IT6900 User Manual-EN
IT6900 User Manual-EN
Supply
IT6900 Series User Manual
Model: IT6922A/IT6932A/IT6933A/IT6942A/
IT6952A/IT6953A/IT6922B/IT6932B/
IT6933B/IT6942B/IT6952B/IT6953B
Version: V5.8
Statement Guarantee
© Itech Electronic, Co., Ltd. 2022 Materials in the document are provided Safety Statement
No part of this manual may be talis qualis and may be changed in
reproduced in any form or by any means future revisions without prior notice. In
(including electronic storage and addition, within the maximum allowable “Caution” signs indicate danger. It is
retrieval or translation into a foreign extent of laws, ITECH is not committed required to pay attention to the contents
language) without prior permission and to any explicit or implied guarantee for of these signs during implementation of
written consent from Itech Electronic, this manual and all information therein, operations.
Co., Ltd. as governed by international including but not limited to the implied The damage to the product or loss of
guarantee on marketability and important data may be caused in case of
copyright laws.
availability for some special purposes. improper operation steps or failure to
Manual Article No. ITECH shall not be responsible for any follow operation steps. Do not continue
error or incidental or indirect losses to implement any improper operation
IT6900-402178 caused by the provision, use or indicated in “Caution” signs when the
application of this documents and specified conditions are not fully
Revision
information therein. If some guarantee understood or these conditions are not
Revision 5, published on clauses in other written agreements satisfied.
Nov. 22th 2022 between ITECH and users are not
consistent with clauses herein, those
Itech Electronic, Co., Ltd. clauses in other written agreements
Trademark Statement shall prevail.
Technology license “Warning” indicates danger. It is required
Pentium is a registered trademark of to pay attention to the contents of these
Intel Corporation in the United States. Hardware and/or software in this signs during implementation of
document cannot be provided without a operation steps. Personal casualties
Microsoft, Visual Studio, Windows and license and can only be used or copied may be caused in case of improper
MS Windows are trademarks of according to the license. operation steps or failure to follow these
Microsoft Corporation in the United operation steps. Do not continue to
States and/or other countries/regions. Restricted permission statement implement any improper operation
indicated in “Warning” signs when the
Restricted permissions of the U.S.
specified conditions are not fully
government. Permissions for software
understood or these conditions are not
and technical data which are authorized
satisfied.
to the U.S. Government only include
those for custom provision to end users.
ITECH follows FAR 12.211 (technical
data), 12.212 (computer software).
DFARS 252.227-7015 (technical Note
data--commercial products) for national
defense and DFARS 227.7202-3 “Instructions” indicates operation
(permissions for commercial computer instructions. It is required to refer to the
software or computer software contents of these signs during operation
documents) while providing the steps. These signs are used for
customized business licenses of providing tips or supplementary
software and technical data. information for operators.
IT6900 User Manual
Warranty Service
ITECH Company will provide one-year warranty services for the product
materials and manufacturing (excluding the following limitations).
When warranty service or repair is needed, please send the product to the
service unit specified by ITECH Company.
When the product is sent to ITECH Company for warranty service, the
customer must pay the one-way freight to the maintenance department of
ITECH, and ITECH will be responsible for return freight.
If the product is sent to ITECH for warranty service from other countries,
the customer will be responsible for all the freight, duties and other taxes.
Limitation of Warranty
This Warranty will be rendered invalid in case of the following:
Damage caused by circuit installed by customer or using customer own
products or accessories;
Product which has been modified or repaired by the customer;
Damage caused by circuit installed by customer or not operating our
products under designated environment;
The product model or serial number is altered, deleted, removed or made
illegible by customer;
Damage caused by accidents, including but not limited to lightning, water,
fire, abuse or negligence.
Safety Signs
Direct current ON (power)
A chassis terminal - -
Safety Precautions
General safety precautions below must be followed in each phase of instrument
operation. In case of failure to follow these precautions or specific warnings in
other parts of the manual, violation against the safety standards related to the
design, manufacture and purpose of the instrument will occur. If the user does
not follow these precautions, ITECH will bear no responsibility arising there
from.
