Cleaning of Microwave Digestion Vessel
Method
Equipment
Type Manufacturer
speedwave Entry Berghof Products + Instruments GmbH
DAP-60K Berghof Products + Instruments GmbH
Reagents Acid Volume
HNO3 (65%) 9.0 mL
HCI (37%) 1.0 mL
Procedure
Put 9.0 mL of HNO3 and 1.0 mL of HCl into the digestion vessel. Close the vessel and heat in
the microwave with the following program.
Temperature Program** Step T [°C] Ta [min] Time [min] Power [%]
1 170 5 5 90
2 200 1 15
3 75 1 10
4
5
Note: To avoid foaming and splashing wait until the vessels have cooled to room temperature (about 20 min). Carefully open the digestion vessel in a
fume hood wearing hand, eye and body protection since a large amount of gas will be produced during the digestion process.
Analysis
Blank concentrations of new DAP-60+ vessels after 3 subsequent blank digestions.
Element 1. Digestion* 2. Digestion* 3. Digestion*
Li µg/l 0.2 <0.025 <0.025
Be µg/l <0.25 <0.25 <0.25
Mg µg/l 20.1 5.0 1.9
Al µg/l 19.8 <0.25 <0.25
V µg/l <0.025 <0.025 <0.025
Cr µg/l 2.6 0.3 <0.05
Mn µg/l 0.6 0.04 <0.05
Co µg/l 0.1 <0.025 <0.025
Ni µg/l 1.3 0.1 <0.025
Cu µg/l 0.53 <0.025 <0.025
APPLICATION NOTE | SPEEDWAVE ENTRY | ENVIRONMENT, GEOLOGY S. 2/2
Note: This application serves only as a guide line and may need to be optimized for your sample.
** The program “Cleaning Vessel” is preinstalled on the controller.