Spectrofluorophotometer Manual
Spectrofluorophotometer Manual
1 OVERVIEW .............................................................................................1
1.6 Menus............................................................................................................................................ 2
2.8 References................................................................................................................................... 10
iii
Spectrofluorophotometer Manual
5 MENUS..................................................................................................49
6 MEASUREMENTS ................................................................................51
iv
Table Of Contents
8 CONTACTING .......................................................................................79
9 APPENDIX ............................................................................................81
v
Spectrofluorophotometer Manual
10 INDEX....................................................................................................85
vi
1 Overview
Thank you for your purchase a Shimadzu spectrofluorophotometer system. The following instruments
are supported:
• RF-5301PC
1.5 Measurements
Any measurements will be carried out from a central measurement window, where instrument
parameters and follow-up actions are configured:
Measurement Window
Before a measurement can be started, the instrument must be initialized:
Instrument Initialization
1
Spectrofluorophotometer Manual
1.6 Menus
RF-5301/RF-1501 menu
1.7 Maintenance
Maintenance
Instrument Parameter
Checking Lamp Time
Lamp Adjustment
Show Instrument Status
1.8 Troubleshooting
Troubleshooting
2
2 What is Fluorescence?
The phenomenon of certain kinds of substance emitting light on absorbing various energies, without
involving heat generation, is called luminescence. That kind of luminescence that is emitted on exposure
to ultraviolet or visible rays is called photo luminescence.
Fluorescence and phosphorescence, representative of photo luminescence, possess hues different from
the reflected or transparent color a substance, emitting longer wavelengths of light than radiated light.
Familiar examples are the green color observed in a red ink in the daylight (eosin aqueous solution) and
a pale color in shirts.
The excited state, however, soon changes back to the base state as the molecule is deactivated by
radiating the energy in the form of heat or light. The molecule then transits, without radiation, to an
excited state having a slightly lower energy level than the excited state S1. The light the molecule emits
as it returns further to the base state S0 is called fluorescence.
Since part of the energy of the light absorbed has been lost as vibration or heat energy, the light covers
a longer wavelength than the light to which the molecule has originally been exposed (Stokes law).
The light the molecule emits as it transits, without radiation, to the triplet state T1 from the excited state
S1, the returns to the base state S0 is called phosphorescence. Phosphorescence has a life longer than
-4
10 seconds because of the need for spin transformation.
3
Spectrofluorophotometer Manual
• Law 2
Generally, fluorescence has a longer wavelength than excitation light.
• Law 3
The quantum yield of fluorescence (Q)is determined by the frequency of non-radiation of the
absorption energy to heat or others.
where
ne: Frequency of light emission
nf: Frequency of non-radiation transition
Law 1 dictates that the task of testing an unknown sample begin with measurement of its absorption
spectrum with a relatively higher concentration. If there is no absorption at all, then it is considered that
fluorescence is not emitted even if the sample is excited at the wavelength. Conversely, fluorescence is
emitted most intensely if the sample is excited with the absorption peak wavelength.
Law 2 indicates that, since part of the energy is reduced, shifting the wavelength of the fluorescence to
longer side. Hence, the task of measuring the fluorescence spectrum can be reduced to a matter of
scanning only the longer wavelength side of the excitation light.
Law 3 uses the quantum yield of fluorescence (Q) to indicate what proportion of energy absorbed is
radiated as fluorescence. The higher the value of Q, the easier the substance produces fluorescence.
The following table lists the quantum yields of typical fluorescent substances:
4
Principles of Fluorescence
5
Spectrofluorophotometer Manual
6
Principles of Fluorescence
7
Spectrofluorophotometer Manual
8
Principles of Fluorescence
where
*
db (λ ): Intensity of the fluorescence observed at wavelength
n: Refractive index
p: Reflection coefficient
Fλ (λ') : True fluorescence intensity at wavelength in the spectrum emitted by the excitation light
at wavelength
B(λ') = K C
Hence, the calibration curve that is proportional to the concentration C is straight. As the concentration
of the sample rises, however, the calibration curve would be curved as explained in "High-concentration
samples" section. The following figure shows a calibration curve observed with a diaminostilbene
aqueous solution:
9
Spectrofluorophotometer Manual
2.8 References
Applied Engineering Division, Shimadzu Corporation. Principles, Application, Application, and
Equipment Structure of Fluorescence Analysis, Shimadzu Fluorescence Analysis Course, Shimadzu
Corporation.
10
3 Installation of the spectrofluorophotometer
11
Spectrofluorophotometer Manual
Danger: The positive electrode of the Xenon lamp is charged to a voltage as high as
30000 V when the lamp is turned ON. The lamp is energized to about 90 V and it
reaches a temperature of 100 °C or more after it has been lit. Make sure that the
cover of the lamp housing is installed in position before operating the
instrument!
