[POWER ELECTRONICS & INSTRUMENTATION 18EC36] [email protected].

in

MODULE 4

Digital Voltmeter: Ramp Technique, Dual slope integrating Type DVM, Direct Compensation
type and Successive Approximations type DVM. Digital Multimeter: Digital Frequency Meter
and Digital Measurement of Time, Function Generator. Bridges: Measurement of resistance:
Wheatstone’s bridge, AC Bridges, Capacitance and Inductance Comparison bridge, Wien’s
bridge.

4.1 Digital Voltmeter: The digital voltmeters referred as DVM, converts the analog signals
into digital and display the voltages to be measured as discrete numerals rather than pointer
deflection, on the digital displays. DVMs can be used to measure ac and dc voltages and with
proper transducer and signal conditioning circuit it can also measure parameters like
pressure, temperature, stress etc. The output voltage is displayed on the digital display on
the front panel.
These DVMs reduces the human reading and interpretation errors and parallax errors.
The DVMs have various features and the advantages, over the conventional analog voltmeters
having pointer deflection on the continuous scale.

There are different types of DVM which differ in number of digits, accuracy, speed of
reading, size, power requirements and cost.

The important performance characteristics of DVM are as follows:


1. The input ranges from 1V to 1000V with provision for range selection and also indicates
the overload condition.
2. Accuracy is high as ±0.005% of reading.
3. Resolution is 1ppm i.e. the meter can read 1µV on a 1V range.
4. Input impedance is around 10MΩ which helps in reducing loading effect.
5. Output is in BCD form and for other forms of output digital processing modules can
be included.

4.2 Ramp Technique:

The operating principle is to measure the time that a linear ramp takes to change the input
level to the ground level, or vice-versa. This time period is measured with an electronic time-
interval counter and the count is displayed as a number of digits on an indicating tube or
display. The operating principle and block diagram of a ramp type DVM are shown in Figs
4.1 and 4.2. The ramp may be positive or negative; in this case a negative ramp has been
selected.

At the start of the measurement a ramp voltage is initiated (counter is reset to 0 and
sampled rate multivibrator gives a pulse which initiates the ramp genera-tor). The ramp
voltage is continuously compared with the voltage that is being measured. At the instant
these two voltage become equal, a coincidence circuit generates a pulse which opens a gate,
i.e. the input comparator generates a start pulse. The ramp continues until the second

MOHAMMED SALEEM | Asst. Prof., Dept. of E & C, PACE 1


[POWER ELECTRONICS & INSTRUMENTATION 18EC36] [email protected]

comparator circuit senses that the ramp has reached zero value. The ground comparator
compares the ramp with ground. When the ramp voltage equals zero or reaches ground
potential, the ground comparator generates a stop pulse. The output pulse from this
comparator closes the gate. The time duration of the gate opening is proportional to the input
voltage value.

Figure 4.1: Voltage to time conversion


In the time interval between the start and stop pulses, the gate opens and the oscillator
circuit drives the counter. The magnitude of the count indicates the magnitude of the input
voltage, which is displayed by the readout. Therefore, the voltage is converted into time and
the time count represents the magnitude of the voltage. The sample
rate multivibrator determines the rate of cycle of measurement. A typical value is 5
measuring cycles per second, with an accuracy of ± 0.005% of the reading. The sample rate
circuit provides an initiating pulse for the ramp generator to start its next ramp voltage. At
the same time a reset pulse is generated, which resets the counter to the zero state.

Figure 4.2: Block diagram of Ramp type DVM


MOHAMMED SALEEM | Asst. Prof., Dept. of E & C, PACE 2
[POWER ELECTRONICS & INSTRUMENTATION 18EC36] [email protected]

4.2.1 Advantages and Disadvantages

The ramp technique circuit is easy to design and its cost is low. Also, the output pulse can
be transmitted over long feeder lines. However, the single ramp requires excellent
characteristics regarding linearity of the ramp and time measurement. Large errors are
possible when noise is superimposed on the input signal. Input filters are usually required
with this type of converter.

4.3 Dual Slope Integrating Type DVM (Voltage to Time Conversion):

Dual Slope Integrating Type DVM – In ramp techniques, superimposed noise can cause large
errors. In the dual ramp technique, noise is averaged out by the positive and negative ramps
using the process of integration.

Principle of Dual Slope Type DVM

As illustrated in Fig. 4.3, the input voltage ’ei’ is integrated, with the slope of the integrator
output proportional to the test input voltage. After a fixed time, equal to t1, the input voltage
is disconnected and the integrator input is connected to a negative voltage — er. The
integrator output will have a negative slope which is constant and proportional to the
magnitude of the input voltage. The block diagram is given in Fig. 4.4.

