UScan Operation Manual CT
UScan Operation Manual CT
Version 2.0.1
May, 2017
Warranty, Exclusions and Limitations
CHOPIN Technologies warrants workmanship and materials to be free from defects for a
period of one year from the date of shipment.
Excluded from the warranty coverage are products a) which have been subjected to
mechanical, electrical or other misuse or abuse b) which have been disassembled or
repaired or attempted to be, by other than CHOPIN Technologies authorized servicing
agents.
The foregoing warranty shall be in lieu of all other warranties, expressed or implied, and the
Corporation expressly disclaims any warranty for merchant ability or fitness for a particular
purpose.
Unless otherwise agreed in writing by the Corporation, repairs and replacement parts
provided by the Corporation under its warranty shall be guaranteed only for the non-expired
portion of the original warranty on the repaired product.
The liability of CHOPIN Technologies to any purchaser or user of its products shall not in any
case exceed the cost of correcting defects in a product as herein provided and, upon
expiration of one year from the date of sale, all such liability shall terminate. In no event 2
shall the Corporation be liable for interruption of operations, loss of profit, or special or
consequential damages.
CHOPIN Technologies
20 avenue Marcellin Berthelot
92390 Villeneuve La Garenne – France
Phone: + 33 1 41 47 50 48
Email: [email protected]
Contents ....................................................................................................................................... 3
1 Introduction .......................................................................................................................... 5
1.1 Supported Hardware ..................................................................................................... 5
1.2 Scope of UScan Software ............................................................................................. 6
1.1 Accessory/Companion Programs ................................................................................ 6
1.4 Multi-Language Support ............................................................................................... 6
2 UScan User Interface............................................................................................................ 7
3 Configuration ........................................................................................................................ 9
3.1 Introduction ................................................................................................................... 9
3.2 Entering Configuration Mode ..................................................................................... 11
3.3 Scan Method Configuration........................................................................................ 11
3.3.1 Create a new Scan Method ................................................................................. 12
3.3.2 Rename a Scan Method ...................................................................................... 14
3.3.3 Edit a Scan Method ............................................................................................. 14 3
3.3.4 Delete a Scan Method ......................................................................................... 15
3.4 Calibration Method Configuration .............................................................................. 16
3.4.1 Create Calibration Method based on PRD, MLR or CVF file .............................. 16
3.4.2 Create a Conformity Calibration Method ............................................................ 23
3.4.3 Create an MLR Calibration Model ....................................................................... 29
3.4.4 Create a Calculated Variable Calibration Model ................................................ 32
3.4.5 Rename a Calibration Method ............................................................................ 37
3.4.6 Edit a Calibration Method .................................................................................... 38
3.4.7 Delete a Calibration Method ............................................................................... 38
3.5 Define Product Configuration ..................................................................................... 39
3.5.1 Create a Product .................................................................................................. 39
3.5.2 Rename a Product ............................................................................................... 47
3.5.3 Edit a Product ...................................................................................................... 48
3.5.4 Delete a Product .................................................................................................. 48
3.6 Global Settings Configuration .................................................................................... 49
3.6.1 UIF Settings ......................................................................................................... 49
The UScan™ program serves as the primary user interface for routine quantitative analysis
applications for the CHOPIN Technologies Spectralab . The primary objective in the
development of the UScan software was to make it as intuitive and easy to operate as
possible without compromising any of the vital features or functionality needed for
demanding NIR analysis.
UScan is designed with a graphical interface for easy operation via touch screen. This
design makes for easy training and operation in both laboratory and at-line process control
operations.
Configuration of the software is protected with a password for secure and reliable operation
in multi-user environments. The configuration is based on re-usable blocks of settings
termed Methods, which are versioned and stored in a database. These settings are linked to
each sample record for complete traceability.
UScan also supports TAS™; True Alignment Spectroscopy technology. TAS is a technology
that allows a customer to recalibrate the instrument to factory alignment using certified
reference standards based on first principles. As such, TAS based instruments are
inherently spectrally compatible with each other and future instrumentation, making data
collection future-proof.
5
This first section of the manual provides introductory information about the UScan product.
Section 2 provides a description of the components of the main UScan screen and an
overview of the fundamental capabilities of the UScan program. Section 3 describes various
UScan configuration options which are typically customized according to application
requirements, user preferences, instrument model, and sampling accessories being used.
Section 4 describes routine operation and diagnostic tests using the UScan program.
Section 5 provides information on miscellaneous other aspects of the software such as
language selection and practical guidelines. Additional useful information can be found in
the Appendix sections. Please note that detailed information about a particular model of
Spectralab spectrometer and its accessories can be found in the operation manual that
corresponds to it.
UScan is primarily intended to be run on the internal computer embedded in Spectralab and
has been developed to work in conjunction with the Spectralab touch screen (a standard
feature of all Spectralab instruments). Additionally, UScan can be installed on a Windows®
computer and operated in Demo mode for training and demonstration purposes. Please
see the appendix for installation and configuration instructions for demo installations.
Although English language text is used in the screen images in this manual, UScan supports
the display of text in a number of additional languages. As this is the first version of UScan,
translated versions of the software are in progress. Please check with your local distributor
for information on translated versions.
UScan is designed around a central graphical user interface (UI) that can easily be operated
with a touch screen for routine operations. Sample analysis and routine diagnostics are
performed from the main UI, with configuration and other advanced operations accessible
behind a password protected icons.
The main UI is presented below, along with a description of its function. Each item is
described in detail in later sections of this manual under the Configuration and Operation
sections.
Descriptions
Configuration
Quit UScan
History View
SPC View
Spectrum View
Operation TAS
8
TAS Diagnostics
Lamp Health
Choose Product
Scan Sample
Print Sample
Instrument Health
Scan Reference
3.1 Introduction
Configuration of UScan for routine analysis involves creating re-usable blocks of settings
termed Methods and combining these methods to create Products. The Products are then
selected and used for routine analysis.
There are two types of methods – Scan Methods and Calibration Methods.
Scan Methods contain settings related to sample cup type and all the scanning settings that
are used to analyze that cup type. UScan installations generally have at least one Scan
Method per cup type that they use.
Calibration Methods contain settings related to a calibration model and how the calibration
model should be used. The calibration model might be a quantitative PLS based model, an
MLR model, or a Calculated Constituent model.
Every Product must have one Scan Method and at least one Calibration Method. More than
one Calibration Method can be combined into one Product.
This modular design for Product configuration has several implications. First, it allows for 9
easy creation and maintenance of important analytical settings. For example, the settings
for using a Forage cup can be created once, saved in a Scan Method and re-used for all
products that require this cup type. If a change is needed to the settings for the Forage cup,
the Scan Method can be updated, and all products using this Scan Method will be
automatically updated with the new settings.
Secondly, it allows for easier tracking of all settings used create each method, and likewise
link these settings to each sample record. UScan creates unique versions when a Method is
created or updated, and links the Methods and Versions to each sample record. In this way,
each sample record can be linked to all analytical settings that were in place at the time of
analysis for a completely traceable analytical record.
To enter configuration mode, select the icon from the upper right section of the main
user interface (UI). A password dialog will open.
The default user and password are both ‘admin.’ The advanced configuration screens will
now be available from buttons across the top of the main UI.
11
To log out from the administrator user, press the icon present in the upper right of the
UI.
To return back to the routine analysis screen, press the Main button from the top buttons
present in the main UI.
The Scan Method contains all of the analytical settings that an instrument needs to collect a
spectrum from a sample. These settings include cup type, cup rotation settings, reference
method and settings, and sample scan settings. In general, most installations have one
Scan Method per cup type.
The Scan Method settings are versioned, meaning that each time a Scan Method is changed,
a new version is created. All Sample records are traceable to the Scan Method and version
in place at the time of analysis for complete traceability.
Below are the settings for the Scan Method configuration. The actual options present are
specific to the instrument configuration, and thus not all options might be present in your
installation.
