0.5º.
Table 6-13 Tilt Limits When Table Below Scan Plane Lower Limit Tests
Chapter 7
Image Quality
Start
Table/Gantry
Alignment Procedure
Introduction
Tomographic Plane
Indication
Basic Preparation
IQ Series
Tube Warm-Up
Fast Calibration System Functional
Test
Center Phantom
Save System State
Prepare the QA
Phantom
End
NOTICE Do not perform the following procedure until instructed to do so in other sections of this
chapter.
This procedure is used to locate the Manufacturing and Installation protocols on the SERVICE tab.
Note: Manufacturing and Service share this Protocol list. Different product option offerings also
use this list. Carefully follow the scan section instructions and verify you acquired the
images with the correct technique before filling out the data sheet. Otherwise you may
troubleshoot an image problem that only exists because you used the wrong technique.
PROCEDURE
1.) Select the NEW PATIENT icon on the left monitor.
2.) Enter a Patient ID (e.g., getest)for easy identification.
3.) Click on the box labeled SERVICE to access the necessary protocols.
• 1 mm wire
4.3 Preparation
4.4 Procedure
CRADLE SETUP
5.) Turn on the alignment lights.
6.) Advance the end of the cradle to the black dot on cradle.
7.) Tape a 100 mm section of tungsten wire on the cradle that aligns with the white cradle center
line.
8.) Using the gantry keypad, set an internal landmark, and then advance the cradle 1000 mm.
9.) Tape second 100 mm section of tungsten wire on the cradle that aligns with the alignment light.
SCANNING SETUP
10.) From the application screen Select NEW PATIENT.
a.) Fill out patient ID: GE Test
b.) Name: Alignment
11.) From the Protocol screen:
a.) Select SERVICE,
b.) Select IMAGE QUALITY,
c.) Select PERPENDICULAR ALIGNMENT.
12.) The red boxes should disappear from the screen. If not, reset an internal landmark.
13.) Select CONFIRM, then press the START SCAN button when lighted.
IMAGE REVIEW
14.) On the Service screen, select IMAGE WORKS
a.) Locate the scanned examination in the Examinations column.
b.) Highlight the Alignment scans
FINALIZATION
20.) Use a calibrated torque wrench to tighten the anchors to 75 ± 6 N-m (55 ± 5 ft.-lb.). Confirm
that the torqued anchor still meets the anchor installation specifications:
a.) Maintain 1 full thread of adjustor showing above the lock ring or table base plate.
b.) Have not more than 1 in. of anchor showing above the nut. Do not cut off any access.
c.) Using a permanent marker, draw a line on the nut and base. Use this line to determine
whether the anchor loosened over time.
21.) Reinstall all table components removed to access the anchors.
tool (Figure 7-1). This will center the tool in the FOV.
b.) Align any of the top lines with the internal laser plane by driving the cradle in/out.
c.) Using the side adjacent screw, tilt phantom until sagittal line strikes both top and bottom
lines. The reading should be 0.0.
d.) Set an internal landmark.
e.) Turn off lasers.
5.) Open a New Patient window at the console.
6.) Enter a Patient ID and Patient Name.
7.) Click the Service tab and select Service Generic Scan.
8.) Choose an Axial 1 second scan and set the following parameters:
a.) Thick Speed: 20.0 mm detector coverage, 0.625/32i, 1.0 rotation
b.) SFOV: Medium body
c.) Interval: 0 mm
d.) 120 kV
e.) 200 mA
f.) Start Location: S9.688
g.) End Location: I9.688
9.) Confirm the scan and press the [Scan] button when it lights up green.
10.) After the scan ends, open Image Works at the console and do the following:
a.) Select the exam and click Viewer on the right of the screen.
b.) Choose the 1-block Format to view 1 image.
c.) Adjust W:400 and L:-300 (using the middle mouse button)
d.) Scroll through the images (using [Page Up] or [Page Down]) until the holes are visible
(Figure 7-2).
(Compare Figure 7-2 with Figure 7-3, where the holes are NOT visible.)
