SPECIFICATIONS.
Physical Specification.
-Dimensions at least 300mmx300mmx250mm.
-Display Resolution (H) x (V) at least 1200x800 pixel
-Screen 12.1 colors active with touch screen.
-Adjustable Brightness
Ventilation Specifications;
-Able to perform for adult, paediatric and neonatal
Functions:
-perform oxygen therapy
-able to perform nebulization
-perform non invasive and invasive ventilation
-difference alarm systems
Ventilation Mode;
-V-A/C (Volume assist/control)
-P-A/C (Pressure assist/control)
-V-SIMV (Volume - Synchronized Intermittent Mandatory Ventilation)
-P-SIMV (Pressure - Synchronized Intermittent Mandatory Ventilation)
-Duo Level (Duo Level Ventilation)
-CPAP (Continuous Positive Airway Pressure)
-PSV (Pressure Support Ventilation)
-APRV (Airway Pressure Release Ventilation)
-PRVC-SIMV (Pressure Regulated Volume Control-Synchronized
Intermittent -Mandatory Ventilation)
-AMV (Adaptive Minute Ventilation)
-NIV (Non-invasive Ventilation)
-Apnea Ventilation
Environmental specifications:
-Temperature; 5 - 40 °C (operating); -20 to 60 °C (storage and transport,
O2 sensor: -20 to 50°C)
-Relative humidity; 10 - 95 % (operating); 10 - 95 % (storage and transport)
-Barometric pressure; 62 - 106 kPa (operating); 50 -106 kPa (storage and
transport)
External AC power supply
-Input voltage; 100 - 240 V
-Input frequency; 50/60 Hz
-Input current; 2.7 - 1.1 A
-Fuse; T3.15 AH/250 V
physical specifications:
-Weight; 3-5kgs (excluded battery and accessories)
-dimensions at least 300mmx220mmx190mm
-display size at least 1200x800pixel with mult operation touch screen
- Up to 8 waveforms or more
ECG.
-Meet standards of IEC 60601-2-27 and IEC 60601-2-25.
-Lead set 3-lead: I, II, III
5-lead: I, II, III, aVR, aVL, aVF, V
12-lead: I, II, III, aVR, aVL, aVF, V1 to V6
-Automatic 3/5/6/12 - lead recognition.
-Input signal range ± 10 mV (p-p)
-Electrode offset potential tolerance ± 800 mV
-Sweep speed 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
-Waveform format Standard, Cabrera
-Bandwidth Diagnostic mode: 0.05 to 150 Hz
-Monitor mode: 0.5 to 40 Hz
-Surgical mode: 1 to 20 Hz
-ST mode: 0.05 to 40 Hz
-CMRR Diagnostic mode: > 90 dB
-Monitor, Surgical, ST mode: > 105 dB
-Pace detection Amplitude: ± 2 mV to ± 700 mV
-Width: 0.1 to 2 ms
-Rise time: 10 to 100 μs
-Defib. Protection Withstand 5000V (360J) defibrillation
-Recovery time <5 s
Heart rate.
-HR range Adult: 15 to 300 bpm
-Pediatric/Neonate: 15 to 350 bpm
-HR resolution 1 bpm
Arrhythmia analysis
-Intended use for adult, pediatric and neonate.
-Multi-lead, 25 classifications. Asystole, VFib/VTac, Vtac, Vent. Brady,
Extreme Tachy, Extreme Brady, Vrhythm, PVCs/min, Pauses/min, Couplet,
Bigeminy, Trigeminy, R on T, Run PVCs, PVC, Tachy, Brady, Missed Beats,
PNP, PNC, Multif. PVC, Nonsus. Vtac, Pause, Irr. Rhythm., Afib (for adult
only).
Respiration.
-Lead I or II, auto
-RR range 0 to 200 rpm
-RR accuracy ± 1 rpm (0 to 120 rpm)
-RR resolution 1 rpm
-Sweep speed 3 mm/s, 6.25 mm/s, 12.5 mm/s, 25 mm/s,
50 mm/s
-Apnea time 10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s
spo2
-Meet standards of ISO 80601-2-61.
-SpO2 range 0 to 100 %
-Perfusion indicator
-refreshing rate ≤ 1 s
pulse rate
-PR range 20 to 300 bpm (from SpO2), 20 to 350 bpm (from IBP), 30 to
300 bpm (from NIBP)
Refreshing rate ≤ 1 s
temperature
-Meet standard of ISO 80601-2-56.
-Technique Thermal resistance
-Channels 2 channels
-Temp range 0 to 50 °C (32 to 122 °F)
Refreshing rate ≤ 1 s
alarms
-Audio/voice alarming.
-video/light alarming.
-Provide Alarm Sight info graphic alarm indicator.
power
-Line voltage 100 to 240 VAC (±10 %)
-Maximum current 2.0A
-Frequency 50/60 Hz (±3 Hz)
-Battery Rechargeable lithium-ion battery, 2600mAh/4500mAh
Rechargeable smart lithium-ion battery 5600mAh
Environment requirement.
-Temperature Operating: 0 to 40 °C , Storage: -30 to 70 °C, Storage: -20 to
60 °C,
-Humidity Operating: 15 to 95 % (non condensing) Storage: 10 to 95 %
(non condensing)
-Barometric Operating: 427.5 to 805.5 mmHg (57 to 107.4 kPa) Storage:
120 to 805.5 mmHg (16 to 107.4 kPa)
physical specifications;
-Dimensions at least (WxHxD) 250x90x150mm
-Net weight not more than 4kgs
display
-Screen 3.5 TFT color
-Adjustable brightness
-information to be displayed on screen a)delivery rate of infusion,
b)current infusion, c)volume to be infused(VTBI), d) IV set brand,
e)pressure limit, f)battery capacity, g)drug name h)remaining time i)
alarms.
Infusion set.
-Automatically recognition of infusion set brand after calibration.
-Automatically and manual bolus delivery
Functional spectrum.
-Automatic calculation of the delivery rate based on different modes
-infusion pause/standby while present parameters remain stored.
-drug library at least up to 1500 drug names.
-History record up to 2000.
- Able to transmit data to computerized materials
Safety systems.
-Detection of occlusion pressure and automatic stop infusion.
-air detection
-multi bubble or single bubble detection
Battery.
-Rechargeable
-Able to work at least 180minutes out from power source.
-At least 4500mAh battery capacity.
physical specifications;
-Dimensions at least (WxHxD) 250x90x150mm
-Net weight not more than 4kgs
display
-Screen 3.5 TFT color
-Adjustable brightness
-information to be displayed on screen a)delivery rate of infusion,
b)current infusion, c)volume to be infused(VTBI), d) IV set brand,
e)pressure limit, f)battery capacity, g)drug name h)remaining time i)
alarms.
Infusion set.
-Automatically recognition of infusion set brand after calibration.
-Automatically and manual bolus delivery
Functional spectrum.
- Able to operate with all size of syringes while Syringe 5 ml: 0.1-100 ml/h
(in 0.01, 0.1, 1, 10 ml/h increments), Syringe 10 ml: 0.1-300 ml/h, Syringe
20 ml: 0.1-600 ml/h, Syringe 30, ml: 0.1-800 ml/h, Syringe 50/60 ml:
0.1-1200 ml/h
-Automatic calculation of the delivery rate based on different modes
-Syringe pumping pause/standby while present parameters remain stored.
-drug library at least up to 1500 drug names.
-History record up to 2000.
- Able to transmit data to computerized materials
Safety systems.
-Detection of occlusion pressure and automatic stop infusion.
-air detection
-multi bubble or single bubble detection
Battery.
-Rechargeable
-Able to work at least 180minutes out from power source.
-At least 4500mAh battery capacity.
SUGGESTED BRAND AND MODE; MINDRAY, BENEFUSION SP5 series
syringe pump
Physical specifications;
-Size & Weight, Sturdy & light weight machine less than 5kgs
- Should have carry handle for portability
- Should be compact
Power Supply
- Compatibility with mains 220-240 V (normal), 50/60 Hz power supply
- High performance Li-ion rechargeable battery with builtin charger.
-Equipment should have sufficient battery backup for taking minimum 100
ECGs without AC power.
- Digital filtering to remove interference from power line, muscle tremor
etc
ECG recording
- ECG recording with 12 leads
- Standard Leads (the limb leads or bipolar Limb leads: I, II & III).
Augmented Limb Leads a. Standard Leads (the limb leads or bipolar Limb
leads: I, II & III) , Augmented Limb Leads- (aVL, aVR and aVF) AND Chest
Leads ( the unipolar or V leads)- from V1 to V6 Simultaneous.
- Recording speed selection of 25 mm/ sec and 50 mm/ sec with facility
for speed selection
- Automatic adjustment of baseline for optimal recording
- Should have different filters like Baseline Filter, EMG Filter & AC Filter
- Multiple operating modes -automatic, manual and rhythm
Standard Accessories
- The machine should be supplied with power cord , patient cable, user
manual and warranty card, Operation Manual with user demonstration
video CD, interpretation manual & 10 thermal recording paper rolls , and at
least 5 bottles of jelly
- And but not limit with Two sets each of: patient cable, chest electrodes
– Both adult and paediatric (2 sets each), limb electrodes – Both adult
and paediatric (2 sets each)
Safety Profile
- Should be provided with terminal for good earth connection to preclude
electric disturbances while recording
- Must have a safety certificate or valid detailed electrical and functional
safety test report from a recognized competent authority
- Copy of the certificate / test report shall be produced along with the
technical bid.
SUGGESTED BRAND AND MODE; MINDRAY, BENEHEART R12
Functions specifications.
-The equipment must be capable of operating in B, M, Doppler, Color flow
and Power Doppler modes
- The machine include at least a 19" LCD monitor for both excellent
images viewing as well as providing for workflow and productivity
features
- The LCD monitor mounted on an articulating arm that moves side-to-side,
forward and backward.
- The machine includes a minimum 10 inch Touch-screen LCD with
context sensitive menus to facilitate productivity.
- The system have minimum three universal active probe Ports in a
convenient, easy to access location to maximize the availability of needed
probes, with pin less connector.
- Ability to store images by patient and include the ability to review images
from different exam dates.
-support the ability to store digital data in complete Raw Form, that allows
to optimize imaging parameters such as B Gain, Dynamic Range, Speckle
Reduction levels, Doppler Gain, Doppler Base Line on old Images & old
loops recalled from the image archive.
- allow for live image and archive images side-by-side or quad display on a
single monitor. This display allows any type of image – B-Mode, Color, or
power Doppler on either side.
- Automatically invoke the correct mode and imaging parameter and
advance to the next step within the examination with a one-bottom
operation. Contrast Ultrasound Capability (CEUS) with Times Intensity
Curve Graphs.
- provide a display zoom function on frozen images.
Physical specificitaions
-should be portable with flexible wheels with locks
-Dimension size at least 35x35x7cm
-Net weight <10kgs
-12 LED screen.
Rechargable battery
- At least 8000mAh built in battery that continue scanning at least 3 hours
without power.
Uses of optinal probe in single active universal port.
-linear probe
-curved probe
-pediatric probe
-cardiac probe.
Output interface
-USB Video output
-SVGA Video output
-PAL Video output.
SUGGESTED BRAND AND NAME: MINDRAY,ME8 ULTRASOUND
Physical specifications.
-ECG Monitoring optional button
-CPR monitoring optional button
-manual mode button
-Self test mode button.
-beside printing optional
-dimensional size of screen at least 220x300 pixels
Shock Sequence.
