SOMATOM Spirit

CT

Software
Installation of SOMARIS/5.5 VB35

This document is valid for:


SOMATOM Spirit © Siemens AG 2007
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.

English
Print No.: CT02-025.816.05.01.02 Doc. Gen. Date: 06.07
Replaces: n.a
0-2 Revision

Chapter Page Revision


0 to 6 all 01

Document revision level


The document corresponds to the version/revision level effective at the time of system delivery. Revi-
sions to hardcopy documentation are not distributed automatically.
Please contact your local Siemens office to order current revision levels.

Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho-
rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempting installation or service procedures.

SOMATOM Spirit CT02-025.816.05 Page 2 of 4 Siemens AG


Rev. 01 06.07 CS PS CT Medical Solutions
Contents 0-3

Page
1 _______General _______________________________________________________ 1 - 1

Definitions and Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2


Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Software not approved by Siemens . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Software kit VB35 for ICS and IRS . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Backup media . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 -5

2 _______SW Installation ICS _____________________________________________ 2 - 1

Conditions for starting the SW installation. . . . . . . . . . . . . . . . . . . . . . . .2-1


Required data carrier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
Required data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Regional Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Service key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Host name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Write down the host name.
System type and serial numbers . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Backup parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Starting the Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5

Installation running . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7


Configure the keyboard language and enable Windows XP Access . . . . . . . .2-7
Restore Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9
Restore 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9
Restore 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 11
Completion of configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 12
Import Reference Images . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 12
Regional Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 12
Restart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 12
Siemens Remote Services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 12
SRS Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 12
2.8.2 SRS Configuration & SRS Basic Settings . . . . . . . . . . . . . . . . . . 2 - 13
2.8.3 SRS Basic Settings Outside NAT . . . . . . . . . . . . . . . . . . . . . . 2 - 14
2.8.4 Activating System Management . . . . . . . . . . . . . . . . . . . . . . . 2 - 15
Remote Access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 15

3 _______SW Installation of IRS ___________________________________________ 3 - 1

Installation of IRS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1


Final steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Software also reinstalled on the ICS. . . . . . . . . . . . . . . . . . . . . . . . .3-2
No software reinstallation on the ICS . . . . . . . . . . . . . . . . . . . . . . . .3-2

4 _______Additional Worksteps ___________________________________________ 4 - 1

Software self test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1


Final Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1

Siemens AG Drucknr_kurz_V Page 3 of 4 Produkt_V


Medical Solutions Rev. _V MM.JJ Abt._V
0-4 Contents

Final Work Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 2


Configure paper printer for e-logbook . . . . . . . . . . . . . . . . . . . . . . . 4 - 2
Note: E-logbook configuration ICON can be used only when the customer has the elog-
book license. It cannot be seen without the license. . . . . . . . . . . . . . . . . 4 - 2
Backup at the ICS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 2
Disable Windows XP Access . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 3
5.1 Chinese. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1
5.1.1 Settings in Local Service . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1
5.1.2 Settings under meduser . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 2
5.1.3 Settings under administrator . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 4
5.2 Japanese . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 4
5.2.1 Settings in Local Service . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 4
5.2.2 Settings under meduser . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 5
5.2.3 Settings under administrator . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 8

5 ______ Changes to previous version _____________________________________6 - 1

Produkt_V Drucknr_kurz_V Page 4 of 4 Siemens AG


Rev. _V MM.JJ Abt._V Medical Solutions
General 1-1

1 General

Siemens AG Page 1 of 6 CT02-025.816.05 SOMATOM Spirit


Medical Solutions CS PS CT Rev. 01 06.07
1-2 General

1.1 Definitions and Abbreviations


Abbreviation Meaning
ACR-NEMA American College of Radiology - National Electrical
Manufacturers Association
BIOS Basic Input Output System, lowest PC configuration settings
AET Application Entity Title - Name of DICOM service participating in the
communication either as user (SCU) or provider (SCP)
CD Compact Disc
CDR Compact Disc Recordable
CT Computed Tomography
DICOM Digital Image Communication in Medicine (ACR-NEMA standard)
DHCP Dynamic Host Configuration Protocol - procedure for assigning
numeric IP address to nodes in which blocks of addresses are leased
to clients on a first-come, first-serve basis. Typically, DHCP is com-
bined with WINS, meaning an IP address is both dynamically
assigned and automatically associated with a domain name. This can
be combined with DNS, resulting in a dynamically assigned numeric
IP address that is automatically associated with both NetBIOS and
Internet names.
DNS Domain Name Services - address resolution procedure that associ-
ates numeric IP addresses with Internet-style names
FTP File Transfer Protocol - a client-server application that transfers files
from one computer to another
Gateway A display that controls traffic between two networks using factors
above the network layer
ICS = Navigator Image Control System - control computer for the entire CT scanner,
with interface to the acquisition system
IMS Imaging System, consisting of ICS, IRS and console
IRS Image Reconstruction System - reconstruction computer
LAN Local Area Network (here, always Ethernet)
LMHOSTS Address resolution procedure which statically associates numeric IP
addresses with NetBIOS names. The LMHOSTS file is checked to
see if the name is being addressed in this file.
MEP Market Entrance Phase
MOD Magneto-Optical Disk
NetBIOS Network Basic Input Output System - a collection of network services
and programming interfaces. NetBIOS services run on network proto-
cols such as TCP/IP.
NIC Network Interface Card - an expansion card that provides a physical
connection to the network

