0% found this document useful (0 votes)
46 views145 pages

DICOM Conformance Statement P20

Manual p20

Uploaded by

gerardomartin
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
46 views145 pages

DICOM Conformance Statement P20

Manual p20

Uploaded by

gerardomartin
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 145

SONOSCAPE MEDICAL CORP DICOM 3.

0 CONFORMANCE STATEMENT

DICOM 3.0 CONFORMANCE STATEMENT

Ultrasound Systems (S40, S40Exp, S30, S12, S22, S9, S9Pro, S8Exp, S50, S50Exp, S50Pro, S45, S45Exp, S45Pro, P50,
P50Exp, P50Pro, P20, P20 Pro, P15, P22, P25, P10 Plus, P30VO, P40VO)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X

2017 SONOSCAPE MEDICAL CORP

All Rights Reserved

Yizhe Building, Yuquan Road, Shenzhen, China

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 1 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

REVISION HISTORY

Document Date of Issue Author Description


Version
1.0 2013-1-31 zhoudh Version for Final Text
2.0 2014-5-31 zhoudh Add DICOM MPPS print
3.0 2017-1-7 yuql Add Structured Report
4.0 2017-5-18 zhoudh Add P50 Ultrasound Systems
5.0 2018-9-1288 zhoudh Add P20 Ultrasound Systems

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 2 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

1 CONFORMANCE STATEMENT OVERVIEW

The Ultrasound systems implements the necessary DICOM services to download work lists from an information system,
save acquired US images to a network storage device or media, print to a networked hardcopy device and inform the
information system about the work actually done.

Table 1-1 provides an overview of the supported network services.

Table 1-1
NETWORK SERVICES
SOP Classes User of Provider of
Service (SCU) Service (SCP)
Transfer
Secondary Capture Image Storage Yes No
Ultrasound Image Storage Yes No
Ultrasound Multi-frame Image Storage Yes No
Comprehensive SR Yes No
Workflow Management
Modality Worklist Yes No
Storage Commitment Push Model Yes No
Modality Performed Procedure Step Yes No
Print Management
Basic Grayscale Print Management Yes No
Basic Color Print Management Yes No

Table 1-2 provides an overview of the Media Storage Application Profiles supported.

Table 1-2
MEDIA SERVICES
Media Storage Application Profile Write Files (FSC or FSU) Read Files (FSR)

Compact Disk - Recordable

STD-US-ID-SF&MF-CDR Yes / Yes No


STD-US-SC-SF&MF-CDR

DVD

STD-US-ID-SF&MF-DVD Yes / Yes No


STD-US-SC-SF&MF-DVD

USB

STD-GEN-USB Yes / Yes No

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 3 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

2 TABLE OF CONTENTS

Revision History ........................................................................................................................................................ 2

1 CONFORMANCE STATEMENT OVERVIEW ........................................................................................................... 3

2 TABLE OF CONTENTS............................................................................................................................................ 4

3 INTRODUCTION ...................................................................................................................................................... 8

3.1 AUDIENCE ................................................................................................................................................ 8

3.2 REMARKS ................................................................................................................................................. 8

3.3 TERMS and Definitions .............................................................................................................................. 8

3.4 Abbreviations ............................................................................................................................................. 9

3.5 REFERENCES ........................................................................................................................................ 10

4 NETWORKING ...................................................................................................................................................... 11

4.1 IMPLEMENTATION MODEL.................................................................................................................... 11

4.1.1 Application Data Flow ..................................................................................................................... 11

4.1.2 Functional Definition of AE’s ........................................................................................................... 12

4.1.3 Sequencing of Real-World Activities ............................................................................................... 13

4.2 AE SPECIFICATIONS ............................................................................................................................. 14

4.2.1 Storage Application Entity Specification .......................................................................................... 14

4.2.2 Workflow Application Entity Specification ........................................................................................ 23

4.2.3 Hardcopy Application Entity Specification ....................................................................................... 33

4.2.4 Verification Application Entity specification...................................................................................... 45

4.3 NETWORK INTERFACES ....................................................................................................................... 48

4.3.1 TCP/IP Stack Supported ................................................................................................................. 48

4.3.2 Physical Network Interface ............................................................................................................. 48

4.3.3 IPv4 and IPv6 Support .................................................................................................................... 48

4.4 CONFIGURATION ................................................................................................................................... 49

4.4.1 Local .............................................................................................................................................. 49


Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 4 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

4.4.2 Remote .......................................................................................................................................... 49

5 MEDIA INTERCHANGE ......................................................................................................................................... 50

5.1 IMPLEMENTATION MODEL.................................................................................................................... 50

5.1.1 Application Data Flow ..................................................................................................................... 50

5.1.2 Functional Definition of AE’s ........................................................................................................... 50

5.1.3 Sequencing of Real-World Activities ............................................................................................... 50

5.1.4 File Meta Information Options ......................................................................................................... 50

5.2 AE SPECIFICATIONS ............................................................................................................................. 51

5.2.1 Media Application Entity Specification ............................................................................................. 51

6 SUPPORT OF CHARACTER SETS ....................................................................................................................... 53

7 SECURTIY ............................................................................................................................................................. 54

8 ANNEXES .............................................................................................................................................................. 55

8.1 IOD CONTENTS...................................................................................................................................... 55

8.1.1 Created SOP Instances .................................................................................................................. 55

8.1.2 Used Fields in received IOD by application ..................................................................................... 75

8.1.3 Attribute mapping ........................................................................................................................... 75

8.1.4 Coerced/Modified Fields ................................................................................................................. 76

8.2 CodED TERMINOLOGY and templates ................................................................................................... 77

8.4 GRAYSCALE IMAGE CONSISTENCY .................................................................................................... 77

8.5 standard EXTENSded / SPECIALIZed / PRIVATe SOP Classes .............................................................. 77

8.5.1 Ultrasound Image Storage SOP Class ............................................................................................ 77

8.6 PRIVATE TRANSFER SYNTAXES .......................................................................................................... 77

APPENDIX A – Structured Reports ................................................................................................................................ 77

A.1 STRUCTURED REPORTS ............................................................................................................................ 77

A.1.1 Introduction ................................................................................................................................................... 77

A.2 TID 5000 OB-GYN ULTRASOUND PROCEDURE REPORT .......................................................................... 78

A.2. 1 TID 5001 OB-GYN Patient Characteristics ............................................................................................. 78

A.2. 2 TID 5002 OB-GYN Procedure Summary Section ................................................................................... 78

A.2. 3 TID 5003 OB-GYN Fetus Summary....................................................................................................... 78

A.2. 4 TID 5004 Fetal Biometry Ratio Section .................................................................................................. 79


Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 5 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

A.2. 5 TID 5005 Fetal Biometry Section ........................................................................................................... 79

A.2. 6 TID 5006 Fetal Long Bones Section ...................................................................................................... 85

A.2. 7 TID 5007 Fetal Cranium Section ........................................................................................................... 86

A.2. 9 TID 5010 Amniotic Sac Section .............................................................................................................. 87

A.2. 10 TID 5011 Early Gestation Section ........................................................................................................ 87

A.2. 11 TID 5012 Ovaries Section ................................................................................................................... 89

A.2. 13 TID 5014 Follicle Measurement Section ............................................................................................... 89

A.2. 14 TID 5015 Pelvis and Uterus Section ..................................................................................................... 91

A.2. 15 TID 5025 OB-GYN Fetal Vascular Ultrasound Measurement Group .................................................... 91

A.2. 16 TID 5026 OB-GYN Pelvic Vasculatrue Section Measurement Group .................................................. 102

A.3 TID 5100 Vascular Ultrasound Report ........................................................................................................ 116

A.3 .1 TID 5101 Vascular Patient Characteristics ................................................................................ 116

A.3 .2 TID 5102 Vascular Procedure Summary Section ...................................................................... 117

A.3 .3 Anatomy Measure Item ............................................................................................................ 117

A.3 .4 Blood Vessel of Head DCID (12105) Intracranial Cerebral Vessels DCID (12106) Unpaired
Intracranial Cerebral Vessels (unilateral) ........................................................................................................ 117

A.3.5 Artery of Neck DCID (12104) Extracranial Arteries ................................................................................ 118

A.3.6 Artery of Lower Extremity DCID (12109) Lower Extremity Arteries ......................................................... 119

A.3. 7 Vein of Lower Extremity DCID (12110) Lower Extremity Veins ............................................................. 119

A.3. 8 Artery of Upper Extremity DCID (12107) Upper Extremity Arteries ........................................................ 120

A.3. 9 Vein of Upper Extremity DCID (12108) Upper Extremity Veins ............................................................. 121

A.3. 10 Vascular Structure of Kidney DCID (12115) Renal Vessels ................................................................ 121

A.3. 11 Artery of Abdomen DCID (12111) Abdominal Arteries (lateral) DCID (12112) Abdominal Arteries (unilateral)
...................................................................................................................................................................... 122

A.3. 12 Vein of Abdomen DCID (12113) Abdominal Veins (lateral) DCID (12114) Abdominal Veins (unilateral)
...................................................................................................................................................................... 122

A.4 TID 5200 Echocardiography Procedure Report ......................................................................................... 123

A.4.1 TID 5201 Echocardiography Patient Characteristics .............................................................................. 123

A.4.2 Left Ventricle DCID (12200) Echocardiography Left Ventricle ................................................................ 123

A.4.3 Right Ventricle DCID (12204) Echocardiography Right Ventricle ........................................................... 129

A.4.4 Left Atrium DCID (12205) Echocardiography Left Atrium ....................................................................... 130

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 6 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

A.4.5 Right Atrium DCID (12206) Echocardiography Right Atrium .................................................................. 131

A.4.6 Aortic Valve DCID(12211) Echocardiography Aortic Valve .................................................................... 132

A.4.7 Mitral Valve DCID(12207) Echocardiography Mitral Valve .................................................................... 134

A.4.8 Pulmonic Valve DCID(12209) Echocardiography Pulmonic Valve.......................................................... 138

A.4.9 Tricuspid Valve DCID(12208) Echocardiography Tricuspid Valve .......................................................... 139

A.4.10 Aorta DCID (12212) Echocardiography Aorta ...................................................................................... 141

A.4.11 Pulmonary artery DCID (12210) