Recent Changes To Reprocessing Validations For Reusable Medical
Recent Changes To Reprocessing Validations For Reusable Medical
Recent Changes To Reprocessing Validations For Reusable Medical
1
Spaulding Classification
Device
Patient Contact Example
Classification
No patient contact or
Non-Critical
intact skin contact
Mucous membranes
Semi-Critical
or non-intact skin
2
Process Overview – Reusable Devices
Low-, Intermediate-,
Sterilization Disinfection or High-Level
3
Overview – Reprocessing Validations
• Updates to Cleaning Validations
• Lifecycle Validations
4
Updates to Cleaning Validations
5
Cleaning Validation Steps
Soiling, Cleaning,
Disinfection/Sterilization
6
Updates to Cleaning Validations
• Cleaning validations have seen a lot of updates over the last few years with updated and new standards.
• Final Guidance for Industry and FDA Staff: Reprocessing Medical Devices in Health Care Settings: Validation
Methods and Labeling
o Initially published in 2015 and revised in 2017.
7
AAMI ST98 Updates and Changes Overview
HIGHLIGHTS OF MAJOR CHANGES:
8
AAMI ST98 – Simulated use Cycles
• Multiple, sequential simulated use and processing
cycles to simulate soil accumulation shall be performed. Contamination
o Sequential (soiling, drying, cleaning and disinfection
/sterilization (FULL CYCLE))
Manual
• Shall consider all worst-case use of the medical devices Sterilizer
(full range of motion/actuation, cauterizing, heating) Cleaning
6x
• Number of cycles NOT DETERMINED
9
AAMI ST98 – Added Controls from the FDA document to standard
• Controls align with FDA guidance document on labelling and validation methods for
reusable devices, 2015 (2017)
Validation
Control Definition
Type
Negative Device
Test Method Device that is not soiled but cleaned
Control
Device that is soiled but not cleaned. This shall be used to
Positive Device Control Test Method demonstrate that an appropriate amount of test soil was
applied to the devices
10
Sample Size and Controls
Sample size
3 samples per
1 sample undefined per 1 sample 1 sample 1-3 devices
ST98*
ST98
Extraction fluid
Testing lowest
extracted in same
Soiled concentration of Soiled Processed processed
container as test
test soil
articles
11
AAMI ST98 – Positive Device Control – Test Method
• For critical and semi-critical devices, a minimum of 3 positive controls
shall be used
• In some instances, a smaller size can be used however will have to be
scientifically justified
12
AAMI ST98 – Sample Size Selection
• Sample size should be determined based on:
• Medical device complexity
• Data reproducibility
• Criticality
13
AAMI ST98 – Data Reproducibility Example
14
Acceptance Criteria Update
AAMI ST98 ISO 15883-5 2021
Clinically Relevant
Recommended Level Clinically Relevant
Analyte Alert Level Action Level
Analyte
Protein ≤ 6.4 μg/cm2
Protein ≥ 3 μg/cm2 ≥ 6.4 μg/cm2
Hemoglobin ≤ 2.2 μg/cm2
Hemoglobin ≥ 1.0μg/cm2 ≥ 2.2 μg/cm2
Total Organic Carbon
≤ 12 µg/cm2
(TOC) Carbohydrate ≥ 0.9 μg/cm2 ≥ 1.8 μg/cm2
Carbohydrate ≤ 1.8 μg/cm2 Endotoxin ≥ 2.2 EU/device ≥ 20 EU/device
Adenosine ≥ 10 fmoles of ≥ 22 fmoles of
≤ 22 femtomoles/cm2 ATP
Triphosphate (ATP) ATP/cm2 ATP/cm2
TOC ≥ 6 μg/cm2 ≥ 12 μg/cm2
15
Non-Critical Devices
AAMI ST98:2022; Section 8 - Endpoints
16
Updates to Disinfection Validations
17
Disinfection Validation steps
Contaminating Disinfecting
Testing for
Samples with through Manual Extraction of the
Microbial
Microorganism or Automated Samples
Reduction
(Bacteria) Methods
18
Spaulding Classification for Chemical Disinfection
Patient Device Disinfection
Examples
Contact Classification Level
v
No patient Low-level or
contact or intact Non-Critical Intermediate-
skin contact Level Disinfection
Mucous
High-Level
membranes or Semi-Critical
Disinfection
non-intact skin
Sterile areas of
the body
Critical Sterilization
including blood
contact
19
Updates to Disinfection Validations
• American Association of Medical Instrumentation (AAMI)
o TIR12 – This document was updated in 2020.
20
Disinfection Levels and Acceptance Criteria
21
AAMI TIR12 – Update to Inoculum Preparation
• The updated AAMI TIR12 instructs to evaluate the four vegetative organisms separately and not in a
mixed suspension.
o Mixed suspensions can engage in competitive inhibition.
Staphylococcus aureus
Escherichia coli
Klebsiella pneumoniae
Pseudomonas aeruginosa
22
Non-Critical Devices in AAMI TIR12
Minimum
Patient Contact Expected Contamination Disinfection
Level
Intact skin of patient
Intermediate-
or Blood or other potentially infectious materials
level disinfection
Health care personnel
Low-level
None Blood or other potentially infectious materials
disinfection
Intact skin of patient
or
Health care personnel Not blood or other potentially infectious
*visibly clean
or materials
None
23
TIR12:2020 – Annex F
• Gives some guidance regarding distinguishing between Intermediate level
disinfection (ILD) and low-level disinfection (LLD).
• ILD
o It is appropriate for use on certain patient care items that contact intact skin and on
environmental or medical equipment surfaces (e.g., adjustment knobs, handles, buttons, or
levers on X-ray equipment, instrument trays and carts, and dental units) which are generally
not intended to come into direct contact with the patient.
o In addition, these environmental and medical equipment surfaces can be touched repeatedly
by health care personnel during procedures involving parenteral or mucous membrane
contact
• LLD
o It is appropriate for use on non-critical devices and equipment surfaces which ordinarily do
not contact patients during routine use and are not expected to be contaminated with blood
or Other Potentially Infectious Materials (OPIMs)
24
Sample Size and Controls
Positive
Test article Neutralization Cytotoxicity
control
Undefined
1 sample per AAMI 1 sample 1-3 samples
TIR12
unprocessed Processed
25
Sample Size
• AAMI TIR12:2010 previously gave a recommended sample size of three test replicates (test articles)
and one concurrent positive control.
26
Updates to Sterilization Validations
27
Steam Sterilization Validation Steps
Sterility Assurance
Level (SAL) Dry Time
Temperature Profiling
Validation with Validation with full
full cycle parameters
half or full cycle cycle parameters
parameters
28
Recent and Upcoming Changes
• AAMI TIR12:2020 - Minimum drying times for steam sterilization were removed from Table B.1 in
Annex B.
• The next revision of ISO 17665 (Sterilization of health care products — Moist heat — Requirements for
the development, validation and routine control of a sterilization process for medical devices) is going
to Final Draft International Standard (FDIS) stage, which is the final stage before it is published.
• ISO 22441 (Sterilization of health care products — Low temperature vaporized hydrogen peroxide —
Requirements for the development, validation and routine control of a sterilization process for medical
devices) has been published and has been recognized by US FDA as a consensus standard.
29
Lifecycle Validations
30
Is End-of-Life Testing Needed?
31
End-of-Life Cycles
• What is the appropriate number of cycles?
Patient use
Cycles?
Disinfection
Cleaning
/sterilization
32
End-of-Life Testing
ISO 17664:2017
• MDM must determine if reprocessing leads to a degree of degradation that will limit the useful
life
• MDM must determine the number of reprocessing cycles or other indication of the end of it’s
life
• Labeling should inform how many times the device can be reused based on testing
• Provide a mechanism or method to ascertain whether the device has exceeded its use life and
if the device is functioning properly
• Include appropriate disposal
33
MDR
• CHAPTER III: Requirements regarding the information supplied with the Device
o Information in the IFU
o Information that allow the user to be informed of any warnings, precautions, measures
to be taken and limitations to use regarding device.
34
End-of-Life Testing
• ISO 10993-1:2018 Section 4.7 and 4.8:
o 4.7 The biological safety of a medical device shall be evaluated by the manufacturer over the whole
life-cycle of a medical device.
o 4.8 For re-usable medical devices, biological safety shall be evaluated for the maximum number of
validated processing cycles by the manufacturer.
35
Standards for Reprocessing
• AAMI • ASTM • ISO
• AAMI ST98 (cleaning validation) • ASTM F3208 (test soils) • ISO 17664 (cleaning/disinfection)
• AAMI TIR12 (cleaning, disinfection, • ASTM F3293 (application of test • ISO 17665 (sterilization)
sterilization) soils)
• ASTM F3438 (Analytes test • ISO 15883 (washer/disinfector)
• AAMI ST58 (HLD &Chemical sterilization) method)
• TIR68 (ILD,LLD) • ASTM F3321 (Extraction test
• AAMI TIR34 -ST108 (water for soils for cleaning)
reprocessing)
• AAMI ST84 (Flexible Endoscope)
• AAMI TIR99 (TEE, dilators)
• AAMI ST79 (Steam sterilization)
36
Curated content from industry thought leaders to help you
mitigate risk and go to market faster.
38