Beam Configuration Reference Guide
Beam Configuration Reference Guide
Eclipse
Publication Title
Beam Configuration Reference Guide
Abstract
This document provides reference information and procedures for using the Beam Configuration application, version 18.1.
This publication is the English-language original.
Notice
Information in this publication is subject to change without notice and does not represent a commitment on the part of
Varian.
Copyright
© 2001-2024 Varian Medical Systems, Inc. All rights reserved. Published in Finland
No part of this publication may be reproduced, translated, or transmitted without the express written permission of Varian
Medical Systems, Inc.
Trademarks
Varian Medical Systems, Inc. or its affiliates own the names of our products and services referenced herein. These names
® ™
are either registered trademarks ( ) or trademarks ( ) in the United States and/or other countries. All other trademarks are
the property of their respective owners. Any rights not expressly granted herein are reserved.
Electronic Labeling
This symbol on the label indicates that the Instructions for Use for the corresponding product are available at
www.MyVarian.com. Access the Instructions for Use in electronic form by logging in with your assigned MyVarian user
credentials.
In compliance with Anvisa and EU regulations, Varian will send Brazil and EU customers a
free printed copy of the Instructions for Use within 7 days. Use the “Paper Publication Re-
quest” form provided on the Varian webpage to order your copy.
WHO
ICD-O codes and terms used by permission of WHO, from:
● International Classification of Diseases for Oncology, Third Edition.
ICD-10 codes and terms used by permission of WHO, from:
● International Statistical Classification of Diseases and Related Health Problems, Tenth Revision (ICD-10).
CAUTION: US Federal law restricts this device to sale by or on the order of a physician.
IEC 62083
™
Eclipse Treatment Planning System is IEC 62083 compliant.
Legal Manufacturer
Legal Information 3
Table of Contents
Introduction ................................................................................................................... 8
Information in This Guide .................................................................................................................. 8
Who Should Read This Guide ........................................................................................................... 8
Visual Cues ....................................................................................................................................... 9
Related Publications .......................................................................................................................... 9
Table of Contents 5
Dose Calculation for Rectangular Fields in Beam Analysis ................................................... 44
Calculations in Beam Analysis ............................................................................................... 44
Define the Dose Calculation Options for Beam Analysis ....................................................... 47
Calculate the Dose for Beam Analysis ................................................................................... 47
Normalize the Depth Dose Comparison for Beam Analysis ................................................... 48
Approving and Unapproving the Configured Beam Data ................................................................ 48
Approve Beam Data ............................................................................................................... 50
Unapprove Beam Data ........................................................................................................... 50
CT Calibration ............................................................................................................. 51
About CT Calibration ....................................................................................................................... 51
About Calibration Protocols .................................................................................................... 51
Converting Image Values to Electron and Mass Densities ............................................................. 53
Modifying the CT Calibration ........................................................................................................... 53
Create a New Calibration Protocol ......................................................................................... 54
Modify a Calibration Protocol ................................................................................................. 55
Remove or Retire a Calibration Protocol ................................................................................ 55
Edit the Calibration Curve ...................................................................................................... 56
Modify the Calibration Curve with the Curve Editor ................................................................ 56
Index .......................................................................................................................... 78
Table of Contents 7
Introduction
WARNING: It is the responsibility of the user to ensure the validity and integrity of
the input data, and to understand that the quality of the output depends
critically on the quality of the input. Any irregularities or uncertainties
about input data, units, identification, or quality of any other nature
shall be thoroughly investigated before the data are used.
CAUTION: Do not use the computer operating system to edit or delete any files
used by the treatment planning application except when the use of the
files is detailed and supported in the user documentation of the
manufacturer of the treatment planning system.
Note: A note describes information that may pertain to only some conditions, readers, or
sites.
Tip: A tip describes useful but optional information such as a shortcut, reminder, or
suggestion, to help get optimal performance from the equipment or software.
Related Publications
● Eclipse Photon and Electron Reference Guide: Provides reference information for using the
Eclipse treatment planning system and instructions for tasks performed outside the daily
treatment planning workflow.
