Turbimax CUS51D
Turbimax CUS51D
Operating Instructions
Turbimax CUS51D
Sensor for turbidity and solids content
Turbimax CUS51D Table of contents
Table of contents
5 Mounting . . . . . . . . . . . . . . . . . . . . . . . . . 15
5.1 Mounting requirements . . . . . . . . . . . . . . . . . 15
5.2 Mounting the sensor . . . . . . . . . . . . . . . . . . . 16
5.3 Post-mounting check . . . . . . . . . . . . . . . . . . . 23
6 Electrical connection . . . . . . . . . . . . . . 24
6.1 Connecting the sensor . . . . . . . . . . . . . . . . . . 24
6.2 Ensuring the degree of protection . . . . . . . . . . 25
6.3 Post-connection check . . . . . . . . . . . . . . . . . . 26
7 Commissioning . . . . . . . . . . . . . . . . . . . . 27
7.1 Function check . . . . . . . . . . . . . . . . . . . . . . . 27
8 Operation . . . . . . . . . . . . . . . . . . . . . . . . . 28
8.1 Adapting the measuring device to the process
conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
10 Maintenance . . . . . . . . . . . . . . . . . . . . . . 39
10.1 Maintenance tasks . . . . . . . . . . . . . . . . . . . . . 39
11 Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
11.1 General notes . . . . . . . . . . . . . . . . . . . . . . . . 40
Endress+Hauser 3
About this document Turbimax CUS51D
4 Endress+Hauser
Turbimax CUS51D About this document
1.4 Documentation
The following manuals, which complement these Operating Instructions, can be found on
the product pages on the Internet:
Technical Information Turbimax CUS51D, TI00461C
Endress+Hauser 5
Basic safety instructions Turbimax CUS51D
6 Endress+Hauser
Turbimax CUS51D Basic safety instructions
During operation:
‣ If faults cannot be rectified:
products must be taken out of service and protected against unintentional operation.
Endress+Hauser 7
Product description Turbimax CUS51D
3 Product description
4
A0030850
1 Deflection of light
1 Light source
2 Light beam
3 Particle
4 Scattered light
The incident light is scattered in many directions, i.e. at different angles to the direction of
propagation. 2 angle ranges are of particular interest here:
• Light scattered at a 90° angle is used primarily for turbidity measurement in drinking
water.
• Light scattered at a 135° angle extends the dynamic range for high particle densities.
8 Endress+Hauser
Turbimax CUS51D Product description
135° 90°
1 2 3
A0030846
If the particle density in the medium is low, most of the light is scattered in the 90°channel
and a small amount of light is scattered in the 135° channel. As the particle density
increases, this ratio shifts (more light in the 135° channel, less light in the 90° channel).
Ir
1
135°
90°
FNU
A0030849
The CUS51D turbidity sensor has 2 sensor units, which are independent of each other and
arranged in parallel. The application-dependent evaluation of both signals leads to stable
measured values.
Endress+Hauser 9
Product description Turbimax CUS51D
4 6
3 5
1 2
A0030845
The sensor covers a broad range of turbidity and solids measurements thanks to the
optical arrangement with 2 light sources, each with 2 light receivers placed at different
angles (90° and 135°).
• As soon as the customer selects an application, e.g. Activated sludge, the optical
method best suited for the particular measuring task is automatically activated in the
sensor (e.g. 90° measurements with both light sources).
• The double sensing system (2 light sources with 2 receivers per source) largely
compensates for measurement errors caused by fouling (4-beam pulsed light method
→ 10).
The sensor types available vary in terms of their measuring ranges and therefore the
range of available applications.
