Usability Engineering For Medical Devices
Usability Engineering For Medical Devices
custom-medical.com/en/knowledge/usability-engineering-for-medical-devices-according-to-iec-62366-1/
How exactly does the Usability Engineering process work? What does IEC 62366-1 require
of you? What exactly do you have to prove?
In this article, we clarify exactly these questions and take a step-by-step look at the Usability
Engineering process. We clarify what you, as a manufacturer of medical devices, have to
achieve in order to be able to prove good Usability. In doing so, we look again and again at
IEC 62366-1 and show you which parts you need to know and observe. By the end of this
article, you will have a comprehensive understanding of the recommended Usability
Engineering process and know what is required of you and how to meet (and prove!) it.
Before we dig deeper into the required process, we would like to clarify what IEC 62366-1
actually defines as Usability Engineering. For this we look at the definition from the text:
Further, the note adds, “Achieving adequate USABILITY can result in acceptable RISK
related to use”.
Usability Engineering for medical devices thus pursues the goal of eliminating unacceptable
risks to patients, users, and third parties through user-centered Usability Design. It aims to
find out Use Errors and then minimize them.
To get your medical devices regulatory approval in Europe, it is a prerequisite that you meet
all the requirements of the MDR. These include the general performance and safety
requirements (Annex I), which you can implement by complying with IEC 62366-1. So the
Usability Engineering process is a very good tool to comply with an important part of the
MDR requirements.
The importance of the IEC 62366-1 Usability Engineering Process for the development of
medical devices becomes clear when you realize that IEC 62366-1 is recognized by the
regulatory bodies of the different markets. The FDA, for example, lists the standard as a
Recognized Consensus Standard.
1/10
The strong normative regulation of product development makes sense when one considers
that misuse of medical devices can lead to death in the worst case scenario. The wrong
interpretation of data, incorrect or too late treatment due to operating errors – all this should
be avoided!
In order to be able to assess the risk of the product correctly, the Usability Engineering
process is closely linked to Risk Management. The Usability experts are responsible for the
safety of use, while the Risk Management is responsible for the safety of the complete
product. The Usability Engineering process (like all other processes) must be accompanied
by certified personnel.
Tip: Since you have to perform Usability Engineering including the proof of safe use (e.g. via
formative and summative user tests), it is also worthwhile for you to think about improving
the User Experience of your medical product. With the right approach, this can be done
simultaneously. This way, you can take advantage of the opportunity that the mandatory
human-centered development process offers you to better position yourself against the
competition with an outstanding UX. However, aspirations for a better UX are not a
requirement of IEC 62366-1.
You have already recognized it by the three exclamation marks: This info is especially
important! To put it bluntly: No (correct) documentation, no clearance of your product.
Therefore, before we look at the process in detail, we would like to point out that all your
undertakings within the Usability Engineering process must be documented in the so-called
Usability Engineering File.
The Usability Engineering File is defined by IEC 62366-1 as: ” set of RECORDS and other
documents that are produced by the USABILITY ENGINEERING PROCESS “. (source: IEC
62366-1:2021-08, chapter 3.18)
The form is relatively free: Chapter 4.2 states that the parts forming the contents of the
Usability Engineering File may be parts of other documents and files. No further information
about the form is given.
2/10
The process – step by step according to IEC 62366-1
Let’s take a look at the Usability Engineering process according to IEC 62366-1. The graphic
may seem a bit overwhelming at first glance, but on closer inspection it is far less intimidating
than initially thought. One thing first: The graphic shows both the process for Usability
Engineering (right area) and that of Risk Management (left area).
We will now ” walk through” this process step by step, taking a look at the standard every
now and then.
3/10
4/10
Step 1: Prepare Use Specification
Even a Usability journey begins with the first step. In the Usability Engineering process this
first step is called “Prepare Use Specification”. Directly with the first step you have to start
the documentation of all actions, because the compliance is checked by inspection of the
Usability Engineering File (source: IEC 62366-1:2021-08) – this means: You have to record
all your steps of the process in a Usability Engineering File (you remember: If you don’t
document, you fail).
