Lc-20at Im en
Lc-20at Im en
May 2005
LC-20AT
INSTRUCTION MANUAL
Read the instruction manual thoroughly before you use the product.
Keep this instruction manual for future reference.
.
Introduction
Thank you for purchasing this instrument. This manual describes: the
installation, operation, hardware validation, cautions for use, and details on the
accessories and options. Read the manual thoroughly before using the
instrument. Use the instrument in accordance with the manual’s instructions.
Keep this manual for future reference.
IMPORTANT • Do not use this instrument before fully understanding the contents of this
manual.
• Provide this documentation to the next user in the event that the instrument is
borrowed or sold.
• If this documentation or the warning labels on the instrument become lost or
damaged, promptly obtain replacements from your Shimadzu representative.
• To ensure safe operation, read the Safety Instructions before using the
instrument.
COPYRIGHT • Copyright (C) 2004-2005 Shimadzu Corporation. All rights reserved. This
publication may not be reproduced in whole or in part without written
permission from Shimadzu Corporation. Since Shimadzu products are
frequently upgraded and improved, information in this publication is subject to
change without notice. We appreciate notification of any errors or omissions.
LC-20AT I
Introduction
Warranty 1. Validity
Please consult your Shimadzu representative for information about the extent
of the warranty.
2. Term
The manufacturer will provide free replacement parts for, or repair free of
charge, any instrument that fails during the warranty period, if the cause can
be attributed to a defect in manufacturing.
After-Sales Service If any problem occurs with this instrument, inspect it and take appropriate
corrective action as described in the Section "6 Troubleshooting". If the
problem persists, or symptoms not covered in the Troubleshooting section
occur, contact your Shimadzu representative.
II LC-20AT
Warranty and After-Sales Service
Replacement Parts Replacement parts for this instrument will be available for a period of seven
(7) years after the discontinuation of the product. Thereafter, such parts may
Availability
cease to be available. Note, however, that the availability of parts not
manufactured by Shimadzu shall be determined by the relevant
manufacturers.
Hardware Validation Each LC component and the entire LC system should be checked
periodically to ensure that they function normally, or the analysis data may
not be reliable. To this end, it is necessary to carry out periodic hardware
validation and keep records of the validation. There are two types of
hardware validation - component validation and system validation. The
purpose of component validation is to check that the individual components of
the system function normally, while the system validation checks that the
system as a whole (the several components in combination) functions
normally.
Before shipment from the factory, this instrument was rigorously inspected.
The results are summarized in the Inspection Certificate accompanying the
instrument. To inspect the instrument performance after installation, repeat
the Hardware Validation as described in "7 Hardware Validation".
"7 Hardware Validation" P. 7-1
LC-20AT III
Introduction
Safety Instructions
• To ensure safe operation of the instrument, read these Safety Instructions carefully before use.
• Observe all of the WARNINGS and CAUTIONS described in this section. They are extremely important for
safety.
• In this manual, warnings and cautions are indicated using the following conventions;
NOTE
Emphasizes additional information that is provided to ensure
the proper use of this instrument.
Application Precautions
WARNING
This instrument is a solvent delivery module for use with a high performance liquid
chromatography system.
Use this instrument ONLY for the intended purpose.
Using this instrument for any other purpose could cause accidents.
IV LC-20AT
Safety Instructions
WARNING
• The solvents used in high performance liquid
chromatograph are flammable and toxic. The room
where the instrument is installed should be well
ventilated; otherwise, solvent vapors could cause
poisoning or ignite and cause a fire.
CAUTION
• The weight of this instrument is 11kg.
During installation, consider the entire weight combined
with other LC components.
The lab table on which this instrument is installed should be strong
enough to support the total weight of the LC system. It should be
level, stable and have depth of at least 600mm.
Otherwise, the instrument could tip over or fall off the table.
At least
• Avoid installation sites that are exposed to corrosive 600mm
gases or excessive dust.
These adverse conditions may be detrimental to maintaining the
instrument performance and may shorten its service life.
LC-20AT V
Introduction
Installation Precautions
WARNING
• Take measures to prevent the instrument from falling
in the event of an earthquake or other disaster.
Strong vibrations could cause the instrument to fall over,
resulting in injury.
VI LC-20AT
Safety Instructions
CAUTION
• When installing the instrument, be careful not to pinch
your fingers between the system components, as this
could result in injury.
Operation Precautions
WARNING
• Take thorough measures to prevent buildup of
static electricity.
"Static Electricity Precautions" P.IX
Static electricity could result in fires or explosions.
LC-20AT VII
Introduction
WARNING
• Unplug the instrument before inspection, maintenance, or
parts replacement.
Otherwise, electrical shock or short-circuit accidents could occur.
• If the power cord plug gets dusty, remove the plug from
the power outlet and wipe away the dust with a dry cloth.
If dust is allowed to accumulate, fire could result.
VIII LC-20AT
Static Electricity Precautions
The major cause of these accidents is static electricity. Devising preventative measures for static can be
difficult, because the symptoms before an accident vary and can be hard to detect, since such accidents
occur as a result of several simultaneous coincidences. Recommended methods for preventing static
electricity accidents are provided below. Take thorough safety measures based on this information.
When liquid is passed through thin tubing at high flow rates, as in liquid
Generation of static electricity chromatograph, the electrostatic charges of the flowing matter generate
static electricity.
Generation of Static Electricity by a Liquid Flowing over a Solid
Flowing liquid A: Charges move with
flow of liquid.
B: Immobile charges,
fixed to surface of
solid.
Solid
Insulated container of
spark polyethylene, etc.
LC-20AT IX
Introduction
CAUTION
• It is important to take multiple preventive measures simultaneously.
• If large amounts of flammable solvents are collected in a large container, implement
preventative measures 1, 2, and 3 below.
Preventive Measure 1
Use a metal container for the waste liquid, and ground the container.
This will ensure that the electrical charges of the container and liquid pass to the ground.
CAUTION
• Be sure to ground the metal waste container properly.
If the grounding wire is not properly attached or connected to the ground, static electricity can build up
in the container.
• Some metal containers have surfaces that are laminated or oxidized, and therefore do
not conduct electricity. After grounding the metal container, use a tester to verify that
electricity is conducted to the ground.
X LC-20AT
Static Electricity Precautions
Preventive Measure 2
Cover the spaces between the tubing and the sides of the inlet and outlet openings of the waste container
with caps or other protective covering. This will prevent any sparks generated outside the container from
getting inside.
Preventive Measure 3
Keep electrostatically charged objects, including the human body, away from the waste liquid container.
To prevent electrostatic charging of the human body, take the following precautions:
CAUTION
• Persons who have not taken anti-static precautions should touch some grounded
metal object before coming near the waste liquid container, in order to drain static
charges.
Preventive Measure 4
Use tubing with an inner diameter of at least 2mm for drain lines with high flow rates.
CAUTION
• Periodically check the tubing connections for leaks.
Air bubbles in liquid can multiply the electrostatic charge by a factor of 20, 30 or more.
LC-20AT XI
Introduction
Preventive Measure 5
If it is not possible to use a conductive waste liquid container, take the following precautions:
• Ensure that the end of the inflow tubing is always submerged inside the container. Also, place some
type of grounded metal object, such as a ground wire connected to the instrument, into the liquid.
CAUTION
The above precaution will be ineffective for low conductivity (less than 10-10 S/m) liquids.
• Use as small a container as possible to minimize damage in the event of fire.
• Keep the room at a proper humidity.
Ambient humidity exceeding 65% will prevent static.
For Reference
Anti-static equipment (anti-static clothing, shoes and matting) and charge measurement equipment
(potentiometer) are sold by specialty manufacturers.
XII LC-20AT
Precautions for Mobile Phase Selection and Use
CAUTION
• If PEEK resin parts are used in the plumbing, do not use the following mobile phases.
These mobile phases weaken the PEEK resin, which could result in cracked plumbing
and mobile phase leaks.
Concentrated sulfuric acid, concentrated nitric acid, dichloroacetic acid, acetone,
tetrahydrofuran (THF), dichloromethane, chloroform, dimethyl sulfoxide (DMSO).
Note: Briefly using a weak solution of less than 0.5% acetone in water (e.g. in order to check gradient
performance) will present no problems.
NOTE
• Use only HPLC grade or comparable mobile phase, and filter it with a filter of 0.45µm mesh or finer
before use to remove particulates and foreign matter.
• Halogen ions can corrode the stainless steel material (SUS316L) used in the plumbing, so avoid, as
much as possible, mobile phases that contain halogen ions - such as KCI, NaCI and NH4CI - or mobile
phases that generate halogen ions in certain reactions. If such mobile phases must be used, clean all
flow lines thoroughly with distilled water immediately after analysis.
• When SPD or a similar UV detector is used for high-sensitivity analysis, be sure to use HPLC grade
mobile phases that have a low absorptivity of UV rays.
• Always degas the mobile phase, as air bubbles may tend to form during solvent mixing or during
temperature or pressure changes. Air bubbles may cause pump malfunctions and detector signal
noise.
• For boiling points, viscosities and other data relating to the mobile phases used,
"9.5 Mobile Phase Characteristics" P. 9-47
LC-20AT XIII
Introduction
Warning Labels
For safety operation, warning labels are affixed to where special attention is required.
Should any of these labels peel off or be damaged, obtain replacements from Shimadzu Corporation.
XIV LC-20AT
Contents
Introduction ......................................................................................................................I
Warranty and After-Sales Service ..................................................................................II
Safety Instructions ........................................................................................................ IV
Application Precautions .............................................................................................. IV
Installation Site Precautions ......................................................................................... V
Installation Precautions ............................................................................................... VI
Operation Precautions ............................................................................................... VII
Precautions for Instrument Inspection, Maintenance, Adjustment and Care ............ VIII
Static Electricity Precautions ........................................................................................ IX
Typical Cause of Static Electricity Accidents .............................................................. IX
Preventing Static Electricity Accidents ......................................................................... X
Precautions for Mobile Phase Selection and Use ...................................................... XIII
Warning Labels ........................................................................................................... XIV
Chapter 1 Configuration
1.1 Overview ..........................................................................................................1-2
1.2 Features ...........................................................................................................1-2
LC-20AT Contents-1
Contents
Chapter 3 Preparation
3.1 Turning Power ON/OFF ...................................................................................3-2
4.5 Plunger Rinsing When Buffer Solution is Used as Mobile Phase ..................4-14
4.5.1 Connecting the Plumbing .............................................................................. 4-14
4.5.2 Rinsing .......................................................................................................... 4-15
Contents-2 LC-20AT
Contents
LC-20AT Contents-3
Contents
Contents-4 LC-20AT
Contents
Setting Pressure Sensor Sensitivity Compensation Factor [PRESS COMP FACT] 5-39
Setting Operation Level of Leak Sensor [LEAK THR] ............................................. 5-40
Setting Timing to replace Plunger Seal [SEAL REPLACEMENT] .......................... 5-40
Selecting Operation Mode [OP MODE] .................................................................. 5-40
Initializing Parameters [INITIALIZE PARAM] .......................................................... 5-40
Changing Passwords [CHANGE PASSWORD] ...................................................... 5-41
Showing/Setting CBM Parameter [CBM PARAMETER] ......................................... 5-41
LC-20AT Contents-5
Contents
Chapter 6 Troubleshooting
6.1 Troubleshooting and Corrective Action ............................................................6-2
Contents-6 LC-20AT
Contents
LC-20AT Contents-7
Contents
Chapter 8 Maintenance
8.1 Periodic Inspection and Maintenance ..............................................................8-2
8.1.1 Prior to Inspection and Maintenance ............................................................... 8-2
8.1.2 List of Periodic Inspection and Maintenance ................................................... 8-3
8.1.3 Check after Inspection and Maintenance ........................................................ 8-4
8.6 Inspection (Replacement) and Ultrasonic Bath Cleaning of Check Valves ....8-19
8.7 Inspection (Replacement) and Ultrasonic Bath Cleaning of Line Filter ..........8-22
Replacing Frit ..........................................................................................................8-23
8.8 Inspection (Replacement) and Ultrasonic Bath Cleaning of Suction Filter .....8-24
Contents-8 LC-20AT
Contents
LC-20AT Contents-9
Contents
Index
Contents-10 LC-20AT
1 Configuration
1
Contents
1.1 Overview
LC-20AT (hereinafter referred to as the instrument) is a dual-plunger tandem-flow solvent delivery module. It
offers improved accuracy and sensitivity in high-performance liquid chromatograph. In addition to the
instrument, a high performance liquid chromatograph system requires an autosampler, column oven, detector
and system controller at a minimum, which should be purchased separately. Consult your Shimadzu
representative regarding the components required for your system.