● Eclipse Proton Reference Guide: Provides reference information for using the Eclipse
treatment planning system and instructions for tasks performed outside the daily proton
treatment planning workflow.
● Eclipse Photon and Electron Algorithms Reference Guide: Describes algorithms supported in
the Eclipse treatment planning system.
● Eclipse Proton Algorithms Reference Guide: Describes algorithms supported in the Eclipse
treatment planning system for proton planning.
● RT and Imaging Online Help: Describes the functions available in Radiation Oncology
applications and provides instructions for using them.
● Varian Service Portal User Rights Reference Guide: Describes the administration of users,
user groups and user rights for Varian ARIA CORE applications, except for 4D Integrated
Treatment Console.
● Varian Service Portal Administration Reference Guide: Describes the global configuration
and management of the shared system, user, licensing, database, security, maintenance,
and radiation therapy settings for ARIA CORE modules and applications.
● DICOM Import and Export Reference Guide: Contains information on using the DICOM
Import and Export application.
● Varian Cybersecurity Administration Reference Guide: Provides reference information on the
Varian policy regarding installation of third-party products on or with Varian products,
reference information on data backup processes and guidance for Varian system-specific
backups, general security recommendations on ransomware and on securing customer-
Introduction 9
purchased Varian products, and clarification on the function, purpose, and intent of the
Mission Critical Application Protection (MICAP) configurations.
Beam Configuration
Beam Configuration is designed for the entry of measured dosimetric beam data and automatic
generation of beam data files for external treatment planning. Measured beam data can be
imported to Beam Configuration from a water phantom device in the w2CAD file format or the
data can be entered manually. The measured beam data is converted into the format required for
dose calculation. The measured beam data and the configured beam data can be analyzed using
the beam analysis tools. Potential deviations between the measured and calculated data are
shown by means of a differential graph and a statistical report.
CAUTION: Be careful that the measured beam data that is used for the
configuration corresponds to the characteristics of the treatment unit.
When modifications are made to the treatment unit, make sure that the
measured beam data still corresponds to the treatment unit.
Note: Full functionality Beam Configuration is restricted to user access right groups
Physicist, Service, System Administrator, and All Rights. Make sure that these user
access rights are not assigned to unauthorized persons. More information on user rights:
Varian Service Portal User Rights Reference Guide.
Beam Configuration supports configuring multiple calculation models and energies of different
modalities and provides integration with the database that contains the treatment machine
definitions.
WARNING: The system uses a central database that allows running the application
on different workstations simultaneously. If multiple simultaneous
users work with the same data, this may lead to situations where
another user has modified and saved the data that you are currently
working with. Therefore, avoid situations with multiple users
concurrently editing the same patient data. If you try to save your
changes, the system will issue a warning message about another user
already editing the same patient. Always carefully verify the
synchronized data, because it may contain changes made by both you
and the other user.
Tip: You can start multiple application sessions from the QuickLinks menu of the
Assistant menu bar. To do this, point at the application name in the menu and then
click the right-pointing arrow shown next to the application name.
1
2
3
5
1. Menu bar
2. Toolbar
3. Context Window
4. Object Window
5. Data Windows
1. Menu bar
2. Toolbar
3. Context Window
4. Object Window
5. Data Windows
1. Scope Window
2. Focus Window
3. Selected object
In the figure, the treatment unit forms the top of the hierarchy. Beneath the treatment unit are
several photon and electron energies each with one or several calculation models configured.
Moreover, beam data objects with an invalid checksum value are shown with a yellow
background in their icon.
The contents of the Focus Window are linked to the object selected in the Scope Window. In
this example, a treatment unit configured for a calculation model is selected in the Scope
Window, and its contents are shown in the Focus Window.
The approval state of the beam data is indicated by a colored frame around the data icons
displayed in the Context Window. Configured beam data can be in three possible states: Empty,
Approved and Unapproved. The green boxes shown in the figure indicate that the beam data is
approved.
