01. SPECIFICATIONS
02. DESCRIPTION
03. INSTALLATION
1. TRACEABILITY CERTIFICATE
INDICATOR
The system consists of 8 Nos seprate platform which is resting on 4No.s of highly
precession Double ended shear beam Loadcell of 10 ton capacity each . the Rails are
mounted on the I beam to avoid shearing of Load. 4 loadcell of platform is connected to the
Junction Box and loadcell output is taken to the programmable digital load indicator .
The Load indicator Model EW/PF/752/14 is a field programmable indicator designed to be used with strain gauge
based transducers to measure parameters like load, pressure, torque etc. It can take an input of 30 mV DC and
can read up to a maximum of 60000 counts with selectable resolution of 0 / 5. Each indicator has different address
starting from 01 to 08. The RS485 output of the indicator is connected to each indicator serialwise and it is
converted to RS232 through a convertor and it connected to the Computer. The Computer software program is
made to identify the each platform and scan the weight.
The indicator can be programmed through keyboard for range, resolution, calibration, zero track
value and decimal point. The front panel has got a digital display and a keyboard. The display
can indicate the following.
LOAD INDICATOR
LOADCELL
******
Programmable parameters
General
The power supply should be made available with a proper earthing within 2 meters of the indicator.
Connect the instrument to ac mains through the connector provided at the rear of the indicator. Mains
voltage to be used is given in the specification
INSTALLATION:_
Connect the transducer to the equipment through the connector provided at the back of the
indicator. Connect the Remote display. Switch on the Indicator.
Now the equipment is ready to be configured for measurement as given in the following
sections.
04. OPERATION & CALIBRATION
4.5. CALIBRATION :
Ensure that there is no load on the transducer. Keep TARE/STR key pressed and switch ON the
instrument with the switch provided at the back panel. The system will display `H-CAL’. Release
TARE/STR key once the ‘H-CAL’ has appeared. Now the equipment is ready for Hardware
TARE & CALIBRATION and will display a reading corresponding to initial unbalance of transducer.
Should the unbalance be more than 5% of the Range then switch OFF the indicator and open the
equipment. Switch ON and adjust TARE potentiometer on ADC card to zero the display.
Place a known weight of more than 60% of the capacity. Now the indicator should read the same value.
If not adjust the SPAN pot meter on the ADC Card to get the correct value an remove the
Load. This completes the Hardware Tare & Calibration operation and close the Indicator.
Switch the equipment ON again. It will show ” EUUI “ momentarily and then zero or some reading.
Now press TARE key, the display will show TARE . Now press the INC(^) key. The reading will be
zeroed and indicator will read zero value.
SOFTWARE CALIBRATION:-
With out any load on the load cell, make the reading zero by pressing the TARE & INC keys and load
the load cell with a known weight approximately 60 – 80% of the capacity. If the displayed value is
equal to the load then no further calibration is required. If the displayed value is not equal to the load
and the difference is more than +/- 10% of the actual load, the Hardware Calibration is required.
If the difference is less than +/- 10%, then proceed with the Software Calibration as given below.
Press the SET key, the display will show go to PASS WORD Mode and will show “ ----“.
Now Press SHIFT(>), INC (^), SET and STR keys successively. Now the display shows ‘S-CAL’
with ‘S’ blinking. If no change is required press SET key and the display will come out of the
Calibration mode will display the load.
If a change is required, press STR key and the display changes to ‘Xxxxx', which is the
previously programmed CAL VALUE, where `X' blinks, using INC (^) and SHIFT(>) keys , set the
display value equal to the Load on the load cell. Now Press `TARE/STR' key. It now comes to
weighing mode, display shows the load on the load cell.
Now the indicator is ready for operation. The indication displayed on the front panel is also
transmitted to the Remote Display and also through RS485 port to a computer.
The RS232 port transmits at 9600 baud rate No Parity 8 Data bits and 1 Stop bit.
The First character is the polarity of the indication. Space for +ve and – for negative readings.
The Second to Seventh characters are the indication from MSD to LSD.
05. SYMPTOMS & FAULTS
2. The display shows `OR’ a. Loadcell cable open (in this condition
(over range) indicator will not sense any key
(operation)
b. System overload.
c. Loadcell not connected in the same
sequence as shown in the drawing.