SOMATOM Spirit CT02-025.816.05 Page 2 of 6 Siemens AG


Rev. 01 06.07 CS PS CT Medical Solutions
General 1-3

Abbreviation Meaning
NT "New Technology" Operating System - trademark of Microsoft, Inc.
OI Operating Instructions
PB Patient Browser
PPP Point-to-Point Protocol - a data link and network protocol designed for
use over serial connections (such as dial-up modems)
RAS Remote Access Services - a Windows NT service that supports dial-
in and dial-out connections using modems and ISDN.
Router A multiport computer or piece of dedicated hardware that examines
network layer addresses of packets on a network and retransmits
them on a different network if necessary
SCU Service Class User
SCP Service Class Provider
TCP/IP Transport Control Protocol/Internet Protocol - the most widely used
standard for networking
UI User Interface
USC Uptime Service Center
WINS Windows Internet Name Service - address resolution procedure that
dynamically associates numeric IP addresses with NetBIOS names.
It is possible to combine DNS and WINS.
XP "eXPerience" Operating System - trademark of Microsoft, Inc.

Siemens AG Page 3 of 6 CT02-025.816.05 SOMATOM Spirit


Medical Solutions CS PS CT Rev. 01 06.07
1-4 General

1.2 Overview
ICS + IRS = Navigator
In the standard configuration, the ICS is the main console for interchangeable scanning
and imaging functions. The IRS is equipped with four computers as well as switchable
keyboard, mouse and monitor.

1.3 Software not approved by Siemens


Warning Unauthorized or incorrect operation or changes to software can
cause malfunctions in your system. This can result in injuries or
system defects. For this reason, the use or installation of soft-
ware not approved by Siemens is strictly prohibited.

1.4 Software kit VB35 for ICS and IRS


The software installation kit for SOMATOM Spirit systems is included in the delivery vol-
ume and comprises the following items required for a complete SW installation:

• CD OEM-Setup English VB35A (1 of 4)


• CD OEM-Setup English VB35A (2 of 4)
• CD MUI_CHN_JPN (CD 2A/4)
• CD SOM/5.5 ICS SW/FW VB35A (CD 3/4)
• CD Som/5.5 ICS SW/FW VB35A (CD 4/4)
• 1x CD Som/5.5 IRS SW/FW VB35A V1.0

SOMATOM Spirit CT02-025.816.05 Page 4 of 6 Siemens AG


Rev. 01 06.07 CS PS CT Medical Solutions
General 1-5

1.5 Backup media


The CD-Writer (drive [-R-] CD-R) is the standard backup device for the ICS. Alternatively,
the MOD may be used, if available.
• Make sure a separate backup media (e.g. "CD Recordable") is on hand for the ICS.

Siemens AG Page 5 of 6 CT02-025.816.05 SOMATOM Spirit


Medical Solutions CS PS CT Rev. 01 06.07
1-6 General

SOMATOM Spirit CT02-025.816.05 Page 6 of 6 Siemens AG


Rev. 01 06.07 CS PS CT Medical Solutions
SW Installation ICS 2-1

2 SW Installation ICS

2.1 Conditions for starting the SW installation

To ensure that all archiving, filming, copying and networking jobs are finished and that the
respective queues are empty:
• Open the Patient Browser.
• Click Transfer -> Select "Local Job Status" and, if necessary, delete all entries.
• Click Transfer -> Select "Network Job Status" and, if necessary, delete all entries.
• Click Patient -> Select "Film Task Status" and, if necessary, delete all entries.

NOTE For safety reasons, the customer needs to archive all images and
raw date prior to installing the software.

NOTE To ensure successful installation of the software, ensure the fol-


lowing:
1. There must be no restart pending due to changes made in the
Configuration. Open LocalService and go to Configuration. Check
whether a "restart pending" is shown in the message bar. If yes,
click HOME to start the missing automatic restart.
2. All applications (e.g. Explorer, Patient Browser) must be closed
and stay closed during the entire update unless requested in the
instructions.
3. Walk through the instructions step-by-step and do not try to
save time by working out your own " installation strategy." WAIT
when it asks you to wait until the ICS is ready.