Right-clicking an object in the Context Window opens a shortcut menu used to modify the
selected object. The contents of the shortcut menu depend on the selected item and the currently
displayed tree structure.
Related Topics
Beam Data and Checksums on page 20
NOTICE: Apart from saving your work regularly, create regular backups of your
system to avoid losing important information. The regular backups
should contain patient database including image directories, dosimetric
data, and beam data.
Note: After making changes to beam data in Beam Configuration, make sure that all
Eclipse applications are restarted to enforce the changes to the beam data that is used in
treatment planning.
1. Parameters
2. Measured Profiles
3. Spectrum
To resolve an invalid checksum and take the affected configured beam data back into use, you
need to unapprove the data, and then verify and revalidate it. In the process of unapproving the
beam data, all data that has dependencies on the affected beam data is deleted, and a message
is shown listing the data to be deleted.
Related Topics
Context Window in Beam Configuration on page 17
NOTICE: Use the same file naming conventions in beam data measurements
(input files) and in Beam Configuration (Therapy Unit name), and
name the input files so that it is easy to match the beam data to the
treatment unit during the configuration process.
Related Topics
Loading Parameter Values from the Database for a Calculation Model on page 36
NOTICE: Use the same file naming conventions in beam data measurements
(input files) and in Beam Configuration (Therapy Unit name), and
name the input files so that it is easy to match the beam data to the
treatment unit during the configuration process.
Related Topics
Assigning Beam Data on page 36
Copying Beam Data on page 38
Related Topics
Configuring Add-on Data on page 33
If you are importing FFF or SRS beam data, make sure that the beam data contains correct
primary fluence mode data for the selected treatment unit.
1. In the Focus Window, select the add-on for which you want to import the measured data.
2. Choose File > Import and then the measured data you wish to import.
2 3 4
1
5
1. Navigate to the folder containing the w2CAD files in this list box
2. Select the w2CAD file or files in this box
3. Displays the measurements in the selected w2CAD files
4. Select the measurements for preview in the data view
5. Select this check box to view the curves during the import
6. Preview the selected measured data in these boxes
3. In the Look in list box, navigate to the folder containing the files to be imported, and select a
file or multiple files.
To see a preview of the measurements contained in the files, select a measurement in the
Measurements in selected files box.
4. To view each individual curve when the data is imported, select the Review data curves
during import check box. You can decide for each curve, if you want to import it or not.
Related Topics
Verifying Configured Beam Data on page 30
1 4
2 5
4. To navigate to the directory containing the Eclipse beam data, click Browse.
Navigate to the level of general data files of the calculation model. This level is above the
folders containing the configured beam data files. For instance, the general data folder may
be named “AAA”, and the configured data folders below it “000”, “001”, “002”, and so on. You
should navigate to the folder named “AAA” in this case.
Related Topics
Verifying Configured Beam Data on page 30
File Hierarchy in Beam Configuration on page 73
Match and Assign Add-On Data on page 34
Adding New Beam Data on page 23
Related Topics
Editing Configured Beam Data on page 31
Note: After making changes to beam data in Beam Configuration, make sure that all
Eclipse applications are restarted to enforce the changes to the beam data that is used in
treatment planning.
Related Topics
Verifying Configured Beam Data on page 30
Related Topics
Verifying Configured Beam Data on page 30
Note: Manual changes made to the measured or configured beam data curves do not
automatically invalidate the previously configured data. If you modify beam data
manually, make sure you take the necessary configuration steps again to re-generate the
configured data from the edited data.
Related Topics
Editing Configured Beam Data on page 31
The lower data window displays the selected curve. The upper data window displays all
curves contained in the selected beam data.
Tip: The normalization type for the displayed curve depends on the selected
algorithm. More information: Eclipse Photon and Electron Algorithms Reference
Guide or Eclipse Proton Algorithms Reference Guide.
3. To show the data points along the curve, click Show Data Points .
4. Review the data curves.
Related Topics
Editing Configured Beam Data on page 31
1 2
The point values along the curve are shown in the left part of the Curve Editor; the data curve
itself in the right part. You can move the points by typing new values.
Edit and View Beam Data Curves with the Curve Editor
1. In the Focus Window under an add-on, select the item containing the curve to edit.
2. In the Object Window, select the curve to edit.
5. Click OK.
Related Topics
Generating Configured Beam Data on page 29
Edit and View Beam Data Curves with the Curve Editor on page 32
Table 1 Data Used for Automatic Add-On Matching for Photon/Electron Treatment Units
Add-On Data
Standard wedge Wedge angles with source-wedge distance
™
Enhanced Dynamic Wedge All data found for the add-on
Electron applicator Applicator sizes
Compensator Compensator materials
Cone set Cone applicators
Elekta motorized wedge All data found for the add-on
Elekta virtual wedge All data found for the add-on
Siemens virtual wedge All data found for the add-on
Table 2 Data Used for Automatic Add-On Matching for Proton Treatment Units
Add-On Data
Option Options
Lateral Spreader Lateral spreaders
Range Shifter Range shifters
Range Modulator Range modulators
Beam Spot Beam spots
Block Block materials
Compensator Compensator materials
MLC MLC materials
Note: To skip an add-on, select None in the list. This may be necessary, if you
later wish to match the add-on manually.
Note: The match is valid for all wedges in the database with the selected angle
and source-wedge distance. Wedges with the same angle but different source-
wedge distance are matched to different add-ons.
Load All Parameter Values from the Database for a Calculation Model
1. Choose Beam Data > Parameter Values > Get All Parameters from Database.
If there are differences between the values saved in Beam Configuration and the default
values stored in the database, a message box opens, listing the parameter values found in
Beam Configuration and values found in the database. If there are no differences, the
message box simply states this.
2. To close the message box and complete the action, click OK.
Related Topics
Copying Beam Data on page 38
Related Topics
Assigning Beam Data on page 36
Related Topics
Match and Assign Add-On Data on page 34
Related Topics
Match and Assign Add-On Data on page 34
Note: The preferred method in making beam data changes is to modify the original
measured beam data files and re-import modified files to Beam Configuration.
1. In the Focus Window, right-click a configured beam data object and then choose Set Data
as Modified.
2. Confirm the modification and click Set to Modified State.
Information Example
Number of measurements $NUMS 1
Direction direction: 0 0 1
In addition, the curve export files contain information about the selected curve, such as the field
size, depth, direction, or the measurement depth, depending on the selected curve.
Information Example
Treatment unit (configured in RT Administration) machine: D_Varian23EX, 6X
In addition, the export files contain information about the particular data, such as the SSD, the
SPD or the calibration date, depending on the type of data exported.
Note: If the calibration curves of the Varian Phantom CT Scanner do not correspond to
unit density at HU = 0, the dose calculation in Beam Analysis will not correspond to a
water-equivalent situation.
where
Dm(x) = Measured dose at the point x
Dc(x) = Calculated dose at the point x
The following quality parameters are calculated to indicate the quality of the beam data:
● Photon and electron beam data
● Distance (in millimeters) to agreement of the 100% dose point (dmax)
● Distance (in millimeters) to agreement of the 50% dose point
● Percent dose difference at 100 mm depth
2 3
1. Graphical overlay
2. Dose differences plot
3. Table of numerical comparison parameters
2 3
1. Graphical overlay
2. Dose difference plot
3. Table of numerical comparison parameters
2
3
3. Select the desired calculation option group on the left side of the Calculation Options dialog
box.
4. Click the desired calculation option on the right side of the dialog box, and do one of the
following:
● Click the down-pointing arrowhead and choose a new value from the list.
● Type a new value in the text box.
5. Click OK.
The Approved status is indicated in the Focus Window by a green frame around the beam data
icons.
1 2
In this example, the beam data on the left is unapproved, and the beam data on the right is
approved. You can approve all beam data of a treatment unit by selecting the treatment unit in
the Focus Window and then approving the data. The Properties dialog shows more details
about the beam data, such as the date and time of approval.
Related Topics
Map File in Beam Configuration on page 75
About CT Calibration
You must create a valid calibration protocol for each imaging device (CT scanner) added to the
database. A calibration protocol has three calibration curves that do the following:
● Convert the image values in an image to electron or mass densities or proton stopping power
during dose calculation.
● Convert image values to electron density during DICOM import, and electron density into
image values during DICOM export.
In Eclipse, you can assign an image value to a structure. During DICOM export or import, the
image value of the structure is converted to relative electron density. The relative electron
density is read from the electron density calibration curve defined for the imaging device. A
calibration curve must be approved before it can be used in dose calculation.
To obtain HU vs. calibration curves for an individual CT scanner, scan a phantom that contains a
sufficient range of materials with known mass density calibration, electron density or proton
stopping power data, or create a curve from theoretical considerations 1, 2, 3.
Note: Set the same image value range for all calibration curves.
1 Chetty I.J., Curran B., Cygler J.E. et al., Report of the AAPM Task Group No. 105: Issues associated with clinical
implementation of Monte Carlo-based photon and electron external beam treatment planning. Med Phys 2007; 34:
4818–4853.
2 Schaffner B. and Pedroni E.: The precision of proton range calculations in proton radiotherapy treatment planning:
experimental verification of the relation between CT-HU and proton stopping power. Phys. Med. Biol. 1998 43,
1579.
3 Constantinou C., Harrington J.C., DeWerd L.A., An electron density calibration phantom for CT-based treatment
planning computers. Med Phys. 1992 Mar–Apr; 19 (2): 325–7.
CT Calibration 51
When you import an image set, DICOM attributes are used to match it to a calibration protocol. If
no perfect match is found, you will be prompted to assign the calibration protocol in the planning
workspace before you start dose calculation. You can view the calibration protocol properties, the
calibration curves, and their approval history in RT Administration and in the Image Properties
in the planning workspace.
WARNING: The CT scanner software version may affect the DICOM metadata of
images imported to Eclipse or ARIA RTM. Ensure that the match rules
defined for the calibration protocol work when the CT scanner software
has been updated. Automatic matching of images with the wrong
calibration protocol may lead to an incorrect dose distribution and
incorrect treatment of the patient.
The different statuses of the calibration protocols and calibration curves are shown in the
following tables. The modifications and status changes of a calibration protocol are shown on the
History tab of the Calibration Protocol Properties dialog box. At least one of the calibration
curves must be approved when you start dose calculation.
Note: The calibration protocol status controls if the calibration protocol can be assigned
to the images (automatically or by the user), whereas the calibration curve status controls
the dose calculation and approval. It is possible to have a plan which has an unapproved
calibration protocol assigned to calculate a dose for that plan. This could happen if you
unapprove a calibration protocol which has already been assigned to the images in the
plan.
The System Default calibration protocol is only used as a template for new calibration protocols
and can only be modified by your system administrator. Phantom images are linked to the Varian
Phantom CT Scanner by default. You need global admin rights to edit the properties and
calibration protocol of the Varian Phantom CT Scanner. The Varian Phantom CT Scanner or its
protocols can not be deleted from the system.
where
ρω,ε = Electron density relative to the electron density of water
NCT = Dependence of the electron density from the electron density of
water on CT number
Related Topics
Edit the Calibration Curve on page 56
NOTICE: Apart from saving your work regularly, create regular backups of your
system to avoid losing important information. The regular backups
should contain patient database including image directories, dosimetric
data, and beam data.
Note: After making changes to beam data in Beam Configuration, make sure that all
Eclipse applications are restarted to enforce the changes to the beam data that is used in
treatment planning.
4 ICRU Report 42, “Use of Computers in External Beam Radiotherapy Procedures with High Energy Photons and
Electrons”, p. 18
CT Calibration 53
Note: When editing a calibration curve, make sure that the values in the Y axis are in
increasing order.
®
Note: Always verify the CT calibration. Varian does not assume responsibility for the
validity of the calibration protocol or the calibration curve.
To verify that the calibration curve is correctly configured, take a CT image of the calibration
phantom, import it in the system and check that the values shown in the Physical Properties tab
of the Point Dose tool are correct.
Information about using the Physical Property tab: Eclipse Photon and Electron Reference
Guide.
4. Click OK.
A copy of the default protocol with three curves is created under the imaging device.
You can also copy an existing calibration protocol from another imaging device and modify it.
Note: You can only edit unapproved calibration protocols. If you wish to edit a
calibration protocol with any other status, change its status to unapproved first. When
you unapprove a calibration protocol, its calibration curves will also become
unapproved.
WARNING: Always test any changes to the calibration protocol match rules
before taking the calibration protocol into clinical use. Automatic
matching of images with the wrong calibration protocol may lead to
an incorrect dose distribution and incorrect treatment of the
patient.
Note: A retired calibration protocol can contain approved calibration curves. Existing
images or plans assigned to the retired protocol can still be recalculated with the
approved curves of the retired protocol.
CT Calibration 55
Edit the Calibration Curve
1. Choose Beam Data or Beam Analysis > CT Calibration.
2. Select a CT scanner and the calibration protocol from the Imaging Devices list to display its
calibration curves.
3. To modify a calibration curve:
a. Select the curve and click Edit Calibration Curve.
b. To view a point on the curve, click a value on the table.
You can also hover on the curve to see the values on it.
c. Modify the points if necessary, and click OK.
Note: You can only edit the calibration curves if the calibration protocol status is
unapproved or commissioning. If you wish to edit the curves of a calibration protocol
with any other status, change its status to unapproved first.
Related Topics
Modify the Calibration Curve with the Curve Editor on page 56
CT Calibration 57
Configuring Distributed Calculation Framework
and Calculation Models
In the above example, Eclipse planning workstations contain a local calculation configuration file
that enables performing calculations on the workstation. In addition to local calculations, these
workstations can also perform remote calculations on other workstations connected to the DCF
network. Eclipse planning workstations also have an agent installed as a system service. Agents
enable Eclipses on other workstations to perform calculations on Eclipse planning workstations.
The agents use calculation configuration files located in the DCF folder on the DCF server.
The installation of an Eclipse system with the DCF includes the DCF server. The location of the
DCF server must be defined on each platform hosting Eclipse.
Note: When defining calculation settings, remember that changing a global calculation
setting affects every workstation or Citrix server connected to the same DCF server.
Local calculation settings override the global calculation settings on the current
workstation or Citrix server.
Note: When defining calculation settings, remember that changing a global calculation
setting affects every workstation or Citrix server connected to the same DCF server.
Local calculation settings override the global calculation settings on the platforms hosting
Eclipse.
Note: When defining calculation settings, remember that changing a global calculation
setting affects every workstation or Citrix server connected to the same DCF server.
Local calculation settings override the global calculation settings on the platforms hosting
Eclipse.
Note: When defining calculation settings, remember that changing a global calculation
setting affects every workstation or Citrix server connected to the same DCF server.
Local calculation settings override the global calculation settings on the platforms hosting
Eclipse.
Debug output
This setting is optional. If the value is set to true, the calculation processes send extra information
of their progress and actions during calculation. The setting should only be used for debug
purposes since it can have an adverse effect on the calculation performance, and the amount of
data sent can be very large. The debug logs are generated in a folder in
C:\VMSOS\Log\Application\RTM\DCF\. Each DCF request generates a new subfolder
named based on the date, time and logged user ([email protected]_UserName)
under the previously mentioned folder and detailed debug files are written to this folder. If the
value is set to false or left empty, no debug output is generated.
Related Topics
Distributed Calculation Framework on page 58
Tip: Removing a calculation model does not remove the configured beam data folder for
the calculation model. If you wish to create the same calculation model, you need to use
a different name for the configured beam data folder.