2.) Using the Shim Calculator Chart below (Figure 7-5), determine the approximate number of
shims. (Choose the closest number.)
3.) Using the example below (Figure 7-6), determine on which end of the alignment tool to place
the shims.
2.) Using the previous method, verify that the top and bottom holes are off by no more than 1
image. If off by more than 1 image, adjust the shims accordingly. (See Figure 7-8).)
3.) Ensure that alignment hasn’t changed from Step 3 in the previous section, then repeat the
image series.
4.5.5 Finalization
1.) Remove any shims from the alignment tool.
2.) Repeat scan as defined in Section 4.5.1.
3.) Make additional adjustments so that the top and bottom holes can be seen in the same 2
consecutive images (0 slices between centerlines). (See Figure 7-10.)
6.2 Procedure
1.) Locate the QA phantom and mount it and the phantom holder on the table.
2.) Use the adjustment knobs on the phantom holder to level the phantom front-to-back and side-
to-side with a bubble level.
3.) Select SCANNER UTILITIES on the left monitor.
4.) Select CENTER PHANTOM.
5.) Follow the on-screen procedures.
Figure 7-11
Scan Start End No. of Thick Interval Gantry SFOV kV mA Total DFOV Reco
Type Locatio Locatio Image Speed (mm) Tilt Exposur n
n n s e Time Type
Helical I3.000 S3.200 32 0.625 0.200 S0.0 Small 120 100 2.1 25.0 Bone
Full 10.62 Body
0.8 sec 0.531:1
11.) Locate the scan plane indicator, the longest bar in the bar pattern on the right side of the
phantom. The right side of the phantom corresponds to the side of the image labeled L on the
display screen. See Figure 7-12.
Slice
location
(I1.0 to
S1.0)
Bar
pattern
12.) On the GE Form e4879 Data Sheet, record the scan location (shown on the image annotation)
of the image with the darkest scan plane indicator (darkest long bar).
13.) If your system meets all the installation and alignment specifications, the image at scan
location zero (S0.0) should contain the scan plane indicator. If scan location S1.0 or scan
location I1.0 has the darkest bar, the system still meets the specification. The scan plane
deviation should equal S0.0 +/- 1.0mm. If necessary, adjust the internal alignment light position
to meet the S0.0 +/- 1.0mm requirement.
14.) Repeat the Tomographic Plane Indication test with the external alignment lights.
a.) Use the external alignment light, and press the external landmark.
b.) Verify the external light lines up along the black line on both the left and right sides of the
QA phantom.
c.) The scan plane indication must fall within the S0.0 + 1.0mm specification.
15.) Initial below.
8.2 Procedures
SETUP
1.) Check that the table cradle is level in all directions. Correct, if necessary.
2.) Check the scan window for proper installation
3.) From the operator’s desktop, select SCANNER UTILITY.
4.) Complete COLLIMATOR CALIBRATION and FAST CALIBRATION.
5.) Complete section 4, CENTER PHANTOM procedure.
6.) Locate the white service CD that ships with the system, located in the tech pub tray of the lean
cart.
7.) Load the CD into the lapto drive. Look under the Functional Checks tab located on the left hand
side. Run the QA procedure.
8.) Record all pass/fail information on the GE e4879 form.
9.1 Preparation
• All table mechanical alignment procedures are completed.
• The IQ test passed.
• All options are installed and operational.
If you encounter a failure during the system tests:
• Record any evidence of artifacts, such as rings, streaks, shading, cupping, noise, or center
artifacts.
• Correct artifacts, system test, or image series failures when they occur.
• Record failure information in the comment section of the GE Form e4879.
9.2 Procedure
1.) Locate the white service CD that ships with the system, located in the tech pub tray of the lean
cart.
2.) Load the CD into the laptop drive. Look under the Functional Checks tab located on the left
hand side. Run the System Scanning Test procedure.
3.) Record all pass/fail information on the GE e4879 form.
9.3.1 Procedure
2.) Turn on the monitor. Follow the monitor self test setup procedure using the document shipped
with the system.
