DICOM Conformance Statement P20
DICOM Conformance Statement P20
0 CONFORMANCE STATEMENT
Ultrasound Systems (S40, S40Exp, S30, S12, S22, S9, S9Pro, S8Exp, S50, S50Exp, S50Pro, S45, S45Exp, S45Pro, P50,
P50Exp, P50Pro, P20, P20 Pro, P15, P22, P25, P10 Plus, P30VO, P40VO)
REVISION HISTORY
The Ultrasound systems implements the necessary DICOM services to download work lists from an information system,
save acquired US images to a network storage device or media, print to a networked hardcopy device and inform the
information system about the work actually done.
Table 1-1
NETWORK SERVICES
SOP Classes User of Provider of
Service (SCU) Service (SCP)
Transfer
Secondary Capture Image Storage Yes No
Ultrasound Image Storage Yes No
Ultrasound Multi-frame Image Storage Yes No
Comprehensive SR Yes No
Workflow Management
Modality Worklist Yes No
Storage Commitment Push Model Yes No
Modality Performed Procedure Step Yes No
Print Management
Basic Grayscale Print Management Yes No
Basic Color Print Management Yes No
Table 1-2 provides an overview of the Media Storage Application Profiles supported.
Table 1-2
MEDIA SERVICES
Media Storage Application Profile Write Files (FSC or FSU) Read Files (FSR)
DVD
USB
2 TABLE OF CONTENTS
2 TABLE OF CONTENTS............................................................................................................................................ 4
3 INTRODUCTION ...................................................................................................................................................... 8
4 NETWORKING ...................................................................................................................................................... 11
7 SECURTIY ............................................................................................................................................................. 54
8 ANNEXES .............................................................................................................................................................. 55
A.2. 15 TID 5025 OB-GYN Fetal Vascular Ultrasound Measurement Group .................................................... 91
A.2. 16 TID 5026 OB-GYN Pelvic Vasculatrue Section Measurement Group .................................................. 102
A.3 .4 Blood Vessel of Head DCID (12105) Intracranial Cerebral Vessels DCID (12106) Unpaired
Intracranial Cerebral Vessels (unilateral) ........................................................................................................ 117
A.3.6 Artery of Lower Extremity DCID (12109) Lower Extremity Arteries ......................................................... 119
A.3. 7 Vein of Lower Extremity DCID (12110) Lower Extremity Veins ............................................................. 119
A.3. 8 Artery of Upper Extremity DCID (12107) Upper Extremity Arteries ........................................................ 120
A.3. 9 Vein of Upper Extremity DCID (12108) Upper Extremity Veins ............................................................. 121
A.3. 10 Vascular Structure of Kidney DCID (12115) Renal Vessels ................................................................ 121
A.3. 11 Artery of Abdomen DCID (12111) Abdominal Arteries (lateral) DCID (12112) Abdominal Arteries (unilateral)
...................................................................................................................................................................... 122
A.3. 12 Vein of Abdomen DCID (12113) Abdominal Veins (lateral) DCID (12114) Abdominal Veins (unilateral)
...................................................................................................................................................................... 122
A.4.2 Left Ventricle DCID (12200) Echocardiography Left Ventricle ................................................................ 123
A.4.3 Right Ventricle DCID (12204) Echocardiography Right Ventricle ........................................................... 129
A.4.4 Left Atrium DCID (12205) Echocardiography Left Atrium ....................................................................... 130
A.4.5 Right Atrium DCID (12206) Echocardiography Right Atrium .................................................................. 131
A.4.11 Pulmonary artery DCID (12210) Echocardiography Pulmonary Artery ................................................. 141
A.4.12 Pulmonary Venous Structure DCID (12214) Echocardiography Pulmonary Veins ................................ 142
A.4.13 Lateral Mitral Annulus (12235) Echocardiography Cardiac Shunt ........................................................ 142
A.4.14 Left Ventricle Outflow Tract DCID (12223) Echocardiography Stroke Volume Origin ............................ 143
A.4.15 Right Ventricle Outflow Tract DCID (12223) Echocardiography Stroke Volume Origin ......................... 144
A.4.16 Medial Mitral Annulus DCID (12235) Mitral Valve Anatomic Sites ........................................................ 144
3 INTRODUCTION
3.1 AUDIENCE
This document is intended for hospital staff, health care system integrators, software designers or implementers. It is
assumed that the reader has a working understanding of DICOM.
3.2 REMARKS
DICOM, by itself, does not guarantee interoperability. However, the Conformance Statement facilitates a first- level
validation for interoperability between different applications supporting the same DICOM functionality.
This Conformance Statement is not intended to replace validation with other DICOM equipment to ensure proper
exchange of information intended.
The scope of this Conformance Statement is to facilitate communication between Philips Medical Systems and other
vendors’ Medical equipment. The Conformance Statement should be read and understood in conjunction with the DICOM
Standard [DICOM]. However, by itself it is not guaranteed to ensure the desired interoperability and successful
interconnectivity.
Abstract Syntax – the information agreed to be exchanged between applications, generally equivalent to a
Service/Object Pair (SOP) Class. Examples:Verification SOP Class, Modality Worklist Information Model Find
SOP Class, Computed Radiography Image Storage SOP Class.
Application Entity (AE) – an end point of a DICOM information exchange, including the DICOM network or media
interface software; i.e., the software that sends or receives DICOM information objects or messages. A single
device may have multiple Application Entities.
Application Entity Title – the externally known name of an Application Entity, used to identify a DICOM application
to other DICOM applications on the network.
Application Context – the specification of the type of communication used between Application Entities. Example:
DICOM network protocol.
Attribute – – a unit of information in an object definition; a data element identified by a tag. The information may be
a complex data structure (Sequence), itself composed of lower level data elements. Examples: Patient ID
(0010,0020), Accession Number (0008,0050), Photometric Interpretation (0028,0004), Procedure Code Sequence
(0008,1032).
Information Object Definition (IOD) – the specified set of Attributes that comprise a type of data object; does not
represent a specific instance of the data object, but rather a class of similar data objects that have the same
properties. The Attributes may be specified as Mandatory (Type 1), Required but possibly unknown (Type 2), or
Optional (Type 3), and there may be conditions associated with the use of an Attribute (Types 1C and 2C).
Examples: US Image IOD, CT Image IOD, Print Job IOD.
Module – a set of Attributes within an Information Object Definition that are logically related to each other. Example:
Patient Module includes Patient Name, Patient ID, Patient Birth Date, and Patient Sex.
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 8 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT
Negotiation – first phase of Association establishment that allows Application Entities to agree on the types of data
to be exchanged and how that data will be encoded.
Presentation Context – the set of DICOM network services used over an Association, as negotiated between
Application Entities; includes Abstract Syntaxes and Transfer Syntaxes.
Protocol Data Unit (PDU) – a packet (piece) of a DICOM message sent across the network. Devices must
specify the maximum size packet they can receive for DICOM messages.
Service Class Provider (SCP) – role of an Application Entity that provides a DICOM network service; typically, a
server that performs operations requested by another Application Entity (Service Class User). Examples: Picture
Archiving and Communication System (image storage SCP, and image query/retrieve SCP), Radiology Information
System (modality worklist SCP).
Service Class User (SCU) – role of an Application Entity that uses a DICOM network service; typically, a client.
Examples: imaging modality (image storage SCU, and modality worklist SCU), imaging workstation (image
query/retrieve SCU)
Service/Object Pair (SOP) Class – the specification of the network or media transfer (service) of a particular type
of data (object); the fundamental unit of DICOM interoperability specification. Examples: Ultrasound Image Storage
Service, Basic Grayscale Print Management.
Service/Object Pair (SOP) Instance – an information object; a specific occurrence of information exchanged in a
SOP Class. Examples: a specific x-ray image.
Tag – a 32-bit identifier for a data element, represented as a pair of four digit hexadecimal numbers, the “group” and
the “element”. If the “group” number is odd, the tag is for a private (manufacturer-specific) data element. Examples:
(0010,0020) [Patient ID], (07FE,0010) [Pixel Data], (0019,0210) [private data element]
Transfer Syntax – the encoding used for exchange of DICOM information objects and messages. Examples:
JPEG compressed (images), little endian explicit value representation.
Unique Identifier (UID) – a globally unique “dotted decimal” string that identifies a specific object or a class of
objects; an ISO-8824 Object Identifier. Examples: Study Instance UID, SOP Class UID, SOP Instance UID.
Value Representation (VR) – the format type of an individual DICOM data element, such as text, an integer, a
person’s name, or a code. DICOM information objects can be transmitted with either explicit identification of the
type of each data element (Explicit VR), or without explicit identification (Implicit VR); with Implicit VR, the receiving
application must use a DICOM data dictionary to look up the format of each data element.
3.4 ABBREVIATIONS
Abbreviations of the terms that are used have been listed here.
AE Application Entity
AET Application Entity Title
CD-R Compact Disk Recordable
CSE Customer Service Engineer
DICOM Digital Imaging and Communications in Medicine
FSC File-Set Creator
FSU File-Set Updater
FSR File-Set Reader
HIS Hospital Information System
IOD Information Object Definition
IPv4 Internet Protocol version 4
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 9 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT
3.5 REFERENCES
Referenced documents are listed below.
NEMA PS3 Digital Imaging and Communications in Medicine (DICOM) Standard, available free at
http://medical.nema.org/
Basic operator’s manual of the product
Service Manual of the product
4 NETWORKING
Remote Application
Send Storage
Entity Receives
US Images & Application
Images & SR, R e tu rn
SR, Commits Entity
s C o m m itm e n ts
Remote
Application
Update Entity Provides
Worklist Worklist Items
Workflow
Application
Entity
Acquire
Remote Application
US Images
Entity Receives
MPPS
Create/Update
Remote
Film Hardcopy Application Entity
US Images Application Prints Film Sheets
Entity
Figure 4.1-1
APPLICATION DATA FLOW DIAGRAM
— The Storage Application Entity sends images to a remote AE. It is associated with the local real-world activity “Send
US Images & SR”. “Send US Images & SR” is performed upon user request for each study completed or for specific
images selected. When activated by user’s settings (auto-send), each marked set of images can be immediately
stored to a preferred destination whenever a Patient/Study is closed by the user. If the remote AE is configured as
an archive device the Storage AE will request Storage Commitment and if a commitment is successfully obtained will
record this information in the local database.
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 11 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT
— The Workflow Application Entity receives Worklist information from and sends MPPS information to a remote AE. It is
associated with the local real-world activities “Update Worklist” and “Acquire Images”. When the “Update Worklist”
local real-world activity is performed the Workflow Application Entity queries a remote AE for worklist items and
provides the set of worklist items matching the query request. ”Update Worklist” is performed as a result of an
operator request or can be performed automatically at specific time intervals. When the “Acquire Images” local
real-world activity is performed the Workflow Application Entity creates and updates Modality Performed Procedure
Step instances managed by a remote AE. Acquisition of images will result in automated creation of an MPPS
Instance. Completion of the MPPS is performed as the result of an operator action.
— The Hardcopy Application Entity prints images on a remote AE (Printer). It is associated with the local real-world
activity “Film US Images”. “Film US Images” creates a print-job within the print queue containing one or more virtual
film sheets composed from images selected by the user.
The Workflow AE performs the creation of a MPPS Instance automatically whenever images are acquired. Further
updates on the MPPS data can be performed interactively from the related MPPS user interface. The MPPS “Complete” or
“Discontinued” states can only be set from the user interface.
1. Query Worklist
2. Receive Worklist
5. Acquire US Images
Figure 4.1-2
SEQUENCING CONSTRAINTS
Under normal scheduled workflow conditions the sequencing constraints illustrated in Figure 4.1-2 apply:
1. Query Worklist
5. Acquire US Images
9. If the Image Manager is configured as an archive device the Storage AE will request Storage Commitment for the
images and SR.
Other workflow situations (e.g. unscheduled procedure steps) will have other sequencing constraints. Printing could
equally take place after the acquired images have been stored. Printing could be omitted completely if no printer is
connected or hardcopies are not required.
4.2 AE SPECIFICATIONS
4.2.1 Storage Application Entity Specification
4.2.1.1 SOP Classes
The system provides Standard Conformance to the following SOP Classes:
Table 4.2-1
SOP CLASSES FOR AE STORAGE
SOP Class Name SOP Class UID SCU SCP
Ultrasound Image 1.2.840.10008.5.1.4.1.1.6.1 Yes No
Storage
Ultrasound 1.2.840.10008.5.1.4.1.1.3.1 Yes No
Multi-frame Image
Storage
Secondary Capture 1.2.840.10008.5.1.4.1.1.7 Yes No
Image Storage
Grayscale Softcopy 1.2.840.10008.5.1.4.1.1.88.33 Yes No
Presentation State
Storage
Storage 1.2.840.10008.1.20.1 Yes No
Commitment Push
Model
Verification 1.2.840.10008.1.1 Yes Yes
Table 4.2-2
DICOM APPLICATION CONTEXT FOR AE STORAGE
Application Context Name 1.2.840.10008.3.1.1.1
Table 4.2-3
NUMBER OF ASSOCIATIONS INITIATED FOR AE STORAGE
Maximum number of simultaneous Associations 1
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 14 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT
The system accepts Associations to receive N-EVENT-REPORT notifications for the Storage Commitment Push Model
SOP Class.
Table 4.2-4
NUMBER OF ASSOCIATIONS ACCEPTED FOR AE STORAGE
Maximum number of simultaneous Associations 1
4.2.1.2.3 Asynchronous Nature
The system does not support asynchronous communication (multiple outstanding transactions over a single Association).
Table 4.2-5
ASYNCHRONOUS NATURE AS A SCU FOR AE STORAGE
Maximum number of outstanding asynchronous 1
transactions
Table 4.2-6
DICOM IMPLEMENTATION CLASS AND VERSION FOR AE STORAGE
Implementation Class UID 1.2.156.112601
Implementation Version Name DCMTK 3.6.0
The Storage AE is invoked by the job control interface that is responsible for processing network archival tasks. The job
consists of data describing the instances marked for storage and the destination. An internal daemon process triggered by
a job for a specific network destination initiates a C-STORE request to store images. If the process successfully
establishes an Association to a remote Application Entity, it will transfer each marked instance one after another via the
open Association. Status of the transfer is reported through the job control interface. Only one job will be active at a time. If
the C-STORE Response from the remote Application contains a status other than Success or Warning, the Association is
aborted and the related Job is switched to a failed state. It can be restarted any time by user interaction or, if configured, by
automated retry.
The Storage AE attempts to initiate a new Association in order to issue a C-STORE request. If the job contains multiple
images then multiple C-STORE requests will be issued over the same Association.
If the Remote AE is configured as an archive device the Storage AE will, after all images and presentation states have been
sent, transmit a single Storage Commitment request (N-ACTION) over the same Association. Upon receiving the
N-ACTION response the Storage AE will delay releasing the Association for a configurable amount of time. If no
N-EVENT-REPORT is received within this time period the Association will be immediately released (i.e. notification of
Storage Commitment success or failure will be received over a separate association). However, the Storage AE is capable
of receiving an N-EVENT-REPORT request at any time during an association provided a Presentation Context for the
Storage Commitment Push Model has been successfully negotiated (i.e. the N-ACTION is sent at the end of one
association and the N-EVENT-REPORT is received during an association initiated for a subsequent send job or during an
association initiated by the Remote AE for the specific purpose of sending the N-EVENT-REPORT).
Storage Image
AE Manager
Manager
1. Open Association
3. C-STORE (SR)
5. C-STORE (SR)
8. Close Association
Figure 4.2-1
SEQUENCING OF ACTIVITY – SEND IMAGES
A possible sequence of interactions between the Storage AE and an Image Manager (e.g. a storage or archive device
supporting the Storage and Storage Commitment SOP Classes as an SCP) is illustrated in Figure 4.2-1:
2. An acquired US image is transmitted to the Image Manager using a C-STORE request and the Image Manager
replies with a C-STORE response (status success).
3. A SR is transmitted to the Image Manager using a C-STORE request and the Image Manager replies with a
C-STORE response (status success).
4. Another acquired US image is transmitted to the Image Manager using a C-STORE request and the Image
Manager replies with a C-STORE response (status success).
5. Another SR is transmitted to the Image Manager using a C-STORE request and the Image Manager replies with a
C-STORE response (status success).
6. An N-ACTION request is transmitted to the Image Manager to obtain storage commitment of previously transmitted
US images and SR. The Image Manager replies with a N-ACTION response indicating the request has been
received and is being processed.
7. The Image Manager immediately transmits an N-EVENT-REPORT request notifying the Storage AE of the status of
the Storage Commitment Request (sent in step 6 using the N-ACTION message). The Storage AE replies with a
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 16 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT
N-EVENT-REPORT response confirming receipt. The Image Manager could send this message at any time or
omit it entirely in favor of transmitting the N-EVENT-REPORT over a separate dedicated association (see note).
NOTE: Many other message sequences are possible depending on the number of images and SR instances to be stored, support for
Storage Commitment and when the SCP sends the N-EVENT-REPORT. The N-EVENT-REPORT can also be sent over a
separate association initiated by the Image Manager (see Section 4.2.1.4.1 on Activity – Receive Storage Commitment
Response).
Table 4.2-7
PROPOSED PRESENTATION CONTEXTS FOR ACTIVITY SEND IMAGES
Presentation Context Table
Abstract Syntax Transfer Syntax
Ext.
Name UID Name List UID List
Role Neg.
Ultrasound Image 1.2.840.10008.5.1. Implicit VR Little Endian 1.2.840.10008.1.2
Storage 4.1.1.6.1 Explicit VR Little Endian 1.2.840.10008.1.2.1 SCU None
Explicit VR Big Endian 1.2.840.10008.1.2.2
Ultrasound 1.2.840.10008.5.1. Implicit VR Little Endian 1.2.840.10008.1.2
Multi-frame Image 4.1.1.3.1 Explicit VR Little Endian 1.2.840.10008.1.2.1 SCU None
Storage Explicit VR Big Endian 1.2.840.10008.1.2.2
Secondary Capture 1.2.840.10008.5.1. Implicit VR Little Endian 1.2.840.10008.1.2
Image Storage 4.1.1.7 Explicit VR Little Endian 1.2.840.10008.1.2.1 SCU None
Explicit VR Big Endian 1.2.840.10008.1.2.2
Comprehensive 1.2.840.10008.5.1.
Implicit VR Little Endian 1.2.840.10008.1.2
Structured 4.1.1.88.33 SCU None
Report Storage Explicit VR Little Endian 1.2.840.10008.1.2.1
Presentation Contexts for Ultrasound Image Storage will only be proposed if the Send Job contains instances for these
SOP Classes.
A Presentation Context for the Storage Commitment Push Model will only be proposed if the Remote AE is configured as an
archive device.
If Ultrasound Image Storage SOP Instances are included in the Send Job and a corresponding Presentation Context is not
accepted then the Association is aborted using AP-ABORT and the send job is marked as failed. The job failure is logged
and reported to the user via the job control application. The behavior of Storage AE when encountering status codes in a
C-STORE response is summarized in the Table below:
Table 4.2-8
STORAGE C-STORE RESPONSE STATUS HANDLING BEHAVIOR
Service
Further Meaning Error Code Behavior
Status
Success Success 0000 The SCP has successfully stored the SOP Instance. If all SOP
Instances in a send job have status success then the job is marked
as complete.
Refused Out of Resources A700-A7FF The Association is aborted using A-ABORT and the send job is
marked as failed. The status meaning is logged and the job failure
is reported to the user via the job control application. This is a
transient failure.
Error Data Set does not A900-A9FF The Association is aborted using A-ABORT and the send job is
match SOP Class marked as failed. The status meaning is logged and the job failure
is reported to the user via the job control application.
Error Cannot Understand C000-CFFF The Association is aborted using A-ABORT and the send job is
marked as failed. The status meaning is logged and the job failure
is reported to the user via the job control application.
Warning Coercion of Data B000 Image transmission is considered successful but the status meaning
Elements is logged.
Warning Data Set does not B007 Image transmission is considered successful but the status meaning
match SOP Class is logged.
Warning Elements Discarded B006 Image transmission is considered successful but the status meaning
is logged.
* * Any other The Association is aborted using A-ABORT and the send job is
status code. marked as failed. The status code is logged and the job failure is
reported to the user via the job control application.
The behavior of Storage AE during communication failure is summarized in the Table below:
Table 4.2-9
STORAGE COMMUNICATION FAILURE BEHAVIOR
Exception Behavior
Timeout The Association is aborted using A-ABORT and the send job
is marked as failed. The reason is logged and the job failure
is reported to the user via the job control application.
Association aborted by the SCP or network layers The send job is marked as failed. The reason is logged and
the job failure is reported to the user via the job control
application.
A failed send job can be restarted by user interaction. The system can be configured to automatically resend failed jobs if
a transient status code is received. The delay between resending failed jobs and the number of retries is also configurable.
The contents of Ultrasound Image Storage SOP Instances created by the system conform to the DICOM Ultrasound Image
IOD definition and are described in section 8.1.
The Storage AE will consider Storage Commitment failed if no N-EVENT-REPORT is received for a Transaction UID within
a configurable time period after receiving a successful N-ACTION response (duration of applicability for a Transaction UID).
The Storage AE does not send the optional Storage Media FileSet ID & UID Attributes or the Referenced Study Component
Sequence Attribute in the N-ACTION.
The behavior of Storage AE when encountering status codes in a N-ACTION response is summarized in the Table below:
Table 4.2-10
STORAGE COMMITMENT N-ACTION RESPONSE STATUS HANDLING BEHAVIOR
Service Further
Error Code Behavior
Status Meaning
Success Success 0000 The request for storage comment is considered successfully sent. A timer is
started which will expire if no N-EVENT-REPORT for the Transaction UID is
received within a configurable timeout period.
* * Any other The Association is aborted using A-ABORT and the request for storage comment is
status marked as failed. The status meaning is logged and reported to the user.
code.
The behavior of Storage AE during communication failure is summarized in the Table below:
Table 4.2-11
STORAGE COMMITMENT COMMUNICATION FAILURE BEHAVIOR
Exception Behavior
Timeout The Association is aborted using A-ABORT and the send job
is marked as failed. The reason is logged and the job failure
is reported to the user via the job control application.