Before removing the cover of the lamp housing be sure to disconnect the power
cable from the outlet!
Remember that even after the power switch is set to the OFF position, the lamp
unit remains electrically charged and can cause electric shock. Disconnect the
power cable from the power outlet, set the power switch to the OFF position
and leave the instrument for five minutes or longer. Only then remove the cover
of the lamp housing!
Danger: The Xenon lamp contains high pressure gas. Impact, vibration or pressure on it
may cause it to burst. Be extremely careful with handling it!
12
Installation of the spectrofluorophotometer
Warning: Be careful not to spill water, organic solvent, etc. on the instrument. Spilling
liquid on the instrument can cause electric shock, fire, damage or malfunction
of the instrument.
Caution: While installing the lamp do not touch the bulb with bar fingers. Finger oil
remaining on the bulb is be baked onto the bulb when the lamp is lit. That can
cause the lamp to burst! Cleaning of the lamp is possible with ethyl alcohol or
the special cleaning agent supplied with the lamp.
Caution: When relocating or shipping the instrument be sure to remove the Xenon lamp!
Store the removed lamp in the special case supplied with the lamp.
Caution: Replace the lamp immediately when ist total service life has exceeded 500
hours (guaranteed service life). Never use a lamp more than 1000 hours! A
lamp used more than 1000 hours is liable to burst. A damage of the lamp unit is
possible.
13
Spectrofluorophotometer Manual
14
Installation of the spectrofluorophotometer
Right-side View
(1) Screw for horizontal positioning of lamp. For adjusting the horizontal position of the lamp.
(2) Screw for vertical positioning of lamp. For adjusting the vertical position of the lamp.
(3) Lamp fixing screw. Is tightened or loosened for installing or removing the Xenon lamp using the
supplied screwdriver which is inserted through the guide tube.
(4) Power switch. Powers the instrument ON/OFF.
(5) Lamp ON/OFF switch. Usually, keep this switch in the ON position.
(6) Voltage selector 1. Must be set according to the site power supply (100/120/220/240 VAC).
(7) Voltage selector 2. Must be set according to the site power supply (100/200 VAC).
(8) Parallel I/F connector. Not used.
(9) RS-232C I/F connector. Used for communications with the computer.
(10) I/O-1 connector. For connecting the optional auto-sampler.
(11) I/O-2 connector. For connecting the optional sipper.
15
Spectrofluorophotometer Manual
Left-side View
(1) Analog OUT terminals. Outputs voltage (1 V/FS) in proportion to the fluorescence intensity. The
Chromatopack or other output devices can be connected here.
(2) Contact IN terminals. Analysis operations can be started by jumpering these terminals.
Rear View
16
Installation of the spectrofluorophotometer
17
Spectrofluorophotometer Manual
18
Installation of the spectrofluorophotometer
19
Spectrofluorophotometer Manual
Rear View
20
Installation of the spectrofluorophotometer
Set an optional interface board or output board into this slot. It is normally covered.
Item Description
Wavelength scan range 220 - 900 nm and zero-order light (EX and EM)
Measuring wavelength range 220 - 750 nm and zero-order light (EX and EM)
Wavelength scan speed SURVEY, SUPER, VERY FAST, FAST, MEDIUM, SLOW,
VERY SLOW
Light source compensation system Monochromatic light monitoring diode feedback system
21
Spectrofluorophotometer Manual
S/N ratio 150 or greater at the water Raman peak of distilled water,
spectral bandwidth 5 nm for both EX and EM, EX 350 nm
Weight Photometer: 43 kg
Item Description
Wavelength scan range 220 - 900 nm and zero-order light (EX and EM)
Measuring wavelength range 220 - 750 nm and zero-order light (EX and EM)
Wavelength accuracy ± 5 nm
Wavelength scan speed Selectable from among SUPER (about 3700 nm/min), FAST,
MEDIUM, and SLOW
22
Installation of the spectrofluorophotometer
Light source compensation system Dynode feedback method by monitored excitation light.
Response 0-98%
0.02, 0.03, 0.1, 0.25, 0.5, 2 and 8 sec
S/N ratio The S/N ratio at the Raman peak of distilled water is 300 or
higher under the measurement conditions:
Excitation wavelength : 350 nm
Emission wavelength : Peak wavelength of the Raman line
Bandwidths : 10 nm for both excitation and emission
Response : 2 sec
• Differential spectra
23
Spectrofluorophotometer Manual
Storage of measurement files Three files stored in the unit's internal memory and 60 per
RAM card
Weight 23 kg
3.5.1 Location
To provide optimal performance of the instrument and to ensure a long trouble-free operating life, be
sure to install the instrument in a location that satisfies the requirements listed below.
• A position not subject to strong vibration or any continuous (even weak) vibration.
• Relative humidity within the range 45 % to 70 %. (If the room temperature is 30 °C or higher
the relative humidity must be more than 70 %.)
• A location free from exposure to corrosive gas or any organic or inorganic gas that has an
absorption band in the UV region.
• Place the spectrofluorophotometer on a table or surface which can withstand the load of the
spectrofluorophotometer (weight of RF 5301PC: 43 kg; weight of RF-1501: 23 kg).
24
Installation of the spectrofluorophotometer
Warning: Be careful not to spill water, organic solvent, etc. on the instrument. Spilling
liquid on the instrument can cause electric shock, fire, damage or malfunction
of the instrument.
25
Spectrofluorophotometer Manual
26
Installation of the spectrofluorophotometer
27
Spectrofluorophotometer Manual
28
Installation of the spectrofluorophotometer
Caution: Do not confuse the positive side of the lamp with the negative side when
installing the lamp. If the instrument is powered ON with the lamp installed in
the wrong direction it will be damaged.
29
Spectrofluorophotometer Manual
Caution: If the positive electrode side is subjected to a pull-force by the power cord
while the lamp is lit, the lamp may be damaged. When securing the terminal of
the positive electrode with the knurled screw make sure that the power cord at
the positive side is a little slack.
30
Installation of the spectrofluorophotometer
• Protect the cards against intense magnetic fields or electrostatic effects. The
stored programs might be destroyed.
• Spectrum mode
in which the excitation or emission spectrum of a sample is measured by wavelength scanning.
31
Spectrofluorophotometer Manual
• Sipper
Please refer to the section "Working with the Sipper Accessory" for details.
• Auto-Sampler
Please refer to the section "Working with the Auto-Sampler Accessory" for details.
• Constant temperature cell holder (currently not supported by the panorama fluorescence
software!)
Install the external accessory according to the steps described in the particular accessory sub-section.
As long as you run the instrument with external accessories together with the panorama fluorescence
software, there is no need for a special setup of the instrument. The software will take over full control
over all accessories and parameter settings.
In case you like to run the RF-1501 unit as a stand-alone instrument, the external accessories need to
be installed separately via the on-screen display. Please refer to your RF-1501 operation instructions
manual, chapter 4.7 for details.
32
Installation of the spectrofluorophotometer
3.5.6 Powering ON
After installing the Xenon lamp and optional IC cards or accessories power ON the instrument. Before
setting the power switch to the ON position, make sure of the following aspects.
1. Grounding
The power cable for the instrument is a three-wire type including a ground wire. Connect the
cable to a three-wire type power outlet, and provide reliable grounding. If the power outlets
available on-site are two-wire type, make sure to provide grounding using the grounding
terminal on the power cable, ground the instrument directly.
2. Connecting the power cable
First, be sure that the power switch is in the OFF position. Then insert the attached power
cable into the power inlet located on the right-side face of the instrument. Make sure that the
setting of the AC voltage selector conforms to the voltage of the site power.
3. Xenon lamp ON/OFF switch
Set the Xenon lamp ON/OFF switch to the ON position.
If all items are satisfied please continue as follows:
1. Set the power switch to the ON position. The Xenon lamp will turn ON automatically.
2. After powering ON the instrument power ON the computer and start MS-Windows.
3. Start the panorama fluorescence software package.
33
Spectrofluorophotometer Manual
34
Installation of the spectrofluorophotometer
5. Click the Read button to update the actual intensity value (optional).
6. Click the Set button to apply actual values.
7. From the RF-5301/RF-1501 menu, select the Lamp Adjustment command.
The instrument shutter is opened and the following dialog is shown:
The lamp time reset function is not available for the RF-1501 Instrument!
The current lamp time is only displayed for the RF-1501 instrument.
35
Spectrofluorophotometer Manual
10. Adjust the screw for horizontal positioning of the lamp so that the beam spot on the white
paper is of maximum brightness.
Caution: Be sure to adjust the horizontal position of the lamp first. Adjusting the
vertical position first may cause the light beam to deviate and make it
impossible to correct the lamp position.
36
Installation of the spectrofluorophotometer
11. Similarly to the horizontal position adjust the screw for vertical positioning of the lamp.
2. Adjust the screw for horizontal positioning of the lamp until the energy value indicates a
maximum.
3. Adjust the screw for vertical positioning of the lamp until the energy value indicates a
maximum.
4. If a new lamp was installed, click the Reset button to reset the lamp timer (optional).
5. Click the Close button to finish the procedure.
37
4 Performance Tests
4.1.1 Prerequisites
A 2D Emission measurement is required in advance to check experimental conditions with the sample.
Please follow the instructions below:
1. Start the panorama fluorescence software and make sure, the instrument has warmed up for
at least 30 minutes.
2. From the RF-5301/RF-1501 menu, select the 2D Emission Measurement command. The
measurement dialog is opened showing the last measured spectrum and last used parameter
settings.
3. In the measurement dialog, set the following parameters:
• Sensitivity: high
39
Spectrofluorophotometer Manual
9. Click the button on top right of the measurement dialog to close it.
The measured 2D emission spectrum is now shown in the main workspace.
10. From the Mathematics menu, select the Find Peaks command.
11. In the Mathematics tab on bottom right, adjust the parameters and click the Calculate
button.
The peaks should be detected automatically and a peak table will be created.
40
Performance Tests
41
Spectrofluorophotometer Manual
42
Performance Tests
2. In the dialog, click the Perform button next to the S/N Ratio Check label.
The Signal Noise Performance Test dialog is opened:
43
Spectrofluorophotometer Manual
44
Performance Tests
2. From the RF-5301/RF-1501 menu, select the 2D Emission Measurement command. The
measurement dialog is opened showing the last measured spectrum and last used parameter
settings:
• Sensitivity: high
• Sampling interval:
RF-5301 PC: 0.2 nm
RF-1501: 1 nm
45
Spectrofluorophotometer Manual
5. After scanning has stopped, check that the resultant peak in the spectrum is within the range
of 435.8 ± 1.5 nm.
6. Click the button on top right of the measurement dialog to close it.
The measured 2D Emission spectrum is shown in the main workspace now.
7. From the Mathematics menu, select the Find Peaks command.
8. In the Mathematics tab on bottom right, adjust the parameters and click the Calculate
button.
The peak should be detected automatically and a peak table will be created.
46
Performance Tests
• Sensitivity: high
47
Spectrofluorophotometer Manual
5. Load a cell filled with distilled water into the cell holder.
6. Close the lid of the sample compartment.
7. Click the Measure button to start scanning.
8. After scanning has stopped, check that the resultant peak in the spectrum is within the range
of 435.8 ± 1.5 nm.
48
5 Menus
• 2D Emission Measurement
• 3D Emission Measurement
• 2D Excitation Measurement
• 3D Excitation Measurement
• 2D Synchro Measurement
49
Spectrofluorophotometer Manual
• Time Measurement
• Photometric Measurement
• Go to Wavelength
• Instrument Parameter
• Lamp Adjustment
• Accessories
• Sipper
50
6 Measurements
The contents of the measurement window are described in more detail in the following:
To show the measurement window, select one of the measurement procedures from the RF-5301/RF-
1501 menu. It looks like this:
51
Spectrofluorophotometer Manual
Abort Button
To abort the current measurement, click this button.
Autozero Button
While working with the instrument adjustment of the photomultiplier must be re-calibrated from time to
time to obtain a constant baseline for spectra.
Click this button to reset the baseline.
Shutter Button
The shutter protects the photomultiplier from permanent exposure to light in order to prevent damage.
Usually, the shutter is controlled by the measurement procedures automatically. For maintenance
purposes it might be useful to control the shutter manually.
Click this button to open or close the shutter manually. The image inside the button shows the actual
shutter condition.
Go to WL Button
Please refer to the Go to Wavelength section for details.
Close Button
Closes the measurement window and returns to the main software.
Reset Button
This button is only available with the photometric measurement. By default, results of photometric
measurements will be appended to the current report on the left pane of the measurement window.
Click Reset to clear the current report and recent measurements and restart with a new report.
52
Measurements
Instrument Parameters
The following parameters are available:
• Sensitivity
This parameter controls the receiver gain. It can be adjusted to High or Low sensitivity. The
high setting is about 50 times more sensitive than the low setting.
Online Display
The following parameters are available:
• Autoscale Display
This parameter controls the display of the 2D and 3D online monitors on the left of the dialog.
On Yes, spectra will be automatically scaled after completion of measurement.
Sample Information
The following parameters are available:
• Folder
This parameter specifies the destination folder in the panorama project, where measured data
are stored. If the folder does not exist, it will be created automatically.
• Project
This parameter specifies the destination panorama project, where measured data are stored. If
the project does not exist, it will be created automatically. For details on how to work with
projects please refer to the Projects and Project Explorer section of the panorama help
manual.
• Sample Name
In the text field the name of the measured spectrum must be entered. In order to allow
automatic naming, some wildcards can be used as described in the following:
• %c
This wildcard specifies a counter starting with 0. It will be increased automatically.
• %d
This wildcard specifies the actual date. The date format of the regional settings of the
operating system is applied.
• %t
This wildcard specifies the actual time. The time format of the regional settings of the
operating system is applied.
53
Spectrofluorophotometer Manual
Scan Parameters
The following parameters are available:
• Response Time
This corresponds to the response speed of the RF-5301 unit relative to the variation in the
fluorescence intensity on a particular sample. The lower the setting (in seconds) the more
swiftly the instrument can follow variations in the fluorescence intensity with time, although the
resulting noise level will be greater. In contrast, the higher the setting, the less rapidly the
instrument can follow variations, but the lower the noise. The allowed settings are
RF-5301 PC: 0.02, 0.03, 0.1, 0.25, 0.5, 2.0, 4.0, 8.0 and Auto.
RF-1501: 0.02, 0.03, 0.1, 0.25, 0.5, 2.0 and 8.0.
• Sampling Interval
Sets up the wavelength intervals for scanning. The allowed settings are
RF-5301 PC: 0.2 nm, 1.0 nm and 2.0 nm.
RF-1501: 1.0 nm and 2.0 nm.
Limitation of settings!
However, that the scanning interval of 2.0 nm is automatically selected only
when the scan speed is set to "Survey" and cannot be used for other speed
settings. Also, the setting of 0.2 nm cannot be selected for speed setting
"Survey", "Super" and "Very Fast".
• Scanning Speed
This parameter specifies the scanning speed for data acquisition. The seven settings are
RF-5301 PC: Survey, Super, Very Fast, Fast, Medium, Slow and Very Slow.
RF-1501: Super, Fast, Medium and Slow.
Sipper Parameters
For a general introduction into functionality of the sipper unit and how to use it with the instrument,
please refer to the section "Working with the Sipper Accessory". The following parameters are available,
which control the sipper accessory functions:
• Use Sipper
Specifies, whether the sipper accessory is enabled and used during measurement or not.
Parameters will be expanded or collapsed in the measurement window according to the
settings.
The parameter can be also adjusted from the Accessories sub menu in the RF-5301/RF-1501
menu.
• Yes
Make sure the sipper accessory is connected to the instrument before you change the
flag status to Yes.
54
Measurements
• No
The sipper accessory is not used.
• Pump Speed
This parameter controls the pump speed of the built-in sipper pump. The following pump speed
levels are applicable:
• Slow
• Medium
• Fast
• Sipping Time
Specifies the pumping duration in seconds.
• Dwell Time
Specifies the rest period after pumping and before performing the measurement. This interval
is applied to let the sample rest in the cuvette before measurement.
• Purging Time
Specifies the purging duration after completion of the measurement in seconds. The sample is
removed from the cuvette by pumping the solvent through the cuvette.
• Number of Rinses
Specifies the number of purge intervals required to clean the cuvette.
• Use Auto-Sampler
• Yes
The auto-sampler is enabled
• No
The auto-sampler is disabled
• Number of Samples
Denotes the number of samples to be measured and provided in the auto-sampler.
55
Spectrofluorophotometer Manual
1. From the RF-5301/RF-1501 menu, select the 2D Emission Measurement command. The
measurement dialog is opened showing the last measured spectrum and last used parameter
settings:
56
Measurements
11. Click the button on top right of the measurement dialog to close it.
The measured spectrum is now shown in the main workspace and a new spectrum node has
been added to the destination project.
57
Spectrofluorophotometer Manual
7. Click the button on top right of the measurement dialog to close it.
The measured spectrum is now shown in the main workspace and a new spectrum node has
been added to the destination project.
58
Measurements
1. From the RF-5301/RF1501 menu, select the 2D Excitation Measurement command. The
measurement dialog is opened showing the last measured spectrum and last used parameter
settings:
2. Set the Start Wavelength to an appropriate start value of the detection range in [nm].
3. Set the Stop Wavelength to an appropriate end value of the detection range in [nm].
4. Set the other parameters in the dialog optionally. For details on general parameter settings,
please refer to the Measurement Window section.
5. Open up the lid of the sample compartment
6. Load a cell filled with a sample into the cell holder.
7. Close the lid of the sample compartment.
8. Click the Autozero button.
9. Click the Measure button to start scanning.
10. During measurement the user can see evolving the spectrum.
11. Click the button on top right of the measurement dialog to close it.
The measured spectrum is now shown in the main workspace and a new spectrum node has
been added to the destination project.
59
Spectrofluorophotometer Manual
60
Measurements
• 3D View
• Top View
• 2D Overlay
12. Click the button on top right of the measurement dialog to close it.
The measured spectrum is now shown in the main workspace and a new spectrum node has
been added to the destination project.
61
Spectrofluorophotometer Manual
1. From the RF-5301/RF1501 menu, select the 3D Emission Measurement command. The
measurement dialog is opened showing the last measured spectrum and last used parameter
settings:
• Acquisition Rate
• Total time.
6. Set the other parameters in the dialog optionally. For details on general parameter settings,
please refer to the Measurement Window section.
7. Open up the lid of the sample compartment
8. Load a cell filled with a sample into the cell holder.
9. Close the lid of the sample compartment.
10. Click the Autozero button.
11. Click the Measure button to start scanning.
During measurement the user can see evolving the 2D spectra and the 3D spectrum.
62
Measurements
• 3D View
• Top View
• 2D Overlay
12. Click the button on top right of the measurement dialog to close it.
The measured spectrum is now shown in the main workspace and a new spectrum node has
been added to the destination project.
63
Spectrofluorophotometer Manual
• 3D View
• Top View
• 2D Overlay
8. Click the button on top right of the measurement dialog to close it.
The measured spectrum is now shown in the main workspace and a new spectrum node has
been added to the destination project.
64
Measurements
1. From the RF-5301/RF-1501 menu, select the 3D Excitation Measurement command. The
measurement dialog is opened showing the last measured spectrum and last used parameter
settings:
• Acquisition Rate
• Total time.
6. Set the other parameters in the dialog optionally. For details on general parameter settings,
please refer to the Measurement Window section.
7. Open up the lid of the sample compartment
8. Load a cell filled with a sample into the cell holder.
9. Close the lid of the sample compartment.
10. Click the Autozero button.
11. Click the Measure button to start scanning.
During measurement the user can see evolving the 2D spectra and the 3D spectrum.
65
Spectrofluorophotometer Manual
• 3D View
• Top View
• 2D Overlay
12. Click the button on top right of the measurement dialog to close it.
The measured spectrum is now shown in the main workspace and a new spectrum node has
been added to the destination project.
6.9 Go to Wavelength
The Go to Wavelength function is used to adjust a particular excitation and emission wavelength
directly. By adjusting the pair of parameters, the response signal can be directly detected. This
procedure is useful for single value detection or instrument maintenance.
To set particular excitation and emission wavelengths, please follow the instructions below:
1. From the RF-5301/RF-1501 menu, select the Go to Wavelength command. The Go to
Wavelength dialog is opened as shown in the following:
2. Set the Excitation Wavelength to an appropriate value in [nm] or move the slider to adjust.
3. Set the Emission Wavelength to an appropriate value in [nm] or move the slider to adjust.
4. Click the Read button to update the actual intensity value (optional).
5. Click the Set button to apply actual values and close the dialog.
Setting a wavelength...
... will be applied instantly! The instrument will move to the selected excitation
and emission wavelength and will keep this settings until new positions are
applied by any other measurement procedure.
66
Measurements
6. Click the Cancel button to discard changes and close the dialog.
• Set the Emission Wavelength to an appropriate value of the detection range in [nm].
6. Set the other parameters in the dialog optionally. For details on general parameter settings,
please refer to the Measurement Window section.
7. Open up the lid of the sample compartment.
8. Load a cell filled with a sample into the cell holder.
9. Close the lid of the sample compartment.
10. Click the Autozero button.
11. Click the Measure button to start scanning.
67
Spectrofluorophotometer Manual
After completion of a measurement, the trend plot and report are updated accordingly.
• Acquisition Rate
• Total time.
5. Set the other parameters in the dialog optionally. For details on general parameter settings,
please refer to the Measurement Window section.
6. Open up the lid of the sample compartment
7. Load a cell filled with a liquid sample into the cell holder.
8. Close the lid of the sample compartment.
9. Click the Measure button to start scanning.
During measurement the user can see the emission evolving.
68
7 Hardware and Maintenance
7.1 Maintenance
Caution: Never use a lamp in excess of 1000 hours. A lamp used for more than 1000
hours may burst, possibly damaging the lamp unit.
69
Spectrofluorophotometer Manual
70
Hardware and Maintenance
2. Enter the serial number of the RF-1501 instrument, which is located on a label on the right
side of the instrument. (RF-1501 System only!)
Caution: Entering the serial number is only required for the RF-1501
instrument!
With the RF-5301PC system, the serial number is available
automatically.
• On (default)
Automatic regulation of Lamp voltage is applied.
• Off
No regulation of Lamp voltage is applied.
PMT Control
This parameter controls the sensitivity of the photomultiplier tube (PMT) with regard to the amount of
light exposure. By default the sensitivity of the PMT will be controlled automatically. For service and
maintenance purposes automatic regulation can be deactivated.
• On (default)
PMT sensitivity is regulated automatically.
• Off
The PMT sensitivity is not regulated.
Auto Shutter
This parameter controls the shutter which protects the photomultiplier from permanent irradiation with
light. By default, the auto shutter is on to avoid damages to the photomultiplier. For some dark or
background measurements carried out by service persons the shutter must be controlled manually.
• On (default)
The shutter opens and closes automatically during measurements.
• Off
The shutter must be opened and closed manually before starting measurements.
71
Spectrofluorophotometer Manual
• Lamp Alignment
Please review the Lamp Adjustment section in the chapter "Hardware and Maintenance" for
details.
Click the Perform button next to the desired operation to start it.
The lamp time reset function is not available for the RF-1501 Instrument!
The current lamp time is only displayed for the RF-1501 instrument.
72
Hardware and Maintenance
• Instrument name
• Serial number
73
Spectrofluorophotometer Manual
If no valid serial number is entered, you will be prompted by the following message when you try
to close the dialog:
• ROM version
• Serial port
74
Hardware and Maintenance
For details on remedial actions please follow the instructions in the error message or refer to the RF-
5301 troubleshooting section of this manual.
Click the Close button to finish.
7.5 Troubleshooting
Cable connection Check the cable connections between the instrument and the computer.
Next, try again to establish a communication.
COM port setting Study the documentation for your computer to check that the COM port in
question has been addressed correctly.
If a normal communication cannot be established in spite of the above remedial actions contact
Shimadzu or its nearest representative.
Instrument is Power cable is Securely insert both end of the cable into the inlet on the
not powered not securely instrument and the outlet at the site.
although the connected.
power switch is
in the ON
position.
75
Spectrofluorophotometer Manual
Lamp is still Allow the lamp to cool down for about 10 minutes.
hot.
S/N ratio does Xenon lamp is After powering ON the instrument wait 30 minutes until the
not satisfy the not stably lit. Xenon lamp is stably lit.
guaranteed
value.
76
Hardware and Maintenance
77
Spectrofluorophotometer Manual
Light emitted from the xenon lamp is concentrated in the entrance slit S1 at the excitation
monochromator by the light-concentrating ellipsoidal mirror M1. Light passing the slit S1 is reflected by
the plane mirror M2 and enters the excitation grating G1 before it is dispersed to enter the exit slit S2.
Light emitted from S2 (excitation light) is concentrated in the sample cells through the lenses L1 and L2.
Part of the excitation light is dispersed by the beam splitter 82 and enters the monitor photomultiplier
PM1 for light source compensation.
Fluorescence passing the slit S3 is dispersed by the emission grating and, passing through the exit slit
S4, enters the fluorescence photomultiplier PM2 for fluorescence intensity measurement.
78
8 Contacting
Albert-Hahn-Str. 6-10
D-47269 Duisburg
Address:
Telephone:
Telefax:
e-mail:
Further headquarters are located around the world. Please find the nearest headquarter in the internet
at
http://eu.shimadzu.de/company/international
79
9 Appendix
• Yes
The auto-sampler is enabled
• No
The auto-sampler is disabled
All samples will be processed and measured one after the other automatically. On any error the system
will stop and an error message is displayed.
81
Spectrofluorophotometer Manual
Choose the automatic sample naming to avoid intermediate breaks and messages!
In the measurement Window setup the following parameters correctly to avoid interruptions:
1. In the Sample Information category set the flag Show Sampler Name Dialog to
No.
2. Setup Sample Name parameter.
For details, please refer to the section Sample Information in the chapter Measurement
Window.
• Checked
The sipper is enabled
• Unchecked
The sipper is disabled
Alternatively, the sipper can be enabled or disabled directly in the Measurement Window. The sipper
unit has a few parameters that can be adjusted in the software as described in the Measurement
Window section.
82
Appendix
• Number of Rinses = 1
4. Click the Sip and Measure button to start the measurement.
5. Put the tube to the ground of the reservoir to avoid bubbles being pumped into the cuvette
chamber and confirm the message box with OK.
The measurement progress and steps are shown in the status dialog:
1. The sample is pumped into the cell.
The following status dialog is shown:
Aborting measurements.
Whenever you like to abort the current measurement, click the Abort
button in the status dialog.
Abort takes some time to complete already started operations. This may
take up to 30 seconds.
The sample remains in the cuvette chamber on abort. It will not be removed
automatically!
2. Pumping stops and the instrument is idle for the specified dwell time in order to let the sample
rest for a while.
83
Spectrofluorophotometer Manual
84
10 Index
Coarse-adjustment 33
2 COM 72
Contact IN 14
2D
Cyclohexane 3
monochromator 44
Cylce 63
2D 39, 44
2D 44
2D 55 D
2D 57
Damages 69
2D 58
Dark Level Correction 69
2D 59
Determined S/N 39
2D 61
Diaminostilbene 3
2D 63
2D 64
2D Emission 39, 44 E
2D Emission Measurement 39, 44, 55
EM 21
2D Excitation
Emission 44
monochromator 44
Emission Start Wavelength 55, 59, 61
2D Excitation 44
Emission Stop Wavelength 55, 59, 61
2D Excitation Measurement 44, 57, 58
Emission Wavelength 33, 57, 63, 64, 66, 68
2D Synchro Measurement 58
Emission/Excitation monochromator 21
EX 21
3 Excitation Start Wavelength 57, 63, 64
Excitation Stop Wavelength 57, 63, 64
3D 59, 61, 63, 64
Excitation Wavelength 33, 55, 59, 61, 66, 68
3D Emission Increment Measurement 59
Excitation-side 77
3D Emission Measurement 61
3D Excitation Increment Measurement 63
3D Excitation Measurement 64 F
Face
A RF-5301PC 24
File menu 39, 44
Absorptiometry 3
Find Peaks 39, 44
Acquisition Rate 61, 64, 68
Fine-adjustment 33
Acquisitions
Fluorescein 3
number 61
Fluorescence 3
Acquisitions 61
Fluorescence Analysis 3
Acquisitions 64
Acquisitions 68
Adjust G
photomultiplier 69
Germany 79
Adjust 69
Grating 3
After coarse-adjustment 33
Analog OUT 14
Anthracene 3, 63 H
Anthracene-9-carboxylic 57
Hardware 69
Aspirin 3
Hardware Specification 21
Autozero button 39, 55, 57, 58, 59, 61, 63, 64
High/low 21
C
I
Calculate button 39, 44
Illuminates 77
Cancel button 66
Indole 3
Carbon Tetrachloride 3
Inspection 11
Chromatopack 14
Installation
Close button 33, 69, 72
RF-5301PC 24
Coarse-adjustment
Installation 24
Xenon 33
85
Spectrofluorophotometer Manual
Installing set 12
Xenon 24, 33 switch 12
Installing 33 OFF 12
Instrument Calibration 69 OFF 14
Instrument Construction 77 OFF 75
Instrument Initialization 33 OK button 39
Instrument Initialization Failure 33 ON
Instrument Parameters 39, 69 powering 24, 33, 75
Instrument Settings 69 ON 12, 14
ON 24
ON 33
J
ON 75
Jumpering 14 ON/OFF 14, 24, 69
L P
Lamp 69 P/N 071-60814-01 11
Lamp Adjustment 33, 72 P/N 072-01663-14 11
Lamp Adjustment dialog 33, 72 P/N 200-81500-01 11
Lamp Alignment 69 P/N 200-94612 11
Lamp ON/OFF 14, 75 P/N 200-94613 11
Lamp Time 72 P/N 206-81601 11
Lits 12 P/N 206-83579-91 11
P/N 980-00071 11
Parallel I/F 14
M Part Names 14
Maintenance 69 Part Number 11
Mathematics 39, 44 Perform button 39, 69
Mathematics menu 39, 44 Performance Tests 69
Measure button 39, 44, 55, 57, 58, 59, 61, 63, 64, 68 Phosphorescence 3
Measurement Mode 55, 57, 58, 59, 61, 63, 64 Photometry 21
Misaligned 75 Photomultiplier
Monitor 21 adjust 69
Monochromator damages 69
2D 44 Photomultiplier 21
2D Excitation 44 Photomultiplier 69
Monochromator 3, 33 Photomultiplier 77
Monochromator 44 Place
Monochromator 77 spectrofluorophotometer 24
Monochromators 44 PMT 69
Position
fine-adjustment 33
N Position 33
Naphthaline 63 Powering
Napthalene 3 ON 24, 33, 75
Neon Lamp 44 Powering 24
Nm 33, 61, 64, 66 Powering 33
Noise report layout Powering 75
Signal 44 Principles, Application 3
Noise report layout 44 Print 39, 44
Noise Test Procedure Purchase
Signal 39 Shimadzu spectrofluorophotometer 1
Noise Test Procedure 39
Number R
Acquisitions 61
Number 61 R212-14 21
R3788-02 21
RAMAN 3, 21, 33
O Rayleigh 3
Occuring 77 Read button 33, 66
OFF Remove
86
Index
87