Dual Trace Oscilloscope

GOS-630FC

USER MANUAL
GW INSTEK PART NO. 82OS-630FCMA1

This manual contains proprietary information, which is protected by


copyrights. All rights are reserved. No part of this manual may be
photocopied, reproduced or translated to another language without
prior written consent of Good Will company.
The information in this manual was correct at the time of printing.
However, Good Will continues to improve products and reserves the
rights to change specification, equipment, and maintenance
procedures at any time without notice.

ISO-9001 CERTIFIED MANUFACTURER

Good Will Instrument Co., Ltd.


No. 7-1, Jhongsing Rd., Tucheng City, Taipei County 236, Taiwan.
TABLE OF CONTENTS GOS-630FC User Manual

APPENDIX ...................................................................... 36

Table of Contents Line Voltage & Fuse Replacement ........ 36


Specifications ...................................... 38
Declaration of Conformity .................... 40
SAFETY INSTRUCTIONS ................................................... 5

PRODUCT OVERVIEW ..................................................... 10


Product Description ............................. 10
Main Features ...................................... 11
Block Diagram ..................................... 12

PANEL OVERVIEW .......................................................... 13


Front Panel Overview ........................... 13
Display Controls .................................. 14
LCD Display ......................................... 15
Horizontal Controls ............................. 16
Vertical Controls .................................. 17
Trigger Controls ................................... 19
Input Terminals.................................... 22
Rear Panel Overview ............................ 23

SETUP............................................................................. 24
Default Settings ................................... 24
Set up & Probe Compensation ............. 26

MEASUREMENT.............................................................. 28
Single Channel (Basic) Measurement .. 28
Dual Channel Measurement................. 29
Waveform Addition/Subtraction ........... 31
Frequency Measurement ...................... 31
X-Y Mode ............................................. 32
Waveform Magnification ...................... 33

FAQ ................................................................................ 34

3
SAFETY INSTRUCTIONS GOS-630FC User Manual

Safety guidelines

SAFETY INSTRUCTIONS
General • Never connect a hazardous live voltage to the
Guideline ground side of the BNC connectors. It might
lead to fire or electric shock.
CAUTION • Do not place heavy objects on the instrument.

This chapter contains important safety instructions • Avoid severe impacts or rough handlings that
that you must follow when operating the lead to damaging the instruments.
instrument and when keeping it in storage. Read • Do not discharge static electricity onto the
the following instructions before operating the instruments.
instrument to ensure your safety and to keep it in
• Use only mating connectors, not bare wires, for
best condition.
the terminals.
• Do not block the cooling fan opening.
Safety symbols
• Do not perform measurements at power
These safety symbols may appear in this manual or on the generating sources and building installation
instrument. sites (See note below).
• Do not disassemble the instrument unless you
Warning: Identifies conditions or practices that
WARNING are technically qualified.
could result in injury or loss of life.
• To prevent a permanent damage to the CRT
Caution: Identifies conditions or practices that phosphor, avoid excessively brightening the
CAUTION trace or holding a light spot for an unreasonably
could result in damage to the instrument or to
other properties. long time.
• Do not operate the instrument in a place where
DANGER: High Voltage strong magnetic or electric field exists as it may
disturb the measurement.
Attention: Refer to the Manual
• Make sure the input voltage does not exceed the
following values.
Protective Conductor Terminal
Input Terminal Maximum Input Voltage

Earth (Ground) Terminal CH1, CH2 input 300Vpeak


EXT TRIG input 300Vpeak

5 6
SAFETY INSTRUCTIONS GOS-630FC User Manual

Probe input 600Vpeak


Cleaning the • Disconnect the power cord before cleaning the
Z AXIS input 30Vpeak
instrument instrument.
(Measurement categories) EN 61010-1:2001 specifies the
measurement categories and their requirements as follows. The • Use a soft cloth dampened in a solution of mild
GOS-630FC falls under category II. detergent and water. Do not spray any liquid
• Measurement category IV is for measurements performed at the into the instrument.
source of low-voltage installations.
• Do not use chemicals containing harsh products
• Measurement category III is for measurements performed in
building installations. such as benzene, toluene, xylene, and acetone.
• Measurement category II is for measurements performed on the
circuits directly connected to low voltage installations. Operation • Location: Indoor, no direct sunlight, dust free,
• Measurement category I is for measurements performed on Environment almost non‐conductive pollution (See note
circuits not directly connected to mains. below)
Power Supply • AC Input voltage: 115V/230V AC, 50/60Hz • Relative Humidity: < 85%, non‐condensing
• The power supply voltage should not fluctuate • Altitude: < 2000m
WARNING
more than 15%.
• Temperature: 0°C to 40°C
• To avoid electric shock, connect the protective (Pollution Degree) EN 61010-1:2001 specifies the pollution degrees
grounding conductor of the AC power cord to and their requirements as follows. The GOS-630FC falls under
degree 2.
an earth ground.
Pollution refers to “addition of foreign matter, solid, liquid, or
gaseous (ionized gases), that may produce a reduction of dielectric
Fuse • Fuse type: T0.63A/250V (AC 115V), strength or surface resistivity”.
T0.315A/250V (AC230V) • Pollution degree 1: No pollution or only dry, non-conductive
WARNING pollution occurs. The pollution has no influence.
• Make sure the correct type of fuse is installed
before power up. • Pollution degree 2: Normally only non-conductive pollution
occurs. Occasionally, however, a temporary conductivity caused
• To ensure fire protection, replace the fuse only by condensation must be expected.
with the specified type and rating. • Pollution degree 3: Conductive pollution occurs, or dry, non-
conductive pollution occurs which becomes conductive due to
• Disconnect the power cord before fuse condensation which is expected. In such conditions, equipment
replacement. is normally protected against exposure to direct sunlight,
precipitation, and full wind pressure, but neither temperature
• Make sure the cause of fuse blowout is fixed nor humidity is controlled.

before replacing the fuse. Storage • Location: Indoor


environment • Relative Humidity: < 70%
• Temperature: –10°C to 70°C

7 8
SAFETY INSTRUCTIONS GOS-630FC User Manual

Power cord for the United Kingdom

PRODUCT OVERVIEW
When using the instrument in the United Kingdom, make sure the
power cord meets the following safety instructions.
NOTE: This lead/appliance must only be wired by competent persons

WARNING: THIS APPLIANCE MUST BE EARTHED


IMPORTANT: The wires in this lead are coloured in accordance with the
following code:
Green/ Yellow: Earth
Product Description
Blue: Neutral
Brown: Live (Phase)
The GOS‐630FC oscilloscope is a portable, dual‐channel oscilloscope.
As the colours of the wires in mains leads may not correspond with the The GOS‐630FC provides maximum 30MHz DC bandwidth and
colours marking identified in your plug/appliance, proceed as follows: 1mV/DIV sensitivity. Its time base reaches 0.2μS/DIV and the sweep
The wire which is coloured Green & Yellow must be connected to the Earth speed 100nS/DIV. The GOS‐630FC uses a 6‐inch rectangular cathode
terminal marked with the letter E or by the earth symbol or coloured Green ray tube display with red internal graticule. This is a sturdy, easy‐to‐
or Green & Yellow. operate, and highly reliabile product. The GOS‐630FC also
The wire which is coloured Blue must be connected to the terminal which is incorporates a separate LCD Display, which can show CH1/CH2
marked with the letter N or coloured Blue or Black. VOLT/DIV, TIME/DIV, X‐Y MODE, and Frequency.
The wire which is coloured Brown must be connected to the terminal marked
with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment or contact
the supplier.
This cable/appliance should be protected by a suitably rated and approved
HBC mains fuse: refer to the rating information on the equipment and/or user
instructions for details. As a guide, cable of 0.75mm2 should be protected by
a 3A or 5A fuse. Larger conductors would normally require 13A types,
depending on the connection method used.
Any moulded mains connector that requires removal /replacement must be
destroyed by removal of any fuse & fuse carrier and disposed of immediately,
as a plug with bared wires is hazardous if a engaged in live socket. Any re-
wiring must be carried out in accordance with the information detailed on this
label.

