FTIR
FTIR
FTIR
THEORY
What is Infrared?
Infrared radiation lies between the visible and microwave portions of the electromagnetic
spectrum. Infrared waves have wavelengths longer than visible and shorter than microwaves, and
have frequencies which are lower than visible and higher than microwaves.
The Infrared region is divided into: near, mid and far-infrared. Near-infrared refers to the part of
the infrared spectrum that is closest to visible light and far-infrared refers to the part that is closer
to the microwave region. Mid-infrared is the region between these two.
The primary source of infrared radiation is thermal radiation (heat). It is the radiation produced
by the motion of atoms and molecules in an object. The higher the temperature, the more the
atoms and molecules move and the more infrared radiation they produce. Any object radiates in
the infrared. Even an ice cube, emits infrared.
Infrared Spectroscopy
The bonds between atoms in the molecule stretch and bend, absorbing infrared energy and
creating the infrared spectrum.
A molecule such as H2O will absorb infrared light when the vibration (stretch or bend) results in
a molecular dipole moment change
A molecule can be characterized (identified) by its molecular vibrations, based on the absorption
and intensity of specific infrared wavelengths.
The interferometer produces a unique type of signal which has all of the infrared frequencies
―encoded‖ into it. The signal can be measured very quickly, usually on the order of one second or
so. Thus, the time element per sample is reduced to a matter of a few seconds rather than several
minutes. Most interferometers employ a beamsplitter which takes the incoming infrared beam
and divides it into two optical beams. One beam reflects off of a flat mirror which is fixed in
place. The other beam reflects off of a flat mirror which is on a mechanism which allows this
mirror to move a very short distance (typically a few millimeters) away from the beamsplitter.
The two beams reflect off of their respective mirrors and are recombined when they meet back at
the beamsplitter. Because the path that one beam travels is a fixed length and the other is
constantly changing as its mirror moves, the signal which exits the interferometer is the result of
these two beams ―interfering‖ with each other. The resulting signal is called an interferogram
which has the unique property that every data point (a function of the moving mirror position)
which makes up the signal has information about every infrared frequency which comes from the
source. This means that as the interferogram is measured, all frequencies are being measured
simultaneously. Thus, the use of the interferometer results in extremely fast measurements.
Because the analyst requires a frequency spectrum (a plot of the intensity at each
individual frequency) in order to make identification, the measured interferogram signal can not
be interpreted directly. A means of ―decoding‖ the individual frequencies is required. This can be
accomplished via a well-known mathematical technique called the Fourier transformation. This
transformation is performed by the computer which then presents the user with the desired
spectral information for analysis.
Attenuated Total Reflectance (ATR) is used for surface analysis of soft samples and liquids.
See figures below. The sample is put on top of the blue crystal. The IR beam picks up a signal
since it passes a few micron into the sample before it is reflected. It is therefore important to
obtain good contact between the crystal and the sample. This can be achieved by pressing the
sample down on the crystal as shown in the right hand image. If the substance under study is a
liquid, it is simply dropped onto the crystal, thereby obtaining optimal contact.
Light path in the ATR setup. Equipment to push the sample onto the crystal.
Potassium bromide (KBr) does not influence the IR spectrum in the wave number range 4000 til
400 cm-1. This technique can be used when the sample material can be crushed or ground. A
couple of milligram of sample is mixed with about 200 mg of KBr powder. The mixture is
pressed into a tablet which is then used in the analysis.
Light path with the KBr-technique. Sample holder with a KBr-tablet, ready
to be analysed in the sample chamber.
Microscope: With a microscope one can analyze very small samples, down to 5 µm x 5 µm.
Both transmission and reflection techniques can be used.
FT-IR Microscope
EXPERIMENTAL PROCEDURE
There are three main classes of triglycerides—saturated fats, and unsaturated trans-fats and cis-
fats. Unsaturation indicates the triglyceride contains one or more carbon–carbon double bonds.
Most natural oils like vegetable oil consist of cis-fats and are poly-unsaturated, meaning they
contain more than one double bond. The cis nature prevents solidification of the fat; trans-fats
solidify more readily, which can lead to blockages in the bloodstream.
Fourier transform infrared (FT-IR) spectroscopy functions well to analyse trans-fats. The
vibrational absorptions of fats show C–H bands around 2800–3000 cm–1 and the C=O band
between 1600 cm–1 and 1800 cm–1. Critically, intra- and intermolecular interactions shift the
peaks, allowing IR to distinguish between cis and trans (and saturated) fats.
Olive oil, margarine and butter will be tested by FTIR/ATR microscope to distinguish them in
relation to their cis and trans fatty acid content. Before testing each sample a background
spectrum will be recorded with an air emittance spectrum. Each sample will be tested by putting
small amount of them on to the crystal and measured between a frequency range 4000-650 cm-1
twice and recorded to computer. Then olive oil, margarine and butter will be heated and heated
samples will be tested for change in cis and trans fatty acid contents.