EN Service Manual Wasi-Steam Classic
EN Service Manual Wasi-Steam Classic
EN Service Manual Wasi-Steam Classic
Wasi-Steam Classic
SERVICE MANUAL
Dear customer,
Thank you for choosing a product from the Wassermann range. Wassermann Dental-
Maschinen incorporates the highest standards of quality and the latest technology.
In order to enjoy maximum performance and years of trouble-free operation, please read
this service manual carefully before you connect this device and start work, and operate
the device according to the recommended guidelines. The operation safety and the
functionality of this device can only be guaranteed if you follow both the general safety
guidelines and the applying laws to prevent accidents as well as the precautions given in
this service manual. We are not liable for any damages which occur due to inappropriate
usage or faulty operation of this device.
Make sure that anyone using this device has read and understood this service
manual.
Keep this service manual in a safe place where it can be referred to as required
at any time.
Company address:
Wassermann Dental-Maschinen GmbH
Rudorffweg 15-17
21031 Hamburg, Germany
Phone : +49 (0)40 730 926 -20/ -24 Fax.: +49 (0)40 730 37 24
[email protected]
www.wassermann.hamburg
www.facebook.com/WassermannDentalmaschinen
Warning!
This is a warning of risk situations and dangers.
Failure to observe this warning could be life-threatening. These warnings has to
be observed.
Information!
This symbol draws your attention to specific features that has to be observed.
3 Safety Guidelines
Configuring and operating this equipment requires precise knowledge and observance
of the instructions in this service manual. The equipment is designed only for its intended
application.
WARNING:
Servicing and repairs should be carried out only by authorised specialists.
Disconnect the power plug before starting any maintenance work.
Make sure that the equipment is connected to the correct power source.
4
● Never touch the steam gun directly at the nozzle – risk of scalding!
● Never point the steam gun at people (living beings) – risk of injury!
● The device can be filled manually, e.g. by means of a beaker, funnel, or the
like. The device must not be filled with pressure and or be attached to a tap.
5
4 Responsibility for Operation or Damage
The responsibility for operating the device lies exclusively with the owner or user if said
device is incorrectly serviced, maintained or altered by persons not employed by an
authorised dealer or if the device is used in a manner contrary to its specified purpose.
The unit has to be maintained and operated in accordance with this service manual.
Wassermann Dental-Maschinen GmbH is not responsible for damage arising from the
nonobservance of these instructions.
Warranty and responsibility provisions contained in the sales and supply conditions of
Wassermann Dental-Maschinen GmbH are not extended by these instructions.
4.1 Warranty
The warranty period for our equipment is 12 months. If faults occur within the
warranty period, contact your dental depot or get in touch directly with our service
department.
Your equipment should only be operated in perfect condition. If faults occur which
could harm operators or third parties, the unit should not be used until it has been
fixed.
This warranty does not cover damage caused by improper use, external mechanical
causes, transport damage or interference with the unit by unauthorized persons.
5 Application
Because of its variable work pressure of up to 5 bar it offers the user a gentle but quick
and thorough cleaning of all small parts in the dental laboratory.
6
6 Before Starting
6.1 Transport
Before transporting the unit, ensure that it has been unplugged from the power socket.
Make sure that it is packed correctly in order to avoid accidental damage.
Be sure to check for any transport damage when unpacking the goods. Note
down any damage if found.
6.2 Installation
Open the box, remove the packing materials, and carefully lift out the device and
accessories. Check the included accessories.
The device must stand on a steady and even surface or may be mounted to a stable wall
by means of the mounting kit. If mounting the unit on the wall, make sure that the wall is
strong enough to take the weight.
See 6.4 Mounting instructions fastening set for wall mounting Wasi-Steam Classic.
Install the device in a place where it will not block the working area and the
functionality (take the dimensions into account).
Please see that the louvres are not set because air circulation (heat flow) is
absolutely necessary.
7
6.3 Storage
If the unit is to be stored for an extended period, protect it from moisture and dust.
The unit location is very important when it comes to workplace safety, even if it is only to
be set up there for a short period. The room should be dry, well ventilated and vibration-
free. An even temperature and wooden supports also help.
The unit should not be stored or set up outdoors.
6.4 Mounting Instruction Fastening Set for Wall Mounting Wasi-Steam Classic
8
Mounting instruction:
Please drill the holes as indicated in the following drilling template. Please make sure
before that the wall is really suitable for mounting a device (aerated concrete walls and
drywall are problematic).
Drilling template
drilling
drilling
9
Secure the enclosed screw anchors as shown in Figure 1.
Screw the 6x50 screws into the two upper holes; keep a distance of 6-7 mm to the wall!
(Figure 2)
Screw
anchor S8
fig. 2
Srew 6 x 50
Screw
anchor S5
fig. 1
10
Screw the securing bracket to the Wasi-Steam Classic as shown in Figure 3.
Do not yet fully tighten the Phillips head screw, however – you still need to be able to move
the securing bracket.
Securing bracket
Washer
Serrated washer
fig. 3
11
Carefully hook in the device. Check, whether the device hangs loosely (i.e. moving back
and forth on the screws). If this is the case tighten the 6x50 screws until the device hangs
securely.