The following information is shown for each calculation model in the dialog box:
● Model Name—Calculation model name. In cases where several calculation models are
defined for a particular beam energy, each model is identified by its name.
● Algorithm—Name of the algorithm on which the model is based.
● Version—Version of the algorithm on which the model is based.
The version denotes the version of the algorithm that is used by this calculation model.
Changing the version here does not update the configured beam data for this model. If the
beam data is incompatible with the new version of the algorithm, you must re-configure the
beam data.
Automatically created calculation models have the _Auto suffix, and the version information
also displays in parenthesis the name of model from which the beam data was copied. For
example, AAA_18.0.0(AAA_17.0.0)_Auto. Automatically created calculation models are not
enabled by default, and the beam data is not approved. You need to update an automatically
created calculation model to be able to use it.
● Beam Data Directory—Directory on the DCF server where the beam data of the calculation
model is stored.
● Particles—Particle type of the calculation model (photon, electron, proton). This information is
read from the installation file of the algorithm and cannot be modified.
● Calculation Types—Calculation capabilities of the calculation model (for example, volume
dose, compensator, optimization). This information is read from the installation file of the
algorithm and cannot be modified.
Related Topics
Update Calculation Models on page 70
File Hierarchy in Beam Configuration on page 73
● To use the beam data of another calculation model configured in the system, select the
Use same beam data as existing calculation model option and select the desired
calculation model from the list. To limit the list of algorithms shown in the list, select Only
show compatible calculation models.
● To copy the beam data from an old calculation model when upgrading data from an older
Eclipse, or a configured beam data folder of a calculation model not currently configured
in the Eclipse system, select the Copy beam data from old beam data directory option
and then type or navigate to the desired directory. Verify that the correct source folder is
always used.
Note: If the selected old beam data has been configured in an earlier version of
Beam Configuration, you are prompted to update the data.
● In the Beam Data Directory text box, type the relative path of the directory where the
parameters from the selected existing algorithm are saved, navigate to the desired
directory, or use the path suggested.
6. To add the new calculation model, click Add.
If you selected the Copy beam data from existing calculation model option or the Copy
beam data from old beam data directory option, the beam data is updated, and a message
is shown describing which data types were modified in the update. Continue with the next
step.
Otherwise this is the last step and the calculation model is added.
7. Click Close in the message box describing which data types were modified in the update.
8. When prompted, do one of the following: To delete all calculated beam data, click Yes.
To keep the calculated beam data, click No.
9. In the message box prompting you to re-calculate all beam data, click OK.
10. Re-calculate the beam data.
Related Topics
Updating Beam Data Between Versions on page 70
File Hierarchy in Beam Configuration on page 73
Note: If you have plans that are calculated using the previously named calculation
model and you want to recalculate these plans, you must recalculate them using the
new renamed calculation model.
Note: The version denotes the version of the algorithm that is used by this
calculation model. Changing the version here does not update the configured beam
data for this model. If the beam data is incompatible with the new version of the
algorithm, you must re-configure the beam data.
5. Click OK.
6. To complete the version change, update the beam data using the Update Calculation Model
command, re-configure the beam data if necessary, and approve the beam data.
Related Topics
Update Calculation Models on page 70
Related Topics
Configuring Calculation Defaults on page 71
Note: It is not possible to remove server side algorithm based models, such as
MSS Leaf Motion Calculator, Varian Leaf Motion Calculator, DVH Estimation, or
Smart LMC. To disable the use of these models, clear the Enabled check box.
Related Topics
Change the Version of the Algorithm for a Calculation Model on page 69
1. In the Configure Calculation Models dialog box, click Add Model to add a new calculation
model to use as a container for the updated beam data.
2. Type a descriptive name for the model.
3. Select the algorithm that the model will be based on.
4. Select the Copy beam data from existing calculation model option and select the existing
calculation model from the list.
To limit the list of algorithms shown in the list, select Only show compatible calculation
models.