To create a new Scan Method, press the Add button. Place a name for the Scan
Method in the dialog below. It is suggested that the name refer to the cup type being
configured.
Cup Type
Next, select the Cup Type appropriate to your sample and calibration. In general, finely
ground samples can be analyzed in smaller cups like ring cups and powder cups, while
coarse or unground materials should be analyzed in larger cups with more surface area to
reduce sampling errors. Please consult your CHOPIN Technologies specialist for further
information.
Depending on the Adapter type and Cup chosen, Cup Rotation, Sample Steps and / or Platter
Speed might be displayed. In general, Stepped rotation with 1 rotation and 12 scans is
recommended. Please consult with your CHOPIN Technologies specialist for specific
recommendations.
A reference every sample means that the reference is scanned before every sample. This
method is good in some situations where the environment is changing but makes the
scanning sequence longer. A timed reference setting is best in situations where the
environment is stable as the scanning sequence is shorter.
Scanning Parameters
The Number of Sample Scans refers to the number of complete scans that are averaged to
make up a sample scan. Increasing the number of scans lengthens the scanning sequence
but lowers the noise. It is recommended to use the default value here unless circumstances
dictate otherwise.
Number of Scans for Internal Reference refers to the number of reference scans that are
averaged to make up the reference scan. This is set in the Global Settings, System Settings
and is listed here only as an indication.
Internal Reference Expiration Time (min) is only used when the Timed reference method is
selected, and is also set in the Global Settings, System Settings.
To edit a Scan Method, select the method desired from the list on the left of the screen and
press the Edit icon. The warning below will indicate that a new version is being created.
Press OK.
The Scan Method dialog will now be editable. Make the desired changes and press the Save
button to save the updated Scan Method. Pressing the Cancel button will cancel the 15
edit session and leave the Scan Method as it was.
press the Delete icon. The warning below will ask you to confirm the deletion:
Press OK.
If the Scan Method is currently part of a Product configuration, it cannot be deleted and the
following warning will be displayed:
The Calibration Method contains all of the analytical settings that an instrument needs to
predict constituent values from a sample spectrum, calculate additional constituent values
from available data, or perform a qualitative conformity analysis on a sample spectrum.
Every Product must have at least one Calibration Method, but more than one Calibration
Method can be combined into a Product to produce the desired analytical results.
Quantitative calibration Methods can be based on PLS (Partial Least Squares) calibrations
present in a .PRD file or an MLR (Multiple Linear Regression) calibration present in a .MLR
file. Calculated Variables can be configured and are found in CVF files. Qualitative models 16
for performing a conformity test are present in PRC files. For more information on these
calibration types, please see the Glossary in the Appendix.
To create a Calibration Method, press the Add button. Place a name for the Calibration
Method in the dialog below.
button to browse in your PRD calibration file and then give it a description that is
informational.
NOTE: If you are configuring a Calibration Method based on an MLR or CVF model, please
change the file extension filter to the proper extension and select the appropriate file.
17
3.4.1.2 Constituents
Click on the Constituents tab to verify the Constituent names. If the incoming constituents
have already been registered in UScan, they will be matched automatically. If any of the
The dialog below shows the incoming constituents from the PRD and how they are matched
to the constituents registered in UScan.
18
The properties for constituents are accessed from the Edit icon on the Constituents
tab.
Display Name – the name displayed in the UScan program and for reporting. The display
name can be useful when working in languages other than English or for customizing the
displayed constituent name as desired, for example Fat to Oil.
Display Priority – the highest priority constituents appear to the top and to the left when
displayed in UScan. Priority can also be used in reports.
Force Zero – negative numbers will be displayed as 0. This can be useful for constituents
that routinely have values close to 0.00 to prevent negative numbers from being displayed.
NOTE: For a definition of moisture basis, dry matter, as received etc., please see the
Glossary in the Appendix.
For example, many calibrations are made using reference values having a dry matter basis.