1.2 Features
• Stable Delivery with Low Flow Pulsation
The unique method used to drive the dual reciprocating plungers provides stable delivery with low flow
pulsation.
1-2 LC-20AT
1.3 Component Parts
LC-20AT − 1
CBM grounding plate (with 3 screws) 228-42299 1 Used when CBM-20Alite is built-in.
LC-20AT 1-3
1. Configuration
Accessory Pack
Parts Part No. Q’ty Remark
Wrench, 8 × 10 086-03006 2
Wrench, 13 × 17 086-03017 1
Lid 228-17644 1
1-4 LC-20AT
1.4 Optional Parts
3-way solenoid
valve
LC-20AT 1-5
1. Configuration
For continuous automatic rinsing of the back side of the plunger seal,
Automatic rinsing pump using a rinsing pump with rinse solution. Used when system uses high-
228-45568-91
20AT salt-concentration buffer solution.
Installed inside of the instrument.
Injector mounting plate 228-35659-92 Plate for installing manual injectors on the instrument.
Column holder 228-35655-94 For use if the column is installed on the instrument.
CBM-20Alite 228-45011-XX CBM-20Alite system controller can be installed inside of the instrument.
1-6 LC-20AT
2 Parts Identification and Function
2
Contents
Keypad
To operate and configure settings.
Press to show the operation keys.
Front cover
To cover pump heads and flow lines.
Power switch
To switch the power ON/OFF.
Press the switch in to turn power ON.
Press again to turn power OFF.
2-2 LC-20AT
2.2 Behind Front Cover
Pump head
To enclose a reciprocating
2
plunger which delivers solvent.
Rinse tubing U
Pump inlet
The filtered solvent line is
connected to this inlet.
LC-20AT 2-3
2. Parts Identification and Function
Shipping screws
To prevent the shock during
Leakage drain outlet transportation.
Connect the provided drain tubing. Remove before installation.
2-4 LC-20AT
2.4 Back
2.4 Back
[REMOTE] connector
Connect to the system controller or an
2
identical pump unit.
Fuse holders
To hold the fuses.
[DGU/SOL.V] connector
Connect to the flow line switching valve (FCV
series) or DGU-20A degasser.
LC-20AT 2-5
2. Parts Identification and Function
Display panel
This instrument is controlled through the keypad.
The display allows verification of the instruments status.
NOTE
Keypad
1 2 3 4
2.5.1 Display Panel
The display panel consists of a display screen and LED
indicators. 0.000mL MAX 10.0
Names and functions of the display screen and the 0.0MPa MIN 0.0
indications are given below.
5 6 7 8 9
No. Display or Indicator Function
Displays set flow rate (in mL/min) in constant flow pumping mode, and set pressure
1 flow/press (in units set with [PRS-UNIT] additional function) in constant pressure pumping
mode.
Displays reading of pressure sensor (in units set with [PRS-UNIT] additional
2 pressure
function).
3 p.max Displays pressure upper limit (in units set with [PRS-UNIT] additional function).
4 p.min Displays pressure lower limit (in units set with [PRS-UNIT] additional function).
pump
5 Illuminates when pump is running.
(Pump indicator)
prog.run
6 Illuminates when program is being executed.
(Program indicator)
remote
7 Illuminates when the instrument is controlled by system controller.
(Remote mode indicator)
G.E.
8 Illuminates when the instrument is running in gradient mode.
(Gradient mode indicator)
Green : when power is ON.
9 Status Indicator Red : when an error is generated.
Orange : during sleep mode.
2-6 LC-20AT
2.5 Names and Functions of Displays and Keypad
2.5.2 Keypad
The 24 keys on the keypad are used to operate the
instrument and set parameters.
Sleep key To turn off the display screen. This key has no effect on operation.
Delete key To delete individual lines of time program on the display screen.
Edit key To activate edit mode of time program (from the initial screen).
LC-20AT 2-7
2. Parts Identification and Function
2-8 LC-20AT
3 Preparation
3
Contents
Fig. 3.1
Initial Screen
Flow rate [flow]
NOTE
• The values displayed on the initial screen are the Pressure display [pressure]
values last set.
• If an error is detected, an alarm sounds and an error
message appears.
"6.2 Error Message" P. 6-5
3-2 LC-20AT
3.2 Preparation before Operation
3
1 Pour approximately 100mL of 2-propanol into a
beaker.
Suction filter
Fig. 3.2
7 Press . Open
The pumps run, and the pump indicator
illuminates.
NOTE
LC-20AT 3-3
3. Preparation
Pump indicator
10 Press .
The pumps run, and the pump indicator
illuminates.
3-4 LC-20AT
3.3 Checking Pressure Value
1 Pour mobile phase in the reservoir, and put the suction filter inside the reservoir.
3
The end of the drain tubing should be in the waste container, which is placed on the floor.
Reservoir
To Autosampler/
Manual injector
Waste container
Drain tubing
Suction filter
Fig. 3.5
NOTE
Fig. 3.6
LC-20AT 3-5
3. Preparation
6 Press .
Pumping begins at a registered flow rate.
"Setting Purge Flow Rate [P-FLOW]" P. 5-17 PURGING LINE
NOTE
3-6 LC-20AT
3.3 Checking Pressure Value
8 Press .
The pump stops, and the pump indicator goes out.
NOTE
LC-20AT 3-7
3. Preparation
3-8 LC-20AT
4 Basic Operation
4
Contents
1.0 ∼ 44.0MPa
(0 ∼ 5.000mL/min)
p.max 0.1 MPa 10 MPa
1.0 ∼ 22.0MPa
Constant Flow
(5.001 ∼ 10.000mL/min)
Delivery
0.0 ∼ 44.0MPa
(0 ∼ 5.000mL/min)
p.min 0.1 MPa 0 MPa
0.0 ∼ 22.0MPa
(5.001 ∼ 10.000mL/min)
4-2 LC-20AT
4.1 Setting Parameters
2 Press .
The cursor blinks in the display field, prompting 4
input of a new value. Flow rate [flow] (with cursor prompting input)
following sequence:
[flow (press)]→ [P.max] → [P.min]. 0.000mL MAX ™0.0
When an item is accessed, the cursor blinks in the 0.0MPa MIN 0.0
display field, prompting value input. Press
further to access the other auxiliary
functions.
[P.min]
LC-20AT 4-3
4. Basic Operation
Setting Procedure
The procedure for setting the maximum pressure limit is as follows.
For example, 15MPa will be set as the maximum pressure limit.
2 Press , and .
The new value appears (as shown on the right) on
the initial screen.
0.000mL MAX 15.0
0.0MPa MIN 0.0
4-4 LC-20AT
4.1 Setting Parameters
4
Setting Procedure
The procedure for setting the minimum pressure limit is as follows.
For example, 2MPa will be set as the minimum pressure limit.
[P.min]
2 Press , , and .
The new value appears (as shown on the right) on
the initial screen. 0.000mL MAX 15.0
0.0MPa MIN 2.0
NOTE
LC-20AT 4-5
4. Basic Operation
4.2 Operation
There are two basic operating modes; one is in constant flow delivery mode and the other is in constant
pressure delivery mode.
3 Press
rate.
once, then, set the desired flow [flow]
4 Press .
The pumping starts, and the pump indicator
illuminates. 1.000mL MAX 10.0
0.0MPa MIN 0.0
5 Observe the pressure value on the screen to
ensure that the pump outlet pressure is rising.
CAUTION
Close the drain valve if necessary. When it is left opened, mobile phase could flow out of the drain due to
atmospheric pressure.
4-6 LC-20AT
4.2 Operation
4
3 Press
Press
.
repeatedly until [MODE CHANG]
appears. 0.000 MODE CHANG
Enter to Change
4 Press .
The mode changes from constant flow delivery
mode to constant pressure delivery mode. 1.0 MODE CHANG
"Plunger Set [P-SET]" P. 5-20 Enter to Change
6 Press .
The cursor blinks in the [press] display field,
[press]
LC-20AT 4-7
4. Basic Operation
8 Press .
The pump indicator illuminates and pumping
starts. 2.0MPa MAX 10.0
2.0MPa MIN 0.0
4-8 LC-20AT
4.3 High-Pressure Gradient Mode
4
1 Press to set the [SYS] parameter of the
master pump unit (hereinafter referred to as pump
Master pump unit (Pump A)
A).
"Setting System Control Parameter [SYS]" P.
0.000 SYS 2
5-22 Set System Param
B).
0.000 SYS 1
Set System Param
LC-20AT 4-9
4. Basic Operation
2 Set the value of flow rate [flow], enter the flow rate
(the total flow rate value of pump A and pump B).
™.000mL MAX 10.0
0.0MPa MIN 0.0
∗ The mobile phase delivered by pump A is
referred to as mobile phase A, and that of pump
B as mobile phase B.
3 Press .
The screen of setting concentration appears.
A:100.0 B:™ 0.0
5 Press on pump A.
Pumping in pump A and pump B starts at the
same time.
Operation of Pump B
• The flow rate of pump B is controlled by pump A; all the other parameter settings of pump B remain effective
for pump B.
• remains active and can be used to stop and start pump B at any time, during execution of a time
program set on the side of pump A.
• is available to execute purging of pump B.
NOTE
4-10 LC-20AT
4.4 Low-Pressure Gradient Mode
and D.
In the example on the right, 5 mL/min is set for the
flow rate.
6 Press .
The screen of setting concentration appears.
Enter the concentrations. A:100.0 B:™ 0.0
C: 0.0 D: 0.0
LC-20AT 4-11
4. Basic Operation
Mobile phase A (%) = 100 (%) - Mobile phase B (%) - Mobile phase C (%) - Mobile phase D (%)
2 Press .
Pumping starts, and the pump indicator
illuminates.
Fig. 4.1
NOTE
There will be some air bubbles in the flow at first. Check air
Wait for a short time until air bubbles disappear. bubbles
disappear.
4-12 LC-20AT
4.4 Low-Pressure Gradient Mode
5 Press .
Pumping stops, and the pump indicator goes out.
<1-Cycle Mode>
The opening and closing times of the solenoid valves are controlled for each individual cycle (intake and
discharge) of pump.
This permits a small unit liquid volume of gradient concentration, giving excellent mixing performance and
allowing use of a small-capacity mixer that provides highly consistent concentrations.
<2-Cycle Mode>
The opening and closing times of the solenoid valves are controlled for every two cycles (intake and
discharge) of pump.
This permits a large unit liquid volume of gradient concentration, providing high concentration precision.
<4-Cycle Mode>
The opening and closing times of the solenoid valves are controlled for every four cycles (intake and
discharge) of pump.
This permits a large unit liquid volume of gradient concentration, providing high concentration precision.
However, the capacity of the mixer must be large in this method, or the liquids will not be mixed properly.
Consequently this method is unsuitable for analysis by low flow rates.
NOTE
In the initial setting, it is set to [LPGE MODE=0] on which 2-Cycle and 4-Cycle Mode are switched each other
automatically based on the flow rate.
LC-20AT 4-13
4. Basic Operation
CAUTION
The seals and plungers must be rinsed frequently when a buffer solution is used as the mobile phase.
Buffer solutions crystallize upon evaporation, and crystals could damage the plungers and plunger seals,
which may cause the shortening of their service life.
2 Put one end of the rinse tubing into a beaker. Fig. 4.3
Rinse tubing
Rinse tubing
Beaker
Disposable syringe
Fig. 4.4
4-14 LC-20AT
4.5 Plunger Rinsing When Buffer Solution is Used as Mobile Phase
4.5.2 Rinsing
LC-20AT 4-15
4. Basic Operation
4.6.1 When New Mobile Phase is Miscible with the Old Mobile Phase
Fig. 4.5
Fig. 4.6
4-16 LC-20AT
4.6 Replacing the Mobile Phase
6 Press .
The old mobile phase will be completely expelled
from the flow lines through the drain tubing.
Drain valve
Drain tubing
4
Fig. 4.7
Waste container
Fig. 4.8
9 Set the flow rate [flow] to about 1-3mL/min. Flow rate [flow]
11 Press .
Pumping stops, and the pump indicator goes out.
∗ The flow rate should be set so that the pressure Fig. 4.9
will be lower than the maximum column
pressure.
LC-20AT 4-17
4. Basic Operation
4.6.2 When New Mobile Phase is not Miscible with the Old Mobile Phase
1 Prepare an intermediate rinse solution (2-propanol or similar) that is miscible with both the old and new
mobile phases.
2 Replace the old mobile phase with the intermediate rinse solution.
"4.6.1 When New Mobile Phase is Miscible with the Old Mobile Phase" P. 4-16, steps 1-11
3 Replace the intermediate rinse solution with the new mobile phase.
"4.6.1 When New Mobile Phase is Miscible with the Old Mobile Phase" P. 4-16, steps 1-11
CAUTION
Distilled or de-ionized water should be used for purging.
Use of organic solvents such as 2-propanol may cause crystals.
2 Pump at least 200mL of the water prepared in step 1 through the flow line.
"4.6.1 When New Mobile Phase is Miscible with the Old Mobile Phase" P. 4-16, steps 1-11
CAUTION
When the instrument is to remain inactive following operation with a buffer solution, be sure to replace the
buffer solution in the flow lines with distilled or de-ionized water to prevent deposition of buffer solution
crystals.
4-18 LC-20AT
5 Application Operation
5
Contents
PRODUCT INFO
Press func or VP
VP function settings
• Mobile phase monitor
• Product information
• Maintenance information
• Validation support
• Calibration support
5-2 LC-20AT
5.1 Display Panel
Initial screen
0.000mL MAX 10.0
0.0MPa MIN 0.0
LC-20AT 5-3
5. Application Operation
0.000 EVENT ™0
0,1,2 or 12
Compressibility compensation
P.5-16
0.000 COMP ™ .45
Input 0.00-3.00
5-4 LC-20AT
5.1 Display Panel
5
Enter to Delete
LC-20AT 5-5
5. Application Operation
5-6 LC-20AT
5.1 Display Panel
LC-20AT 5-7
5. Application Operation
Initial screen
0.000mL MAX 10.0
0.0MPa MIN 0.0
5-8 LC-20AT
5.1 Display Panel
5
through right seal P.5-33
R SEAL DELIVERED
™ 0 / 180 L
LC-20AT 5-9
5. Application Operation
Time P.5-35
TIME HH:MM:SS
™ 0:00:00
5-10 LC-20AT
5.1 Display Panel
Password P.5-39
∗ If password is input incorrectly, subsequent screens
Calibration support group cannot be accessed.
LC-20AT 5-11
5. Application Operation
About files:
The instrument can store up to 10 files of flow rate and other parameter values, and time programs in
memory .
File No. 0 to 9 are assigned to each file. The auxiliary functions other than [SV] and [EVENT] are common
functions.
Each value can be switched by [FILE NUM] in the File Operation Group.
∗ Operation in the table head shows the types of operation described below.
Display : Check the monitor.
: Press to activate the function.
Numeric keypad : Press - to enter a value and press to determine the value.
5-12 LC-20AT
5.2 Parameter in Auxiliary Functions
∗ Operation in the table head shows the types of operation described below.
Display : Check the monitor.
: Press to activate the function.
Numeric keypad : Press - to enter a value and press to determine the value.
LC-20AT 5-13
5. Application Operation
1 Press .
Initial screen appears.
0.000mL MAX 10.0
0.0MPa MIN 0.0
0.000 SV ™
Select Port No.
5 Press
functions.
or to move to the other
6 Press
Press
to show the group screen.
to return to the initial screen.
5-14 LC-20AT
5.2 Parameter in Auxiliary Functions
0.000 PARAMETER
Enter to Select
0.000 SV ™
Select Port No.
5
Setting Solenoid Valve Position [SV]
Selects the position of the solenoid valve when using the
low-pressure gradient unit or FCV-11AL reservoir
switching valve (option). Enter the value corresponding
to the selected mobile phase and press .
1 Solvent A
Low-pressure 2 Solvent B
gradient unit 3 Solvent C
4 Solvent D
NOTE
LC-20AT 5-15
5. Application Operation
12 Relay 1 ON Relay 2 ON
Water 0.45
Acetonitrile 1.20
Methanol 1.25
Hexane 1.60
5-16 LC-20AT
5.2 Parameter in Auxiliary Functions
1 1-cycle mode
2 2-cycle mode
4 4-cycle mode
5
Setting Purge Time [P-TIMER]
Sets purge time.
Enter a time using the numeric keypad and press
. 0.000 P-TIMER ™3
Setting range is 1-20 (minutes) with 1 minutes step as Input 1 - 20 min
setting unit.
LC-20AT 5-17
5. Application Operation
5-18 LC-20AT
5.2 Parameter in Auxiliary Functions
0.000 FILE
Enter to Select
LC-20AT 5-19
5. Application Operation
0.000 CONTROL
Enter to Select
2 Press .
NOTE
5-20 LC-20AT
5.2 Parameter in Auxiliary Functions
0.000 SYSTEM
Enter to Select
LC-20AT 5-21
5. Application Operation
NOTE
NOTE
5-22 LC-20AT
5.2 Parameter in Auxiliary Functions
1 FCV-11AL
1 kgf/cm2
2 bar
3 psi
Examples:
• When set value is 1, the full scale is 5.0MPaFS.
• When set value is 10, the full scale is 50.0MPaFS.
LC-20AT 5-23
5. Application Operation
0.000 MONITOR
Enter to Select
5-24 LC-20AT
5.2 Parameter in Auxiliary Functions
Monitoring Flow Rate in Constant Pressure Solvent Delivery Mode [MON FLOW]
Monitors the flow rate in the constant pressure solvent
delivery mode.
Enter a set value and press . 0.000 MON FLOW ™
0:Clear 1:Set
Set value Function
LC-20AT 5-25
5. Application Operation
5-26 LC-20AT
5.3 VP Functions
5.3 VP Functions
VP functions support the validation of the instrument by check functions or displaying the instrument
information.
There are five groups for VP functions: Mobile Phase Monitor, Product Information, Maintenance Information,
Validation Support, and Calibration Support.
Display/
Function
ALARM LEVEL Numeric keypad To set the alarm level of remaining mobile phase. P.5-31
∗1 Mobile phase monitor functions are available only for one type of liquid in standard mode. However, 4 types of liquide (liquide A-D) can
be available in low-pressure gradient mode.
SERIAL NUMBER Display To show the serial number of the instrument P.5-32
∗ Operation in the table head shows the types of operation described below.
Display : Check the monitor.
: Press to activate the function.
Numeric keypad : Press - to enter a value and press to determine the value.
LC-20AT 5-27
5. Application Operation
To show the total volume of flow rate through left seal, and
L SEAL DELIVERED Display/ P.5-33
volume at a timing to replace a seal
PART REPLACEMENT Numeric keypad To record the replaced parts number P.5-33
Display/Numeric
DATE To show/set the date P.5-35
keypad
Display/Numeric
TIME To show/set the time P.5-35
keypad
Numeric keypad/
PULSE CHECK To run the pulsation check P.5-36
∗ Operation in the table head shows the types of operation described below.
Display : Check the monitor.
: Press to activate the function.
Numeric keypad : Press - to enter a value and press to determine the value.
5-28 LC-20AT
5.3 VP Functions
FLOW COMP FACT Numeric keypad To enter the flow rate compensation factor (ALPHA) P.5-39
PRESS COMP FACT Numeric keypad To enter pressure sensor sensitivity compensation factor P.5-39
LEAK THR Numeric keypad To enter leak sensor operation level P.5-40
SEAL REPLACEMENT Numeric keypad To set the volume at which seals are replaced P.5-40
∗1 If the password is not input correctly, the functions after [FLOW COMP FACT] in the Calibration Support Group cannot be accessed,
even is pressed. 5
∗ Operation in the table head shows the types of operation described below.
Display : Check the monitor.
: Press to activate the function.
Numeric keypad : Press - to enter a value and press to determine the value.
LC-20AT 5-29
5. Application Operation
1 Press .
The initial screen appears.
0.000mL MAX 10.0
0.0MPa MIN 0.0
MOBILE PHASE
Press func or VP
MOBILE PHASE
0 / ™000 mL
5-30 LC-20AT
5.3 VP Functions
MOBILE PHASE
Press func or VP
WARN : M.PHASE
Press CE Key
LC-20AT 5-31
5. Application Operation
SERIAL NUMBER
L20110000000
TOTAL OP TIME
1(h)
5-32 LC-20AT
5.3 VP Functions
1 2
∗ The left plunger seal is [L], as viewed from the front of
the instrument.
If less than 10 logs are recorded, the screen displays the LOG 1 03-05-12
message as shown on the right. P/N:228-30000-91
Press to return to the title screen.
No more Logs
LC-20AT 5-33
5. Application Operation
No more Logs
If less than 10 logs are recorded, the screen displays the LOG 1 03-05-12
message as shown on the right. CHANGE PASSWORD
Press to return to the title screen.
No more Logs
5-34 LC-20AT
5.3 VP Functions
Date YY-MM-DD
03-01-02
2 Press .
LC-20AT 5-35
5. Application Operation
Result
MEMORY CHECK
ROM OK / RAM OK
3 Press .
The date and results are recorded.
1.000 wait...
1.99min left
NOTE
PULSE 0.17MPa OK
Enter to Memory
5-36 LC-20AT
5.3 VP Functions
1 Press .
Pumping starts at 1mL/min. Prepare for
measuring. FLOW CHECK
Enter to Start
2 When the preparation for measuring is done,
press again to measure the required time
for flowing liquid to fill 5mL of measuring flask.
Setting Time Program to Check Concentration Accuracy of Gradient Mode [GE TEST
PROGRAM]
Enter a file number 0-9 of time program in which the
default gradient test program is to be copied. This file is
automatically loaded for execution. GE TEST PROGRAM
Copy to File ™
∗ The test program is the gradient time program
described in P.7-22.
LEAKAGE TEST
2 Press during the pump stopped. Enter to Start
LC-20AT 5-37
5. Application Operation
Leak sensor
Fig. 5.1
NOTE
After wetting and testing the leak sensor, wipe away the
water on the leak tray completely.
"8.11 Wiping of Leak Tray" P. 8-28
5-38 LC-20AT
5.3 VP Functions
NOTE CALIBRATION
Press func or VP
The instrument is adjusted before leaving the factory. Do
not change values unnecessarily.
LC-20AT 5-39
5. Application Operation
0 CBM-20A/20Alite
1 SCL-10Avp
2 SCL-10A
5-40 LC-20AT
5.3 VP Functions
1 Press .
Input screen appears.
CHANGE PASSWORD
Enter to Change
NOTE
LC-20AT 5-41
5. Application Operation
NOTE
5-42 LC-20AT
5.3 VP Functions
1 Not use
LC-20AT 5-43
5. Application Operation
2 To connect to CLASS-VP
3 To connect to LCsolution
11 To connect to C-R8A
12 To connect to C-R7A/C-R5A
13 To connect to C-R4A
NOTE
5-44 LC-20AT
5.4 Creating Time Program
5
FLOW (Available only in constant flow solvent delivery 0 − 10.000mL/min unit: P.4-3
mode) 0.001mL/min
Minimum setting
• 1.0 ∼ 40.0 MPa units:
To set pressure
• 10 ∼ 408 kgf/cm2 0.1MPa
PRESS (Available only in constant pressure solvent P.4-7
• 10 ∼ 400 bar • 1kgf/cm2
delivery mode)
• 142 ∼ 5804 psi • 1bar
• 1psi
0 - 255
Setting 0 will repeat Minimum setting
LOOP To repeat program P.5-54
the program 256 units: 1
times.
LC-20AT 5-45
5. Application Operation
1 Press .
Initial screen appears.
0.000mL MAX 10.0
0.0MPa MIN 0.0
2 Press .
The screen appears as shown on the right.
Number of steps already created USED LEFT
Number of remaining steps 10 310
This example shows that 10 steps of time program
have been set, and that 310 steps remain.
1 2
3 Press .
Editing screen appears.
1 2 3
5-46 LC-20AT
5.4 Creating Time Program
Initial screen
Input Stored
™5.00 FLOW 4.000
0.01 - 999.99min
LC-20AT 5-47
5. Application Operation
3 30.00 STOP
Min.
Fig. 5.2
1 Press .
Initial screen appears.
1.000mL MAX 10.0
0.0MPa MIN 0.0
2 Press .
The number of steps already created and
remaining steps appear. USED LEFT
0 320
Time setting
3 Press .
The setting screen appears.
™IME FUNC VALUE
0.01 - 999.99min
Step 1
15.00 ™LOW
0.000 - 10.000mL
5-48 LC-20AT
5.4 Creating Time Program
5 Press repeatedly until [FLOW] appears. Input the value of the command (Flow rate)
5
Step 2
8 Press
time.
, and to enter the Determination of step 2
Step 3
30.00 ™LOW
0.000 - 10.000mL
30.00 STOP
11 Press . 0.01 - 999.99min
LC-20AT 5-49
5. Application Operation
NOTE
5-50 LC-20AT
5.4 Creating Time Program
Concentration of Mobile phase A (%) = 100 (%) - Concentration of Mobile phase B [BCNC] (%)
The value of [BCNC] command can be set to any concentration between 0 and 100%.
Minimum setting unit is 0.1%.
40.00 STOP
0 20 30 40 min
LC-20AT 5-51
5. Application Operation
Mobile phase A% = 100 (%) - Mobile phase B% [BCNC] - Mobile phase C% [CCNC] - Mobile phase D%
[DCNC]
The values of [BCNC], [CCNC] and [DCNC] can be set to any value between 0 and 100%.
Minimum setting units: 0.1%.
2 Press .
Step 1 will be deleted, and the subsequent step
appears. ™0.00 FLOW 5.000
0.01 - 999.99min
5-52 LC-20AT
5.4 Creating Time Program
Fig. 5.4
LC-20AT 5-53
5. Application Operation
30.00 LOOP 3
NOTE
• After the time program is repeated as set for [LOOP], how the time program will automatically stop depends
on whether [GOTO] command following [LOOP] has been entered or not as follows.
With [GOTO] command:
Until the time set for [GOTO] command is reached, time program maintain the setting values when [LOOP]
is ended, and then the next time program starts.
Without [GOTO] command:
The time program will stop when [LOOP] has finished.
• Steps set after [LOOP] command will deactivate except for [GOTO] command.
• Value of [LOOP] command can be set to 255. When [0] is set, time program will be repeated 256 times.
Event cable
Fig. 5.5
5-54 LC-20AT
5.4 Creating Time Program
5
5 Set the EVENT output terminal of the newly called
up file specified in step 4 as [1].
"Setting EVENT Output Terminals [EVENT]" 0.000 EVENT 1
P. 5-16 0,1,2 or 12
LC-20AT 5-55
5. Application Operation
5.5.1 Preparation
To control the instrument by the CBM-20A or CBM-20Alite system controller, set the parameters as follows:
5-56 LC-20AT
5.6 Control by SCL-10Avp or SCL-10A System Controller
5.6.1 Preparation
To control the instrument by the SCL-10Avp or SCL-10A system controller, set the parameters as follows:
1 : VP ∗1
OP MODE "Selecting Operation Mode [OP MODE]" P. 5-40
2 : 10A∗2
∗1 To connect to SCL-10Avp : the instrument is recognized as LC-10ATvp.
∗2 To connect to SCL-10A : the instrument is recognized as LC-10AT.
5
5.6.2 Basic Parameters
A max.of three pump units can be connected to a SCL-10Avp or SCL-10A. SCL-10Avp/SCL-10A can set flow
rate (pressure), maximum pressure limit, minimum pressure limit, solenoid valves, pumping ON/OFF, time
program, etc. and control binary high-pressure gradient, ternary high-pressure gradient and quaternary low-
pressure gradient. Refer to the SCL-10Avp/SCL-10A instruction manual.
5.6.3 Attention
When the instrument is connected to a SCL-10Avp or SCL-10A, the instrument will work in compatibility mode
for LC-10ATvp or LC-10AT. In this case, the following applies.
1) Maximum limit of flow rate is 9.999mL/min.
2) Maximum value of [P.MAX] is 43.2MPa.
3) Only [0, 1, 4] can be set for [LPGE MODE]. However, the operation in [0] setting is slightly different from the
one of LC-10AT(vp).
LC-20AT 5-57
5. Application Operation
WARNING
• Before connecting the cable, turn off the power and unplug the instrument.
• Use only the specified cable.
• Connect as specified.
Otherwise, fire, electric shock or malfunction may occur.
Fig. 5.6
5-58 LC-20AT
5.7 Connection to External Input/Output Terminals
peel 10mm
NOTE
The instrument provides one event cable. When more than 2 cables are required, use the following cables.
• Cable with single wire : φ 0.4 to φ 1.2 (AWG26 to 16)
• Cable with stranded wire : 0.3mm2 to 1.25mm2 (AWG22 to 16), diameter of single wire thicker than φ 0.18.
The cable with stranded wire is suitable to prevent disconnection.
NOTE
0.000 EXT-S 0
Input 0 - 3
LC-20AT 5-59
5. Application Operation
5-60 LC-20AT
6 Troubleshooting
6
Contents
Power does not turn Power plug is disconnected. • Connect plug correctly. P.9-5
ON even after
Power cord internal wires are • Replace with a new cord of the same type.
switching ON power. P.1-3
cut.
Power supply does not meet • Use power supply that meets specifications
specifications for this for this instrument. P.9-5
instrument.
Flow rate is set to [0]. • Set a flow rate other than 0. P.4-3
Error message (P.MAX, P.MIN • Press to clear the message and make
P.6-5
etc.) appears. corrective action for errors.
The instrument is Air bubbles are generated inside • Press to purge out the bubbles.
running, but no liquid pump head. • Insert the disposable syringe into the drain P.3-7
is pumped. tubing outlet and draw out the bubbles.
Air bubbles are produced • Make sure that the filter bushing is tightly
through suction filter and pump fitted. P.9-13
inlet.
Check valve is not working • Push in 2-propanol etc. from the check valve
properly. inlet.∗2
6-2 LC-20AT
6.1 Troubleshooting and Corrective Action
Pumping is unstable, Air bubbles are generated inside • Press to purge out the bubbles.
with large pump pump head. • Insert the disposable syringe into the drain P.3-7
pulsation. tubing outlet and draw out the bubbles.
Previous mobile phase is still • Press , and thoroughly purge the old
remaining inside pump head. mobile phase.
Air bubbles are generated in • Press , and thoroughly purge the old
filter tubing. mobile phase.
Check valve is not working • Push in 2-propanol etc. from the check valve
properly. inlet.∗2
Liquid is leaking from flow line • Retighten the male nuts. Otherwise, replace
P.9-8
connection. the nuts and ferrules.
Flow line is (partially) clogged. • Clean the line filter in an ultrasonic bath or
P.8-22
replace it.
LC-20AT 6-3
6. Troubleshooting
Flow rate is below set Check valve does not work • Pump 2-propanol to clean check valve. P.8-18
value. properly.
• Pump 2-propanol to clean check valve.
Otherwise, clean the valve in an ultrasonic P.8-19
bath or replace it.
Retention time cannot Check valve does not work • Pump 2-propanol to clean check valve. P.8-18
be reproduced. properly.
• If cleaning check valve is not effective,
replace the valve or clean it in an ultrasonic P.8-19
bath.
In a high-pressure Zero points of the pressure • Use [ZERO ADJ] auxiliary function to execute
gradient system, the sensors of two pump units are zero adjustment of the pressure sensors. P.5-20
two pump units show not the same.
different pressures.
Line filter of one of the pump • Clean the line filter in an ultrasonic bath, or
(Difference of up to P.8-22
units is clogged. replace it.
2.0MPa or 5% is
normal.) Flow line(s) is/are clogged at the • Identify the clogged lines and replace it.
junction of the two pump unit
flow lines.
Pressure does not Drain valve open. • Close drain valve. P.2-2
rise.
Liquid is leaking from flow line • Retighten the male nuts. Otherwise, replace
P.9-8
connections. the nuts and ferrules.
Pressure rises too Line filter is clogged. • Clean the line filter in an ultrasonic bath or
P.8-22
much. replace it.
(Remove the column
Flow lines are clogged. • Identify the clogged parts and replace them.
to check.)
Inner diameter of tubings is too • Use tubings with the specified inner diameter.
small.
6-4 LC-20AT
6.2 Error Message
NOTE
ERROR OVER HEAT Fatal Cause: Interior temperature has risen to an abnormal
level.
(Overheating)
Action: Check to be sure that the internal fan can move,
and that the rear exhaust vent is not blocked. If
this message still appears, turn the power OFF
and contact your Shimadzu representative.
OUT OF MAXSTEPS Alarm Cause: More than 320 steps of time program have been
set.
(Max. number of steps exceeded)
Action: Delete unnecessary programs.
"5.4 Creating Time Program" P. 5-45
LC-20AT 6-5
6. Troubleshooting
ERROR P-MAX Alarm Cause: Pump discharge pressure has exceeded the set
max. pressure limit, [P.MAX]. Pumping will
(Max. pressure limit error)
automatically stop, unless [S-PROT] function has
been activated, in which case pumping will
continue at half flow rate.
"Setting System Protection [S-PROT]" P. 5-
23
Action: Inspect the clogging in the flow line. If no clogging
is found, set the appropriate [P.MAX] value.
"4.1.3 Setting Maximum Pressure Limit" P. 4-
4
ERROR P-MIN Alarm Cause: Pump discharge pressure has fallen below the set
min. pressure limit, [P.MIN]. Pumping will
(Min. pressure limit error)
automatically stop.
Action: Inspect the leakage in the flow line. If no leakage
is found, set the appropriate [P.MIN] value. Note
that [P.MIN] function is not activated for first one
minute of pumping.
"4.1.4 Setting Minimum Pressure Limit" P. 4-5
ERROR HOME POS Alarm Cause: The motor home position cannot be detected, the
motor does not run, or the motor is slipping.
(Home position error)
Action: Turn the power OFF and ON again, and press
. If this message still appears, turn the
power OFF and contact your Shimadzu
representative.
OPEN DRAIN VALVE Alarm Cause: Pump discharge pressure has exceeded [P-
PMAX] during purging.
(Purge error)
Action: Open the drain valve before pressing .
If an error message appears during auto-purging
by system controller, inspect the clogging in the
flow line. If no clogging is found, set the
appropriate [P-FLOW] and [P-PMAX] value.
"Setting Purge Flow Rate [P-FLOW]" P. 5-17
"Setting Maximum Pressure Limit during
Purging [P-PMAX]" P. 5-18
"5.5 Control by CBM-20A or CBM-20Alite
System Controller" P. 5-56
6-6 LC-20AT
6.2 Error Message
ERROR EXTERNAL Alarm Cause: An error signal was sent from the external device
connected to the external input/output terminal
(External equipment error)
[ERR IN].
Action: Inspect the external device, and eliminate the
cause of the error.
WARN : M.PHASE Warning Cause: The remaining volume of mobile phase is less
than the alarm level.
(Mobile phase alarm)
Action: Replace the mobile phase and reset the amount of
mobile phase.
"Showing/Setting Mobile Phase Volume
[MOBILE PHASE]" P. 5-31
LC-20AT 6-7
6. Troubleshooting
6-8 LC-20AT
7 Hardware Validation
7
This chapter provides instruction on hardware validation, which verifies the
performance of individual components and the instrument as a whole.
Contents
The operational protocol and criteria for this component and the HPLC system are described in this chapter to
assist the user in conducting validation. Refer to each the instruction manual for each component for
operational protocol of that specific component.
7-2 LC-20AT
7.2 Implementation of Hardware Validation
7
After any maintenace, component performance must be re-validated. The type of validation depends on the
actual work done.
If the maintenance inspection cannot be performed solely by the specific component validation, system
validation is required.
NOTE
Maintenance information and results of hardware validation must be recorded and kept for future reference.
LC-20AT 7-3
7. Hardware Validation
7.3.1 Environment
Instrument performance may be affected by abrupt changes in ambient temperature such as drafts from
heating and air conditioning vents.
The equipment should be installed in a room with minimal (< 2°C) temperature fluctuation and away from
sources of drafts and air currents.
WARNING
• Provide ample ventilation with no fire sources in vicinity
When flammable or toxic solvents are used as the mobile phase, the room must be properly ventilated.
When flammable solvents are used, open flame or other fire sources must be strictly prohibited.
CAUTION
• Avoid dust or corrosive gas
Avoid installing the instrument in places subject to excessive dust or corrosive gas since service life
and performance levels may be affected.
• Keep away from strong magnetic fields
Do not install the instrument near equipment that generates strong magnetic fields. If the power supply
line is subject to high electrical noise, use a commercially-available power surge protector.
• Provide adequate installation surface and space
The weight of LC-20AT is 11kg. During installation, consider the entire weight combined with other LC
components.
The lab table on which this instrument is installed should be strong enough to support the total weight
of the LC system. It should be level, stable and have depth of at least 600mm.
If these precautions are not followed, the instrument could tip over or fall off the table.
When components are installed side by side, maintain a keep space of at least 30 mm between the
components.
• Regulate room temperature and humidity
The room temperature should be between 4 and 35°C, with minimal temperature variations throughout
the course of a day. Humidity should be kept within 20-85%.
• Position instrument properly in the room
Install the instrument in a location that is free from vibration and away from sunlight, and heat/air
conditioning drafts.
7-4 LC-20AT
7.4 Equipment Required for Validation
Testing Equipment
A list of testing equipment required for hardware validation is shown below. A certificate ensuring traceability
or inspection results should accompany each item of testing equipment that is used.
Equipment Description
For inspection of the temperature setting accuracy for the column oven and the
autosampler's sample cooler.
Thermo recorder
The thermo recorder must be certified as having an accuracy rating of ±1.0°C for the
required temperature range (0°C to 50°C) at the time of inspection.
For inspection of the temperature accuracy for the column oven and autosampler's sample
cooler.