10 Endress+Hauser
Turbimax CUS51D Product description
S1 E1-135 E1-90
S2 E2-135 E2-90
A0030847
P a
S
A0030852
Endress+Hauser 11
Product description Turbimax CUS51D
Is = I0 · C · A· f(a)
IS
E1 a P
I0
S1
A0030855
12 Endress+Hauser
Turbimax CUS51D Incoming acceptance and product identification
4.2.1 Nameplate
The nameplate provides you with the following information on your device:
• Manufacturer identification
• Order code
• Extended order code
• Serial number
• Safety information and warnings
Product page
www.endress.com/cus51d
Endress+Hauser 13
Incoming acceptance and product identification Turbimax CUS51D
Manufacturer's address
Endress+Hauser Conducta GmbH+Co. KG
Dieselstraße 24
70839 Gerlingen
Germany
14 Endress+Hauser
Turbimax CUS51D Mounting
5 Mounting
5.1.1 Dimensions
(0.81)
20.5
12.5 (0.49)
82.7 (3.26)
NPT ¾“
G1
36
13.5 (0.53)
(1.42)
Ø 40
(1.57)
188.5 (7.42)
A0030853
Endress+Hauser 15
Mounting Turbimax CUS51D
A B 72 (2.83)
(0.70)
17.8
(0.70)
17.8
77.3 (3.04)
65.8 (2.59)
11.5 (0.45)
48 (1.89)
Ø 40.1 (1.58)
Ø 50 (1.97)
A0030854
16 Endress+Hauser
Turbimax CUS51D Mounting
3
2 4
A0051207
A0030856
Endress+Hauser 17
Mounting Turbimax CUS51D
2
1
4 3
A0030843
Pipe installation
A0051206
The installation angle α must not exceed 90° → 13, 18. The recommended
installation angle is 75°. The optical windows of the sensor must be aligned along the flow
direction.
The medium pressure may not exceed 2 bar (29 psi) for manual assembly retraction.
18 Endress+Hauser
Turbimax CUS51D Mounting
A0035858
The installation angle is 90°. For turbidity measurements < 200 FNU, the backscattering of
the internal surfaces of the assembly causes distortions in the measured values.
The following diagram shows different installation scenarios in pipes, indicating whether
or not they are permitted.
3
5
1 4
A0030848
• When using reflective materials (e.g. stainless steel), the pipe diameter must be at least
100 mm (3.9 in). An onsite calibration is recommended.
• Install the sensor in places with consistent flow conditions.
• The best installation location is in the ascending pipe (item 1). Installation is also
possible in the horizontal pipe (item 5).
• Do not install in places where air spaces or bubbles occur (item 3) or where
sedimentation may occur (item 2).
• Avoid installation in the down pipe (item 4).
• When measuring turbidity < 200 FNU, the backscattering of the pipe wall causes
distortions in the measured values. For this reason, measured value adjustment with an
offset is recommended here.
• Avoid fittings downstream from pressure reduction stages which can lead to outgassing.
Endress+Hauser 19
Mounting Turbimax CUS51D
Immersion operation
3
3
2 4
4 1
6
1
A0051207
A0013383
20 Endress+Hauser
Turbimax CUS51D Mounting
4
3
3 2 4
2
6
1
7
A0013384 A0051208
The chain retainer is particularly suitable for applications that require a sufficient distance
between the mounting location and the wall of the aeration basin. As the assembly is
freely suspended, any vibration of the upright post is avoided. The swinging of the chain
enhances the self-cleaning effect of the optics.
For detailed information on mounting with the wastewater assembly, see the
Operating Instructions BA00432C
Endress+Hauser 21
Mounting Turbimax CUS51D
A0031105
The cleaning unit is particularly suitable for clear water or media with a high fat content
that tend to cause heavy buildup.
2
1 1
A0030861
A0030860
22 Endress+Hauser
Turbimax CUS51D Mounting
Endress+Hauser 23
Electrical connection Turbimax CUS51D
6 Electrical connection
LWARNING
Device is live!
Incorrect connection may result in injury or death!
‣ The electrical connection may be performed only by an electrical technician.
‣ The electrical technician must have read and understood these Operating Instructions
and must follow the instructions contained therein.
‣ Prior to commencing connection work, ensure that no voltage is present on any cable.
PK
85
2DS
1
GY
86
PK
85
1
85
GY
86
86
85
2
Sensor 1
86
BN
98 97 88 87
WH
Sensor 1 Sensor GN
YE
98 97 88 87
GN
YE
Sensor 2
98 97 88 87
Sensor
A0033092
24 Endress+Hauser
Turbimax CUS51D Electrical connection
Cable sample (does not necessarily correspond to the original cable supplied)
1 2
3
4
24 Terminated cable
A0045763 A0045764
1 Outer shield (exposed) 25 Connect the cable to the 26 Press the cable into the
2 Cable cores with ferrules grounding clamp grounding clamp
3 Cable sheath (insulation)
4 Grounding clamp
The cable shield is grounded using
the grounding clamp 1)
1) Please note the instructions in the "Ensuring the degree of protection" section
Endress+Hauser 25
Electrical connection Turbimax CUS51D
Is the outside of the sensor, assembly or cable free ‣ Perform a visual inspection.
from damage?
Are the mounted cables strain-relieved and not ‣ Perform a visual inspection.
twisted? ‣ Untwist the cables.
Is a sufficient length of the cable cores stripped, and ‣ Perform a visual inspection.
are the cores positioned in the terminal correctly? ‣ Pull gently to check they are seated correctly.
Are the power supply and signal lines connected ‣ Use the transmitter wiring diagram.
correctly?
26 Endress+Hauser
Turbimax CUS51D Commissioning
7 Commissioning
Endress+Hauser 27
Operation Turbimax CUS51D
8 Operation
8.1.1 Applications
The sensor permits measurements in a wide variety of applications. The measuring
method is set automatically by selecting the relevant application.
A
B
C
g/l
0 10 20 30 40 50 60 70 80 90 100 110 120 130 140 150 160
D
FNU
1000
2000
3000
4000
5000
6000
7000
8000
9000
9999
0
A0050651
28 Endress+Hauser
Turbimax CUS51D Operation
A
B
C
D
E
g/l
0 10 20 30 40 50 60 70 80 90 100
A0050652
The Thin sludge application enables measurements in any sludge applications from
0 to 5 g/l (0 to 0.04 lb/gal). Measurements in numerous sludge applications from
0 to 50 g/l (0 to ⁰.⁴⁄ lb/ga) (e.g. SBR) are possible with the Sludge, general application.
These applications can be calibrated at a single point in the process during operation.
Fields of use and associated applications → 29
NOTICE
Multiple scattering in the following applications: formazine, kaolin and thin sludge
If the specific operational range is exceeded, the measured value displayed by the sensor
can decrease despite increasing turbidity or increasing TS content. The indicated
operational range is reduced in the case of highly absorbing (e.g. dark) media.
‣ In the case of highly absorbing (e.g. dark) media, determine the operational range
experimentally beforehand.
8.1.2 Calibration
The sensor is precalibrated on leaving the factory. As such, it can be used in a wide range
of applications (e.g. clear water measurement) without any further calibration. The factory
calibrations are based on a three-point calibration in each case. The Formazine
application is already fully calibrated and can be used without any further calibration.
All other applications are precalibrated with reference samples and require calibration to
the corresponding application.
In addition to the factory calibration data, which cannot be modified, the sensor has five
other data records to be used for storing process calibrations.
Application selection
‣ During initial commissioning or calibration at the transmitter, select the appropriate
application for your field of application and measuring range.
Application: Wastewater
Field of application Range Application Recommended type of
calibration
Inlet < 5 g/l Thin sludge [mg/l, g/l] Single-point (in the
Formazine process)
[FNU, NTU]
Endress+Hauser 29
Operation Turbimax CUS51D
Primary sludge 3 to approx. Excess sludge [g/l, %TS] Two-point (outside the
extraction, primary 50 g/l process)
clarification
> approx. Digested sludge[g/l, %TS] Two-point (outside the
50 g/l process)
Aeration basin 0 to 5 g/l Thin sludge [mg/l, g/l] Single-point (in the
process)
Sequencing batch 0 to approx. Sludge, general [mg/l, g/l, %TS] Single-point (in the
reactors 50 g/l For applications with wide dynamic range, process)
from clear water to high solids content
Recirculation pipe 3 to approx. Excess sludge [g/l, %TS] Two-point (outside the
50 g/l process)
Waste activated sludge 3 to approx. Excess sludge [g/l, %TS] Two-point (outside the
extraction 50 g/l process)
Sludge thickener 3 to approx. Excess sludge [g/l, %TS] Two-point (outside the
(primary sludge) 50 g/l process)
Digester inlet 3 to approx. Excess sludge [g/l, %TS] Two-point (outside the
50 g/l process)
Digester outlet (sludge) > 5 g/l Digested sludge [g/l, %TS] Two-point (outside the
process)
WWTP outlet 0 to 5 g/l Formazine [FNU, NTU], Thin sludge [mg/l, Single-point (in the
g/l] process)
Kaolin
[mg/l, g/l]
Sand filter monitoring 0 to 5 g/l Formazine [FNU, NTU], Thin sludge [mg/l, Single-point (in the
g/l] process)
Silicon dioxide process 0 to 5 g/l Formazine [FNU, NTU], Single-point (in the
water Thin sludge process)
[mg/l, g/l], Kaolin [mg/l, g/l])
Silicon dioxide process 5 to 100 g/l SiO2 [ppm, g/l] Two-point (outside the
sludges process)
30 Endress+Hauser
Turbimax CUS51D Operation
Titanium dioxide 1 to 150 g/l TiO2 [ppm, g/l] Two-point (outside the
process sludges process)
Kaolin process water/ 0 to 5 g/l Kaolin [mg/l, g/l] Single-point (in the
process water sludges process)
b
a
signal c
0 2 4 6 8
g/l
A0050659
29 Single-point calibration
a Factory calibration curve
b New calibration curve
c Calibration point
A single-point calibration causes a change in the slope of the factory calibration curve
programmed into the device.
Two-point calibration
a
b
signal
0 c c
0 2 4 6 8
g/l
A0050661
30 Two-point calibration
a Factory calibration curve
b New calibration curve
c Calibration points
A two-point calibration causes a change in the slope and zero point of the factory
calibration curve programmed into the device. This type of calibration is recommended as
Endress+Hauser 31
Operation Turbimax CUS51D
the standard method as it produces robust calibration curves and good measurement
results with minimum calibration effort.
1. Select the two calibration points at the limits of the expected measuring range.
2. Do not select any calibration points outside the specified measuring range for the
application.
Three-point calibration
signal
b
0 c c c
0 2 4 6 8
g/l
A0050664
31 Three-point calibration
a Factory calibration curve
b New calibration curve
c Calibration points
With a three-point calibration, a new calibration curve is plotted through all 3 calibration
points, resulting in a high level of accuracy in the calibrated range.
1. Within the measuring range, select calibration points that are as far apart as
possible.
2. Do not select any calibration points outside the specified measuring range for the
application.
If the calibration points selected are inappropriate, the curve profile will be distorted
to an extent that could result in implausible measured values.
Five-point calibration
c
Δ5
a
Δ4
signal Δ2 Δ3
c c
b Δ1 c
0 c
0 2 4 6 8
g/l
A0050665
32 Five-point calibration
a Factory calibration curve
b New calibration curve
c Calibration points
32 Endress+Hauser
Turbimax CUS51D Operation
With a four- or five-point calibration, the calibration curve is plotted between the
calibration points. Avoid this type of calibration if possible, as it does not significantly
improve accuracy.
1. For the laboratory measurement, take a sample of the medium in the direct vicinity
of the sensor.
2. Give the sample to the laboratory so that the turbidity or solids content can be
determined.
3. Select a data record on the CM44x transmitter.
4. If possible, start the calibration at the same time as the sampling procedure and enter
the laboratory value of the sample as the set point.
5. Enter an approximate value as the set point if no laboratory value is available during
calibration.
As soon as the laboratory value is available, amend the set point on the
transmitter.
Endress+Hauser 33
Operation Turbimax CUS51D
A0020482
A0035855
A0035856
Siphon off the excess water (if possible) in order to increase the concentration of the
sample.
4. Stir the sample to make it more homogeneous.
5.
A0020485
A0020486
Transfer a defined amount of the sample (e.g.2 l (0.5 gal)) to the calibration vessel
(bucket).
7. Continue stirring the sample to maintain homogeneity.
34 Endress+Hauser
Turbimax CUS51D Operation
A0020487
1
~10 (0.39)
>10
(0.39)
A0030900
Endress+Hauser 35
Operation Turbimax CUS51D
Performing a calibration:
Using the example of a two-point calibration in the expected measuring range 2 to 6 g/l.
1. At the CM44x transmitter, select a free data record and the suitable application.
2. Wait at least 1 minute (to stabilize).
3.
A0020489
Start the calibration for measuring point 1 (e.g. 2 l (0.5 gal). Sample with a
concentration of 6 g/l (0.05 lb/gal)).
4. Enter the value of the sample determined in the laboratory as the set point (e.g.
6 g/l (0.05 lb/gal)) or edit the value later.
5.
H2O
A0030902
Perform a 1:3 dilution of the sample. Add water (4 l (1.1 gal)); in the example this
results in 2 g/l (0.02 lb/gal).
6. Avoid air bubbles beneath the sensor.
7. Calibrate measuring point 2. Enter a third of the laboratory value as the set point.
Stability criterion
During calibration, the measured values provided by the sensor are checked to ensure they
are constant. The maximum deviations that may occur in measured values during a
calibration are defined in the stability criterion.
The specifications comprise the following:
• The maximum permitted deviation in temperature measurement
• The maximum permitted deviation in measured value as a %
• The minimum time frame in which these values must be maintained
The calibration resumes as soon as the stability criteria for signal values and temperature
have been reached. If these criteria are not met in the maximum time frame of 5 minutes,
no calibration is performed - a warning is issued.
36 Endress+Hauser
Turbimax CUS51D Operation
The stability criteria are used to monitor the quality of the individual calibration points in
the course of the calibration process. The aim is to achieve the highest possible calibration
quality in the shortest possible time frame while taking external conditions into account.
For calibrations in the field in adverse weather and environmental conditions, the
measured value windows selected can be suitably large and the time frame selected
can be suitably short.
Strong Strong filtering, low sensitivity, slow response to changes (25 seconds)
Endress+Hauser 37
Diagnostics and troubleshooting Turbimax CUS51D
Blank display, no sensor reaction • Line voltage at transmitter? ‣ Connect mains voltage.
• Sensor connected correctly? ‣ Establish correct connection.
• Buildup on optical windows? ‣ Clean sensor.
Display value too high or too low • Buildup on optical windows? ‣ Clean device.
• Sensor calibrated? ‣ Calibrate device.
Display value fluctuating greatly Is the mounting location correct? ‣ Select a different mounting location.
‣ Adjust measured value filter.
Pay attention to the troubleshooting information in the Operating Instructions for the
transmitter. Check the transmitter if necessary.
38 Endress+Hauser
Turbimax CUS51D Maintenance
10 Maintenance
LCAUTION
Acid or medium
Risk of injury, damage to clothing and the system!
‣ Switch off cleaning before the sensor is removed from the medium.
‣ Wear protective goggles and safety gloves.
‣ Clean away splashes on clothes and other objects.
‣ You must perform maintenance tasks at regular intervals.
We recommend setting the maintenance times in advance in an operations journal or log.
The maintenance cycle primarily depends on the following:
• The system
• The installation conditions
• The medium in which measurement takes place
‣ To ensure reliable measurements, clean the sensor at regular intervals. The frequency
and intensity of the cleaning depend on the medium.
Lime deposits ‣ Immerse the sensor in 1 to 5% hydrochloric acid (for several minutes).
Dirt particles on the optics ‣ Clean the optics with a cleaning cloth.
After cleaning:
‣ Rinse the sensor thoroughly with water.
Endress+Hauser 39
Repair Turbimax CUS51D
11 Repair
11.3 Return
The product must be returned if repairs or a factory calibration are required, or if the
wrong product was ordered or delivered. As an ISO-certified company and also due to legal
regulations, Endress+Hauser is obliged to follow certain procedures when handling any
returned products that have been in contact with medium.
11.4 Disposal
The device contains electronic components. The product must be disposed of as electronic
waste.
40 Endress+Hauser
Turbimax CUS51D Accessories
12 Accessories
The following are the most important accessories available at the time this documentation
was issued.
Listed accessories are technically compatible with the product in the instructions.
1. Application-specific restrictions of the product combination are possible.
Ensure conformity of the measuring point to the application. This is the responsibility
of the operator of the measuring point.
2. Pay attention to the information in the instructions for all products, particularly the
technical data.
3. For accessories not listed here, please contact your Service or Sales Center.
12.1.1 Assemblies
FlowFit CUA120
• Flange adapter for mounting turbidity sensors
• Product Configurator on the product page: www.endress.com/cua120
Technical Information TI096C
Flexdip CYA112
• Immersion assembly for water and wastewater
• Modular assembly system for sensors in open basins, channels and tanks
• Material: PVC or stainless steel
• Product Configurator on the product page: www.endress.com/cya112
Technical Information TI00432C
Cleanfit CUA451
• Manual retractable assembly made of stainless steel with ball valve shut-off for turbidity
sensors
• Product Configurator on the product page: www.endress.com/cua451
Technical Information TI00369C
Flowfit CYA251
• Connection: See product structure
• Material: PVC-U
• Product Configurator on the product page: www.endress.com/cya251
Technical Information TI00495C
12.1.2 Cable
Memosens data cable CYK11
• Extension cable for digital sensors with Memosens protocol
• Product Configurator on the product page: www.endress.com/cyk11
Technical Information TI00118C
Endress+Hauser 41
Accessories Turbimax CUS51D
12.1.3 Holder
Flexdip CYH112
• Modular holder system for sensors and assemblies in open basins, channels and tanks
• For Flexdip CYA112 water and wastewater assemblies
• Can be affixed anywhere: on the ground, on the coping stone, on the wall or directly
onto railings.
• Stainless steel version
• Product Configurator on the product page: www.endress.com/cyh112
Technical Information TI00430C
12.1.5 Cable
Memosens data cable CYK11
• Extension cable for digital sensors with Memosens protocol
• Product Configurator on the product page: www.endress.com/cyk11
Technical Information TI00118C
42 Endress+Hauser
Turbimax CUS51D Technical data
13 Technical data
13.1 Input
CUS51D-**D1 Application
Solids content 0 to 300 g/l (0 to 2.5 lb/gal) Solids content depending on the
0 to 30 % selected application (see list)
Endress+Hauser 43
Technical data Turbimax CUS51D
Maximum measured error Turbidity < 2% of measured value or 0.1 FNU (the greater value applies in each case).
Solids < 5% of the measured value or 1% of the upper range value (the greater
value applies in each case); applies to sensors that are calibrated for the
observed measuring range.
The measurement error encompasses all inaccuracies of the measuring chain (sensor
and transmitter). However, it does not include the inaccuracy of the reference
material used for calibration.
For solids, the achievable measurement errors depend very much on the media that
are actually present and may differ from the specified values. Extremely
inhomogeneous media cause the measured value to fluctuate and increase the
measurement error.
Drift Working on the basis of electronic controls, the sensor is largely free of drifts.
13.4 Environment
44 Endress+Hauser
Turbimax CUS51D Technical data
13.5 Process
Endress+Hauser 45
Index Turbimax CUS51D
Index
0…9 R
4-beam pulsed light method . . . . . . . . . . . . . . . . . . . 10 Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
90° scattered light method . . . . . . . . . . . . . . . . . . . . . 11 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
135° backscattered light method . . . . . . . . . . . . . . . . 11
S
A Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Sensor structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
C Signal filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Certificates, approvals . . . . . . . . . . . . . . . . . . . . . . . . 14 Stability criterion . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37, 39 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Cyclic cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
T
D Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 U
Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
E
Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . 24 W
Energy supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
F
Function check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
I
Immersion operation . . . . . . . . . . . . . . . . . . . . . . . . . 20
Incoming acceptance . . . . . . . . . . . . . . . . . . . . . . . . . 13
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Installation examples . . . . . . . . . . . . . . . . . . . . . . . . 18
Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
M
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Measuring methods . . . . . . . . . . . . . . . . . . . . . . . . . 10
Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Mechanical construction . . . . . . . . . . . . . . . . . . . . . . 45
Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
N
Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
P
Performance characteristics . . . . . . . . . . . . . . . . . . . . 43
Pipe installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Post-connection check . . . . . . . . . . . . . . . . . . . . . . . . 26
Post-installation check . . . . . . . . . . . . . . . . . . . . . . . 23
Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Product description . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Product design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Product identification . . . . . . . . . . . . . . . . . . . . . . . . 13
Product safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
46 Endress+Hauser
*71623201*
71623201
www.addresses.endress.com