This means that the constituent values are expressed as a percentage of the sample weight
with all of the moisture removed. Dry matter based values can be useful as the constituent
values remain constant across various sample moistures. However, many customers need
to know how much of a constituent is in the sample as it is or as received. Moisture
correction allows you to display the constituents in the desired moisture basis.
To use moisture correction, the check box at the top of the dialog must be checked. This
enables the moisture correction dialog. 19
Any calibration moisture basis can be displayed as any other moisture basis
There must be a Moisture constituent or a Dry Matter constituent present to base the
correction on.
Constant Moisture with a value of 0 is the same as Dry Matter
1) Determine which constituent will be used to base the corrections on. This must be a
moisture or a dry matter constituent. Locate this constituent in the list of
constituents and select the proper Input basis and Output basis. If it is a moisture
constituent, select Moisture Variable in both input and output. If it is a dry matter
constituent, select Dry Matter Variable in both input and output.
2) For all the constituents that you wish to be corrected, use the dialog to select the
proper Input Basis, Input Moisture (if Constant Moisture is selected), Output Basis
and Output Moisture (if Constant Moisture is selected).
3) Some constituents should not be corrected. An example might be a property like
digestibility or hardness. The input and output basis should be the same for these
constituents.
20
21
A transfer file is created after scanning standards on both your instrument and the master
instrument. The collected spectra are compared, and a standardization file containing the
correction is saved with an extension of .STF. Please follow the calibration originator’s
instructions on the use and scope of the STF file.
Transfer files are added to the Calibration Method. Place a check mark in the Use Transfer
File check box, browse in the STF file and enter a description. The transfer file will then be
used during routine analysis.
The types of outliers are listed below (Please see the Glossary in the Appendix for more
information):
When a PRD calibration is imported, default values are entered for all of the outlier types. It
is recommended that these values be used initially as outlier criteria. The Outlier limits may
23
When the above tabs of the Calibration Method are properly filled in, press the Save
icon to save the Calibration Method.
To create a Calibration Method, press the Add button. Place a name for the Calibration
Method in the dialog below.
button to browse in your PRD calibration file and then give it a description that is
informational. 24
NOTE: Please change the file extension filter to the proper extension (PRC) and select the
appropriate file.
At the bottom of the Constituents tab, there are radio buttons to select the Conformity Value
to use for the Conformity test. This is a very important setting and will determine what value
will be used globally in UScan for the Conformity evaluations. This is the value that will be
displayed on the Result and History Views, and this is the value that will determine the
Sample PASS or FAIL based on the Product Limits set. The two options are:
Probability Value: A Probability value gives the probability that your newly scanned sample is
the same as some standard value. Intuitively, a probability of 1.0 would indicate that your
unknown will CERTAINLY be a match for the standard value.
Standardized Value: In the Standardized value system, perfection occurs when the
standardized value is 0.0. Values above (or below) zero, are presented in standard deviation
terms.
Click to select the Conformity Value to use. Depending on your application and users, you
might prefer Probability Value or Standardized Value. Please ask your CHOPIN
Technologies Product Specialist if you have questions.
25
NOTE: All Calibration Methods in a Product must have the same Transfer file or not use a
transfer file at all.
A transfer file is created after scanning standards on both your instrument and the master
instrument. The collected spectra are compared, and a standardization file containing the
correction is saved with an extension of .STF. Please follow the calibration originator’s
instructions on the use and scope of the STF file.
Transfer files are added to the Calibration Method. Place a check mark in the Use Transfer
File check box, browse in the STF file and enter a description. The transfer file will then be
used during routine analysis.
26
27
When the above tabs of the Calibration Method are properly filled in, press the Save
icon to save the Calibration Method.
Probability Value
A Probability Value range is 0 to 1, with perfection being 1. Thus a Warning Limit must be
less than 1, with an Action Limit lower than the Warning Limit. Probability Value Limits are
added using the Lower Action and Lower Warning Limits as shown below.
The range of Standardized Value results from a Conformity test is 0 to ∞, with 0 being
perfection. Thus a Warning Limit must be above 0, with an Action Limit above the Warning 28
Limit. Standardized Value Product Limits are added using the Upper Warning and Upper
Action Limits as shown below.
The wavelengths and coefficients of the calibration must be known to create an MLR
calibration.
These MLR calibrations define a relationship between a spectrum and a constituent value
based on only a few datapoints and frequently must be modified with a bias and / or slope
correction to obtain good values. Filter instruments also have a large degree of variability
with the wavelength and quality of light passing through them. It is recommended that full
spectrum PLS calibrations be created for more difficult constituents and more robust
predictions.
Creating an MLR Calibration Method involves 2 steps: Creating the MLR File and then
creating the Calibration Method to use that file.
Click to select an existing Constituent name, or enter in a new Constituent name and press
Add New Constituent.
30
It is recommended that you save the file in the default C:\Unity\models directory. Save the
file with a .MLR extension. The MLR calibration is now ready for configuration in a
Calibration Method.
Calculations for sample properties like Energy that are based on other constituent
values
Evaluations of constituents for further processing ie if a constituent is below a
threshold then apply a correction factor. These are useful for forage evaluations and
other applications
Estimating growth rates or potentials
Displaying additional moisture bases on the results screens
From the Calibration Method tab, click the Model Types button.
33
Click to select the Calculated Variables tab, then press the Select button as shown below to
select the name of the Calculated Variable result. To edit an existing CVF file, press the
must be selected. Press the Add button and select the Constituents from the dialog
below. Multiple constituents can be selected by holding down the <ctrl> key. An example is
35
The selected Constituents will be present in the table to the left. These are the constituents
that are available for building the equation. It is recommended that you double click the Test
Value column for each constituent and place an example value as shown below. These
values will be used to test the equation. Choose values that make the result easy to predict.
NOTE: Decimal values must be added with a . period as the decimal. For detailed guidelines
on syntax for numeric and logical expressions, please see the Appendix 6.1 Calculated
36
Constituent Expression Formula Guidelines.
Note the Result value in the lower right as you are creating the equation. If the expression is
invalid, an error will be displayed. Make sure that the result is what you expect based on the
Test Values that you entered above.
When the equation is complete, press the icon and enter in a name for the CVF model
as shown below.
All Products that are configured to use this Calibration Method are immediately
updated to use the newest version of the Calibration Method.
All Samples scanned are traceable to the specific version of the Calibration Method
that was in place at the time of analysis.
To edit a Calibration Method, select the method desired from the list on the left of the screen
and press the Edit icon. The warning below will indicate that a new version is being
created. Press OK.
38
The Calibration Method dialog will now be editable. Make the desired changes and press
the Save button to save the updated Calibration Method. Pressing the Cancel button will
cancel the edit session and leave the Calibration Method as it was.
screen and press the Delete icon. The warning below will ask you to confirm the
deletion:
A Product is the collection of settings and calibrations that a UScan user selects to analyze
a sample. The Product was designed to be a versatile container that would allow users to
create custom configurations of a wide collection of settings yet be nimble to maintain and
use.
A Product must have one (1) Scan Method, and one (1) or more Calibration Methods. These
methods may be updated individually, or be swapped in and out during Product updates.
When a sample is analyzed, all the settings in place at the time of analysis are traceable to
the sample record in the database.
When a Product is saved, it may be placed in a Main Category, or in custom categories that
allow Products to be placed in logical groups for more intuitive use by routine operators. All
Products and Categories have associated icons for easy operation with a touch screen.
Users can choose from an included library of icons or import their own image files.
To create a Product, press the Add button. Place a name for the Product in the dialog
below.
NOTE: There must be exactly one (1) Scan Method and one (1) or more Calibrations
Methods in each product. If more than one Calibration Method is configured, the following
rules apply:
40
Use Date and Counter – A field is provided to the immediate right of this option to enter the
initial counter value. A name will be automatically generated using a date code prefix
followed by the next value in the counter sequence. The date is formatted according to the
“Short Date Format” configured within the Date tab of the Windows Control Panel’s Regional
Options. Any “/” characters within the date are replaced with the “-” character. InfoStar will
automatically reset the counter to 1 when generating the name of the first report file for a
new day. The maximum value for this counter is 999999. Example sample number would
be “06-03-2010_1” for the first sample analyzed on the date of June 3, 2010 with a regional
“Short Date Format” of MM/dd/yyyy or MM-dd-yyyy.
