Manual TFT8210H Tecnofluid V3.0
Manual TFT8210H Tecnofluid V3.0
Manual TFT8210H Tecnofluid V3.0
II
1. Product Function Specification
1.1 Base Function
Implementation standards
《JJG-1033-2007 electromagnetic flowmeter》。
Suitable to Size of the Sensor (mm):
1,2,3、6、8,10、15、20、25、32、40、50、65、80、100、125、150、200、250、300、350、400、
450、500、600、650,700、800、900、1000、1100,1200、1300,1400、1500,1600、1700,1800、2000、
2200、2400、2600、2800、3000
Power Supply
Power supply :85VAC --- 265VAC or or 110vac to 420vac or 24vdc or 3.6v lithium battery
The Velocity Range:
0.01 - 15 m / s, velocity resolution: 0.5 mm / sec
Requirement For Sensor And Excitation (AC and DC power supply type)
Excitation current can be set to 125mA、160mA、250mA
250mA excitation current is suitable to the resistance of two coil:35~ 65Ω;
125mA excitation current is suitable to the resistance of two coil: 65 ~ 120Ω
excitation frequency:25HZ;12.5HZ,6.25HZ,3.125HZ
Empty-Pipe Check And Full-Pipe Check
Automatic & continuous measure the condition of liquid and diplay the empty pipe and full pipe message.
do not need the full -pipe calibration and Avoid false alarm
Output Signal (AC and DC power supply type)
Current output: 4 to 20mA,load;resistance:0~750Ω,Base deviation:0.1%±10µA。
Frequency output: Frequency range is 100~5000Hz;Photoelectric isolation, isolation voltage:> 1000VDC;
Pulse equivalent output: user defined pulse width,automatic conversion to square wave at high
requency ;Photoelectric isolation, isolation voltage:> 1000VDC;
Alarm Output
Alarm output contact:H-ALM and L-ALM ; Photoelectric isolation, isolation voltage:> 1000VDC;Output
driver:Maximum withstand voltage 36VDC, maximum load current 30mA.
Communcation
Communication :RS485 (standard)、HART (option)
MODBUS interface:RTU format,Physical interface:RS-485,Electrical isolation:1000V;
HART interface:Support standard HART protocol
Language And LCD Dispaly
English ,With LCD,display flow rate , total flow , velocity etc
three internal accumulator, record positive total flow and reverse total flow and net total
3.6V battery power emf type with backligh 。
3
Nonlinear Correction Function:Multi segment linear correction,suitable to variety of sensors
Automatic Zero Calibration Function
Quick Response, Response time of 0.3 seconds
Electrode Self-Cleaning Function
Protection
with lightning protection circuit design. High efficiency anti interference circuit, suitable for all kinds of harsh .
Grade of Protection: IP65 or IP68
1.2 Working conditions
Ambient temperature:–20~+65℃;Relative humidity:5%~90%;Power: less than 10W (after connecting the
sensor).
1.3 Connection type with sensor
Integral type and Separate type (converter is connected to the junction box of the sensor through the cable )
1.4 Transmitter outline dimension and type
According to the size of the sensor, we offer four kinds of transmitter housing, As shown in the figure below,
please note: the PCBboard, software and menus and functions are be the same, only the size of housing and
shape of the housing is different..
⚫ Integral type(TFT8210HM1)
⚫ Integral type(TFT8210HM2)
4
⚫ Integral type(TFT8210HS)
⚫ Integral type(TFT8210HB)
⚫ Separate type(TFT8210R)
5
2. Transmitter Operation And Parameter Setting
2.1 Keyboard Definition and Display
Figure A
Figure B
Left shift, parameter setting confirmation key and exit sub directory key
6
2.2 Transmitter Menu Structure
option: L/s L/m L/h m3/s m3/m m3/h USG/s USG/m USG/h Kg/s
Kg/m Kg/h t/s t/m t/h
PV Unit
default = m3/h ;define the unit of the flow rate
L (liter), h(hour), t(ton), s(second) , m(minute)
option: 0 1 2 3 ,default = 1
PV Decimal
define the decimal point position of the flow rate.
7
Float point: 99.00 ~ 0.00 % ,default = 0.0 %
This value is a percentage of flow range. for example, if this value is
setup to ten(10), then Equivalent to ten percent(10%) of flow range. If
Low alarm
the Absolute value of instantaneous flow < (flow range × 10%),then the
reansmitter willl otput the high alarm signal ,the contact of high alarm will
close 。
8
⚫ Caculation:Ajust analog output and zero value of flow meter
9
option: Low(Active low) High(active high)
default = Low(Active low)
this parameter is related to the "pulse width".
When set to“ Low”, the set value of the pulse width is used as the
Active level
pulse width of the pulse output low level.
When set to “high”, the set value of the pulse width is used as the
pulse width of the pulse output high level.
In general, this parameter does not need to be set.
option: 7 8 ,default = 8
Data Bit
Note: if use RTU protocol,prohibit to choose “7”
option: 1,2,3、6、8,10、15、20、25、32、40、50、65、80、
100、125、150、200、250、300、350、400、450、500、550、600、
Sensor Size 650,700、800、900、1000、1100,1200、1300,1400、1500,1600、
1700,1800、2000、2200、2400、2600、2800、3000 mm
default value = 50 mm
10
Floating point: 9.9000~0.0100 ,default = 0.1
This parameter is determined when the real flow calibration is carried
out. This parameter is only related to the sensor, which indicates the
characteristic value of the sensor.
k-factor
New Sensor K = the old Sensor K ×〔flowrate of the Calibrated
flowmeter ÷actual flow rate〕
After you finished the test by water , then need setup the final
K-fator at here.
default:0.0 , unit:mv.(votage)
When the fluid is still in the pipeline and completely filled with the
pipeline, the zero value is equal to the voltage difference between the
two electrodes of the electromagnetic flowmeter.
Actual Zero
Note: after the implementation of zero calibration, if the zero value
is high, please check the sensor。
In general, the zero value can not exceed 2mV. If the zero value is
more than 2mv , then the accuaracy of the flow mter is not ok .
the correction range: the default value is 0 ~ 2m/s (refers to the scope
of the role of nonlinear correction value)。
Linear Correction
default value =0.0 mv
修正值=(被校表流量-标准表流量)/标准表流量*当前被校表电压值
,
Float point: 999999~0.1 kg/m3 default = 1000.0 kg/m3
Density When you need to use the quality unit, set the exact density!
If the unit is Volume unit,then this item have no any affect.
11
If you setup the “Lower limit of empty-pipe” to 3500, the meaning
is that the flowmeter determines the current state to “empty
pipe(no water in pipe)” when the actual measurement value >3500.
If the actual measurement value <3500, then the flowmeter
determines that the pipe is filled with liquid.
The actual measurement value is appeared in the left of this menu.
Lower limit of empty
In some cases it may be possible to change this setting.
pipe (as shown in right
picture)
12
Filter time: default =0,option : 0、1、2
Note: In the normal state, the filtering time is “0”, and meet the
requirements of the corresponding time for 0.3 seconds.
If the flowrate is very instable , and the sensor is ok ,and this instability
condition is caused by the fluid, then you can change the ”filter time”
to “1”.
Filter-parameter setup
Filtering threshold:default =1,option: 1、2、3、4、5
⚫ Energy Parameter Setting:Define the relevant parameters of the energy meter, and please
note: this menu will appear when you purchase the energy meter.
13
Flow Rate Setup
Total Flow Setup
Caculation
Output Setup
Communcation
Figure 2
After the interface is shown in Figure 2,press or then you can choose different set items。
Press or move the cursor to the sub menu,press go to sub menu to setup the parameter,
for example , we need setup the “flowrate parameter”,when this item become light,pree , then will display
the menu as shown in figure 3 :
Figure 3
Press or to select the item which you want to modify,The selected item will brighten,if need return
to the menu as shown in figure 2,then press ;If you need to enter the next level of items,then press
to setup the parameters as shown in figure 4:
Figure 4
In this case , press or to modify the parameter,for example, as shown in figure 4, you need
change the flowrate unit from " m³/h ” to " m³/m ”,then press ,the flowrate unit will become to “ m³/m ”,
as shown in figure 5:
14
Figure 5
After you modify the parameter , if you need save,then press ,The system will prompt the "confirmation"
and "exit" option, as shown in Figure 6:
图6
In this case,press ,then save the setup and back(as shown in figure 3);if do not need save the
15
3. Wiring Diagram And Output Define
3.1 Integral Type Wiring Diagram (AC and DC Power Supply Type)
Pulse or requency
meter Power
supply
Current
meter
Note:
The driving ability of high alarm
and low alarm is 24VDC/30mA,
24vdc power
and can drive more than 800 ohm
impedance relays, valves and
alarm lights, etc。
24vdc power
16
3.2 Separate Type Wiring Diagram (AC and DC Power Supply Type)
thermal resistance
17
3.3 Battery Power Flow Transmitter wiring Diagram (3.6V lithium battery)
two core line (connect to the coil of sensor, three core wire with shield connection to the electrode of sensor .
