7th Semester
7th Semester
7th Semester
and chemical changes. It involves the study of reactions where electrons are transferred between
species, typically involving redox (reduction-oxidation) reactions. Electrochemistry encompasses a
variety of processes and technologies, such as batteries, fuel cells, electrolysis, corrosion, and
electroplating.
Electrochemical Cells
• Galvanic (Voltaic) Cells: Generate electrical energy from spontaneous redox reactions.
o Anode: The electrode where oxidation occurs (negative terminal).
o Cathode: The electrode where reduction occurs (positive terminal).
• Electrolytic Cells: Use electrical energy to drive non-spontaneous redox reactions.
o Anode: The electrode where oxidation occurs (positive terminal).
o Cathode: The electrode where reduction occurs (negative terminal).
4. Nernst Equation
• Describes the relationship between the cell potential and the concentrations of the
reactants and products.
1. Electrodes
• Working Electrode: The electrode at which the redox reaction of interest occurs.
• Reference Electrode: Maintains a constant potential for comparison (e.g., Standard Hydrogen
Electrode, Silver/Silver Chloride Electrode).
• Auxiliary Electrode: Completes the circuit in three-electrode systems, typically inert.
The indicator electrode is the electrode at which the redox reaction of interest occurs. It responds
to the changes in the concentration of the analyte and provides the necessary signal for
measurement.
1. Metal Electrodes:
o Platinum (Pt): Used for a wide range of redox reactions due to its inertness.
o Gold (Au): Often used in studies involving thiol chemistry and biological molecules.
o Silver (Ag): Common in chloride ion measurements.
2. Carbon Electrodes:
o Glassy Carbon: Known for its wide potential window and chemical stability.
o Graphite: Cost-effective and widely used in various applications.
3. Ion-Selective Electrodes (ISEs):
o pH Electrode: Selectively measures hydrogen ion concentration.
o Fluoride ISE: Measures fluoride ion concentration.
Reference Electrode
The reference electrode provides a stable and known reference potential against which the
potential of the indicator electrode can be measured. It is essential for ensuring accurate and
reproducible electrochemical measurements.
• Principle: Measure the potential (voltage) of an electrochemical cell without drawing significant
current.
• Application: Used to determine the concentration of ions in a solution by measuring the
electrode potential of an ion-selective electrode.
• pH meter
Basic Principle
The potential difference between the indicator electrode and the reference electrode is measured.
This potential difference is related to the concentration of the ion of interest through the Nernst
equation:
Voltammetric Methods
• Principle: Measure the current as a function of an applied potential. The potential is varied, and
the resulting current is measured to provide information about the redox processes of the
analyte.
• Application: Used to study redox properties, determine concentration, and analyze reaction
mechanisms.
• Types:
o Linear Sweep Voltammetry (LSV): Linear variation of potential with time.
o Cyclic Voltammetry (CV): Potential is swept linearly in forward and reverse directions.
o Differential Pulse Voltammetry (DPV): Superimposes a series of pulses on a linear
potential sweep.
o Square Wave Voltammetry (SWV): Uses a square wave modulation on a staircase
potential.
Amperometric Methods
• Principle: Measure the current at a fixed potential over time. The current is proportional to the
concentration of the analyte undergoing redox reactions
Coulometric Methods
• Principle: Measure the total charge passed during the complete electrolysis of an analyte. The
charge is proportional to the amount of analyte present.
Metallic Indicator Electrodes
Metallic indicator electrodes are made of metals that directly participate in the redox reactions of
the analyte or establish equilibrium with the analyte ions.
Membrane indicator electrodes, also known as ion-selective electrodes (ISEs), use a selective
membrane to respond to specific ions in the solution. These electrodes provide a high degree of
selectivity and sensitivity.
Ion-exchange membrane electrodes utilize a membrane that selectively allows the passage of
specific ions while excluding others.
Characteristics:
• Composition: Membranes are typically made from ion-exchange resins or polymers doped with
ion-exchange materials.
• Fluoride Ion-Selective Electrode:
Characteristics:
4. Bio-Membrane Electrodes
Potentiometric titrations are a type of titration method in which the potential difference
(voltage) between two electrodes is measured as a reagent is added to a solution. The method is used to
determine the concentration of an analyte by observing changes in the electrical potential of the
solution as the titration progresses.