Use Base Name and Counter – Allows a base name to be specified, and provides a
counter that will automatically increment and append to the base name. A field is provided 41
to the immediate right of this option to enter the initial counter value. A field below is
provided for entry of the base name. For file and/or sample name sorting purposes, it is
sometimes helpful to set the counter to a large value, such as 1000. This will cause the next
8999 measurements to have names with the same number of characters. The maximum
value for this counter is 999999.
Sample Full Name – If one of the automatic sample naming schemes is in use, this
provides an example of what the name of the next sample in the sequence will be. If “Enter
Name for Each Sample” is selected, then this field will indicate “Manual Entry”
Enable Repacks
If a value of three were used for this setting, then the final spectrum would be an average of
the three spectra measured, one per repack, and the final results for the sample would be
the average of the three separate NIR predictions, one corresponding to each repack.
NOTE: The reported value for a repack sample might be slightly different than what would
be predicted using the calibration and the average spectrum because the reported value is
based on an average of the predictions.
NOTE: Slope adjustments have a significant and sometimes unexpected effect on predicted
values and are rarely needed in most NIR installations. Please contact a CHOPIN
Technologies product specialist if it appears that you require a slope adjustment.
To enter a bias, double click on the Intercept field next to the Constituent name and enter the
bias value.
43
NOTE: Conformity Values have special rules for Product Limits. Please see 3.4.2.6 Special
Considerations for Product Limits for Conformity Values for more information.
44
To select a Display Parameter set, click Edit icon. The dialog below will open.
press the Add button and enter in a name for the Display Parameter set. Place a check
45
mark in the checkboxes of the items you wish displayed. Press Save to save the
Display Parameter set.
For an explanation of each statistical term, please see the Glossary in the Appendix.
Place a check mark in the check boxes next to the UIFs that you wish to have linked to this
Product.
Product Categories
UScan has a Main product category that can be used to organize Products. If the UScan
installation is to have many Products, it might be useful to create additional categories. For
example, a flour mill might want to have a category for finished flours and a category for by-
products.
To add a Category, press the Add button and enter a name for the Category. Browse in
Product Image
GIF or PNG may be browsed in and used. To add a Product image, press the Add
button and enter a name for the Product Image. Browse in an image icon file and press
Save .
47
Once the category and image are selected, the Product can be saved and is ready for routine
analysis.
To edit a Product, select the Product desired from the list on the left of the screen and press
the Edit icon. The warning below will indicate that a new version is being created.
Press OK.
48
The Define Product dialog will now be editable. Make the desired changes and press the
Save button to save the updated Product. Pressing the Cancel button will cancel the
edit session and leave the Product as it was.
Delete icon. The warning below will ask you to confirm the deletion:
The Global Settings tab contains settings that have a program-wide effect on the
configuration of UScan.
UScan has 8 UIFs available, and the configuration of each one is the same. Products can
then be configured to use any combination of the 8 UIFs, depending on need. Each UIF can
be made a required field, meaning that the Sample record cannot be saved until the operator
fills in that UIF. Making a UIF required ensures that the field is filled out with every sample
analyzed.
UIFs can be configured in two ways: Manual Entry and Select Values. The examples below
explain each one. Each UIF needs to be configured and saved individually.
Manual Entry
UIFs configured for manual entry present a text box for the user to fill in. These fields are
useful to capture information that is unique or always changing, like a Purchase Order or
Sales Order.
To configure a UIF for manual entry, simply click Manual Entry and fill in the name as shown
below. If this field is to be required, place a check in the Check Box for Required field. Press
save to save the UIF configuration.
To configure a UIF for Select Values, click this selection and fill in the UIF name as shown
below. If this field is to be required, place a check in the Check Box for Required field.
