Lecture Notes XRF
Lecture Notes XRF
K shell - 2 electrons
L shell - 8 electrons
M shell - 18 electrons
N shell - 32 electrons
• Electrons occupy specific energy levels, which are quantized. For example, in a
hydrogen atom, these levels are designated by the principal quantum number 𝑛n
(1, 2, 3, etc.).
• Each element has a unique set of energy levels, meaning that the X-rays emitted
will have specific energies characteristic of that element. This is what allows for
the identification of elements in the sample.
• The emitted X-rays are detected by an XRF instrument, which analyzes the
energies of these X-rays. By measuring the energies and intensities of the emitted
X-rays, the elemental composition of the sample can be determined
Bohr Atomic Model
The term fluorescence is applied to phenomena in
which the absorption of higher-energy radiation
results in the re-emission of lower-energy
radiation.
Characteristic radiation
• Each element has specific
energy levels for its
electrons (characteristic
energy).
300
Intensity (cps)
200
As K line
11.73 keV
100
0
0 5 10 15 20 25 30 35 40
Energy (keV)
500
Intensity (cps)
200
100
0
0 5 10 15 20 25 30 35 40
Energy (keV)
• The presence of Pb in this sample is confirmed through observation of two peaks centered
at energies very close (within ±0.05 keV) to their tabulated (reference) line energies
• These same two peaks will appear in XRF spectra of different lead-based materials
(i.e., lead arsenate, tetraethyl lead, etc.)
Instrumentation components
• X-ray source (tube or radioactive source)
• Sample holder
• Detector (e.g., Si-PIN, Ge)
• Multichannel analyzer
BOX DIAGRAM OF XRF INSTRUMENT
X-ray Digital Pulse XRF Results
Source Detector Spectrum (elements
Processor software
(cps vs keV) and conc’s)
Sample
• X-ray tube source
✓ High energy electrons fired at anode (usually made from Ag or Rh)
✓ Can vary excitation energy from 15-50 kV and current from 10-200 A
• collimator
• crystal
• detector
• path
Schematic diagram XRF instrument
The Hardware
• Sources
• Optics
• Filters & Targets
• Detectors
Side Window X-Ray Tube
Be Window
Glass Envelope Target (Ti, Ag, HV Lead
Rh, etc.)
Electron beam
Copper Anode
Filament
Silicone Insulation
Soller Collimators
Sample
X-Ray
Source
Source Filters
Filters perform one of two functions
– Background Reduction
– Improved Fluorescence
Source Filter
Detector
X-Ray
Source
Filter Fluorescence Method
With Zn Source filter Target peak
ENERGY (keV) Fe
Region
device. Sample
Detector
Collimators
X-Ray
Source
Diffraction Device
Energy Dispersive Electronics
Signal to Electronics
Element A
Element B
DETECTOR
Element C
Element D
Multi-Channel Analyser
• Detector current pulses are translated into counts (counts per second, “CPS”).
• Pulses are segregated into channels according to
energy via the MCA (Multi-Channel Analyser).
Intensity (# of CPS
per Channel)
Channels, Energy
Signal from Detector
Detector Principles
➢A detector is composed of a non-conducting or semi- conducting material
between two charged electrodes.
➢X-ray radiation ionizes the detector material causing it to become conductive,
momentarily.
➢The newly freed electrons are accelerated toward the detector anode to
produce an output pulse.
➢In ionized semiconductor produces electron-hole pairs, the number of pairs
produced is proportional to the X-ray photon energy
E
n =
e
where : n = n u m b e r o f e l ec t r o n - h o l e p ai r s p r o d u c e d
E = X-ray ph oton ener gy
e = 3 . 8 ev f or S i a t L N 2 t e m p e r a t u r e s
Detectors
• Si(Li)
• PIN Diode
• Silicon Drift Detectors
• Proportional Counters
• Scintillation Detectors
Si(Li) Detector
FET
Window
Super-Cooled Cryostat
Dewar
filled with
Si(Li) LN2
Pre-Amplifier
crystal
Anode Filament
Window: Be or Al
Count Rates: 10,000 to 1,000,000+ cps
Resolution: >1000 eV
Spectral Comparison - Au
Solids:
✓ Orient surface patterns in same manner so as minimise scatter affects.
