Method 1700 Conductivity Water USP 645 257478-001 A
Method 1700 Conductivity Water USP 645 257478-001 A
Method 1700 Conductivity Water USP 645 257478-001 A
Note: The stirrer probe, Cat. No. 096019, has been added as optional equipment for Stage 2 if the measured
conductivity does not meet the requirements of Stage 1 of the USP <645> Water Conductivity procedures. A ROSS
Ultra pH electrode and pH 4.01, 7.00 and 10.01 buffers have been included in this kit for Stage 3 if the measured
conductivity does not meet the requirements of Stage 2.
1 of 4
Conductivity of
Water USP <645>
Sample Preparation
1. Using a 150 mL beaker, collect about 100 mL of purified water or water for injection and measure it
immediately (within 5 seconds). Alternately, use the flow cell that is included with the conductivity probe.
2. Repeat step 1 for additional samples.
Results
Five aliquots of deionized water were measured and the obtained conductivity results demonstrated good
repeatability. The temperature value was 22 °C for the above results. From the Physical Tests, <645> Water
Conductivity Stage 1 Table for Temperature and Conductivity Requirements, 1.1 μS/cm is the limit at
20 °C. Better results may be obtained using the flow cell included with the 013016MD conductivity probe if
the measured conductivity is less than 1 μS/cm. If the measured conductivity is higher than the table value,
proceed with Stage 2 as described in the USP <645> Water Conductivity procedures.
Deionized Water Conductivity (μS/cm)
Sample # 1: 0.09
Sample # 2: 0.07
Sample # 3: 0.08
Sample # 4: 0.08
Sample # 5: 0.07
Mean: 0.08
Standard Deviation: 0.01
% CV 10.73
Stage 1 Table for Temperature and Conductivity Requirements – For non-temperature compensated
conductivity measurements only
2 of 4
Conductivity Probe Storage
1. For overnight and longer, cell should be stored clean and dry.
2. Storing the probe in water is acceptable between measurements.
Equipment Set Up
Electrode Setup – Conductivity Probe
1. Thoroughly rinse the conductivity probe with deionized water. Gently shake off any excess water.
2. Enter the cell constant value that is printed on the cable of the conductivity probe.
a. I n the measurement mode, press the LINE SELECT key until the arrow icon points to the middle line,
press the UP ARROW key until the μS/cm or mS/cm icon is shown and press the CALIBRATE key.
b. T
he manual calibration screen will display the cell constant on the top line, a conductivity value on the
middle line and “CELL” on the bottom line. To change the cell constant, within 5 seconds press the
DIGIT key until the first digit to be changed is flashing, press the UP/DOWN ARROW key to change
the value of the flashing digit and continue to change the digits until the value matches the cell constant
printed on the cable. Once the value is set, press the DIGIT key until the decimal point is in the
correct location.
c. Press the MEASURE key to save and end the calibration.
Note: In the calibration screen, start changing the cell constant within five seconds or the meter will proceed to
the automatic/direct calibration screen. If this occurs, press and hold the MEASURE key to abort the calibration
and then repeat the calibration.
Table 2 – Expected Conductivity Value of the 100 µS/cm Conductivity Standard at Various Temperatures
Cond Cond Cond Cond Cond Cond Cond
Temp Temp Temp Temp Temp Temp Temp
(µS/ (µS/ (µS/ (µS/ (µS/ (µS/ (µS/
(°C) (°C) (°C) (°C) (°C) (°C) (°C)
cm) cm) cm) cm) cm) cm) cm)
18.0 86.4 19.9 90.1 21.8 93.8 23.7 97.5 25.6 101.3 27.4 104.9 29.2 108.6
18.1 86.6 20.0 90.3 21.9 94.0 23.8 97.7 25.7 101.5 27.5 105.1 29.3 108.8
18.2 86.8 20.1 90.5 22.0 94.2 23.9 97.9 25.8 101.7 27.6 105.3 29.4 109.0
18.3 87.0 20.2 90.7 22.1 94.4 24.0 98.1 25.9 101.9 27.7 105.5 29.5 109.2
18.4 87.2 20.3 90.9 22.2 94.6 24.1 98.3 26.0 102.1 27.8 105.7 29.6 109.4
18.5 87.4 20.4 91.0 22.3 94.7 24.2 98.5 26.1 102.3 27.9 105.9 29.7 109.6
18.6 87.6 20.5 91.2 22.4 94.9 24.3 98.7 26.2 102.5 28.0 106.1 29.8 109.8
18.7 87.8 20.6 91.4 22.5 95.1 24.4 98.9 26.3 102.7 28.1 106.3 29.9 110.0
18.8 88.0 20.7 91.6 22.6 95.3 24.5 99.1 26.4 102.9 28.2 106.5 30.0 110.2
18.9 88.2 20.8 91.8 22.7 95.5 24.6 99.3 26.5 103.1 28.3 106.7 30.1 110.4
19.0 88.3 20.9 92.0 22.8 95.7 24.7 99.5 26.6 103.3 28.4 106.9 30.2 110.6
19.1 88.5 21.0 92.2 22.9 95.9 24.8 99.7 26.7 103.5 28.5 107.1 30.3 110.8
19.2 88.7 21.1 92.4 23.0 96.1 24.9 99.9 26.8 103.7 28.6 107.3 30.4 111.0
19.3 88.9 21.2 92.6 23.1 96.3 25.0 100.1 26.9 103.9 28.7 107.5 30.5 111.2
19.4 89.1 21.3 92.8 23.2 96.5 25.1 100.3 27.0 104.1 28.8 107.8 30.6 111.4
19.5 89.3 21.4 93.0 23.3 96.7 25.2 100.5 27.1 104.3 28.9 108.0 30.7 111.6
19.6 89.5 21.5 93.2 23.4 96.9 25.3 100.7 27.2 104.5 29.0 108.2 30.8 111.9
19.7 89.7 21.6 93.4 23.5 97.1 25.4 100.9 27.3 104.7 29.1 108.4 30.9 112.1
19.8 89.9 21.7 93.6 23.6 97.3 25.5 101.1
©2008 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc.
and its subsidiaries. ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. 257478-001 Rev. A 07-08
Environmental Instruments North America Europe Asia Pacific
Water Analysis Instruments 166 Cummings Center Denmark House, Angel Drove Blk 55, Ayer Rajah Crescent
Beverly, MA 01915 USA Ely, Cambridgeshire #04-16/24, Singapore 139949
Toll Free: 1-800-225-1480 England, CB7 4ET Tel: 65-6778-6876
Tel: 1-978-232-6000 Tel: 44-1353-666111 Fax: 65-6773-0836
Dom. Fax: 1-978-232-6015 Fax: 44-1353-666001
Int’l Fax: 978-232-6031 www.thermo.com/water