CAT02-1 (CAMI Amplifier Load Cells)

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

ANALOGUE TRANSMITTER

CAT02

Analogue amplifier for load cells.

Designed for industrial


applications and laboratories.

High stability and accuracy

Available in version with 2kHz


frequency for measurement with
fast transients.

TECHNICAL FEATURES

Voltage supply 24 V dc ± 15 %
Power consumption max 5W
Insulation class Class III
Working temperature -10°C ÷ +40°C (humidity max 85%)
Storage temperature -20°C ÷ +50°C

Dimensions 76 mm x 90 mm x 65 mm (l x h x p)
Mounting Support for DIN or OMEGA bars
Material mounting support Polyamide 6.6 UL 94V-0
Connections Extractable screwed terminal boards - 5.08

Load cells voltage supply 10 Vdc / 120 mA (max 4 load cells 350Ω in parallel) protection
against short ciurcuit.
Linearity 0.02 % full scale
Temperature drift 0.005 % fondo scala / °C
Measurement range ± 1 mV/V, ± 2 mV/V, ± 3 mV/V selectable
Analogue filter Adjustable by trimmer
Zero and span setting Zero setting from –8% to + 3% (F.S.)
Span setting: ± 5% (F.S.)
Zero and span setting by trimmer 18 turns.

Output Voltage: 0 ÷ 10 V / 0 ÷ 5 V Current: 0 ÷ 20 mA / 4 ÷ 20 mA


Impedance Voltage: 10KΩ min.; Current: 300Ω µαξ

Standard EN50081-1, EN50082-2, EN61010-1

CAMI S.R.L.
MEASUREMENT INSTRUMENTS
20090 TREZZANO S/N
Tel. 02 4840 0166 - Fax 02 4840 1905
www.cami-it.com - e-mail [email protected]
CONNECTIONS TABLE

NUM. Terminal Board MA 9P - 5.08


1 + 24 Vcc ± 15% (Voltage supply)
2 GND
3 + Load cell voltage supply
4 - Load cell voltage supply
5 - Load cell signal
6 + Load cell signal
7 Current analogue output 0-20 mA / 4-20 mA
8 Negative analogue output (voltage / current)
9 Voltage analogue output 0-10 Vcc

CAMI S.R.L.
MEASUREMENT INSTRUMENTS
20090 TREZZANO S/N
Tel. 02 4840 0166 - Fax 02 4840 1905
www.cami-it.com - e-mail [email protected]

You might also like