X-ray photoelectron spectroscopy principle

The characteristic binding energies of core electrons in atoms keep them in their
orbitals. The characteristics of adjacent atoms, orbital kinds, and atom types all
influence the binding energies. The binding energies of electrons emitted from the
sample surface are measured by XPS.

The mechanism of photoionization is utilized to eject electrons from surfaces. For


this, the sample surface is irradiated by soft x-ray photons usually 1486 eV from an
Al-anode. The photon penetrates the surface to a depth of approximately 1µm.
When a photon interacts with a sample atom, the photon’s energy ionizes core
electrons, causing them to be ejected from the atoms. Within the bulk of the
sample, the mean free path of the ejected electrons is very small due to interaction
with the surrounding atoms. But, electrons expelled from atoms in the sample’s top
5nm layer have a high chance of leaving without further interaction.
These electrons will pass into the ultrahigh vacuum(UHV) conditions within the
spectrometer to ensure that electrons are not scattered by gas molecules and to
limit surface contamination during the analysis. An electron energy
analyzer(spectrometer) measures the kinetic energy Ek of the ejected electrons.
The original binding energy Eb experienced by an electron in the sample can be
calculated from Ek using the simple relationship.

Ek= hv-Eb– ϕ

where Ek= kinetic energy of the ejected electron

Eb= binding energy of the electron

hv= incident photon energy

ϕ= work function of spectrometer

The different atoms have different electron shells with different binding energies.
Each element has unique binding energy (Eb) which can be used to identify the
elemental composition of the sample surface. The number of emitted
photoelectrons with various Ek is counted, and an XPS spectrum is generated,
which facilitates the identification of all elements in all states except H and He.

X-ray photoelectron spectroscopy instrumentation

The basic five components of the XPS instrument are listed as,

1. An ultrahigh vacuum chamber(UHV) with the monochromatic soft x-ray


source
2. Lens system
3. Energy analyzer
4. Detector
5. Monitor

X-ray source of XPS

Al or Mg-coated anode is struck by electrons from a high voltage around 10-15


kV, AlKα or MgKα radiation lines are produced at energies of 1486.6 eV and
1256.6 eV respectively.
Ultra-High Vacuum(UHV)

10-9– 10-10 torr or high vacuum condition is necessary to reduce the amount of
noise in the XPS spectrum and to avoid any extra peaks from contaminations on
the sample surface.

Electrons lens

The electron lens collects and slows the ejected photoelectrons from the sample
surface before entering the energy analyzer of XPS.

Energy analyzer of XPS

It is the hemispherical sector, which consists of two concentric hemispheres with a


voltage applied between them.

Detector

A multichannel display detector which is equivalent to a photographic plate allows


simultaneous detection of the kinetic energy of the ejected electrons.

Monitors/Recorder

The high-performance computer is used as a recorder/monitor which controls


voltage to the lense and analyzer. It also produces an XPS spectrum. The XPS
spectra are generated by plotting Ek intensity vs binding energies, by computers
automatically.

Application of x-ray photoelectron spectroscopy

XPS can be used for both qualitative as well as quantitative analysis.


Qualitative analysis

Elemental analysis: The electron energy levels of an atom can be divided into two
types, core levels, which are bound to the nucleus, and valence levels, which are
only weakly bound. The valence levels of an atom are the ones that interact with
the valence levels of other atoms to form chemical bonds in molecules and
compounds. Their character(valence bond character) and energy are changed by
this process, becoming characteristics of the new species formed.

The core-level electrons of an atom have energies that are nearly independent of
the chemical species in which the atoms are bound since they are not involved in
the bonding process. Thus, the identification of Core level binding energies thus
provides unique signatures of the elements. All elements in the periodic table can
be identified in this manner, except for H and He which have no core levels.

Quantitative analysis

Quantitative analysis requires the measurement of relative peak intensities and the
intensity of the peak depends on the atomic concentration(N) of an element present
in the surface film.

Chemical shift (chemical state analysis)

Chemical shift is the difference in binding energy between the same atom in two
chemically distinct states in the same molecule or in two separate compounds. It is
characteristic of the nature and the number of atoms surrounding the ionized atoms
of the sample. It can be used to see what kind of chemical bonding of atoms is on
the sample surface.
 It concerns the chemical bonding nature, the nature of functional groups, and
the oxidation states of atoms.
 An important capability of XPS is its ability to distinguish the same element
in different chemical states, because of BE of an electron is different due to
changes in charge density around that element by oxidation

In-depth analysis

Estimation of the in-depth distribution of elements in thin surface film is called in-
depth surface analysis. It is done to know whether the thin film is homogeneous or
not.