Association aborted by the SCP or network layers The send job is marked as failed. The reason is logged and
the job failure is reported to the user via the job control
application.
Upon receipt of a N-EVENT-REPORT the timer associated with the Transaction UID will be canceled.
The behavior of Storage AE when receiving Event Types within the N-EVENT-REPORT is summarized in the Table below.
Table 4.2-12
STORAGE COMMITMENT N-EVENT-REPORT BEHAVIOUR
The reasons for returning specific status codes in a N-EVENT-REPORT response are summarized in the Table below.
Table 4.2-13
STORAGE COMMITMENT N-EVENT-REPORT RESPONSE STATUS REASONS
Service Further Error
Reasons
Status Meaning Code
Success Success 0000 The storage commitment result has been successfully received.
Failure Unrecognized 0211H The Transaction UID in the N-EVENT-REPORT request is not recognized (was
Operation never issued within an N-ACTION request).
Failure Resource 0213H The Transaction UID in the N-EVENT-REPORT request has expired (no
Limitation N-EVENT-REPORT was received within a configurable time limit).
Failure No Such Event 0113H An invalid Event Type ID was supplied in the N-EVENT-REPORT request.
Type
Failure Processing 0110H An internal error occurred during processing of the N-EVENT-REPORT. A short
Failure description of the error will be returned in Error Comment (0000,0902).
Failure Invalid 0115H One or more SOP Instance UIDs with the Referenced SOP Sequence (0008,1199)
Argument or Failed SOP Sequence (0008,1198) was not included in the Storage Commitment
Value Request associated with this Transaction UID. The unrecognized SOP Instance
UIDs will be returned within the Event Information of the N-EVENT-REPORT
response.
Storage Image
AE Manager
Manager
1. Open Association
3. Close Association
Figure 4.2-2
SEQUENCING OF ACTIVITY - RECEIVE STORAGE COMMITMENT RESPONSE
A possible sequence of interactions between the Storage AE and an Image Manager (e.g. a storage or archive device
supporting Storage Commitment SOP Classes as an SCP) is illustrated in the Figure above:
1. The Image Manager opens a new association with the Storage AE.
2. The Image Manager sends an N-EVENT-REPORT request notifying the Storage AE of the status of a previous
Storage Commitment Request. The Storage AE replies with a N-EVENT-REPORT response confirming receipt.
3. The Image Manager closes the association with the Storage AE.
The Storage AE may reject association attempts as shown in the Table below. The Result, Source and Reason/Diag
columns represent the values returned in the appropriate fields of an ASSOCIATE-RJ PDU (see PS 3.8, Section 9.3.4).
The contents of the Source column is abbreviated to save space and the meaning of the abbreviations are:
a) 1 – DICOM UL service-user
Table 4.2-14
ASSOCIATION REJECTION REASONS
Result Source Reason/Diag Explanation
2– c 2 – local-limit-exceeded The (configurable) maximum number of
rejected-transient simultaneous associations has been reached.
An association request with the same
parameters may succeed at a later time.
2– c 1 – temporary-congestion No associations can be accepted at this time
rejected-transient due to the real-time requirements of higher
priority activities (e.g. during image acquisition
no associations will be accepted) or because
insufficient resources are available (e.g.
memory, processes, threads). An association
request with the same parameters may
succeed at a later time.
1– a 2– The association request contained an
rejected-permanent application-context-name-not-supported unsupported Application Context Name. An
association request with the same parameters
Table 4.2-15
ACCEPTABLE PRESENTATION CONTEXTS FOR
ACTIVITY RECEIVE STORAGE COMMITMENT RESPONSE
Presentation Context Table
Abstract Syntax Transfer Syntax
Ext.
Name UID Name List UID List
Role Neg.
Storage
1.2.840.10008.1.20 Implicit VR Little Endian 1.2.840.10008.1.2
Commitment Push SCU None
.1 Explicit VR Little Endian 1.2.840.10008.1.2.1
Model
1.2.840.10008.1.1 Implicit VR Little Endian 1.2.840.10008.1.2
Verification SCP None
Explicit VR Little Endian 1.2.840.10008.1.2.1
The Storage AE will prefer to select the Explicit VR Little Endian Transfer Syntax if multiple transfer syntaxes are offered.
The Storage AE will only accept the SCU role (which must be proposed via SCP/SCU Role Selection Negotiation) within a
Presentation Context for the Storage Commitment Push Model SOP Class.
The behavior of Storage AE when receiving Event Types within the N-EVENT-REPORT is summarized in Table 4.2-12.
The reasons for returning specific status codes in a N-EVENT-REPORT response are summarized in Table 4.2-13.
Table 4.2-16
SOP CLASSES FOR AE WORKFLOW
SOP Class Name SOP Class UID SCU SCP
Table 4.2-17
DICOM APPLICATION CONTEXT FOR AE WORKFLOW
Table 4.2-18
NUMBER OF ASSOCIATIONS INITIATED FOR AE WORKFLOW
Maximum number of simultaneous
1
Associations
Table 4.2-19
ASYNCHRONOUS NATURE AS A SCU FOR AE WORKFLOW
Maximum number of outstanding asynchronous
1
transactions
Table 4.2-20
DICOM IMPLEMENTATION CLASS AND VERSION FOR AE WORKFLOW
1.2.156.112601
Implementation Class UID
The interactive Patient Worklist Query will display a dialog for entering data as search criteria. When the Query is started on
user request, only the data from the dialog will be inserted as matching keys into the query.
With automated worklist queries (including “Worklist Update”) the system always requests all items for a Scheduled
Procedure Step Start Date (actual date), Modality (US) and Scheduled Station AE Title. Query for the Scheduled Station AE
Title is configurable by a Service Engineer.
Upon initiation of the request, The system will build an Identifier for the C-FIND request, will initiate an Association to send
the request and will wait for Worklist responses. After retrieval of all responses, the system will access the local database to
add or update patient demographic data. To protect the system from overflow, the system will limit the number of processed
worklist responses to a configurable maximum. During receiving the worklist response items are counted and the query
processing is canceled by issuing a C-FIND-CANCEL if the configurable limit of items is reached. The results will be
displayed in a separate list, which will be cleared with the next worklist update.
The system will initiate an Association in order to issue a C-FIND request according to the Modality Worklist Information
Model.
Workflow Department
AE Scheduler
1. Open Association
6. Close Association
Figure 4.2-3
SEQUENCING OF ACTIVITY – WORKLIST UPDATE
A possible sequence of interactions between the Workflow AE and a Departmental Scheduler (e.g. a device such as a RIS
or HIS which supports the Modality Worklist SOP Class as an SCP) is illustrated in the Figure above:
2. The Worklist AE sends a C-FIND request to the Departmental Scheduler containing the Worklist Query attributes.
3. The Departmental Scheduler returns a C-FIND response containing the requested attributes of the first matching
Worklist Item.
4. The Departmental Scheduler returns another C-FIND response containing the requested attributes of the second
matching Worklist Item.
5. The Departmental Scheduler returns another C-FIND response with status Success indicating that no further
matching Worklist Items exist. This example assumes that only 2 Worklist items match the Worklist Query.
7. The user selects a Worklist Item from the Worklist and prepares to acquire new images.
Table 4.2-21
PROPOSED PRESENTATION CONTEXTS FOR ACTIVITY WORKLIST UPDATE
Table 4.2-22
MODALITY WORKLIST C-FIND RESPONSE STATUS HANDLING BEHAVIOR
Service
Further Meaning Error Code Behavior
Status
Success Matching is 0000 The SCP has completed the matches. Worklist items are available for
complete display or further processing.
Refused Out of Resources A700 The Association is aborted using A-ABORT and the worklist query is
marked as failed. The status meaning is logged and reported to the user if
an interactive query. Any additional error information in the Response will
be logged.
Failed Identifier does not A900 The Association is aborted using A-ABORT and the worklist query is
match SOP Class marked as failed. The status meaning is logged and reported to the user if
an interactive query. Any additional error information in the Response will be
logged.
Failed Unable to Process C000 – The Association is aborted using A-ABORT and the worklist query is
CFFF marked as failed. The status meaning is logged and reported to the user if
an interactive query. Any additional error information in the Response will be
logged.
Cancel Matching FE00 If the query was cancelled due to too may worklist items then the SCP has
terminated due to completed the matches. Worklist items are available for display or further
Cancel request processing. Otherwise, the Association is aborted using A-ABORT and the
worklist query is marked as failed. The status meaning is logged and
reported to the user if an interactive query.
Pending Matches are FF00 The worklist item contained in the Identifier is collected for later display or
continuing further processing.
Pending Matches are FF01 The worklist item contained in the Identifier is collected for later display or
continuing – further processing. The status meaning is logged only once for each
Warning that one C-FIND operation.
or more Optional
Keys were not
supported
* * Any other The Association is aborted using A-ABORT and the worklist is marked as
status code. failed. The status meaning is logged and reported to the user if an
interactive query. Any additional error information in the Response will be
logged.
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 26 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT
The behavior of the system during communication failure is summarized in the Table below.
Table 4.2-23
MODALITY WORKLIST COMMUNICATION FAILURE BEHAVIOR
Exception Behavior
Timeout The Association is aborted using A-ABORT and the worklist query marked as failed.
The reason is logged and reported to the user if an interactive query.
Association aborted by the SCP The worklist query is marked as failed. The reason is logged and reported to the user if
or network layers an interactive query.
Acquired images will always use the Study Instance UID specified for the Scheduled Procedure Step (if available). If an
acquisition is unscheduled, a Study Instance UID will be generated locally.
The Table below provides a description of the system Worklist Request Identifier and specifies the attributes that are copied
into the images. Unexpected attributes returned in a C-FIND response are ignored.
Requested return attributes not supported by the SCP are set to have no value. Non-matching responses returned by the
SCP due to unsupported optional matching keys are ignored. No attempt is made it filter out possible duplicate entries.
Table 4.2-24
WORKLIST REQUEST IDENTIFIER
Module Name Tag VR M R Q D IOD
Attribute Name
SOP Common
Specific Character Set (0008,0005) CS x
Scheduled Procedure Step
Scheduled Procedure Step Sequence (0040,0100) SQ x
> Scheduled Station AET (0040,0001) AE (S) x x
> Scheduled Procedure Step Start Date (0040,0002) DA S x x
> Scheduled Procedure Step Start Time (0040,0003) TM x x
> Modality (0008,0060) CS S x
> Scheduled Performing Physician’s Name (0040,0006) PN x x x x
> Scheduled Procedure Step Description (0040,0007) LO x x x
> Scheduled Station Name (0040,0010) SH S x
> Scheduled Procedure Step Location (0040,0011) SH S x
> Scheduled Protocol Code Sequence (0040,0008) SQ x x
> Pre-Medication (0040,0012) LO x x
> Scheduled Procedure Step ID (0040,0009) SH x x x
> Requested Contrast Agent (0032,1070) LO x x
Requested Procedure
Requested Procedure ID (0040,1001) SH S x x x x
Requested Procedure Description (0032,1060) LO x x x
Study Instance UID (0020,000D) UI x x
Requested Procedure Priority (0040,1003) SH x
Patient Transport Arrangements (0040,1004) LO x
Referenced Study Sequence (0008,1110) SQ x x
Requested Procedure Code Sequence (0032,1064) SQ x x
Module Name: The name of the associated module for supported worklist attributes.
Attribute Name: Attributes supported to build an Worklist Request Identifier.
Tag: DICOM tag for this attribute.
VR: DICOM VR for this attribute.
M: Matching keys for (automatic) Worklist Update. A "S" will indicate that the system will supply an attribute
value for Single Value Matching, a “R” will indicate Range Matching and a “*” will denote wildcard
matching. It can be configured if “Scheduled Station AE Title” is additionally supplied “(S)” and if Modality
is set to US or SC.
R: Return keys. An "x" will indicate that the system will supply this attribute as Return Key with zero length
for Universal Matching. The system will support retired date format (yyyy.mm.dd) for "Patient’s Birth
Date" and "Scheduled Procedure Step Start Date" in the response identifiers. For "Scheduled Procedure
Step Start Time" also retired time format as well as unspecified time components are supported.
Q: Interactive Query Key. An “x” " will indicate that the system will supply this attribute as matching key, if
entered in the Query Patient Worklist dialog. For example, the Patient Name can be entered thereby
restricting Worklist responses to Procedure Steps scheduled for the patient.
D: Displayed keys. An “x” indicates that this worklist attribute is displayed to the user during a patient
registration dialog. For example, Patient Name will be displayed when registering the patient prior to an
examination.
IOD: An "x" indicates that this Worklist attribute is included into all Object Instances created during
performance of the related Procedure Step.
The default Query Configuration is set to "Modality" (US) and "Date" (date of today). Optionally, additional matching for the
own AET is configurable.
A manual update can be performed with the MPPS user interface where is it possible to set the final state of the MPPS to
“COMPLETED” or “DISCONTINUED”. In the “Discontinued” case the user can also select the discontinuation reason from a
list corresponding to Context Group 9300. A MPPS Instance that has been sent with a state of “COMPLETED” or
“DISCONTINUED” can no longer be updated.
The system will support creation of “unscheduled cases” by allowing MPPS Instances to be communicated for locally
registered Patients.
The system only supports a 0-to-1 relationship between Scheduled and Performed Procedure Steps.
— N-CREATE request according to the CREATE Modality Performed Procedure Step SOP Instance operation or a
— N-SET request to update the contents and state of the MPPS according to the SET Modality Performed Procedure Step
Information operation.
Workflow Department
AE Scheduler
1. Open Association
3. Close Association
4. Acquire Images
5. Open Association
7. Close Association
Figure 4.2-4
SEQUENCING OF ACTIVITY – ACQUIRE IMAGES
A possible sequence of interactions between the Workflow AE and a Departmental Scheduler (e.g. a device such as a RIS
or HIS which supports the MPPS SOP Class as an SCP) is illustrated in Figure 4.2-4:
2. The Worklist AE sends an N-CREATE request to the Departmental Scheduler to create an MPPS instance with
status of “IN PROGRESS” and create all necessary attributes. The Departmental Scheduler acknowledges the
MPPS creation with an N-CREATE response (status success).
6. The Worklist AE sends an N-SET request to the Departmental Scheduler to update the MPPS instance with status
of “COMPLETED” and set all necessary attributes. The Departmental Scheduler acknowledges the MPPS update
with an N-SET response (status success).
Table 4.2-25
PROPOSED PRESENTATION CONTEXTS FOR REAL-WORLD ACTIVITY ACQUIRE IMAGES
Presentation Context Table
Abstract Syntax Transfer Syntax
Ext.
Name UID Name List UID List
Role Neg.
Implicit VR Little Endian 1.2.840.10008.1.2
Modality Performed 1.2.840.10008.3.1.
Explicit VR Little Endian 1.2.840.10008.1.2.1 SCU None
Procedure Step 2.3.3
Explicit VR Big Endian 1.2.840.10008.1.2.2
Table 4.2-26
MPPS N-CREATE / N-SET RESPONSE STATUS HANDLING BEHAVIOR
Service Error
Further Meaning Behavior
Status Code
Success Success 0000 The SCP has completed the operation successfully.
Failure Processing Failure – 0110 The Association is aborted using A-ABORT and the MPPS is
Performed Procedure Step marked as failed. The status meaning is logged and reported to
Object may no longer be the user. Additional information in the Response will be logged
updated (i.e. Error Comment and Error ID).
Warning Attribute Value Out of Range 0116H The MPPS operation is considered successful but the status
meaning is logged. Additional information in the Response
identifying the attributes out of range will be logged (i.e. Elements in
the Modification List/Attribute List)
* * Any other The Association is aborted using A-ABORT and the MPPS is
status marked as failed. The status meaning is logged and reported to
code. the user.
The behavior of the system during communication failure is summarized in the Table below:
Table 4.2-27
MPPS COMMUNICATION FAILURE BEHAVIOR
Exception Behavior
Timeout The Association is aborted using A-ABORT and MPPS marked as failed. The
reason is logged and reported to the user.
Association aborted by the SCP or The MPPS is marked as failed. The reason is logged and reported to the user.
network layers
Table 4.2-28 provides a description of the MPPS N-CREATE and N-SET request identifiers sent by the system. Empty
cells in the N-CREATE and N-SET columns indicate that the attribute is not sent. An “x” indicates that an appropriate value
will be sent. A “Zero length” attribute will be sent with zero length.
Table 4.2-28
MPPS N-CREATE / N-SET REQUEST IDENTIFIER
Attribute Name Tag VR N-CREATE N-SET
Specific Character Set (0008,0005) CS “ISO_IR 100” or “ISO_IR 144”
Modality (0008,0060) CS US
Referenced Patient
(0008,1120) SQ Zero length
Sequence
From Modality Worklist or user
input (all 5 components). The
Patient’s Name (0010,0010) PN
user can modify values
provided via Modality Worklist.
From Modality Worklist or user
input. The user can modify
Patient ID (0010,0020) LO
values provided via Modality
Worklist.
From Modality Worklist or user
input. The user can modify
Patient’s Birth Date (0010,0030) DA
values provided via Modality
Worklist.
From Modality Worklist or user
input. The user can modify
Patient’s Sex (0010,0040) CS
values provided via Modality
Worklist.
Table 4.2-29
SOP CLASSES FOR AE HARDCOPY
SOP Class Name SOP Class UID SCU SCP
Table 4.2-30
DICOM APPLICATION CONTEXT FOR AE HARDCOPY
Table 4.2-31
NUMBER OF ASSOCIATIONS INITIATED FOR AE HARDCOPY
Table 4.2-32
ASYNCHRONOUS NATURE AS A SCU FOR AE HARDCOPY
Table 4.2-33
DICOM IMPLEMENTATION CLASS AND VERSION FOR AE HARDCOPY
1.2.156.112601
Implementation Class UID
The Hardcopy AE is invoked by the job control interface that is responsible for processing network tasks. The job consists of
data describing the images and graphics to be printed as well as the requested layout and other parameters. The film
sheet is internally processed, converted to a STANDARD/1,1 page and then the page image is sent. If no association to
the printer can be established, the print-job is switched to a failed state and the user informed.
Hardcopy Printer
AE
1. Open Association
2. N-GET (Printer)
Figure 4.2-5
SEQUENCING OF ACTIVITY – FILM IMAGES
A typical sequence of DIMSE messages sent over an association between Hardcopy AE and a Printer is illustrated in Figure
4.2-5:
2. N-GET on the Printer SOP Class is used to obtain current printer status information. If the Printer reports a status
of FAILURE, the print-job is switched to a failed state and the user informed.
4. N-CREATE on the Presentation LUT SOP Class creates a Presentation LUT (if supported by the printer).
5. N-CREATE on the Film Box SOP Class creates a Film Box linked to the Film Session. A single Image Box will be
created as the result of this operation (Hardcopy AE only uses the format STANDARD\1,1)
6. N-SET on the Image Box SOP Class transfers the contents of the film sheet to the printer. If the printer does not
support the Presentation LUT SOP Class, the image data will be passed through a printer-specific correction LUT
before being sent.
7. N-ACTION on the Film Box SOP Class instructs the printer to print the Film Box
9. The Printer asynchronously reports its status via N-EVENT-REPORT notification (Printer SOP Class). The printer
can send this message at any time. Hardcopy AE does not require the N-EVENT-REPORT to be sent. Hardcopy
AE is capable of receiving an N-EVENT-REPORT notification at any time during an association. If the Printer
reports a status of FAILURE, the print-job is switched to a failed state and the user informed.
10. N-DELETE on the Film Session SOP Class deletes the complete Film Session SOP Instance hierarchy.
Status of the print-job is reported through the job control interface. Only one job will be active at a time for each separate
hardcopy device. If any Response from the remote Application contains a status other than Success or Warning, the
Association is aborted and the related Job is switched to a failed state. It can be restarted any time by user interaction or, if
configured, by automated retry.
Table 4.2-34
PROPOSED PRESENTATION CONTEXTS FOR ACTIVITY FILM IMAGES
Presentation Context Table
Abstract Syntax Transfer Syntax
Ext.
Name UID Name List UID List
Role Neg.
Basic Grayscale Implicit VR Little Endian 1.2.840.10008.1.2
1.2.840.10008.5.1.
Print Management Explicit VR Little Endian 1.2.840.10008.1.2.1 SCU None
1.9
Meta Explicit VR Big Endian 1.2.840.10008.1.2.2
Implicit VR Little Endian 1.2.840.10008.1.2
Basic Color Print 1.2.840.10008.5.1.
Explicit VR Little Endian 1.2.840.10008.1.2.1 SCU None
Management Meta 1.18
Explicit VR Big Endian 1.2.840.10008.1.2.2
Implicit VR Little Endian 1.2.840.10008.1.2
Basic Annotation 1.2.840.10008.5.1.
Explicit VR Little Endian 1.2.840.10008.1.2.1 SCU None
Box SOP Class 1.15
Explicit VR Big Endian 1.2.840.10008.1.2.2
Implicit VR Little Endian 1.2.840.10008.1.2
1.2.840.10008.5.1.
Print Job SOP Class Explicit VR Little Endian 1.2.840.10008.1.2.1 SCU None
1.14
Explicit VR Big Endian 1.2.840.10008.1.2.2
Implicit VR Little Endian 1.2.840.10008.1.2
1.2.840.10008.5.1.
Presentation LUT Explicit VR Little Endian 1.2.840.10008.1.2.1 SCU None
1.23
Explicit VR Big Endian 1.2.840.10008.1.2.2
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 36 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT
4.2.3.3.1.3 Common SOP Specific Conformance for all Print SOP Classes
The general behavior of Hardcopy AE during communication failure is summarized in the Table below. This behavior is
common for all SOP Classes supported by Hardcopy AE.
Table 4.2-35
HARDCOPY COMMUNICATION FAILURE BEHAVIOR
Exception Behavior
Timeout The Association is aborted using A-ABORT and the print-job
is marked as failed. The reason is logged and the job failure
is reported to the user via the job control application.
Association aborted by the SCP or network layers The print-job is marked as failed. The reason is logged and
the job failure is reported to the user via the job control
application.
— N-GET
— N-EVENT-REPORT
Details of the supported attributes and status handling behaviour are described in the following subsections.