Figure 4.3: Basic principle of dual slope type DVM

At the start a pulse resets the counter and the F/F output to logic level ‘0’. Si is closed
and Sr is open. The capacitor begins to charge. As soon as the integrator output exceeds zero,
the comparator output voltage changes state, which opens the gate so that the oscillator
clock pulses are fed to the counter. (When the ramp voltage starts, the comparator goes to
state 1, the gate opens and clock pulse drives the counter.) When the counter reaches
maximum count, i.e. the counter is made to run for a time ‘t1‘ in this case 9999, on the next
clock pulse all digits go to 0000 and the counter activates the F/F to logic level ‘1’. This
activates the switch drive, ei is disconnected and –er is connected to the integrator. The
integrator output will have a negative slope which is constant, i.e. integrator output now
decreases linearly to 0 volts. Comparator output state changes again and locks the gate. The
MOHAMMED SALEEM | Asst. Prof., Dept. of E & C, PACE 3
[POWER ELECTRONICS & INSTRUMENTATION 18EC36] [email protected]

discharge time t2 is now proportional to the input voltage. The counter indicates the count
during time t2. When the negative slope of the integrator reaches zero, the comparator
switches to state 0 and the gate closes, i.e. the capacitor C is now discharged with a constant
slope. As soon as the comparator input (zero detector) finds that e0, is zero, the counter is
stopped. The pulses counted by the counter thus have a direct relation with the input voltage.

Figure 4.4: Block diagram of dual slope type DVM

During charging

(4.1)

During discharging

(4.2)

Subtracting Eqs 5.2 from 5.1 we have

(4.3)

MOHAMMED SALEEM | Asst. Prof., Dept. of E & C, PACE 4


[POWER ELECTRONICS & INSTRUMENTATION 18EC36] [email protected]

If the oscillator period equals T and the digital counter indicates n1 and n2 counts
respectively,

Now, n1 and er are constants.

(4.4)

From Eq. 4.3 it is evident that the accuracy of the measured voltage is independent of
the integrator time constant. The times t1 and t2 are measured by the count of the clock given
by the numbers n1 and n2 respectively. The clock oscillator period equals T and if n1 and
er are constants, then Eq. 5.4 indicates that the accuracy of the method is also independent
of the oscillator frequency.
The dual slope technique has excellent noise rejection because noise and
superimposed ac are averaged out in the process of integration. The speed and accuracy are
readily varied according to specific requirements; also an accuracy of ± 0.05% in 100 ms is
available.

4.4 Successive Approximations

The successive approximations principle can be easily understood using a simple example;
the determination of the weight of an object. By using a balance and placing the object on
one side and an approximate weight on the other side, the weight of the object is determined.
If the weight placed is more than the unknown weight, the weight is removed and
another weight of smaller value is placed and again the measurement is performed. Now if it
is found that the weight placed is less than that of the object, another weight of smaller value
is added to the weight already present, and the measurement is performed. If it is found to
be greater than the unknown weight the added weight is removed and another weight of
smaller value is added. In this manner by adding and removing the appropriate weight, the
weight of the unknown object is determined.

The successive approximation DVM works on the same principle. Its basic block
diagram is shown in Fig. 4.5. When the start pulse signal activates the control circuit, the
successive approximation register (SAR) is cleared. The output of the SAR is 00000000. Vout of
the D/A converter is 0. Now, if Vin > Vout the comparator output is positive. During the first
clock pulse, the control circuit sets the D7 to 1, and Vout jumps to the half reference voltage.
The SAR output is 10000000. If V out is greater than Vin the comparator output is negative

MOHAMMED SALEEM | Asst. Prof., Dept. of E & C, PACE 5


[POWER ELECTRONICS & INSTRUMENTATION 18EC36] [email protected]

and the control circuit resets D7. However, if Vin is greater than Vout the comparator output
is positive and the control circuits keep D7 set. Similarly the rest of the bits beginning from
D7 to D0 are set and tested. Therefore, the measurement is completed in 8 clock pulses.

Table 4.1

At the beginning of the measurement cycle, a start pulse is applied to the start-stop
multivibrator. This sets a 1 in the MSB of the control register and a 0 in all bits (assuming
an 8-bit control) its reading would be 10000000. This initial setting of the register causes the
output of the D/A converter to be half the reference voltage, i.e. 1/2 V. This converter output
is compared to the unknown input by the comparator. If the input voltage is greater than the
converter reference voltage, the comparator output produces an output that causes the
control register to retain the 1 setting in its MSB and the converter continues to supply its
reference output voltage of 1/2 Vref.

Figure 4.5: Successive approximation DVM

MOHAMMED SALEEM | Asst. Prof., Dept. of E & C, PACE 6


[POWER ELECTRONICS & INSTRUMENTATION 18EC36] [email protected]

The ring counter then advances one count, shifting a 1 in the second MSB of the control
register and its reading becomes 11000000. This causes the D/A converter to increase its
reference output by 1 increment to 1/4 V, i.e. 1/2 V + 1/4 V, and again it is compared with
the unknown input. If in this case the total reference voltage exceeds the unknown voltage,
the comparator produces an output that causes the control register to reset its second MSB
to 0. The converter output then returns to its previous value of 1/2 V and awaits another
input from the SAR. When the ring counter advances by 1, the third MSB is set to 1 and the
converter output rises by the next increment of 1/2 V + 1/8 V. The measurement cycle thus
proceeds through a series of successive approximations. Finally, when the ring counter
reaches its final count, the measurement cycle stops and the digital output of the control
register represents the final approximation of the unknown input voltage.

4.5 Digital Multimeter Principles:

Analog meters require no power supply, they give a better visual indication of changes and
suffer less from electric noise and isolation problems. These meters are simple and
inexpensive. Digital meters, on the other hand, offer high accuracy, have a high input
impedance and are smaller in size. They gives an unambiguous reading at greater viewing
distances. The output available is electrical (for interfacing with external equipment), in
addition to a visual readout. The three major classes of Digital Multimeter Principles
are panel meters, bench type meters and system meters.

All Digital Multimeter Principles employ some kind of analog to digital (A/D) converters
(often dual slope integrating type) and have a visible readout display at the converter output.

Panel meters are usually placed at one location (and perhaps even a fixed range), while
bench meters and system meters are often multimeters, i.e. they can read ac and dc voltage
currents and resistances over several ranges.

Figure 4.6: (a) Digital Multimeter

MOHAMMED SALEEM | Asst. Prof., Dept. of E & C, PACE 7


[POWER ELECTRONICS & INSTRUMENTATION 18EC36] [email protected]

The basic circuit shown in Fig. 4.6(a) is always a dc voltmeter. Current is converted to
voltage by passing it through a precision low shunt resistance while alternating current is
converted into dc by employing rectifiers and filters. For resistance measurement, the meter
includes a precision low current source that is applied across the unknown resistance; again
this gives a dc voltage which is digitized and readout as ohms.

Bench meters are intended mainly for stand-alone operation and visual operation
reading, while system meters provide at least an electrical binary coded decimal output (in
parallel with the usual display), and perhaps sophisticated interconnection and control
capabilities, or even microprocessor based computing power.

A basic Digital Multimeter Principles (DMM) is made up of several A/D converters,


circuitry for counting and an attenuation circuit. A basic Digital Multimeter Block Diagram
is shown in Fig. 4.6 (b). The current to voltage converter shown in the block diagram of Fig.
4.6 (b) can be implemented with the circuit shown in Fig. 4.6 (c).

Figure 4.6: (b) Block diagram of a basic digital multimeter

The current to be measured is applied to the summing junction (Σi) at the input of the
opamp. Since the current at the input of the amplifier is close to zero because of the very
high input impedance of the amplifier, the current IR is very nearly equal to Ii, the current
IR causes a voltage drop which is proportional to the current, to be developed across the
resistors. This voltage drop is the input to the A/D converter, thereby providing a reading
that is proportional to the unknown current.
Resistance is measured by passing a known current, from a constant current source,
through an unknown resistance. The voltage drop across the resistor is applied to the A/D
converter, thereby producing an indication of the value of the unknown resistance.

MOHAMMED SALEEM | Asst. Prof., Dept. of E & C, PACE 8


[POWER ELECTRONICS & INSTRUMENTATION 18EC36] [email protected]

Figure 4.6: (c) Current to Voltage Converter

4.6 Digital Frequency Meter:

The Principle of Operation: The signal waveform is converted to trigger pulses and applied
continuously to an AND gate, as shown in Fig. 4.7. A pulse of 1 s is applied to the other
terminal, and the number of pulses counted during this period indicates the frequency.

Figure 4.7: Principle of digital frequency meter

The signal whose frequency is to be measured is converted into a train of pulses, one
pulse for each cycle of the signal. The number of pulses occurring in a definite interval of
time is then counted by an electronic counter. Since each pulse represents the cycle of the
unknown signal, the number of counts is a direct indication of the frequency of the signal
(unknown). Since electronic counters have a high speed of operation, high frequency signals
can be measured.

4.6.1 Basic Circuit of a Digital Frequency Meter:

The block diagram of a basic circuit of a digital frequency meter is shown in Fig. 4.8.

Figure 4.8: Basic circuit of a digital frequency meter

The signal may be amplified before being applied to the Schmitt trigger. The Schmitt
trigger converts the input signal into a square wave with fast rise and fall times, which is

MOHAMMED SALEEM | Asst. Prof., Dept. of E & C, PACE 9


[POWER ELECTRONICS & INSTRUMENTATION 18EC36] [email protected]

then differentiated and clipped. As a result, the output from the Schmitt trigger is a train of
pulses, one pulse for each cycle of the signal.

The output pulses from the Schmitt trigger are fed to a START/STOP gate. When this
gate is enabled, the input pulses pass through this gate and are fed directly to the electronic
counter, which counts the number of pulses.

When this gate is disabled, the counter stops counting the incoming pulses. The
counter displays the number of pulses that have passed through it in the time interval
between start and stop. If this interval is known, the unknown frequency can be measured.

4.6.2 Basic Circuit for Frequency Measurement:

The basic circuit for frequency measurement is as shown in Fig. 4.9. The output of the
unknown frequency is applied to a Schmitt trigger, producing positive pulses at the output.
These pulses are called the counter signals and are present at point A of the main gate.
Positive pulses from the time base selector are present at point B of the START gate and at
point B of the STOP gate.

Figure 4.9: Basic circuit for measurement of frequency showing gate control flip-flop

Initially the Flip-Flop (F/F-1) is at its logic 1 state. The resulting voltage from output Y
is applied to point A of the STOP gate and enables this gate. The logic 0 stage at the output
Y of the F/F-1 is applied to the input A of the START gate and disables the gate.

As the STOP gate is enabled, the positive pulses from the time base pass through the
STOP gate to the Set (S) input of the F/F-2 thereby setting F/F-2 to the 1 state and keeping
it there.

The resulting 0 output level from Y of F/F-2 is applied to terminal B of the main gate.
Hence no pulses from the unknown frequency source can pass through the main gate.

MOHAMMED SALEEM | Asst. Prof., Dept. of E & C, PACE 10


[POWER ELECTRONICS & INSTRUMENTATION 18EC36] [email protected]

In order to start the operation, a positive pulse is applied to (read input) reset input of
F/F-1, thereby causing its state to change. Hence Y = 1, Y= 0, and as a result the STOP gate
is disabled and the START gate enabled. This same read pulse is simultaneously applied to
the reset input of all decade counters, so that they are reset to 0 and the counting can start.

When the next pulse from the time base arrives, it is able to pass through the START
gate to reset F/F-2, therefore, the F/F-2 output changes state from 0 to 1, hence Y changes
from 0 to 1. This resulting positive voltage from Y called the gating signal, is applied to input
B of the main gate thereby enabling the gate.

Now the pulses from the unknown frequency source pass through the main gate to the
counter and the counter starts counting. This same pulse from the START gate is applied to
the set input of F/F-1, changing its state from 0 to 1. This disables the START gate and
enables the STOP gate. However, till the main gate is enabled, pulses from the
unknown frequency continue to pass through the main gate to the counter.

The next pulse from the time base selector passes through the enabled STOP gate to
the set input terminal of F/F-2, changing its output back to 1 and fi = 0. Therefore the main
gate is disabled, disconnecting the unknown frequency signal from the counter. The counter
counts the number of pulses occurring between two successive pulses from the time base
selector. If the time interval between this two successive pulses from the time base selector
is 1 second, then the number of pulses counted within this interval is the frequency of the
unknown frequency source, in Hertz.

The assembly consisting of two F/Fs and two gates is called a gate control F/F. The
block diagram of a digital frequency meter is shown in Fig. 4.10.

Figure 4.10: Block diagram of a digital frequency meter

MOHAMMED SALEEM | Asst. Prof., Dept. of E & C, PACE 11


[POWER ELECTRONICS & INSTRUMENTATION 18EC36] [email protected]

The input signal is amplified and converted to a square wave by a Schmitt trigger
circuit. In this diagram, the square wave is differentiated and clipped to produce a train of
pulses, each pulse separated by the period of the input signal. The time base selector output
is obtained from an oscillator and is similarly converted into positive pulses.

The first pulse activates the gate control F/F. This gate control F/F provides an enable
signal to the AND gate. The trigger pulses of the input signal are allowed to pass through the
gate for a selected time period and counted. The second pulse from the decade frequency
divider changes the state of the control F/F and removes the enable signal from the AND
gate, thereby closing it. The decimal counter and display unit output corresponds to the
number of input pulses received during a precise time interval; hence the counter display
corresponds to the frequency.

4.7 Digital Measurement of Time:


Principle of Operation: The beginning of the time period is the start pulse originating from
input 1, and the end of the time period is the stop pulse coming from input 2.

The oscillator runs continuously, but the oscillator pulses reach the output only
during the period when the control F/F is in the 1 state. The number of output pulses
counted is a measure of the time period.

4.7.1 Time Base Selector:


It is clear that in order to know the value of frequency of the input signal, the time interval
between the start and stop of the gate must be accurately known. This is called time base.

The time base consist of a fixed frequency crystal oscillator, called a clock oscillator,
which has to be very accurate. In order to ensure its accuracy, the crystal is enclosed in a
constant temperature oven. The output of this constant frequency oscillator is fed to a
Schmitt trigger, which converts the input sine wave to an output consisting of a train of
pulses at a rate equal to the frequency of the clock oscillator. The train of pulses then passes
through a series of frequency divider decade assemblies connected in cascade. Each decade
divider consists of a decade counter and divides the frequency by ten. Outputs are taken
from each decade frequency divider by means of a selector switch; any output may be
selected.

Figure 4.11: Time Base Selector


MOHAMMED SALEEM | Asst. Prof., Dept. of E & C, PACE 12
[POWER ELECTRONICS & INSTRUMENTATION 18EC36] [email protected]

The circuit of Fig. 4.11 consists of a clock oscillator having a 1 MHz frequency. The
output of the Schmitt trigger is 106 pulses per second and this point corresponds to a time
of 1 microsecond. Hence by using a 6 decade frequency divider, a time base with a range of
1μs—10μs—100μs —1ms—10ms—100ms—1s can be selected using a selector switch.

4.7.2 Measurement of Time (Period Measurement):

In some cases it is necessary to measure the time period rather than the frequency. This is
especially true in the measurement of frequency in the low frequency range. To obtain good
accuracy at low frequency, we should take measurements of the period, rather than make
direct frequency measurements. The circuit used for measuring frequency (Fig. 4.10) can be
used for the measurement of time period if the counted signal and gating signal are
interchanged.

Figure 4.12 shows the circuit for measurement of time period. The gating signal is
derived from the unknown input signal, which now controls the enabling and disabling of
the main gate. The number of pulses which occur during one period of the unknown signal
are counted and displayed by the decade counting assemblies. The only disadvantage is that
for measuring the frequency in the low frequency range, the operator has to calculate the
frequency from the time by using the equation f = 1/T.

Figure 4.12: Basic Block diagram of time measurement

For example, when measuring the period of a 60 Hz frequency, the electronic counter
might display 16.6673 ms, whence the frequency is

The accuracy of the period measurement and hence of frequency can be greatly
increased by using the multiple period average mode of operation. In this mode, the main
gate is enabled for more than one period of the unknown signal. This is obtained by passing
the unknown signal through one or more decade divider assemblies (DDAs) so that the period
is extended by a factor of 10,000 or more.

MOHAMMED SALEEM | Asst. Prof., Dept. of E & C, PACE 13


[POWER ELECTRONICS & INSTRUMENTATION 18EC36] [email protected]

Hence the digital display shows more digital of information, thus increasing accuracy.
However, the decimal point location and measurement units are usually changed each time
an additional decade divider is added, so that the display is always in terms of the period of
one cycle of the input signal, even though the measurements may have lasted for 10,100 or
more cycles.

Figure 4.13 show the multiple average mode of operation. In this circuit, five more
decade dividing assemblies are added so that the gate is now enabled for a much longer
interval of time than it was with single DDA.

Figure 4.13: Block diagram of a single and multiple period (average) measurement.

4.8 Function Generator

A Function Generator Block Diagram produces different waveforms of adjustable frequency.


The common output waveforms are the sine, square, triangular and sawtooth waves. The
frequency may be adjusted, from a fraction of a Hertz to several hundred kHz.
The various outputs of the generator can be made available at the same time. For
example, the generator can provide a square wave to test the linearity of an amplifier and
simultaneously provide a sawtooth to drive the horizontal deflection amplifier of the CRO to
provide a visual display.

Capability of Phase Lock: The function generator can be phase locked to an external source.
One function generator can be used to lock a second function generator, and the two output
signals can be displaced in phase by adjustable amount.

In addition, the fundamental frequency of one generator can be phase locked to a


harmonic of another generator, by adjusting the amplitude and phase of the harmonic,
almost any waveform can be generated by addition. The function generator can also be phase
locked to a frequency standard and all its output waveforms will then have the same accuracy
and stability as the standard source.
MOHAMMED SALEEM | Asst. Prof., Dept. of E & C, PACE 14
[POWER ELECTRONICS & INSTRUMENTATION 18EC36] [email protected]

The Function Generator Block Diagram is illustrated in Fig. 4.14. Usually the
frequency is controlled by varying the capacitor in the LC or RC circuit. In this instrument
the frequency is controlled by varying the magnitude of current which drives the integrator.
The instrument produces sine, triangular and square waves with a frequency range of 0.01
Hz to 100 kHz.

Figure 4.14: Function generator

The frequency controlled voltage regulates two current sources. The upper current
source supplies constant current to the integrator whose output voltage increases linearly
with time, according to the equation of the output signal voltage.

An increase or decrease in the current increases or decreases the slope of the output
voltage and hence controls the frequency. The voltage comparator multivibrator changes
states at a pre-determined maximum level of the integrator output voltage. This change cuts
off the upper current supply and switches on the lower current supply.

The lower current source supplies a reverse current to the integrator, so that its output
decreases linearly with time. When the output reaches a predetermined minimum level, the
voltage comparator again changes state and switches on the upper current source.

The output of the integrator is a triangular waveform whose frequency is determined


by the magnitude of the current supplied by the constant current sources.

The comparator output delivers a square wave voltage of the same frequency. The
resistance diode network alters the slope of the triangular wave as its amplitude changes and
produces a sine wave with less than 1% distortion.
MOHAMMED SALEEM | Asst. Prof., Dept. of E & C, PACE 15
[POWER ELECTRONICS & INSTRUMENTATION 18EC36] [email protected]

4.9 Wheatstone Bridge:


Introduction: A Bridge Circuit in its simplest form consists of a network of four resistance
arms forming a closed circuit, with a dc source of current applied to two
opposite junctions and a current detector connected to the other two junctions, as shown in
Fig. 4.15.

Figure 4.15: Wheatstone bridge


Wheatstone Bridge Circuit are extensively used for measuring component values such
as R, L and C. Since the bridge circuit merely compares the value of an unknown component
with that of an accurately known component (a standard), its measurement accuracy can be
very high. This is because the readout of this comparison is based on the null indication at
bridge balance, and is essentially independent of the characteristics of the null detector. The
measurement accuracy is therefore directly related to the accuracy of the bridge component
and not to that of the null indicator used.
The basic dc bridge is used for accurate measurement of resistance and is called
Wheatstone’s bridge.

4.9.1 Wheatstone Bridge Circuit (Measurement of Resistance):


Wheatstone’s bridge is the most accurate method available for measuring resistances and is
popular for laboratory use. The circuit diagram of a typical Wheatstone Bridge Circuit is given
in Fig. 4.15. The source of emf and switch is connected to points A and B, while a sensitive
current indicating meter, the galvanometer, is connected to points C and D. The
galvanometer is a sensitive microammeter, with a zero centre scale. When there is no current
through the meter, the galvanometer pointer rests at 0, i.e. mid-scale. Current in one
direction causes the pointer to deflect on one side and current in the opposite direction to
the other side.

MOHAMMED SALEEM | Asst. Prof., Dept. of E & C, PACE 16


[POWER ELECTRONICS & INSTRUMENTATION 18EC36] [email protected]

When SW1 is closed, current flows and divides into the two arms at point A, i.e. I1 and I2. The
bridge is balanced when there is no current through the galvanometer, or when the potential
difference at points C and D is equal, i.e. the potential across the galvanometer is zero.
To obtain the bridge balance equation, we have from the Fig. 4.15.

I1R1 = I2R2 (4.5)

For the galvanometer current to be zero, the following conditions should be satisfied.

(4.6)

(4.7)
Substituting in Eq. (4.2)

This is the equation for the bridge to be balanced.


In a practical Wheatstone Bridge Circuit, at least one of the resistance is made
adjustable, to permit balancing. When the bridge is balanced, the unknown resistance
(normally connected at R4) may be determined from the setting of the adjustable resistor,
which is called a standard resistor because it is a precision device having very small
tolerance.
Hence

(4.8)
4.9.2 Sensitivity of a Wheatstone Bridge
When the bridge is in an unbalanced condition, current flows through the galvanometer,
causing a deflection of its pointer. The amount of deflection is a function of the sensitivity of
the galvanometer. Sensitivity can be thought of as deflection per unit current. A more
sensitive galvanometer deflects by a greater amount for the same current. Deflection may be

MOHAMMED SALEEM | Asst. Prof., Dept. of E & C, PACE 17


[POWER ELECTRONICS & INSTRUMENTATION 18EC36] [email protected]

expressed in linear or angular units of measure, and sensitivity can be expressed in units of
S = mm/μA or degree/µA or radians/μA.

Therefore it follows that the total deflection D is D = S x I, where S is defined above and
I is the current in microamperes.

4.9.3 Unbalanced Wheatstone’s Bridge


To determine the amount of deflection that would result for a particular degree of unbalance,
general circuit analysis can be applied, but we shall use Thevenin’s theorem.

Since we are interested in determining the current through the galvanometer, we wish
to find the Thevenin’s equivalent, as seen by the galvanometer.