-Constant or escalating, Programmable (factory settings).
Functions
-Adult and pediatric settings. For pediatric 50j (Average for 50 Ω at low
level energy), 75j (Average for 50 Ω at high level energy). For adult
170j (Average for 50 Ω at low level energy), 270j (Average for 50 Ω at
high level energy)
-Biphasic output waveform
-Automatic ECG (VT/VF) detection and analysis.
-Automatic impedance compensation based on patient.
-PaediatricsPaediatric dose attenuation.
-Automatic switch between AED and CPR modes based on analysis
Battery
-Rechargeable, Battery type LiSO2 or LiM no (depending on the model
supplied).
-The internal battery, when full supports at least 140 full discharges at
200J, or two hours continuous ECG monitoring.
Environmental conditions.
-Operating conditions: 0°C - 50°C
- Storage conditions: -10°C - 50°C
SUGGESTED BRAND AND MODE; MINDRAY, BENEHEART D3
Physical specifications.
-Bar code scanning for cartridge insertion
-User Input button or Keyboard or screen touch
-USB port for data transfer
-Cartridge insertion port
-Attached or separated printer for scanning and print results
-screen size at least 10x10cm
-Dimension < 25x10x5cm
Functional specifications,
-Able to perform blood gases, chemistry, electrolytes and hematology
tests.
-Multi-parameter card Single-use test card offer a broad menu of tests
on a single, portable platform. Each test card has a unique
combination of biosensors to suit a wide range of clinical needs.
-Analysis Time less that 2 min
-Calibration every sample
Built in battery that may perform at least 3 hours out from power source.
Sample
-Arterial or venous whole blood in either 1-10cc syringe.
Physical specifications.
-4 rails, 2 at top, left and right and 2 at bottom, left and right
-size dimensional at least 210cm length, 100cm width and 60cm
height.
-Material, plastic mixed with Mild Steel
-Number Of Wheels 4.
-Mattress size may correspond with mentioned bed size.
-working load at least 250kg.
-IV pole 4 hooks/drip stand pole, one in each corner of the
bed
- At least 2 Drainage/urinary bag hook left and right side of
the bed.
- Has degree angle indicator.
-Has auto weighing measure.
Functions specifications.
-using remote control and built-in button control to perform.
-perform height adjustment, backrest, footrest, Trendelenburg, and
anti-Trendelenburg.
-Has cardiac set and CPR set to quick cardiac arrest performance
-Head Elevation up to 0~85°
-Knee Elevation up to 0~32°
-Auto-chair
-Outer / Inner Siderail Controls
-Normal Lockable Castor
-Volts & Frequency AC100 ~ 240V ; 50 / 60 Hz
Built-in battery able to last at least 3 hour out from source of power
Physical specifications.
-Monitor Size: 3 Inch
-Video Aspect Ratio: 4:3
-Power: 3.7VDC
-Internal Memory: 16G TF Card
-Rotation Angle: Multi-rotational display
-Rechargeable Battery
-Camera Chip: CMOS
-Light Source: White LED
-Video Resolution at least 320 x 240 (QVGA) pixels per frame
-Video Screen: OLED Display
-Video Screen at least Size 6.096cm / 2.4" diagonal
Functions specifications
-Able to apply to Adult and Neonatal.
-Attached Adult baton with disposable Mac Blades of size 2, 3, 4.
-Attached Neonatal baton with disposable Miller Blades of size 0, 00,
1.
-Computerized Power Management System: Auto Shut off, Auto white
balancing.
-Portability of hand-held video laryngoscope.
Battery.
-Batteries: 3AAA
-Battery Life: <90 min. (confirm with Power Indicator status)
Physical specifications.
-3 projection sizes
Attachment
- should has Vein Viewer
Functions specifications.
-Five projection colors and inverse color mode for every color
sizes
Technical specifications.
-Weight <450 g
-Dimensions: 240 mm x 60 x 60 mm
at least 5 hours