Table 4.2-36
PRINTER SOP CLASS N-GET REQUEST ATTRIBUTES
Presence of
Attribute Name Tag VR Value Source
Value
Printer Status (2110,0010) CS Provided by Printer ALWAYS Printer
Printer Status Info (2110,0020) CS Provided by Printer ALWAYS Printer
2. If Printer status (2110,0010) is FAILURE, the print-job is marked as failed. The contents of Printer Status Info
(2110,0020) is logged and reported to the user via the job control application.
3. If Printer status (2110,0010) is WARNING, the print-job continues to be printed. The contents of Printer Status Info
(2110,0020) is logged and reported to the user via the job control application.
The behavior of Hardcopy AE when encountering status codes in a N-GET response is summarized in the Table below:
Table 4.2-37
PRINTER SOP CLASS N-GET RESPONSE STATUS HANDLING BEHAVIOR
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 37 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT
Service Further
Error Code Behavior
Status Meaning
Success Success 0000 The request to get printer status information was success.
* * Any other The Association is aborted using A-ABORT and the print-job is marked as
status code. failed. The status meaning is logged and reported to the user.
The behavior of Hardcopy AE when receiving Event Types within the N-EVENT-REPORT is summarized in the Table
below:
Table 4.2-38
PRINTER SOP CLASS N-EVENT-REPORT BEHAVIOUR
Event Type Event Behavior
Name Type ID
Normal 1 The print-job continues to be printed.
Warning 2 The print-job continues to be printed. The contents of Printer Status Info (2110,0020) is
logged and reported to the user via the job-control application.
Failure 3 The print-job is marked as failed. The contents of Printer Status Info (2110,0020) is logged
and reported to the user via the job-control application.
* * An invalid Event Type ID will cause a status code of 0113H to be returned in a
N-EVENT-REPORT response.
The reasons for returning specific status codes in a N-EVENT-REPORT response are summarized in the Table below:
Table 4.2-39
PRINTER SOP CLASS N-EVENT-REPORT RESPONSE STATUS REASONS
Service Further Error
Reasons
Status Meaning Code
Success Success 0000 The notification event has been successfully received.
Failure No Such Event 0113H An invalid Event Type ID was supplied in the N-EVENT-REPORT request.
Type
Failure Processing 0110H An internal error occurred during processing of the N-EVENT-REPORT. A short
Failure description of the error will be returned in Error Comment (0000,0902).
4.2.3.3.1.5 SOP Specific Conformance for the Film Session SOP Class
Hardcopy AE supports the following DIMSE operations for the Film Session SOP Class:
— N-CREATE
— N-DELETE
Details of the supported attributes and status handling behavior are described in the following subsections.
Table B4.2-40
FILM SESSION SOP CLASS N-CREATE REQUEST ATTRIBUTES
Presence of
Attribute Name Tag VR Value Source
Value
Number of
(2000,0010) IS 1 .. 10 ALWAYS User
Copies
Print
(2000,0020) CS LOW, MED or HIGH ALWAYS User
Priority
Medium Type (2000,0030) CS BLUE FILM, CLEAR FILM or PAPER ALWAYS User
Print
(2000,0020) CS LOW, MED or HIGH ALWAYS User
Priority
Film Destination (2000,0040) CS MAGAZINE or PROCESSOR ALWAYS User
Film Session
(2000,0050) LO User
Label
The behavior of Hardcopy AE when encountering status codes in a N-CREATE response is summarized in the Table below:
Table 4.2-41
FILM SESSION SOP CLASS N-CREATE RESPONSE STATUS HANDLING BEHAVIOR
Service Further Error
Behavior
Status Meaning Code
Success Success 0000 The SCP has completed the operation successfully.
Warning Attribute Value 0116H The N-CREATE operation is considered successful but the status meaning is
Out of Range logged. Additional information in the Response identifying the attributes out of
range will be logged (i.e. Elements in the Modification List/Attribute List)
Warning Attribute List 0107H The N-CREATE operation is considered successful but the status meaning is
Error logged. Additional information in the Response identifying the attributes will be
logged (i.e. Elements in the Attribute Identifier List)
* * Any other The Association is aborted using A-ABORT and the print-job is marked as failed.
status The status meaning is logged and reported to the user.
code.
Table 4.2-42
PRINTER SOP CLASS N-DELETE RESPONSE STATUS HANDLING BEHAVIOR
Service Further
Error Code Behavior
Status Meaning
Success Success 0000 The SCP has completed the operation successfully.
* * Any other The Association is aborted using A-ABORT and the print-job is marked as
status code. failed. The status meaning is logged and reported to the user.
4.2.3.3.1.6 SOP Specific Conformance for the Presentation LUT SOP Class
Hardcopy AE supports the following DIMSE operations for the Presentation LUT SOP Class:
— N-CREATE
Details of the supported attributes and status handling behavior are described in the following subsections.
Table 4.2-43
PRESENTATION LUT SOP CLASS N-CREATE REQUEST ATTRIBUTES
Presence of
Attribute Name Tag VR Value Source
Value
Presentation LUT
(2050,0020) CS IDENTITY ALWAYS Auto
Shape
The behavior of Hardcopy AE when encountering status codes in a N-CREATE response is summarized in the Table below:
Table 4.2-44
PRESENTATION LUT SOP CLASS N-CREATE RESPONSE STATUS HANDLING BEHAVIOR
Service
Further Meaning Error Code Behavior
Status
Success Success 0000 The SCP has completed the operation successfully.
Warning Requested Min Density or Max B605H The N-CREATE operation is considered successful but the
Density outside of printer’s status meaning is logged.
operating range
* * Any other The Association is aborted using A-ABORT and the
status code. print-job is marked as failed. The status meaning is
logged and reported to the user.
Service Further Meaning Error Code Behavior
Status
Success Success 0000 The SCP has completed the operation successfully.
4.2.3.3.1.7 SOP Specific Conformance for the Film Box SOP Class
Hardcopy AE supports the following DIMSE operations for the Presentation LUT SOP Class:
— N-CREATE
— N-ACTION
Details of the supported attributes and status handling behavior are described in the following subsections.
Table 4.2-45
FILM BOX SOP CLASS N-CREATE REQUEST ATTRIBUTES
Presence of
Attribute Name Tag VR Value Source
Value
STANDARD\1,1 depending on printer. Default is
Image Display
(2010,0010) CS displayed, and user editable. Edit only when a ALWAYS Auto/User
Format
vaild substitute value is known.
Referenced Film
Session (2010,0500) SQ ALWAYS Auto
Sequence
>Referenced
(0008,1150) UI 1.2.840.10008.5.1.1.1 ALWAYS Auto
SOP Class UID
>Referenced
SOP Instance (0008,1155) UI From created Film Session SOP Instance ALWAYS Auto
UID
Film Orientation (2010,0040) CS PORTRAIT or LANDSCAPE ALWAYS User
8INX10IN, 8_5INX11IN, 10INX12IN,
10INX14IN, 11INX14IN, 11INX17IN;
Film Size ID (2010,0050) CS ALWAYS User
14INX14IN, 14INX17IN, 24CMX24CM,
24CMX30CM, A4, A3
Magnification
(2010,0060) CS REPLICATE, BILINEAR, CUBIC or NONE ALWAYS User
Type
Border Density (2010,0100) CS BLACK or WHITE ALWAYS User
Max Density (2010,0130) US 0 .. 310 Default is displayed, and user editable. ALWAYS Auto
Min Density (2010,0120) US 0 .. 50 Default is displayed, and user editable. ALWAYS Auto
Smoothing Type (2010,0080) US 0 .. 100 Default is displayed, and user editable. ALWAYS User
Empty Image
(2010,0110) CS BLACK or WHITE ALWAYS User
Density
Trim (2010,0140) CS YES or NO ALWAYS User
Configuration Default is displayed, and user editable. Edit only
(2010,0150) ST ALWAYS User
Information when a vaild substitute value is known.
Referenced
Only sent if Presentation LUT SOP Class has
Presentation LUT (2050,0500) SQ ANAP Auto
been negotiated.
Sequence
>Referenced
(0008,1150) UI 1.2.840.10008.5.1.1.23 ALWAYS Auto
SOP Class UID
>Referenced
SOP Instance (0008,1155) UI From created Presentation LUT SOP Instance ALWAYS Auto
UID
The behavior of Hardcopy AE when encountering status codes in a N-CREATE response is summarized in the Table below:
Table 4.46
FILM BOX SOP CLASS N-CREATE RESPONSE STATUS HANDLING BEHAVIOR
Service
Further Meaning Error Code Behavior
Status
Success Success 0000 The SCP has completed the operation successfully.
Warning Requested Min Density or Max B605H The N-CREATE operation is considered successful but the
Density outside of printer’s status meaning is logged.
operating range
* * Any other The Association is aborted using A-ABORT and the
status code. print-job is marked as failed. The status meaning is
logged and reported to the user.
The behavior of Hardcopy AE when encountering status codes in a N-ACTION response is summarized in the Table below:
Table 4.2-47
FILM BOX SOP CLASS N-ACTION RESPONSE STATUS HANDLING BEHAVIOR
Service Error
Further Meaning Behavior
Status Code
Success Success 0000 The SCP has completed the operation
successfully. The film has been accepted for
printing.
Warning Film Box SOP Instance hierarchy does not B603H The Association is aborted using A-ABORT and
contain Image Box SOP Instances (empty the print-job is marked as failed. The status
page) meaning is logged and reported to the user.
Warning Image size is larger than Image Box size. The B604H The N-ACTION operation is considered
image has been demagnified. successful but the status meaning is logged.
Warning Image size is larger than Image Box size. The B609H The N-ACTION operation is considered
image has been cropped to fit. successful but the status meaning is logged.
Warning Image size or Combined Print Image Size is B60AH The N-ACTION operation is considered
larger than Image Box size. The image or successful but the status meaning is logged.
combined Print Image has been decimated to
fit.
Failure Unable to create Print Job SOP Instance; print C602 The Association is aborted using A-ABORT and
queue is full. the print-job is marked as failed. The status
meaning is logged and reported to the user.
Failure Image size is larger than Image Box size. C603 The Association is aborted using A-ABORT and
the print-job is marked as failed. The status
meaning is logged and reported to the user.
Failure Combined Print Image Size is larger than Image C613 The Association is aborted using A-ABORT and
Box size. the print-job is marked as failed. The status
meaning is logged and reported to the user.
* * Any other The Association is aborted using A-ABORT and
status the print-job is marked as failed. The status
code. meaning is logged and reported to the user.
4.2.3.3.1.8 SOP Specific Conformance for the Image Box SOP Class
Hardcopy AE supports the following DIMSE operations for the Image Box SOP Class:
— N-SET
Details of the supported attributes and status handling behavior are described in the following subsections.
Table 4.2-48-1
BASIC GRAYSCALE IMAGE BOX SOP CLASS N-SET REQUEST ATTRIBUTES
Presence of
Attribute Name Tag VR Value Source
Value
Image Position (2020,0010) US 1 ALWAYS Auto
Basic Grayscale
(2020,0110) SQ ALWAYS Auto
Image Sequence
>Samples Per Pixel (0028,0002) US 1 ALWAYS Auto
>Photometric
(0028,0004) CS MONOCHROME2 ALWAYS Auto
Interpretation
>Rows (0028,0010) US Depends on film size ALWAYS Auto
>Columns (0028,0011) US Depends on film size ALWAYS Auto
>Pixel Aspect Ratio (0028,0034) IS 1\1 ALWAYS Auto
>Bits Allocated (0028,0100) US 8 ALWAYS Auto
>Bits Stored (0028,0101) US 8 ALWAYS Auto
>High Bit (0028,0102) US 7 ALWAYS Auto
>Pixel
(0028,0103) US 0 ALWAYS Auto
Representation
>Pixel Data (7FE0,0010) OB Pixels of rendered film sheet ALWAYS Auto
Table 4.2-48
BASIC GRAYSCALE OR BASIC COLOR IMAGE BOX SOP CLASS N-SET REQUEST ATTRIBUTES
Presence of
Attribute Name Tag VR Value Source
Value
Image Position (2020,0010) US 1 ALWAYS Auto
Basic Grayscale
(2020,0110) SQ Used for BW (MONOCHROME2) print ALWAYS Auto
Image Sequence
Basic Color Image
(2020,0111) SQ Used for color (RGB) print ALWAYS Auto
Sequence
1 for MONOCHROME2
>Samples Per Pixel (0028,0002) US ALWAYS Auto
3 for RGB
>Photometric
(0028,0004) CS MONOCHROME2 or RGB ALWAYS Auto
Interpretation
Planar
(0028,0006) US Always “1”,only used for RGB print ANAP Auto
Configuration
>Rows (0028,0010) US Depends on film size ALWAYS Auto
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 43 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT
The behavior of Hardcopy AE when encountering status codes in a N-SET response is summarized in the Table below:
Table 4.2-49
IMAGE BOX SOP CLASS N-SET RESPONSE STATUS HANDLING BEHAVIOR
Service Error
Further Meaning Behavior
Status Code
Success Success 0000 The SCP has completed the operation
successfully. Image successfully stored in
Image Box.
Warning Image size is larger than Image Box size. The B604H The N-SET operation is considered successful
image has been demagnified. but the status meaning is logged.
Warning Requested Min Density or Max Density outside B605H The N-SET operation is considered successful
of printer’s operating range. but the status meaning is logged.
Warning Image size is larger than Image Box size. The B609H The N-SET operation is considered successful
image has been cropped to fit. but the status meaning is logged.
Warning Image size or Combined Print Image Size is B60AH The N-SET operation is considered successful
larger than Image Box size. The image or but the status meaning is logged.
combined Print Image has been decimated to
fit.
Failure Image size is larger than Image Box size. C603 The Association is aborted using A-ABORT and
the print-job is marked as failed. The status
meaning is logged and reported to the user.
Failure Insufficient memory in printer to store the C605 The Association is aborted using A-ABORT and
image. the print-job is marked as failed. The status
meaning is logged and reported to the user.
Failure Combined Print Image Size is larger than Image C613 The Association is aborted using A-ABORT and
Box size. the print-job is marked as failed. The status
meaning is logged and reported to the user.
* * Any other The Association is aborted using A-ABORT and
status the print-job is marked as failed. The status
code. meaning is logged and reported to the user.
Table 4.2-50
SOP CLASSES FOR AE
VERIFICATION
SOP Class Name SOP Class UID SCU SCP
Table 4.2-51
DICOM APPLICATION CONTEXT FOR AE
VERIFICATION
Table 4.2-52
NUMBER OF ASSOCIATIONS INITIATED FOR AE VERIFICATION
Table 4.2-53
NUMBER OF ASSOCIATIONS ACCEPTED FOR AE VERIFICATION
Table 4.2-54
ASYNCHRONOUS NATURE AS A SCU FOR AE VERIFICATION
Table 4.2-55
DICOM IMPLEMENTATION CLASS AND VERSION FOR AE VERIFICATION
1.2.156.112601
Implementation Class UID
SCP: The system listens on the port configured on the System Settings Configuration screen for Verification requests
initiated by other remote devices. The calling device AE must already be configured as a remote device in the
system or the association is rejected.
Verification Verification
SCU SCP
Manager
1. Open Association
2. C-ECHO Request
3. C-ECHO Response
4. Close Association
Figure 4.2-6
SEQUENCING OF ACTIVITY – ISSUE VERIFY
Verification Verification
SCP SCU
Manager
1. Open Association
2. C-ECHO Request
3. C-ECHO Response
4. Close Association
Figure 4.2-7
SEQUENCING OF ACTIVITY – RECEIVE VERIFY
4.2.4.3.3 Proposed Presentation Contexts
The system will propose Presentation Contexts as shown in the following table:
Table 4.2-56
PROPOSED PRESENTATION CONTEXTS FOR ACTIVITY VERIFICATION
Presentation Context Table
Abstract Syntax Transfer Syntax
Ext.
Name UID Name List UID List
Role Neg.
Implicit VR Little Endian 1.2.840.10008.1.2 SCU
Verification 1.2.840.10008.1.1 None
Explicit VR Little Endian 1.2.840.10008.1.2.1 /SCP
A message will appear on the user interface if the system receives any other SCP response status than “Success.”
Table 4.2-57
VERIFICATION C-ECHO RESPONSE STATUS HANDLING BEHAVIOR
Service
Further Meaning Error Code Behavior
Status
Success 0000 Device Status is set to: Verified
Refused Out of Resources A700 Device Status is set to: Not Verified
Failed Unable to Process C000 – CFFF Same as “Refused” above.
* * Any other status code. Same as “Refused” above.
4.2.4.3.4 Association Acceptance Policy
4.2.4.3.5.1 Verification SOP Class Notifications
Possible Responses when “Echo” is used in Global Device configuration:
Note: A given “Archive” server may not support all of the SOP Classes requested in the Verification request.
Receiving failures (“Not Verified”) responses for SOP Classes outside the scope or capability of the server will not
result in a communications failure.
For example, if the correct Image Store SOP Classes are supported and Structured Report is not, then Image Storage
will work successfully. If multiframe is not supported and loops are acquired, the transfer will fail. In this case however,
the single frame images will transfer.
Table 4.3-1
SUPPORTED PHYSICAL NETWORK INTERFACES
Ethernet 10/100/1000*BaseT, RJ-45, UTP, STP; AutoDetect Speed, Full or Half Duplex
4.3.3 IPv4 and IPv6 Support
This product only supports IPv4 connections.
4.4 CONFIGURATION
DICOM and networking parameters can be configured for both the local, the system device and remote DICOM service
class providers through the System Settings DICOM Menu.
4.4.1 Local
The system local network parameters are configurable. The following network parameters can be configured for a device:
• Host Name
• IP address
• Subnet IP mask
• Default Gateway
• DICOM Application Entity Title
4.4.2 Remote
Multiple DICOM service class providers can be configured through the system presets.
4.4.2.1 DICOM Store Configuration
Configuration of DICOM Modality Worklist remote devices is enabled in the system.
• DICOM Device Application Entity Title
• IP address
• Port Number
• Max retries, Retry interval, Timeout
• Send After Every Image Stored or Send At End Of Exam
4.4.2.2 DICOM Storage Commitment Configuration
Configuration of DICOM Storage Commitment remote devices must be performed separately from DICOM Store
Configuration. The system supports Storage Commitment to the same remote device as Store or to a different device.
• DICOM Device Application Entity Title
• IP address
• Port Number
• Max retries, Retry interval, Timeout
4.4.2.3 DICOM Modality Worklist Configuration
Configuration of DICOM Modality Worklist remote devices is enabled in the system.
• DICOM Device Application Entity Title
• IP address
• Port Number
• Max retries, Retry interval, Timeout
• Maximum number of downloaded entries
• Port Number
• Max retries, Retry interval, Timeout
• Configuration for each job according to attribute description in Section 4.2.3.3.1.5 of this document
5 MEDIA INTERCHANGE
Export
to Media
Media Media
Application Storage
Entity Medium
Update
to Media
Figure 5.1-1
APPLICATION DATA FLOW DIAGRAM FOR MEDIA STORAGE
— The Media Application Entity exports images and Structured Reports to a disk Storage medium. It is associated with
the local real-world activity “Export to Media” and “Update to Media”. “Export to Media” and “Update to Media” is
performed upon user request for selected patients, studies, series or instances (images or Structured Reports).
Table 5.1-1
DICOM IMPLEMENTATION CLASS AND VERSION FOR MEDIA STORAGE
1.2.156.112601
Implementation Class UID
5.2 AE SPECIFICATIONS
5.2.1 Media Application Entity Specification
The Media Application Entity provides standard conformance to the Media Storage Service Class. The Application
Profiles and roles are listed below:
Table 5.2-1
APPLICATION PROFILES, ACTIVITIES AND ROLES FOR MEDIA
Application Profiles Supported Real World Activity Role
STD-US-ID-SF&MF-CDR Export to Media FSC
STD-US-ID-SF&MF-DVD Update to Media FSU
STD-US-SC-SF&MF-CDR
STD-US-SC-SF&MF-DVD
STD-GEN-USB
A dialogue will be presented allowing the user to modify the suggested media label and provides control over the available
media capacity. If the contents of the current selection do not fit on a single media an automatic separation into multiple
export jobs will be suggested which can be adapted by the user.
The user will be prompted to insert an empty media for each export job. The contents of the export job will be written
together with a corresponding DICOMDIR to media. Writing in multi-session mode is not supported. The user can cancel an
export job in the job queue.
The system user selects exams from the system’s directory for transfer to media that already contains data. The
DICOMDIR is updated allowing access to original and new data.
DVD+RW or CDR+RW media may be erased at any time, removing all previously recorded data.
5.2.1.2.1.1.1 Options
The Media Application Entity supports the SOP Classes and Transfer Syntaxes listed in the Table below:
Table 5.2-2
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 51 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT
Information Object Definition SOP Class UID Transfer Syntax Transfer Syntax UID
Explicit VR Little
Media Storage Directory Storage 1.2.840.10008.1.3.10 1.2.840.10008.1.2.1
Endian
Explicit VR Little
Ultrasound Image Storage 1.2.840.10008.5.1.4.1.1.6.1 1.2.840.10008.1.2.1
Endian
7 SECURTIY
It is assumed that the system is used within a secured environment. It is assumed that a secured environment includes at
a minimum:
a. Firewall or router protections to ensure that only approved external hosts have network access to the system.
b. Firewall or router protections to ensure that the system only has network access to approved external hosts and
services.
c. Any communication with external hosts and services outside the locally secured environment use appropriate secure
network channels (e.g. such as a Virtual Private Network (VPN))
Other network security procedures such as automated intrusion detection may be appropriate in some environments.
Additional security features may be established by the local security policy and are beyond the scope of this conformance
statement.
8 ANNEXES
Table 8.1-17 specifies the attributes of a Comprehensive Structured Report transmitted by the system storage application.
The following tables use a number of abbreviations. The abbreviations used in the “Presence of …” column are:
VNAP Value Not Always Present (attribute sent zero length if no value is present)
ANAP Attribute Not Always Present
ALWAYS Always Present
EMPTY Attribute is sent without a value
NOTE: All dates and times are encoded in the local configured calendar and time. Date, Time and Time zone are configured using
the Service/Installation Tool.
Table 8.1-2
PATIENT MODULE
Presence of
Attribute Name Tag VR Value Source
Value
From Modality Worklist or user input.
Values supplied via Modality Worklist
will be entered as received. Values MWL/
Patient’s Name (0010,0010) PN VNAP
supplied via user input will contain all 5 USER
components (some possibly empty). .
Maximum 64 characters.
From Modality Worklist or user input. MWL/
Patient ID (0010,0020) LO VNAP
Maximum 64 characters. USER
MWL/
Patient’s Birth Date (0010,0030) DA From Modality Worklist or user input VNAP
USER
MWL/
Patient’s Sex (0010,0040) CS From Modality Worklist or user input VNAP
USER
Referenced Patient (0008,1120) Not Used
SQ
Sequence
>Referenced SOP Class (0008,1150) Not Used
UI
UID
>Referenced SOP (0008,1155) Not Used
UI
Instance UID
Patient's Birth Time (0010,0032) TM Not Used
Other Patient IDs (0010,1000) LO Not Used
Other Patient Names (0010,1001) PN Not Used
Ethnic Group (0010,2160) SH Not Used
From User Input. As entered in the
Comments field on the New Study
Patient Comments (0010,4000) LT VNAP USER
screen but not sent if zero length.
Maximum 1024 characters.
Table 8.1-3
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 56 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT
Table 8.1-4
PATIENT STUDY MODULE
Presence
Attribute Name Tag VR Value Source
of Value
Admitting
Diagnosis (0008,1080) LO From Modality Worklist VNAP MWL
Description
Calculated from DOB input on
Patient’s Age (0010,1010) AS ALWAYS AUTO
base of actual Date
From Modality Worklist or user MWL/
Patient’s Weight (0010,1030) DS VNAP
input USER
Additional Patient’s From Modality Worklist or user
(0010,21B0) LT VNAP MWL
History input
Table 8.1-5
GENERAL SERIES MODULE
Presence of
Attribute Name Tag VR Value Source
Value
Modality (0008,0060) CS US ALWAYS AUTO
Series Instance UID (0020,000E) UI Generated by device ALWAYS AUTO
Series Number (0020,0011) IS Generated by device ALWAYS AUTO
Series Date (0008,0021) DA <yyyymmdd> ALWAYS AUTO
Series Time (0008,0031) TM <hhmmss> ALWAYS AUTO
Performing Physician’s Physician field in Study list. Maximum 64
(0008,1050) PN VNAP USER
Name characters.
Protocol Name (0018,1030) LO Exam Type ALWAYS AUTO
Organ from Study list. Maximum 512
Series Description (0008,103E) LO VNAP USER
characters.
Operator field in Study list. Maximum 64
Operator’s Name (0008,1070) PN VNAP USER
characters.
Referenced Study (0008,1111) SQ Not Used
Component Sequence
>Referenced SOP Class (0008,1150) UI Not Used
UID
>Referenced SOP Instance (0008,1155) UI Not Used
UID
Body Part Examined (0018,0015) CS Not Used
Patient Position (0018,5100) CS Not Used
Smallest Pixel Value in (0028,0108) US Not Used
Series or
SS
Largest Pixel Value in (0028,0109) US Not Used
Series or
SS
Referenced Performed Identifies the MPPS SOP Instance to
(0008,1111) SQ ALWAYS MPPS
Procedure Step Sequence which this image is related
>Referenced SOP Class
(0008,1150) UI MPPS SOP Class UID ALWAYS MPPS
UID
>Referenced SOP Instance
(0008,1155) UI MPPS SOP Instance UID ALWAYS MPPS
UID
Request Attributes
(0040,0275) SQ Zero or 1 item will be present ALWAYS AUTO
Sequence
>Requested Procedure ID (0040,1001) SH From Modality Worklist VNAP MWL
>Scheduled Procedure
(0040,0009) SH From Modality Worklist VNAP MWL
Step ID
>Scheduled Procedure
(0040,0007) LO From Modality Worklist VNAP MWL
Step Description
>Scheduled Protocol Code (0040,0008) SQ From Modality Worklist VNAP MWL
Sequence
Performed Procedure Step
(0040,0253) SH Same as MPPS. ALWAYS MPPS
ID
Performed Procedure Step
(0040,0244) DA Same as MPPS ALWAYS MPPS
Start Date
Performed Procedure Step
(0040,0245) TM Same as MPPS ALWAYS MPPS
Start Time
Performed Procedure Step Same as MPPS. From user input.
(0040,0254) LO VNAP MPPS
Description Maximum 64 characters.
Performed Protocol Code
(0040,0260) SQ Same as MPPS ALWAYS MPPS
Sequence
Comments on the Same as MPPS. From user input.
(0040,0280) LO VNAP MPPS
Performed Procedure Step Maximum 64 characters.
Table 8.1-6
GENERAL EQUIPMENT MODULE
Presence
Attribute Name Tag VR Value Source
of Value
Manufacturer (0008,0070) LO SONOSCAPE ALWAYS AUTO
Concatenated Hospital Name
and Department from system
Institution Name (0008,0080) LO VNAP CONFIG
configuration (see section).
Always sent even if zero length.
Station Name (0008,1010) SH From Configuration ALWAYS CONFIG
Manufacturer’s S40, S40Exp, S30, S12, S22,
(0008,1090) LO ALWAYS AUTO
Model Name S9, S9Pro, S8Exp
Device Serial
(0018,1000) LO From Configuration ALWAYS CONFIG
Number
Software Version (0018,1020) LO From Configuration ALWAYS CONFIG
Spatial (0018,1050) Not Used
DS
Resolution
Date of Last (0018,1200) Not Used
DA
Calibration
Time of Last (0018,1201) Not Used
TM
Calibration
Pixel Padding (0028,0120) US Not Used
Value or
SS
Table 8.1-7
GENERAL IMAGE MODULE
Attribute Presence of
Tag VR Value Source
Name Value
Instance (0020,0013) IS Generated by device ALWAYS AUTO
Number
Patient
(0020,0020) CS Zero length EMPTY AUTO
Orientation
<yyyymmdd>
Content Date (0008,0023) DA ALWAYS AUTO
Date when image was acquired
<hhmmss>
Content Time (0008,0033) TM ALWAYS AUTO
Time when image was acquired
Image Type (0008,0008) CS ORIGINAL/PRIMARY/<Analysis Type*> for
uncompressed,
DERIVED/PRIMARY/ < Analysis Type *> if
compressed
Acquisition
(0020,0012) IS Not Used ALWAYS AUTO
Number
Acquisition (0008,0022) Not Used
DA
Date
Acquisition (0008,0032) Not Used
TM
Time
Image Not Used
(0020,4000) LT VNAP USER
Comments
Referenced (0008,1140) SQ Not Used
Image
Sequence
>Referenced (0008,1150) UI Not Used
SOP Class
UID
>Referenced (0008,1155) UI Not Used
SOP Instance
UID
>Referenced (0008,1160) IS Not Used
Frame Number
Derivation (0008,2111) ST Not Used
Description
Source Image (0008,2112) SQ Not Used
Sequence
>Referenced (0008,1150) UI Not Used
SOP Class
UID
>Referenced (0008,1155) UI Not Used
SOP Instance
UID
>Referenced (0008,1160) IS Not Used
Frame Number
Images in (0020,1002) IS Not Used
Acquisition
Image (0020,4000) LT Not Used
Comments
Table 8.1-8
IMAGE PIXEL MODULE
Presence of
Attribute Name Tag VR Value Source
Value
Samples per Pixel (0028,0002) 1 A value of 1 for “MONOCHROME2” or a
ALWAYS AUTO
value of 3 for “RGB”
Photometric (0028,0004) 1 A value of “MONOCHROME2” or “RGB”
ALWAYS AUTO
Interpretation
Rows (0028,0010) 1 Image height in pixels:640,1024,1180 ALWAYS
Columns (0028,0011) 1 Image width in pixels:480,768,768 ALWAYS
Bits Allocated (0028,0100) 1 Always 0008H ALWAYS AUTO
Bits Stored (0028,0101) 1 Always 0008H ALWAYS AUTO
High Bit (0028,0102) 1 Always 0007H ALWAYS AUTO
Pixel (0028,0103) 1 Always 0000H
ALWAYS AUTO
Representation
Pixel Data (7FE0,0010) OW /
ALWAYS AUTO
OB
Planar (0028,0006) 1C Always 0000H
Configuration
Pixel Aspect Ratio (0028,0034) 1C Not Used
Smallest Image (0028,0106) 3 Not Used
Pixel Value
Largest Image Pixel (0028,0107) 3 Not Used
Value
Red Palette Color (0028,1101) 1C Not Used
Lookup Table
Descriptor
Green Palette Color (0028,1102) 1C Not Used
Lookup Table
Descriptor
Blue Palette Color (0028,1103) 1C Not Used
Lookup Table
Descriptor
Red Palette Color (0028,1201) 1C Not Used
Lookup Table Data
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 61 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT
Table 8.1-9
CINE MODULE
Attribute Presence
Tag VR Value Source
Name of Value
Frame Time (0018,1063) DS Only if multi-frame. ANAP AUTO
Recommende
d Display (0008,2144) IS Only if multi-frame Same as Cine Rate ANAP AUTO
Frame Rate
Cine Rate (0018,0040) IS Only if multi-frame ANAP AUTO
Table 8.1-10
MULTI-FRAME MODULE
Attribute Presence
Tag VR Value Source
Name of Value
Number of
(0028,0008) IS Only if multi-frame ANAP AUTO
Frames
Table 8.1-11
CONTRAST/BOLUS MODULE
Presence of
Attribute Name Tag VR Value Source
Value
Contrast/Bolus
(0018,0010) US Only if multi-frame ANAP AUTO
Agent
Contrast/Bolus
(0018,0012) SQ Not used
Agent Sequence
Contrast/Bolus
(0018,1040) Not used
Route
Contrast/Bolus
Administration (0018,0014) SQ Not used
Route Sequence
Contrast/Bolus
(0018,1041) Not used
Volume
Contrast/Bolus
(0018,1042) Not used
Start Time
Contrast/Bolus
(0018,1043) Not used
Stop Time
Contrast/Bolus
(0018,1044) Not used
Total Dose
Contrast Flow
(0018,1046) Not used
Rate(s)
Contrast Flow
(0018,1047) Not used
Duration(s)
Contrast/Bolus
(0018,1048) Not used
Ingredient
Contrast/Bolus
Ingredient (0018,1049) Not used
Concentration
Table 8.1-12
PALETTE COLOR LOOKUP TABLE MODULE
Presence of
Attribute Name Tag VR Value Source
Value
Red Palette Color Only used when reading
(0028,1101) US or SS
Lookup Table Descriptor Palette images.
Only used when reading
Green Palette Color
(0028,1102) US or SS Palette images.
Lookup Table Descriptor
Only used when reading
Blue Palette Color
(0028,1102) US or SS Palette images.
Lookup Table Descriptor
Palette Color Lookup
(0028,1199) UI Not used
Table UID
Only used when reading
Red Palette Color
(0028,1201) OW Palette images.
Lookup Table Data
Only used when reading
Green Palette Color
(0028,1202) OW Palette images.
Lookup Table Data
Only used when reading
Blue Palette Color
(0028,1203) OW Palette images.
Lookup Table Data
Segmented Red Palette
Not used
Color Lookup Table Data (0028,1221) OW
Segmented Green
Palette Color Lookup Not used
(0028,1222) OW
Table Data
Table 8.1-13
US REGION CALIBRATION MODULE
Presence of
Attribute Name Tag VR Value Source
Value
Sequence of Ultrasound (0018,6011) A sequence is present for
Regions SQ each region on the system
display, except for ECG
Table 8.1-14
US IMAGE MODULE
Presence of
Attribute Name Tag VR Value Source
Value
Samples Per Pixel (0028,0002) A sequence is present for each
region on the system display, except
SQ for ECG regions. Only when set for
“Display Area”. No scaling for “Full
Screen” images
2D Image = 0003H = CM
US
M mode = 0003H = CM
Number of Views in (0008,212A) A bit mask 0-5 if: none, 2d, M mode,
Stage IS spectral Doppler. See DICOM PS3.3
C.8.5.5.1.1
R Wave Time Vector (0018,6060) US Enumerated Value. See DICOM
PS3.3 C.8.5.5.1.2
Ultrasound Color Data (0028,0014) US Bit mask. See DICOM PS3.3
Present C.8.5.5.1.3
Stage Name (0008,2120) SH
>Include “Code
Sequence Macro”
Table 8.8-1.
Stage Number (0008,2122)
Include 'General
Anatomy Optional
Macro' Table 10-7
View Code Sequence (0054,0220)
>Include ‘Code
Sequence Macro’
Table 8.8-1
>View Modifier Code (0054,0222)
Sequence
>>Include ‘Code
Sequence Macro’
Table 8.8-1
Acquisition DateTime (0008,002A)
Table 8.1-15
MODALITY LUT MODULE
Attribute Presence of
Tag VR Value Source
Name Value
Modality LUT
(0028,3000) SQ present if (0028,1040) = LOG ANAP AUTO
Sequence
> LUT
(0028,3002) US <1024,0,16> ANAP AUTO
Descriptor
> Modality
(0028,3004) LO US ANAP AUTO
LUT Type
> LUT Data (0028,3006) US LUT ANAP AUTO
Window
(0028,1050) DS 0…1023 ALWAYS AUTO
Center
Window Width (0028,1051) DS 1…1024 ALWAYS AUTO
Table 8.1-16
SOP COMMON MODULE
Attribute Presence of
Tag VR Value Source
Name Value
Specific
(0008,0005) CS “ISO_IR 100” ALWAYS CONFIG
Character Set
SOP Class 1.2.840.10008.5.1.4.1.1.6.1
(0008,0016) UI ALWAYS AUTO
UID 1.2.840.10008.5.1.4.1.1.6
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 67 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT
1.2.840.10008.5.1.4.1.1.7
SOP Instance
(0008,0018) UI Generated by device ALWAYS AUTO
UID
Instance (0008,0012) <yyyymmdd>
Creation Date DA ALWAYS AUTO
Date when image was acquired
Instance (0008,0013) <hhmmss>
Creation TM ALWAYS AUTO
Time Time when image was acquired
Table 8.1-18
PATIENT MODULE
Presence of
Attribute Name Tag VR Value Source
Value
From Modality Worklist or user input.
Values supplied via Modality Worklist
will be entered as received. Values MWL/
Patient’s Name (0010,0010) PN VNAP
supplied via user input will contain all 5 USER
components (some possibly empty). .
Maximum 64 characters.
From Modality Worklist or user input. MWL/
Patient ID (0010,0020) LO VNAP
Maximum 64 characters. USER
MWL/
Patient’s Birth Date (0010,0030) DA From Modality Worklist or user input VNAP
USER
MWL/
Patient’s Sex (0010,0040) CS From Modality Worklist or user input VNAP
USER
Referenced Patient (0008,1120) SQ Not Used
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 68 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT
Sequence
>Referenced SOP Class (0008,1150) Not Used
UI
UID
>Referenced SOP (0008,1155) Not Used
UI
Instance UID
Patient's Birth Time (0010,0032) TM Not Used
Other Patient IDs (0010,1000) LO Not Used
Other Patient Names (0010,1001) PN Not Used
Ethnic Group (0010,2160) SH Not Used
From User Input. As entered in the
Comments field on the New Study
Patient Comments (0010,4000) LT VNAP USER
screen but not sent if zero length.
Maximum 1024 characters.
Table 8.1-19
GENERAL STUDY MODULE
Presence of
Attribute Name Tag VR Value Source
Value
From Modality Worklist or generated by MWL/
Study Instance UID (0020,000D) UI ALWAYS
device AUTO
Study Date (0008,0020) DA <yyyymmdd> ALWAYS AUTO
Study Time (0008,0030) TM <hhmmss> ALWAYS AUTO
Referring Physician’s
(0008,0090) PN From Modality Worklist VNAP MWL
Name
Requested Procedure ID from Worklist MWL/
Study ID (0020,0010) SH VNAP
or User Input USER
MWL/
Accession Number (0008,0050) SH From Modality Worklist or user input VNAP
USER
Comment text box in study list.
Study Description (0008,1030) LO VNAP USER
Maximum 1024 characters.
Physician(s) of Record (0008,1048) Not Used
Referenced Study
(0008,1110) SQ From Modality Worklist VNAP MWL
Sequence
>Referenced SOP Class
(0008,1150) UI From Modality Worklist VNAP MWL
UID
>Referenced SOP
(0008,1155) UI From Modality Worklist VNAP MWL
Instance UID
Procedure Code (0008,1032) MWL Requested Procedure Code
Sequence SQ Sequence (0032,1064) Absent if ANAP MWL
unscheduled.
>Code Value (0008,0100) SH Same value as MWL attribute VNAP MWL
>Coding Scheme
(0008,0102) SH Same value as MWL attribute VNAP MWL
Designator
>Coding Scheme Version (0008,0103) SH Same value as MWL attribute VNAP MWL
Table 8.1-20
PATIENT STUDY MODULE
Presence
Attribute Name Tag VR Value Source
of Value
Admitting
Diagnosis (0008,1080) LO From Modality Worklist VNAP MWL
Description
Calculated from DOB input on
Patient’s Age (0010,1010) AS ALWAYS AUTO
base of actual Date
From Modality Worklist or user MWL/
Patient’s Weight (0010,1030) DS VNAP
input USER
Additional Patient’s From Modality Worklist or user
(0010,21B0) LT VNAP MWL
History input
Table 8.1-21
GENERAL SERIES MODULE
Presence of
Attribute Name Tag VR Value Source
Value
Modality (0008,0060) CS US ALWAYS AUTO
Series Instance UID (0020,000E) UI Generated by device ALWAYS AUTO
Series Number (0020,0011) IS Generated by device ALWAYS AUTO
Series Date (0008,0021) DA <yyyymmdd> ALWAYS AUTO
Series Time (0008,0031) TM <hhmmss> ALWAYS AUTO
Performing Physician’s Physician field in Study list. Maximum 64
(0008,1050) PN VNAP USER
Name characters.
Protocol Name (0018,1030) LO Exam Type ALWAYS AUTO
Organ from Study list. Maximum 512
Series Description (0008,103E) LO VNAP USER
characters.
Operator field in Study list. Maximum 64
Operator’s Name (0008,1070) PN VNAP USER
characters.
Referenced Study (0008,1111) SQ Not Used
Component Sequence
>Referenced SOP Class (0008,1150) UI Not Used
UID
>Referenced SOP Instance (0008,1155) UI Not Used
UID
Body Part Examined (0018,0015) CS Not Used
Patient Position (0018,5100) CS Not Used
Smallest Pixel Value in (0028,0108) US Not Used
Series or
SS
Table 8.1-22
GENERAL EQUIPMENT MODULE
Presence
Attribute Name Tag VR Value Source
of Value
Manufacturer (0008,0070) LO SONOSCAPE ALWAYS AUTO
Concatenated Hospital Name
and Department from system
Institution Name (0008,0080) LO VNAP CONFIG
configuration (see section).
Always sent even if zero length.
Station Name (0008,1010) SH From Configuration ALWAYS CONFIG
Manufacturer’s
(0008,1090) LO pj0905 ALWAYS AUTO
Model Name
Device Serial
(0018,1000) LO From Configuration ALWAYS CONFIG
Number
Table 8.1-23
SR DOCUMENT GENERAL MODULE
Presence
Attribute Name Tag VR Value Source
of Value
Instance (0020,0013) IS Unique number ALWAYS AUTO
Number
Completion Flag (0040,A491) CS PARTIAL VNAP AUTO
Verification Flag (0040,A493) CS UNVERIFIED ALWAYS AUTO
Content Date (0008,0023) DA Date content created. ALWAYS AUTO
Content Time (0008,0033) TM Time content created. ALWAYS AUTO
Used when Send of
Predecessor Demand is used and Send
(0040,A360) SQ
Documents Structured Reports with
Sequence Send on Demand is ALWAYS AUTO
selected. Will indicate the
preceding SR sent in the
study.
>Study Instance (0020,000D) UI Study’s UID
ALWAYS AUTO
UID
>Referenced
(0008,1115) SQ Identifies the Series containing ALWAYS AUTO
Series
the referenced SR
Sequence
>>Reference
(0008,1199) SQ SOP Instance UID for SR ALWAYS AUTO
d SOP
Series in the study
Sequence
>>> Comprehensive SR SOP Class
(0008,1150) UI ALWAYS AUTO
Referenced 1.2.840.10008.5.1.4.1.1.88.33
SOP Class
>>>
(0008,1155) UI SOP Instance UID of the ALWAYS AUTO
Referenced
preceding SR in the study
SOP Instance
UID
Identifies Requested
Referenced ALWAYS AUTO
(0040,A370) SQ Procedures being
Request
fulfilled (completely or partially)
Sequence
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 72 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT
by creation of this
Document.
>Study Instance (0020,000D) UI Same value as in MWL or auto ALWAYS AUTO
UID generated
>Referenced
(0008,1110) SQ 1 item per item in MWL, zero ALWAYS AUTO
Study
length if unscheduled
Sequence
>>Reference
(0008,1150) UI Identifies the Referenced SOP ALWAYS AUTO
d SOP Class
Class
UID
>>Reference
(0008,1155) UI Instance UID ALWAYS AUTO
d SOP
Instance UID
Table 8.1-24
SR DOCUMENT CONTENT MODULE
Presence
Attribute Name Tag VR Value Source
of Value
Content (0040,A504) SQ
Template ALWAYS AUTO
Sequence
The Root Content Item
>Template (0040,DB00) CS identifies TID 5000 (OB- ALWAYS AUTO
Identifier GYN), 5100 (Vascular).
5200 (Echo) or SONO1000
(Other).
>Mapping (0008,0105) CS DCMR ALWAYS AUTO
Resource
Content (0040,A730) SQ ALWAYS AUTO
Sequence
See Template ID
5000 for
ALWAYS AUTO
>Relationship (0040,A010) CS OB-GYN,
Type Template ID
5100 for
Vascular
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 73 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT
Template ID
5200 for Adult
Echo and
Template ID
SONO1000 for
Other
See Template ID ANAP AUTO
Document 5000 for
Relationship OB-GYN,
Macro Table Template ID
5100 for
Vascular
Template ID
5200 for Adult
Echo and
Template ID
SONO1000 for
Other
See Template ID
Document 5000 for
ALWAYS AUTO
Content OB-GYN,
Macro Template ID
5100 for
Vascular
Template ID
5200 for Adult
Echo and
Template ID
SONO1000 for
Other
Value Type (0040,A040) CS CONTAINER, ALWAYS AUTO
always first tag
of SR
Additional
values used:
TEXT, NUM, DATE, CODE,
IMAGE, SCOORD
Concept Name
(0040,A043) SQ ALWAYS AUTO
Code
Sequence
>Code Value (0008,0100) 125000, 125100 125200 or ALWAYS AUTO
SONO1000
>Coding
(0008,0102) DCM ALWAYS AUTO
Scheme
Designator
“OB-GYN Ultrasound
Procedure Report”,
>Code Meaning (0008,0104) ALWAYS AUTO
“Vascular Ultrasound
Procedure Report”,
“Adult
Echocardiography
Procedure Report”
Or Other Procedure Report
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 74 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT
Table 8.1-25
SOP COMMON MODULE
Presence
Attribute Name Tag VR Value Source
of Value
SOP Class UID (0008,0016) UI 1.2.840.10008.5.1.4.1.1.88.33 ALWAYS AUTO
SOP Instance (0008,0018) UI Generated by device
ALWAYS AUTO
UID
Specific
(0008,0005) CS ISO_IR 100
Character ALWAYS CONFIG
Set
Table 8.1-31
ATTRIBUTE MAPPING BETWEEN MODALITY WORKLIST, IMAGE AND MPPS
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 75 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT
The contents of Performed Procedure Step Discontinuation Reason Code Sequence (0040,0281) for a discontinued MPPS
will be filled with a code selected by the user from a fixed list corresponding to Context Group 9300.
This Appendix contains tables of the measurement and calculations displayed on the system calculation screens and the
data dictionary of all code values, coding scheme designators and code meanings associated with those measurements
and labels. There is no attempt to provide all tags that will be sent in the SR object. For those specifics, please refer to
DICOM PS3-3, 3-16 and 3-17 for complete discussion on Comprehensive SOP Class support and the Templates
described here.
Part 17 of the DICOM Standard includes tree diagrams showing graphic examples of the structure of each template.
Briefly, an SR document will contain only the measurements, calculations and observations made during the exam. Its
exact structure is therefore determined by two main components, the measurements that are available within the context
of the template and its referenced Templates and Context groups, and those measurements and calculations that are
implemented on the system creating the report. Supplementing these constraints are private and user defined
measurements and calculations, which may be added if the Root Container Template is extensible.
In the system, the calculation packages contain a combination of template and user defined measurements and
calculations. This Appendix contains a description of the calculation packages on each system and the related templates
that support them. See the table in the Clinical Scope section below.
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 77 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT
In each template section, there is a brief description of the mapping of measurement and calculation labels as they
appear on the system when the Analysis feature is enabled.
(LN, 11746-5)
SONO-5005-42)
GYN/GYN_FOLLICLE1/Lt-Fol1-D1
GYN/GYN_FOLLICLE1/Lt-Fol1-D2
GYN/GYN_FOLLICLE1/Lt-Fol1-D3
GYN/GYN_FOLLICLE1/Rt-Fol1-D1
GYN/GYN_FOLLICLE1/Rt-Fol1-D2
GYN/GYN_FOLLICLE1/Rt-Fol1-D3
…
Follicles All Follicle Diameter(LN, 11793-7)
GYN/GYN_FOLLICLE20/Lt-Fol20-D1
GYN/GYN_FOLLICLE20/Lt-Fol20-D2
GYN/GYN_FOLLICLE20/Lt-Fol20-D3
GYN/GYN_FOLLICLE20/Rt-Fol20-D1
GYN/GYN_FOLLICLE20/Rt-Fol20-D2
GYN/GYN_FOLLICLE20/Rt-Fol20-D3
GYN/GYN_FOLLICLE1/Lt-Fol1-Avg
GYN/GYN_FOLLICLE1/Rt-Fol1-Avg
GYN/GYN_FOLLICLE20/Rt-Fol20-Avg
GYN/GYN_FOLLICLE20/Lt-Fol20-Avg
GYN/GYN_FOLLICLE1/Lt-Fol1- Vol
GYN/GYN_FOLLICLE1/Rt-Fol1- Vol
GYN/GYN_FOLLICLE20/Rt-Fol20- Vol
GYN/GYN_FOLLICLE20/Lt-Fol20- Vol
GYN/GYN_FIBROID1/Fib1-D1
GYN/GYN_FIBROID1/Fib1-D2
GYN/GYN_FIBROID1/Fib1-D3
GYN/GYN_FIBROID20/Fib20-D1
GYN/GYN_FIBROID20/Fib20-D2
GYN/GYN_FIBROID20/Fib20-D3
GYN/GYN_FIBROID1/Fib1-Avg
GYN/GYN_FIBROID20/Fib2-Avg
GYN/GYN_FIBROID1/Fib1-Vol
GYN/GYN_FIBROID20/Fib20-Vol
Pelvis and
Finding Site: Pelvis and Uterus (DCM, 125011)
Uterus
Endometrium
GYN/GYN_UT_THICKN/Ut-Endom.Th Endometrium Thickness (LN, 12145-9)
Thickness
OB/DUCTUSART/D_DA_PS/DA-PS
OB/DUCTUSART/D_DA_RI/DA-PS
Ductus Art PS Peak Systolic Velocity(LN, 11726-7)
OB/DUCTUSART/D_DA_PI/DA-PS
OB/DUCTUSART/D_DA_PERS/DA-PS
OB/DUCTUSART/D_DA_AT/DA-PS
OB/DUCTUSART/D_DA_MT/DA-PS
OB/AO/D_AO_PS/Ao-PS/
OB/AO/D_AO_RI/Ao-PS
OB/AO/D_AO_PI/Ao-PS
Ao PS
OB/AO/D_AO_PERS/Ao-PS
OB/AO/D_AO_AT/Ao-PS
OB/AO/D_AO_MT/Ao-PS
OB/MCA/D_MCA_RI/Lt-MCA-PS
OB/MCA/D_MCA_RI/Rt-MCA-PS
OB/MCA/D_MCA_PI/Lt-MCA-PS
OB/MCA/D_MCA_PI/Rt-MCA-PS
OB/MCA/D_MCA_PERS/Lt-MCA-PS
Lt MCA(Rt MCA) PS
OB/MCA/D_MCA_PERS/Rt-MCA-PS
OB/MCA/D_MCA_AT/Lt-MCA-PS
OB/MCA/D_MCA_AT/Rt-MCA-PS
OB/MCA/D_MCA_MT/Lt-MCA-PS
OB/MCA/D_MCA_MT/Rt-MCA-PS
OB/SMA/D_SMA_RI/SMA-PS
OB/SMA/D_SMA_PI/SMA-PS
SMA PS OB/SMA/D_SMA_PERS/SMA-PS
OB/SMA/D_SMA_AT/SMA-PS
OB/SMA/D_SMA_MT/SMA-PS
OB/CELIACA/D_CELA_RI/Celiac A.-PS
OB/CELIACA/D_CELA_PI/Celiac A.-PS
OB/CELIACA/D_CELA_AT/Celiac A.-PS
OB/CELIACA/D_CELA_MT/Celiac A.-PS
OB/CAROTID/D_CAR_RI/Lt-Car-PS
OB/CAROTID/D_CAR_RI/Rt-Car-PS
OB/CAROTID/D_CAR_PI/Lt-Car-PS
OB/CAROTID/D_CAR_PI/Rt-Car-PS
OB/CAROTID/D_CAR_PERS/Lt-Car-PS
Carotid PS
OB/CAROTID/D_CAR_PERS/Rt-Car-PS
OB/CAROTID/D_CAR_AT/Lt-Car-PS
OB/CAROTID/D_CAR_AT/Rt-Car-PS
OB/CAROTID/D_CAR_MT/Lt-Car-PS
OB/CAROTID/D_CAR_MT/Rt-Car-PS
OB/DUCTUSART/D_DA_ED/DA-ED
OB/DUCTUSART/D_DA_RI/DA-ED
OB/DUCTUSART/D_DA_PI/DA-ED
Ductus Art ED
OB/DUCTUSART/D_DA_PERS/DA-ED
OB/DUCTUSART/D_DA_AT/DA-ED
OB/DUCTUSART/D_DA_MT/DA-ED
OB/AO/D_AO_ED/Ao-ED
OB/AO/D_AO_RI/Ao-ED
OB/AO/D_AO_PI/Ao-ED
Ao ED End Diastolic Velocity(LN, 11653-3)
OB/AO/D_AO_PERS/Ao-ED
OB/AO/D_AO_AT/Ao-ED
OB/AO/D_AO_MT/Ao-ED
OB/MCA/D_MCA_ED/Lt-MCA-ED
OB/MCA/D_MCA_ED/Rt-MCA-ED
OB/MCA/D_MCA_RI/Lt-MCA-ED
MCA ED
OB/MCA/D_MCA_RI/Rt-MCA-ED
OB/MCA/D_MCA_PI/Lt-MCA-ED
OB/MCA/D_MCA_PI/Rt-MCA-ED
OB/MCA/D_MCA_PERS/Lt-MCA-ED
OB/MCA/D_MCA_PERS/Rt-MCA-ED
OB/MCA/D_MCA_AT/Lt-MCA-ED
OB/MCA/D_MCA_AT/Rt-MCA-ED
OB/MCA/D_MCA_MT/Lt-MCA-ED
OB/MCA/D_MCA_MT/Rt-MCA-ED
OB/SMA/D_SMA_ED/SMA-ED
OB/SMA/D_SMA_RI/SMA-ED
OB/SMA/D_SMA_PI/SMA-ED
SMA ED
OB/SMA/D_SMA_PERS/SMA-ED
OB/SMA/D_SMA_AT/SMA-ED
OB/SMA/D_SMA_MT/SMA-ED
OB/CELIACA/D_CELA_ED/Celiac A.-ED
OB/CELIACA/D_CELA_RI/Celiac A.-ED
OB/CELIACA/D_CELA_PI/Celiac A.-ED
Celiac A. ED
OB/CELIACA/D_CELA_PERS/Celiac A.-ED
OB/CELIACA/D_CELA_AT/Celiac A.-ED
OB/CELIACA/D_CELA_MT/Celiac A.-ED
OB/CAROTID/D_CAR_ED/Lt-Car-ED
OB/CAROTID/D_CAR_ED/Rt-Car-ED
OB/CAROTID/D_CAR_RI/Lt-Car-ED
OB/CAROTID/D_CAR_RI/Rt-Car-ED
OB/CAROTID/D_CAR_PI/Lt-Car-ED
Carotid ED
OB/CAROTID/D_CAR_PI/Rt-Car-ED
OB/CAROTID/D_CAR_PERS/Lt-Car-ED
OB/CAROTID/D_CAR_PERS/Rt-Car-ED
OB/CAROTID/D_CAR_AT/Lt-Car-ED
OB/CAROTID/D_CAR_AT/Rt-Car-ED
OB/CAROTID/D_CAR_MT/Lt-Car-ED
OB/CAROTID/D_CAR_MT/Rt-Car-ED
OB/DUCTUSART/D_DA_RI/DA-RI
OB/DUCTUSART/D_DA_PERS/DA-RI
Ductus Art RI
OB/DUCTUSART/D_DA_AT/DA-RI
OB/DUCTUSART/D_DA_MT/DA-RI
OB/AO/D_AO_RI/Ao-RI
OB/AO/D_AO_PERS/Ao-RI
Ao RI
OB/AO/D_AO_AT/Ao-RI
OB/AO/D_AO_MT/Ao-RI
OB/MCA/D_MCA_RI/Lt-MCA-RI
OB/MCA/D_MCA_RI/Rt-MCA-RI
OB/MCA/D_MCA_PERS/Lt-MCA-RI
OB/MCA/D_MCA_PERS/Rt-MCA-RI
MCA RI
OB/MCA/D_MCA_AT/Lt-MCA-RI
Resistivity Index(LN, 12023-8)
OB/MCA/D_MCA_AT/Rt-MCA-RI
OB/MCA/D_MCA_MT/Lt-MCA-RI
OB/MCA/D_MCA_MT/Rt-MCA-RI
OB/SMA/D_SMA_RI/SMA-RI
OB/SMA/D_SMA_PERS/SMA-RI
SMA RI
OB/SMA/D_SMA_AT/SMA-RI
OB/SMA/D_SMA_MT/SMA-RI
OB/CELIACA/D_CELA_RI/Celiac A.-RI
OB/CELIACA/D_CELA_PERS/Celiac A.-RI
Celiac A. RI
OB/CELIACA/D_CELA_AT/Celiac A.-RI
OB/CELIACA/D_CELA_MT/Celiac A.-RI
OB/CAROTID/D_CAR_RI/Lt-Car-RI
Carotid RI
OB/CAROTID/D_CAR_RI/Rt-Car-RI
OB/CAROTID/D_CAR_PERS/Lt-Car-RI
OB/CAROTID/D_CAR_PERS/Rt-Car-RI
OB/CAROTID/D_CAR_AT/Lt-Car-RI
OB/CAROTID/D_CAR_AT/Rt-Car-RI
OB/CAROTID/D_CAR_MT/Lt-Car-RI
OB/CAROTID/D_CAR_MT/Rt-Car-RI
OB/DUCTUSART/D_DA_RI/DA-PI
OB/DUCTUSART/D_DA_PI/DA-PI
Ductus Art PI
OB/DUCTUSART/D_DA_AT/DA-PI
OB/DUCTUSART/D_DA_MT/DA-PI
OB/AO/D_AO_RI/Ao-PI
OB/AO/D_AO_PI/Ao-PI
Ao PI
OB/AO/D_AO_AT/Ao-PI
OB/AO/D_AO_MT/Ao-PI
OB/MCA/D_MCA_RI/Lt-MCA-PI
OB/MCA/D_MCA_RI/Rt-MCA-PI
OB/MCA/D_MCA_PI/Lt-MCA-PI
OB/MCA/D_MCA_AT/Rt-MCA-PI
OB/MCA/D_MCA_MT/Lt-MCA-PI
OB/MCA/D_MCA_MT/Rt-MCA-PI
OB/SMA/D_SMA_RI/SMA-PI
OB/SMA/D_SMA_PI/SMA-PI
SMA PI
OB/SMA/D_SMA_AT/SMA-PI
OB/SMA/D_SMA_MT/SMA-PI
OB/CELIACA/D_CELA_RI/Celiac A.-PI
OB/CELIACA/D_CELA_MT/Celiac A.-PI
OB/CAROTID/D_CAR_RI/Lt-Car-PI
OB/CAROTID/D_CAR_RI/Rt-Car-PI
OB/CAROTID/D_CAR_PI/Lt-Car-PI
OB/CAROTID/D_CAR_PI/Rt-Car-PI
Carotid PI
OB/CAROTID/D_CAR_AT/Lt-Car-PI
OB/CAROTID/D_CAR_AT/Rt-Car-PI
OB/CAROTID/D_CAR_MT/Lt-Car-PI
OB/CAROTID/D_CAR_MT/Rt-Car-PI
OB/DUCTUSART/D_DA_PERS/DA-S/D
OB/DUCTUSART/D_DA_MT/DA-S/D
OB/AO/D_AO_PERS/Ao-S/D
Ao-S/D OB/AO/D_AO_AT/Ao-S/D
OB/AO/D_AO_MT/Ao-S/D
OB/MCA/D_MCA_PERS/Lt-MCA-S/D
OB/MCA/D_MCA_PERS/Rt-MCA-S/D
OB/MCA/D_MCA_AT/Lt-MCA-S/D
MCA-S/D
OB/MCA/D_MCA_AT/Rt-MCA-S/D
Systolic to Diastolic Velocity Ratio(LN,
12144-2)
OB/MCA/D_MCA_MT/Lt-MCA-S/D
OB/MCA/D_MCA_MT/Rt-MCA-S/D
OB/SMA/D_SMA_PERS/SMA-S/D
SMA-S/D OB/SMA/D_SMA_AT/SMA-S/D
OB/SMA/D_SMA_MT/SMA-S/D
OB/CELIACA/D_CELA_PERS/Celiac A.-S/D
OB/CELIACA/D_CELA_MT/Celiac A.-S/D
OB/CAROTID/D_CAR_PERS/Lt-Car-S/D
Carotid-S/D
OB/CAROTID/D_CAR_PERS/Rt-Car-S/D
OB/CAROTID/D_CAR_AT/Lt-Car-S/D
OB/CAROTID/D_CAR_AT/Rt-Car-S/D
OB/CAROTID/D_CAR_MT/Lt-Car-S/D
OB/CAROTID/D_CAR_MT/Rt-Car-S/D
OB/DUCTUSART/D_DA_AT/DA-TAmean
Ductus Art-TAmean
OB/DUCTUSART/D_DA_MT/DA-TAmean
OB/AO/D_AO_AT/Ao-TAmean
Ao-TAmean
OB/AO/D_AO_MT/Ao-TAmean
OB/MCA/D_MCA_PI/Lt-MCA-TAmean
OB/MCA/D_MCA_PI/Rt-MCA-TAmean
OB/MCA/D_MCA_AT/Lt-MCA-TAmean
OB/MCA/D_MCA_AT/Rt-MCA-TAmean
MCA-TAmean
OB/MCA/D_MCA_MT/Lt-MCA-TAmean
OB/MCA/D_MCA_MT/Rt-MCA-TAmean
OB/MCA/D_MCA_VOL/Lt-MCA-TAmean
Time averaged mean velocity(LN,
OB/MCA/D_MCA_VOL/Rt-MCA-TAmean 20352-1)
OB/SMA/D_SMA_AT/SMA-TAmean
SMA-TAmean
OB/SMA/D_SMA_MT/SMA-TAmean
OB/CELIACA/D_CELA_AT/Celiac A.-TAmean
Celiac A.-TAmean
OB/CELIACA/D_CELA_MT/Celiac A.-TAmean
OB/CAROTID/D_CAR_AT/Lt-Car-TAmean
OB/CAROTID/D_CAR_AT/Rt-Car-TAmean
OB/CAROTID/D_CAR_MT/Lt-Car-TAmean
Carotid-TAmean
OB/CAROTID/D_CAR_MT/Rt-Car-TAmean
OB/CAROTID/D_CAR_VOL/Lt-Car-TAmean
OB/CAROTID/D_CAR_VOL/Rt-Car-TAmean
OB/D_IVC_S/IVC-S
Peak Velocity During Ventricular
IVC OB/D_IVC_SAPLI/IVC-S
Systole(SONO, SONO-5025-11)
OB/D_IVC_PVIV/IVC-S
OB/AO/D_AO_AT/Ao-TAmax
Ao-TAmax
OB/AO/D_AO_MT/Ao-TAmax
OB/MCA/D_MCA_PI/Lt-MCA-TAmax
OB/MCA/D_MCA_PI/Rt-MCA-TAmax
OB/MCA/D_MCA_AT/Lt-MCA-TAmax
MCA-TAmax
OB/MCA/D_MCA_AT/Rt-MCA-TAmax
OB/MCA/D_MCA_MT/Lt-MCA-TAmax
OB/MCA/D_MCA_MT/Rt-MCA-TAmax
OB/SMA/D_SMA_AT/SMA-TAmax
Time averaged peak velocity(LN,
SMA-TAmax
11692-1)
OB/SMA/D_SMA_MT/SMA-TAmax
OB/CELIACA/D_CELA_PI/Celiac A.-TAmax
OB/CELIACA/D_CELA_MT/Celiac A.-TAmax
OB/CAROTID/D_CAR_PI/Lt-Car-TAmax
OB/CAROTID/D_CAR_PI/Rt-Car-TAmax
OB/CAROTID/D_CAR_AT/Lt-Car-TAmax
Carotid-TAmax
OB/CAROTID/D_CAR_AT/Rt-Car-TAmax
OB/CAROTID/D_CAR_MT/Lt-Car-TAmax
OB/CAROTID/D_CAR_MT/Rt-Car-TAmax
OB/DUCTUSART/D_DA_AT/DA-AT
Ductus Art-AT
OB/DUCTUSART/D_DA_MT/DA-AT
OB/AO/D_AO_AT/Ao-AT
Ao-AT
OB/AO/D_AO_MT/Ao-AT
OB/MCA/D_MCA_AT/Lt-MCA-AT
OB/MCA/D_MCA_AT/Rt-MCA-AT
MCA-AT
OB/MCA/D_MCA_MT/Lt-MCA-AT
OB/MCA/D_MCA_MT/Rt-MCA-AT
Acceleration Time(LN, 20168-1)
OB/SMA/D_SMA_AT/SMA-AT
SMA-AT
OB/SMA/D_SMA_MT/SMA-AT
OB/CELIACA/D_CELA_AT/Celiac A.-AT
Celiac A.-AT
OB/CELIACA/D_CELA_MT/Celiac A.-AT
OB/CAROTID/D_CAR_AT/Lt-Car-AT
OB/CAROTID/D_CAR_AT/Rt-Car-AT
Carotid-AT
OB/CAROTID/D_CAR_MT/Lt-Car-AT
OB/CAROTID/D_CAR_MT/Rt-Car-AT
OB/DUCTUSART/D_DA_AT/DA-VTI
Ductus Art -VTI
OB/DUCTUSART/D_DA_MT/DA-VTI
OB/AO/D_AO_AT/Ao-VTI
Ao-VTI
OB/AO/D_AO_MT/Ao-VTI
OB/MCA/D_MCA_AT/Lt-MCA-VTI
OB/MCA/D_MCA_AT/Rt-MCA-VTI
MCA-VTI Velocity Time Integral(LN, 20354-7)