Thevenin’s equivalent voltage is found by disconnecting the galvanometer from the


Wheatstone Bridge Circuit, as shown in Fig. 4.16, and determining the open-circuit voltage
between terminals a and b.

Figure 4.16: Unbalanced Wheatstone’s bridge

Applying the voltage divider equation, the voltage at point a can be determined as follows

Therefore, the voltage between a and b is the difference between Ea and Eb, which represents
Thevenin’s equivalent voltage.

Therefore

MOHAMMED SALEEM | Asst. Prof., Dept. of E & C, PACE 18


[POWER ELECTRONICS & INSTRUMENTATION 18EC36] [email protected]

Thevenin’s equivalent resistance can be determined by replacing the voltage source E with
its internal impedance or otherwise short-circuited and calculating the resistance looking
into terminals a and b. Since the internal resistance is assumed to be very low, we treat it as
0 Ω. Thevenin’s equivalent resistance circuit is shown in Fig. 4.17.

Figure 4.17: Thevenin’s resistance Figure 4.18: Thevenin’s equivalent

The equivalent resistance of the circuit is R1//R3 in series with R2//R4 i.e. R1//R3 + R2//R4.

Therefore, Thevenin’s equivalent circuit is given in Fig. 4.18. Thevenin’s equivalent


circuit for the bridge, as seen looking back at terminals a and b in Fig. 4.16, is shown in Fig.
4.18.

If a galvanometer is connected across the terminals a and b of Fig.4.16, or its Thevenin


equivalent Fig. 4.18 it will experience the same deflection at the output of the bridge. The
magnitude of current is limited by both Thevenin’s equivalent resistance and any resistance
connected between a and b. The resistance between a and b consists only of the galvanometer
resistance Rg. The deflection current in the galvanometer is therefore given by

(4.9)

4.9.4 Slightly Unbalanced Wheatstone’s Bridge

If three of the four resistor in a bridge are equal to R and the fourth differs by 5% or less, we
can develop an approximate but accurate expression for Thevenin’s equivalent voltage and
resistance.

Consider the circuit in Fig. 4.19.

MOHAMMED SALEEM | Asst. Prof., Dept. of E & C, PACE 19


[POWER ELECTRONICS & INSTRUMENTATION 18EC36] [email protected]

Figure 4.19: Slightly unbalanced Wheatstone’s bridge

The voltage at point a is

The voltage at point b is

Thevenin’s equivalent voltage between a and b is the difference between these voltages.

Therefore
If Δ r is 5% of R or less, Δ r in the denominator can be neglected without introducing
appreciable error. Therefore, Thevenin’s voltage is

MOHAMMED SALEEM | Asst. Prof., Dept. of E & C, PACE 20


[POWER ELECTRONICS & INSTRUMENTATION 18EC36] [email protected]

Figure 4.20: Thevenin’s equivalent of slightly unbalanced Wheatstone’s bridge


The equivalent resistance can be calculated by replacing the voltage source with its
internal impedance (for all practical purpose short-circuit). The Thevenin’s equivalent
resistance is given by

Again, if Δr is small compared to R, Δ r can be neglected. Therefore, R R

Using these approximations, the Thevenin’s equivalent circuit is as shown in Fig. 4.20. These
approximate equations are about 98% accurate if Δr ≤ 0.05 R.

4.9.5 Application of Wheatstone’s Bridge


A Wheatstone bridge may be used to measure the dc resistance of various types of wire, either
for the purpose of quality control of the wire itself, or of some assembly in which it is used.
For example, the resistance of motor windings, transformers, solenoids, and relay coils can
be measured.

Wheatstone Bridge Circuit is also used extensively by telephone companies and others
to locate cable faults. The fault may be two lines shorted together, or a single line shorted to
ground.

4.9.6 Limitations of Wheatstone’s Bridge

MOHAMMED SALEEM | Asst. Prof., Dept. of E & C, PACE 21


[POWER ELECTRONICS & INSTRUMENTATION 18EC36] [email protected]

For low resistance measurement, the resistance of the leads and contacts becomes significant
and introduces an error. This can be eliminated by Kelvin’s Double bridge.

For high resistance measurements, the resistance presented by the bridge becomes so
large that the galvanometer is insensitive to imbalance. Therefore, a power supply has to
replace the battery and a dc VTVM replaces the galvanometer. In the case of high resistance
measurements in mega ohms, the Wheatstone’s bridge cannot be used.

Another difficulty in Wheatstone Bridge Circuit is the change in resistance of the bridge
arms due to the heating effect of current through the resistance. The rise in temperature
causes a change in the value of the resistance, and excessive current may cause a permanent
change in value.

4.10 Comparison Bridge:

There are two types of Comparison Bridge:

 Capacitance Comparison Bridge


 Inductance Comparison Bridge

4.10.1 Capacitance Comparison Bridge

Figure 4.21 shows the circuit of a capacitance comparison bridge. The ratio arms R1, R2 are
resistive. The known standard capacitor C3 is in series with R3. R3 may also include an added
variable resistance needed to balance the bridge. Cx is the unknown capacitor and Rx is the
small leakage resistance of the capacitor. In this case an unknown capacitor is compared
with a standard capacitor and the value of the former, along with its leakage resistance, is
obtained.

Figure 4.21: Capacitance comparison bridge

Hence,

MOHAMMED SALEEM | Asst. Prof., Dept. of E & C, PACE 22


[POWER ELECTRONICS & INSTRUMENTATION 18EC36] [email protected]

The condition for balance of the bridge is

Two complex quantities are equal when both their real and their imaginary terms are equal.
Therefore,

(4.10)

(4.11)

Since R3 does not appear in the expression for Cx, as a variable element it is an obvious
choice to eliminate any interaction between the two balance controls.

4.10.2 Inductance Comparison Bridge

Figure 4.22 gives a schematic diagram of an inductance comparison bridge. In this, values
of the unknown inductance Lx and its internal resistance Rx are obtained by comparison with
the standard inductor and resistance, i.e. L3 and R3.

MOHAMMED SALEEM | Asst. Prof., Dept. of E & C, PACE 23


[POWER ELECTRONICS & INSTRUMENTATION 18EC36] [email protected]

Figure 4.22: Inductance comparison bridge

The equation for balance condition is

The inductive balance equation yields

(4.12)
and resistive balance equations yields

(4.13)
In this bridge R2 is chosen as the inductive balance control and R3 as the resistance
balance control. (It is advisable to use a fixed resistance ratio and variable standards).
Balance is obtained by alternately varying L3 or R3. If the Q of the unknown reactance is
greater than the standard Q, it is necessary to place a variable resistance in series with the
unknown reactance to obtain balance.

If the unknown inductance has a high Q, it is permissible to vary the resistance ratio
when a variable standard inductor is not available.

4.11 Wien’s Bridge:

The Wien Bridge Circuit Diagram shown in Fig. 4.23 has a series RC combination in one arm
and a parallel combination in the adjoining arm. Wien’s bridge in its basic form, is designed
to measure frequency. It can also be used for the measurement of an unknown capacitor
with great accuracy.

Figure 4.23: Wien’s bridge


The impedance of one arm is
MOHAMMED SALEEM | Asst. Prof., Dept. of E & C, PACE 24
[POWER ELECTRONICS & INSTRUMENTATION 18EC36] [email protected]

The admittance of the parallel arm is

Using the bridge balance equation,

we have

Therefore,

Equating the real and imaginary terms we have

(4.14)

(4.15)

(4.16)

The two conditions for bridge balance, (4.14) and (4.16), result in an expression
determining the required resistance ratio R2/R4 and another expression determining the

MOHAMMED SALEEM | Asst. Prof., Dept. of E & C, PACE 25


[POWER ELECTRONICS & INSTRUMENTATION 18EC36] [email protected]

frequency of the applied voltage. If we satisfy Eq. (4.14) and also excite the bridge with the
frequency of Eq. (4.16), the bridge will be balanced.

In most Wien Bridge Circuit Diagram, the components are chosen such that R1 = R3 =
R and C1 = C3 = C. Equation (4.14) therefore reduces to R2/R4 = 2 and Eq. (4.16) to f=1/2 RC,
which is the general equation for the frequency of the bridge circuit.

The bridge is used for measuring frequency in the audio range. Resistances R1 and
R3 can be ganged together to have identical values. Capacitors C1 and C3 are normally of
fixed values.

The audio range is normally divided into 20 — 200 — 2 k — 20 kHz ranges. In this
case, the resistances can be used for range changing and capacitors C1 and C3 for fine
frequency control within the range. The Wien Bridge Circuit Diagram can also be used for
measuring capacitances. In that case, the frequency of operation must be known.

The bridge is also used in a harmonic distortion analyzer, as a Notch filter, and in
audio frequency and radio frequency oscillators as a frequency determining element.

An accuracy of 0.5% — 1% can be readily obtained using this bridge. Because it is


frequency sensitive, it is difficult to balance unless the waveform of the applied voltage is
purely sinusoidal.

MOHAMMED SALEEM | Asst. Prof., Dept. of E & C, PACE 26


[POWER ELECTRONICS & INSTRUMENTATION 18EC36] [email protected]

MOHAMMED SALEEM | Asst. Prof., Dept. of E & C, PACE 27