2.2 Required data carrier


To perform the SW installation of the Navigator (ICS), the following data carriers are
required:

CD OEM-Setup English VB35A (1 of 4)


CD OEM-Setup English VB35A (2 of 4)
CD SOM/5.5 ICS SW/FW VB35A (CD 3/4)
CD Som/5.5 ICS SWFW VB35A (CD 4/4)

Siemens AG Page 1 of 16 CT02-025.816.05 SOMATOM Spirit


Medical Solutions CS PS CT Rev. 01 06.07
2-2 SW Installation ICS

CD MUI_CHN_JPN (CD2A/4)
Backup CD-R
CD System Management CA Unicenter

2.3 Required data


The ICS has to be switched on.
Read the following configuration data and write it down in the fields provided. These data
are required during the SW installation of the ICS and will not be restored.

It may be useful to store the following file to floppy in case of problems with the backup
files or during installation. This file contains the complete configuration of the system.

C:\Somaris\Service\html\Reports\report.htm

SOMATOM Spirit CT02-025.816.05 Page 2 of 16 Siemens AG


Rev. 01 06.07 CS PS CT Medical Solutions
SW Installation ICS 2-3
2.3.1 Regional Settings
• From the syngo user interface, select -> Options -> Configuration.
• In the configuration panel, double-click the regional and language options icon.
• Write down the following data:

Regional Options Language


Customized Time Time separator:
Date Date separator:
Languages Input locales

2.3.2 Service key


• From the syngo user interface, select Options -> Local Service.
• Write down the service key.

Service key (level=7)


Service key = 20 characters ---------------------- 14 characters ---------------- --- last 6 characters ---

2.3.3 Host name


• In the main menu of the service software, select Configuration.
• Click Next.
• Select Local Host -> TCP/IP LAN.

2.3.4 Write down the host name.

Computer name
IP address

NOTE Setting at Local Host- TCP/IP LAN a DNS-Domain e.g "med.sie-


mens.de" will not be in the backup. Please note the configuration.

System type and serial numbers

Name Identifier for Material No. Serial No.


backup files
SOMATOM Spirit P15A 100 45 692 xxx

2.3.5 Backup parameters


• Start the service software and select the Backup & Restore function in the main menu.

Siemens AG Page 3 of 16 CT02-025.816.05 SOMATOM Spirit


Medical Solutions CS PS CT Rev. 01 06.07
2-4 SW Installation ICS
• Insert the backup CD-R for the ICS.
• From Command, select Backup.
• From Drives, select the CD drive, e.g., [-R-] CD-R.
• From Packages, select CT-Settings-XXX and click Go.
- BackupRestore Running ...
- Wait until Ready is displayed.
• Close the CD tray.
• From Packages, select SW-Settings02-XXX and click Go.
- BackupRestore Running ...
- Wait until Ready is displayed.
• Close the CD tray.
• From Packages, select Security Settings XXX and click Go.
- BackupRestore Running ...
- Wait until Ready is displayed.
• When "Ready," remove the backup media.
• Click Home and terminate the service software by closing the browser window.

SOMATOM Spirit CT02-025.816.05 Page 4 of 16 Siemens AG


Rev. 01 06.07 CS PS CT Medical Solutions
SW Installation ICS 2-5

2.4 Starting the Installation


• Insert CD OEM-Setup English VB35A (1 of 4) in the ICS.
• From the ICS, select System /End and shut down the system.
• Switch off the system at the UPS.
• Switch on the system at the UPS.
The computer boots from the CD-ROM.

After a short while, the following window appears:


( e.g., it’s only a sample)

Siemens Medical Solutions syngo CT 2005A installation


SELECTION MENU INFORMATION WINDOW

Turn off the computer to interrupt

or press any key to continue Creating a new primary partition on


first hard drive. This partition will
be converted to NTFS during Windows XP
installation.
CONTINUE INSTALLATION

HARDWARE DETECTION IN PROGRESS

Checking first hard drive

• Press <Enter> to continue the installation (the ICS is beeping). This reboots the ICS.
• Select distribution type: CUSTOMER (default).
• Enter the machine name (host name) as noted by replacing the default entry "MedPC"
and press Enter.

NOTE Please note that the computer name must not contain blanks or
underscores. Use the following naming convention for the ICS
computer name: ct<serial number> (e.g. ct123456).

• Select Keep data on second hard drive or


select Clear data on second hard drive (deletes patient database).
Use ENTER to change the selection.

NOTE Do not use Clear all harddrives 1-2. This option will format all hard
drives only (removes all software). Reinstallation of SW will not be
started.

Siemens AG Page 5 of 16 CT02-025.816.05 SOMATOM Spirit


Medical Solutions CS PS CT Rev. 01 06.07
2-6 SW Installation ICS

NOTE Using Keep Data on other Hard Drive will extend the software
installation time by up to 3 hours! The duration depends on the
number of images stored in the database.

• Use CONTINUE INSTALLATION to confirm the start of installation (ICS is beeping if


"Clear data" is selected).
• Press any key.
All necessary files are copied from CD OEM-Setup English VB35A (1 of 4) to
the hard drive.
• When the computer is beeping, take out OEM-Setup English VB35A (1 of 4) and insert
OEM-Setup English VB35A (2 of 4) , then press any key.
Chinese and Japanese language users: please read the note at the bottom before
the CD is ejected!

After approximately 4 minutes, the CD OEM-Setup English VB35A ( 2 of 4 ) is ejected


and has to be removed.

For Chinese and Japanese language users only:


• Insert the additional CD MUI_CHN_JPN (CD 2A/4). All other users proceed with the next
step.

NOTE CD MUI_CHN_JPN (CD 2A/4) has to be inserted directly after the


OEM CD was ejected.
If you missed to insert this CD on time, the computer asks for CD
SOM/5.5 ICS SW/FW VB35A (CD 3/4).
You have to start the installation at the beginning.
For CD MUI_CHN_JPN (CD 2A/4), no warning is generated as it is for
the other CD’s, if you did not insert it on time.

• Insert the CD SOM/5.5 ICS SW/FW VB35A (CD 3/4).


The installation continues automatically.

NOTE If you missed inserting the requested CD on time, a warning will come
up:
No package has been found. Please insert the CD "........" in the CD
drive to continue the installation.
Insert the requested CD and press any key to continue.

SOMATOM Spirit CT02-025.816.05 Page 6 of 16 Siemens AG


Rev. 01 06.07 CS PS CT Medical Solutions
SW Installation ICS 2-7

2.5
Installation running
From this point on, the ICS installation runs automatically.

After approximately 3 minutes, CD SOM/5.5 ICS SW/FW VB35A (CD 3/4) is ejected and
has to be removed.

Remove CD SOM/5.5 ICS SW/FW VB35A (CD 3/4) and insert


CD Som/5.5 ICS SWFW VB35A (CD 4/4) and close the CD drive. The system
reboots automatically after finishing the copy job.

NOTE While the installation is running, you may start the installation of
the IRS, if necessary (refer to chapter 3).

Completion of the ICS installation (approx. 50 minutes) is indicated by a pop-up window


displaying the message Please enter license.
• Click ’Click here to start the Local Service Configuration’ to start the service software
(complete service license required).

2.5.1 Configure the keyboard language and enable Windows XP Access

Perform the following steps to configure keyboard languages other than English.

• From the syngo user interface, select > Options > Configuration.
• In the configuration panel, click the Regional and Language Options icon.
• Click Regional and Language Options.
• Select the tab card Languages > Details tab card
• Click Add and select the correct Input Language and confirm with OK.
• Select the English keyboard language, click Remove and confirm with OK.
• A message window appears informing you that removal is only possible after Reboot.
Confirm with OK
• Click OK in the remaining windows or close the remaining windows.

Perform the following steps to enable Windows XP access at the time of update.

• Enter the Service Key and open the Local Service main menu.
• From the home menu of the service software , select Utilities.
• Under Source, select Escape to OS and at Command, NT Command Interpreter.
• Under Parameters, type c:\Somaris\Tools\bin\KeyFilter.bat disable.

Siemens AG Page 7 of 16 CT02-025.816.05 SOMATOM Spirit


Medical Solutions CS PS CT Rev. 01 06.07
2-8 SW Installation ICS
• Activate with Return or GO.
Only a reboot makes the change effective. This reboot will be performed after
Restore 1!
• Click Home.

SOMATOM Spirit CT02-025.816.05 Page 8 of 16 Siemens AG


Rev. 01 06.07 CS PS CT Medical Solutions
SW Installation ICS 2-9

2.6 Restore Steps

NOTE If you install the IRS software, wait until the IRS installation fin-
ishes and ensure that the IRS is ready before performing the sub-
sequent steps.
Reason: after Restore 1, the ICS tries to transfer the IRS "default
tables" to the IRS. If the IRS is not ready at that time, the transfer
will fail and the IRS "default tables" will have to be transferred
manually.

2.6.1 Restore 1
• In the main menu of the service software, select the Backup & Restore function.
• Insert the Navigator backup CD.
• From Command, select Restore.
• From Drives, select the corresponding drive.
• From Archive, select <hostname>_SW-Settings02-XXX.ar.
(This is only an example!)

Command Drives Groups


Restore [-R-] CD-R UserSettings
Service\LocalHost\Users
Archive Service\TCPIP\LAN-Settings
<hostname>_SW-Settings02-VB35_date&time.ar Service\TCPIP\RAS-Settings
Service\Database/Satellite Console
Service\Database\DBOffset
Service\Applications\kWorklistResults

• From Groups, select all except for User Settings and Service\Database\DBOffset and
click Go.
BackupRestore Running ...; wait until Ready is displayed.

Siemens AG Page 9 of 16 CT02-025.816.05 SOMATOM Spirit


Medical Solutions CS PS CT Rev. 01 06.07
2 - 10 SW Installation ICS
• From Archive, select <hostname>_CT-Settings-XXX.ar.
(This is only an example!)