9 10
PRODUCT OVERVIEW GOS-630FC User Manual

Main Features Block Diagram


High intensity The CRT incorporates high beam transmission,
and high intensity, and a 2kV high acceleration voltage.
acceleration The CRT displays clearly readable traces even at
higher sweep speeds.
CRT
Wide In addition to the DC‐30MHz (‐3dB) wide
bandwidth and bandwidth, the oscilloscope provides a 1mV/DIV
sensitivity high sensitivity. The 30MHz frequency range is
obtained by an improved triggering
synchronization.
Frequency A built‐in 5‐digit frequency counter offers ±0.02%
Counter accuracy between 1kHz and 30MHz, and ±0.05%
between 50Hz and 1kHz.
Automatic Pressing the AUTO TIMEBASE key automatically
timebase adjusts the sweep time to an appropriate range.
adjustment
Alternate Even when observing two waveforms in different
triggering frequencies, both waveforms can be stably
triggered using the alternate triggering mode.
TV sync The oscilloscope has a sync separator circuit for
triggering: the TV‐V and TV‐H signal triggering.

CH1 output The CH1 signal is output from a rear panel


terminal, which can be applied to external devices.
Z-axis input Display intensity can be controlled via the TTL
compatible z‐axis input on the rear panel.
X-Y mode The X‐Y mode compares the CH1 and CH2 signal
amplitude. It is useful for comparing the phase
difference.
Buzzer alarm Incorrect operation generates a warning beep.

11 12
PANEL OVERVIEW GOS-630FC User Manual

Vertical Controls the vertical scale, vertical position, display


Controls mode, CH2 inversion, and alternate display mode.

PANEL OVERVIEW
Trigger Controls the trigger mode, trigger level, trigger
Controls coupling source, trigger slope, and alternate
triggering mode. Accepts the external trigger input.
Input Accepts the CH1 and CH2 input signals and ground
Terminals wire. Controls the input signal coupling mode.

Front Panel Overview Display Controls

1 2 3 4 5 6
1 CAL Output Generates the probe compensation signal; 2Vp‐p,
1kHz, positive square wave. For probe
compensation details, see page26.
2 INTEN Knob Controls the brightness of a light spot or trace in
the display.
3 FOCUS Controls the focus (sharpness) of the waveforms in
Detailed descriptions of each block start from the next page.
Knob the display.
Main Display Shows the waveforms of input signals. 4 TRACE Controls the alignment of the horizontal trace in
Display Controls power on/off, display configuration, and ROTATION parallel with graticule lines.
Controls the probe compensation output. Point
LCD Display Shows the vertical scale, horizontal scale, X‐Y 5 AUTO Automatically adjusts the horizontal scale to an
display mode, and waveform frequencies. TIMEBASE appropriate range according to the input signal.
Horizontal Controls the horizontal scale, horizontal position, Key
Controls sweep length, and x10 magnification.
6 POWER Turns on or off the oscilloscope’s main power.
Switch When the power is on, the LED lights.

13 14
PANEL OVERVIEW GOS-630FC User Manual

LCD Display Horizontal Controls

1 Horizontal Controls the horizontal position of traces and light


1 CH1 Shows the CH1 vertical scale. POSITION spots.
Vertical Knob
Scale 2 ×10 MAG Magnifies the horizontal scale by a factor of 10. For
2 Horizontal Shows the horizontal scale. Switch horizontal magnification details, see page33.
Scale 3 SWP VAR Adjusts the horizontal scale.
3 X-Y Mode When turned on, indicates that the X‐Y mode is Knob At the minimum position, the horizontal scale
activated. For X‐Y mode details, see page32. becomes 2.5 times wider than the original value
4 Signal Shows the waveform frequency. For frequency selected by the TIME/DIV knob. For example, if the
Frequency measurement details, see page31. original scale is 1mS/DIV, the adjusted scale
becomes 2.5mS/DIV.
5 CH2 Shows the CH2 vertical scale.
Vertical At the maximum (CAL) position, there is no
change in the horizontal scale.
Scale
4 TIME/DIV Controls the horizontal scale from 0.2 μs/div to 0.5
Knob s/div in 20 steps.
When “X‐Y” is selected, the oscilloscope works in
the X‐Y mode. For X‐Y mode details, see page32.