The power supply is provided with a three-core power cord during delivery
and should be connected to a three-core junction box. Before operation, be
sure that the power supply is well grounded.
Use electric wires of appropriate load. All loading wires should be capable
of bearing maximum short-circuit of electronic load without overheating.
Ensure the voltage fluctuation of mains supply is less than 10% of the
working voltage range in order to reduce risks of fire and electric shock.
To prevent burnout, please pay special attention to positive and negative
polarities of power supply during connection!
Do not use damaged equipment. Please check the housing before using
the equipment. Check whether the instrument is subject to cracking or is
lack of plastic. Do not operate the instrument in the environment with
explosive gas, steam or dust.
Observe all tags on the equipment before connection.
Do not install alternative parts on the instrument or perform any
unauthorized modification.
Do not use the equipment when the removable cover is dismantled or
loose.
Please use the power adapter supplied by the manufacturer to avoid
accidental injury.
We do not accept responsibility for any direct or indirect financial damage
or loss of profit that might occur when using the instrument.
This instrument is used for industrial purposes. Do not apply this product to
IT power supply system.
Do not use the equipment on the life support system or other equipment
with safety requirements.
Environmental Conditions
The IT6900 series power supply can only be used indoors or in low
condensation areas. The following table shows general environmental
requirements for this instrument.
Environmental conditions Requirement
Operating temperature 0°C - 40°C
Operating humidity 20% - 80% (non condensing)
Storage temperature -20°C - 70 °C
Altitude Operating up to 2,000 meters
Pollution Grade 2 pollution
Installation category
Ⅱ
Note
To make accurate measurements, allow the instrument to warm up for 30 min before
operation.
Regulation Tag
The CE tag shows that the product complies with the
provisions of all relevant European laws (if the year is
shown, it indicates that the year when the design is
approved).
The UKCA tag shows that the product complies with the
provisions of all relevant United Kingdom laws (if the year
is shown, it indicates that the year when the design is
approved).
Compliance Information
Complies with the essential requirements of the following applicable European
Directives, and carries the CE marking accordingly:
Electromagnetic Compatibility (EMC) Directive 2014/30/EU
Low-Voltage Directive (Safety) 2014/35/EU
EMC Standard
IEC 61326-1:2012/ EN 61326-1:2013 ¹²³
Reference Standards
CISPR 11:2015+A1:2016 Ed 6.1
IEC 61000-3-2: 2018 RLV
IEC 61000-3-3: 2013+A1:2017
IEC 61000-4-2:2008
IEC 61000-4-3 2006+A1:2007+A2:2010/ EN 61000-4-3 A1:2008+A2:2010
IEC 61000-4-4:2012
IEC 61000-4-5:2014+A1:2017
IEC 61000-4-6:2013+cor1:2015
IEC 61000-4-11:2004+A1:2017
Safety Standard
IEC 61010-1:2010+A1:2016
Content
Certification and Quality Assurance ........................................................................................................... i
Warranty Service ........................................................................................................................................ i
Limitation of Warranty ................................................................................................................................. i
Safety Signs................................................................................................................................................ i
Safety Precautions ..................................................................................................................................... ii
Environmental Conditions .......................................................................................................................... iii
Regulation Tag ........................................................................................................................................... iii
Waste Electrical and Electronic Equipment (WEEE) Directive.................................................................. iv
Compliance Information ................................................................................................. 错误!未定义书签。
Appendix .................................................................................................................................................... 39
IT6942B/IT6952B/IT6953B
USB cable x1 - -
NOTE
After confirming that package contents are consistent and correct, please appropriately
keep package box and related contents. The package requirements should be met when
the instrument is returned to factory for repair.
IT6900 series power supply is supplied with the following optional accessories
(sold separately):
Please select appropriate space for installation based on the electronic load
size.
IT6900 series power supply different models are not the same size, the detail
size of the power supply is shown as below.