What does the Use Specification consist of? IEC 62366-1 requires the following content to
be specified:
The Intended Medical Purpose. This is where you specify what the medical device is to
be used for. The intended medical purpose can also include an exclusion (i.e. clear
information about what the medical device must not be used for).
Intended User Profile. This is about who the product is intended to be used by.
However, it is not enough just to name the user group, you must also describe it in
terms of the relevant criteria.
The Intended Patient Population. On whom will the medical device be used? The
standard provides exemplary delimitation criteria for the patient groups. These include
health status and age.
The Intended Use Environment. In which context will the medical device be used? The
Intended Use Environment must also be described in the relevant criteria.
The Type of Tissue/ Part of Body. The purpose here is to specify the body part or type
of tissue on which the medical device is intended to be used.
The operating principle of the medical device. How does the medical device work and
how does it perform?
The standard does not specify how you can develop this content. We recommend a well-
documented context of use analysis, especially in the area of the intended user and patient
groups and the environment of use.
Chapter 5.2. requires you to identify, as part of the risk analysis, all parts and characteristics
of the User Interface that are related to the safety of the medical device. (source: IEC 62366-
1:2021-08; chapter 5.2).
This means: In the course of your Usability Engineering you have to define and find out
which properties of the User Interface can be hazardous for the user, for patients and for
third parties. From this, you then derive use errors that can occur during use. An example:
5/10
You design an Interface that displays critical measurement data. An assumed operating error
would be an incorrect or omitted interpretation of displayed data due to a lack of prior
knowledge on the part of some users.
For risk analysis, the standard refers to IEC 14971:2019 (Risk management). In this
standard, you will find further guidance on how to more easily identify the security-related
aspects of your Interface.
A very useful tool for identifying use errors is the Task Analysis, as well as the Use Related
Risk Analysis, which is required to be performed for the US market. This helps you to find out
the safety-related characteristics of your User Interface. The goal of the Task Analysis is to
find out the security risks of the usage. The safety risks of the entire product (i.e. also factors
such as possible radiation, etc.) are found out by Risk Management.
Chapter 5.3 of the standard deals with “Identifying known or foreseeable hazards and
hazardous situations”. Once again, you are responsible for the aspects of the product that
are relevant to use ( i.e., everything Usability-related).
In determining the hazards and hazardous situations, the standard requires you to consider
the following three things:
– All information about Hazards and Hazardous Situations known from medical devices of
similar type. This can include competitor products as well as own products of an
older generation.
All foreseeable hazards and hazardous situations that may affect users, but also patients
and third parties, must be identified.
The associated hazard-related use scenarios are now derived from the hazards and
hazardous situations identified in the previous step. For your Usability Engineering File, you
must identify and describe these. According to the standard, the description must include
the following:
6/10
– The sequence of the tasks.
So not only do you need to identify the Hazard-Related Use Scenarios, there also needs to
be an assessment of the severity of the potential harm. Risk Management then decides
which of the risks are acceptable and for which risks you need to address their reduction and
control (i.e., it identifies the parts of your product that need to be revised to control the risk
during use). Remember, the MDR requires you to keep the risk as low as possible without
negatively impacting the risk-benefit ratio.
But which scenarios do you have to choose? Either you test all hazard-related scenarios, or
you test a reasoned selection based on the severity of potential harm and other
circumstances specific to the medical device and the manufacturer (source: IEC 62366-
1:2021-08, Chapter 5.5).
According to the standard, you must document not only the selection of Hazard Related Use
Scenarios for your Summative Evaluation in the Usability Engineering File, but also the
chosen selection scheme and sufficient justification for the choice of the respective scheme.
The standard requires three content points from you for the User Interface Specification:
7/10
The User Interface Evaluation Plan is a document that defines how you will test whether the
Interface of your medical device poses hazards. In other words, you determine which parts of
the Interface will be tested and how. You must record the methods for all formative
evaluations and the summative evaluation. The User Interface Evaluation Plan must be part
of the Usability Engineering File.
If you decide to use Usability tests, IEC 62366-1 requires detailed documentation of the test
conditions. This includes the intended user groups (including a justification for each
grouping), the test environment (and other usage conditions), possible supporting
documentation during the test, and (if necessary) the user training measures and their time
interval between training and application.
The standard leaves it open to you whether you test in a simulated or real use environment.
You need to evaluate many parts of your interface with Usability tests (or other appropriate
methods). For example, you may need to test the correct user response to an error
message. Some parts, however, can also be verified by formally matching the requirements
to the Interface (e.g., a specific size of a design element). This can be done by a
Usability verification.
The goal here is to implement all the design recommendations derived from the previous
steps into an initial design that will allow you to conduct the formative evaluations that will
soon follow.
Your goal now is to find out whether your product generates (critical) use errors. These must
then be minimized or eliminated (where possible) through the design. So, you also check if
your product development is on the right track. Your Mindset: You want to learn and
understand which use errors your product still brings to light. The goal is to make your
product better and, above all, safer.
8/10
You are free to choose the methods for implementation. However, it is recommended to test
formatively with usability tests, if the summative evaluation is also a usability test. This way
you are safe from unpleasant surprises during the final test.
Were critical use errors found? Are further product improvements necessary? If you have
answered “yes” to either of these questions at this point, go back as far as necessary in the
Usability Engineering process and iterate through it from this point on until no new and
critical problems occur during use. If this is the case, you are ready for the last big step:
performing the summative evaluation.
Step 10: Perform summative Evaluation of the Usability of the User Interface:
The last step of the Usability Engineering process is the execution of the summative
evaluation including its post-processing. For the final and evaluative test, your users must be
trained as they would be after market launch. Likewise, your product must be ready or in pilot
production to perform the summative evaluation.
The goal is to prove that the medical device (and its Interface) can be used without
unacceptable residual risks for users, patients and third parties. The hazard-related use
scenarios previously defined by you are tested.
In addition, you can also use existing data: For the Summative Evaluation, the manufacturer
may use data obtained from summative evaluations of devices with comparable user
interface, together with a justification of how these data can be applied. (Source: IEC 62366-
1:2015, Chapter 5.9)
After implementation, the data is evaluated to the effect that you have analyzed all the use
errors that occurred and the resulting (possible) effects. If a hazardous situation could arise
as a result of one of the use errors made, a root cause analysis must be performed. To do
this, use the data obtained from the observations and follow-up interviews of the
study participants.
If new use errors, hazards or hazardous situations have occurred, you must weigh up
whether or not these make it necessary to revise the product. Here again is the urgent
recommendation to extensively test all hazard-related use scenarios in formative tests to
ensure that no unpleasant and costly surprises arise in the summative evaluation.
If your product fails, the summative evaluation becomes formative and the Usability
Engineering process continues until your product is safe enough for the summative study.
The standard then requires you to document why improvement is not necessary or not
practical. (source: IEC 62366-1:2021-08, Chapter 5.9)
9/10
If the residual risk has been determined based on the data, you have reached the end of the
Usability Engineering process.
For more tips on conducting summative evaluation according to IEC 62366-1, see our article
“17 Dos & Dont’s for the execution of the summative Usability evaluation“.
Conclusion
The Usability Engineering process is described in great detail in IEC 62366-1 and is always
closely linked to Risk Management. Correct implementation and complete documentation are
important! Everything is geared towards the assessment and control of use hazards.
If you have gone through the process conscientiously and documented everything correctly,
there will be no more stumbling blocks for your medical device in the area of Usability.
For which medical device do you need to apply Usability Engineering? Comment on this post
or get in touch via our contact form.
You are welcome to discuss your specific case with us in a free strategy meeting. We look
forward to hearing from you!