Thermocouple
The thermocouple must have a testing accuracy of ±0.6°C for the required temperature
range (0°C to 50°C) at the time of inspection.
DC voltage/current
For the hardware validation of the chromatopac.
The DC voltage/current generator must be certified as having an accuracy rating of
7
generator
±0.15% at the time of testing.
For inspection of the flow rate accuracy for the solvent delivery module.
Stopwatch
The stopwatch must be certified at 5'30" ±0.3sec at the time of inspection.
For inspection of the flow rate accuracy for the solvent delivery module.
Measuring flask
Obtain a 5mL-measuring flask.
LC-20AT 7-5
7. Hardware Validation
7-6 LC-20AT
7.5 Validation: Pump
7.5.2 ROM, RAM Self Diagnosis Checks whether the memory (ROM, RAM) functions correctly.
7.5.4 Display, LED Test Checks the operation of display and LEDs.
7.5.5 Flow Rate’s Control Parameter Checks parameters to control the flow rate.
7.5.6 Pumping Stability Test Checks that the pump's pulse width does not exceed the rated value.
7.5.9 Flow Rate Accuracy Test Checks the flow rate accuracy by using a measuring flask.
Check Procedure
2 Press
[VALIDATION].
four times to display
3 Press
CHECK].
three times to display [MEMORY
MEMORY CHECK
ROM OK / RAM OK
4 Press .
"Checking Memory [MEMORY CHECK]" P.
5-36
LC-20AT 7-7
7. Hardware Validation
Check Procedure
2 Press
number.
two times to display the version
"Showing S/W Version No. [S/W ID]" P. 5-32 S/W ID: V*.**
LC-20AT
Check Procedure
CHECK CRITERIA : All the dots and LEDs on the screen illuminate.
7-8 LC-20AT
7.5 Validation: Pump
Check Procedure
2 Press
PASSWORD].
once to display [Input
Input PASSWORD
™
5 Press once.
Make a note of the displayed two values (1: [PRS-
1] and 5: [PRS-5]) of [PRESS COMP FACT] . PRESS COMP FACT
"Setting Pressure Sensor Sensitivity 1:1456 5:8535
Compensation Factor [PRESS COMP
FACT]" P. 5-39
LC-20AT 7-9
7. Hardware Validation
Resistor tube
Check Procedure
5 Press ⋅ ⋅
to set the value as[0.45].
⋅ ⋅
6 Press
screen.
two times to return to the initial
7 Press
[VALIDATION].
four times to display
8 Press
CHECK].
four times to display [PULSE
7-10 LC-20AT
7.5 Validation: Pump
9 Press ⋅ ⋅
set [CRITERIA] as [0.30MPa].
⋅ to
PULSE CHECK
10 Press .
The pump operates at 1mL/min and measurement
CRITERIA:0.30MPa
11 Press .
The result and date having measured will be
stored at the instrument memory.
CHECK CRITERIA :
The measurement result for pressure fluctuation width does not exceed 0.30MPa.
7
Resetting after Check
2 Press or
[COMP] appears.
repeatedly until
LC-20AT 7-11
7. Hardware Validation
Resistor tube
Check Procedure
4 Press to activate setting of the flow rate. Cursor (positioned to prompt input)
5 Press
[1mL/min].
and to set the flow rate to ™.000mL MAX 10.0
0.0MPa MIN 2.0
6 Press
[P.MAX].
two times to activate input of
7 Press ,
[20.0MPa] as [P.MAX].
and to set
1.000mL MAX ™0.0
0.0MPa MIN 2.0
7-12 LC-20AT
7.5 Validation: Pump
8 Press
[P.MIN].
three times to activate input of
Illuminates
7
12 Change the flow rate to [1.5mL/min], and observe
the pressure.
When the actual pressure exceeds the [P.MAX]
value set by step 11, [ERROR P-MAX] should ERROR P-MAX
appear. 11.1MPa MIN 9.0
LC-20AT 7-13
7. Hardware Validation
ERROR P-MIN
5.0MPa MIN 9.0
Check Procedure
Leak sensor
Fig. 7.1
7-14 LC-20AT
7.5 Validation: Pump
NOTE
After wetting and testing the leak sensor, wipe away the water on the leak tray completely.
"8.11 Wiping of Leak Tray" P. 8-28
Resistor tube
Stopwatch
Check Procedure
7
1 Connect the resistor tube to the pump outlet
instead of HPLC column.
4 Press
[1mL/min].
and to set the flow rate to
LC-20AT 7-15
7. Hardware Validation
7 Press ⋅ ⋅
to set [0.45] as [COMP].
⋅ and
Ready ?
Enter to Measure
7-16 LC-20AT
7.5 Validation: Pump
Measuring flask
Fig. 7.2
< For a measuring cylinder>
1) Bring the end of the tube into contact with the
cylinder wall.
2) Determine a suitable water level inside the
cylinder at which to start measurement. Start the
stopwatch when the water reaches the
determined start level.
3) Stop the stopwatch when the water is pooled Measuring
5mL. cylinder
7
NOTE
Fig. 7.3
Make sure that the end of the tube remains above the
water in the graduated cylinder throughout the
measurement.
LC-20AT 7-17
7. Hardware Validation
2 Press or
[COMP] appears.
repeatedly until
Check Procedure
3 Press ⋅
mode as isocratic.
and set the pumping
4 Press
screen.
two times to return to the initial
5 Press
rate.
once to activate setting of the flow
7-18 LC-20AT
7.5 Validation: Pump
7
CHECK CRITERIA : Valve switches correctly.
LC-20AT 7-19
7. Hardware Validation
∗ Both mobile phases must be degassed using helium degasser or degasser unit.
Check Procedure
7-20 LC-20AT
7.6 Validation: Gradient Pump System (For Gradient System)
A: 50.0 B:™50.0
7
• UV detector Wavelength 272nm
RESPONSE 3 (0.5sec)
AUX RANGE 2 (1.0AU/V)
• Data processor 1,024 mAU full scale (for Chromatopac, ATTEN=10)
(With LCsolution, the intensity axis is set automatically.)
Chart speed 5mm/min (for Chromatopac, SPEED=5)
the right.
• If a system controller or LC workstation is used,
select time program from each menu, and
create the program.
Time (min.)
Fig. 7.4
LC-20AT 7-21
7. Hardware Validation
For C-R8A/R7A/R4A
8 If a Chromatopac is used as the data processor,
create the BASIC programs given on the right,
LINE PROGRAM
10 LPRINT LEVEL(1);
which print the signal level at every 30 seconds
(according to the model). 20 WAIT 30
30 GOTO 10
40 END
For C-R3A/R6A
LINE PROGRAM
10 PRINT LEVEL;
20 WAIT 30
30 GOTO 10
40 END
For C-R5A
LINE PROGRAM
10 LPRINT LEVEL;
20 WAIT 30
30 GOTO 10
40 END
7-22 LC-20AT
7.6 Validation: Gradient Pump System (For Gradient System)
NOTE
<Calculation method>
LC-20AT 7-23
7. Hardware Validation
Resistor tube
A Distilled water
(How to prepare)
<When using Shimadzu's standard reagents for validation>
Mobile Weigh 20mg of caffeine (250mg/L, Part No. 228-45725-06) and
phase∗ bring it to 500mL volume with water.
500mL B,C,D Caffeine solution (10mg/L) <In case of preparation by customer>
for each Weigh 20mg of anhydrous caffeine, transfer them to a 100mL
measuring flask and bring it to100mL volume with water.
Then take 25mL of the solution, and add water to bring it to 500mL
volume.
∗ Both mobile phases must be degassed using helium degasser or degasser unit.
Check Procedure
7-24 LC-20AT
7.6 Validation: Gradient Pump System (For Gradient System)
A: 25.0 B:™25.0
C: 25.0 D: 25.0
the right.
• If a system controller or LC workstation is used,
select time program from each menu, and
create the program.
Time (min.)
Fig. 7.5
LC-20AT 7-25
7. Hardware Validation
For C-R8A/R7A/R4A
8 If a Chromatopac is used as the data processor,
create the BASIC programs given on the right,
LINE PROGRAM
10 LPRINT LEVEL(1);
which print the signal level at every 30 seconds
(according to the model). 20 WAIT 30
30 GOTO 10
40 END
For C-R3A/R6A
LINE PROGRAM
10 PRINT LEVEL;
20 WAIT 30
30 GOTO 10
40 END
For C-R5A
LINE PROGRAM
10 LPRINT LEVEL;
20 WAIT 30
30 GOTO 10
40 END
NOTE
7-26 LC-20AT
7.6 Validation: Gradient Pump System (For Gradient System)
<Calculation method>
7
phases A and C, also A and D.
LC-20AT 7-27
7. Hardware Validation
20mg/L caffeine solution (included in Caffeine set (5 concentrations) Part No. 228-45725-91)
<Preparation>
Sample Weigh 20mg of anhydrous caffeine, transfer to a 100mL volumetric flask and dilute to volume with
water.
Transfer 1mL of the solution to a 10mL volumetric flask, and dilute to volume with water.
7-28 LC-20AT
7.7 System Validation
2
Coupling
Check the LC system plumbing. 1.6C
Ensure that the tubing between (a) the
autosampler outlet and the column inlet, (b) the
column outlet and the detector inlet, has an I.D. of
less than 0.3mm, and is shorter than 300mm.
Keep the liquid volume that is not in the column as
low as possible.
Fig. 7.6
LC-20AT 7-29
7. Hardware Validation
Validation Procedure
procedures.
1.000mL MAX 20.0
0.0MPa MIN 0.0
setting procedures.
SET TEMP 40
Input 4 - 85
Validation" P. 7-31
See the detector’s instruction manual for setting 272nm 0.000 AU
procedures. D2 1.0000AUFS
7-30 LC-20AT
7.7 System Validation
Fig. 7.8
Parameter Settings for Isocratic System Validation
The parameters to be set for the various devices when validation analysis of an isocratic system is performed
are given below.
CHECK CRITERIA
The RSD (C.V.)'s obtained must satisfy the following criteria:
Retention time RSD must not exceed 0.5%.
Peak area RSD must not exceed 1.0%.
LC-20AT 7-31
7. Hardware Validation
A: Distilled water
B: Methanol
Mobile phases
A /B =60%/40%
∗Both the water (distilled) and the methanol should be HPLC grade.
Shim-pack VP-ODS (Part No. 228-34937-91), LUNA C18 (2) (Part No. 00F-4252-E0) or
Column
equivalent ODS column (Particle size 5µm, Column Dimension : I.D. 4.6mm × length 150mm)
20mg/L caffeine solution (included in Caffeine set (5 concentrations) Part No. 228-45725-91)
< Preparation >
Sample Weigh 20mg of anhydrous caffeine, transfer to a 100mL volumetric flask and dilute to volume
with water.
Transfer 1mL of the solution to a 10mL volumetric flask, and dilute to volume with water.
2
Coupling
Check the LC system plumbing.
1.6C
Ensure that the tubing between (a) the
autosampler outlet and the column inlet, (b) the
column outlet and the detector inlet, has an I.D. of
less than 0.3mm, and is shorter than 300mm.
Keep the liquid volume that is not in the column as
low as possible.
Fig. 7.9
7-32 LC-20AT
7.7 System Validation
Validation Procedure
Pump’s display screen
LC-20AT 7-33
7. Hardware Validation
setting procedures.
SET TEMP 40
Input 4 - 85
Validation" P. 7-35
See the detector's instruction manual for setting
272nm 0.000 AU
procedures. D2 1.0000AUFS
Fig. 7.11
7-34 LC-20AT
7.7 System Validation
CHECK CRITERIA
The RSD (C.V.)'s obtained must satisfy the following criteria:
Retention time RSD must not exceed 0.5%.
Peak area RSD must not exceed 1.0%.
LC-20AT 7-35
7. Hardware Validation
• Check whether any consumable items have reached the end of their service life:
The cause of failure to satisfy check criteria could be a consumable part that is no longer usable. Check
consumable parts and replace them if necessary.
• Perform troubleshooting:
It is possible that some minor problem (such as air bubbles) has caused the system to fail the criteria.
Perform troubleshooting to check for such problems, and take action to eliminate any problems found.
For troubleshooting procedures for individual system components, see the applicable instruction
manuals.
7-36 LC-20AT
8 Maintenance
8
Contents
WARNING
• Unless the instructions here specified, turn off the power always and unplug the instrument prior to
performing inspections and maintenance. Otherwise, fire, electric shock or malfunction may occur.
CAUTION
• When replacing parts, use only the parts listed in "1.3Component Parts" and "9.3Maintenance Parts".
If any other parts are used, injury or malfunction may occur.
• Never remove the main cover. Otherwise, injury or malfunction may occur.
Contact your Shimadzu representative to remove the main cover.
8-2 LC-20AT
8.1 Periodic Inspection and Maintenance
CAUTION
The replacement and maintenance periods listed in this table are presented only as guidelines. These are
not guarantee periods. These will vary depending on usage conditions.
6 1 2 3
Inspection/Maintenance Item Remark Page
months year years years
• Sealing efficiency decreases when
seals are worn.
• Replace when the plunger is replaced.
• As a guideline, seals should be replaced
Replacement of plunger seal × after delivery of 180L when 2-propanol P.8-5
is pumped at pressure of 10MPa
(102kgf/cm2).
(VP function, [L(R) SEAL DELIVERED],
shows the cumulative pumping volume.)
Cleaning and inspection
× P.8-11
(replacement) of Plunger
Replacement of rinse flow line • Sealing efficiency decreases when
× P.8-16
seal seals are worn.
Inspection (replacement) and
ultrasonic bath cleaning of ×
inlet check valve
Inspection (replacement) and
ultrasonic bath cleaning of ×
P.8-19
8
primary-inlet check valve
Inspection (replacement) and • Particulates in mobile phase clog the
× P.8-22
cleaning of line filter filter in prolonged use.
Inspection (replacement) and • Particulates in mobile phase clog the
× P.8-24
cleaning of suction filter filter in prolonged use.
• Sealing efficiency decreases when the
Replacement of drain valve × P.8-25
drain valves are worn.
Pump ASSY lubrication × • Contact your Shimadzu representative. -
Fuse replacement × P.8-27
LC-20AT 8-3
8. Maintenance
Plunger
To pump the mobile phase.
Plunger surfaces may be damaged by foreign
matter from the mobile phase or by buffer
solution crystals.
Fuse
Plunger seal
To increase sealing efficiency to keep
Primary-inlet check valve the constant pumping
pressure.
Line filter
To filter particulates from the
mobile phase in order to Inlet check valve
protect the column.
Suction filter
To filter particulates from the
mobile phase to eliminate.
Fig. 8.1
8-4 LC-20AT
8.2 Replacement of Plunger Seal
Necessary parts
8
2 Press repeatedly until [CONTROL]
appears, and press .
[P-SET] display
Cursor blinks
[P-SET] appears on the screen and the cursor
blinks in the [P-SET] input field, prompting input.
0.000 P-SET ™
0:Clear 1:Set
3 Press and .
The pump indicator illuminates for a short while,
then goes out. When it goes out, the right and left
plungers move approximately to the same 0.000 P-SET 1
position. 0:Clear 1:Set
"Plunger Set [P-SET]" P. 5-20
LC-20AT 8-5
8. Maintenance
SUS tubing
Bush
Remove the lower male nut from the right pump head
with provided 8×10 wrench by holding the inlet check
valve.
Remove the bush of primary-inlet check valve with your
hand.
Wrench 8×10
Hold
Fig. 8.3
Fig. 8.4
8-6 LC-20AT
8.2 Replacement of Plunger Seal
CAUTION
Plunger
Ease the pump head out gently, keeping it aligned
with the plunger at all times. If it is pulled out
forcefully, the plunger could be bent.
Pump head
Fig. 8.5
Seal
installer/
remover
Flanged end
Plunger
seal
8
Spacer S
Fig. 8.6
Fig. 8.7
LC-20AT 8-7
8. Maintenance
CAUTION
After removing the seal, wipe the plunger head inner
side and plunger seal outlet with clean gauze soaked
in 2-propanol.
If seal material adheres to these surfaces, the
sealing efficiency of the plunger cannot be
maintained.
Spacer S
Fig. 8.8
the seal.
After installation, check if the seal has been
straight inserted all the way in the pump head.
Fig. 8.9
8-8 LC-20AT
8.2 Replacement of Plunger Seal
Pin
CAUTION
Groove
Ease the pump head in gently, keeping it aligned with
the plunger at all times. If it is pushed in forcefully, Plunger
the plunger could be bent. Pump head
Fig. 8.10
2 Place the two hex socket head bolts into the holes
in the pump head, and screw them in alternately
and uniformly with provided Allen wrench.
NOTE
8
Allen wrench
Fig. 8.11
Fig. 8.12
LC-20AT 8-9
8. Maintenance
NOTE
When the phenomena above happen even after replacement of the plunger seals, the surfaces of the plunger
may be scratched or nicked, in which case the plunger must be replaced.
"8.3 Cleaning and Inspection (Replacement) of Plungers" P. 8-11
8-10 LC-20AT
8.3 Cleaning and Inspection (Replacement) of Plungers
Necessary parts
Plunger
Pump head 8
Fig. 8.13
Fig. 8.14
LC-20AT 8-11
8. Maintenance
CAUTION
Ease the seal holder ASSY out gently, keeping it
aligned with the plunger at all times. If it is pulled out Plunger
Fig. 8.15
Fig. 8.16
Head holder
Allen wrench
Fig. 8.17
8-12 LC-20AT
8.3 Cleaning and Inspection (Replacement) of Plungers
CAUTION
When inserting the plunger tool into the plunger Plunger
holder, be careful not to bend the plunger.
Plunger tool
Fig. 8.18
Clean/Inspect
Plunger
8
Fig. 8.19
CAUTION
Plunger surfaces may be damaged by foreign matter
(from the mobile phase) or buffer solution crystals
inside the head holder.
LC-20AT 8-13
8. Maintenance
Plunger
Plunger tool
Fig. 8.20
CAUTION
Ease the seal holder in gently, keeping them
horizontal and aligned with the plunger at all times. If Plunger
it is pushed in forcefully, the plunger could be bent.
8-14 LC-20AT
8.3 Cleaning and Inspection (Replacement) of Plungers
Hex socket
Allen wrench head bolt
Fig. 8.22
LC-20AT 8-15
8. Maintenance
Necessary parts
Plunger
Fig. 8.23
8-16 LC-20AT
8.4 Replacement of Rinse Flow Line Seal
CAUTION
Ease the seal holder ASSY in gently, keeping
Plunger
horizontal and aligned with the plunger at all times.
If they are pushed in forcefully, the plunger could be
bent.
Seal holder ASSY
Fig. 8.26
LC-20AT 8-17
8. Maintenance
Male nut
Resistor tube
Fig. 8.27
6 Press once.
The cursor blinks in the flow rate display field,
Flow rate setting screen
Cursor blinks
prompting input of the flow rate.
7 Press
flow rate.
and to set [2.000] for the
8 Press .
Pumping of 2-propanol will begin at a rate of 2mL/
2.000mL MAX 30.0
min, and the pump indicator will illuminate. 20.0MPa MIN 0.0
Pump for at least 1 hour.
8-18 LC-20AT
8.6 Inspection (Replacement) and Ultrasonic Bath Cleaning of Check Valves
NOTE
In order to prevent mobile phase from flowing out of the suction tubing or pump head when plumbing is
disconnected from the pump, take one of the following precautions:
• Place the reservoir lower than the pump inlet.
• Empty the reservoir and tubing of all mobile phase, and unscrew and remove the filter bushing from the
pump inlet.
8
2 Unscrew the male nuts on both ends of the upper
SUS tubing using another wrench, and remove
the upper SUS tubing.
Wrench 8×10 Hold
Fig. 8.28
Primary-inlet
check valve
Bushing
Fig. 8.29
LC-20AT 8-19
8. Maintenance
CAUTION
Do not disassemble the check valves for any reason.
If they are disassembled, their performance may be Primary-inlet
affected. check valve
Fig. 8.30
Open
Drain valve
Fig. 8.31
8-20 LC-20AT
8.6 Inspection (Replacement) and Ultrasonic Bath Cleaning of Check Valves
10 Press once
The cursor blinks in the [flow] display field,
Flow rate setting screen
Cursor blinks
prompting input of the flow rate.
11 Press
[flow].
and to set [2.000] for the
12 Press .
The pump indicator illuminates, and liquid flows
from the drain tubing. Pump indicator illuminates
If liquid flows continuously, the check valves are
functioning normally.
If the liquid does not flow continuously from the
drain tubing, the check valves must be replaced.
LC-20AT 8-21
8. Maintenance
8×10 Wrench
Hold
2 Unscrew the male nuts on the end of the SUS
tubing using another wrench, and remove the
SUS tubing. SUS tubing
8×10 Wrench
Fig. 8.32
Drain valve
4 Wipe any dirt from the line filter hole with a cotton
swab.
Line filter
Fig. 8.33
6 Screw the line filter back into the line filter Drain valve
Male nut
Fig. 8.34
8-22 LC-20AT
8.7 Inspection (Replacement) and Ultrasonic Bath Cleaning of Line Filter
Replacing Frit
Nippers
Frit
8
Fig. 8.35
Filter body
Fig. 8.36
LC-20AT 8-23
8. Maintenance
Suction tubing
Suction filter
Fig. 8.37
8-24 LC-20AT
8.9 Replacement of Drain Valve ASSY
NOTE
Fig. 8.38
8
NOTE
Seal
Fig. 8.39
Fig. 8.40
LC-20AT 8-25
8. Maintenance
Cap
Screw
Knob
Fig. 8.41
8-26 LC-20AT
8.10 Replacement of Fuse
Coin
Unscrew by
using a coin
8
Fig. 8.42
Fuse holder
New fuse
Coin
Fig. 8.43
LC-20AT 8-27
8. Maintenance
NOTE
When a buffer solution is used as the mobile phase, it may deposit crystals on the leak tray when it
evaporates. Wipe the dirt on the leak tray with cloth soaked in water, as below.
Protective plate
2 Wipe away the leakage around the leak sensor
and on the leak tray completely.
Leak sensor
NOTE
Fig. 8.44
CAUTION
• Make sure to reinstall the protective plate after the wiping.
Leak sensor may not detect leakage without the protective plate.
8-28 LC-20AT
8.12 Exterior Cleaning
1 Dip a piece of cloth in a dilute neutral detergent and twist firmly to remove excess liquid. Use this cloth
to scrub the soiled area of the exterior surface of the instrument.
2 Dip a piece of cloth into water and twist firmly to remove excess liquid. Use this cloth to wipe away all
the remaining detergent. Use a dry cloth to remove all moisture from the exterior surface of the
instrument.
NOTE
Do not allow spilled water to remain on the instrument surface, and do not use alcohol or thinner-type solvents
to clean the surfaces. These can cause rusting and discoloration.
LC-20AT 8-29
8. Maintenance
8-30 LC-20AT
9 Technical Information
9
Contents
9.1 Installation
WARNING
• Ample ventilation
The solvents used with the HPLC system are often flammable and toxic.
Therefore, the room where the instrument is installed must be well-ventilated.
• No fire sources used near the instrument
The solvents used with the HPLC are often flammable. Therefore, the use of open flame where the
instrument is installed must be strictly prohibited. Also, do not install in the same room with equipment
that emits or could potentially emit sparks.
• Fire extinguishers permanently available
Have fire extinguishers permanently available in case of fire.
• Protective equipment provided near the instrument
If solvent gets into the eyes or onto the skin, it must be flushed away immediately.
Provide equipment, such as eye wash stations and safety showers, as close to the instrument as
possible.
CAUTION
• Avoid dust or corrosive gas
To ensure a long service life of the instrument and preserve its performance levels, avoid installing it in
places subject to large amounts of dust or corrosive gas.
• Keep away from equipment generating strong magnetic fields
To ensure proper operation, do not install the instrument in places subject to strong magnetic fields.
If the power supply line is subject to high electrical noise, install a surge protector.
• Install the instrument in the location that satisfies the following conditions to preserve the
performance:
• room temperature is between 4 and 35°C, with minimal temperature variation through a day.
• air currents from heating or air conditioning equipment are not directed on the instrument.
• sunlight does not shine directly on the instrument.
• there is no vibration.
• humidity stays within 20 - 85%.
• place without condensation
9-2 LC-20AT
9.1 Installation
CAUTION
• The weight of this instrument is 11kg. During installation, consider the entire weight combined
with other LC components.
The lab table on which this instrument is installed should be strong enough to support the total weight
of the LC system. It should be level, stable and have depth of at least 600mm.
If these precautions are not followed, the instrument could tip over or fall off the table.
• Keep at least 100mm between the rear of the instrument and the wall.
This allows for sufficient air circulation to provide cooling and prevent the instrument from overheating
and impairing the performance.
Typical system configurations and required installation spaces are shown in the figures below.
(Unit : mm)
Fig. 9.1
LC-20AT 9-3
9. Technical Information
9.1.2 Installation
Remove the Shipping Screws
In order to prevent shock during transportation, the instrument is fixed with the shipping screws.
Remove these screws prior to installation.
NOTE
When the instrument is used without removing the shipping screws, system makes noise because of
vibration.
Installation
The instrument is designed for stacking with other
Shimadzu HPLC components.
"9.4 Introduction to HPLC System" P. 9-44
CAUTION
When the LC-20A series components are stacked on
each other, the clearance between the components
is only 5 mm.
Use caution to avoid pinching your fingers between
the components.
Fig. 9.2
Stacking Brackets
The use of commercially available stacking brackets is
recommended. These brackets limit the possibility of the
instrument falling off the lab table during an earthquake
or the like. Various grades of stacking brackets are
Safety
available.
bracket
Fig. 9.3
9-4 LC-20AT
9.1 Installation
228-45001-31 AC100-120V(100-120V∼)
228-45001-38 AC220-240V(220-240V∼)
Verify that the power outlet to be used for connection has sufficient capacity. If capacity is insufficient, a power
outage or voltage drop can occur, affecting not only this instrument, but other instruments connected to the
same power supply.
WARNING
Handle the power cord with care, and observe the
following precautions to avoid cord damage, fire,
electric shock or instrument malfunction.
• Do not place heavy objects on the cord.
• Keep hot items away from the cord.
• Do not modify the cord.
• Do not bend the cord excessively or pull on it.
• To unplug the instrument, pull the plug itself, NOT
the cord.
If the cord is damaged, replace it immediately.
Fig. 9.4 9
CAUTION Power cord connector
Before plugging in the instrument, make sure that the
power switch is OFF.
Plug side
Fig. 9.5
LC-20AT 9-5
9. Technical Information
Grounding
WARNING
The three-line type power cable provided as an
accessory includes the grounding wire.
Be sure to ground through this cable in order to
prevent electrical shock and to ensure stable
operation of the instrument.
Cutting Tubings
Cut provided tubing to the proper lengths for installation.
NOTE
Fig. 9.6
9-6 LC-20AT
9.1 Installation
Fig. 9.7
CAUTION
• Make the cut surface at right angle. Otherwise,
dead volume will be created and may cause
chromatographic peak broadening.
• Make sure that the inner diameter of the tubing is
not deformed. Otherwise, the tubing may be
clogged.
9
Cut at right
angle
Fig. 9.8
LC-20AT 9-7
9. Technical Information
Connecting Tubings
Tubing
CAUTION Male nut
Fig. 9.9
male nut.
The ferrule will be secured on the tubing.
Fig. 9.10
CAUTION
• Insert the tubing completely into the opening, until
it butts against the end of the opening.
Otherwise, dead volume will be created and may
cause chromatographic peak broadening.
• Do not overtighten the male nut.
Otherwise, the threads will be damaged.
NOTE
Male nut
• For an SUS male nut:
Use the open-end wrench (provided) to tighten and
loosen the nut.
If the nut is to be connected to a union or other part
that is not secured, use a second wrench to secure the
union.
• For a resin male nut:
Tighten and loosen the nut by hand. Coupling
Fig. 9.11
9-8 LC-20AT
9.1 Installation
Moves
Fig. 9.12
Protective Plugs
Inlets and outlets of the instrument are fitted with protective plugs (bushings, stop plugs, caps and similar
items) to keep out dirt and dust during shipment.
When the instrument is not connected to other LC system components, replace the protective plugs.
Otherwise, dirt and dust may cause clogging of the instrument.
Keep the plugs, and replace them when the instrument will be left not in use for a long time.
NOTE
LC-20AT 9-9
9. Technical Information
9.1.5 Plumbing
CAUTION
• Before plumbing, turn OFF the power supply to all the system components and unplug them.
• For plumbing, use the appropriate parts listed in "1.3Component Parts".
• Connect only the tubing described in the instructions.
Otherwise, injury or equipment failure may cause.
NOTE
For details on plumbing when the optional degasser, low-pressure gradient unit and/or reservoir switching
valve is/are used, see the instruction manual(s) for these options.
Reservoir preparation
P.9-11
9-10 LC-20AT
9.1 Installation
WARNING
Do not use cracked or damaged bottles.
φ29
The reservoir should be made of glass and have a Pitch of screw 4.23
φ26.5±0.3
capacity of at least 500mL. Width of screw 2.0-2.2
φ24.5±0.3 No. of turns of screw 1·1/2 turns
φ17.5±0.4
2.5-3.0
CAUTION
The waste container must be positioned lower than
11
the instrument (for example, on the floor). If it is
positioned higher than the instrument, liquid will not
4
drain, and will leak from the connections. φ25.4
Fig. 9.14
Front Cover Removal
The front cover must be removed in order to perform the
plumbing.
9
Front cover
Fig. 9.15
NOTE
Fig. 9.16
LC-20AT 9-11
9. Technical Information
NOTE
Prevent any dirt from adhering to the filter portion of the filter when installing it.
Dirt on the filter could cause clogging and flow rate fluctuations.
NOTE
In order to ensure stable analysis at all times, the mobile phase should be degassed.
For details of the degassers, "1.4 Optional Parts" P. 1-5.
Bushing
suction port.
Suction tubing
Fig. 9.17
NOTE Lid
Fig. 9.18
9-12 LC-20AT
9.1 Installation
NOTE
Pump inlet
Shipping bushing
Fig. 9.19
NOTE
Bushing
Fig. 9.20
Fig. 9.21
LC-20AT 9-13
9. Technical Information
Drain tubing
3 Put the other end of the drain tubing into the waste connection port
container.
NOTE
Waste container
Fig. 9.22
9-14 LC-20AT
9.1 Installation
Bottom of Instrument
Fig. 9.24
Silicone tubing
Fig. 9.25
L-Joint
NOTE
Fig. 9.26
To ensure a smooth flow of liquid, insert the silicone
tubing into the container with its tip pointing downward.
LC-20AT 9-15
9. Technical Information
NOTE
Leaks from the column oven are drained separately (See column oven Instruction Manual.). If any
components are installed on top of the column oven, carry out the same procedure as in "Installation on top of
the column oven" on next page. Also when the bottom unit has no leakage hole ("Fig. 9.28"), carry out the
same procedure as in "Installation on top of the column oven" on next page.
45°
Drain adapter
Fig. 9.27
Drain outlet
Upper unit
Drain adapter
Fig. 9.28
9-16 LC-20AT
9.1 Installation
NOTE
When the bottom unit has no leakage hole ("Fig. 9.28"), carry out the same procedure described bellow.
Fig. 9.29
Silicone tubing
• To ensure a smooth flow of liquid, insert the silicone tubing into the container with its tip pointing downward.
• Set the silicone tubing so that its edge does not touch the liquid surface in the waste container. When the tip
of tubing touches into the liquid of waste container, the drainage becomes difficult to flow.
LC-20AT 9-17
9. Technical Information
Front cover
Fig. 9.31
9-18 LC-20AT
9.1 Installation
Injector installation
plate
Fig. 9.32
Fig. 9.33
Fig. 9.34
LC-20AT 9-19
9. Technical Information
Column inlet
Fig. 9.35
NOTE
For connecting ports 1 to 6 of the manual injector, use the male nuts (with long bushing) and ferrules, provided
as manual injector standard accessories.
Sample loop
Fig. 9.36
9-20 LC-20AT
9.1 Installation
4 Route the other ends of the waste liquid tubing Waste liquid
tubing
through the tubing opening and into the vial.
Tubing hole
Vial
Fig. 9.37
NOTE
Waste liquid tubings
To prevent liquid from flowing out due to the siphon
effect, position the ends of the waste liquid tubing level
with the needle port.
Height of
Needle port
Fig. 9.38
Connect to
port 2
Fig. 9.39
LC-20AT 9-21
9. Technical Information
NOTE
Stop plug
Column inlet
Fig. 9.40
Column
inlet
SUS tubing
Fig. 9.41
9-22 LC-20AT
9.1 Installation
Detector
Cell inlet
Waste container
Unscrew and
remove stop plug.
LC-20AT 9-23
9. Technical Information
Preparation of Mixer
The mixer may be plumbed to accommodate three volumes; 0.5mL, 1.7mL and 2.6mL.
Select the appropriate capacity for the application. (The standard plumbing, installed at delivery, is for 2.6mL
capacity.)
Pre-mixer
Mixer
NOTE
Attach caps
Attach cap
to Autosampler /
Manual Injector
to Autosampler /
to Autosampler / Manual Injector
Manual Injector
Attach cap
Fig. 9.44
9-24 LC-20AT
9.1 Installation
Fig. 9.45
Fig. 9.46
Mixer
2 Install the mixer on this instrument using the
unscrewed screws.
Fig. 9.47
LC-20AT 9-25
9. Technical Information
Inlet A
Inlet B
Cap
Premixer
Fig. 9.48
Stop plug
Fig. 9.49
the following:
• Pump outlet and premixer inlet
• Mixer outlet and injector
9-26 LC-20AT
9.1 Installation
NOTE
to Port 2
Fig. 9.51
LC-20AT 9-27
9. Technical Information
NOTE
Pump outlet
SUS tubing
Pump outlet
Inlet A
from degasser or reservoir Inlet B
SUS tubing Inlet C (Stop plug)
switching valve
Fig. 9.53
9-28 LC-20AT
9.1 Installation
NOTE
from Mixer
outlet
Fig. 9.54
Mixer Cover Installation
9
Mixer cover
Fig. 9.55
LC-20AT 9-29
9. Technical Information
9.1.9 Wiring
WARNING
• Before performing wiring, turn OFF all components and unplug the power cables.
• Do not use any other than specified cables for wiring.
• Do not perform any other than the indicated wiring operations.
Failure to observe the above cautions could result in fire, electric shock or instrument malfunction.
Connectors
• [REMOTE] connector is used in cases below.
for connection to system controller
for connecting dual pumps together without using a system controller in a high-pressure gradient system
• [DGU/SOL. V] connector ............... For connection of optional reservoir selection valve/degasser
• [PUMP PRESS] connector ............ For connection to Chromatopac (or other integrator) or recorder.
Outputs voltage proportional to pressure.
• External input/output terminals ...... For connection to external equipment. For connection instructions, see
"5.7 Connection to External Input/Output Terminals" P. 5-58.
• [DGU PRESS] connector............... For connection to degasser to enter the pressure signal of vacuum
pump inside of degasser.
Use the connectors above needed for the system. Connection instructions are provided on the following
pages.
Fig. 9.56
9-30 LC-20AT
9.1 Installation
CAUTION
The caps on the [REMOTE] connectors prevent dirt
or dust from getting into the connector.
If a [REMOTE] connector is not used, leave the cap
on it to prevent dirt or dust from interfering with
communication.
When a cap is removed, keep it in a safe place for
Cap [REMOTE] connector
future use.
Fig. 9.57
CAUTION
Gray
• Make sure there is no dirt or dust on the plug.
Dirt or dust on the plug will get inside the
[REMOTE] connector.
Plug
• Be careful not to insert the plug across two
different channels. Insert until it clicks.
Failure to follow these precautions above could
result in malfunction or communication problems.
9
Fig. 9.58
CAUTION
• Do not bend the optical cable less than 35 mm in
radius.
• When inserting and removing the plug, grip the
plug itself, not the cable.
• Do not bend the cable where it joins the plug.
Failure to follow these above precautions could
result in damage to the plug or a broken wire in the
cable.
Fig. 9.59
LC-20AT 9-31
9. Technical Information
NOTE
Optical cable
Channels between 3 and 8 of the system controller
[REMOTE] connector are typically used for this purpose.
[REMOTE] connector
Fig. 9.60
0.000 ADRS 3
Input 1 - 12,85
9-32 LC-20AT
9.1 Installation
WARNING
• Only Shimadzu service representative is responsible for installation of CBM-20Alite.
There is a risk of electric shock or short.
• By the installation of the instrument, remove the power plug of the instrument from the power outlet.
There is a risk of electric shock or short.
NOTE 9
It is not necessary to connect the optical cable, because 0.000 ADRS 5
the instrument controlled by the internal-connected Fixed by Int-CBM
CBM-20Alite.
In case of internal connection, the link address [ADRS] is
the fixed value [5].
"Setting Link Address [ADRS]" P. 5-21
LC-20AT 9-33
9. Technical Information
NOTE
[REMOTE] connector
0.000 ADRS 85
Input 1 - 12,85
9-34 LC-20AT
9.1 Installation
[DGU/SOL.V] connector
CAUTION
Make sure that the power switch of the instrument is
OFF before connecting or disconnecting the special
cable.
Otherwise, the instrument could be damaged.
FCV-11AL
0.000 SV ™
Select Po rt No.
LC-20AT 9-35
9. Technical Information
Signal cable
Signal cable
(accessory for Chromatopac) Black (−)
White (+)
Fig. 9.63
9-36 LC-20AT
9.1 Installation
WARNING
• Only Shimadzu service representative is responsible for installation.