✓ Polishing surfaces will also minimise scatter affects.
✓ Flat samples are optimal for quantitative results.
Liquids:
✓ Samples should be fresh when analysed and analysed with short analysis time –
if sample is evaporative.
✓ Sample should not contain precipitants/solids
XRF methodology
First determine the following:
• what elements are to be measured
• what are their concentration ranges
• what accuracy is required
• how many samples are to be measured
• are suitable standards available
XRF methodology
Elements to be measured
• low Z will require careful preparation
• low Z may have lower accuracy
• low Z may require fusion of powders
• semi-quant does not measure the very light
elements (Be to N)
XRF methodology
concentration ranges
as the concentration range for each element
increased, accuracy generally decreases
Zn
Ga
Ge
As
Se
10
Intensity
(cps)
0
5 6 7 8 9 10 11 12 13 14 15
Energy (keV)
∞ As ∞ Pb
N 4s2p3d10f14 N 4s2p3d10f14
L 12.55 keV
M 3s2p3d10 M 3s2p3d10
K 11.73 keV
L 10.61 keV
L 2s2p6
As Pb
L 2s2p6
>15.21 keV
(absorption edge)
K 10.53 keV
K 1s2 K 1s2
>11.86 keV
(absorption edge - minimum amount of energy needed to remove electron)
• Since XRF affects inner shell and not bonding electrons, the XRF spectrum of an
element is independent of its chemical form (i.e., spectra of lead, lead arsenate, and
tetraethyl lead will ALL show peaks at 10.61 and 12.55 keV)
K LINE SERIES
~10% As in Chinese supplement
400
As K line
10.53 keV
200
As K line
11.73 keV
100
0
0 5 10 15 20 25 30 35 40
Energy (keV)
• L lines not observed (1.28 and 1.32 keV - too low in energy to be excited)
• K and K peak energies are often close together (1.2 keV apart for As)
• K lines observed for low to medium Z elements (i.e., Cl, Fe, As)
• K and K peaks have typical ratio of ~ 5 to 1
L LINE SERIES
~10% Pb in imported Mexican tableware
700
Pb L line Pb L line
10.55 keV 12.61 keV
600
500
Intensity (cps)
400
300
200
100 Pb L line
0
0 5 10 15 20 25 30 35 40
Energy (keV)
• K lines not observed (75.0 and 94.9 keV - too high in energy to be excited)
• L and L peak energies are often further apart (2.1 keV apart for Pb)
• L lines observed for high Z elements (i.e., Hg, Pb, Th)
• L and L peaks have typical ratio of ~ 1 to 1
MORE COMPLEX XRF SPECTRUM
Chinese supplement containing 4% As and 2% Hg
120
As K line
10.53 keV
100
80
Intensity (cps)
Overlapping lines
60
As K 11.73 keV
Hg L 11.82 keV
Hg L line
40 9.99 keV
20
0
9 10 11 12 13
Energy (keV)
• Line overlaps are possible and users must evaluate spectrum to confirm the
presence or absence of an element
EFFECT OF DETECTOR RESOLUTION
Spectra of 900 ppm Pb added into Pepto-Bismol
K-Lines
L-lines
Rh X-ray Tube
Scatter
Sample
• Some of the source X-rays
strike the sample and are
scattered back at the detector.
Sometimes called
“backscatter”
Detector Source
Rayleigh Scatter
EO
EI
RAYLEIGH PEAKS Elastic scattering from metal alloy sample
50
metal sample
45 Cr, Fe, Ni
peaks from
40
metal sample
35
30 Rayleigh Peaks
Intensity (cps)
25
(Rh L and L lines)
20
15
10
0
0 5 10 15 20 25 30 35 40
1 Energy (keV)
E0 = initial energy of X-ray from target element in x-ray tube source.
E1 = energy of X-ray elastically scattered from (typically dense) sample
• Peaks arising from target anode in X-ray tube source (Rh in this case) that may appear in all XRF spectra.
• No energy is lost in this process so peaks show up at characteristic X-ray energies (Rh L and L at
20.22 and 22.72 keV in this case)
• Typically observed in spectra of dense samples as weak peaks (due to increased absorption of X-ray source
photons by sample)
Compton Scatter
EI EO
COMPTON PEAKS: Inelastic scattering from cellulose sample
100
90
cellulose sample Compton Peaks
(E’s < Rh L and L lines )
80
70
PHOTO
ELECTRON 60
Intensity (cps)
50
Rayleigh Peaks
40 (Rh L and L lines)
30
20
• Peaks arising from target element in X ray tube (again, Rh in this case) that may appear in all XRF spectra.
• Some energy is lost in this process so peaks show up at energies slightly less than characteristic X-ray tube
target energies
• Typically observed in spectra of low density samples as fairly intense peaks (note these peaks are wider than
Rayleigh peaks)
Brehmstrahlung
40 kV
E0 > 90
20 kV
10 kV
80
70
Bremsstrahlung
Intensity (cps)
60
50
40
30
20
10
0
0 5 10 15 20 25 30 35 40
E1 , E2 = energy of X-ray
• Very broad peak due to backscattering of X-rays from sample to detector that may appear in all XRF spectra
• Maximum energy of this peak limited by kV applied to X-Ray tube, maximum intensity of this peak is ~
2/3 of the applied keV
• More prominent in XRF spectra of less dense samples which scatter more of X-ray source photons back to the
detector
Interferences
❖ Spectral Interferences
❖ Environmental Interferences
❖ Matrix Interferences
Spectral Interferences
• Spectral interferences are peaks in the
spectrum that overlap the spectral peak
220 eV Resolution (region of interest) of the element to be
analyzed.
• Examples:
– K & L line Overlap - S & Mo,
Cl & Rh, As & Pb
– Adjacent Element Overlap -
Al & Si, S & Cl, K & Ca...
• Resolution of detector
determines extent of overlap.
Environmental Interferences
• Solution:
– Purge instrument with He (less dense than air
= less attenuation).
– Evacuate air from analysis chamber via a
vacuum pump.
• Either of these solutions also eliminate
interference from Ar (spectral overlap to Cl).
Argon (Ar) is a component of air.
Matrix Interferences
Absorption/Enhancement Effects
ence
Absorption-Enhancement Affects
• The interaction of X-ray fluorescence photons from the sample with the
detector can generate several different types of artifact peaks in an XRF
spectrum which may include the following:
➢Sum peaks
➢Escape peaks
SUM PEAKS Example from analysis of Fe sample
Detector
Fe K photon Fe K peak
6.40 keV 6.40 keV
Sum peak
12.80 keV
Fe K photon
6.40 keV
• Artifact peak due to the arrival of 2 photons at the detector at exactly the same
time (i.e.,K + K, K + K )
• More prominent in XRF spectra that have high concentrations of an element
• Can be reduced by keeping count rates low
Sum Peaks
▪ 2 photons strike the detector at the same time.
Intensity (cps)
400 Pb escape peak
Escape peak (from L)
8.81 keV 300
Pb L photon
Pb escape peak
10.55 keV 200
(from L)
100
Escape Peak = Pb – Si
8.81 = 10.55 – 1.74 0
0 5 10 15
Adapted from Thermo Scientific Quant’X EDXRF training manual
• Artifact peak due to the absorption of some of the energy of a photon by Si atoms in the detector
(Eobserved = Eincident –Esi where ESi = 1.74 keV)
• More prominent in XRF spectra that have high concentrations of an element and for lower Z
elements
• Can be reduced by keeping count rates low
ARTIFACT PEAKS DUE TO BLANK MEDIA
20
18 cellulose blank
16
14
12
Intensity
10
Artifact Peaks
(cps)
0 0 5 10 15 20 25 30 35 40
Energy (keV)
• May observe peaks due to contaminants in XRF cups, Mylar film, and matrix
• In this case, the cellulose matrix is highly pure and the peaks are due to trace elements in
the XRF analyzer window and detector materials
• This can complicate interpretation (false positives)
SUMMARY OF FACTORS THAT COMPLICATE
INTERPRETATION OF XRF SPECTRA
• Elements in the sample may produce 2 or more lines
✓ K, K L, L (we use simplified nomenclature and discussed only and lines)
✓ L, L, L, L (can also have and lines, and lines, lines, etc.)
• Peak overlaps arising from the presence of multiple elements in the sample and
limited detector
resolution
• Sum peaks (two X-ray photons arriving at the detector at the same
time)
• Escape peaks (Si in the detector absorbing some of the energy from a
X-ray)
Fe sum peak
12.80 keV
• User acquired sample spectrum near lid (>10% Fe), which gave Fe sum peak at 6.40 keV * 2
photons = 12.80 keV
• Vendor algorithm incorrectly identified Pb in this sample at over 2000 ppm (detection and
quantitation based on signal at the Pb L line at 12.61 keV, zero intensity of Pb L line at
10.55 keV not considered by algorithm)
• Be wary of analyzer software and be sure to avoid potential false positives such as this by
evaluating the spectrum to confirm the presence of an element
QUALITATIVE ANALYSIS
Spectra for positive, tentative, and negative identifications
6
blank As K line
5 10 ppm As and Hg 10.54 keV
100 ppm As and Hg
Overlapping lines
Intensity
3 As K 10.54 keV
(cps)
Hg L 11.82 keV
2 Hg L line
9.99 keV
1
0
9 10 11 12 1
3
Energy (keV)
• As and Hg clearly present in blue spectrum (see both and peaks)
• As and Hg possibly present in purple spectrum ( peaks barely > blank)
• As and Hg not present in black spectrum (no visible peaks)
CONCLUSIONS ON QUALITATIVE ANALYSIS
• Vendor software in commercial XRF analyzers are usually reliable in identifying which
elements are present in a sample, but are not foolproof and an occasional false positive or
false negative is possible
• Ca = f(Ia, Ib)
Quantitative Analysis
• XRF is a reference method, so
standards are required for
quantitative results.
• Are the element CONCENTRATIONS within the detection range of XRF (% to ppm
levels)?
• What sort of SAMPLE PREP is required (can samples be analyzed as is or do they need
to be ground up)?
• particle size
• inhomogeneity
Semi-quant (standardless
analysis)
• Accuracy of the semi-quantitative method
can be as good as 1% relative;
STANDARD ADDITIONS
• Involves adding known amounts of element of interest into the sample
• Provides most accurate results as the standards are prepared in thesample
matrix as the sample
• Usually reserved for laboratory analyses by trained analysts, and even then
used only as needed as this is labor intensive and time consuming
EFFECT OF CONCENTRATION
Spectra of As standards in cellulose
80
1000 ppm
70 100 ppm
10 ppm
blank
60
Intensity (cps) 50
40
30
As K line
20 11.7 keV
As K line
10.5 keV
10
0
9 10 11 12 13
Energy (keV)
800
400
200
0
0 2000 4000 6000 8000 10000 12000
Concentration (ppm)
40000
)
m
p p
3
CN response ( 0000
y = 4.1359x - 847.3
R2 = 0.9973
10000
0
0 2000 4000 6000 8000 10000 12000
Concentration (ppm)
6
-7% error
4
-11% error
2
-5% error
0
0 2 4 6 8 10 12 14
True %Cr
• Although Fundamental Parameters based quantitation gives fairly accurate
results, it also gives determinate error (consistently negative errors)
• Determination of Cr in surgical grade stainless steel samples using an XRF
analyzer calibrated with these standards gave results that were statistically
equivalent to flame atomic absorption spectrophotometry
• For determining % levels of an element, use Fundamental Parameters mode
QUANTITATIVE ANALYSIS AT LOW CONC’S
As, Hg, Pb, and Se standards in cellulose for supplement analysis
1800
Hg
y = 1.6371x - 19.023
1600
As
Hg CN mode with R2 = 0.9993
Pb
Se
empirical calibration Se
y = 1.3598x - 5.3975
1400 R2 = 0.9998
As
y = 1.2494x - 1.5008
Pb
1000 y = 1.091x - 3.0741
9% error R2 = 0.9999
800
600
8% error
400
200
4% error
0
0 200 400 600 800 1000 1200
True concentration (ppm)
y = 0.0003x + 0.0106
R2 = 0.9928
0.015
0.010
• Typically gives more reliable quantitative results as this method involves matrix
matching (the sample is “converted” into standards by adding known amounts of the
element of interest)
• This process is more time consuming (requires analysis of sample “as is” plus two or
more samples to which known amounts of the element of interest have been added)
EFFECT OF MEASUREMENT TIME
Longer analysis times give better precision and lower LODs
30
20
For field applications, the sample is often analyzed “as is” and some
accuracy is sacrificed in the interest of shorter analysis times and higher
sample throughput, as the more important issue here is sample triage
(identifying potential samples of interest for more detailed lab analysis)
4. APPLICATIONS OF XRF
Screening for toxic elements in large numbers of samples
Accurate quantitative analysis of target elements in various matrices
5. CONCLUSIONS
XRF advantages and limitations
References and additional reading
FOUR KEY ADVANTAGES OF XRF
FOR MANY APPLICATIONS
SIMPLICITY
• Relatively simple theory, instrument, and spectra (versus IR, MS, NMR)
PORTABILITY
• Instrument can be brought to the samples
ANALYTICAL PROCESS STREAM
More intelligent
• Use XRF for “sample triage” (sort into analysis protocol
“detects” and “non-detects”) Screen
product via
XRF
warranted
TYPICAL APPLICATIONS OF XRF
Geology
• Major, precious, trace element analysis
• Characterization of rocks, ores, and soils
Environmental Remediation
• Pb in paint
• Heavy metals in soil (EPA method 6200)
Recycling
• Alloy identification
• Waste processing
Miscellaneous
• Art and archeology
• Industrial hygiene
• Forensics
Pb and other elements are
still causing problems
Pb and U
detected in ceramic material
imported from Mexico
• Ceramic plates may contain toxic elements that can leach into food
• XRF can be used to quickly identify elements and their concentrations in
tableware, glazes, and base ceramic material, and food
Handheld XRF analysis of
Kervorkian-designed biopsy
forceps
• ICP-MS showed 6.85% As (note low value here versus XRF may be due to
inability of acid digestion procedures to dissolve realgar)
30
y = 0.8176x + 0.0926
20
R2 = 0.9964
10
0
0 5 10 15 20 25 30 35 40 45 50
ppm Pb via ICPMS
• t test indicates no significant differences at the 95% confidence level between lab-
grade XRF and conventional ICP-MS method
• Such data demonstrate that XRF can give accurate quantitative results
(impressive considering most samples contain these elements at concentration
that are very close to the detection limit)
OUTLINE
1. INTRODUCTION
The electromagnetic spectrum and X-rays
Basic theory of XRF and simple XRF spectra
Different types of XRF instruments
4. APPLICATIONS OF XRF
Screening for toxic elements in large numbers of samples
Accurate quantitative analysis of target elements in various matrices
5. CONCLUSIONS
XRF advantages and limitations
References and additional reading
ADVANTAGES OF XRF
Speed: Minimal sample prep (analyze “as is” or homogenize and transfer to cup)
Fast analysis times (typically seconds to minutes)
To get more accurate results, one must homogenize the samples and
calibrate instrument response using authentic standards
Multiply wt% ELEMENT by numerical value
• Ag2O 1.0741 • below for equivalent
CuO 1.2518 • MnO 1.2912 • RhO 1.5555 • Tm2O3 1.1421
• Dy2O3 1.1477 • MnO2 1.5825 • RuO 1.1583
expressed as OXIDE
• Al2O3 1.8895 • UO2 1.1344
• As2O3 1.3203 • Er2O3 1.1435 • MoO3 1.5003 • SO3 2.4972 • UO3 1.2017
• As2O5 1.5339 • Eu2O3 1.1579 • N2O5 3.8551 • Sb2O5 1.3284 • U3O8 1.1792
• Au2O 1.0406 • FeO 1.2865 • Na2O 1.3480 • Sc2O3 1.5338 • V2O5 1.7852
• B2O3 3.2202 • Fe2O3 1.4297 • Nb2O5 1.4305 • SeO3 1.6079 • WO3 1.2610
• BaO 1.1165 • Ga2O3 1.3442 • Nd2O3 1.1664 • SiO2 2.1392 • Y2O3 1.2699
• BeO 2.7758 • Gd2O3 1.1526 • NiO 1.2725 • Sm2O3 1.1596 • Yb2O3 1.1387
• Bi2O5 1.1914 • GeO2 1.4408 • OsO 1.0841 • SnO2 1.2696 • ZnO 1.2448
• CO2 3.6644 • HfO2 1.1793 • P2O5 2.2916 • SrO 1.1826 • ZrO2 1.3508
• CaO 1.3992 • HgO 1.0798 • PbO 1.0772 • Ta2O5 1.2211
• Therefore the optical path is simpler than for WDXRF spectrometers because no
crystals or goniometers are needed and the fluorescence photons from the
sample hit the detector directly.
• The ED-XRF only consists of two basic units, the excitation source and the
spectrometer / detection system.
ED-XRF
• The samples are normally irradiated by X-rays from a tube with lower
power than that used with the WD-XRF spectrometers.
• They can also use miniature X-ray tubes or gamma sources. This
makes them cheaper and allows miniaturization and portability.
ED-XRF
• This type of instrument is commonly used for portable quality control
screening applications, such as testing toys for Lead (Pb) content, sorting
scrap metals, and measuring the lead content of residential paint.
• On the other hand, the low resolution and problems with low count rate
and long dead- time makes them inferior for high-precision and high-speed
analysis.
Feature WD-XRF ED-XRF
Uses a crystal to disperse X-rays by wavelength, Uses a semiconductor detector to measure X-ray
Principle
allowing for high-resolution detection. energies directly.
Generally provides better energy resolution due to Lower energy resolution compared to WD-XRF,
Resolution
wavelength dispersion. but faster analysis.
Quantitative Highly accurate quantitative analysis; well-suited Good quantitative capabilities, but less precise
Analysis for trace element detection. than WD-XRF for trace elements.
Element Effective for heavier elements (higher atomic Suitable for a broader range of elements,
Range number); less effective for lighter elements. including lighter elements.
Typically requires solid samples; can analyze Can analyze solids, liquids, and powders;
Sample Type
powders, thin films, and some liquids. versatile sample types.
Speed of Slower due to the complexity of the wavelength Faster analysis time; ideal for routine testing and
Analysis selection process. high-throughput applications.
Cost and More complex and generally more expensive due to Simpler and typically less expensive; easier to
Complexity the required optics and components. maintain.
Used in applications requiring high precision, such Commonly used in industries for quality control,
Applications as mineralogy, metallurgy, and environmental environmental monitoring, and material
analysis. characterization.
Energy-Dispersive XRF
(ED-XRF)