OB/MCA/D_MCA_MT/Lt-MCA-VTI
OB/MCA/D_MCA_MT/Rt-MCA-VTI
OB/SMA/D_SMA_AT/SMA-VTI
SMA-VTI
OB/SMA/D_SMA_MT/SMA-VTI
OB/CELIACA/D_CELA_AT/Celiac A.-VTI
Celiac A.-VTI
OB/CELIACA/D_CELA_MT/Celiac A.-VTI
OB/CAROTID/D_CAR_AT/Lt-Car-VTI
OB/CAROTID/D_CAR_AT/Rt-Car-VTI
Carotid-VTI
OB/CAROTID/D_CAR_MT/Lt-Car-VTI
OB/CAROTID/D_CAR_MT/Rt-Car-VTI
OB/DUCTUSART/D_DA_AT/DA-HR
OB/DUCTUSART/D_DA_HR/DA-HR
OB/AO/D_AO_AT/Ao-HR
Ao-HR OB/AO/D_AO_MT/Ao-HR
OB/AO/D_AO_HR/Ao-HR
OB/MCA/D_MCA_AT/Lt-MCA-HR
OB/MCA/D_MCA_AT/Rt-MCA-HR
OB/MCA/D_MCA_MT/Lt-MCA-HR
MCA-HR
OB/MCA/D_MCA_MT/Rt-MCA-HR
OB/MCA/D_MCA_HR/Lt-MCA-HR
MCA-HR OB/SMA/D_SMA_MT/SMA-HR
OB/SMA/D_SMA_HR/SMA-HR
OB/CELIACA/D_CELA_AT/Celiac A.-HR
OB/CELIACA/D_CELA_HR/Celiac A.-HR
OB/CAROTID/D_CAR_AT/Lt-Car-HR
OB/CAROTID/D_CAR_AT/Rt-Car-HR
OB/CAROTID/D_CAR_MT/Lt-Car-HR
Carotid-HR
OB/CAROTID/D_CAR_MT/Rt-Car-HR
OB/CAROTID/D_CAR_HR/Lt-Car-HR
OB/CAROTID/D_CAR_HR/Rt-Car-HR
Ao-T OB/AO/D_AO_TIME/Ao-T
OB/MCA/D_MCA_TIME/Lt-MCA-T
MCA-T
OB/MCA/D_MCA_TIME/Rt-MCA-T
Acceleration Time (LN, 20168-1)
SMA-T OB/SMA/D_SMA_TIME/SMA-T
OB/CAROTID/D_CAR_TIME/Lt-Car-T
Carotid-T
OB/CAROTID/D_CAR_TIME/Rt-Car-T
OB/MCA/D_MCA_VOL/Rt-MCA-Vol. Flow
Volume flow(LN, 33878-0)
Carotid-Vol.Flow OB/CAROTID/D_CAR_VOL/Lt-Car-Vol. Flow
OB/CAROTID/D_CAR_VOL/Rt-Car-Vol. Flow
UmbV-Dia OB/UMBILICALART/D_UMBA
Umbilical Vein (SRT,
Umbilical Vein T-F1820)
OB/UMBVD/UmbV-Diam
Diam
OB/UMBILICALART/D_UMBA_PS/UmbA-PS
OB/UMBILICALART/D_UMBA_RI/UmbA-PS
OB/UMBILICALART/D_UMBA_PI/UmbA-PS
Peak Systolic
Umbilical Art.-PS
Velocity(LN, 11726-7)
OB/UMBILICALART/D_UMBA_PERS/UmbA-PS
OB/UMBILICALART/D_UMBA_AT/UmbA-PS
OB/UMBILICALART/D_UMBA_MT/UmbA-PS
OB/UTERINEART/D_UTA_PS/Lt-UtA-PS
OB/UTERINEART/D_UTA_PS/Rt-UtA-PS
OB/UTERINEART/D_UTA_RI/Lt-UtA-PS
OB/UTERINEART/D_UTA_RI/Rt-UtA-PS
OB/UTERINEART/D_UTA_PI/Lt-UtA-PS
OB/UTERINEART/D_UTA_PI/Rt-UtA-PS
Uterine Art.-PS
OB/UTERINEART/D_UTA_PERS/Lt-UtA-PS
OB/UTERINEART/D_UTA_PERS/Rt-UtA-PS
OB/UTERINEART/D_UTA_AT/Lt-UtA-PS
OB/UTERINEART/D_UTA_AT/Rt-UtA-PS
OB/UTERINEART/D_UTA_MT/Lt-UtA-PS
OB/UTERINEART/D_UTA_MT/Rt-UtA-PS
OB/UMBILICALARTLT/D_UMBALT_PS/Lt-UmA-PS
OB/UMBILICALARTLT/D_UMBALT_PS/Rt-UmA-PS
OB/UMBILICALARTLT/D_UMBALT_RI/Lt-UmA-PS
OB/UMBILICALARTLT/D_UMBALT_RI/Rt-UmA-PS
OB/UMBILICALARTLT/D_UMBALT_PI/Lt-UmA-PS
OB/UMBILICALARTLT/D_UMBALT_PI/Rt-UmA-PS
UmA-PS
OB/UMBILICALARTLT/D_UMBALT_PERS/Lt-UmA-PS
OB/UMBILICALARTLT/D_UMBALT_PERS/Rt-UmA-PS
OB/UMBILICALARTLT/D_UMBALT_AT/Lt-UmA-PS
OB/UMBILICALARTLT/D_UMBALT_AT/Rt-UmA-PS
OB/UMBILICALARTLT/D_UMBALT_MT/Lt-UmA-PS
OB/UMBILICALARTLT/D_UMBALT_MT/Rt-UmA-PS
OB/UMBILICALART/D_UMBA_ED/UmbA-ED
OB/UMBILICALART/D_UMBA_RI/UmbA-ED
End Diastolic
Umbilical Art.-ED OB/UMBILICALART/D_UMBA_PI/UmbA-ED
Velocity(LN, 11653-3)
OB/UMBILICALART/D_UMBA_PERS/UmbA-ED
OB/UMBILICALART/D_UMBA_AT/UmbA-ED
OB/UMBILICALART/D_UMBA_MT/UmbA-ED
OB/UTERINEART/D_UTA_ED/Lt-UtA-ED
OB/UTERINEART/D_UTA_ED/Rt-UtA-ED
OB/UTERINEART/D_UTA_RI/Lt-UtA-ED
OB/UTERINEART/D_UTA_RI/Rt-UtA-ED
OB/UTERINEART/D_UTA_PI/Lt-UtA-ED
OB/UTERINEART/D_UTA_PI/Rt-UtA-ED
Uterine Art.-ED
OB/UTERINEART/D_UTA_PERS/Lt-UtA-ED
OB/UTERINEART/D_UTA_PERS/Rt-UtA-ED
OB/UTERINEART/D_UTA_AT/Lt-UtA-ED
OB/UTERINEART/D_UTA_AT/Rt-UtA-ED
OB/UTERINEART/D_UTA_MT/Lt-UtA-ED
OB/UTERINEART/D_UTA_MT/Rt-UtA-ED
OB/UMBILICALARTLT/D_UMBALT_ED/Lt-UmA-ED
OB/UMBILICALARTLT/D_UMBALT_ED/Rt-UmA-ED
OB/UMBILICALARTLT/D_UMBALT_RI/Lt-UmA-ED
OB/UMBILICALARTLT/D_UMBALT_RI/Rt-UmA-ED
OB/UMBILICALARTLT/D_UMBALT_PI/Lt-UmA-ED
OB/UMBILICALARTLT/D_UMBALT_PI/Rt-UmA-ED
UmA-ED
OB/UMBILICALARTLT/D_UMBALT_PERS/Lt-UmA-ED
OB/UMBILICALARTLT/D_UMBALT_PERS/Rt-UmA-ED
OB/UMBILICALARTLT/D_UMBALT_AT/Lt-UmA-ED
OB/UMBILICALARTLT/D_UMBALT_AT/Rt-UmA-ED
OB/UMBILICALARTLT/D_UMBALT_MT/Lt-UmA-ED
OB/UMBILICALARTLT/D_UMBALT_MT/Rt-UmA-ED
OB/UMBILICALART/D_UMBA_RI/UmbA-RI
OB/UMBILICALART/D_UMBA_PERS/UmbA-RI
Resisitivity Index(LN,
Umbilical Art.-RI
12023-8)
OB/UMBILICALART/D_UMBA_AT/UmbA-RI
OB/UMBILICALART/D_UMBA_MT/UmbA-RI
OB/UTERINEART/D_UTA_RI/Lt-UtA-RI
OB/UTERINEART/D_UTA_RI/Rt-UtA-RI
OB/UTERINEART/D_UTA_PERS/Lt-UtA-RI
OB/UTERINEART/D_UTA_PERS/Rt-UtA-RI
Uterine Art.-RI
OB/UTERINEART/D_UTA_AT/Lt-UtA-RI
OB/UTERINEART/D_UTA_AT/Rt-UtA-RI
OB/UTERINEART/D_UTA_MT/Lt-UtA-RI
OB/UTERINEART/D_UTA_MT/Rt-UtA-RI
OB/UMBILICALARTLT/D_UMBALT_RI/Lt-UmA-RI
OB/UMBILICALARTLT/D_UMBALT_RI/Rt-UmA-RI
OB/UMBILICALARTLT/D_UMBALT_PERS/Lt-UmA-RI
OB/UMBILICALARTLT/D_UMBALT_PERS/Rt-UmA-RI
UmA-RI
OB/UMBILICALARTLT/D_UMBALT_AT/Lt-UmA-RI
OB/UMBILICALARTLT/D_UMBALT_AT/Rt-UmA-RI
OB/UMBILICALARTLT/D_UMBALT_MT/Lt-UmA-RI
OB/UMBILICALARTLT/D_UMBALT_MT/Rt-UmA-RI
OB/UMBILICALART/D_UMBA_RI/UmbA-PI
OB/UMBILICALART/D_UMBA_PI/UmbA-PI
Umbilical Art.-PI
OB/UMBILICALART/D_UMBA_AT/UmbA-PI
OB/UMBILICALART/D_UMBA_MT/UmbA-PI
OB/UTERINEART/D_UTA_RI/Lt-UtA-PI
OB/UTERINEART/D_UTA_PI/Rt-UtA-PI
Uterine Art.-PI
OB/UTERINEART/D_UTA_AT/Lt-UtA-PI
OB/UTERINEART/D_UTA_AT/Rt-UtA-PI
OB/UTERINEART/D_UTA_MT/Lt-UtA-PI
OB/UTERINEART/D_UTA_MT/Rt-UtA-PI
OB/UMBILICALARTLT/D_UMBALT_RI/Lt-UmA-PI
OB/UMBILICALARTLT/D_UMBALT_RI/Rt-UmA-PI
OB/UMBILICALARTLT/D_UMBALT_PI/Lt-UmA-PI
OB/UMBILICALARTLT/D_UMBALT_PI/Rt-UmA-PI
UmA-PI
OB/UMBILICALARTLT/D_UMBALT_AT/Lt-UmA-PI
OB/UMBILICALARTLT/D_UMBALT_AT/Rt-UmA-PI
OB/UMBILICALARTLT/D_UMBALT_MT/Lt-UmA-PI
OB/UMBILICALARTLT/D_UMBALT_MT/Rt-UmA-PI
OB/D_UMBV_TAMAX/UmbV-TAmax
OB/UMBILICALART/D_UMBA_PI/UmbA-TAmax
Umbilical
Art.-TAmax OB/UMBILICALART/D_UMBA_AT/UmbA-TAmax
OB/UMBILICALART/D_UMBA_MT/UmbA-TAmax
OB/UTERINEART/D_UTA_PI/Lt-UtA-Tamax
OB/UTERINEART/D_UTA_PI/Rt-UtA-TAmax
Uterine OB/UTERINEART/D_UTA_AT/Lt-UtA-TAmax
Art.-TAmax
OB/UTERINEART/D_UTA_AT/Rt-UtA-TAmax Time averaged peak
velocity(LN, 11692-1)
OB/UTERINEART/D_UTA_MT/Lt-UtA-TAmax
OB/UTERINEART/D_UTA_MT/Rt-UtA-TAmax
OB/UMBILICALARTLT/D_UMBALT_PI/Lt-UmA-TAmax
OB/UMBILICALARTLT/D_UMBALT_PI/Rt-UmA-TAmax
OB/UMBILICALARTLT/D_UMBALT_AT/Lt-UmA-TAmax
UmA-TAmax
OB/UMBILICALARTLT/D_UMBALT_AT/Rt-UmA-TAmax
OB/UMBILICALARTLT/D_UMBALT_MT/Lt-UmA-TAmax
OB/UMBILICALARTLT/D_UMBALT_MT/Rt-UmA-TAmax
OB/UMBILICALART/D_UMBA_AT/UmbA-TAmean
Umbilical
Art.-TAmean
OB/UMBILICALART/D_UMBA_MT/UmbA-TAmean Time averaged mean
velocity(LN, 20352-1)
Uterine OB/UTERINEART/D_UTA_AT/Lt-UtA-TAmean
Art.-TAmean
OB/UTERINEART/D_UTA_AT/Rt-UtA-TAmean
OB/UTERINEART/D_UTA_MT/Lt-UtA-TAmean
OB/UTERINEART/D_UTA_MT/Rt-UtA-TAmean
OB/UTERINEART/D_UTA_VOL/Lt-UtA-TAmean
OB/UTERINEART/D_UTA_VOL/Rt-UtA-TAmean
OB/UMBILICALARTLT/D_UMBALT_AT/Lt-UmA-TAmean
OB/UMBILICALARTLT/D_UMBALT_AT/Rt-UmA-TAmean
OB/UMBILICALARTLT/D_UMBALT_MT/Lt-UmA-TAmean
UmA-TAmean
OB/UMBILICALARTLT/D_UMBALT_MT/Rt-UmA-TAmean
OB/UMBILICALARTLT/D_UMBALT_VOL/Lt-UmA-TAmean
OB/UMBILICALARTLT/D_UMBALT_VOL/Rt-UmA-TAmean
OB/UMBILICALART/D_UMBA_PERS/UmbA-S/D
OB/UMBILICALART/D_UMBA_MT/UmbA-S/D
OB/UTERINEART/D_UTA_PERS/Lt-UtA-S/D
OB/UTERINEART/D_UTA_PERS/Rt-UtA-S/D
OB/UTERINEART/D_UTA_AT/Lt-UtA-S/D
Uterine Art.-S/D
OB/UTERINEART/D_UTA_AT/Rt-UtA-S/D
Systolic to Diastolic
OB/UTERINEART/D_UTA_MT/Lt-UtA-S/D Velocity Ratio(LN,
12144-2)
OB/UTERINEART/D_UTA_MT/Rt-UtA-S/D
OB/UMBILICALARTLT/D_UMBALT_PERS/Lt-UmA-S/D
OB/UMBILICALARTLT/D_UMBALT_PERS/Rt-UmA-S/D
OB/UMBILICALARTLT/D_UMBALT_AT/Lt-UmA-S/D
UmA-S/D
OB/UMBILICALARTLT/D_UMBALT_AT/Rt-UmA-S/D
OB/UMBILICALARTLT/D_UMBALT_MT/Lt-UmA-S/D
OB/UMBILICALARTLT/D_UMBALT_MT/Rt-UmA-S/D
OB/UMBILICALART/D_UMBA_AT/UmbA-AT
Umbilical Art.-AT
OB/UMBILICALART/D_UMBA_MT/UmbA-AT Acceleration Time(LN,
20168-1)
OB/UTERINEART/D_UTA_AT/Lt-UtA-AT
Uterine Art.-AT
OB/UTERINEART/D_UTA_AT/Rt-UtA-AT
OB/UTERINEART/D_UTA_MT/Lt-UtA-AT
OB/UTERINEART/D_UTA_MT/Rt-UtA-AT
OB/UMBILICALARTLT/D_UMBALT_AT/Lt-UmA-AT
OB/UMBILICALARTLT/D_UMBALT_AT/Rt-UmA-AT
UmA-AT
OB/UMBILICALARTLT/D_UMBALT_MT/Lt-UmA-AT
OB/UMBILICALARTLT/D_UMBALT_MT/Rt-UmA-AT
OB/UMBILICALART/D_UMBA_AT/UmbA-VTI
Umbilical Art.-VTI
OB/UMBILICALART/D_UMBA_MT/UmbA-VTI
OB/UTERINEART/D_UTA_AT/Lt-UtA-VTI
OB/UTERINEART/D_UTA_AT/Rt-UtA-VTI
Uterine Art.-VTI
OB/UTERINEART/D_UTA_MT/Lt-UtA-VTI Velocity Time
Integral(LN, 20354-7)
OB/UTERINEART/D_UTA_MT/Rt-UtA-VTI
OB/UMBILICALARTLT/D_UMBALT_AT/Lt-UmA-VTI
OB/UMBILICALARTLT/D_UMBALT_AT/Rt-UmA-VTI
UmA-VTI
OB/UMBILICALARTLT/D_UMBALT_MT/Lt-UmA-VTI
OB/UMBILICALARTLT/D_UMBALT_MT/Rt-UmA-VTI
OB/UMBILICALART/D_UMBA_AT/UmbA-HR
OB/UMBILICALART/D_UMBA_HR/UmbA-HR
OB/UTERINEART/D_UTA_AT/Lt-UtA-HR
OB/UTERINEART/D_UTA_AT/Rt-UtA-HR
OB/UTERINEART/D_UTA_MT/Lt-UtA-HR
Uterine Art.-HR
OB/UTERINEART/D_UTA_MT/Rt-UtA-HR Heart Rate(LN, 8867-4)
OB/UTERINEART/D_UTA_HR/Lt-UtA-HR
OB/UTERINEART/D_UTA_HR/Rt-UtA-HR
OB/UMBILICALARTLT/D_UMBALT_AT/Lt-UmA-HR
OB/UMBILICALARTLT/D_UMBALT_AT/Rt-UmA-HR
UmA-HR
OB/UMBILICALARTLT/D_UMBALT_MT/Lt-UmA-HR
OB/UMBILICALARTLT/D_UMBALT_MT/Rt-UmA-HR
OB/UMBILICALARTLT/D_UMBALT_HR/Lt-UmA-HR
OB/UMBILICALARTLT/D_UMBALT_HR/Rt-UmA-HR
OB/UTERINEART/D_UTA_TIME/Lt-UtA-T
Uterine Art.-T
Acceleration Time(LN,
OB/UTERINEART/D_UTA_TIME/Rt-UtA-T
20168-1)
OB/UMBILICALARTLT/D_UMBALT_TIME/Lt-UmA-T
UmA-T
OB/UMBILICALARTLT/D_UMBALT_TIME/Rt-UmA-T
OB/UMBILICALARTLT/D_UMBALT_VOL/Lt-UmA-Vol. Flow
Volume flow(LN,
UmA-Vol.Flow 33878-0)
OB/UMBILICALARTLT/D_UMBALT_VOL/Rt-UmA-Vol. Flow
Umbilical
OB/D_UMBV_TAMAX/UmbV-VenFlow
Vein-Vol.Flow
GYN/GYN_M_OVARIANL/GYN_M_OVARIAN_VESSEL/Rt-OvA-D
GYN/GYN_M_UTERINE_ART/GYN_M_UTERINE_VESSEL/Lt-UtA-D
GYN/GYN_M_OVARIANL/GYN_M_OVARIAN_STENOSIS/Lt-OvA-D1
GYN/GYN_M_UTERINE_ART/GYN_M_UTERINE_STENOSIS/Rt-UtA-D1
GYN/GYN_M_OVARIANL/GYN_M_OVARIAN_STENOSIS/Lt-OvA-D2
GYN/GYN_M_UTERINE_ART/GYN_M_UTERINE_STENOSIS/Rt-UtA-D2
GYN/GYN_M_UTERINE_ART/GYN_M_UTERINE_STENOSIS/Lt-%Sten
GYN/GYN_M_UTERINE_ART/GYN_M_UTERINE_STENOSIS/Rt-%Sten
GYN/GYN_M_OVARIANL/GYN_M_OVARIAN_TIME/Lt-OvA-T
GYN/GYN_M_UTERINE_ART/GYN_M_UTERINE_TIME/Rt-UtA-T
GYN/GYN_M_OVARIANL/GYN_M_OVARIAN_HR/Lt-OvA-HR
GYN/GYN_M_OVARIANL/GYN_M_OVARIAN_HR/Rt-OvA-HR
Heart Rate Heart Rate(LN, 8867-4)
GYN/GYN_M_UTERINE_ART/GYN_M_UTERINE_HR/Lt-UtA-HR
GYN/GYN_M_UTERINE_ART/GYN_M_UTERINE_HR/Rt-UtA-HR
GYN/OVARIANART/GYN_D_OVARIAN_PS/Lt-OvA-PS
GYN/OVARIANART/GYN_D_OVARIAN_PS/Rt-OvA-PS
GYN/OVARIANART/GYN_D_OVARIAN_RI/Lt-OvA-PS
GYN/OVARIANART/GYN_D_OVARIAN_RI/Rt-OvA-PS
GYN/OVARIANART/GYN_D_OVARIAN_PI/Lt-OvA-PS
GYN/OVARIANART/GYN_D_OVARIAN_GROUP/Rt-OvA-PS
GYN/OVARIANART/GYN_D_OVARIAN_AT/Lt-OvA-PS
GYN/OVARIANART/GYN_D_OVARIAN_AT/Rt-OvA-PS
GYN/OVARIANART/GYN_D_OVARIAN_MT/Lt-OvA-PS
GYN/OVARIANART/GYN_D_OVARIAN_MT/Rt-OvA-PS
GYN/OVARIANART/GYN_D_OVARIAN_ED/Lt-OvA-ED
GYN/OVARIANART/GYN_D_OVARIAN_ED/Rt-OvA-ED
GYN/OVARIANART/GYN_D_OVARIAN_RI/Lt-OvA-ED
End Diastolic
Ovarian Art.-ED GYN/OVARIANART/GYN_D_OVARIAN_RI/Rt-OvA-ED
Velocity(LN, 11653-3)
GYN/OVARIANART/GYN_D_OVARIAN_PI/Lt-OvA-ED
GYN/OVARIANART/GYN_D_OVARIAN_PI/Rt-OvA-ED
GYN/OVARIANART/GYN_D_OVARIAN_GROUP/Lt-OvA-ED
GYN/OVARIANART/GYN_D_OVARIAN_GROUP/Rt-OvA-ED
GYN/OVARIANART/GYN_D_OVARIAN_AT/Lt-OvA-ED
GYN/OVARIANART/GYN_D_OVARIAN_AT/Rt-OvA-ED
GYN/OVARIANART/GYN_D_OVARIAN_MT/Lt-OvA-ED
GYN/OVARIANART/GYN_D_OVARIAN_MT/Rt-OvA-ED
GYN/OVARIANART/GYN_D_OVARIAN_RI/Lt-OvA-RI
GYN/OVARIANART/GYN_D_OVARIAN_RI/Rt-OvA-RI
GYN/OVARIANART/GYN_D_OVARIAN_GROUP/Lt-OvA-RI
GYN/OVARIANART/GYN_D_OVARIAN_AT/Rt-OvA-RI
GYN/OVARIANART/GYN_D_OVARIAN_MT/Lt-OvA-RI
GYN/OVARIANART/GYN_D_OVARIAN_MT/Rt-OvA-RI
GYN/OVARIANART/GYN_D_OVARIAN_MT/Lt-OvA-TAmax
GYN/OVARIANART/GYN_D_OVARIAN_MT/Rt-OvA-TAmax
GYN/OVARIANART/GYN_D_OVARIAN_AT/Lt-OvA-TAmax
GYN/OVARIANART/GYN_D_OVARIAN_AT/Rt-OvA-TAmax
GYN/OVARIANART/GYN_D_OVARIAN_MT/Lt-OvA-TAmean
GYN/OVARIANART/GYN_D_OVARIAN_AT/Rt-OvA-TAmean
GYN/OVARIANART/GYN_D_OVARIAN_RI/Lt-OvA-PI
GYN/OVARIANART/GYN_D_OVARIAN_RI/Rt-OvA-PI
GYN/OVARIANART/GYN_D_OVARIAN_AT/Lt-OvA-PI
GYN/OVARIANART/GYN_D_OVARIAN_AT/Rt-OvA-PI
GYN/OVARIANART/GYN_D_OVARIAN_MT/Lt-OvA-PI
GYN/OVARIANART/GYN_D_OVARIAN_MT/Rt-OvA-PI
GYN/OVARIANART/GYN_D_OVARIAN_GROUP/Lt-OvA-S/D
GYN/OVARIANART/GYN_D_OVARIAN_GROUP/Rt-OvA-S/D
GYN/OVARIANART/GYN_D_OVARIAN_MT/Lt-OvA-S/D
GYN/OVARIANART/GYN_D_OVARIAN_MT/Rt-OvA-S/D
GYN/OVARIANART/GYN_D_OVARIAN_AT/Lt-OvA-VTI
GYN/OVARIANART/GYN_D_OVARIAN_MT/Rt-OvA-VTI
GYN/OVARIANART/GYN_D_OVARIAN_AT/Lt-OvA-AT
GYN/OVARIANART/GYN_D_OVARIAN_MT/Rt-OvA-AT
GYN/OVARIANART/GYN_D_OVARIAN_AT/Lt-OvA-HR
GYN/OVARIANART/GYN_D_OVARIAN_AT/Rt-OvA-HR
GYN/OVARIANART/GYN_D_OVARIAN_MT/Lt-OvA-HR
Ovarian Art.-HR Heart Rate(LN, 8867-4)
GYN/OVARIANART/GYN_D_OVARIAN_MT/Rt-OvA-HR
GYN/OVARIANART/GYN_D_OVARIAN_HR/Lt-OvA-HR
GYN/OVARIANART/GYN_D_OVARIAN_HR/Rt-OvA-HR
GYN/D_UTERINEART_A/GYN_D_UTERINE_PS/Lt-UtA-PS
GYN/D_UTERINEART_A/GYN_D_UTERINE_PS/Rt-UtA-PS