15 16
PANEL OVERVIEW GOS-630FC User Manual

DUAL The CH1 and CH2 signals are displayed


Vertical Controls simultaneously.
ADD The CH1 and CH2 signals are added or
subtracted, and then the result is
displayed. For CH1/CH2
addition/subtraction details, see page31.
5 VAR Knob Adjusts the vertical scale.
At the minimum position, the vertical scale
becomes 2.5 times wider than the original value
selected by the VOLTS/DIV knob. For example, if
the original scale is 1mV/DIV, the adjusted scale
1 2 3 4 5 6
becomes 2.5mV/DIV.
1 VOLTS/DIV Controls the CH1/CH2 vertical scale from At the maximum (CAL) position, there is no
Knob 1mV/DIV to 5V/DIV in 12 steps. change in the vertical scale.
2 Vertical Controls the vertical position of traces and light 6 CH2 INV Inverts the CH2 input signal vertically. When the
POSITION spots for CH1/CH2.
Switch vertical mode switch is in the ADD position, CH2 is
Knob subtracted from CH1 instead of being added to
CH1. For CH1/CH2 addition/subtraction details,
3 ALT/CHOP Selects how the CH1 and CH2 signals appear in the
see page31.
Switch display, in the DUAL trace mode.

CHOP The two signals are chopped and


displayed simultaneously (generally
used in slow sweep, 1ms/DIV or
slower).
ALT The two signals are displayed
alternatively (generally used in fast
sweep, 0.5ms/DIV or faster).
4 Vertical Selects CH1 and CH2 display modes.
MODE
Switch
CH1/CH2 The CH1 or CH2 signal is displayed
independently.

17 18
PANEL OVERVIEW GOS-630FC User Manual

TV-V The oscilloscope triggers when a


Trigger Controls vertical video synchronization signal
appears. For triggering on the field,
select 2ms/DIV as the horizontal scale;
for triggering on the frame (two
interlaced fields), 5ms/DIV.
TV-H The oscilloscope triggers when a
horizontal video synchronization
signal appears. For triggering on the
line, select 10us/DIV as the horizontal
scale. Use the SWP VAR knob to
control the number of waveforms.

1 Trigger ALT When pressed, the TRIG ALT switch constantly Note • For TV‐V and TV‐H trigger, the
oscilloscope responds only to
Switch toggles the trigger source between CH1 and CH2
signal, so that both signals can be clearly viewed. negative polarity signals.

The TRIG ALT switch works when the vertical


mode is in the DUAL position and also the trigger
source is in the CH1 or CH2 position.
• The oscilloscope cannot trigger
Note • The TRIG ALT switch does not work when the
ALT/CHOP switch is in the CHOP position. input signals when their
frequencies are less than 25Hz.
• The frequency counter in the LCD display does
not work when the TRIG ALT switch is pressed. 3 Trigger Changes the trigger level vertically.
LEVEL Knob The trigger level moves up when the trigger LEVEL
2 Trigger The TRIGGER MODE switch selects when the
knob is turned clockwise.
MODE oscilloscope responds to the trigger conditions.
Switch The trigger level moves down when the trigger
LEVEL knob is turned counterclockwise.
AUTO The oscilloscope sweeps regardless of
the existence of trigger conditions.
NORM The oscilloscope sweeps only when a
trigger condition occurs.

19 20
PANEL OVERVIEW GOS-630FC User Manual

4 Trigger Selects the triggering slope.


SLOPE When in the “+” position ( ), the oscilloscope Input Terminals
Switch triggers when the positive slope of the trigger
source signal crosses the trigger level.
When in the “–” position ( ), the oscilloscope
triggers when the negative slope of the trigger
source signal crosses the trigger level.

1 CH 1 (X) Accepts the CH1 input signal. In X‐Y mode, the


Input CH1 input signal becomes the X‐axis.
Terminal
2 AC/GND/ Selects the coupling mode for the input signal.
DC Switch
5 Trigger Selects the signal on which the oscilloscope sweeps. AC The oscilloscope block DC components
SOURCE included in the input signal.
Switch
GND Shows the ground (zero volt) level on
CH1 CH1 signal becomes the trigger the display. This mode is only for
source. checking the reference level; input
CH2 CH2 signal becomes the trigger signal does not appear on the display.
source. DC The oscilloscope displays all of the
LINE AC power line signal becomes the input signal.
trigger source. Useful when the input 3 GND Accepts a ground wire. The GND terminal is
signal synchronizes with the power Terminal connected to the oscilloscope mainframe.
line frequency.
4 CH 2 (Y) Accepts the CH2 input signal. In X‐Y mode, the
EXT The external trigger input (TRIG IN Input CH2 input signal becomes the Y‐axis.
terminal) signal becomes the trigger
Terminal
source.
6 TRIG IN Accepts an external trigger source signal. The • For X‐Y mode details, see page32.
Terminal signal becomes active when the trigger SOURCE
switch is in the EXT position.
Input impedance: 1MΩ // 25pF

21 22
PANEL OVERVIEW GOS-630FC User Manual

Rear Panel Overview

SETUP
Default Settings
Before powering up the oscilloscope, set up the front panel as
follows.

1 2 3 4 POWER Switch Off

1 FUSE & Line Holds the AC mains fuse and selects the AC line
INTEN Knob Center
Voltage voltage, 115V or 230V.
Selector
FOCUS Knob Center
2 AC Power Accepts the AC power cord.
Input Vertical MODE CH1
Connector Switch
3 CH1 Outputs the CH1 signal; approximately 20mV/DIV
ALT/CHOP ALT ( )
OUTPUT when terminated with 50Ω.
Switch
Terminal
CH 2 INV Released (Inversion disabled)
4 Z AXIS Accepts an external intensity modulation (Z‐axis)
signal; 1kHz square wave, DC – 2MHz. Positive Switch
INPUT
Terminal going reduces intensity. For detailed specifications, CH1/CH2 Center of the display
see page38. Vertical
POSITION
• For AC line voltage selection and fuse replacement, see page36.
Knob
CH1/CH2 50mV/DIV
VOLTS/DIV
Knob

23 24
SETUP GOS-630FC User Manual

CH1/CH2 CAL
VARIABLE Set up & Probe Compensation
Knob
Follow these steps and properly set up the oscilloscope, make sure it
CH1/CH2 AC is functional, and compensate the probe.
Coupling
Switch 1 Power on Press the Power switch and turn
Trigger CH1 on the oscilloscope’s power. The
LED lights when the power is on.
SOURCE Switch
The trace line starts to appear after
Trigger SLOPE + (Positive slope) 20 – 30 seconds.
Switch
2 Intensity Use the INTEN knob and FOCUS
TRIG ALT Released (alternating trigger
and focus knob to adjust the trace line
Switch disabled)
adjustment appearance to the best condition.
Trigger MODE AUTO
Switch 3 Probe Connect the probe between the
connection CAL output and the CH1 input
terminal. Select x10 (attenuation)
TIME/DIV 0.5ms/DIV
on the probe.
Knob
4 Probe Select x10 (attenuation) on the probe.
attenuation

x10
Horizontal

x1
CAL
SWP.VAR Knob

Horizontal Center of the display


POSITION 5 Coupling Select the AC position in the CH1
Knob mode AC‐DC‐GND switch. A square
selection waveform appears in the display.
x10 MAG Released (x10 maginification
Switch disabled)

25 26
SETUP GOS-630FC User Manual

6 Vertical Use the CH1 VOLTS/DIV knob

MEASUREMENT
adjustment and CH1 vertical POSITION knob
to adjust the vertical scale and
position of the waveform.