Flow Rate
F= Upper Frequency Limit
Flow Range
Frequency output mode is generally used to control the occasion, because it reflects the percentage of traffic,
if the user is used for measurement occasions, it should choose the pulse output mode。
The frequency output is provided with an internal 24VDC power supply and NPN way。
If you need use the frequency output mode , then three parameters must setup:
⚫ Setup “upper frequency limit”
⚫ Setup “flow range” (corresponding to upper frequency limit)
⚫ Setup “Pulse equivalent value” = 0
18
pulse equivalent selection is too small, it will cause the upper limit of the pulse output, so the pulse output
frequency should be limited to the following 3000Hz. If the flowrate is small and the pulse equivalent is too large, it
will cause the meter to output a pulse for a long time。In addition, it must be explained that the pulse output is
different with the frequency output, the pulse output is a pulse equivalent to output a pulse, therefore, the pulse
output is not very uniform. The general measurement of the pulse should use the counter meter, not choose the
frequency meter。
The pulse output is provided with an internal 24VDC power supply and NPN way。
Flow Rate
I0 = 16 + 4.0
Flow Range
For 4 ~ 20mA signal system, the current zero is 4mA. Therefore, in order to improve the resolution of the
output analog current, the flow range of the flowmeter should be selected properly.
3.6V battery power EMF type have no 4 to 20 Ma signal output .
Grounding resistance
less than 10 ohm
Inet
19
20
4.2 Power On And Preheat
The flow meter need to be warm up for at least 15 minutes after power on .
Battery Life
Battery Type
Size ≤DN150 DN200≤Size ≤DN350 DN400≤Size ≤DN600
Internal Battery 33 months 30 months 26 months
External Battery 72 months 72 months 72 months
If your sensor size is big ,and need long battery life , then need install external battery..and different size , the
external battery capacity is different. If the size is greater than DN600, then the internal battery life is short than 26
months , so suggest use external battery .if you need install external battery , then ask usd , we can provide the
battery parameters and Installation method.
21
6.Appendix
Appendix 1: Flow Meter Transmitter Mode
P Only PCB
S Integral type, with 1608 shell
M1 Integral type, with 84mm-Abb shell
M2 Integral type, with 97mm-Abb shell
B Integral type, with big -size shell
1 3.6v battery
2 24vdc power supply
3 86vac to 265vac power supply
4 100vac to 420 vac power supply
R RS485 communcation
H HART communcation
22
Appendix 2: Electromagnetic Flow Meter Mode (TFT8210)
23
Appendix 3: Outline Dimension (Insert Electromagnetic Flowmeter)
24
Appendix "RS485 Communication Address Table
Note:
⚫ The start address data is 16 binary , if need 10 binary , then need convert.
⚫ The data bit is 32 bit.
⚫ If connect to PLC ,maybe the start address will add “1” , for example , if the start address is 0708 ,
then will becom 0709 ( for PLC)
25
Appendix 6: Fault Code
0x01 Invalid instruction code
0x02 Invalid register address
0x30 Parameter over limit
0x31 Super lower limit of parameters
0x32 Parameter selection error
0x40 Invalid register length
The register does not support the current
0x41
instruction code
0x42 Register not specified
0x43 Instantaneous flow unit does not exist
0x44 Total unit does not exist
0x45 Maximum frequency output over limit
0x46 Minimum frequency output
0x47 Maximum flow rate over the upper limit
0x48 Duty cycle over limit
0xFE Data frame confusion
0xFF Data frame check error
26
Appendix 8: Symbolic Code Definition
RTU 0
Modbus mode
ASCII 1
Even check 0
Check mode Parity check 1
No parity 2
1200bps 0
2400bps 1
4800bps 2
Baud rate
9600bps 3
19200bps 4
38400bps 5
7 bit 0
Data bit
8 bit 1
小数点后0位 0
小数点后1位 1
Decimal
小数点后2位 2
小数点后3位 3
DN1 0 DN600 27
DN1.5 1 DN700 28
DN2 2 DN750 29
DN3 3 DN800 30
DN4 4 DN900 31
DN5 5 DN1000 32
DN6 6 DN1100 33
DN8 7 DN1200 34
DN10 8 DN1300 35
DN15 9 DN1350 36
DN20 10 DN1400 37
DN25 11 DN1500 38
DN32 12 DN1600 39
DN40 13 DN1700 40
DN50 14 DN1800 41
DN65 15 DN2000 42
DN80 16 DN2100 43
27
DN100 17 DN2200 44
DN125 18 DN2300 45
DN150 19 DN2400 46
DN200 20 DN2500 47
DN250 21 DN2600 48
DN300 22 DN2700 49
DN350 23 DN2800 50
DN400 24 DN2900 51
DN450 25 DN3000 52
DN500 26
28