Command Drives Groups


Restore [-R-] CD-R
Mode
Archive Tube
<hostname>_CT-Settingsr-VB35_date&time.ar Service
IRS
General
Firmware

• From Groups, select all except IRS and click Go.


BackupRestore Running ...; wait until Ready is displayed.

• From Archive, select <hostname>_SecuritySettings-XXX.ar.


• From Groups, select all and click Go.
• Wait until "Ready".

• Click Home.

NOTE At this point the SW installation of the IRS has to be completed.


• If not already done, remove the CDs from the IRS computer.

SOMATOM Spirit CT02-025.816.05 Page 10 of 16 Siemens AG


Rev. 01 06.07 CS PS CT Medical Solutions
SW Installation ICS 2 - 11
2.6.2 Restore 2
• In the main menu of the service software, select the Backup & Restore function.
• From Command, select Restore.
• From Drives, select the corresponding drive.
• From Archive, select <hostname>_SW-Settings02-XXX.ar.

(This is only an example!)

Command Drives Groups


Restore [-R-] CD-R UserSettings
Service\LocalHost\Users
Archive Service\TCPIP\LAN-Settings
<hostname>_SW-Settings02-VB35_date&time.ar Service\TCPIP\RAS-Settings
Service\Database/Satellite Console
Service\Database\DBOffset
Service\Applications\kWorklistResults

NOTE The following red error message can be ignored during the
restore of SW-Settings
cmd /C C:\Somaris\Config\BaReUp\CapPb\update.bat

• From Groups, select User Settings and click Go.


BackupRestore Running ...; wait until Ready is displayed.

• From Archive, select <hostname>_CT-Settings-XXXS.


(This is only an example!)

Command Drives Groups


Restore [-R-] CD-R Mode
Tube
Archive Service
<hostname>_CT-Scanner-VB35_date&time.ar General
IRS
Firmware

• From Groups, select IRS and click Go.


BackupRestore Running ...; wait until Ready is displayed.
• Take out the backup CD.
• Click Home.

Siemens AG Page 11 of 16 CT02-025.816.05 SOMATOM Spirit


Medical Solutions CS PS CT Rev. 01 06.07
2 - 12 SW Installation ICS

2.7 Completion of configuration


2.7.1 Import Reference Images
This is mandatory if "Clear data on second hard drive" was selected in the installation
mask.
• From the Patient Browser, select Transfer -> Import from Offline.
• From the drop-down list, select the path H:\SiteData\ReferenceImages.
• Select all images and click OK.
After the import, the images are available in the Patient Browser.

2.7.2 Regional Settings


• From the syngo user interface, select > Options > Configuration.
• In the configuration panel, click the Regional and Language Options icon.
• In the Regional Options tab card, select the language that will be used - Chinese (PRC)
or German(German)
• Under Location, select the country where you are.
• Click Apply.
• Click Customize on the Regional Options tab card
• Configure Date and Time as noted

2.7.3 Restart
• Perform System/End -> Restart the System
(A restart is necessary to enable options and make the option tab cards available in the
customer UI (user interface)).

2.8 Siemens Remote Services

2.8.1 SRS Setup


• Go to Options > Service > Remote Service
• Set the Remote Service Access Control Mode to Limited access permanent
• Open the LocalService Home Menu and go to Configuration
• Mark the System Management checkbox under list of system options
• Click Next
- Confirm the message informing you that the configuration has been changed
• Click Next or > until you reach the end of all "black" entires
- At the end, all "black" menu items have changed to yellow and the Finish button ap-
pears.
• Click Finish.

SOMATOM Spirit CT02-025.816.05 Page 12 of 16 Siemens AG


Rev. 01 06.07 CS PS CT Medical Solutions
SW Installation ICS 2 - 13

NOTE Starting with VB35, the configuration item SRS has been added.
SRS provides the page SRS Basic Settings where site-specific
parameters are entered. When saving the SRS Basic Settings, all
SRS functions such as FTP and mail, Somatom LifeNet and CA
Unicenter (only partly) are configured automatically.
Entries in SRS Basic Settings need to be restored manually according to
the previous note.
Note: If the CT uses the default gateway (CTP/IP LAN) to reach the SRS
gateway, the SRS Gateway is deselected.

Navigator

2.8.2 SRS Configuration & SRS Basic Settings 2

Select Service > SRS


1 Outside NAT: normally outside NAT is
deselected. If outside NAT is used,
continue with 3.5.3

2 SRS Target Host (TCP/IP Address):


- Furth for sites in Europe/Africa
- Newark for sites in America
- Singapore for sites in Asia/Australia

3 SRS Gateway (static route):


If a static route was configured in VB15B,
the IP address of the SRS gateway is
displayed under the TCP/IP address
(192.168.217.253 is only an example)
Press <Save> to save the Site Data.
Click to next page via ">" to perform a
SRS connection check.
The connection test is successful if
both the connection to the SRS server
and the connection to the Somatom
Life Extranet are OK.