1. Acid-Base Titrations:
o Principle: Measures the change in pH as an acid or base is added.
o Example: Titration of hydrochloric acid (HCl) with sodium hydroxide (NaOH).
2. Redox Titrations:
o Principle: Measures the change in potential as an oxidizing or reducing agent is added.
o Example: Titration of ferrous ion (Fe²⁺) with potassium permanganate (KMnO₄).
3. Complexometric Titrations:
o Principle: Measures the change in potential as a complexing agent is added.
o Application: Determine the concentration of metal ions in solution.
o Example: Titration of calcium ions (Ca²⁺) with EDTA (ethylenediaminetetraacetic acid).
1. Electrochemical Cell:
o Working Electrode: The electrode where the electrochemical reaction takes place. It can
be a planar, cylindrical, or other forms depending on the design of the cell.
o Counter Electrode: Completes the circuit by allowing the flow of current. Often made
from inert materials such as platinum or graphite.
o Reference Electrode: Provides a stable reference potential. Common reference
electrodes include the Saturated Calomel Electrode (SCE) or Silver/Silver Chloride
Electrode (Ag/AgCl).
2. Power Supply:
o Provides a constant current or potential to drive the electrochemical reaction. In
potentiostatic coulometry, a constant potential is applied, while in amperostatic
coulometry, a constant current is used.
3. Current Measuring Device:
o Measures the current flowing through the electrochemical cell. This is crucial for
determining the total charge passed.
o Devices include ammeters or more advanced instruments that can integrate current
over time to calculate total charge.
4. Integration Device:
o Calculates the total charge (Q) by integrating the current over time.
o This charge is used to determine the quantity of the analyte based on the
electrochemical reaction.
5. Data Acquisition System:
o Records and processes data from the current measuring device and integration device.
o Often includes software for analyzing and interpreting the results.
Types of Coulometry
1. Potentiostatic Coulometry:
o Principle: A constant potential is applied to the electrochemical cell, and the resulting
current is measured. The total charge is integrated to determine the quantity of analyte.
2. Amperostatic Coulometry:
o Principle: A constant current is applied to the electrochemical cell, and the potential
change is measured. The total charge is calculated based on the current and time.
In this process, the analyte (the element or compound of interest) is deposited electrolytically on an
electrode. The mass of the deposited analyte is subsequently measured, enabling the determination of the
analyte's concentration in the initial solution
Instrumentation of Electrogravimetry:
A basic electrogravimetry setup typically includes an electrolytic cell, an ammeter, a voltage source, an
electrode, and a balance
The analyte is added to an electrolytic solution contained in the cell, and the electrode is held in the
solution. An external voltage is applied to the electrode, causing electrolysis, i.e., the reduction of the
oxidized analyte at the cathode (negative electrode) and the oxidation of other substances at the anode
(positive electrode)
When the desired amount of analyte has been deposited on the cathode, the cell is disconnected from the power
source, and the cathode is weighed to determine the mass of the deposited analyte
This mass, combined with the known cell volume, initial analyte concentration, and Faraday's laws of
electrolysis, can be used to calculate the analyte's concentration in the original solution
Consequences Of Electrogravimetry
Electrogravimetry, while a powerful technique, can be influenced by several factors that impact its
accuracy. Ohmic drop, activation overpotential, concentration polarization, and gas polarization are key
issues to address:
• Ohmic drop: Ohmic drop, also known as iR drop, is the voltage drop due to the electrical resistance in the cell
components [4,9]. It can cause inaccuracies in potential measurements.
• Activation overpotential: Activation overpotential is the additional potential required to overcome the
activation energy barrier for the electrochemical reaction [5]. It leads to a decrease in the reaction rate.
• Concentration polarization: This phenomenon occurs due to the accumulation of reaction products near the
electrode surface, hindering further reaction [5]. It causes a decrease in the reaction rate and an increase in the
overpotential.
• Gas polarization: Gas polarization refers to the formation of gas bubbles on the electrode surface,
which can interfere with the electrochemical reaction. Gas polarization is a hindrance to the
electrochemical reaction caused by the slow rate at which gaseous reactants or products can diffuse through the
electrolyte [2]. It leads to a decrease in the reaction rate and an increase in the overpotential.
Chromatographic Processes
1. Sample Introduction: The sample mixture is introduced into the chromatographic system.
2. Elution: The mobile phase carries the sample through the stationary phase.
3. Separation: Compounds in the sample interact differently with the stationary phase, leading to
separation.