7 Horizontal Use the CH1 TIME/DIV knob and


adjustment CH1 horizontal POSITION knob to
adjust the horizontal scale and
position of the waveform.
Single Channel (Basic) Measurement
Steps 1. Connect the input signal to the
8 Probe Adjust the compensation point on the probe so CH1 or CH2 terminal.

compensa- that the waveform becomes square.


2. Select the vertical coupling
tion
between AC (DC components
are blocked) or DC (all signal
components appear).

3. Configure the trigger settings.


For details, see page19.
Under compensation Right amount Over compensation
4. Use the VOLTS/DIV knob and
vertical POSITION knob to
adjust the vertical scale and
position of the waveform.

5. Press the AUTO TIMEBASE


key to automatically adjust the
9 Completion Now setting up the oscilloscope is completed. For horizontal scale according to
more advanced measurements, see page28. the input signal.

6. If necessary, use the TIME/DIV


knob to adjust the horizontal
scale manually. Use the
horizontal POSITION knob to
adjust the position of the
waveform.

27 28
MEASUREMENT GOS-630FC User Manual

6. Use the vertical ALT/CHOP


Dual Channel Measurement switch to select how the two
waveform appear on the
Steps 1. Connect both CH1 and CH2 display: alternately (ALT mode)
signals to the input terminals. or simultanouesly (CHOP
mode). See page17 for details.
2. Select the vertical coupling 7. If necessary, use the
between AC (DC components VOLTS/DIV knob and vertical
are blocked) or DC (all signal POSITION knob to adjust the
components appear) for both vertical scale and position of
channels. the waveform.
3. Configure the trigger settings. 8. If necessary, use the TIME/DIV
For details, see page19. knob and horizontal POSITION
Make sure the trigger SOURCE knob to adjust the horizontal
switch is in either the CH1 or scale and position of the
CH2 position. waveform.
4. Set the vertical MODE switch to
the DUAL position.
Note The trigger ALT switch does not
work when the ALT/CHOP switch
5. Both the CH1 and CH2 signals appear on the is in the CHOP position.
display.

29 30
MEASUREMENT GOS-630FC User Manual

Waveform Addition/Subtraction X-Y Mode


Steps 1. Make sure both CH1 and CH2 Background The X‐Y mode compares the amplitude of two
waveforms appear on the signals (CH1 and CH2), one as X‐axis (CH1) and
display. For details, see page29. the other as Y‐axis (CH2). The X‐Y mode is useful
for measuring the phase difference of two signals,
2. Set the vertical MODE switch to video color patterns, and frequency response.
the ADD position. The two
waveforms are added and Steps 1. Make sure both CH1 and CH2
appear on the display as a waveforms appear on the
single waveform. display. For details, see page29.

3. To subtract the CH2 signal from 2. Move the TIME/DIV knob to the
the CH1 signal, invert the CH2 X‐Y position.
signal by pressing the CH2 INV
switch.
3. The CH1 and CH2 signals appear in the X‐Y
Frequency Measurement mode and the X‐Y indicator in the LCD display
( X‐Y ) turns on.
Frequency The frequency of the input signal appears on the
shown on the LCD display and is constantly updated.
LCD

Frequency The CH1 signal frequency can be


measurement measured using an external device 4. To adjust the X‐axis position and
using the CH1 such as frequency counter, via the deflection, use the horizontal
rear panel terminal. POSITION knob (position) and
output
CH1 VOLTS/DIV knob
(deflection).

31 32
MEASUREMENT GOS-630FC User Manual

5. To adjust the Y‐axis position and


deflection, use the CH2 vertical

FAQ
POSITION knob (position) and
CH2 VOLTS/DIV knob
(deflection).

Waveform Magnification
Background The oscilloscope can magnify the waveform by ten The probe waveform is distorted.
times in the horizontal direction. The
magnification is useful for observing complex You might need to compensate the probe. For details, see page26.
signals. Note that frequency accuracy and duty factor are not specified for
the probe compensation waveform and therefore it should not be
Steps 1. Make sure the waveform is shown on the used for reference purpose.
display.
The trace line does not appear on the display.
2. Press the ×10 MAG switch. The
waveform becomes magnified Make sure that the trigger mode is in the AUTO mode. In the
by ten times in the horizontal NORMAL mode, the trace does not appear unless a trigger
direction. condition occurs.

x10 magnification The alternate trigger (TRIG ALT switch) does not work.

Make sure that the ALT/CHOP switch is released (ALT position).


The TRIG ALT switch does not work in the CHOP mode.

The frequency counter in the LCD display does not work.

Make sure that the TRIG ALT switch is not pressed. The frequency
counter does not work in the alternate trigger mode.

The TV trigger does not work.

Any part can be covered by using the POSITION control Make sure that the video synchronization signal is positive. The TV‐
V/TV‐H trigger works only when the synchronization signal is
negative.

33 34
FAQ GOS-630FC User Manual

The input signal does not appear on the display.

APPENDIX
Check the following settings.
• The coupling mode is not set at the GND mode, in which the
waveform does not appear on the display. See page22 for details.
• The appropriate trigger source is selected. See page19 for details.

The oscilloscope accuracy does not match the specifications.

Make sure the oscilloscope is powered on for at least 30 minutes,


Line Voltage & Fuse Replacement
within +20°C – +30°C. This is necessary to stabilize the oscilloscope.
1. Take off the power cord.

For more information, contact your local dealer, Good Will


Instruments website www.gwinstek.com.tw, or
[email protected]. 115

230

2. Use a small minus driver and pull out the fuse socket.

115

230

3. Rotate the holder so that the target voltage marking moves to the
upper side.
AC115V AC230V
115
115 230

230
230 115

35 36
APPENDIX GOS-630FC User Manual

4. If necessary, replace the fuse that is attached to the fuse holder.


Specifications
The specifications apply when the GOS‐630FC is powered on for at
115
least 30 minutes under +20°C – +30°C.
Vertical Sensitivity 1mV/DIV to 2mV/DIV: ±5%
accuracy 5mV/DIV to 5V/DIV: ±3%
230

Bandwidth 1mV/DIV to 2mV/DIV: DC to 7MHz


5mV/DIV to 5V/DIV: DC to 30MHz
5. Push the fuse holder back into the socket. AC coupling > 10Hz (reference: 100kHz, 8DIV, –3dB)
Rise time 1mV/DIV to 2mV/DIV: Approx. 50nS
5mV/DIV to 5V/DIV: Approx. 11.7nS
Input impedance Approx. 1M ohm // 25pF.
230 Square wave ≤ 5% Overshoot at 10mV/DIV
characteristics Other ranges: 5% added to the above
115 Linearity ±0.1DIV when moving 2 DIV at center
Vertical mode CH1, CH2, DUAL, ADD
Chop frequency Approx. 250kHz
Input coupling AC, GND, DC
Fuse rating AC 115V: T0.63A/250V, AC 230V: T0.315A/250V Max input voltage CAT II 300Vpeak (AC: ≤ 1kHz)
Max effective Probe1:1 40Vpp (14Vrms Sine wave)
readout Probe10:1 400Vpp (140Vrms Sine wave)
Common mode ≥ 50:1 at 50kHz sine wave
rejection ratio (CH1 and CH2 vertical scales are equal)
Channel isolation >1000:1 at 50kHz
@ 5mV/DIV >30:1 at 30MHz
CH1 signal output ≥ 20 mV/DIV @ 50Ω, 50Hz to 5MHz
CH2 INV BAL. ≤1 DIV (Reference at center graticule)
Trigger Trigger source CH1, CH2, LINE, EXT
Coupling AC: 20Hz to full bandwidth
Slope +/-
Sensitivity 20Hz to 2MHz : 0.5 DIV
(TRIG-ALT:2 DIV, EXT: 200mV)
2MHz to 30MHz: 1.5 DIV
(TRIG-ALT:3 DIV, EXT: 800mV)
TV: Sync pulse > 1 DIV (EXT: 1V)
Triggering modes AUTO, NORM, TV-V, TV-H
EXT TRIG input Input impedance: Approx.:1MΩ // 25pF
Max. input: CAT II 300Vpeak < 1kHz

37 38
APPENDIX GOS-630FC User Manual

Horizontal Axis Sweep time 0.2µS to 0.5S/DIV, ±3%


accuracy
x10MAG accuracy ±5% (20/50nS uncalibrated)
Declaration of Conformity
Linearity ±3%, x10:±5% (20/50nS uncalibrated) We
X-Y Mode Sensitivity Same as vertical axis GOOD WILL INSTRUMENT CO., LTD.
Bandwidth DC to at least 500kHz (1) No.7-1, Jhongsing Rd., Tucheng City, Taipei County, Taiwan
Phase Difference ≤3 degrees at DC to 50kHz (2) No. 69, Lu San Road, Suzhou City (Xin Qu), Jiangsu Sheng, China
Z-Axis Sensitivity 5Vpp (Positive going reduces intensity) declare, that the below mentioned product
Bandwidth DC to 2MHz Type of Product: Oscilloscope
Input resistance Approx. 47kΩ Model Number: GOS-630FC
Max input voltage CAT II 30Vpeak (AC ≤1kHz) are herewith confirmed to comply with the requirements set out in the
Council Directive on the Approximation of the Law of Member States
Probe Waveform Positive-going square wave
relating to Electromagnetic Compatibility (2004/108/EC, 92/31/EEC,
Compensation Frequency Approx. 1 kHz
93/68/EEC) and Low Voltage Directive (2006/95/EC).
Signal Duty Ratio Within 48:52
For the evaluation regarding the Electromagnetic Compatibility and
Output voltage 2 Vp-p ±2%.
Low Voltage Directive, the following standards were applied:
Output impedance Approx. 1 kΩ
◎ EMC
Frequency Display Digits Maximum 5 digit decimal
EN 61326-1: Electrical equipment for measurement, control and laboratory
Counter Frequency Range 50Hz to 30MHz
use –– EMC requirements (2006)
Accuracy 50Hz to 1kHz: ±0.05%
Conducted Emission Electrical Fast Transients
1kHz to 30MHz: ±0.02%
Radiated Emission EN 61000-4-4: 2004
Sensitivity > 2DIV EN 55011: 1998 + A1:1999 +
LCD Display VOLT/DIV, TIME/DIV, X-Y, Frequency A2:2002, ClassB
Backlight Orange Current Harmonics Surge Immunity
CRT Type 6-inch rectangular, internal graticule EN 61000-3-2: 2000 + A2:2005 EN 61000-4-5: 1995 + A1:2001
Phosphor P 31 Voltage Fluctuations Conducted Susceptibility
Acceleration Volt. Approx. 2kV EN 61000-3-3: 1995 + A1: 2001 EN 61000-4-6: 1996 + A1:2001
Screen Size 8 × 10 DIV (1 DIV = 10mm (0.39in)) +A2: 2005
Power Source AC115V/230V±15%, 50/60Hz Electrostatic Discharge Power Frequency Magnetic Field
Power Approx. 40VA, 35W(max.) EN 61000-4-2: 1995 + A1:1998 + EN 61000-4-8: 1993 + A1:2001
Operation Indoor use, Altitude ≤ 2000 m A2:2001
Environment Ambient temperature : Radiated Immunity Voltage Dip/ Interruption
To satisfy specifications:10℃ to 35℃ (50°F to 95°F) EN 61000-4-3: 2002 + A1:2002 EN 61000-4-11: 2004
Maximum operating range: 0℃ to 40℃ (32°F to 104°F) ◎ Safety
Relative humidity: 85% RH(max.) non-condensing Low Voltage Equipment Directive 2006/95/EC
Installation Category II, Pollution degree 2 Safety Requirements
Storage Ambient temperature: -10℃ to 70℃ IEC/EN 61010-1: 2001
Environment Relative humidity: 70%RH (maximum)
Accessories Power cord x 1, User manual x 1, Probe x 2
Dimensions 310 (W) x 150 (H) x 455 (D) mm
Weight Approx.8.2kgs (18.0lbs)