TSU109 e

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

A Meggitt company

vibro-meter
certified by

TSU 109

Charge Generator, Type TSU 109

FEATURES DESCRIPTION
• Universal charge and voltage generator, The widespread use of airborne, nuclear and
0 to 1000 pC, 0 to 1000 mV industrial vibration monitoring systems based on
piezoelectric transducers has brought about the need
• Frequency : Selectable 1 to 999 Hz for a robust, portable test set capable of completely
• Simulates piezoelectric transducers or velocity checking the installation.
transducers Vibro-Meter's TSU 109 M1-01 Charge Generator
permits in-situ testing and calibration of remote
• Allows system re-calibration and transmission
charge converters, continuity testing of cables,
line check
connectors, electronic units, and their alarm settings
• Robust portable execution, reliable in operation and filter functions.

REAR VIEW OF THE TSU 109

LEMO connector
Type F1S 303

©Vibro-Meter SA / 280-001 / 11.02 / E 1/2


Charge Generator TSU 109

SPECIFICATIONS
Voltage : Banana plug SHIELD ...mV
OUTPUT CHARACTERISTICS
Charge : 0 to 1000 pC (0 to 100 pC) Supply : 2 batteries 9 V, 0.38 Ah, type
Voltage : 0 to 1000 mV (0 to 100 mV) 6F22 (ref. CEI - IEC)
Output ranges : Selectable with x1 / x10 toggle Autonomy : ‘OPERATION’ mode min.
switch 20 hours at +23°C
(intermittent average: 10 min.)
Fine adjustment : 10-turn vernier with decade
indication PHYSICAL CHARACTERISTICS
ADJUSTABLE LEVELS Dimensions
For range (x1) : 0 to 100 pC peak ±1% • Height : 80 mm (3.15 in)
0 to 100 mV peak ±3% • Width : 145 mm (5.71 in)
For range (x10) : 0 to 1000 pC peak ±1% • Length : 239 mm (9.06 in)
0 to 1000 mV peak ±3%
WEIGHT : 1.1 kg (2.4 lbs)
THERMAL STABILITY
ENVIRONMENT
Charge output :
0.05% per °C
Temperature range
Voltage output :
0.03% per °C
• Operation : 0 to +55°C (32 to 131°F)
Frequency range :
1 to 999 Hz
• Storage : -20 to +70°C (-4 to 158°F)
Accuracy :
±0.1% typical at +23°C
Stability ≤ 0.08% per °C
:
Frequency selection :
Press to adjust frequency
selector / indicator
Charge generator output : Symmetrical / asymmetrical
switchable SYM / ASYM
Symmetry error to shield : ≤ 1%

OUTPUT IMPEDANCE
Symmetrical charge : 500 pF serial with 3 kΩ
Asymmetrical charge : 1000 pF serial with 1.5 kΩ
Voltage : Impedance 10 Ω
limited current 1 mA

OUTPUT CONNECTION
Charge : Cable length 0.7 m with LEMO
connector F1S303N, mating
connector LEMO RC1S303N,
banana plug SHIELD •

APPLICATIONS
- Complete functional check of piezoelectric vibration or - System re-calibration without removal from machine.
dynamic pressure monitoring systems.
- Preventive maintenance work. - Immediate identification of a faulty monitoring component.

ORDERING INFORMATION
To order please specify : Type Designation Ordering Number
TSU 109 Charge Generator 264-109-000-102

In this publication, a dot (.) is used as the decimal separator and thousands are separated by spaces. Example : 12 345.678 90
Although care has been taken to assure the accuracy of the data presented in this publication, we do not assume liability for errors or omissions.
We reserve the right to alter any part of this publication without prior notice.

Head Office Your Local Agent Sales Offices


Vibro-Meter SA Sales offices in :
Rte de Moncor 4, P.O. Box, • Germany • France
• USA • Canada
CH-1701 Fribourg, Switzerland • Singapore • United Kingdom
Phone :+41 26 407 11 11 • Russia • Ukraine
Agents in over 30 countries
Fax : +41 26 407 13 01
Vibro-Meter is a member of the
www.vibro-meter.com Meggitt Aerospace Systems Division

©Vibro-Meter SA / 280-001 / 11.02 / E 2/2

You might also like