To add items to the list, press the Add button to create an entry in the list box, and then
click on this entry to edit it. Repeat for each item in the list. Items can be deleted from the
list as well. When finished, press Save to save the UIF configuration.
For more information on creating a demo setup, please see the UScan Quick Start Guide.pdf
in C:\Program Files\Unity Scientific\UScan\Help .
Enabling the LIMS export function is configured in the Product Settings tab of Define
Product. Thus each Product can have the LIMS export enabled individually.
The LIMS Configuration tab contains the global settings that define the fields and format of
the LIMS export file.
Settings
The Settings tab defines where the LIMS output file will be stored and format settings.
Press the browse button and select the directory and filename of the LIMS output file.
Place a check in the check box if you wish the file to be appended with each sample record
(recommended in most situations).
Select the Start Character and Delimiter for the data format. These settings help the external
data system properly parse the file. Consult your IT professional if you have questions on
the proper format.
51
Place a check in the check boxes next to the parameters that should be exported to the LIMS
file.
52
This setting keeps the internal reference scan current for reliable and consistent data
collection. 30 minutes is the default and good for most installations. For extremely variable
environments, 15 or 20 minutes (or a reference setting of before every sample in the Scan
Method) might be appropriate.
4 Routine Analysis
UScan is designed to offer a flexible user interface for operators so that the analytical flow
and data presentation are optimized for the particular analytical situation. The setup of
UScan is addressed in the Configuration section of this manual. Please refer to that section
for more complete information on configuration options.
There are 2 methods for maintaining a current reference scan: timed reference and a
reference before each sample. This is configured in the Scan Settings the Scan Method. At
any time, the reference scan can be updated by pressing the Scan Reference button.
2. Place the sample on the instrument in the proper sample cup. The sample cup type
is displayed in the status section of the main window below the Product name.
54
Pressing the print sample icon to get a hard copy of the results. The document
will be sent to the Windows default printer.
In Options, the operator can define how to select the samples to display, display options and
moving average options.
The Product Limits for each product determine the scaling of the y-axis. If no Product Limits
are configured, the y-axis starts at y=0.
possible to overplot other sample spectra by pressing the button and selecting
Samples to plot.
56
UScan stores all sample data in one central database, with all sample records traceable to
the exact configuration settings in place at the time of analysis. UScan has a collection of
tools available for managing and reporting sample data.
To activate the Sample Search, press the Search button on the History View. The dialog
below will be presented:
57
Each of the fields in the dialog above acts as a filter to all the samples in the database.
Enter in the number of samples, sample name or date range to select samples from the
listed product. To change Products, click the Change Product icon and select the Product
desired.
When using Sample Name as a filter, use % as a wildcard. For example, 114% will select
114
11456778
114 Retest
114943#1
Sample labels such as Sample Number, Comment and User Information Fields (UIFs) can be
edited in the Sample Record. These fields are outlined in red below. Simply click in the field
of interest and edit the data. Other data such as the Sample DateTime of analysis and
Product type are fixed and cannot be changed.
The Configuration Methods are also displayed in the Sample Details. Each Method is listed
along with the version in place at the time of analysis. For example, the sample below was
analyzed with version 5 of the Puppy Chow product.
58
When finished, press Save to save the updated information or to cancel the
update.
To Export Sample data, select the samples to export using one of the methods above. Press
Export to initialize the export, and the dialog below will appear:
59
Select the report template from the list at the left and press View Report. The sample data
will be exported based on the template selected and the Window below will be displayed:
From this dialog is it possible to export the report to a variety of formats including XLS, PDF,
RTF, TXT or CSV. It is also possible to print the report from the icon circled in red. It is also
possible to send this report to any Windows printer. Controls across the top allow for
navigating within the report.
If more than on sample is selected, the sample data will be exported to an XLSX file in
C:\Unity\Reports. This file can be opened by Microsoft Excel or Open Office from Apache
(https://www.openoffice.org/ ).
To export spectra, select the samples to be exported with the methods described above.
