Manual - Solution Delivery System Local-ME12568
Manual - Solution Delivery System Local-ME12568
TABLE OF CONTENTS
PAGE
CHAPTER ONE: GENERAL INFORMATION 1-1
CHAPTER ONE:
GENERAL INFORMATION
SHOCK HAZARD:
Connect this device to a proper ground connection in accordance with the
National Electrical Code. DO NOT under any circumstances remove the ground
wire or ground prong from any power plug. DO NOT use an extension cord with
this equipment.
DEVICE LABELING:
Do not under any circumstances remove any Caution, Warning or any other
descriptive labels from the devices until the conditions warranting the label are
eliminated. Do not remove model number, part number, or serial number labels
for any reason.
DO NOT OPERATE IN A FLAMMABLE ATMOSPHERE:
To avoid fire or explosion, do not operate this device in an explosive environment
or near flammable anesthetics.
1.3 Applications
The Solution Distribution System (SDS) is intended to be used for mixing and
distributing sodium bicarbonate (bicarb) as well as acid concentrates to remote
points of use.
Bicarb and acid concentrates are used with purified water at the point of use to
create the dialysate solutions used for hemodialysis.
NOTE:
The SDS is not a proportioning device. It is only a fluid delivery system.
WARNING:
The SDS is not equipped with monitoring devices for conductivity, pH or
other parameters. The operator must follow the testing recommendations
of the solution, chemical or product manufacturer. The operator is
responsible for the correct application of these procedures.
1.4 Contraindications
The Solution Distribution System is not designed, sold or intended for use
outside of the device's stated applications, specifications or limitations.
WATER:
The SDS requires water that meets AAMI or local hemodialysis water
quality standards. Additionally, it is necessary to have adequate flow rates
and pressures. The minimum flow rate required for operation of the SDS
is ½ gallon per minute.
NOTE:
Three to five gallons per minute flow rate is recommended. Lower flow
rates will prolong the filling of the mix tank.
DRAIN:
The SDS requires a drain outlet. The drain must have a minimum
capacity of twenty gallons per minute of continuous flow. The maximum
height for the drain is six inches.
TEMPERATURE:
The equipment should be located in an environment that will protect the
equipment from freezing or excessive heat. Refer to the bicarb
manufacturer‟s recommendations for the preferred water temperature.
NOTE:
Low water temperature may cause difficulty in mixing and/or will result in
extended mixing times.
NOISE:
It is advisable to locate the system away from the patient area. Hard walls
and floors will reflect noise and may make the unit seem louder.
Flow Meter:
The flow meter, with integral needle valve, allows for visual indication and
adjustments of fill rate. This device is not supplied on all machines.
Mix Tank:
The mix tank is a sealed tank with a conical shaped bottom. It is made of
high purity polyethylene. The mix tank has volume marks molded into its
side. The 70-gallon tank has 5-gallon and 10-liter graduations. The 100-
gallon tank has 5-gallon graduations. The mix tank is made up of the
following components:
The low-level control switch prevents the mix pump from operating
when the mix tank has insufficient volume for safe pump operation.
A spray head for rinsing all internal surfaces.
A sampling port is located at the bottom of the tank for testing/sampling
purposes.
A removable cover with O-ring seal is provided for access into the
tank.
An eductor (mix nozzle) is located in the mix tank to aid in the mixing of
powdered bicarb.
The tank outlet is screened to protect the pump from small particles,
and a cartridge vent filter inhibits dust or other debris from entering the
tank.
Drain:
A drain valve is provided to allow discharge of unused solution or rinse
water.
Mix Pump:
The mix pump provides recirculation to mix the bicarb powder into
solution.
Flow Switch:
The flow switch is located on the output side of the mix pump to prevent
damage to the pump. The flow switch will automatically turn the mix pump
off during insufficient flow situations.
Transfer Pump(s):
Transfer pump(s) provide automatic refilling of the head tank(s). An
individual transfer pump is provided for each solution.
Dip Tube:
Acid is drawn from a 55-gallon drum using a Dip Tube with an integrated
level switch. An audible alarm sounds and the transfer pumps are
disabled when the tank is empty. When bulk acid storage is used, a level
switch with connections to the tank provides the same low level alarm
function as the dip tube.
LRDP Switch:
The LRDP Switch powers the bicarb loop distribution pump (after a 1
minute delay). The LRDP, on the bicarb loop, purges air from the
distribution system and primes distribution piping.
Transfer Pump Manual Override Switch:
The Transfer Pump Manual Override Switch allows the user to transfer
residual bicarb to the head tank manually. Once initiated, the transfer
pump will operate for 5 minutes or until the head tank is full.
Mix Tank Auto Fill Switch:
The Mix Tank Auto Fill Switch (a momentary switch) initiates automatic or
manual filling of the mix tank. When the AUTO position is selected, the fill
timer and solenoid are activated, filling the mix tank. When manual filling
is required, the switch must be held in the MANUAL position until the
required water level in the mix tank is achieved.
UV Unit Switch:
The UV Unit Switch signals the UV unit to illuminate only when the bicarb
head tank is adequately filled.
Alarm Mute Switch:
The Alarm Mute Switch silences any audible alarm and changes any
flashing alarm light to solid. The mute switch will silence the audible alarm
for 5 minutes before the alarm sounds again unless the alarm situation
has been corrected.
Mix Tank Low-Level Indicator:
The Mix Tank Low-Level Indicator red light flashes and an audible alarm
sounds when the bicarb solution is below the mix tank low-level switch.
The light will turn solid when the alarm mute switch is activated.
Auto Fill Flow Meter with Needle Valve:
The flow meter, with integral needle valve, allows for visual indication and
adjustments of fill rate. The flow meter can be read in gallons per minute
(gpm) and liters per minute (lpm). This device is not supplied on all
machines.
Acid Tank Switch:
The Acid Tank Switch energizes the transfer pump system controlling the
automatic refill of the acid head tank(s). SDS units with acid pumps may
have up to three Acid Tank Switches.
Pump Run Indicator:
A green pump run indicator will illuminate whenever the transfer pump is
operating.
Low Level Indicator:
A red low-level indicator will illuminate, accompanied by an audible alarm
when a low-level condition in the acid tank (55 gallon drum or bulk acid
tank) occurs.
Dip Tube:
A dip tube, with an integrated level switch, sounds an audible alarm and
disables the transfer pumps when the tank is empty. When bulk acid
storage is used, a level switch and connections in the tank provide the
same functionality as the dip tube.
Flow Switch:
The flow switch is located on the output side of the mix pump to prevent
damage to the pump. The flow switch will automatically stop the mix
pump during insufficient flow situations or if the line becomes clogged.
Mix Tank Low Level Switch:
The low-level float switch, located in the mix tank, prevents the mix pump
from operating when the mix tank has insufficient volume for safe pump
operation.