4. Detection: The separated compounds are detected and quantified using various detectors (e.g.,
UV-Vis, mass spectrometry).
Rate Theory of Chromatography
The rate theory of chromatography describes the movement of solute molecules through the
chromatographic column and the factors affecting their separation. It focuses on the kinetics of
solute transfer between the mobile phase and the stationary phase and the band broadening that
occurs during this process.
The Van Deemter equation is a key component of the rate theory of chromatography. It describes
the relationship between the height equivalent to a theoretical plate (HETP or simply H), the
linear velocity of the mobile phase (u), and various factors contributing to band broadening. The
equation is given by:
H = A + B/u + C⋅u
Where:
General Principle
In gas-liquid chromatography, the mobile phase is a gas (often an inert gas like helium or
nitrogen), and the stationary phase is a liquid coated on a solid support or the inner walls of a
capillary column. The basic principle involves the differential partitioning of analytes between
the mobile phase and the stationary phase.
Sample Preparation/Derivatization
1. Sample Introduction:
o Direct Injection: For liquid samples that are directly injectable.
o Headspace Analysis: For volatile compounds in complex matrices (e.g., blood, food).
2. Derivatization: Enhancing the volatility or detectability of analytes that are not naturally
volatile or detectable.
o Silylation: Adding silyl groups to hydroxyl or amino groups.
o Acylation: Adding acyl groups to enhance volatility.
o Methylation: Converting carboxylic acids to their methyl esters.
Separation Process
1. Sample Injection:
o The sample is vaporized in the injector and carried into the column by the mobile phase.
o Injection modes include split, splitless, on-column, and programmed temperature
vaporizing (PTV).
2. Column:
o The column is the heart of the GC system. It can be packed or capillary.
o Packed Columns: Filled with solid support coated with the liquid stationary phase.
o Capillary Columns: Narrow, with stationary phase coated directly on the inner wall.
3. Oven Temperature Programming:
o The column is housed in an oven where the temperature can be precisely controlled.
o Isothermal (constant temperature) or temperature-programmed (ramped) methods can
be used to improve separation.
4. Detection:
o As analytes elute from the column, they are detected and quantified.
o Common detectors include Flame Ionization Detector (FID), Thermal Conductivity
Detector (TCD), Electron Capture Detector (ECD), and Mass Spectrometer (MS).
Instrumental Aspects
1. Carrier Gas:
o Inert Gases: Helium, nitrogen, or hydrogen.
o Flow Control: Precise control of flow rate is essential for reproducibility.
2. Injector:
o Vaporizing Injectors: For liquid samples, often with split or splitless options.
o Headspace Injectors: For volatile analytes in complex matrices.
3. Columns:
o Packed Columns: Suitable for high sample capacity but lower efficiency.
o Capillary Columns: Higher resolution, commonly used in modern GC.
4. Oven:
o Temperature Control: Crucial for reproducible retention times.
o Temperature Programming: Enhances separation of analytes with a wide boiling point
range.
5. Detectors:
o FID (Flame Ionization Detector): Universal for organic compounds, sensitive, and widely
used.
o TCD (Thermal Conductivity Detector): Measures changes in thermal conductivity, less
sensitive but universal.
o ECD (Electron Capture Detector): Highly sensitive for halogenated compounds.
o MS: Provides structural information, highly sensitive and specific. he MS detects
compounds by ionizing them and measuring the mass-to-charge ratio (m/z) of the
resulting ions.
Applications
1. Environmental Analysis:
o Monitoring pollutants and volatile organic compounds (VOCs) in air, water, and soil.
2. Food and Beverage Analysis:
o Identifying and quantifying flavors, fragrances, and contaminants.
3. Clinical and Pharmaceutical Analysis:
o Analyzing biological samples (e.g., blood, urine) for drugs, metabolites, and biomarkers.
o Quality control of pharmaceuticals.
4. Petrochemical Industry:
o Analyzing hydrocarbons, fuels, and lubricants.
5. Forensic Science:
o Analyzing drugs, toxins, and residues in forensic samples.
6. Industrial Applications:
o Quality control of industrial products and monitoring process streams.