Siemens AG Page 13 of 16 CT02-025.816.05 SOMATOM Spirit


Medical Solutions CS PS CT Rev. 01 06.07
2 - 14 SW Installation ICS

Navigator

2.8.3 SRS Basic Settings Outside NAT 2

In an outside NAT environment, the customer network administrator only allows certain
IP addresses in his network. If a network node, e.g., the CT system wants to communicate
with the "outside world", for example, the SRS Access Server in Fürth, it cannot use the
real IP address 194.138.39.18. Instead it has to use a NAT address given by the network
administrator. For SRS, the adminstrator provides a NAT address for the SRS Access
Server and for the 3 CA Unicenter Server. The NAT addresses have to be entered in the
SRS Basic Settings
Outside NAT is used very seldom, except in countries like Sweden, where the complete
SRS environment is built up on Outside NAT functionality.
• Select Service > SRS
• Selected 1 Outside NAT.
- SRS Target Host:
select the SRS server where your sys-
tem is connected to (Fürth for Europe/
Africa, Newark for America, etc.)
- for configuring the remaining IP ad-
dresses contact the hospital network
administrator.
In our example, the NAT address is
10.10.10.18 - 22
• Press <Save> to save the Site Data.
• Click to the next page via ">" to perform
a SRS connection check.
The connection test is successful
if both, the connection to the SRS
server and the connection to the
Somatom Life Extranet are OK.

Navigator

SOMATOM Spirit CT02-025.816.05 Page 14 of 16 Siemens AG


Rev. 01 06.07 CS PS CT Medical Solutions
SW Installation ICS 2 - 15

2.8.4 Activating System Management 2

Select System Mgmt > Agent_Control.


Enable the "System
Management Base - Activated" checkbox
and press <Go> to start the CA Unicenter
installation.
If the CA installation is successful, the Sys-
tem Management status is changed to run-
ning after appr. 10 minutes indicating that
the CA Unicenter is successfully installed.
If the CA installation fails, the status is ’not
running.’ The user can click the status to
gain additional information.
If CA Unicenter is successfully installed ,
the SRS configuration is completed.

2.9 Remote Access


• Set up Remote Access as it was prior to the installation.

Siemens AG Page 15 of 16 CT02-025.816.05 SOMATOM Spirit


Medical Solutions CS PS CT Rev. 01 06.07
2 - 16 SW Installation ICS

SOMATOM Spirit CT02-025.816.05 Page 16 of 16 Siemens AG


Rev. 01 06.07 CS PS CT Medical Solutions
SW Installation of IRS 3-1

3 SW Installation of IRS

3.1 Installation of IRS


• Power the IRS off and on.

Insert the IRS SW Installation CD into the CD drive.

- To reboot the IRS, press and hold the ON/OFF switch for about 4 seconds. After the
IRS is switched off, switch it back on again.

NOTE When booting to the IRS CD, the automatic and unattended instal-
lation of Windows XP and of SOMARIS/5.5 image reconstruction
software starts up (approx. 40 min.).

• Remove the IRS-SW CD from the CD drive immediately after it is ejected.


• Press the ON/OFF switch twice again to reboot the IRS. The installation will now run
automatically.
• When the IRS software installation is finished (approx. 40 min.), continue with the final
steps

Siemens AG Page 1 of 2 CT02-025.816.05 SOMATOM Spirit


Medical Solutions CS PS CT Rev. 01 06.07
3-2 SW Installation of IRS

3.2 Final steps


3.2.1 Software also reinstalled on the ICS

Notice After Restore 1 (refer to chapter 2), the ICS and IRS synchronize
and the IRS "default tables" are transferred automatically from the
ICS to the IRS. With Restore 2 (refer to chapter 2), the "site-spe-
cific tune-up tables" are restored to the IRS. No further steps are
necessary.

3.2.2 No software reinstallation on the ICS

Notice During the IRS software installation, the IRS "default tables" are
deleted. The IRS "default tables" have to be transferred manually
to the IRS. As a result, the batch file IrsTablesInstall.bat has to be
executed. Afterward, the "site-specific tune-up tables" have to be
restored.

• From the Service Software Home Menu, select Utilities.


• Under Source, select Escape to OS and at Command, NT Command Interpreter.
• Under Parameters, type c:\Somaris\utils\IrsTablesInstall.bat
• Activate with Return or GO.
The "default tables" are transferred to the IRS.

• In the Service Software Main Menu, select the Backup & Restore function.
• Insert the Navigator backup CD.
• From Command, select Restore.
• From Drives, select the corresponding drive.
• From Archive, select CT-Settings-VB35 XXX.
(This is only an example!)

Command Drives Groups


Restore [-R-] CD-R Bolus
Tube
Archive Service
CT-Settings-VB35 date&time.ar IRS
General
Firmware

• From Groups, select IRS and click Go.


The "site-specific tune-up tables" are transferred to the IRS.

• When "Ready" is displayed, remove the backup CD Navigator and restart the complete
system.

SOMATOM Spirit CT02-025.816.05 Page 2 of 2 Siemens AG


Rev. 01 06.07 CS PS CT Medical Solutions
4 Additional Worksteps 4-1

ICS

4.1 Software self test


• Open the Local Service Home Menu and go to Test Tools
• Click Software SelfTest and execute all sub-tests
- Environment
- Configuration
- Security
- Database
- Misc
In case of errors, refer to the hints for troubleshooting

4.2 Final Test


• Open the Local Service Home Menu and go to Configuration.
• Click Next.
• Go to Local Host > Technical
• Select UPS item.
• Select Country for Quality
- USA for USA
- other for all other countries
• Click Save and OK in the upcoming window
• Go to Local Service> Configuration>AutoTransfer, active "AutoTransfer".
• Click Home
• In the Local Service Home Menu go to Quality Assurance
• Click Constancy
• Select Reference
• Click Go and proceed according to the program instructions.
• Perform scans in the customer mode and expose a film to ensure the camera is
functioning properly
• Check the functions of the A and B keys at the GPC.
• Check the system information and configurations using the notes written on the paper
before. If they are correct, modify that information and configuration, if necessary.
Note : for the Chinese ICS software version, please follow the chapter "special language
settings" to disable the Chinese input language.
• Reconnect the network cable to ICS computer.

Siemens AG Page 1 of 4 CT02-025.816.05 SOMATOM Spirit


Medical Solutions CS PS CT Rev. 01 06.07
4-2 Additional Worksteps

4.3 Final Work Steps


4.3.1 Configure paper printer for e-logbook

Note: E-logbook configuration ICON can be used only when the customer has the
elogbook license. It cannot be seen without the license. 4

• From the syngo user interface, select Options > Service > Loc
• Install the printer drive software in Window XP of ICS referring to the chapter
"Configuration Printer Devices" in Configuration of SOMARIS/5.5 (CT02-
025.843.01.xx.02)
• Select printer for E-logbook :
Syngo>Options>Configuration>e-logbook configuration
Select the installed printer in the "Default printer" list
Click "apply"
• Print e-logbook for testing the printer

4.3.2 Backup at the ICS


• Start the service software and select the Backup & Restore function in the main menu.
• Insert the backup CD-R for the ICS.
• From Command, select Backup.
• From Drives, select the CD drive, e.g., [-R-] CD-R.
• From Packages, select CT-Settings-VB35 and click Go.
BackupRestore Running ...

Command Drives
Backup [-R-] CD-R

Packages
CT-Settings-VB35

Wait until Ready is displayed.


• Close the CD tray.

SOMATOM Spirit CT02-025.816.05 Page 2 of 4 Siemens AG


Rev. 01 06.07 CS PS CT Medical Solutions
Additional Worksteps 4-3
• From Packages, select SW-Settings02--VB35 and click Go.
BackupRestore Running ...

Command Drives
Backup [-R-] CD-R

Packages
SW-Settings02-VB27

Wait until Ready is displayed.


• Close the CD tray.
• From Packages, select Security-Settings-VB35 and click Go.
BackupRestore Running ...

Command Drives
Backup [-R-] CD-R

Packages
Security-Settings-VB27

Wait until Ready is displayed.

• When "Ready" is displayed, remove the backup media.


• Click Home and terminate the service software by closing the Patient Browser window.

NOTE The backup files have been saved correctly when ok is added at
the end of the backup filename.
e.g. <hostname>-SW-Settings02-VB35-date_time_ok.ar

4.3.3 Disable Windows XP Access


Perform the following steps to disable Windows XP access for the customer. Administrator
privileges are required; you must use the command shell of the service software.
• From the Service Software Home Menu, select Utilities.
• Under Source, select Escape to OS and at Command, NT Command Interpreter.
• Under Parameters, type c:\Somaris\Tools\bin\KeyFilter.bat enable.
• Activate with Return or GO.
• Reboot to make the change effective.

To enable Windows XP access, type in


c:\Somaris\Tools\bin\KeyFilter.bat disable

Siemens AG Page 3 of 4 CT02-025.816.05 SOMATOM Spirit


Medical Solutions CS PS CT Rev. 01 06.07
4-4 Additional Worksteps
This page intentionally left blank.

SOMATOM Spirit CT02-025.816.05 Page 4 of 4 Siemens AG


Rev. 01 06.07 CS PS CT Medical Solutions
-1

5.1 Chinese 4

5.1.1 Settings in Local Service 4

NOTICE If the dialog window disappears, do not try to restart the


"Regional Language Options" dialog. Press Alt + TAB to move the
dialog to the foreground.

• In the main menu of the service software, select the Configuration function.
• Go to DICOM > Character Set
• Set allow only one character set
• Select ISO_IR_100
• To save the settings, click Save.

Fig.1
• Click Finish and Home.
➪ Starting Report generation ..., wait until ready
• Do not accept the reboot; click Cancel.

Siemens AG Page 1 of 8 CT02-025.816.05 SOMATOM Spirit


Medical Solutions CS PS CT Rev. 01 06.07 System
-2
5.1.2 Settings under meduser 4

• From the syngo user interface, select > Options > Configuration.
In the configuration panel, click the Regional and Language Options icon.
Note: If the window disappears, click with the left mouse button to bring the window into
the foreground.
• In the Regional Options tab card, select Chinese (PRC)
• Under Location, select the country where you are.
• Click Apply.
• If necessary, press Alt + Tab to bring the dialog window into the foreground.

• To modify the appearance of the date,


time or numbers, click Customize.

Note: change the formats according to


the notes you made before the update.
There is only one restriction: the
maximum number of characters is 2 for
the month in the date format,
e.g, YYYY-MM-DD
- Confirm any change with Apply.
- Click OK

Fig.2

• Next, go to the Languages tab card.

Siemens AG Page 2 of 8 CT02-025.816.05 SOMATOM Spirit


Medical Solutions CS PS CT Rev. 01 06.07
-3

• Select Chinese as the "Language used in


menus and dialogs".
• Click Detail...

Fig.3

• Set the default input language to


English
• Select advanced card and select " turn
off advanced test services" combo box
• "Apply" these changes and Exit the
options panel

Fig.4

Siemens AG Page 3 of 8 CT02-025.816.05 SOMATOM Spirit


Medical Solutions CS PS CT Rev. 01 06.07 System
-4
5.1.3 Settings under administrator 4

• Login as administrator
- Select System > End > Restart Application.
- Press Shift until the login window appears and log in as administrator
• Go to Start> Settings > Control Panel > Regional and language options

• Select the "Advanced" Card


• Under "Language for non unicode
programs" select "Chinese (PRC)"
• Select the box at Default user account
settings (see fig.5)
• Apply and click OK for exit

Fig.5

5.2 Japanese 4

5.2.1 Settings in Local Service 4

NOTICE If the dialog window disappears, do not try to restart the


"Regional Language Options" dialog. Press Alt + TAB to move the
dialog to the foreground.

• In the main menu of the service software, select the Configuration function.
• Go to DICOM > Character Set
• Set allow only one character set
• Select ISO_IR_100
• To save the settings, click Save.

Siemens AG Page 4 of 8 CT02-025.816.05 SOMATOM Spirit


Medical Solutions CS PS CT Rev. 01 06.07
-5

Fig.1
• Click Finish and Home.
➪ Starting Report generation ..., wait until ready
• Do not accept the reboot; click Cancel.

5.2.2 Settings under meduser 4

• From the syngo user interface, select > Options > Configuration.
In the configuration panel, click the Regional and Language Options icon.
Note: If the window disappears, click with the left mouse button to move the window into
the foreground.
• In the Regional Options tabc ard, select Japanese
• Under Location, select the country where you are.
• Click Apply.
• If necessary, press Alt + Tab to move the dialog window into the foreground.

Siemens AG Page 5 of 8 CT02-025.816.05 SOMATOM Spirit


Medical Solutions CS PS CT Rev. 01 06.07 System
-6

• To modify the appearance of the date,


time or numbers, click Customize.

Note: change the formats according to


the notes you made before the update.
There is only one restriction: the
maximum number of characters is 2 for
the month in the date format
e.g YYYY-MM-DD
- Confirm any change with Apply.
- Click OK

Fig.2

• Next, go to the Languages tab card.

Siemens AG Page 6 of 8 CT02-025.816.05 SOMATOM Spirit


Medical Solutions CS PS CT Rev. 01 06.07
-7

• Select Japanese as the "Language used


in menus and dialogs".
• Click Detail...

Fig.3

• Set the default input language to


English
• Select advanced card and select " turn
off advanced test services" combo box
• "Apply" these changes and exit the
options panel

Fig.4

Siemens AG Page 7 of 8 CT02-025.816.05 SOMATOM Spirit


Medical Solutions CS PS CT Rev. 01 06.07 System
-8
5.2.3 Settings under administrator 4

• Login as administrator
- Select System > End > Restart Application.
- Press Shift until the login window appears and log in as administrator
• Go to Start > Settings > Control Panel > Regional and language options

• Select the "Advanced" Card


• Under "Language for non unicode
programs" select "Japanese"
• Select the box at Default user account
settings (see fig.5)
• Apply and click OK for exit

Fig.5

Siemens AG Page 8 of 8 CT02-025.816.05 SOMATOM Spirit


Medical Solutions CS PS CT Rev. 01 06.07
Changes to previous version 5-1

Changes to previous version 5

No changes; initial publication.

Siemens AG Page 1 of 2 CT02-025.816.05 SOMATOM Spirit


Medical Solutions CS PS CT Rev. 01 06.07
5-2 Changes to previous version

SOMATOM Spirit CT02-025.816.05 Page 2 of 2 Siemens AG


Rev. 01 06.07 CS PS CT Medical Solutions