Audible Alarm:
An audible alarm will sound when any alarm condition is activated. The
sound may be muted using the mute switch on the control panel.
PLC:
The PLC, or „Programmable Logic Controller‟ is the Input/Output logic
device that controls almost all functions of the SDS machine. It accepts
various input signals from the hand switches on the control box, as well as
signals from the various switches located in the mix tank, head tank(s),
and drums (acid bulk storage). It then processes these inputs and, if the
correct input signals are received (and certain other conditions are met) it
will send an output signal to the appropriate component (acid transfer
pump, bicarb mix pump, etc.) to activate that component. It also monitors
the status of the various systems and triggers an audible and/or visual
alarm when necessary. The operating program is factory installed and
needs no user adjustment.
1.9 Options
Number of Acids:
The various models of the SDS are capable of delivering up to three
different acid concentrates.
Acid Loop Auxiliary Pump:
Mounted in the unit or separately, the Acid Loop Auxiliary Pump purges air
from the acid distribution system and primes the distribution piping.
WARNING:
The ozone system is intended for equipment disinfection only. It is not
intended for the disinfection of bicarb solution.
1.10 Specifications
1.10.1 Input Water Requirements
Minimum Maximum
Input Flow Rate (AAMI Quality Water) 0.5 GPM 5.0 GPM
Temperature (Water)* 10°C (50°F) 30°C (86°F)
pH (Water)* 3.0 11.0
Inlet Pressure 20 PSI 100 PSI
Drain Capacity 20 GPM N/A
Drain Height N/A 6 in. (0.15 m)
* Refer to the bicarbonate powder manufacturer‟s recommendation for water
temperature and pH level.
NOTE:
The user/operator should recognize that moisture can be caused by
condensation and is not necessarily an equipment leak. This equipment
will function in the presence of condensation.
1.10.4 Dimensions
Mixing Tank: 70 or 100 gallon
Head Tank: 10, 15 or 30 gallons
SDS with 70 gallon Mix Tank* 92”H x 112”W x 58”D
SDS with 100 gallon Mix Tank* 92”H x 118”W x 64”D
*Additional space is required for operator access and acid drums/bulk storage
tank.
1.10.5 Approvals
This equipment is listed by the Canadian Standards Association as certified.
This medical equipment has been tested and demonstrated compliance with the
following standards:
CAN/CSA Standard C22.2 No 601.1
UL 2601-1
1.11 Disposal
Disposal of this product or parts must be carried out according with local disposal
codes.
NOTES:
CHAPTER 2:
INSTALLATION GUIDELINES AND INSTRUCTIONS
Inlet and outlet connections are shown in the illustration below. The inlet SDSL
port is field connected from the loop supply feed. The outlet SDSL port is field
connected to the loop return line.
CAUTION:
Overflow tubes must be routed in a manner such that the tubes are self-
draining; i.e., no sags. Sags in overflow lines can fill with fluid, restricting
the head tank from draining properly.
NOTE:
Fittings are provided on the acid dip tube and the bicarb mix tank for
alternate overflow connections. If the bicarb mix tank fitting is not used,
replace the fitting with a plug.
1. Connect ½” blue tube from the bicarb transfer pump to the BICARB
HEAD TANK SUPPLY fitting.
2. Connect ½” blue tube from LOOP DRAIN TEE to the BICARB LOOP
SUPPLY fitting.
3. Connect clear ½” ID vinyl tube to the HEAD TANK OVERFLOW fitting
and route to drain.
6. Connect the bicarb distribution loop #1 and #2 to the tube unions on the
Bicarb Isolation Valve assemblies.
7. Route the bicarb distribution loop return tubes from the distribution loop to
the right side of the SDS frame. Support the tube with clamps and tie-
wraps as required.
2.9 Bulk Acid Supply Connections to the SDS Frame (User Supplied)
NOTES:
CHAPTER 3:
SYSTEM OPERATION
WARNING:
Thoroughly follow all recommendations for mixing and testing the solution
that the manufacturer suggests.
WARNING:
Perform the following procedure only after the recommended disinfection
and rinse procedures have been completed and a NEGATIVE residual
disinfectant test is attained. Due to the possibility of disinfectant rebound,
perform a residual disinfectant test before initiating the bicarb mixing
procedure.
11. After the mix tank water level has been verified correct, close the following
valves:
V-2 (Spray Head Supply Valve)
V-3 (Inlet Water Valve)
WARNING:
The Inlet Water Valve (V-3) should be CLOSED AT ALL TIMES except
when the mix tank is being filled with water. If this valve is left open during
treatments and the inlet solenoid valve fails, the bicarb solution will be
diluted, causing interruption of treatment.
12. Remove the tank lid.
13. Turn the Mix Pump Switch to the MIX position.
14. Add the appropriate amount of bicarb powder to the mix tank.
15. Replace the mix tank lid.
16. Completely mix the bicarb solution.
17. Verify that the bicarb solution is fully mixed by testing the sample to the
manufacturer‟s specifications.
A sample can be taken from the Mix Tank Sample Valve (V-7).
18. Once the bicarb solution mix has been verified, turn the Mix Pump Switch
to the TRANSFER position.
NOTE:
An audible alarm will sound if the bicarb head tank low-level switch is not
activated within two (2) minutes of the initiation of the transfer pump.
Subsequent alarms indicate that bicarb transfer is not occurring or the
transfer is too slow. During initial fill of the bicarb head tank, this is a
normal condition. Pressing the alarm mute will silence the alarm.
CAUTION:
All dialysis machine bicarb concentrate lines must be disconnected from
the wall dispenser concentrate access ports prior to operation the loop
rapid distribution pump.
19. Turn the Loop Rapid Distribution Pump and the UV Unit Hand Switch to
the ON position.
NOTE:
The Loop Rapid Distribution Pump will start following a one-minute delay,
after the bicarb head tank low-level switch is satisfied.
NOTE:
The bicarb solution transfers automatically until a low-level alarm condition
is reached in the mix tank.
27. The system is ready for normal operation.
WARNING:
During the second bicarb mix procedure, the bicarb head tank automatic
refill function will be cancelled. Operator must monitor the head tank levels
and manually transfer bicarb as necessary once the mix tank low level
alarm has been activated to ensure the head tank does not completely
empty.
NOTE:
When the Mix Pump Switch is in the TRANSFER position, the bicarb mix
tank low level switch will initiate an audible and visual alarm when the mix
tank reaches the low level. The mix tank will have approximately 10
gallons of bicarb solution remaining.
1. Push the Alarm Mute button to silence the audible and cancel the low
level alarm, if required.
2. Turn the Mix Pump Switch to the OFF position.
NOTE:
The transfer process must be continuously monitored, because the bicarb
head tank high-level float switch will disable the transfer pump in this
mode.
3. PUSH the Transfer Pump Manual Override Push Button (only if the
solution in the head tank is below the high-level float switch). This will
transfer the solution from the mix tank to the head tank for five minutes or
until the bicarb head tank high-level switch is activated. Depress the
button as necessary to transfer the remaining bicarb solution to the head
tank. If the button is pushed when the high level switch is satisfied, the
transfer pump will run for 5 minutes, possibly overfilling the head tank.
WARNING:
Do not begin the tank fill procedure until all the bicarb has been
transferred to the head tank and the mix tank has been drained of all
residual solution.
4. Open the Mix Tank Drain Valve (V-5) and drain any remaining bicarb
solution.
5. Verify that the mix tank is empty and CLOSE the Mix Tank Drain Valve (V-
5).
6. Determine the amount of bicarb solution to be made:
How much bicarb will be required?
How much water (gallons) will be required?
NOTE:
If the batch size is not the same size as the first batch, the Auto Fill Mode
Time Preset must be adjusted to provide the correct amount of water.
7. Open the following valves:
(V-2) Spray Head Supply Valve
(V-3) Inlet Water Valve
WARNING:
Do not initiate the Auto Fill function of the SDS until the bicarb head tank
transfer pump has stopped. The transfer pump will operate for five
minutes after the Manual Override push button has been pressed or until
the bicarb head tank level reaches the high level switch.
8. Turn the Mix Tank Auto Fill Switch momentarily to the START position. If
the SDS does not have the Mix Tank Auto Fill Switch, fill tank with
appropriate amount of water and proceed to Step 10.
9. Check the fluid level in the mix tank. If an additional amount of water is
required to bring the mix tank to the required amount of water, either:
Turn and hold the Mix Tank Auto Fill Switch in the MANUAL
position to bring the amount of water to the correct level.
OR:
Reset the Auto Fill Mode Time Preset for the time needed to bring
the mix tank to the correct water level.
10. If water level is too high, OPEN the Mix Tank Drain Valve (V-5) and drain
the water until the level is correct.
11. After the mix tank water level has been verified, close the following valves:
(V-2) Spray Head Supply Valve
(V-3) Inlet Water Valve
WARNING:
The Inlet Water Valve (V-3) should be CLOSED AT ALL TIMES except
when the mix tank is being filled with water. If this valve is left open during
treatments, and the inlet solenoid valve fails, the bicarb solution will be
diluted, causing interruption of treatment.
12. Remove the mix tank lid.
13. Turn the Mix Pump Switch to the MIX position.
14. Add the appropriate amount of bicarb powder to the mix tank.
15. Replace the mix tank lid.
16. Completely mix the bicarb solution.
17. Verify that the bicarb solution is fully mixed by testing the sample per the
manufacturer‟s specifications.
A sample can be taken from the Mix Tank Sample Valve (V-7).
18. Once the bicarb solution mix has been verified, turn the Mix Pump Switch
to the TRANSFER position.
NOTE:
An audible alarm will sound if the bicarb head tank low-level switch is not
activated within two minutes of the initiation of the transfer pump,
indicating that insufficient fluid transfer has occurred.
NOTE:
If the bicarb head tank empties (completely) during this procedure, the
bicarb loop must be purged using the Loop Rapid Distribution Pump.
Refer to Section 3.5 for the “Bicarb Distribution Loop Purge Procedure.”
WARNING:
All dialysis machine concentrate lines must be disconnected from the wall
dispenser concentrate access ports prior to initiating the loop purge (rapid
loop) procedure.
1. Ensure the following conditions of the SDS unit:
Main Power Switch is ON.
Acid Tank Switch is ON.
Head Tank 2 Distribution Valve is ON.
Acid Head Tank level above the low-level float switch.
Bulk acid supply is available.
2. Ensure that the outlet end of the Acid Loop Drain Valve is connected to
drain, then OPEN the valve.
DUAL LOOP INSTRUCTIONS:
Systems that have more than one acid distribution loop will require that
each loop be independently purged.
Each loop is supplied with an isolation valve. To purge a loop, close
one of the loop isolation valves and perform Steps 3 and 4 on each
loop, then repeat on second loop.
3. Depress the Auxiliary Acid Loop Pump Switch button (located on the left
side of unit) and hold in the ON position.
4. Purge any air from the acid distribution loop(s). Allow the Loop Rapid
Distribution Pump to run for a minimum of five (5) minutes per ten (10)
patient stations.
5. RELEASE Auxiliary Acid Loop Pump Push Button Switch.
6. CLOSE the Acid Loop Drain Valve.
7. Purge any air from the acid distribution loop. Allow the pump to run a
minimum of five (5) minutes per ten (10) patient stations.
NOTE:
If no fluid is pumped to drain within 1 minute, stop the pump. The pump
will need to be installed at one of the first patient stations at the beginning
of the loop. Reattempt purge procedure.
8. CLOSE the Acid Loop Drain Valve and unplug the purge pump power
cord.
9. Momentarily OPEN the Acid Loop Drain Valve to relieve pressure and
disconnect the acid line from the pump inlet and restore it to the original
configuration.
NOTE:
The External Purge Pump should be rinsed with clean water prior to
storage.
WARNING:
Dialysis Equipment must not be connected to the SDS system during de-
calcifying and/or disinfection. VERIFY THAT ALL DISTRIBUTION
POINTS OF USE ARE DISCONNECTED.
1. Place the end of the Bicarb Wand in the Loop Drain QD on the SDS unit.
2. CLOSE the following valves:
V-1 (Mix Pump Discharge Valve)
V-2 (Spray Head Supply Valve)
3. OPEN the following valves:
V-3 (Inlet Water Valve)
V-4 (Mix Pump Suction Valve)
V-5 (Mix Tank Drain Valve)
Loop Drain Valve
4. Verify that the mix tank and head tank are empty by checking to see that
no flow is present from the drain lines. If flow is still present from the drain
lines, wait until the flow has ceased before proceeding.
21. After 1 minute, momentarily turn the Mix Tank Auto Fill switch to the
START position to close the inlet water solenoid valve and allow the
system to drain.
22. After the system has completely drained, CLOSE the Mix Tank Drain
Valve V-5.
23. Momentarily turn the Mix Tank Auto Fill switch to the START position to
open the inlet water solenoid valve and begin filling the bicarb mix tank.
24. Fill the bicarb mix tank with 30-50 gallons of water. Confirm the mix tank
water level.
25. Close Inlet Water Valve V-3.
WARNING:
The Inlet Water Valve (V-3) should be CLOSED AT ALL TIMES except
when the mix tank is being filled with water. If this valve is left open during
treatments, and the inlet solenoid valve fails, the bicarb solution will be
diluted, causing interruption of treatment.
26. Turn the Mix Pump Switch to the TRANSFER position.
27. Transfer water to the head tank. Verify water is flowing to drain through
the Loop Drain Valve.
NOTE:
Under certain conditions an air lock may develop in the head tank. It may
be necessary to momentarily operate the loop rapid distribution pump to
overcome the air lock to initiate flow to drain. The Loop Drain Valve must
be closed prior to running the loop rapid distribution pump. Reopen the
Loop Drain Valve after flow has initiated.
28. After 3-5 minutes, turn the Mix Pump Hand Switch to the OFF position and
allow tank to drain.
29. CLOSE the Loop Drain Valve.
30. Turn the Mix Pump Hand Switch to the TRANSFER position.
31. Turn the Loop Rapid Distribution Pump Switch to the ON position.
NOTE:
The loop rapid distribution pump will start following a one-minute delay
after the bicarb head tank low-level float switch is activated.
32. Pump water to drain for 10 minutes, and then begin testing to confirm that
there is no residual bicarb, decalcifier, or disinfectant at the Loop Drain
Valve.
33. Confirm the absence of bicarb, de-calcifier, or disinfectant from all bicarb
distribution loop points of use.
34. If a residual exists, continue to rinse the bicarb distribution loop. If more
water is required OPEN V-3 (Inlet Water Valve) and repeat Steps 23
through 25 to refill mix tank and Steps 30 through 33 to purge the
distribution loop until a negative residual is achieved.
35. Turn the Mix Pump Switch to the OFF position.
36. Turn the Loop Rapid Distribution Pump Switch to the OFF position.
37. Turn the Main Power switch to the OFF position.
38. OPEN the following valves:
V-5 (Mix Tank Drain Valve)
Loop Drain Valve
39. When the head tank and mix tank are completely drained, the system is
now rinsed.
WARNING:
We recommend that the disinfection procedure be performed every day
that the bicarb system is used.
WARNING:
Label all the wall dispenser stations and the SDS with appropriate warning
signs, such as “DO NOT USE, CONTAINS DISINFECTANT.”
WARNING:
Dialysis equipment must not be connected to the SDS system during de-
calcifying and disinfection. VERIFY THAT ALL DISTRIBUTION POINTS
OF USE ARE DISCONNECTED.
CAUTION:
Prior to disinfection/de-calcifying the system must be rinsed. Refer to the
Bicarb System Rinse Procedure in Section 3.8.
1. OPEN the following valves:
V-1 (Mix Pump Discharge Valve)
V-2 (Spray Head Supply Valve)
V-3 (Inlet Water Valve)
V-4 (Mix Pump Suction Valve)
V-5 (Mix Tank Drain Valve)
2. Connect the end of the Bicarb Wand to the Loop Drain QD port on the
SDS unit.
3. Turn the Mix Tank Auto Fill Switch to the MANUAL position.
4. Open the Flow Meter Needle Valve (V-8) to achieve maximum flow rate.
Do not exceed 5 GPM (20 LPM).
5. CLOSE the following valves:
(V-5) Mix Tank Drain Valve
Loop Drain Valve
6. Set the Auto Fill Mode Time Preset for the time duration required to fill the
mix tank with 30 gallons.
7. Momentarily turn the Mix Tank Auto Fill switch to the START position.
NOTE:
As the timer counts down, the display will count down from a full black bar
to a dash-lined white bar. It will display „OFF OUTPUT‟ when finished.
8. Check the fluid level in the mix tank. If an additional amount of water is
required to bring the mix tank to the required amount of water, either:
Turn and hold the Mix Tank Auto Fill Switch in the MANUAL
position to bring the amount of water to the correct level.
OR:
Reset the Auto Fill Mode Time Preset for the time duration to bring
the mix tank water level to the correct amount of water.
9. If water needs to be removed from the mix tank open the Mix Tank Drain
Valve (V-5) and lower to the correct level.
10. After the mix tank water level has been verified correct, CLOSE the Inlet
Water Valve V-3.
WARNING:
The Inlet Water Valve (V-3) should be CLOSED AT ALL TIMES except
when the mix tank is being filled with water. If this valve is left open during
treatments and the inlet solenoid valve fails, the bicarb solution will be
diluted, causing interruption of treatment.
11. Remove the mix tank lid.
12. Select any ONE chemical for disinfection or decalcifying.
WARNING:
DO NOT MIX CHEMICALS TOGETHER.
NOTE:
Biosan® is not recommended for use in SDS systems.
NOTE:
Add quantities listed to 30 gallons of water.
13. Gradually add the disinfectant or de-calcifier to the mix tank.
WARNING:
Wear gloves, eye protection and protective clothing as required by your
facility. Refer to the chemical manufacturer‟s safety recommendations.
WARNING:
Place a label prominently on the mix tank stating, “WARNING! DE-
CALCIFIER OR DISINFECTANT IN USE.”
14. Replace the mix tank lid.
15. Turn the Mix Pump Switch to the MIX position.
16. Mix the disinfectant or de-calcifier for 1 minute. Read the disinfectant or
de-calcifier manufacturer‟s recommendations and adjust the mix time as
required.
17. CLOSE the Mix Pump Discharge Valve, V-1.
18. Allow the solution to mix for an additional 1 minute.
19. OPEN the following valve(s):
V-1 (Mix Pump Discharge Valve)
Loop Drain Valve
NOTE:
Dual loop systems must ensure that both Isolation Valves are OPEN
during recirculation
CAUTION:
During recirculation, confirm the presence of the disinfectant/de-calcifier at
all of the bicarb distribution loop ports. Use appropriate test methods for
disinfectants.
27. After the circulation time is complete, place the Bicarb Wand in the Loop
Drain QD port on the SDS unit.
28. Turn Mix Pump Switch to the OFF position.
29. CLOSE the following valves:
V-1 (Mix Pump Discharge Valve)
V-2 (Spray Head Supply Valve)
V-3 (Inlet Water Valve)
30. OPEN the following valves:
V-4 (Mix Pump Suction Valve)
V-5 (Mix Tank Drain Valve)
Loop Drain Valve
31. Verify that the mix tank and head tank are empty by checking that there is
no flow present from the drain lines. If flow is still present, wait until flow
has ceased before proceeding.
32. The system is now disinfected or cleaned.
WARNING:
The system must be rinsed and confirmed to be free of disinfectant/de-
calcifier before patient use. Refer to the Bicarb System Rinse Procedure
in Section 3.8.
NOTES:
CHAPTER FOUR:
ROUTINE MAINTENANCE
1. When checking the wall box and loop fittings for leaks, ensure that no
liquid or salt crust formations are present at any connection.
2. If salt crust is present, soak the fitting with a cloth wetted with hot water
and wipe away, allowing the tube to be removed from the fitting.
3. Cut approximately ¾” off the end of the tube and reinsert it into the fitting.
In most cases, this will correct the problem. If the leak persists, replace
the fitting.
4. Completely remove the threaded connectors, clean thoroughly, and apply
a small layer of food-grade silicone caulk (FDA approved) to the threads of
the fitting. Reinsert the fitting into the port. This will help prevent the leak
from reoccurring.
Tube connections:
1. Inspect all of the connections and fittings in the SDS base unit for any
leaks. Repair as necessary to prevent damage to other components.
2. Check the acid and bicarb head tanks (including connections) for leaks.
Repair leaks as necessary.
3. Check the head tanks for leaks at the threaded connections into the tanks
and the tube connections into the fittings. Repair as necessary.
Check the Bicarb Mix Tank Vent filter for the presence of moisture.
1. Remove the filter and shake it with a downward motion, causing any water
to be expelled. Replace this filter if any water is expelled.
Clean the SDS unit frame with water to remove any acid and/or bicarb
residue.
Procedure:
1. Using aseptic technique, open the sampling port and allow approximately
100 cc or more of mixed bicarb concentrate to flow before taking the
sample.
2. With solution flowing, insert the sample container in the stream until
adequately filled.
NOTE:
Closing the sampling port before removing the container from the stream
will expose the test solution to „un-rinsed‟ sides of the sampling valve and
may cause an erroneous culture result.
3. Fill the container with the appropriate amount of solution and remove it
from the solution stream.
4. Close the sampling port.
5. Cap the specimen container immediately.
6. Label the specimen appropriately with:
Test to be performed - culture.
Sample source - bicarb solution, machine number or sample port
area.
Time and date sample obtained.
Person who obtained specimen.
Any other pertinent information or procedures required by
laboratory.
Threaded Connections:
1. Prior to performing maintenance on any threaded fitting, relieve the
pressure in the piping/tubing system; i.e., turn off the pump, open a drain
port, etc.
2. Remove the connecting tube or disconnect the piping at the nearest union
joint.
3. Tape the pipe threads with 2 or 3 wraps of Teflon tape.
4. Insert the threaded fitting into the opening and hand tighten. Be careful to
avoid cross threading. Tighten using a wrench only if necessary, about
one-half turn.
5. Re-connect piping, pressurize, and inspect for leaks.
6. If the leak was not corrected, de-pressurize system and tighten an
additional quarter turn. Excessive tightening can crack threaded fittings.
Pressurize the system and inspect for leaks.
7. If the leak continues, it may be necessary to replace the threaded fitting.
Tubing Connections:
NOTE:
Yearly replacement is recommended due the abrasive nature of
bicarbonate powder.
WARNING:
Prior to performing any maintenance, unplug the SDS power cord from the
wall receptacle.
1. Disconnect the unions at the outlet and inlet piping on the pump.
2. Loosen the bolts that connect the pump casing and the motor.
WARNING:
When performing maintenance, do not touch any exposed electrical
connections.
3. Carefully separate the pump casing from the motor.
16. Install the pump casing (large) O-ring onto the casing cover.
17. Replace pump casing cover. Ensure not to pinch or crush O-ring.
Reinstall the eight 10 mm hex head bolts. Tighten until snug.
18. Reconnect the unions on the inlet and outlet of the pump.
19. The pump is now ready for testing to ensure there are no leaks and
operation is correct.
NOTE:
Prior to quartz sleeve replacement, drain the bicarb delivery system.
WARNING:
Prior to performing any electrical maintenance unplug the SDS power cord
from the wall receptacle.
WARNING:
The UV sanitizer is designed to operate only with the cover in place. Do
not attempt to operate the sanitizer with the cover removed. Do not
attempt to remove, modify or defeat the safety switch. Do not look directly
at a lighted UV lamp. Observe the lamp operation only by means of the
lamp view port located on the front cover.
1. Remove the two screws on either end of the UV unit to remove the
exterior cover. This will expose the UV lamp and quartz sleeve.
2. With two fingers on each end of the UV lamp, rotate the lamp either
clockwise or counter-clockwise so that the pins on each end align with the
slot in the socket. Lift or pull the UV bulb out of the UV unit.
3. The bulb will have a rubber ring on each end that is connected to an „H‟
shaped stainless steel separator plate. Remove this separator.
4. The wrench needed to remove the quartz sleeve from the UV unit is
attached to the back of the unit with two plastic clips. Remove the wrench.
Completely loosen the large nut on the left end of the quartz sleeve by
rotation downward.
5. Completely loosen the large nut on the right end of the quartz sleeve by
loosening the nut upward.
6. Wearing gloves slide the quartz sleeve into one end of the inlet/outlet
bushing and remove the quartz sleeve from the opposite bushing by lifting
out. Then remove the entire quartz sleeve.
CAUTION:
Do not touch the UV lamp glass with your fingers. Any place that is
touched will cause the light wavelength to be altered. If you touch the UV
lamp glass, use a lint and abrasive free cloth and alcohol to remove the
fingerprints from the glass.
7. The quartz sleeve is extremely fragile and should be treated very gently.
The exterior and interior surface of the quartz sleeve should be cleaned.
Clean sleeve by soaking in a vinegar solution and running either a
cotton swab or non-abrasive brush through the interior of the quartz
sleeve.
8. Rinse all of the vinegar or cleaning solution off of the quartz sleeve and
nut fittings and reassemble by installing one end of the quartz sleeve into
the bushing. Then install the other end of the quartz sleeve into the
bushing.
9. Slide both nuts out towards the bushings and hand start/thread until they
are firm. Ensure that the nut is not cross-threaded and that the quartz
sleeve is centered in both ends of the nut. You can do this by looking
through the quartz sleeve and into the bushing portion. You will be able to
see the O-ring seal as well as the end of the quartz sleeve. It should be
uniformly spaced on each end.
10. Using the wrench, tighten until firm, but do not over tighten the nut on the
bushing. Ensure that all fingerprints and dusts are removed from the
quartz sleeve and the reflector.
11. Look at the „H‟ shaped stainless steel separator plate. You will notice that
one of the cutouts is deeper than the other cutout. Install the deeper
cutout over the quartz sleeve, one on each end and just in front of the nut.
12. Remove the rubber grommet from each end of the old UV lamp and install
onto the new UV lamp.
13. Align the rubber gasket with the „H‟ shaped slot in the separator on both
ends. Align the pins of the lamp with the slots in the socket and install the
lamp holder on each end of the UV unit. Rotate the bulb clockwise or
counterclockwise until the pins are 90 degrees from their installation
position.
14. Fill the UV unit with water from the SDS machine. Check for any leaks
around the quartz sleeve, nut, and sealing ring. If no leaks are found,
install the cover over the UV unit. Reinstall the two screws to attach the
cover.
15. Reinstall the wrench onto the back of the UV unit.
16. Place a sticker on the UV unit indicating the date for the next UV
maintenance or bulb replacement.
17. After the completion of the UV lamp quartz sleeve replacement, the SDS
should be disinfected. Refer to the SDS Routine Maintenance Procedure
in this section. During the disinfection, verify the operation of the UV lamp.
CAUTION:
Be careful not to damage the level control switch on the lower end of the
dip tube assembly.
NOTE:
Whenever fluid level is below a float switch, the corresponding LED should
be OFF, and whenever fluid level is above a float switch, the
corresponding LED should be ON.
1. Ensure that the head tank has a fluid level at least four inches above the
low-level float switch. Open the SDS control box door and observe the
PLC. Verify that PLC Input LED #15 is OFF (fluid level below switch) and
#14 is ON.
2. As the fluid level drops below the low-level switch, the Input LED #14
should turn OFF.
3. Verify that the acid transfer pump begins operating and the acid head tank
begins to refill.
4. Continue to observe Input LED‟s #14 and #15.
As the fluid level in the head tank rises, the low-level float switch
should trip and Input LED #14 should turn ON.
As the level continues to rise, the high-level float switch should trip,
Input LED #15 should turn ON, and the acid transfer pump should
turn OFF.
Second Acid Head Tank (Orange Tube)
NOTE:
This check is most easily done during the course of a normal operating
day. The acid drum/bulk tank fluid level should be at least six inches
above the low-level float switch. If any function failure occurs, refer to
Chapter 5, “Troubleshooting.”
NOTE:
Whenever fluid level is below a float switch, the corresponding LED should
be OFF, and whenever fluid level is above a float switch, the
corresponding LED should be ON.
1. Ensure that the head tank has a fluid level at least four inches above the
low-level float switch. Open the SDS control box door and observe the
PLC-XM. Verify that the PLC-XM Input LED #4 is OFF (fluid level below
switch) and PLC-XM #3 is ON.
2. As the fluid level drops below the low level switch, the PLC-XM LED #3
should turn OFF.
3. Verify that the acid transfer pump begins operating and the acid head tank
begins to refill.
4. Continue to observe PLC-XM Input LEDs #3 and #4.
As the fluid level in the head tank rises, the low-level float switch
should trip and PLC-XM Input LED #3 should turn ON.
As the level continues to rise, the high-level float switch should trip,
PLC-XM Input LED #4 should turn ON, and the acid transfer pump
should turn OFF.
NOTE:
Whenever fluid level is below the float switch, the corresponding LED
should be OFF, and whenever fluid level is above the float switch, the
corresponding LED should be ON.
1. Ensure that the head tank has a fluid level at least four inches above the
low-level float switch. Open the SDS control box door and locate the PLC-
XM. Verify that the PLC-XM Input LEDs #8 is OFF (fluid level below
switch) and #7 is ON.
2. As the fluid level drops below the low level switch, the PLC-XM LED #7
should turn OFF.
3. Verify that the acid transfer pump begins operating and the acid head tank
begins to refill.
4. Continue to observe PLC-XM Input LEDs #7 and #8.
As the fluid level in the head tank rises, the low-level float switch
should trip and Input LED #7 should turn ON.
As the level continues to rise, the high-level float switch should trip,
Input LED #8 should turn ON, and the acid transfer pump should
turn OFF.
NOTES:
CHAPTER FIVE:
TROUBLESHOOTING
Table of Contents
Problem B: The loop auxiliary pump motor runs, but does not pump.
1. Suction leak – Air entering pump inlet tubing – Correct leak (refer to
Section 4.2.3 in Chapter 4, “Leak Repair”).
2. Acid head tank empty – Refill (refer to Section 5.2, “Acid Transfer
System”).
3. Pump head defective – Rebuild or replace.
4. Loop pump bypass check valve defective – Replace.
Problem C: The auxiliary transfer pump runs and stops repeatedly in rapid
succession, or runs briefly, then stops.
1. The acid drain valve is CLOSED.
OPEN the acid loop drain valve (usually located near the SDS main
frame, on the acid loop return line).
2. The loop tubing is kinked, pinched, or otherwise obstructed.
Connect a drain line to one or more wall dispenser acid port(s) and
OPEN the acid loop drain valve to relieve pressure. Locate and
correct the kink or blocked loop tube.
Problem C: The acid head tank is dry (3rd Acid Head Tank – yellow tubing
[multiple acid systems only])
1. Is the hand switch ON?
NO Turn ON the hand switch.
YES GO TO 2
7. TEST VOLTAGE: Check for voltage (115 VAC) at the following points, in
order. Reference any terminal „N‟ on the main terminal barrier strip as the
neutral or common connection.
CR6-9
No voltage – OPEN fuse (FU-7) – Replace.
CR6-5
No voltage – Defective relay CR6 – Replace.
TB1-10
No voltage – Loose wire/poor connection – Repair as necessary.
TB1-10
Yes, voltage confirmed, but motor does not operate:
a) Loose wire/poor connection – Repair as necessary.
b) Defective pump/motor unit – Replace.
NOTE:
Try a sharp rap on side of motor. It may operate temporarily.
Replace pump/motor unit as soon as reasonably possible.
Problem D: Acid transfer pump runs, but does not pump. (This applies to
all acid concentrate systems.)
1. The pump has lost its prime. Ensure an adequate fluid level in the drum
or bulk tank and perform the procedure in Section 3.6, “Acid Distribution
Loop Purge Procedure (Internal Pump)” or the procedure in 3.7, “Acid
Distribution Loop Purge Procedure (External Pump)” in Chapter 3.
2. There is a suction (air) leak in the acid transfer pump inlet tubing.
a) Defective fitting or O-ring in any input-side tubing connection.
Check the Acid Drum Draw Tube Fitting, Tank 2, Tank 3, Tank 4,
Bulkhead Fitting (both sides), and Transfer Pump Inlet Fitting.
Replace suspect fittings.
OR:
Perform the “Leak Repair Procedure” in Section 4.2.3 of
Chapter 4.
b) Defective acid drum draw tube assembly (dip tube) – Replace.
3. Defective pump head assembly
a) Replace pump head assembly (or entire pump/motor unit).
b) (OPTION) Rebuild pump head assembly.
4. Excessive distance between bulk storage and SDS mix frame.
a) Accessory remote concentrate pump may be required. Refer to the
“Optional Equipment” section in Chapter 1.
NOTE:
After replacing tube fittings, pump head, or performing the “Leak Repair
Procedure,” it may be necessary to perform the “Head Tank Pump Re-
Prime” procedure to restore system to normal operation. Refer to Section
3.6, “Acid Distribution Loop Purge Procedure (Internal Pump),” or the
procedure in 3.7, “Acid Distribution Loop Purge Procedure (External
Pump),” in Chapter 3.
Problem C: The loop rapid pump runs and stops repeatedly in rapid
succession, or runs briefly and stops.
1. The bicarb wand assembly is not connected or is improperly connected to
the quick disconnect port.
Ensure the bicarb wand is fully connected to either the TANK or
DRAIN QD port to allow flow. Pump should not be turned on while
the wand is in the OPERATE QD.
2. Inspect the tubing for kinks or obstructions.
3. Inspect the male bicarb wand fitting for obstructions.
With the wand assembly disconnected from the port, point the
wand toward the drain and push the button tip in.
4. If multiple bicarb loops are installed, make certain the loop return valves are
fully OPEN.
3. Is PLC Input LED #9 ON (mix tank low level float switch ON Indicates level
OK)?
NO a) Fill the tank above the level switch.
b) If the tank is adequately filled (two inches above the float
switch), check the switch with an OHM meter. If the switch
is open, replace the float switch. If the switch is closed,
check the wiring between the switch and the PLC as follows
and repair as necessary:
TB1-COM and C1 on PLC
TB1-COM and PLC Input #9
TB1-COM and TB1-33
TB1-COM and low level float switch normally open
contact
c) Defective PLC – Replace.
YES GO TO 4
4. Is PLC Output LED #12 ON (mix pump run signal)?
NO a) Defective PLC – Replace.
YES GO TO 5
5. Is relay C-1 (motor contactor) energized (mix pump power relay)?
Toggle the mix pump switch ON/OFF a few times while observing for
movement of the C-1 relay arm.
WARNING:
This relay contains exposed metal parts, which normally carry high voltage
electrical power (115 VAC). Do not touch the contactor relay except as
required, using appropriate, insulated test meter leads.
NO Relay C-1 problem.
TEST: Connect a volt meter across the contactor (C-1) connections
1 and 2. with the mix pump switch ON. If voltage (24 VAC) is
confirmed, the relay is defective – Replace.
YES GO TO 6
6. TEST VOLTAGE: Check for voltage (115 VAC) at the following points, in
order. Reference any terminal „N‟ on the main terminal barrier strip as the
neutral or common connection.
TB1-4
YES Voltage is confirmed, but the pump-motor does not run:
a) Loose wire/poor connection at TB1-4 or TB1-N –
Repair as necessary.
b) Remove motor cover and confirm voltage (115 VAC)
at the motor connections.
Voltage present but motor not running
indicates a defective motor/pump unit –
Replace.
NO Voltage present indicates loose wire/poor connection –
Repair as necessary.
C1-5
No voltage - Circuit breaker tripped – Close breaker switch.
C1-3
No voltage - Defective relay C1 – Replace.
Problem B: The mix pump starts, but shuts down after 5-10 seconds.
NOTE:
When the pump operates correctly, the pump output flow trips a flow switch
in the pump output piping (PLC Input #10 on indicates that flow is OK).
1. Is PLC Input LED #10 ON (while the pump is operating)?
NO a) The pump inlet (V-4) or outlet (V-1, V-2) valve(s) is closed,
causing a NO FLOW condition – Ensure that the valves are
OPEN (LED #10 may light momentarily even if valves are
closed).
b) Mix tank outlet screen occluded - Clean as necessary.
c) Loose wire/poor connection - Replace as necessary.
Check continuity between TB1-COM and TB1-34 and TB1-
34 and PLC Input #10.
d) Flow switch stuck or defective – Replace.
Open the mix tank and verify pump flow immediately after
starting the mix pump. Flow indicates faulty switch -
Replace the flow switch.
YES LED #10 should be ON when pump is operating; normal conditions.
2. If LED #10 is ON while the pump is not operating:
Disconnect the flow switch from its wiring and measure the resistance
across the flow switch leads (with the pump not running):
a) Infinite resistance = Normal – GO TO (c) BELOW
b) Near-zero resistance = Switch is stuck in UP position or defective –
Repair or replace float switch.
c) Defective PLC – Replace.
Problem C: Mix pump leaks.
1. Inlet or outlet pipe fittings
a) Remove the pipe. Replace the Teflon tape (3-4 turns); turn the
pipe clockwise to retighten into the female pump head port.
Operate and check for leaks.
Problem B: Bicarb transfer pump runs, but does not pump any fluid.
1. The pump has lost its prime. Ensure an adequate fluid level in the mix tank
and perform the “Head Tank Pump Re-Prime Procedure.” Refer to Section
3.5, “Bicarb Distribution Loop Purge Procedure,” in Chapter 3.
2. There is a suction (air) leak in the acid transfer pump inlet tubing.
a) Defective fitting or O-ring in any input-side tubing connection -
Bicarb Suction Fitting, Bicarb Head Tank Bulkhead Fitting (both
sides), or Transfer Pump Inlet
Fitting - Replace suspect fittings
OR:
Perform the “Leak Repair Procedure.” Refer to Section 4.2.3 in
Chapter 4.
3. Defective pump head assembly
a) Replace pump head assembly (or entire pump/motor unit).
b) (OPTION) Rebuild pump head assembly.
NOTE:
After replacing tube fittings, pump head, or performing “Tube Fitting
Maintenance Procedure,” it may be necessary to perform the “Head Tank
Pump Re-Prime” procedure to restore system to normal operation.
Problem C: Head tank leaks.
1. Inspect and repair the outlet tubing fitting and float switch connections as
necessary.
2. The high-level float switch malfunctions (tank overfilling).
3. Defective head tank – Cracked? – Replace.
Problem D: Head tank leaks only while transfer pump operates.
1. Defective seal on the top cover plate – Replace seal.
2. Check the fill tube fitting and tube connections.
3. The loop flow is low (acid or bicarb) in normal mode (gravity feed).
a) Air in loop - Refer to Chapter 3, “System Operation.”
b) Kink or occluded head tank air vent tube.
c) Kink or occluded loop tubing.
d) Defective check valve (rapid loop pump bypass) stuck closed –
Replace.
NOTE:
This valve is installed in all bicarb systems, and only in acid systems with
the Acid Loop Auxiliary Pump (Internal) installed.
Problem B: High Conductivity
WARNING:
Distribution loop problems may result in improper quantities of concentrate
fluids delivered to the dialysis machines, which can result in low or high
conductivity dialysate. If the SDS system is found to be functioning
properly, refer to the dialysis equipment for conductivity issues.
1. Incorrect acid concentrate.
2. Incorrect bicarb concentrate mixture.
3. Water flow is low (dialysate make-up water) or low pressure.
WARNING:
Do not operate either the acid loop auxiliary pump or bicarb rapid loop
distribution pump with dialysis machine connected to concentrate delivery
system. All dialysis machine concentrate lines must be disconnected from
wall dispenser concentrate access ports prior to initiation of the loop purge
procedure.
Problem C: Fluid Leaks
1. Defective fittings:
a) O-rings – Repair/replace as necessary.
b) Cracked and/or damaged fittings – Repair/replace as necessary.
2. Scratched/scored tubing - Repair/replace as necessary.
3. Fluid leaks (external) - May indicate possible locations of vacuum leaks,
which in turn cause air in loop problems.
5.10 Leaks
1. General information: Any leak from any part of the system should be
corrected as soon as possible. Left unattended, leaks can cause secondary
damage, necessitating expensive repairs. For instance, bicarb solution
leaking out of the mix pump head can enter the motor, causing damage to
the motor with subsequent expense of replacement, including parts, labor,
and system down time.
2. Refer to the “Maintenance Procedures for Leak Repair.”
A: Normal Operation
1. When the main power switch is turned ON, line power is supplied to the
PLC at L and N.
2. Three (3) LED indicators on the front of the PLC should illuminate: PWR,
OK and RUN.
NOTE:
Other LED indicators may also light, depending upon the status of the
hand switches and float switches, but the three LEDs listed in A.2 relate to
the functional status of the PLC itself. Specific input and output details are
covered in the individual component/ system sections of this problem-
solving guide.
2. TEST VOLTAGE: Check for voltage (115 VAC) at the following points, in
order. Reference any terminal „N‟ on the main terminal barrier strip as the
neutral or common connection:
a) TB1-H No Voltage
Loose wire/poor connection - Replace as necessary.
Power cord unplugged or defective.
No power at receptacle.
b) TB1-1 Voltage
NO Loose wire/poor connection - Replace as necessary.
OPEN fuse (FU-3) (primary transformer power supply)
Faulty transformer – Verify proper input and output
voltages – Replace as necessary.
OPEN fuse (FU-4) (transformer output)
YES Defective relay?
Test 1: Toggle the power switch ON/OFF a few times while
observing MCR1 for movement of the contact arm.
NO Defective MCR1 – Replace.
YES GO TO 3
3. Are the manual valves CLOSED? Ensure that the following valves are
OPEN:
RO Water Supply
SDS Water Inlet Valve (V-3)
Flow Meter Adjust Valve (V-8)
YES Normal operation. If no flow occurs, check valves.
NOTES:
CHAPTER SIX:
DRAWINGS
Illustration Notes:
1. Schematic 3004025 is a three acid system representing all available
components and is provided for reference only. Not all components
shown will be present on all systems. Specific schematics are provided
with each device. Schematic 3004018 is for single-acid systems. Not all
components shown will be present on all systems.
2. This manual includes drawings that cover all components and
configurations available. Some drawings may not be applicable to you
unit. Systems are shown with maximum configuration. User‟s system will
depend on individual order.
ME12568 Rev. E
6-12
Drawings
Solution Distribution System
2 BLACK
1
L1
3 BROWN
LINE
4
L2
P
WHITE
GROUNDED
THE FOLLOWING ARE THE STEPS REQUIRED TO RECONFIGURE A REPLACEMENT MOTOR FROM 230 TO 115
VOLT OPERATION:
1) MOTOR CAN BE WIRED FOR 230 VOLT OR 115 VOLT OPERATION. THE SDS ELECTRICAL
REQUIRES 115 VOLT CONFIGURATION.
WARNING: DISCONNECT ELECTRICAL POWER PRIOR TO SERVICING.
2) REMOVE COVER. DISCONNECT BROWN WIRE FROM TERMINAL MARKED P, THE WHITE WIRE
FROM TERMINAL MARKED 3, AND THE BLACK WIRE FROM TERMINAL 3.
3) RECONNECT WHITE WIRE TO TERMINAL MARKED 4.
4) RECONNECT BROWN WIRE TO TERMINAL MARKED 3.
5) RECONNECT BLACK WIRE TO TERMINAL MARKED 2.
6) CONNECT THE BLACK WIRE FROM 18-3 CORD (P/N ME60011) TO TERMINAL 1. CONNECT THE
WHITE WIRE TO TERMINAL 4, AND CONNECT THE GREEN WIRE TO THE GROUND SCREW.
WARNING: ROUTE WIRES AWAY FROM ROTATING CENTRIFUGAL SWITCH.
7) REPLACE COVER. REMOVE 230 VOLT STICKER FROM HOUSING. THIS WIRING IS FOR 115 VOLTS.
ME12568 Rev. E
6-24
Solution Distribution System
Drawings
Solution Distribution System
6-26
Drawings
Solution Distribution System
Illustration 6-25: Tank 2 (Acid #1) (with Auxiliary Acid Pump) Tube Connections (Red Tube)
Illustration 6-26: Tank 2 (Acid #1) (W/O Auxiliary Acid Pump) Tube Connections (Red Tube)
ME12568 Rev. E
6-32
Solution Distribution System
Drawings
ME12568 Rev. E
6-33
Drawings
Solution Distribution System
ME12568 Rev. E
6-37
Solution Distribution System
Drawings
ME12568 Rev. E
6-38
Solution Distribution System
Drawings
ME12568 Rev. E
6-39
Solution Distribution System
Drawings
Solution Distribution System
6-44
Drawings
Solution Distribution System
ME12568 Rev. E
6-48
Drawings
Solution Distribution System
NOTES:
CHAPTER SEVEN:
SPARE PARTS LIST
SDS REMOTE
SPARE PARTS LIST (Alphabetical)
FITTINGS (GENERAL)
ME40104 QD, FEMALE, ¾” BARB, CAM-LOC
ME40864 QD, FEMALE, 1” BARB, CAM-LOC
ME42092* QD, MALE, ⅜” TUBE (BICARB WAND)
ME40103 QD, MALE, ¾” MNPT, CAM-LOC
ME40863 QD, MALE, 1” MNPT, CAM-LOC
APPENDIX A:
TECHNOTES/LOG SHEETS
Technotes
In this section, you will find a variety of information related to your Mar Cor
Purification equipment and to dialysis water use in general.
Extra copies of Technotes may be ordered from Customer Service at (800) 633-
3080.
Please reference the number of the Technote, preceded by the letters TN; i.e.,
Technote 133 = TN133.
ME12568 Rev. E
Disinfectant/ Dwell Time
Confirmed Test Testing
De-Calcifier
At All Confirmed
Used
Date Time Initials Start Stop Points Of Use Post-Rinse CFU/mL EU/mL
APPENDIX A
* Refer to current standard or facility protocol for action levels
Solution Distribution System
Technotes/Log Sheets
Solution Distribution System
ME12568 Rev. E
APPENDIX A
Solution Distribution System
Technotes/Log Sheets
Solution Distribution System
ME12568 Rev. E
AAMI Bacterial Monitoring
Bicarb/Acid Loop Leak Inspection
SDS Mix Frame Leak Inspection
Bicarb/Acid Head Tank Leak Inspection
Check Bicarb Mix Tank Vent Filter
SDS Mix Frame Wash Down
** Quarterly Requirements **
Bicarb System Decalcification
Check Overflow Line for Obstructions
Check Mix Tank Seal
APPENDIX A
Check Inlet Water Filter
Check Remote Monitor (if applicable)
** Semi-Annual **
Replace SDS Inlet Water Filter
Check All Indicator Lights and Alarms
Replace Bicarb Wand Quick Disconnect
** Annual **
Change UV Lamp
Tighten All Terminals in Controller
Verify Float Switch Operation
Replace Mix Pump Seals and O-ring
Solution Distribution System
Technotes/Log Sheets
** Bi-Annual **
Rebuild Acid/Bicarb Delivery Pumps
Solution Distribution System