3.) From the APPLICATION screen, select NEW PATIENT. Fill out:
- Patient ID: GE Test
- Name: Cardiac Functional System Functional Test
- Select from Protocol Menu
* User
* Chest
- On the dark blue bar on the scan monitor select Gating "On"”
4.) Check for presence of these items:
a.) Heart rate on the gantry display board
b.) Cardiac pulses shown on the screen
c.) Gating BPM displayed on the screen
d.) ECG trace highlighted on the screen
9.4.1 Procedure
PATIENT SCHEDULER
1.) On the scan desktop select Patient scheduler
2.) Select the button labeled Update
3.) The customer list should be present select Export data
4.) Send a test image to a workstation AWW or what ever is available.
9.5 Finalization
Chapter 8
Customer Options Installation & Verification
Note: Only use the Installation manual that arrives with your system for installation. Any other
revisions of this manual may not exactly match your system.
1.2 Preparation
1.3 Procedures
7.) Select the correct Archive Node choice for the device. The Archive Node selection field
defines the ability of the remote host to act as a DICOM Storage/Commitment provider and
indicate to the operator that a study/series/image was archived. Select:
- AUTO to have the CT system automatically check to see if the designated remote host is
a DICOM Storage/Commitment Provider.
- YES if the device is the hospital designated DICOM Storage/Commitment Provider.
During an Application Study Archive process, the local browser screen will indicate
Archive Status = Y to the operator.
- NO if the device is not a DICOM Storage/Commitment Provider.
8.) Select the correct Access to local host: settings. These two selections allow you to
selectively block the remote host from using the LightSpeed DICOM services as a provider
(image push-to destination, and a Query/Retrieve provider).
- Send Images? Set to YES if the customer wants the CT system to be able to have
images pushed to the system from the applicable remote host. Set to NO if the customer
wants to block an image push from the applicable remote host.
- Query/retrieve images? Set to YES if the customer wants the remote host to be
able to review the image database (query) and pull selected images from the database.
Set to NO if the customer does not want the remote host to have this ability.
9.) Select the correct Custom search? setting. This selection allows the CT scanner to
selectively search through the remote host's image database when the operator is using
remote browser screen to query the remote host. The search parameters that the CT system
allows the customer to use are: last name contains, patient ID, exam number, accession
number, and exam date.
- Select ON if the device supports custom searches as part of the devices Query/Retrieve
DICOM Provider service.
- Select OFF if the device does not support custom searches.
10.) Record all the remote host network parameters for each remote host in the Software
Installation Procedures Document.
11.) Select SAVE to store the parameter settings of the remote host.
Refer to the appropriate service manual provided with the Advantage NET Protocol device or
system to find instructions how to declare the CT System as an Advantage NET remote host.
Refer to the appropriate Service Manual provided with the DICOM protocol device or system to find
instructions how to declare the CT System as a DICOM remote host.
The CT System provides two DICOM Services as a provider to remote hosts:
• A remote host can push images to the CT image database.
• A remote host can review the CT image database (query) and pull selected images (retrieve).
Use the following parameter information to configure the DICOM device/system to either push
images to the CT scanner and/or perform a Query/Retrieve operation:
• Hostname: Provided by the Hospital Network Administrator. Exactly the same scanner
assigned hostname entered in Network Configuration Screen.
Most hospital HIS/RIS systems are not DICOM compatible and require a DICOM HIS/RIS Worklist
Interface to provide patient scheduling information to the CT system. Contact your local HNS sup-
port engineer to determine exactly what DICOM HIS/RIS Interface is appropriate for the customer.
In addition, the CT system must have the ConnectPRO software option installed to utilize the
DICOM Protocol Worklist capability.
5.) Insert the ConnectPRO Options MOD into the MOD drive.
6.) Click on OK to continue.
7.) On the Software Option window, select the CONNECTPRO option in the Available Options
window.
8.) Click on the INSTALL button.
The console may display a message box during the software loading operation. When the
system has completed loading the software:
- the Installed Options window displays the ConnectPRO option, and
- the console displays the ConnectPro Setup window as shown in Figure 8-7.