The selected directory must contain the mapfile.txt file, which defines the treatment unit/
energy combination that each beam data subdirectory has been configured for. If the new
Eclipse version contains a treatment unit that has the same ID as the older version, the
correct beam data is automatically assigned to this treatment unit. Otherwise, you can
separately assign beam data for each treatment unit/energy combination.
5. Click Add.
6. When prompted, do one of the following: To delete all calculated beam data, click Yes.
To keep the calculated beam data, click No.
7. In the message box prompting you to re-calculate all beam data, click OK.
8. Re-calculate the beam data.
Related Topics
Assigning Beam Data on page 36
Add Calculation Models to the Configuration on page 67
The table describes the algorithm files that are created, edited and shown in Beam
Configuration and contained in the algorithm folders.
File Description
ConfigurationCalculationSpecification.xml
Related Topics
Update Calculation Models on page 70
Distributed Calculation Framework on page 58
CAUTION: Do not use the computer operating system to edit or delete any files
used by the treatment planning application except when the use of the
files is detailed and supported in the user documentation of the
manufacturer of the treatment planning system.
Note: The contents of the map file are managed by Beam Configuration and should
not be modified using a text editor. If you encounter a need to modify the file, inform
Varian Medical Systems.
The filename of the map file is mapfile.txt. The syntax of the map file is described below. The
text is read from the syntax file of the beam data files.
Data set tags, repeated for each treatment unit and associated add-ons
Other parameters, depend- Each entry contains the file name of the parameters file, approval
ing on the data type specifi- timestamp (date and time, approver, time zone, version of Beam
cation Configuration used for creating the file), approval code and
checksum.
AddOn: Data ID of the add-on. For instance, <00>.
Add-on entry, for instance, File name of the add-on file, approval timestamp (date and time,
<DPR>, or <WDD> approver, time zone), approval code and checksum. For instance,
DPR:<001_00_DPR.xml><ApproveTime="07.06.2013
12:21:45" Approver="physicist"
TimeZone="UTC+03:00 FLE Daylight
Time"><1><1892969726>
It is recommended to use unambiguous date and time formatting. This can be achieved with the
Region and Language settings in Windows. Using a format like dd MMM yyyy for short dates in
the Customize Format dialog box will produce unambiguous date displays for the English locale,
for instance, “05 Sep 2013”. Use either the 24-hour display or 12-hour display together with
am/pm designator from the operating system. By convention, noon is 12:00 p.m. and midnight is
12:00 a.m.
Index 79
see DCF (Distributed Calculation L
Framework)
dose linear attenuation
checksum 20, 21 configuring 33
dose (Beam Configuration) local calculation settings 60
calculating 47 calculation 62
calculation in Beam Analysis 44 general 60
checksum 15, 17, 38, 48 performance 63
comparing with configured beam data 44 trouleshooting 64
dose calculation lockinginfo.txt file 73
see calculation (Beam Configuration) logfile.txt file 73
dosimetric leaf gap 33
M
E
mapfile.txt file 73, 75
electron density mass density and image values 53
calibration 53 matching
conversion formula 53 add-on data 34
converting image values 53 automatic 34
electron density and image values 53 measured beam data
exporting adding 25
beam data 42, 43 adding manually 28
beam data export files 42 importing 26
curve data 42, 43
curve data export files 42
N
navigating in workspaces 16
F
normalization (Beam Configuration)
Focus Window 17 compared depth doses 48
configuring options 67
G
P
general parameters
defining for add-ons 22 profile data
updating between versions 70, 71 adding manually 28
global calculation settings 60 printing 40
calculation 62 progressfile.txt file 73
general 60 publication
performance 63 intended audience 8
troubleshooting 64
S
I
Scope Window 17
information in this guide 8 starting application
intended audience first session 14
publication 8
intended use of product 8
U
unapproving
beam data 21, 48
user
view department of 15
V
verification
configured beam data 30
using ASCII format 42
using W2DAC format 42
visual cues 9
Index 81