01. DESCRIPTION
02. SPECIFICATIONS
03. INSTALLATION
RESOLUTION : 10 Kg
NO. OF INDICATORS : 8
****
CALIBRATION CERTIFICATE GROUP = D
INDICATOR LOADCELLS
JOB ORDER : 049/08 MODEL NO : EW/PF/752/14 MODEL NO : EDSBS
CUSTOMER: M/s B.E.M.L SERIAL NO : 44708 CAPACITY : 10TON
CAPACITY : 7500Kg SL. NO. : 081157, 081158
RESOLUTION : 10Kg 081159, 081160
______________________________________________________________________________
Actual load Observed Reading Reference Indicator Error
in Kg in Kg reading in Kg in Kg
______________________________________________________________________________
TESTED BY :-
H.O.D :-
DATE :-
CALIBRATION CERTIFICATE GROUP = H
INDICATOR LOADCELLS
JOB ORDER : 049/08 MODEL NO : EW/PF/752/14 MODEL NO. : EDSBS
CUSTOMER : M/s B.E.M.L SERIAL NO : 44808 CAPACITY. : 10TON
CAPACITY : 7500Kg SERIAL NO. : 081169,081170
RESOLUTION : 10Kg 081171,081172
___________________________________________________________________________________
Actual load Observed Reading Reference Indicator Error
in Kg in Kg reading in Kg in Kg
___________________________________________________________________________________
TESTED BY :-
H.O.D :-
DATE :-
CALIBRATION CERTIFICATE GROUP = E
INDICATOR LOADELLS
JOB ORDER : 049/08 MODEL NO : EW/PF/752/14 MODEL NO. : EDSBS
CUSTOMER : M/s B.E.M.L SERIAL NO : 44908 CAPACITY. : 10TON
CAPACITY : 7500Kg SERIAL NO. : 081145,081146
RESOLUTION : 10 Kg 081147, 081148
_____________________________________________________________________________________
Actual load Observed Reading Reference Indicator Error
in Kg in Kg reading in Kg in Kg
______________________________________________________________________________________
TESTED BY :-
H.O.D :-
DATE :-
CALIBRATION CERTIFICATE GROUP = C
TESTED BY :-
H.O.D :-
DATE :-
CALIBRATION CERTIFICATE GROUP = F
INDICATOR LOADCELLS
JOB ORDER : 049/08 MODEL NO : EW/PF/752/14 MODEL NO. : EDSBS
CUSTOMER : M/s B.E.M.L SERIAL NO : 45108 CAPACITY. : 10 TON
CAPACITY : 7500Kg SERIAL NO. : 081149,081150
RESOLUTION : 10 Kg 081151,081152
_____________________________________________________________________________________________
Actual load Observed Reading Reference Indicator Error
in Kg in Kg reading in Kg in Kg
______________________________________________________________________________________
TESTED BY :-
H.O.D :-
DATE :-
CALIBRATION CERTIFICATE GROUP = A
INDICATOR LOADCELLS
JOB ORDER :049/08 MODEL NO : EW/PF/752/14 MODEL : EDSBS
CUSTOMER :M/s B.E.M.L SERIAL NO : 45208 CAPACITY : 10TON
CAPACITY : 7500Kg SERIAL NO. : 081165,081166
RESOLUTION : 10Kg 081167,081168
___________________________________________________________________________________
Actual load Observed Reading Reference Indicator Error
in Kg in Kg reading in Kg in Kg
____________________________________________________________________________________
TESTED BY :-
H.O.D :-
DATE :-
INDICATOR LOADCELLS
JOB ORDER : 049/08 MODEL NO : EW/PF/752/14 MODEL. : EDSBS
CUSTOMER : M/s B.E.M.L SERIAL NO : 45308 CAPACITY. : 10 TON
CAPACITY : 7500Kg SERIAL NO.: 081161,081162
RESOLUTION. : 10 Kg 081163,081164
___________________________________________________________________________________
Actual load Observed Reading Reference Indicator Error
in Kg in Kg reading in Kg in Kg
___________________________________________________________________________________
TESTED BY :-
H.O.D :-
DATE :-
CALIBRATION CERTIFICATE GROUP = B
INDICATOR LOADCELLS
JOB ORDER : 049/08 MODEL NO : EW/PF/752/14 MODEL NO. : EDSBS
CUSTOMER : M/s B.E.M.L SERIAL NO : 45408 CAPACITY. : 10 TON
CAPACITY : 7500Kg SERIAL NO. : 081153,081154
RESOLUTION : 10 TON 081155,081156
___________________________________________________________________________________
Actual load Observed Reading Reference Indicator Error
in Kg in Kg reading in Kg in Kg
____________________________________________________________________________________
TESTED BY :-
H.O.D :-
DATE :-
CALIBRATION CERTIFICATE
12000 12010 10
36000 36010 10
TESTED BY :-
H.O.D :-
DATE :-
JOB NO : 049/08
CUSTOMER : M/s B.E.M.L
SYSTEM : RAIL WHEEL WEIGHING SYSTEM
TESTED BY :-
H.O.D :-
DATE :-
JOB NO : 019/08
CUSTOMER : M/s B.E.M.L
SYSTEM : LORRY WEIGHBRIDGE
TESTED BY :-
H.O.D :-
DATE :-
07. 1. TRACEABILITY CERTIFICATE
07. 2. SYSTEM CALIBRATION REPORT
07. 3. LOADCELL CALIBRATION REPORT