Finally tighten the securing bracket. First at the wall, then tighten the screw on the device
itself. (Figure 4)
tighten
Phillips head screw
in a final step
Screw 4 x 16
fig. 4
7 Start-up/ Installation
Before you use the steam cleaner, please put on heat-resistant gloves and wear
safety goggles.
Switch the unit off if it will not be used for an extended period.
Note: Optimal and fast pressurization is carried out only if the adjusting knob is in the
"max." position, since the heater with this is activated via a push button switch, i.e.
immediately at the start of every blasting action. In case of the pressure choice options (1-
4), the heater is activated via a thermostat, i.e. even in case of a massive decrease in
pressure it may take several minutes until the heater is activated due to residual heat of
the container for example. Therefore this setting is preferable only in case of occasional
work.
12
Please note the details on the equipment stickers!!
2 1
13
● Open the screw top of the container carefully while pressing down the screw top
● Fill with at most 2 litres of water
● Operation with distilled / softened water is possible and is recommended
● Tighten container screw top sturdily again (hand-screwed)
● Connect power plug
● Switch on main switch (4)
● Choose pressure range from 1 – 5 according to the scale or choose max. with
adjusting knob (3)
● Heat-up phase approx. 30 minutes (full container).
Briefly deaerate the device after approx. 15 minutes by operating the steam gun.
If the heating light (2) turns off, the device is ready for use. Now steam blasting via
the gun is possible by operating the pushbutton until the temperature delimiter
activates (button under black cap on the right side of the device) and the red light
(1) lights up.
Then proceed by referring to point 8.1.1 "refill water").
8 Operation
All instructions for using the unit, whether in verbal or written form, are based on our own
experience and experimentation and can only be regarded as guidelines.
Note: Do not refill more than 2 litres of water, if the red light (water
shortage) lights up (otherwise the container will be overfilled).
Operation with distilled / softened water is possible and is
recommended .
14
8.1.1 Refill
● After the red light (1) lights up, wait approx. 45 min. (cooling down
period) until you refill again, or else there is a risk of scalding as long as
the temperature remains above 100°C.
● Container must be pressureless; produce a pressure balance via the jet
nozzle by operating the steam gun.
● Only now open the screw top of the container carefully while pressing
down the screw top.
● Refill water (2 litres at most).
● Operation with distilled / softened water is possible and is recommended.
● Fully tighten the screw top of the container again (hand-screwed).
● Switch on the temperature delimiter (button under black cap on the right
side of device) again as follows; first unscrew the black cap, then push in
the pushbutton, the red light (1) goes out and the device heats up to the
adjusted pressure. Finally, screw the black cap back on.
● Heat-up phase (see p. 14).
● Operational; i.e. steam blasting via the gun by operating the pushbutton is
possible until the temperature delimiter activates again and the red light
(1) comes on.
15
9 Troubleshooting
If the above recommendations do not solve the problem, contact your dental depot
or our service department.
16
10 Care and Maintenance
10.1 Cleaning
Remove external dirt from time to time with some form of cold cleaner.
Use only cold cleaners to avoid damaging the paintwork or removing the lettering.
The equipment should be cleaned at regular intervals to ensure trouble-free
operation.
It requires only normal cleaning (sponge, damp cloth, mild detergent) and no further
chemical additives.
17
10.2 Maintenance
The Wasi-Steam Classic should, depending on frequency of use, but at least twice a year,
be decalcified in accordance with the following instructions. Depending on water hardness
a more frequent decalcification could be required. Operation with distilled / softened water
is possible and is recommended. As a basic principle, the device must always be kept
clean.
10.2.1 Decalcification
• Turn off the main switch and unplug the power plug
• Empty / pour out container in cold state
• Mix approx. 50 cc lime scale remover with 2 litres of water and fill into the device
(depending on lime content, the ratio can be increased)
• Heating regulator must be set to less than 1 bar
• Close tank cap
• Switch on the device and let it heat up briefly until the green light goes out
• Turn off the device and allow the lime scale remover to react for approx 8 hours
(lime remover should not react longer than 12 hours)
• Pour out the solution and rinse with clear water (approx. 1 litre) at least two times by
shaking the device
• Refill only maximal 2 litres of water (otherwise the tank could overfill)
• Operation with distilled / softened water is possible and is recommended
• The steam cleaner is now ready for operation
10.3 Repairs
Servicing or repairs to the unit has only to be carried out by qualified technicians.
Only original spare parts are to be used. Responsibility for the product is voided
if unauthorised persons alter it or if inappropriate components are installed.
18
10.4 Spare Parts
11 Technical Data
19
12 Disposing of the Unit
The unit has to be disposed by an authorized recycling operation. The selected company
has to be informed of all possibly health-hazardous residues in the unit.
20
13 EU Declaration of Conformity
in accordance with 2014/35/EU (Low Voltage Directive) and 2014/30/EU (Electromagnetic
Compatibility Directive) and 2006/42/EC (Machinery Directive) and 2011/65/EU (RoHS
Directive) and 2014/68/EU (Pressure Units Directive)
Signature : ________________________
Sven Wassermann
(Managing Director)
21
14 Wiring Diagram
22
15 Spare Part Diagram
23