GYN/D_UTERINEART_A/GYN_D_UTERINE_RI/Lt-UtA-PS
Peak Systolic
Uterine Art.-PS GYN/D_UTERINEART_A/GYN_D_UTERINE_RI/Rt-UtA-PS
Velocity(LN, 11726-7)
GYN/D_UTERINEART_A/GYN_D_UTERINE_PI/Lt-UtA-PS
GYN/D_UTERINEART_A/GYN_D_UTERINE_PI/Rt-UtA-PS
GYN/D_UTERINEART_A/GYN_D_UTERINE_GROUP/Lt-UtA-PS
GYN/D_UTERINEART_A/GYN_D_UTERINE_GROUP/Rt-UtA-PS
GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Lt-UtA-PS
GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Rt-UtA-PS
GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Lt-UtA-PS
GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Rt-UtA-PS
GYN/D_UTERINEART_A/GYN_D_UTERINE_ED/Lt-UtA-ED
GYN/D_UTERINEART_A/GYN_D_UTERINE_ED/Rt-UtA-ED
GYN/D_UTERINEART_A/GYN_D_UTERINE_RI/Lt-UtA-ED
GYN/D_UTERINEART_A/GYN_D_UTERINE_RI/Rt-UtA-ED
GYN/D_UTERINEART_A/GYN_D_UTERINE_PI/Lt-UtA-ED
GYN/D_UTERINEART_A/GYN_D_UTERINE_GROUP/Rt-UtA-ED
GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Lt-UtA-ED
GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Rt-UtA-ED
GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Lt-UtA-ED
GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Rt-UtA-ED
GYN/D_UTERINEART_A/GYN_D_UTERINE_RI/Lt-UtA-RI
GYN/D_UTERINEART_A/GYN_D_UTERINE_RI/Rt-UtA-RI
GYN/D_UTERINEART_A/GYN_D_UTERINE_GROUP/Lt-UtA-RI
GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Rt-UtA-RI
GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Lt-UtA-RI
GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Rt-UtA-RI
GYN/D_UTERINEART_A/GYN_D_UTERINE_PI/Lt-UtA-TAmax
GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Rt-UtA-TAmax
GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Lt-UtA-TAmax
GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Rt-UtA-TAmax
GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Lt-UtA-TAmean
GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Rt-UtA-TAmean
GYN/D_UTERINEART_A/GYN_D_UTERINE_VOL/Lt-UtA-TAmean
GYN/D_UTERINEART_A/GYN_D_UTERINE_VOL/Rt-UtA-TAmean
GYN/D_UTERINEART_A/GYN_D_UTERINE_RI/Lt-UtA-PI
GYN/D_UTERINEART_A/GYN_D_UTERINE_RI/Rt-UtA-PI
GYN/D_UTERINEART_A/GYN_D_UTERINE_PI/Lt-UtA-PI
GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Rt-UtA-PI
GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Lt-UtA-PI
GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Rt-UtA-PI
GYN/D_UTERINEART_A/GYN_D_UTERINE_GROUP/Lt-UtA-S/D
GYN/D_UTERINEART_A/GYN_D_UTERINE_GROUP/Rt-UtA-S/D
GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Lt-UtA-S/D
GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Rt-UtA-S/D
GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Lt-UtA-AT
GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Rt-UtA-AT
GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Lt-UtA-VTI
Velocity Time
Uterine Art.-VTI GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Rt-UtA-VTI
Integral(LN, 20354-7)
GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Lt-UtA-VTI
GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Rt-UtA-VTI
GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Lt-UtA-HR
GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Rt-UtA-HR
GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Lt-UtA-HR
Uterine Art.-HR Heart rate(LN, 8867-4)
GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Rt-UtA-HR
GYN/D_UTERINEART_A/GYN_D_UTERINE_HR/Lt-UtA-HR
GYN/D_UTERINEART_A/GYN_D_UTERINE_HR/Rt-UtA-HR
GYN/VESSEL_D/GYN_D_VESSEL_PS/Ves-PS
GYN/VESSEL_D/GYN_D_VESSEL_RI/Ves-PS
GYN/VESSEL_D/GYN_D_VESSEL_AT/Ves-PS
GYN/VESSEL_D/GYN_D_VESSEL_MT/Ves-PS
GYN/VESSEL_D/GYN_D_VESSEL_ED/Ves-ED
GYN/VESSEL_D/GYN_D_VESSEL_RI/Ves-ED
GYN/VESSEL_D/GYN_D_VESSEL_AT/Ves-ED
GYN/VESSEL_D/GYN_D_VESSEL_MT/Ves-ED
GYN/VESSEL_D/GYN_D_VESSEL_RI/Ves-RI
GYN/VESSEL_D/GYN_D_VESSEL_MT/Ves-RI
GYN/VESSEL_D/GYN_D_VESSEL_AT/Ves-PI
GYN/VESSEL_D/GYN_D_VESSEL_MT/Ves-PI
GYN/VESSEL_D/GYN_D_VESSEL_PI/Ves-TAmax
Time averaged peak
Vessel-TAmax GYN/VESSEL_D/GYN_D_VESSEL_AT/Ves-TAmax
velocity(LN, 11692-1)
GYN/VESSEL_D/GYN_D_VESSEL_MT/Ves-TAmax
GYN/VESSEL_D/GYN_D_VESSEL_AT/Ves-AT
Acceleration Time(LN,
Vessel-AT GYN/VESSEL_D/GYN_D_VESSEL_MT/Ves-AT
20168-1)
GYN/VESSEL_D/GYN_D_VESSEL_TIME/Ves-T
GYN/VESSEL_D/GYN_D_VESSEL_GROUP/Ves-S/D
Systolic to Diastolic
Vessel-S/D GYN/VESSEL_D/GYN_D_VESSEL_AT/Ves-S/D Velocity Ratio(LN,
12144-2)
GYN/VESSEL_D/GYN_D_VESSEL_MT/Ves-S/D
GYN/VESSEL_D/GYN_D_VESSEL_AT/Ves-HR
GYN/VESSEL_D/GYN_D_VESSEL_HR/Ves-HR
A.3 .4 Blood Vessel of Head DCID (12105) Intracranial Cerebral Vessels DCID (12106) Unpaired Intracranial
Cerebral Vessels (unilateral)
Laterality Label Anatomy Item ID Code Meaning (Coding Scheme
Designator, Code Value)
Vein
Left
(SRT ,G-A
101) RA Renal Artery Renal Artery (SRT, T-46640)
Right(SRT
,G-A100)
A.3. 11 Artery of Abdomen DCID (12111) Abdominal Arteries (lateral) DCID (12112) Abdominal Arteries (unilateral)
Laterality Label Anatomy Item ID Code Meaning (Coding Scheme
Designator, Code Value)
A.3. 12 Vein of Abdomen DCID (12113) Abdominal Veins (lateral) DCID (12114) Abdominal Veins (unilateral)
Laterality Label Anatomy Item ID Code Meaning (Coding Scheme
Designator, Code Value)
2D mode(SRT,G-03A2)
Card/CARDIAC_LV_MASS_AL/C/LV
LVMI Left Ventricle Mass Index(SONO, SONO-5202-28)
MI(A-L)
Card/CARDIAC_LV_BULLET/B/LVA
LVAd Left Ventricular Diastolic Area(LN, G-0375)
d
Card/CARDIAC_LV_BULLET/B/LVA
LVAs Left Ventricular Systolic Area(LN, G-0374)
s
Card/CARDIAC_LV_SIMP/B/EDV(A4
C)
Card/CARDIAC_LV_SIMP/B/EDV(A2
EDV Left Ventricular End Diastolic Volume (LN, 18026-5)
C)
Card/CARDIAC_LV_SIMP/B/EDV(B
P)
Card/CARDIAC_LV_SIMP/B/ESV(A4
C)
Card/CARDIAC_LV_SIMP/B/ESV(A2
ESV Left Ventricular End Systolic Volume (LN, 18148-7)
C)
Card/CARDIAC_LV_SIMP/B/ESV(BP
)
Card/CARDIAC_LV_SIMP/B/SV(A4C
)
Card/CARDIAC_LV_SIMP/B/SV(BP)
Card/CARDIAC_LV_SIMP/B/CO(A4
C)
Card/CARDIAC_LV_SIMP/B/CO(BP)
Card/CARDIAC_LV_SIMP/B/EF(A4C
)
Card/CARDIAC_LV_SIMP/B/EF(BP)
Card/CARDIAC_LV_SIMP/B/SI(A4C)
Card/CARDIAC_LV_SIMP/B/SI(BP)
Card/CARDIAC_LV_SIMP/B/CI(A4C)
Card/CARDIAC_LV_SIMP/B/CI(BP)
Cube Method(DCM,125206)
CUBE CUBE
2D mode(SRT,G-03A2)
Card/CARDIAC_LV_MASS_CUBE
LVM Left Ventricle Mass(LN, 18087-0)
/B/LVM(Cube)
Card/CARDIAC_LV_MASS_CUBE/B
LVMI Left Ventricle Mass Index(SONO, SONO-5202-30)
/LVMI(Cube)
Teichholz(DCM,125209)
TEICHHOLZ TEICHHOLZ
2D mode(SRT,G-03A2)
Card/CARDIAC_DIMENSION1/B/ED
EDV Left Ventricular End Diastolic Volume (LN, 18026-5)
V(Teich)
Card/CARDIAC_DIMENSION1/B/ES
ESV Left Ventricular End Systolic Volume (LN, 18148-7)
V(Teich)
Teich)
Card/CARDIAC_DIMENSION1/B/CO
CO Cardiac Output (SRT, F-32100)
(Teich)
Card/CARDIAC_DIMENSION1/B/EF(
EF Left Ventricular Ejection Fraction (LN, 18043-0)
Teich)
Card/CARDIAC_DIMENSION1/B/SI(
SI Stroke Index (SRT, F-00078)
Teich)
Card/CARDIAC_DIMENSION1/B/CI(
CI Cardiac Index (SRT, F-32110)
Teich)
Card/CARDIAC_DIMENSION1/B/FS(
FS Left Ventricular Fractional Shortening(LN, 18051-3)
Teich)
Card/CARDIAC_DIMENSION1/B/IVS
IVSs Interventricular Septum Systolic Thickness(18158-6)
s
Card/CARDIAC_DIMENSION1/B/
LVM(Cube) Left Ventricle Mass(LN, 18087-7)
LVM(Cube)
Cube Method(DCM,125206)
CUBE CUBE
M mode(SRT,G-0394)
/M/LVPWd 18152-9)
Card/CARDIAC_M_VOLUME_CUBE
EDV Left Ventricular End Diastolic Volume (LN, 18026-5)
/M/EDV(Cube)
Card/CARDIAC_M_VOLUME_CUBE
ESV Left Ventricular End Systolic Volume (LN, 18148-7)
/M/ESV(Cube)
Card/CARDIAC_M_VOLUME_CUBE
SV Stroke Volume (SRT, F-32120)
/M/SV(Cube)
Card/CARDIAC_M_VOLUME_CUBE
CO Cardiac Output (SRT, F-32100)
/M/CO(Cube)
Card/CARDIAC_M_VOLUME_CUBE
EF Left Ventricular Ejection Fraction (LN, 18043-0)
/M/EF(Cube)
Card/CARDIAC_M_VOLUME_CUBE
FS Left Ventricular Fractional Shortening (LN, 18051-3)
/M/FS(Cube)
Card/CARDIAC_M_VOLUME_CUBE
SI Stroke Index (SRT, F-00078)
/M/SI(Cube)
Card/CARDIAC_M_VOLUME_CUBE
CI Cardiac Index (SRT, F-32110)
/M/CI(Cube)
Teichholz(DCM,125209)
TEICHHOLZ TEICHHOLZ
M mode(SRT,G-0394)
Card/CARDIAC_M_DIMENSION1/M/
EDV Left Ventricular End Diastolic Volume (LN, 18026-5)
EDV(Teich)
Card/CARDIAC_M_DIMENSION1/M/
ESV Left Ventricular End Systolic Volume (LN, 18148-7)
ESV(Teich)
Card/CARDIAC_M_DIMENSION1/M/
SV Stroke Volume (SRT, F-32120)
SV(Teich)
Card/CARDIAC_M_DIMENSION1/M/
CO Cardiac Output (SRT, F-32100)
CO(Teich)
Card/CARDIAC_M_DIMENSION1/M/
EF Left Ventricular Ejection Fraction (LN, 18043-0)
EF(Teich)
Card/CARDIAC_M_DIMENSION1/M/
FS Left Ventricular Fractional Shortening (LN, 18051-3)
FS(Teich)
Card/CARDIAC_M_DIMENSION1/M/
SI Stroke Index (SRT, F-00078)
SI(Teich)
Card/CARDIAC_M_DIMENSION1/M/
CI Cardiac Index (SRT, F-32110)
CI(Teich)
Card/CARDIAC_LV_SIMP/C/EDV(B
EDV Left Ventricular End Diastolic Volume(LN, 18026-5)
P)
Card/CARDIAC_LV_SIMP/C/ESV(B
ESV Left Ventricular End Systolic Volume(LN, 18148-7)
P)
Card/CARDIAC_LV_SIMP/C/EF(A4C
)
Card/CARDIAC_LV_SIMP/C/EF(BP)
Card/CARDIAC_LV_SIMP/C/SV(A4C
)
Card/CARDIAC_LV_SIMP/C/SV(BP)
Card/CARDIAC_LV_SIMP/C/CO(A4
C)
Card/CARDIAC_LV_SIMP/C/CO(BP)
Card/CARDIAC_LV_SIMP/C/CI(A4C)
Card/CARDIAC_LV_SIMP/C/CI(BP)
Card/CARDIAC_LV_SIMP/C/SI(A4C)
Card/CARDIAC_LV_SIMP/C/SI(BP)
2D mode(SRT,G-03A2)
M mode(SRT,G-0394)
Card/CARDIAC_D_TV_RV_TEI/D/R
RVET(b) Pulmonic Valve Ejection Time(LN, 18042-2)
VET(b)
Card/CARDIAC_D_TV_RV_TEI/D/R
RV TEI Tei Index(Tei Index(SONO, SONO-5201-7))
V TEI
Card/CARDIAC_D_PV_RVET/D/RV
RVET Pulmonic Valve Ejection Time(LN, 18042-2)
ET
Card/CARDIAC_D_PV_RVPEP/D/R
RVPEP Pre-Ejection Period(LN, 18068-7)
VPEP
2D mode(SRT,G-03A2)
Card/CARDIAC_DIMENSION3/B/LA/
AO Left Atrium to Aortic Root Ratio(LN, 17985-3)
AO
Card/CARDIAC_LA_SIMP/B/LA ESV
(A4C)
Card/CARDIAC_LA_SIMP/B/LA ESV
ESV Left Atrial ES Volume(DCM, 122408)
(BP)
Card/CARDIAC_LA_SIMP/B/LA ESV
(A2C)
Card/CARDIAC_DIMENSION3/C/LA/
AO Left Atrium to Aortic Root Ratio(LN, 17985-3)
AO
Card/CARDIAC_LA_SIMP/C/LA ESV
(A4C)
Card/CARDIAC_LA_SIMP/C/LA ESV
ESV Left Atrial ES Volume(DCM, 122408)
(BP)
Card/CARDIAC_LA_SIMP/C/LA ESV
(A2C)
M mode(SRT,G-0394)
Card/CARDIAC_M_DIMENSION3/M/L
AO Left Atrium to Aortic Root Ratio(LN, 17985-3)
A/AO
2D mode(SRT,G-03A2)
Doppler Pulsed(SRT,R-409E4)
Card/CARDIAC_D_RVSP_TRVMAX/
RAP Right Atrium Systolic Pressure(LN, 18070-3)
D/RAP
Card/CARDIAC_D_PV_PAEDP/D/RA
RAP Right Atrium Systolic Pressure(LN, 18070-3)
P
2D mode(SRT,G-03A2)
M mode(SRT,G-0394)
Card/CARDIAC_M_DIMENSION_AS
ACS Aortic Valve Cusp Separation(LN, 17996-0)
C/M/ACS
Card/CARDIAC_DIMENSION_ASC/C
ACS Aortic Valve Cusp Separation(LN, 17996-0)
/ACS
Card/CARDIAC_DIMENSION_AVA/C
AVA Cardiovascular Orifice Area(LN, G-038E)
/AVA
Card/CARDIAC_PISA_AR_RAD/C/R
Rad Flow Radius(SONO, SONO-5202-21)
ad
Doppler Pulsed(SRT,R-409E4)
Card/CARDIAC_D_AV_ARVMAX/D/A
AR Vmax Maximum Velocity(LN, 20351-3)
R Vmax
Card/CARDIAC_D_PISA_MS/D/AV
AV Vmax Peak Systolic Velocity(LN, 11726-7)
Vmax
Card/CARDIAC_D_PISA_MS/D/AV
AV VTI Peak Systolic Velocity(LN, 11726-7)
VTI
Card/CARDIAC_D_PISA_MS/D/AR
AR VTI Peak Systolic Velocity(LN, 11726-7)
VTI
R DecT
Card/CARDIAC_D_AV_ARPHT/D/AR
AR PHT Pressure Half-Time(LN, 20280-4)
PHT
Card/CARDIAC_D_AV_AVVTI/D/AV
AV Vmax Peak Systolic Velocity(LN, 11726-7)
Vmax
Card/CARDIAC_D_AV_AVVTI/D/AV
AV PGmax Peak Gradient(LN, 20247-3)
PGmax
Card/CARDIAC_D_AV_AVVTI/D/AV
AV Vmean Mean Velocity(LN, 20352-1)
Vmean
Card/CARDIAC_D_AV_AVVTI/D/AV
AV PGmean Mean Gradient(LN, 20256-4)
PGmean
Card/CARDIAC_D_AV_AVVTI/D/AV
AV VTI Velocity Time Integral(LN, 20354-7)
VTI
Card/CARDIAC_D_AVA_VTI/D/AV
AV Vmax Maximum Velocity(LN, 11726-7)
Vmax
Card/CARDIAC_D_AVA_VTI/D/AV
AV PGmax Peak Gradient(LN, 20247-3)
PGmax
Card/CARDIAC_D_AVA_VTI/D/AV
AV Vmean Mean Velocity(LN, 20352-1)
Vmean
Card/CARDIAC_D_AVA_VTI/D/AV
AV PGmean Mean Gradient(LN, 20256-4)
PGmean
Card/CARDIAC_D_AVA_VTI/D/AVA(
AVA(VTI) Cardiovascular Orifice Area(SRT, G-038E)
VTI)
Card/CARDIAC_D_AVA_VTI/D/AV
AV VTI Velocity Time Integral(LN, 20354-7)
VTI
Card/CARDIAC_D_AV_AVVMAX/D/A
AV Vmax Maximum Velocity(LN, 11726-7)
V Vmax
Card/CARDIAC_D_AV_AVVMAX/D/A
AV PGmax Peak Gradient(LN, 20247-3)
V PGmax
Card/CARDIAC_D_AV_AR/D/AR
AR Vmax Maximum Velocity(LN, 11726-7)
Vmax
Card/CARDIAC_D_AV_AR/D/AR
AR PGmax Peak Gradient(LN, 20247-3)
PGmax
Card/CARDIAC_D_AV_AR/D/AR
AR Vmean Mean Velocity(LN, 20352-1)
Vmean
Card/CARDIAC_D_AV_AR/D/AR
AR PGmean Mean Gradient(LN, 20256-4)
PGmean
AR Card/CARDIAC_D_AV_AR/D/AR
Flow Radius(SONO, SONO-5202-21)
Radius(2D) Radius(2D)
AR Flow Card/CARDIAC_D_AV_AR/D/AR
Peak Instantaneous Flow Rate(LN, 34141-2)
Rate Flow Rate
Card/CARDIAC_D_AV_AR/D/AR
AR ERO Cardiovascular Orifice Area(SRT, G-038E)
ERO
Card/CARDIAC_D_AV_AR/D/AR
AR Volume Volume Flow(LN, 33878-0)
Volume
Card/CARDIAC_D_AV_AR/D/AR
AR Fraction Regurgitant Fraction(SRT, G-0390)
Fraction
Card/CARDIAC_DIMENSION_EPSS/B
EPSS Mitral Valve EPSS, E wave (LN, 18036-4)
/EPSS
Card/CARDIAC_DIMENSION_MCS/B/
MCS Mitral Valve Cusp Separation(SONO, SONO-5202-16)
MCS
Card/CARDIAC_DIMENSION_MVDIA
MV Diam Cardiovascular Orifice Diameter(SRT, G-038F)
M/B/MV Diam
Card/CARDIAC_DIMENSION_MVA/B/
MVA Cardiovascular Orifice Area(SRT, G-038E)
MVA
M mode M mode(SRT,G-0394)
Card/CARDIAC_M_DIMENSION_MCS
MCS Mitral Valve Cusp Separation(SONO, SONO-5202-22)
/M/MCS
MV E-F Card/CARDIAC_M_TSLOPE_MVEF/M
Mitral Valve E-F Slope by M-Mode(LN, 18040-6)
Slope /MV E-F Slope
MV E Card/CARDIAC_M_TSLOPE_MVEAM
Amplitude E Wave Mmode(SONO, SONO-5202-24)
Amp P/M/MV E Amp
MV A Card/CARDIAC_M_TSLOPE_MVAAM
Amplitude A Wave Mmode(SONO, SONO-5202-25)
Amp P/M/MV A Amp
Card/CARDIAC_DIMENSION_EPSS/C
EPSS Mitral Valve EPSS, E wave (LN, 18036-4)
/EPSS
Card/CARDIAC_DIMENSION_MCS/C/
MCS Mitral Valve Cusp Separation (SONO, SONO-5202-16)
MCS
Card/CARDIAC_DIMENSION_MVDIA
MV Diam Cardiovascular Orifice Diameter(SRT, G-038F)
M/C/MV Diam
Card/CARDIAC_DIMENSION_MVA/C/
MVA Cardiovascular Orifice Area(SRT, G-038E)
MVA
Card/CARDIAC_PISA_MR_RAD/C/Ra
Rad Dome Radius(SONO, SONO-5202-20)
d
Card/CARDIAC_PISA_MS_RAD/C/Ra
Rad Dome Radius(SONO, SONO-5202-20)
d
Doppler Pulsed(SRT,R-409E4)
E Vel Card/CARDIAC_D_MV/D/MV E Vel Mitral Valve E-Wave Peak Velocity (LN, 18037-2)
A Vel Card/CARDIAC_D_MV/D/MV A Vel Mitral Valve A-Wave Peak Velocity (LN, 17978-8)
Card/CARDIAC_D_MV_APHT/D/MV
PHT Pressure Half-Time(LN, 20280-4)
PHT
MVA(PH Card/CARDIAC_D_MV_APHT/D/MVA(
Cardiovascular Orifice Area(SRT, G-038E)
T) PHT)
Card/CARDIAC_D_MV_EDUR/D/MV E
E Dur Mitral Valve E-Wave Duration (SONO, SONO-5202-4)
Dur
Card/CARDIAC_D_MV_ADUR/D/MV A
A Dur Mitral Valve A-Wave Duration (SRT, G-0385)
Dur
Card/CARDIAC_D_MVA_VTI/D/MV
MV VTI Velocity Time Integral(LN , 20354-7)
VTI
Card/CARDIAC_D_MVA_VTI/D/MV
MV Vmax Maximum Velocity(LN, 11726-7)
Vmax
MV Card/CARDIAC_D_MVA_VTI/D/MV
Peak Gradient(LN, 11726-7)
PGmax PGmax
MV Card/CARDIAC_D_MVA_VTI/D/MV
Mean Velocity(LN, 11726-7)
Vmean Vmean
MV Card/CARDIAC_D_MVA_VTI/D/MV
Mean Gradient(LN, 11726-7)
PGmean PGmean
MVA(VTI Card/CARDIAC_D_MVA_VTI/D/MVA(
Cardiovascular Orifice Area(SRT, G-038E)
) VTI)
MR Flow Card/CARDIAC_D_MVA_VTI/D/MR
Peak Instantaneous Flow Rate(LN, 34141-2)
Rate Flow Rate
Card/CARDIAC_D_MVA_VTI/D/MR
MR VTI Peak Systolic Velocity(LN, 11726-7)
VTI
Card/CARDIAC_D_MVA_VTI/D/MR
MR ERO Cardiovascular Orifice Area(SRT, G-038E)
ERO
MR Card/CARDIAC_D_MVA_VTI/D/MR
Volume Flow(LN, 33878-0)
Volume Volume
MR Card/CARDIAC_D_MVA_VTI/D/MR
Regurgitant Fraction(SRT, G-0390)
Fraction Fraction
Card/CARDIAC_D_MV_DECT/D/MV
MV DecT Deceleration Time(LN, 20217-6)
DecT
MR
Card/CARDIAC_D_MV/D/MR Vmax Peak Systolic Velocity(LN, 11726-7)
Vmax
Card/CARDIAC_D_MVA(VTI)/D/MV
MV VTI Maximum Velocity(LN, 20351-3)
VTI
MVA(PH Card/CARDIAC_D_MV_APHT/D/MVA(
Cardiovascular Orifice Area(SRT, G-038E)
T) PHT)
T) PHT)
MR Card/CARDIAC_D_MV_VMAX/D/MR
Maximum Velocity(LN, 11726-7)
Vmax Vmax
MR Card/CARDIAC_D_MV_VMAX/D/MR
Peak Gradient(LN, 11726-7)
PGmax PGmax
MR Card/CARDIAC_D_PISA_MR/D/MR
Maximum Velocity(LN, 11726-7)
Vmax Vmax
MR Card/CARDIAC_D_PISA_MR/D/MR
Peak Gradient(LN, 11726-7)
PGmax PGmax
MR Card/CARDIAC_D_PISA_MR/D/MR
Mean Velocity(LN, 11726-7)
Vmean Vmean
MR Card/CARDIAC_D_PISA_MR/D/MR
Mean Gradient(LN, 11726-7)
PGmean PGmean
Card/CARDIAC_D_PISA_MR/D/MR
MR VTI Velocity Time Integral(LN, 11726-7)
VTI
Card/CARDIAC_D_PISA_MR/D/MR
MR Rad Flow Radius(SONO, SONO-5202-21)
Rad
MR Flow Card/CARDIAC_D_PISA_MR/D/MR
Peak Instantaneous Flow Rate(LN, 34141-2)
Rate Flow Rate
Card/CARDIAC_D_PISA_MR/D/MR
MR ERO Cardiovascular Orifice Area(SRT, G-038E)
ERO
MR Card/CARDIAC_D_PISA_MR/D/MR
Volume Flow(LN, 33878-0)
Volume Volume
MR Card/CARDIAC_D_PISA_MR/D/MR
Regurgitant Fraction(SRT, G-0390)
Fraction Fraction
Card/CARDIAC_D_PISA_MS/D/MS
MS Rad Flow Radius(SONO, SONO-5202-21)
Rad
Card/CARDIAC_D_PISA_MS/D/MS
MS Vmax Maximum Velocity(LN, 20247-3)
Vmax
MS Flow Card/CARDIAC_D_PISA_MS/D/MS
Peak Instantaneous Flow Rate(LN, 34141-2)
Rate Flow Rate
Card/CARDIAC_D_PISA_MS/D/MS
MS ERO Cardiovascular Orifice Area(SRT, G-038E)
ERO
MS Card/CARDIAC_D_PISA_MS/D/MS
Volume Flow(LN, 33878-0)
Volume Volume
Card/CARDIAC_D_PISA_MS/D/MS
MS VTI Velocity Time Integral(LN, 20351-3)
VTI
Doppler Pulsed(SRT,R-409E4)
Card/CARDIAC_D_PV_PRVMAX/D/PR
PR Vmax Maximum Velocity(LN, 20351-3)
Vmax
Card/CARDIAC_D_PV_PVACCT/D/PV
PV AccT Acceleration Time(LN, 20168-1)
AccT
Card/CARDIAC_D_PV_PAEDP/D/PR
PR Ved End Diastolic Velocity(LN, 11653-3)
Ved
Card/CARDIAC_D_PV_PAEDP/D/PR
PR PGed Peak Gradient(LN, 20247-3)
PGed
PVA(Vma Card/CARDIAC_D_PV_PVAVMAX/D/P
Cardiovascular Orifice Area(SRT, G-038E)
x) VA(Vmax)
Card/CARDIAC_D_PV_PVAVMAX/D/P
PV Vmax Maximum Velocity(LN, 11726-7)
V Vmax
Card/CARDIAC_D_PV_PVAVTI/D/PV
PV Vmax Maximum Velocity(LN, 11726-7)
Vmax
PV Card/CARDIAC_D_PV_PVAVTI/D/PV
Peak Gradient(LN, 20247-3)
PGmax PGmax
Card/CARDIAC_D_PV_PVAVTI/D/PV
PV Vmean Mean Velocity(LN, 20352-1)
Vmean
PV Card/CARDIAC_D_PV_PVAVTI/D/PV
Mean Gradient(LN, 20256-4)
PGmean PGmean
Card/CARDIAC_D_PV_PVAVTI/D/PV
PV VTI Velocity Time Integral(LN, 20354-7)
VTI
Card/CARDIAC_D_PV_PVAVTI/D/PVA
PVA(VTI) Cardiovascular Orifice Area(SRT, G-038E)
(VTI)
PR
Card/CARDIAC_D_PR/D/PR PGmax Peak Gradient(LN, 20247-3)
PGmax
PR
Card/CARDIAC_D_PR/D/PR PGmean Mean Gradient(LN, 20256-4)
PGmean
PR Flow
Card/CARDIAC_D_PR/D/PR Flow Rate Peak Instantaneous Flow Rate(LN, 34141-2)
Rate
PR
Card/CARDIAC_D_PR/D/PR Volume Volume Flow(LN, 33878-0)
Volume
PR
Card/CARDIAC_D_PR/D/PR Fraction Regurgitant Fraction(SRT, G-0390)
Fraction
2D mode(SRT,G-03A2, SRT)
Card/CARDIAC_DIMENSION_TVDIA
TV Diam Cardiovascular Orifice Diameter(SRT, G-038F)
M/B/TV Diam
Card/CARDIAC_DIMENSION_TVDIA
TV Diam Cardiovascular Orifice Diameter(SRT, G-038F)
M/C/TV Diam
Doppler Pulsed(SRT,R-409E4)
TV Vmax Card/CARDIAC_D_TV_VMAX/D/TV
Maximum Velocity(LN, 11726-7)
Vmax
TV Card/CARDIAC_D_TV_VMAX/D/TV
Peak Gradient(LN, 20247-3)
PGmax PGmax
TV Vmax Card/CARDIAC_D_TV_VTI/D/TV
Maximum Velocity(LN, 11726-7)
Vmax
TV Card/CARDIAC_D_TV_VTI/D/TV
Peak Gradient(LN, 20247-3)
PGmax PGmax
TV Card/CARDIAC_D_TV_VTI/D/TV
Mean Velocity(LN, 20352-1)
Vmean Vmean
TV Card/CARDIAC_D_TV_VTI/D/TV
Mean Gradient(LN, 20256-4)
PGmean PGmean
TV C-O Card/CARDIAC_D_TV_RV_TEI/D/TV
Tricuspid Valve Closure to Opening Time(SRT, G-0389)
Dur(a) C-O Dur(a)
Card/CARDIAC_D_RVSP_TRVMAX/D
TR Vmax Maximum Velocity(LN, 11726-7)
/TR Vmax
TR Card/CARDIAC_D_RVSP_TRVMAX/D
Peak Gradient(LN, 20247-3)
PGmax /TR PGmax
TR
Card/CARDIAC_D_TR/D/TR PGmax Peak Gradient(LN, 20247-3)
PGmax
TR
Card/CARDIAC_D_TR/D/TR Vmean Mean Velocity(LN, 20352-1)
Vmean
TR
Card/CARDIAC_D_TR/D/TR PGmean Mean Gradient(LN, 20256-4)
PGmean
TR
Card/CARDIAC_D_TR/D/TR Rad(2D) Flow Radius(SONO, SONO-5202-21)
Rad(2D)
TR Flow
Card/CARDIAC_D_TR/D/TR Flow Rate Peak Instantaneous Flow Rate(LN, 34141-2)
Rate
TR
Card/CARDIAC_D_TR/D/TR Volume Volume Flow(LN, 33878-0)
Volume
TR
Card/CARDIAC_D_TR/D/TR Fraction Regurgitant Fraction(SRT, G-0390)
Fraction
M mode M mode(SRT,G-0394)
Card/CARDIAC_DIMENSION_M
MPA Main Pulmonary Artery Diameter(LN, 18020-8)
PA/B/MPA
Card/CARDIAC_DIMENSION_P
PV Diam Cardiovascular Orifice Diameter(LN, G-038F)
VDIAM/B/PV Diam
Card/CARDIAC_DIMENSION_M
MPA Main Pulmonary Artery Diameter(LN, 18020-8)
PA/C/MPA
Card/CARDIAC_DIMENSION_P
PV Diam Cardiovascular Orifice Diameter(LN, G-038F)
VDIAM/C/PV Diam
MPA Card/CARDIAC_D_PV_MPAVM
Main Pulmonary Artery Peak Velocity(SRT, G-038A)
Vmax AX/D/MPA Vmax
LPA Card/CARDIAC_D_PV_LPAVMA
Left Pulmonary Artery Peak Velocity(SONO, SONO-5202-28)
Vmax X/D/LPA Vmax
Card/CARDIAC_D_PV_PAEDP/
PAEDP Pulmonary Artery Pressure(SRT, F0212C)
D/PAEDP
Card/CARDIAC_D_PLUM_SVEL/D/
Pulm S Vel Pulmonary Vein Systolic Peak Velocity(LN, 29450-4)
Pulm S Vel
Card/CARDIAC_D_PLUM_SVTI/D/P
Pulm S VTI Pulmonary Vein S-Wave Velocity Time Integral(SRT, G-038C)
ulm S VTI
Card/CARDIAC_D_PLUM_DVEL/D/
Pulm D Vel Pulmonary Vein Diastolic Peak Velocity(LN, 29451-2)
Pulm D Vel
Card/CARDIAC_D_PLUM_DVTI/D/P
Pulm D VTI Pulmonary Vein D-Wave Velocity Time Integral(SRT, G-038D)
ulm D VTI
Card/CARDIAC_D_PLUM_ADUR/D/
Pulm A Dur Pulmonary Vein A-Wave Duration(SRT, G-038B)
Pulm A Dur
Card/CARDIAC_D_PLUM_DECT/D/
Pulm DecT Deceleration Time(LN, 20217-6)
Pulm DecT
Card/CARDIAC_D_MV/D/E/E'(Late
ral)
Card/CARDIAC_D_TDI_LATERAL/
D/E/E'(Lateral)
Card/CARDIAC_D_TDI_LATERAL/
Aa(Lateral) Mitral Valve DTI Ea to Aa Ratio(SONO, SONO-5201-32)
D/Ea/Aa(Lateral)
A.4.14 Left Ventricle Outflow Tract DCID (12223) Echocardiography Stroke Volume Origin
Label Item ID Code Meaning (Coding Scheme Designator, Code Value)
Card/CARDIAC_D_MVA_VTI/D/LV
LVOT Vmax Maximum Velocity(LN, 11726-7)
OT Vmax
Card/CARDIAC_D_MVA_VTI/D/LV
LVOT PGmax Peak Gradient(LN, 11726-7)
OT PGmax
Card/CARDIAC_D_MVA_VTI/D/LV
LVOT Vmean Mean Velocity(LN, 11726-7)
OT Vmean
LVOT Card/CARDIAC_D_MVA_VTI/D/LV
Mean Gradient(LN, 11726-7)
PGmean OT PGmean
Card/CARDIAC_D_MVA_VTI/D/LV
LVOT VTI Velocity Time Integral(LN, 11726-7)
OT VTI
Card/CARDIAC_D_MVA(VTI)/D/LV
LVOT VTI Peak Systolic Velocity(LN, 11726-7)
OT VTI
Card/CARDIAC_D_AVA_VTI/D/LV
LVOT Vmax Maximum Velocity(LN, 11726-7)
OT Vmax
Card/CARDIAC_D_AVA_VTI/D/LV
LVOT PGmax Peak Gradient(LN, 20247-3)
OT PGmax
Card/CARDIAC_D_AVA_VTI/D/LV
LVOT Vmean Mean Velocity(LN, 20352-1)
OT Vmean
LVOT Card/CARDIAC_D_AVA_VTI/D/LV
Mean Gradient(LN, 20256-4)
PGmean OT PGmean
Card/CARDIAC_D_AVA_VTI/D/LV
LVOT VTI Velocity Time Integral(LN, 20354-7)
OT VTI
Card/CARDIAC_D_AVA_VTI/D/LV
LVOT(2D) Cardiovascular Orifice Diameter(SRT, G-038F)
OT(2D)
Card/CARDIAC_D_AVA_VTI/D/SV(
SV(LVOT) Stroke Volume(SRT, F-32120)
LVOT)
Card/CARDIAC_D_AVA_VTI/D/CO
CO(LVOT) Cardiac Output(SRT, F-32100)
(LVOT)
Card/CARDIAC_D_AVA_VMAX/D/
LVOT(2D) Cardiovascular Orifice Diameter(SRT, G-038F)
LVOT(2D)
Card/CARDIAC_D_AVA_VMAX/D/
AVA(Vmax) Cardiovascular Orifice Area Area(SRT, G-038E)
AVA(Vmax)
Card/CARDIAC_D_AVA_VMAX/D/
LVOT Vmax Peak Systolic Velocity(LN, 11726-7)
LVOT Vmax
A.4.15 Right Ventricle Outflow Tract DCID (12223) Echocardiography Stroke Volume Origin
Label Item ID Code Meaning (Coding Scheme Designator, Code Value)
2D mode(SRT,G-03A2, SRT)
Card/CARDIAC_DIMENSION_RVO
RVOT Cardiovascular Orifice Diameter(SRT, G-038F)
T/B/RVOT
Card/CARDIAC_DIMENSION_RVO
RVOT Cardiac Output(SRT, F-32100)
T/C/ RVOT
M mode(SRT,G-0394)
Card/CARDIAC_M_DIMENSION_R
RVOT Cardiovascular Orifice Diameter(SRT, G-038F)
VOT/M/RVOT
Doppler Pulsed(SRT,R-409E4)
Card/CARDIAC_D_PV_PVAVMAX/
RVOT(2D) Cardiovascular Orifice Diameter(SRT, G-038F)
D/RVOT(2D)
Card/CARDIAC_D_PV_PVAVMAX/
RVOT VMax Maximum Velocity(LN, 11726-7)
D/RVOT VMax
Card/CARDIAC_D_PV_PVAVTI/D/
RVOT(2D) Cardiovascular Orifice Diameter(SRT, G-038F)
RVOT(2D)
Card/CARDIAC_D_PV_PVAVTI/D/
RVOT Vmax Maximum Velocity(LN, 11726-7)
RVOT Vmax
Card/CARDIAC_D_PV_PVAVTI/D/
RVOT PGmax Peak Gradient(LN, 20247-3)
RVOT PGmax
Card/CARDIAC_D_PV_PVAVTI/D/
RVOT Vmean Mean Velocity(LN, 20352-1)
RVOT Vmean
RVOT Card/CARDIAC_D_PV_PVAVTI/D/
Mean Gradient(LN, 20256-4)
PGmean RVOT PGmean
Card/CARDIAC_D_PV_PVAVTI/D/
RVOT VTI Velocity Time Integral(LN, 20354-7)
RVOT VTI
A.4.16 Medial Mitral Annulus DCID (12235) Mitral Valve Anatomic Sites
Label Item ID Code Meaning (Coding Scheme Designator, Code Value)
Card/CARDIAC_D_TDI_MEDIA
Sa Medial Left Ventricular Peak Systolic Tissue Velocity(SRT, G-037D)
L/D/Sa Medial
Card/CARDIAC_D_TDI_MEDIA
Ea Medial Left Ventricular Peak Early Diastolic Tissue Velocity(SRT, G-037A)
L/D/Ea Medial
Card/CARDIAC_D_TDI_MEDIA
Aa(Medial) Mitral Valve DTI Ea to Aa Ratio(SONO, SONO-5201-31)
L/D/Ea/Aa(Medial)
Card/CARDIAC_D_TDI_LATER
Sa Lateral Left Ventricular Peak Systolic Tissue Velocity(SRT, G-037D)
AL/D/Sa Lateral
Card/CARDIAC_D_HEP_SVEL/
S Vel Hepatic Vein Systolic Peak Velocity(LN, 29471-0)
D/Hep S Vel
Card/CARDIAC_D_HEP_DVEL
D Vel Hepatic Vein Diastolic Peak Velocity(LN, 29472-8)
/D/Hep D Vel
Card/CARDIAC_D_HEP_AVEL/
A Vel Hepatic Vein Atrial Contraction Reversal Peak Velocity(LN, 29474-4)
D/Hep A Vel
Card/CARDIAC_D_HEP_ADUR
A Dur Hepatic Vein A-Wave Duration(SONO, SONO-5202-27)
/D/Hep A Dur