S8000 Instructions
S8000 Instructions
S8000 Instructions
Threads mate
Mechanical Installation with knurled
Submersion Installation With ATC nut
A) Apply silicone or other waterproof o-ring grease to all three
o-rings of the electrode cartridge (S8000CD, S8000CD-LC).
Install electrode cartridge into housing (EA899TC) by pushing
it up into the housing until you feel resistance at the connec-
tor. Next, rotate the cartridge clockwise until it stops turning.
The electrode should only be installed hand-tight. FIG 3 and
FIG 3A show before and after images of this operation.
INSTRS800011162023
Page 2 of 12
PRODUCT INSTRUCTION SHEET
8-pin
Connector
Grease these
o-rings
Knurled
nut
Signal
Conditioner
INSTRS800011162023
Page 3 of 12
PRODUCT INSTRUCTION SHEET
B) Install the S853 cable assembly onto the top of the pH Grease these
sensor and housing assembly by aligning the eight-pin o-rings
connector receptacle in the cable assembly with the eight-
pin mating connector on the housing. Hand-tighten knurled
nut onto threads of sensor housing until it stops turning.
Before and after images are shown in FIG 8 and FIG 8A.
Electrode
Cartridge
Knurled
nut
8-pin
Connector
Adapter
threads
INSTRS800011162023
Page 4 of 12
PRODUCT INSTRUCTION SHEET
Internal BNC
(inside S653)
1/4
turn
INSTRS800011162023
Page 5 of 12
PRODUCT INSTRUCTION SHEET
Knurled
Nut
Grease
these
o-rings
INSTRS800011162023
Page 6 of 12
PRODUCT INSTRUCTION SHEET
C) Take housing and tee assembly from FIG 11A and install
signal converter as shown in FIG 13 and FIG 13A.
D) Install the cable assembly onto the top of the signal con-
verter by aligning the eight-pin connector receptacle in the
cable assembly with the eight-pin mating connector on
the housing. Hand-tighten knurled nut onto threads of
sensor housing until it stops turning. Before and after
images are shown in FIG 14 and FIG 14A.
In-line Installation Notes: Make sure to install tee at least 45 FIG. 14 FIG.14A
degerees above horizontal position. Do not install horizontally.
See FIG 15.
S855 or
S856 Grease
these
o-rings
INSTRS800011162023
Page 7 of 12
PRODUCT INSTRUCTION SHEET
Product Specifications
Electrode Specifications:
S8000CD
pH Range: 0-14 (sodium ion error above pH 12.3)
Temp Range: 0-100C (derated based on pressure)
0-80C (in FC800)
0-70C with EM modules
Pressure Range: 0-100psig (derated based on temp)
Speed of Response: 95% in 5 seconds
Wetted Materials: PPS (Body), HDPE(junction), pH glass,
Viton o-rings
S8000CD-HF
pH Range: 0-12
Temp Range: 0-50C (derated based on pressure)
Pressure Range: 0-100psig (derated based on temp)
Speed of Response: 95% in 5 seconds
Wetted Materials: PPS (Body), HDPE(junction), HF-resis-
tant pH glass, Viton o-rings
S8000CD-ORP
Range: +/- 1000mV
Temp Range: 0-100C (de-rated based on pressure)
0-80C (in FC800 or with EM modules)
Pressure Range: 0-100psig (derated based on temp)
Wetted Materials: PPS (Body), HDPE(junction), Platinum,
Viton o-rings
Cable Connections:
S853*
Coaxial: Center = pH or ORP *
Braid = Reference **
S855* S856 **
Coaxial: Center = pH or ORP * Red: V+
Braid = Reference ** Black: GND or signal for 4-20mA
Red: Temperature White: Modbus A (+)
Black: Temperature Green: Modbus B(-)
Green: Solution Ground
Mounting Hardware Specifications:
FC800, FC801, FC875/S, FC875/T
Temp Range: 0-80C Notes:
(derated based on pressure) * extending cable is not recommneded due to noise sensitivity of
Pressure Range: 0-100psig (derated based on temp) measurements for pH
Electronics Modules Specifications: ** cable can be extended. Recommend 24AWG shielded wire
EM800, EM801, EM805
Temp Range: 0- 70C
(derated based on pressure) INSTRS800011162023
Pressure Range: 0-100psig (derated based on temp)
Page 8 of 12
PRODUCT INSTRUCTION SHEET
Two-Point Calibrations
Two-point calibrations correct for both the pH electrode's
FIG. 17
offset and span errors. Since both the offset and span vary
with time the two-point method is the onr preferred. Choose
buffer pH 7 for zero-point and a second buffer close to your
normal operating range (usually pH4.01 or pH 10.00). See FIG
16 A-C.
Page 9 of 12
PRODUCT INSTRUCTION SHEET
Intermittent Operation
Some facilities are only operated part of the time. When out FIG. 18
of operation, electrodes must not be allowed to be exposed
to air and become dry. Electrodes should be removed from
such systems and stored in their bottles or caps or in a bea-
ker filled, preferably, with pH 4.0 Buffer (SEE FIG 10). In some
instances, power to the meter is shut off; this condition can
be harmful to electrodes. Electrodes should be disconnected
from un-powered meters.
Page 10 of 12
PRODUCT INSTRUCTION SHEET
4 5 (-)
(+) 2
INSTRS800011162023
Page 11 of 12
PRODUCT INSTRUCTION SHEET
Electrode Troubleshooting
See FIG 24 table for typical electrode symptoms, causes and
corrective actions.
Page 12 of 12