Water Vapor Content of Gaseous Fuels Using Electronic Moisture Analyzers
Water Vapor Content of Gaseous Fuels Using Electronic Moisture Analyzers
Water Vapor Content of Gaseous Fuels Using Electronic Moisture Analyzers
for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D5454 − 11 (2020)
4. Significance and Use 5.1.4 Power Supply—Analyzers for field use will have
4.1 Water content in fuel gas is the major factor influencing rechargeable or easily replaceable batteries. (Warning—
internal corrosion. Hydrates, a semisolid combination of hy- Analyzers for use in hazardous locations because of combus-
drocarbons and water, will form under the proper conditions tible gas must be certified as meeting the appropriate require-
causing serious operating problems. Fuel heating value is ments.)
reduced by water concentration. Water concentration levels are 6. Calibration
therefore frequently measured in natural gas systems. A com- 6.1 A calibration technique is described in Practice D4178
mon pipeline specification is 4 to 7 lb/MMSCF. This test that should be used to verify the accuracy of the analyzer. This
method describes measurement of water vapor content with method uses the known vapor pressure of water at 0 °C and
direct readout electronic instrumentation. mixes wet gas and dry gas to make up the total pressure so that
a standard gas of known water concentration is achieved.
5. Apparatus 6.1.1 Instruments very sensitive to sample flow must be
5.1 The moisture analyzer and sampling system will have compensated for barometric pressure.
the following general specifications: 6.2 A commercially made water vapor calibrator is shown in
5.1.1 Sampling System—Most errors involved with moisture Fig. 1, which uses essentially the same technique. This method
analysis can be eliminated with a proper sampling system. is useful only between 5 to 50 lb/MMSCF.
5.1.1.1 A pipeline sample should be obtained with a probe
6.3 Low-range water vapor standards may be obtained by
per Method D1145. The sample temperature must be main-
the use of water permeation tubes. Permeation rates must be
tained 2 °C (3 °F) above the dewpoint of the gas to prevent
established by tube weight loss.
condensation in the sample line or analyzer. Use of insulation
or heat tracing is recommended at cold ambient temperatures. 6.4 Compressed gas water vapor standards may be used,
5.1.1.2 Analyzer sensors are very sensitive to contamina- provided they are checked by an independent method once a
tion. Any contaminants injurious to the sensor must be re- month.
moved from the sample stream before reaching the sensor. This 6.5 Calibrate the analyzer using one of the standards in 6.3
must be done with minimum impact on accuracy or time of and 6.4 and respective procedures. Calibration must be at two
response. If the contaminant is an aerosol of oil, glycol, and so points, one higher and one lower than average expected
forth, a coalescing filter or semipermeable membrane separator readings. Some analyzers can have large nonlinear errors. Use
must be used. the calibration adjustment if applicable.
5.1.2 Construction—Sampling may be done at high or low
7. Procedure
pressure. All components subject to high pressure must be
rated accordingly. To minimize diffusion and absorption, all 7.1 Preparation—The analyzer operation and calibration
materials in contact with the sample before the sensor must be should be checked according to the manufacturer’s recommen-
made of stainless steel. Tubing of 1⁄8-in. stainless steel is dations prior to use. See Section 6. Verification of a dry
recommended. (Warning—Use appropriate safety precautions instrument using dry compressed nitrogen to get a reading
when sampling at high pressure.) below 1 lb/MMSCF is recommended before field use.
5.1.2.1 Pressure gages with bourdon tubes should be 7.2 Sample Procedure—Sample as in 5.1.1.1. Use as short a
avoided as a result of water accumulation in the stagnant sample line as practical. Purge the sample for 2 min before
volume. valving to the sensor.
5.1.2.2 Sample purging is important to satisfactory response 7.3 Reading—The time for a sensor to come to equilibrium
time. There must be a method to purge the sample line and is variable depending on its type and condition. The analyzer
sample cleanup system. may require 20 min to stabilize. Some analyzers have an
5.1.3 Electronics—Output from the sensor will be linearized external recorder output, and these can be used with a chart
for analog or digital display in desired units (usually lb/ recorder to become familiar with the true equilibrium response
MMSCF). There must be an adjustment for calibration accu- time.
racy available that can be used in the field if a suitable standard
is available. (This does not apply to instruments that assume 8. Precision and Bias
complete chemical reaction of water. Their accuracy still must 8.1 Precision data is being prepared for this test method by
be verified as in Section 6.) an interlaboratory study.
2
D5454 − 11 (2020)
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or [email protected] (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/