High-Performance Liquid Chromatography (HPLC) is a powerful analytical
technique used to separate, identify, and quantify components in liquid samples. HPLC operates
on the principle of liquid-liquid or liquid-solid chromatography where the sample is dissolved in
a liquid (the mobile phase) and passed through a column packed with a solid adsorbent (the
stationary phase). Components in the sample interact differently with the stationary phase,
leading to separation.
• Mobile Phase: A liquid solvent or mixture of solvents that carries the sample through the
column.
• Stationary Phase: A solid material, often a packed bed of particles or a coated surface
within the column.
Sample Preparation
1. Filtration: Removing particulates from the sample to prevent clogging the column.
2. Dilution: Adjusting the sample concentration to fall within the detection range.
3. pH Adjustment: Adjusting the pH to enhance analyte stability and optimize interactions with the
stationary phase.
4. Extraction and Cleanup: Removing interfering substances using techniques like solid-phase
extraction (SPE) or liquid-liquid extraction (LLE).
5. Derivatization: Enhancing the detectability of certain analytes by chemically modifying them.
Separation Process
Principle:
• The stationary phase is polar (e.g., silica), and the mobile phase is non-polar (e.g., hexane).
Separation Mechanism:
• Polar compounds interact more strongly with the polar stationary phase and elute more slowly.
• Non-polar compounds elute faster as they have weaker interactions with the stationary phase.
Mobile Phase:
• Typically consists of non-polar solvents like hexane, with possible additives to adjust polarity.
Principle:
• The stationary phase is non-polar (e.g., C18-bonded silica), and the mobile phase is polar (e.g.,
water with organic modifiers like methanol or acetonitrile).
Separation Mechanism:
• Non-polar compounds interact more strongly with the non-polar stationary phase and elute
more slowly.
• Polar compounds elute faster as they have weaker interactions with the stationary phase.
Mobile Phase:
• Typically consists of water and organic solvents like methanol, acetonitrile, or tetrahydrofuran,
often with buffer additives.
Instrumentation
1. Solvent Reservoirs:
o Contain the mobile phase solvents.
2. Pump:
o Delivers the mobile phase at a high pressure (up to 6000 psi) and at a precise flow rate.
3. Injector:
o Introduces the sample into the mobile phase stream, typically through an autosampler
or manual injector.
4. Column:
o Packed with the stationary phase, where the separation of compounds occurs. Columns
vary in length, diameter, and particle size of the packing material.
5. Detector:
o Detects the separated compounds as they elute from the column. Common detectors
include UV-Vis absorbance, fluorescence, refractive index, and mass spectrometry.
6. Data System:
o Collects, processes, and analyzes the data, often through dedicated software.
Method Development
Applications
1. Pharmaceutical Analysis:
o Quality control of active pharmaceutical ingredients (APIs), impurities, and degradation
products.
o Bioanalytical assays for drug metabolism and pharmacokinetics.
2. Environmental Monitoring:
o Detection of pollutants, pesticides, and contaminants in water, soil, and air.
3. Food and Beverage Testing:
o Analysis of additives, preservatives, contaminants, and nutritional components.
4. Clinical and Biomedical Research:
o Quantification of biomolecules (e.g., proteins, peptides, metabolites) in biological
samples.
5. Chemical and Petrochemical Industries:
o Analysis of raw materials, intermediates, and final products.
6. Forensic Science:
o Detection of drugs, toxins, and other substances in forensic samples.
EOF is the bulk flow of liquid through the capillary, driven by the applied electric field. It occurs
due to the interaction between the electric field and the charged walls of the capillary. The
direction and magnitude of EOF can influence the migration of analytes.
In CE, the overall velocity of an ion is the sum of its electrophoretic mobility and the electro-
osmotic flow. Separation is achieved because different ions have different electrophoretic
mobilities, causing them to migrate at different rates and thus separate over time.
Instrumentation
1. Capillary:
o Typically made of fused silica, with an inner diameter of 25-100 µm and lengths of 20-
100 cm.
o Filled with an electrolyte buffer.
2. Power Supply:
o Provides a high voltage (up to 30 kV) to create the electric field.
3. Buffer Reservoirs:
o Contain the electrolyte solution at both ends of the capillary.
4. Injection System:
o Introduces the sample into the capillary. Methods include electrokinetic injection and
pressure injection.
5. Detection System:
o Detects separated analytes as they exit the capillary. Common detectors include UV-Vis
absorbance, fluorescence, and mass spectrometry.