The following is a summary of troubleshooting information for DICOM print that was gathered during
software testing and validation of the DICOM print feature.
There is also a significant amount of additional troubleshooting procedures, and the theory of the
DICOM print feature in the System Service Manual. Should you have problems installing a DICOM
print camera, first read the information in System Service Manual.
1. J31 RPM
2. CB3 ON/OFF
3. RUN #056 #2 Ground
4. G22
5. G21
6. J22
7. J21
8. J4
9. J5 Media tower
10. J2 Scan Monitor
11. J3 Display Monitor
12. J1 RUN # 053 Console Power
13. J20 RUN #101 Scan INTL (Interlock)
14. J25 RUN # 103 DAS Data (Fiber)
15. J26 HSP (hospital)
16. J27 TGP RUN # 102 LAN (RJ45)
17. J28 UPS
18. J29 AW
19. J30 EKG Cat 5 J45
US PROCESS OVERVIEW
The United States network connectivity requirement for this product is broad-band. The US process
relies on the Install Specialist to select a Customer Champion and identify an IT contact for the site.
Together, those individuals then complete a site assessment to gauge what tasks are needed to
fulfill the connection.
Anyone can contact the GE Connectivity team at 800.321.7937, Option #3, with questions.
CUSTOMER BROADBAND RESPONSIBILITIES
Provide GE Healthcare Project Manager of Installation with an accurate site address, telephone
number, contact name, and e-mail address for the:
• Customer Champion
- Coordinate VPN activities between Radiology/Cardiology and the Information
Technology (IT) departments
- Act as a focal point in assuring site broadband infrastructure meets GE Healthcare
requirements for connection as determined by a mutual assessment with the GE
Healthcare Connectivity team.
• IT Contact
- Complete an equipment assessment with GE Healthcare Connectivity team to determine
site readiness for broadband
- Work with the Customer Champion to complete any identified infrastructure changes
- Provide IP addresses for new CT equipment
- Provide a VPN compatible appliance that will support the IPSec tunneling protocol and
3DES data encryption
- To utilize an Internet Service Provider that supports static routing
Chapter 9
System-Level Safety Tests
You must complete these tests after all options are installed. They cover three safety and leakage
current checks:
• Patient Touch Current Test (completed after installation)
• System Ground Resistance Measurement (completed during installation)
• Ground Current Typical (completed after installation - optional)
1.3 Safety
NOTICE Follow ALL required safety and PPE procedures customary for your organization, when
working on this product.
1.4 Preparation
1.5 Procedures
4.) Plug the Dale 600 Meter into the outlet on the gantry left side.
Confirm the outlet is correctly wired per the three LED indicators on the meter.
5.) Connect one end of the shorter, black lead to the chassis plug and connect the other end to the
table ground bus.
6.) Connect the longer, red test lead (or the longer black lead) to the external plug on the top of
the Dale 600 meter.
7.) Set the function switch on the Dale meter to external. Use the external lead to touch the meter’s
test terminal, to test that the meter is operational.
8.) The black lead is connected to the table base ground bus, and the read lead is connected to
the devices (components) under test.
Note: Your meter may have two black leads that are keyed for chassis and the ground connection.
A valid calibration sticker must be present on the meter you use. Record this information on
the GE Form e4879.
9.) Leakage current is tested with power ON.:
Measurement Area Current Not to Exceed
Critical care ares (invasive) 10µa CT Systems
General care areas 20µa
Not intended for patient area 50µa
Table 9-1 Maximum Leakage Current Measurements
10.) Complete testing between the system reference ground point (table base) to unit reference
ground points (i.e. gantry and table, see Figure 9-3).
11.) Test all conductive surfaces and components within patient reach or within 1.9 m (6 ft) of the
table and 2.3 m (7.5 ft) above the table.
- Measure at table maximum travels.
- At some sites, wall outlet cover plates and sinks may become an issue.
12.) Test all Optional components such as in-room monitors, injector, overhead monitor
suspensions, and table options.
Record these results in the data sheet.
Gnd Bus to Install <10µA
Any table ground points within the 6’ range
Any gantry ground points within the 6’ range
Injector assembly metal surface
Boom-in-Room metal surface
Monitors or metal surface
Sink or metal surface
Installed table accessories
(other)
(other)
(other)
NOTICE Be aware of Static Discharge from: scan window, keypads, displays, touch pads, or
other plastic surfaces.
1.7 Finalization
No finalization required.
2.3 Safety
NOTICE Follow ALL required safety, PPE, and arc-flash procedures customary for your organization,
when working on this product.
2.4 Procedures
Follow LOTO and other safety procedures found in this manual before starting this procedure.
1.) Remove the Gantry top and side covers. Guide rails on each side of the top cover will aid in
removal. Refer to Appendix A: Gantry Cover Removal and Dolly Setup on page 167 for more
details.
Note: Fans located on the top cover will make that cover heavier.
2.) Power down the console and follow the GEH LOTO procedure.
3.) Remove the Gantry front cover, see cover procedure in Book 1.
4.) Remove the Table foot switch top cover to gain access to the ground cables and ground bar.
Note: Do NOT disconnect ANY grounds at this time.
5.) Confirm that all system grounds are securely attached to the system ground buss and NOT
the Table base.
6.) The electrician will remove all external electrical connections made during installation,
including:
a.) Main system ground at PDU
b.) Power feeder flex connection at PDU
c.) Room door interlocks and room warning light connections
d.) Any and all other external ground connections to the system.
7.) Confirm that all external gantry, table, console, and PDU connections have been removed.
Note: Some wires such as the room warning light may have external power and wire nuts, which
should be installed to protect from arching.
8.) Follow the LOTO procedure for re-energizing power and boot to application level. If not
already done, remove the footswitch assembly top cover while the system boots.
9.) Plug the Dale 600 / 601 Leakage Current Meter into one of the outlets on the gantry.
10.) Connect the meter leads to the meter as follows:
a.) Connect one end of the shorter black lead to the chassis plug and the other to the table
ground bus.
b.) Connect the longer red test lead to the external plug on top of the Dale 600 / 601 meter.
11.) Set the function switch on the Dale 600 / 601 meter to EXTERNAL. Using the external lead,
touch the meter’s test terminal to confirm that the meter is operational.
Note: For more detailed information, refer to the Dale 600 / 601 Operator’s Manual or see Figure 9-
1, for a quick overview.
With the system at application-level and all components functional, test the system ground
wires as follows:
a.) Remove a system ground wire.
b.) Test that system ground wire.
c.) Replace that system ground wire.
d.) Repeat, testing all system ground wires ONE-AT-A-TIME. A list of the system ground
wires appears in Table 9-2.
Note: The measured leakage current must not exceed 5 MA in any ground wire.
Components Results Date Tester
PDU Leakage 1/0
Gantry Leakage 1/0
Console Leakage #2
Table Leakage #10
Option Grounds, if present
12.) After completing all tests, follow the LOTO procedure to power down the system.
13.) The electrician will reinstall all electrical connections, conduits, cables, and wires removed in
Step 6 and will secure all connections per NEC code. Check that all connections are securely
tightened.
14.) Reinstall all removed system covers, except for the gantry right-side cover, located by the
service switch panel.
2.5 Finalization
No finalization required.
Appendix D
System Configuration Data Sheets
_________:_________:_________:_________:_________:_________
Destination:
Orientation:
Medium Type:
Magnification Type:
Table D-5 DICOM Print Camera Configuration
Appendix E
Informational Support Procedures
3
335-1 Three-phase Alternating Current
3N
335-1 Three-phase Alternating Current with
neutral conductor
Direct Current
Emergency Stop
417-5009 Standby
Start
Table Set
Abort
Intercom
This appendix contains the following procedures that may be useful during a software load:
• CDROM — Mount Process
• CDROM — Unmount Process
3.2 Procedures
2.5.1 Preparation
1.) Insert a CD into the DVD ROM drive.
2.) Open a Shell window.
3.) Type: su - ENTER
4.) Type the password and press ENTER
www.gehealthcare.com
312