Press the Export Spectra button and the dialog below will appear:
Next, press the destination folder icon and enter in the filename for the export. Press
OK and the file will be exported to the destination entered.
UScan has a complete set of instrument hardware and performance diagnostic procedures
to ensure proper performance and functionality of your instrument. Many of these
diagnostic tests are instrument specific, and thus are covered in detail in the instrument
operating manual. Please consult this manual for more information. A summary of each
diagnostic test is presented below.
5.1 TAS
TAS, or True Alignment™ Spectroscopy, is a patent pending technology for aligning and
maintaining instrument calibration. TAS routines include a procedure to evaluate instrument
status and performance, as well as a procedure to align the instrument back to factory
specifications.
62
The TAS Instrument Status evaluation will prompt the user to place specific standards on
the instrument. When done, a status page with pass / fails will be displayed. If any tests fail,
proceed to run the TAS Alignment procedure below.
The TAS Alignment procedure will prompt the user to place specific standards on the
instrument. After scanning the samples, the computer will calculate the optimal alignment
settings. When done, a procedure complete message will be displayed.
NOTE: The final section of the alignment procedure can take 4 – 5 minutes of computer
time, and it might appear that the program is not doing anything. Please do not interrupt this
process. The program will tell you when it is done.
Press this button to diagnose the lamp. The elapsed hours, current and other settings are
displayed.
NOTE: The hours remaining are approximate based on the estimated lifetime of the lamp.
63
The Instrument Health button will test various electrical and mechanical parts in the
instrument and return pass/fail indications for each test. If any component fails, please call
your CHOPIN Technologies product specialist.
6.1 Glossary
The terms and definitions below are simple definitions based on the usage with UScan and
CHOPIN Technologies products. For more complete definitions, please consult reference
sources.
(1) Numerical items (“numerics”) (1.0, 2, -0.003, etc.) Note that numerics cannot
contain any commas. If a decimal marker is to be used, it must be a period “.”.
For integers, the decimal marker is optional.
(2) Constituent Names - any constituent name on the internal constituent list can be
used. There is some logic in the software that does arrange the computational
order so that each calculated variable can be computed. Even with that logic in
place, it is still possible to compose an expression that cannot be computed. In
general, .prd constituents, .mlr constituents , and .prc constituents are
determined prior to any calculated variable constituent. The exact spelling of
constituents must be observed. Since constituent names are always forced to
upper case, the constituent names in expression are not case sensitive. For best
results, avoid internal spaces in constituent names (use the underscore "_"
instead).
(3) Operators - operators include + - * / % (for modulus) ^ (for exponents). UScan
also recognizes some logical operators (== != >= <= < > & |). See section below
for more on operators.
(4) Functions - There is a limited selection of Functions. There are unary (one
66
argument) functions, binary (two arguments) functions, and one ternary (three
arguments) function. Functions that start with "I" are logical functions that return
either 1 or 0 as a result. See section 4 for more on Functions.
(5) Parenthesis - Parenthesis must come in pairs. For each left parenthesis, there
must be a corresponding right parenthesis. Note that functions also include
paired parenthesis.
When entering the expression formula, do not use an equal sign unless it is part of a logic
operator. Do not begin an expression formula with an equal sign. Note that an equal sign
can only be used as part of a double-character operator (== != >= <= ).
Order of operations - Parenthesis contents are evaluated first. Since all functions contain
parenthesis, expressions within function parenthesis are evaluated as if they were simple
parenthesis expression.
Feel free to overuse parenthesis to enforce the operation order that you desire.
The unary operator "-" (numerical negate) converts the op argument to -1*argument.
The unary operator "!" converts the op argument to either 0 (when argument is ,0.5) or 1
(when argument>=0.5).
Binary operations must lie between two sub expressions that evaluate to numerics.
Logical Comparisons
Logical Math
& Logical And A&B returns 1.0 if BOTH A and B are greater than 0.49999999, 0
otherwise.
Order of operations
SQR(argument) returns the square root of the absolute value of the argument
Ternary Function