IT6922A/IT6932A/IT6942A/IT6922B/IT6932B/IT6942B Models
IT6933A/IT6933B Models
IT6952A/IT6953A/IT6952B/IT6953B Models
1. VFD display
2. Rotary knob
3. Compound key, the local switch key and power switch
4. Number keys and ESC
5. Function keys
6. UP/DOWN/LEFT and RIGHT key, to move cursor
7. Output terminals
I-set Current set key, set the output current/menu function key,
/Menu to set the relevant Parameters for the power supply.
Callback key to call up a set value of system parameters
Recall /Save already stored / storage key, to save system parameter
settings.
Meter key, to switch from value set panel and the actual
Meter /DVM output value display / voltage meter function keys, to
switch to the measure state of the voltage meter.
Enter /Trigger Enter key, to confirm the number entered and operation /
trigger button, which is used to trigger the List test.
On/Off Output on (off) keys, control power output state / keypad
/Lock lock function keys, used to lock the panel buttons.
Left and right movement keys, used to set the value, to
adjust the cursor to the specified location.
Up and down keys, used to select a item in the menu or
increase (decrease) the output voltage or current
values.
1. Cooling window
2. RS232 Communication cable interface
3. USB Communication cable interface
4. GPIB Communication cable interface (IT6900(G) series only)
5. DVM input terminal, Remote measurement terminal and the output
terminal
6. AC power socket (fuse contained)
1. Cooling window
2. RS232 Communication cable interface
3. USB Communication cable interface
4. GPIB Communication cable interface (IT6900(G) series only)
5. Remote measurement terminal and DVM input terminal
Attention: The + and - terminals here are only used for shorting
between S+ and S-, not the output terminals, so do not connect to
the DUT.
6. Output terminal
7. Fuse
8. AC power socket
1. Cooling window
2. RS232 Communication cable interface
3. USB Communication cable interface
4. GPIB Communication cable interface (IT6900(G) series only)
5. AC power socket (fuse contained)
6. Output Sync signal interface and RS485 Communication cable
interface
7. DVM input terminal, Remote measurement terminal and the output
terminal
Copyright ©ITECH Electronic Co., Ltd. 10
Quick Start
1. Cooling window
2. RS232 Communication cable interface
3. USB Communication cable interface
4. Output Sync signal interface and RS485 Communication cable
interface
5. GPIB Communication cable interface (IT6900(G) series only)
6. Analog control interface
7. Remote measurement terminal and DVM input terminal
Attention: The + and - terminals here are only used for shorting
between S+ and S-, not the output terminals, so do not connect to
the DUT.
8. Output terminal
9. Fuse
10. AC power socket
To avoid burning out, be sure to confirm that power voltage matches with
supply voltage.
Be sure to connect the main power socket to the power outlet of protective
grounding. Do not use terminal board without protective grounding. Before
operation, be sure that the power supply is well grounded.
To avoid burning out, pay attention to marks of positive and negative
polarities before wiring.
Self-test steps
Normal self-test procedures:
1. Correctly connect the power cord. Press Power key to start up.
2. After self-test, if the power supply is normal, then VFD will display the
output voltage and current status as below:
OFF
0.000V 0.0000A
Exception handling
If the power supply cannot start normally, please check and take measures by
reference to steps below.
1. Check whether the power cord is correctly connected and confirm whether
the power supply is powered.
Correct wiring of power cord => 2
Incorrect wiring of power cord => Re-connect the power cord and check
whether the exception is removed.
NOTE
V-set 0 9
The third way: press button and number key( to ) to set
voltage value.
I-set 0 9
The third way: press button and number key( to ) to set
current value.
pressing Meter button. When this button is lit, screen displays actual output
value and the indicator light “meter” will be lit on the VFD board. In other words,
when the button is not lit, the front panel displays setting value.
can use the button + Recall (Save) button or SCPI order *SAV, *RCL to
achieve this function. Saving parameters include: setting voltage/setting current
Saving method:
Press + Recall (save) button, and then input the group number you want
to save through number key board. Press enter button to confirm. If you want to
recall the saved parameters, press Recall button and corresponding group
number (number1-9).At last press enter button to confirm.
After you edit a list file, press + Enter (Trigger) to give a trigger signal.
During the running process, Enter button will be lit all the time.
NOTE
ESC
Pressing button can enable you to quit any function setting.
The range of setting voltage is from 0V to rated voltage. You can press +
I-set (Menu) button to enter the menu, then press direction key to
select >MAX VOLT item. Press Enter button to confirm. After you set the
max voltage value, the output voltage value can only be set less than the max
voltage. Our default max voltage value is the rated value.
Communication (>COMM)
Our unit has provided standard communication interfaces: RS232/USB. In this
option, you can select the communication interface according to your demands.
The range of GPIB (IT6900(G) series only) address is 0-30. Besides, we have
multi-baud rate to be chosen in RS232 mode 4800, 9600, 19200, 38400, 57600,
11.52K. Data bit is 8,Check digit have three choices: NONE, ODD, EVEN.
Before you begin to carry out communication, please make sure the configure
in our unit agrees with PC configure.
GRP1: This indicates saving power supply parameters in 1-9 groups. Press
+ Recall (Save) and the group number (1-9) can save the parameters in
corresponding groups.
NOTE
When AD speed of power supply is set to MID or FAST, the measurement accuracy is too
low to reach the middle accuracy in the specification. Only when AD speed of power
supply is set to SLOW can the middle accuracy in the specification be reached.
Reset (>RESET)
This item is used to reset all items in the menu. If you select >YES, then unit will
be restored to factory setting. If you select >NO, all setting in the menu will
remain unchanged.
Press + I-set button to enter the menu, then press direction key to
select >SYST SET option, after that please push Enter button to confirm. At
last please press direction key to select >Trig MANUAL and push Enter
button to confirm.
You can make the output change order by editing every step value of list
operation. The parameters you need to edit includes: single-step voltage,
single-step current, single-step delay time and whether to go on the next step.
Besides, you also need to set the repeat times and save list sequence file. After
you finish the editing process, at this time if you receive a trigger signal, power
supply will begin to work according to the sequence steps you’ve edited. Now
we take five steps for an example:
Operation steps:
2. VFD display >MAX VOLT, press to select >LIST SET, press Enter to
confirm
3. VFD display >LIST STATE, press to select >LIST EDIT, press Enter
to confirm
4. VFD display >TIME SEC, press Enter to confirm, go to the next step,
you can also through button to select >TIME MIN time unit, press
Enter to confirm.
0 9
5. VFD display >VSET 0.000, press number key to or through
rotary knob to set voltage, after that press Enter to confirm.
0 9
6. VFD display ISET 0.0000, press number key to or rotary knob to
set the single-step current, press Enter to confirm.
0 9
7. VFD display SET 0.1, press number key to or rotary knob to set
single-step delay time, press Enter to confirm.
8. VFD display NEXT >YES, press Enter to confirm.
9. Repeat the steps from 5) to 8) and set the four steps’ voltage/current and
delay time separately. When screen display NEXT>YES in the fourth step
0 9
10. VFD display REPET 1, press number key to or rotary knob to set
the repeat times, press Enter to confirm.
11. VFD display SAVE >NO, press Enter to confirm, in this circumstance,
the list file is not saved but can run for one time, or you can press
button to select >SAVE FILE0, saving the list test file in FILE0~FILE9,
press Enter to confirm. You can recall the file in the following utilization.
12. If you do not save the list test file, VFD will display LIST EDIT; if you select
to save the test file, VFD will display SAVE DONE for three seconds, and
then display LIST EDITL.
14. VFD display LIST >OFF, press to select >LIST >ON, press Enter to
confirm. Now Enter button will be lit. This indicates that list operation
function has been opened.
Esc
15. VFD display >LIST STATE, pressing button can quit the operation.
16. Press On/Off button to open the output, press + Enter (Trigger) to
give a trigger signal.
17. If you have edited several list files, you can select LIST LOAD item to recall
Esc
the file you need. And then press to quit this operation. Press
On/Off button to open the output. Now you only need to press
+ Enter (Trigger) to give a trigger signal, the list file can be ran.
18. In LIST mode, voltage set and current set button can’t be used, In LIST
Copyright ©ITECH Electronic Co., Ltd. 23
Function and Features
STATE item, choose LIST>OFF will enable you to quit list mode.
For IT6952A/IT6953A/IT6952B/IT6953B:
NOTE
In order to ensure the stability of the system, using armored twisted pair cable between
the remote sense terminal of IT6900 and load.
Please note that the positive and negative polarity when wiring, otherwise it will damage
the instrument!
Press shift + I-set button to entry into the menu, user need to set the
parameters of Baud rate, Data bit, Stop bit, parity bit and address, the setting is
the same with RS232 communication setting.
If external analog control is not used, it must be disabled before normal front
panel operation can resume. Follow the steps below to enable or disable
external analog control.(Entry into the system menu----Enter button to
confirm-----select Ext-C item---choose ANA item and enable this mode. If
choose DIGIT, it means enable the external digital control function.)
Pin # Label Description
1 REF_10V 10V DC reference output
2 Vs/D0 Composite terminal: Voltage programming input/
Digital D0 input terminal
3 Is/D1 Composite terminal: Current programming input/
Digital D1 input terminal
4 D2 Digital D2 input terminal
5 ON/OFF Control output state
6 Vm Voltage monitoring output
7 Im Current monitoring output
8 NULL None
9 GND Ground pin
connection, you need to set the front panel composite key + I-set
NOTE
The RS232 settings must match the settings in front panel system information. If any
+ I-set key on the front panel. Parity options are stored in nonvolatile
memory.
Baud Rate
The front panel + I-set button allows the user to select a baud rate
which is stored in the non-volatile memory: 4800 /9600/ 19200 /38400/
57600 /115200.
RS-232 Troubleshooting:
If there is RS-232 connection problem, check the following:
Computer and power supply must configure the same baud rate,
parity, data bits and flow control options. Note that the power
configuration as a start bit and a stop bit (these values are fixed).
As described before in RS-232 connector, you must use the correct
interface cable or adapter. Note that even if the cable has the right
plug, the internal wiring may be wrong.
Interface cable must be connected to the correct serial port on the
computer (COM1, COM2, etc.).
Communication Settings
Before communication, you should first make the following parameters of
power supply and PC matches.
Baud Rate: 9600 (4800,9600,19200,38400,57600,115200). You can
enter the system menu from the front panel, and then set the baud rate.
Calibration (none, even, odd)
EVEN 8 data bits, have even parity
ODD 8 data bits have odd parity
NONE 8 data bits, no parity
Mode: SIGNAL, MUX
SIGNAL (Single connection mode)
MUX (Multiple connection mode) ADDR: Local address (0 to 30, the
factory default setting is 0)
can set by the function key on the front panel, press the + I-set key
to enter the system menu function, find the GPIB address setting by
▼ Enter
button, type the address, key to confirm. GPIB address is
stored in nonvolatile memory line.
On the front panel, press (Shift)+ I-set keys to enter into the system
menu, and the related setting is the same as the RS232 communication setting.
User can set the following parameters of the RS485 interface:
Baud rate: 9600(4800/9600/19200/38400/57600/115200)
Parity and data bit: NONE/8BIT, EVEN/8BIT, ODD/8BIT
EVEN 8 data bits, have even parity
ODD 8 data bits have odd parity
NONE 8 data bits, no parity
Copyright ©ITECH Electronic Co., Ltd. 30
Remote Operation Mode
Set each unit with a different Address (0 to 30). Then by using RS485, connect
the first power supply in the chain to a PC. Now, multiple units daisy-chained
via RS485 can be controlled by one PC by using the commands specific for
multi-unit connection. See “Programming Guide” section for details.