There is a risk of electric shock or short.
• By the installation of the instrument, remove the power plug of the instrument from the power outlet.
There is a risk of electric shock or short.
to IN Rinsing pump
9
Rinse tubing
Fig. 9.64
1 Prepare the small bottle provided with automatic rinsing pump filled with distilled water about four-fifths
and close the cap.
2 Cut the rinse tubing provided with the instrument into two lengths, appropriate for connecting (a) the
instrument and the rinsing pump, and (b) the instrument and the bottle. Then connect plumbing as in
the figure.
NOTE
Do not pull on the rinse tubing connected to the pump heads or any of the rinse pump tubing, or you could
disconnect the tubing.
LC-20AT 9-37
9. Technical Information
Fig. 9.65
9-38 LC-20AT
9.2 Specifications
9.2 Specifications
Item Specification
Commands for flow rate, pressure, [EVENT] functions, [LOOP] (for program
Time Program
Weight 11kg
LC-20AT 9-39
9. Technical Information
Maximum program steps Max. 400 steps per program. 20 programs files
± 1%
Concentration accuracy
(for binary gradient with water/caffeine solution, at 0.1-3mL/min, 1.0-40MPa)
Maximum program steps Max. 320 steps per program. 10 program files.
± 1%
Concentration accuracy
(for binary gradient with water/caffeine solution, at 0.1-3mL/min, 1.0-40MPa)
9-40 LC-20AT
9.2 Specifications
Maximum program steps Max. 400 steps per program. 20 program files.
± 1%
Concentration accuracy
(for binary gradient with water/caffeine solution, at 0.1-2mL/min, 1-20MPa)
0.001-2mL/min (rate over 2mL/min possible, but will shorten life span of low-
Flow rates possible
pressure gradient unit’s solvent valves)
Maximum program steps Max. 320 steps per program. 10 program files
9
± 1%
Concentration accuracy
(for binary gradient with water/caffeine solution, at 0.1-2mL/min, 1-20MPa)
0.001-2mL/min (rate over 2mL/min possible, but will shorten life span of low-
Flow rate possible
pressure gradient unit’s solvent valves)
LC-20AT 9-41
9. Technical Information
9-42 LC-20AT
9.3 Maintenance Parts
Electrical Parts
Part Part No. Remark
Left Panel 20AT ASSY 228-43200-92 Display panel and keypad assembly
Spacer S 228-42700 2
Frit 228-32744 2
LC-20AT 9-43
9. Technical Information
1 Reservoir bottle
2 Degasser
1
3 Pump unit
Manual
4 injector
5 Column
4 5
2
6 Detector
7
7 Waste container 3
1 Mobile phase is drawn out of the reservoir bottle and pumped through the tubing by the pump.
2 The degasser removes dissolved air from the mobile phase, preventing air bubbles and consequent rise,
drift or other baseline irregularities caused by dissolved air.
3 The pump sends the mobile phase through the manual injector, column and detector, in that order, and
finally into the waste container.
4 Samples are injected into the system by the manual injector, with a syringe.
5 In the column, the components are separated by means of the mutual interactions of the mobile phase and
the column packing (stationary phase).
6 The detector detects the components eluted from the column, and sends the signal data to a Chromatopac
or PC.
7 Mobile phase from the detector drains into the waste container.
9-44 LC-20AT
9.4 Introduction to HPLC System
1 Reservoir bottle 1
2 Degasser
2
Low-pressure
3 gradient unit∗
4 8
3
4 Pump unit
5 Mixer
6 Autosampler
7 Column 7
8 Detector 9
6
9 Waste container 5
9
1 Mobile phase is drawn out of the reservoir bottles and pumped through the tubing by the pump.
2 The degasser removes dissolved air from the mobile phase, preventing air bubbles and consequent rise,
drift or other baseline irregularities caused by dissolved air.
3 The low-pressure gradient unit mixes up to 4 mobile phases that have been degassed by the degasser.
(∗This item is necessary for a low-pressure gradient system.)
4 The pump sends the mobile phase through the autosampler, column and detector, in that order, and finally
into the waste container.
5 The mixer enhances the mixing efficiency of the mobile phases. This item is required for low or high-
pressure gradient system.
6 The autosampler automatically injects the sample into the flow lines. By adding a rack changer, it is possible
to automatically change the autosampler racks.
7 In the column, the components are separated by means of the mutual interactions of the mobile phase and
the column packing (stationary phase).
8 The detector detects the components separated in the column, and sends the signal data to a Chromatopac
or PC.
9 Mobile phase from the detector drains into the waste container.
LC-20AT 9-45
9. Technical Information
1 Reservoir bottle
1
2 Degasser
2
3 Pump unit
9
4 Mixer 3
5 Autosampler
7
3
6 Column
7 Detector
8 Waste container
8
5
1 Mobile phase is drawn out of the reservoir bottles and pumped through the tubing by the pump.
2 The degasser removes dissolved air from the mobile phase, preventing air bubbles and consequent rise,
drift or other baseline irregularities caused by dissolved air.
3 The pump sends the mobile phase through the autosampler, column and detector, in that order, and finally
into the waste container.
4 The mixer enhances mixing efficiency of the mobile phases.
5 The autosampler automatically injects the sample into the flow lines. By adding a rack changer, it is possible
to automatically change the autosampler racks.
6 In the column, the components are separated by means of the mutual interactions of the mobile phase and
the column packing (stationary phase).
7 The detector detects the components eluted from the column, and send the signal data to a Chromatopac
or PC.
8 Mobile phase from the detector drains into the waste container.
9 The CBM-20A system controller can control a maximum of 8 LC components (12 LC components as an option)
including a maximum of 4 pump units.
9-46 LC-20AT
9.5 Mobile Phase Characteristics
LC-20AT 9-47
9. Technical Information
(7)
(1) Water
Solvent (3) Boiling Solubility (8) (9)
(*)η<.5cP, > 45°C (2) UV (4) Point Viscosity (5) (6) %Win20°C Dielectric p'+
(**)η<.5cP,< 45°C Source Cutoff R.I.25° (°C) (cP,25°C) p' e°a Solvent Constant e20 0.25e
34 t-Butanol 1.385 82 3.6 4.1 0.7 miscible 12.5
35 n-Butanol LC 210 1.397 118 2.6 3.9 0.7 20.1 17.5 8.3
36 n-Propanol LC 240 1.385 97 1.9 4.0 0.82 miscible 20.3
37 Tetrahydrofuran(*) LC 212 1.405 66 0.46 4.0 0.57 miscible 7.6
38 Propylamine(*) 1.385 48 0.35 4.2 miscible 5.3
39 Ethylacetate(*) LC 256 1.370 77 0.43 4.4 0.58 8.8 6.0 5.8
40 i-Propanol LC 205 1.384 82 1.9 3.9 0.82 miscible 20.3
41 Chloroform(*) LC 245 1.443 61 0.53 4.1 0.40 0.072 4.8 5.6
42 Acetophenone 1.532 202 1.64 4.8 17.4 8.7
43 Methylethyl LC 329 1.376 80 0.38 4.7 0.51 23.4 18.3 9.1
44 Cyclohexanone 1.450 156 2.0 4.7 18.3 9.1
45 Nitrobenzene 1.550 211 1.8 4.4 34.8 13.2
46 Benzonitrile 1.536 191 1.2 4.8 25.2 10.9
47 Dioxane LC 215 1.420 101 1.2 4.8 miscible 2.2
48 Tetramethyl urea LC 265 1.449 175 6.0 0.56 23.0 10.7
49 Quinoline 1.625 237 3.4 5.0 9.0 7.4
50 Pyridine 1.507 115 0.88 5.3 miscible 12.4
51 Nitroethane 380 1.390 114 0.64 5.2 0.9
Acetone(*) 1.356 56 0.30 5.1
52 Benzyl alcohol LC 330 0.71 miscible 13.1 8.8
1.538 205 5.5 5.7
Tetramethyl
53 6.1 0.6
guanidine
54 Methoxyethanol LC 210 1.400 125 1.60 5.5 miscible 19.9
Tris(cyanoethoxy)
55 propane GC 6.6 0.56
Propylene
56 LC 6.1
carbonate
57 Ethanol LC 210 1.359 78 10.8 4.3 miscible 24.6
58 Oxydipropionitrile GC 6.8
59 Aniline 1.584 184 3.77 6.3 6.9 8.1
60 Acetic acid 1.370 118 1.1 6.0 miscible 6.2
61 Acetonitrile(*) LC 190 1.341 82 0.34 5.8 miscible 37.5
62 N,N-dimethylaceta-mide LC 268 1.436 166 0.78 6.5 0.88 37.8
63 Dimethylformamide LC 268 1.428 153 0.80 6.4 36.7
64 Dimethylsulfoxide LC 268 1.477 189 2.00 7.2 0.62 miscible 4.7
65 N-methyl-2-pyrolidone LC 285 1.468 202 1.67 6.7 32
Hexamethyl phosphoric
66 acid triamide 1.457 233 3 7.4 0.65 30
9-48 LC-20AT
9.5 Mobile Phase Characteristics
(7)
(1) Water
Solvent (3) Boiling Solubility (8) (9)
(*)η<.5cP, > 45°C (2) UV (4) Point Viscosity (5) (6) %Win20°C Dielectric p'+
(**)η<.5cP,< 45°C Source Cutoff R.I.25° (°C) (cP,25°C) p' e°a Solvent Constant e20 0.25e
71 Ethylene glycol 1.431 182 16.5 6.9 miscible 37.7
72 Formamide 1.447 210 3.3 9.6 miscible 111
73 Water LC 1.333 100 0.89 10.2 80
(1) An asterisk (*) indicates solvents most suitable for LC, with low boiling points (>45 °C) and low viscosity
(<0.5cP).
Double asterisks (**) indicates solvents with a very low viscosity and boiling point.
(2) “LC” indicates that a grade of solvent specifically for LC is commercially available from companies like the
following:
Burdick & Jackson, Baker Chemical, Mallinkrodt Chemical, Fischer Scientifie, Waters Associate,
Manufacturing Chemists. Inc.
“GC” indicates that a solvent is used as a stationary phase for gas chromatography, and can be
purchased from companies selling GC columns and stationary phases. (These solvents are used as
stationary phase in liquid-to-liquid LC.)
(3) The wavelength below which the solvent becomes opaque.
(4) Refractive index at 25°C.
(5) Polarity parameter of solvent.
(6) Solvent’s strength parameter in relation to liquid-to-solid adsorption in alumina.
(7) Water solubility (%W) at 20°C of solvent used in liquid-to -solid adsorption.
(8) Value at 20°C.
(9) Function consisting of P’ (proportional to solvent strength) plus the dielectric constant, in ion
chromatography.
LC-20AT 9-49
9. Technical Information
9-50 LC-20AT
Index
Index
LC-20AT Index-1
Index
M
R
Maintenance Information Group .... 5-9, 5-28, 5-32
MAINTENANCE LOG ................................... 5-33 R SEAL DELIVERED .................................... 5-33
Maintenance Parts ........................................ 9-42 RANGE .......................................................... 5-23
Maximum Pressure Limit ................................. 4-4 REMOTE Connector .............................. 2-5, 9-31
MEMORY CHECK ........................................ 5-36 Replacement of Check Valves ...................... 8-19
Minimum Pressure Limit .................................. 4-5 Replacement of Drain Valve ASSY ............... 8-25
Mixer ...................................................... 1-6, 9-24 Replacement of Fuse .................................... 8-27
MOBILE PHASE ........................................... 5-31 Replacement of Line Filter ............................ 8-22
Mobile Phase Monitor Group ........ 5-8, 5-27, 5-31 Replacement of Plunger Seal .......................... 8-5
MODE CHANG ............................................. 5-20 Replacement of Plungers .............................. 8-11
MON FLOW .................................................. 5-25 Replacement of Rinse Flow Line Seal .......... 8-16
MON ID ......................................................... 5-25 Replacement of Suction Filter ....................... 8-24
MON REV ..................................................... 5-26 Replacement Parts ........................................ 9-42
MON SV ........................................................ 5-26 Replacing the Mobile Phase .......................... 4-16
MON TIME .................................................... 5-24
Monitor Settings Group ................. 5-7, 5-13, 5-24
S
Index-2 LC-20AT
Index
T
TIME ............................................................. 5-35
Time Program ............................................... 5-45
TOTAL OP TIME ........................................... 5-32
Troubleshooting and Corrective Action ........... 6-2
V
Validation Support Group ............ 5-10, 5-28, 5-35
Validation:Pump .............................................. 7-7
VP Functions .......................................... 5-8, 5-27
W
Wiring ............................................................ 9-30
Z
ZERO ADJ .................................................... 5-20
LC-20AT Index-3
Index
Index-4 LC-20AT