6. Data Acquisition System:
o Collects and analyzes data from the detector.
Modes of Operation
1. Pharmaceutical Analysis:
o Quality control, impurity profiling, and enantiomeric separations.
2. Biotechnology:
o Analysis of proteins, peptides, nucleic acids, and other biomolecules.
3. Environmental Monitoring:
o Detection of pollutants and contaminants in water, soil, and air samples.
4. Clinical Diagnostics:
o Analysis of metabolites, drugs, and biomolecules in biological fluids.
5. Food and Beverage Testing:
o Detection of additives, contaminants, and nutritional components.
6. Forensic Science:
o Analysis of drugs, toxins, and other substances in forensic samples.
AAS is based on the principle that free atoms in the ground state can absorb light of a specific
wavelength. When a sample containing metal atoms is atomized, these atoms absorb light from a
source, resulting in a decrease in the intensity of the light beam. The amount of light absorbed is
proportional to the concentration of the metal atoms in the sample.
According to Beer's Law, the absorption (A) of light by a sample is directly proportional to the
concentration (C) of the absorbing species, the path length (l) of the sample, and the molar
absorptivity (ε):
A=εlC
In AAS, this relationship allows for the quantification of metal ions in a sample by measuring the
absorbance at a specific wavelength.
Quantitative Methodology
1. Calibration Curve: A series of standard solutions with known concentrations are prepared. The
absorbance of each standard is measured, and a calibration curve (absorbance vs.
concentration) is plotted.
2. Sample Measurement: The absorbance of the sample is measured.
3. Concentration Determination: The concentration of the metal in the sample is determined by
comparing the sample absorbance to the calibration curve.
Instrumentation for Atomic Absorption Spectrophotometry
Radiation Sources
• Hollow Cathode Lamps (HCL): Most common source, producing narrow lines of specific
wavelengths for different elements.
• Electrodeless Discharge Lamps (EDL): Provide higher intensity light for some elements.
Atomizers
• Flame Atomizers: Sample is aspirated into a flame, where it is vaporized and atomized.
• Graphite Furnaces: Sample is placed in a graphite tube and subjected to high temperatures to
achieve atomization.
• Electrothermal Atomizers: Use electrical heating to atomize the sample.
Wavelength Selectors
• Monochromators: Used to isolate the specific wavelength of light absorbed by the analyte.
• Filters: Sometimes used to select a broader range of wavelengths.
Detectors
• Photomultiplier Tubes (PMT): Convert the light signal into an electrical signal.
• Charge-Coupled Devices (CCD): Used for multi-element detection.
• Deuterium Lamp Correction: A secondary light source to correct for background absorption.
• Smith-Hieftje Method: Uses modulation of the lamp current to differentiate between analyte
absorption and background absorption.
• Spectral Interferences: Occur when other elements or compounds absorb at the same
wavelength.
• Chemical Interferences: Caused by chemical reactions that alter the absorption characteristics
of the analyte.
• Physical Interferences: Result from differences in sample viscosity, surface tension, and other
physical properties.
• Digestion: Samples are often digested with acids to bring them into solution.
• Use of Organic Solvents: Enhances the solubility of certain samples.
Sample Introduction Methods
• Environmental Analysis: Detection of trace metals in water, soil, and air samples.
• Clinical Analysis: Measurement of metal ions in biological samples such as blood and urine.
• Pharmaceuticals: Analysis of metal contaminants in drugs.
• Food and Beverage: Determination of metal content in food and drinks.
• Mining and Metallurgy: Quantification of metals in ores and processed materials.
Introduction
The principle of AES is based on the fact that when atoms or ions in a sample are excited to
higher energy levels (by heat, electrical discharge, or other energy sources), they emit light as
they return to their ground state. The wavelength and intensity of this emitted light are
characteristic of specific elements, allowing for their identification and quantification.
Plasma sources are often used in AES because they provide high temperatures that can
efficiently excite atoms and ions in the sample. The most common plasma sources include
inductively coupled plasma (ICP), direct current plasma (DCP), and microwave-induced plasma
(MIP).
Plasma is an ionized gas containing free electrons, ions, and neutral particles. It has unique
properties such as high temperature, which makes it an excellent medium for exciting atoms to
emit light. Plasma is characterized by:
ICP is created by passing an inert gas (usually argon) through a radio frequency (RF) coil, which
generates a magnetic field. This field ionizes the gas, forming plasma. ICP is favored for its
stability, high temperature, and ability to excite a wide range of elements.