The figure below is a schematic diagram of the connection of 4 machines.
Connect PC
Parameters IT6933A
voltage 0-150V
Rated values
current 0-5A
(0 °C~40 °C)
Power 200W
voltage 220VAC ±10%/110VAC ±10%
AC Input
Frequency 47 Hz ~63Hz
Maximum input current 2A
Maximum input apparent
500VA
power
Efficiency 0.55
Power Factor 0.7
Load regulation voltage ≤0.01%+20mV
±(%of output+offset) current ≤0.01%+6mA
Line regulation voltage ≤0.01%+20mV
±(%of output+offset) current ≤0.01%+6mA
1mV(<100V),
voltage
Setup resolution 10mV(≥100V)
current 0.1mA
1mV(<100V),
voltage
Readback resolution 10mV(≥100V)
current 0.1mA
Setup accuracy voltage ≤0.04%+30mV
(within twelve months)
(25°C±5°C) current ≤0.1%+10mA
±(%of output+offset)
Read back resolution voltage ≤0.04%+30mV
(within twelve months)
(25°C±5°C) current ≤0.1%+10mA
±(%of output+offset)
voltage ≤30mVp-p
Ripple(20Hz ~20MHz)
current ≤6mArms
Rise time(No-load) voltage 200ms
Rise time(Full-load) voltage 300ms
Fall time(No-load) voltage 2s
Fall time(Full-load) voltage 150ms
Transient Response Time 200us
Remote Sense
1V
Compensation
Command Response Time 10~600ms
Difference-mode
50mV
voltage(Vpp)
Difference-mode current
20mA
(Arms)
Setup Temp.coefficient voltage 50 PPM/°C +30mV
Copyright ©ITECH Electronic Co., Ltd. 33
Technical Specification
Setup accuracy
voltage ≤0.03%+5mV ≤0.03%+5mV
(within twelve months)
(25°C±5°C)
current ≤0.1%+5mA ≤0.1%+10mA
±(%of output+offset)
Read back resolution
voltage ≤0.03%+5mV ≤0.03%+5mV
(within twelve months)
(25°C±5°C)
current ≤0.1%+5mA ≤0.1%+10mA
±(%of output+offset)
Parameters IT6933B
voltage 0~150V
Rated values
current 0~5A
(0 °C~40 °C)
Power 200W
Load regulation voltage ≤0.01%+20mV
±(%of output+offset) current ≤0.01%+6mA
1mV(<100V)
voltage
Readback resolution 10mV(≥100V)
current 0.1mA
Setup accuracy voltage ≤0.04%+30mV
(within twelve months)
(25°C±5°C) current ≤0.1%+10mA
±(%of output+offset)
Read back resolution voltage ≤0.04%+30mV
(within twelve months)
(25°C±5°C) current ≤0.1%+10mA
±(%of output+offset)
Ripple voltage ≤30mVp-p
(20Hz ~20MHz) current ≤6mArms
Transient Response
50%-100% load recovery to 75mV (Freq=1K) 200us
Time
Remote Sense
1V
Compensation
Setup
50ppm/℃+30mV
voltage
Readback
50ppm/℃+30mV
Temp.coefficient voltage
(0 °C~40 °C) Setup
50ppm/℃+30mA
current
Readback
50ppm/℃+30mA
current
Sample rate 10HZ/S
Working temperature 0℃~40℃
Storage temperature -10℃~70℃
Insulation voltage of
240V
the output to ground
Dimension (mm) 214.5mmW×88.2mmH×354.6mmD
Weight (net) 7.7Kg
Parameters IT6953B
voltage 0~150V
Rated values
current 0~10A
( 0 °C~40 °C)
power 600W
Load regulation voltage ≤0.01%+25mV
±(%of output+offset) current ≤0.5%+10mA
Line regulation voltage ≤0.01%+25mV
±(%of output+offset) current ≤0.5%+10mA
Copyright ©ITECH Electronic Co., Ltd. 37
Technical Specification