Manual de Serviços
Manual de Serviços
───Service Manual
Version/ Page 1 OF
File name S1-K12 Service Manual A/0
Revision number 10
Date of
File number PN/WI08-S1-K12-01 2023.8.8
compilation
1 Purpose
To facilitate effective analysis and repair of equipment faults by after-sales service personnel.
2 Scope
Repair of equipment during the after-sales process.
3 Responsibilities
The after-sales department is responsible for implementing maintenance of the company’s
S1-K12 equipment in accordance with this service manual.
4 Procedures
Number Catalog
10 No power output
15 Power indicator lights up during startup, but display screen remains dark
21 No speaker sound
Digital multimeter, clamp meter, laser power meter, testing tooling, load module, end-face tester,
4.2.2 Operations
a) Cause analysis
1.Improper foot switch connection settings in the APP interface, i.e., set as wireless connection.
2.Internal soldering wire detachment or short circuit in the foot switch metal connector.
b)Repair measures
2.If the customer has another foot switch, try replacing it to confirm if the issue lies with the foot
switch.
Pioon Laser Technology Co.,Ltd.───Service Manual
Version/ Page 3 OF
File name S1-K12 Service Manual A/0
Revision number 10
Date of
File number PN/WI08-S1-K12-01 2023.8.8
compilation
a) Cause analysis
b) Maintenance measures
Pull up or rotate the emergency stop switch. If the alarm persists, check customer operations for
further action.
a) Cause analysis
1.Fiber optic not tightened or individual SMA905 connectors are too short.
b) Maintenance measures
1.Check if the fiber optic is tightly connected. Replace the fiber optic if necessary.
2.Replace the main control board. If the fiber optic alarm continues, it indicates laser abnormality.
In such cases, set the fiber optic alarm to be blocked in the engineer interface to satisfy the
customer’s preference.
3.If the customer does not accept the solution, send the equipment for repairs.
a) Cause analysis
Malfunctioning laser.
b) Maintenance measures
If the customer does not accept the solution, send the equipment for repairs.
a) Cause analysis
b) Maintenance measures
Disconnect and reconnect the remote interlock or replace it with a new one. If the alarm persists,
a) Cause analysis
b) Maintenance measures
1.Connect the adapter, turn on the power switch, and check if the status indicator light is off. If
2.If replacing the boat-type switch does not solve the problem, it suggests a malfunctioning main
control board. Replace the main board and recalibrate the power.
a) Cause analysis
b) Maintenance measures
a) Cause analysis
2.Malfunctioning laser.
b) Maintenance measures
1.In the app interface, adjust the laser alignment to the third level. If there is still no laser
2.If the issue persists after replacing the main control board, it confirms a laser malfunction.
Pioon Laser Technology Co.,Ltd.───Service Manual
Version/ Page 5 OF
File name S1-K12 Service Manual A/0
Revision number 10
Date of
File number PN/WI08-S1-K12-01 2023.8.8
compilation
a) Cause analysis
Malfunctioning laser.
b) Maintenance measures
In the app interface, adjust the laser alignment to the third level. Then, in the engineer interface,
adjust the initial range values of the laser alignment to 60, 80, and 100. If there is no
a) Cause analysis
2.Laser malfunction;
b) Maintenance measures
1.Use a clamp meter to test the current at the laser output of the main control board. If there is no
2.If there is current but no power output, it can be confirmed as a laser malfunction, and it needs
a) Cause analysis
b) Maintenance measures
1.Set the equipment to the preparation stage and observe whether the laser spot is circular and
2.After the bare fiber spot is normal, if the power is still low, use an end-face inspection
instrument to observe the SMA905 end face. If it is contaminated, use a lint-free paper to clean it.
3.If the bare fiber is normal, set the equipment to maximum power, use a clamp meter to test the
working current of the laser output. If the current reaches 7A and the output power is still low, it
can be confirmed as a laser malfunction and needs to be sent back for repair. If the current does
a) Cause analysis
b) Maintenance measures
1.First, replace the qualified lithium battery. If charging is normal, it can be confirmed as a
a) Cause analysis
b) Maintenance measures
First, replace the qualified lithium battery. If the startup is normal, it can be confirmed as a
a) Cause analysis
2.Unstable collection of charging and fully charged states of the lithium battery, resulting in
Pioon Laser Technology Co.,Ltd.───Service Manual
Version/ Page 7 OF
File name S1-K12 Service Manual A/0
Revision number 10
Date of
File number PN/WI08-S1-K12-01 2023.8.8
compilation
b) Maintenance measures
1.First, replace the qualified lithium battery and connect the adapter for charging. Observe the
2.If the charging process and display are normal, it can be confirmed as a lithium battery
malfunction. If the display is still abnormal, software and hardware modification is required. It is
4.2.2.15 Status indicator light displays normally at startup, but the screen is not lit
a) Cause analysis
1.Loose gray ribbon cable connecting the display board and the main control board;
3.Screen malfunction;
b) Maintenance measures
1.Disassemble the device and observe if the gray ribbon cable is loose;
2.If there is no looseness, replace the display board. If the screen can display normally, confirm it
as a display board malfunction, and then re-enter the power calibration data;
3.If the screen still does not light up, it can be confirmed as a screen malfunction. It is
recommended to replace the front screen (screen, display board, upper cover).
a) Cause analysis
Screen malfunction;
b) Maintenance measures
Disassemble the device and replace the display screen. If it is not possible for the end customer
to send it back for repair, it is recommended to replace the front screen (display screen + display
4.2.2.17 The display screen shows normally but cannot enter the system
a) Cause analysis
b) Maintenance measures
a) Cause analysis
b) Maintenance measures
Replace the touch screen and upper cover (the display screen and touch screen are an integrated
unit). If it is not possible for the end customer to send it back for repair, it is recommended to
a) Cause analysis
2.For USB upgrade, the APK file is not placed in the root directory of the USB drive;
b) Maintenance measures
1. If automatic upgrade is selected, make sure the Wifi is connected properly. If the Wifi
connection is stable but the upgrade is still not possible, use a USB drive for the upgrade.
2. USB drive upgrade: Place the APK file in the root directory of the USB drive, then follow the
a) Cause analysis
b) Maintenance measures
2. If the network signal is normal but the device still can't connect to Wifi, check if the antenna is
3. If the antenna is not loose but the device still can't connect to Wifi, it indicates a Wifi module
malfunction. It is recommended to replace the display board. If the end customer is unwilling to
send the device for repair, consider replacing the front screen (display screen + display board +
upper cover).
a) Cause analysis
1. Speaker malfunction.
b) Maintenance measures
1. Open the device, but do not rush to remove the speaker. Instead, disconnect the speaker
2. If the external speaker works properly, replace the faulty speaker. If the speaker still does not
a) Cause analysis
2. Fan malfunction.
b) Maintenance measures
2. If there is no looseness and the fan still does not spin, connect an external fan for verification.
Pioon Laser Technology Co.,Ltd.───Service Manual
Version/ Page 10
File name S1-K12 Service Manual A/0
Revision number OF 10
Date of
File number PN/WI08-S1-K12-01 2023.8.8
compilation
a) Cause analysis
b) Maintenance measures
Replace the button cell battery or connect to Wifi to synchronize the time.