DCP is generated by a direct current discharge between electrodes in the presence of an inert gas.
It is less commonly used than ICP but can be effective for certain applications.
MIP is produced by applying microwave radiation to a gas, usually argon. MIP systems are less
expensive and simpler than ICP systems but typically have lower temperatures and are used for
specific applications.
• It is inert and does not react with the sample or the equipment.
• It has a high ionization efficiency, which helps generate a stable plasma.
• It has a relatively low thermal conductivity, allowing it to maintain high temperatures.
1. Sample Introduction System: Introduces the sample into the plasma. Common methods
include nebulization (converting the liquid sample into an aerosol) and direct injection.
2. ICP Torch: Where the plasma is generated. The sample aerosol is introduced into the
plasma, causing the atoms to ionize.
3. Interface: Extracts the ions from the plasma and directs them into the mass spectrometer.
It typically consists of cones (sampler and skimmer) that allow the ions to pass through
while maintaining a vacuum.
4. Mass Spectrometer: Separates ions based on their mass-to-charge ratio (m/z). It
typically includes a quadrupole or time-of-flight (TOF) analyzer.
5. Detector: Measures the intensity of ions, which is proportional to the concentration of the
element in the sample.
6. Data System: Records and processes the data, providing quantitative and qualitative
analysis.
Applications of AES
AES is used in various fields for the analysis of trace and major elements:
Atomic fluorescence occurs when atoms in a sample absorb light at a specific wavelength,
become excited to higher energy states, and then emit light as they return to their ground state.
This emitted light (fluorescence) has a wavelength characteristic of the specific element and can
be measured to determine the concentration of the element in the sample.
The atomic fluorescence spectrum consists of sharp emission lines corresponding to the
transitions of electrons between different energy levels in an atom. The wavelengths of these
emission lines are unique to each element, making atomic fluorescence a powerful tool for
elemental analysis.
1. Resonance Fluorescence: Emission of light at the same wavelength as the absorbed light.
2. Stokes Fluorescence: Emission of light at a longer wavelength (lower energy) than the absorbed
light.
3. Anti-Stokes Fluorescence: Emission of light at a shorter wavelength (higher energy) than the
absorbed light, typically less common.
4. Stepwise Excitation and Fluorescence: Involves multiple absorption and emission steps, leading
to emission at different wavelengths.
1. Excitation: Atoms in the sample are excited by absorbing light from an external source.
2. Emission: The excited atoms emit light (fluorescence) as they return to their ground state.
3. Detection: The emitted fluorescence is measured, and its intensity is used to determine the
concentration of the element in the sample.
Fluorescence Intensity and Analyte Concentration
The intensity of the fluorescence emission is directly proportional to the concentration of the
analyte in the sample, provided that the excitation source is consistent and the sample matrix
does not interfere significantly. The relationship can be described by:
If=k⋅C
where IFI_FIF is the fluorescence intensity, kkk is a constant, and CCC is the analyte
concentration.
1. Light Source: Typically, a hollow cathode lamp (HCL) or a laser is used to provide the excitation
light.
2. Atomizer: Converts the sample into free atoms in the gas phase, commonly using a flame or a
graphite furnace.
3. Monochromator: Isolates the specific wavelength of fluorescence emission.
4. Detector: Measures the intensity of the fluorescence light, commonly using a photomultiplier
tube (PMT) or a charge-coupled device (CCD).
5. Data System: Records and processes the signal to determine the concentration of the element
in the sample.
AFS is widely used in various fields due to its high sensitivity and specificity for trace element
analysis:
• Environmental Analysis: Detection of trace metals in water, soil, and air samples.
• Clinical Analysis: Measurement of trace elements in biological samples such as blood and urine.
• Food and Beverage: Analysis of trace metals in food and drinks.
• Pharmaceuticals: Detection of metal contaminants in drugs.
• Industrial: Quality control of metal compositions in materials.
1. Spectral Interferences: Overlapping emission lines from other elements or molecular species.
2. Chemical Interferences: Reactions that alter the atomic state of the analyte, affecting
fluorescence.
3. Physical Interferences: Variations in sample introduction, such as differences in viscosity or
surface tension.
Merits:
Limitations: