SYSTEM ADMINISTRATION
System Administration
Caution: If a user does not have any of the rights listed next slide, they still have
( mostly read-only ) access to System Administration. Controls on CBCT
Reconstructor tab of System Administration are available and allow CBCT
Reconstructor management.
Edit External Beam Allows changing the Custom User Settings in Service Preferences tab
Service Preferences
Sub-tabs Tabs
Configuration area
Taskbar
Caution: Enabling or Disabling hardware devices (such as Varian Respiratory Gating, Laser Guard II,
etc. ), or otherwise changing machine configuration, requires further actions in Service mode (and
System Administration - 3rd party ADI devices), see Figure 6.
Changing device configuration in System Administration modifies the copy of configuration file of
the control system node that controls the device, which resides on the workstation. Configuration
file residing on the affected control system node has to be updated in SMC, Network -> Conn Status
tab and further actions may be necessary (node reboot, initialization, etc.).
Permission Description
level
The operator does not have to use hand pendant, side panels or control
Instant
console.
Most permissive level – Axes can be moved from control console or treatment
Remote
room ( automatic or manual motions )
Axes can be moved inside the treatment room only – automatically using Auto
InRoom Motion buttons on hand pendants and /or side panels or manually using
thumbwheels
6 | VARIAN ONCOLOGY SYSTEMS on hand pendants.
Least permissive level – moving axes requires using thumbwheels on the hand
None
pendants
System Administration
Configuration/General (cont.)
• Auto Goto – Treatment application includes GoTo button, which can be used
to move the machine to a specific position other than defined in plan; when
pressed, the operator can define machine positions using edit boxes enabled
in Treatment application
Note: Setting Auto Goto to None allows the operator to use hand pendant and the thumbwheels only,
to move the machine. if this option is used, the edit box for the axis set to None will not be displayed
in Treatment application after pressing GoTo button
Note: Auto Setup for couch linear axes is set to InRoom and cannot be changed. The ability to apply
couch shifts remotely after image guided setup verification process can be configured in Treatment tab
of System Administration and is also affected by tolerance table setup levels
Configuration/General (cont.)
Important Note: the system also looks at the field tolerance table to decide where
the machine axis can be moved to. The most restrictive option from Tolerance
table and System Administration Auto Setup is then used for each axis= None
is.
= InRoom
= None
= Remote
Configuration/Accessories
Configuration/Energy Configuration
Configuration/ Synchronization
(same features as Tools/Utilities @ Home screen)
Service Preferences
Auto Start Time – can be used to specify the time and days for the TrueBeam system to
automatically cycle from StandBy to On
Tip: The system starts only if the treatment room door is closed. The operator can
override the automatic start by leaving the treatment room door open
Tools
• Containts 6 sub-tabs primarily used by Varian service personnel
Treatment/
Clinical
A
B C
A
General Preference Description
Yes – allow field grouping for treatment automation. Operator also must have appropriate right to add
Allow Automation automation for a treatment session.
No – do not allow field grouping for treatment automation. Service mode still allows automation.
Yes – requires the therapist to sign off before closing a session, after a patient has been treated. Operator
Close Patient Signoff also must have appropriate right to be allowed to signoff a treatment session.
NO – does not require the therapist to sign off before closing a session, after a patient has been treated
Yes – indicates that Enhanced Dynamic Wedge (EDW) functionality has been commissioned and enables
EDW Commissioned EDW treatments.
No – disables EDW treatments in the TrueBeam System. Service mode still allows EDW fields.
Set tolerance for photon energy override ( 0 – 3 MV ); affects machine override capability when a plan is
transferred from one machine to another.
Photon Energy
Override Tolerance For example, if the plan has been written for 6X energy, and this machine offers 8X instead, setting this
14 | VARIAN ONCOLOGY SYSTEMS
(MV) tolerance to 2 or more will allow the energy to be overridden at machine override if user also has
appropriate override user rights
verification of conical Barcode Conical Collimator Verification (BCCV) system, this setting has no effect.
collimator No – disables ICVI A
System Administration
• Allow Remote Motion – setting determines if couch shifts calculated as a result of image-
guided patient setup verification can be applied remotely from control console, if within
defined limits and allowed by Auto Setup Level for couch parameter(s) of field tolerance
table
• Limits – allows setting limits for remote motions if those are allowed. Shifts larger than
limits can only be applied to couch inside in the treatment room
Important Note: If calculated shifts are smaller than Limits it may be necessary to sign them off. The
sign-off behavior for couch shifts has changed in TrueBeam 2.0 compared to earlier versions. It is
required to sign-off the couch shifts, if they are larger than plan tolerance = tightest of plan fields
tolerance tables. ( tolerance values are compared to difference between acquisition couch position and
target couch position, not, as previously, the difference between plan couch values and target couch
values). ‘Override Setup Correction Threshold’ right is needed for sign-off
• Film Imaging Preferences – allows setting preferences for port films, such as energy, MU
and dose rate
Setup Notes
C
Treatment/Advanced
Treatment/MU limits
• To be able to treat plan with these limits the operating limits set in Eclipse TPS RT
Administration, Radiation and Imaging Devices workspace and Technique tab for the
TrueBeam machine should be the same
CBCT Reconstructor
PVA/Acquisition
• Settings on the PVA tab allow you to configure imaging function of the
TrueBeam system. The Acquisition sub-tab provides settings to control
imaging devices and to configure other acquisition parameters
• IPS – if enabled saves acquired raw images (no pts. Info) to specified IP address, supports
features like Auto Beam Hold, multi-scan, etc.
• Intrafraction Motion Management – this section contains options for Intrafraction Motion
Review (IMR) and Auto Beam Hold (ABH) imaging features
• RPM Import – this option identifies the location for importing RPM data for respiratory gating
plans
• Patient Breathing Pattern – the system automatically re-learns the breathing trace, when
couch motion is detected
25 | VARIAN ONCOLOGY SYSTEMS
Use 6DoF for Matching – if enabled, the matching workspace displays 6 DoF . Affects 4DoF
couches only
Pitch / Roll Tolerance [deg] – sets the threshold for an alert at the treatment console.
Affects 4DoF couches only
Sign-off when Couch Corrections differ from Match – if match requires shift more than
set limit, operator with sign-off is capable to proceed with the max limited value of shift (like 3
degree for Pitch)
26 | VARIAN ONCOLOGY SYSTEMS
This sub-tab allows selecting the oncology information system (OIS) connected to
TrueBeam system and can allow selecting supported features for the selected OIS
• This tab provides debugging and troubleshooting tools for Varian service and
product support specialists. Information saved in the debug files is typically of
no interest to clinical users
Disclaimer
Notice
• This information is an educational aid only and is intended solely as a
supplemental document. This provides information concerning the use of the
TrueBeam. For complete information concerning components, safety
instructions, installation, maintenance, troubleshooting, etc., refer to the
applicable Varian product documentation. This information does not replace the
Varian product documentation.
• Carefully read all instructions prior to use. Observe all contraindications,
warnings and precautions noted in these instructions. Failure to do so may result
in patient and/or user complications.
• This information is subject to change without notice and does not represent a
commitment on the part of Varian. Varian is not liable for errors contained in this
information or for incidental or consequential damages in connection with
furnishing or use of this material. This contains proprietary information protected
by copyright. No part of this may be reproduced, translated, or transmitted
without the express written permission of Varian Medical Systems, Inc.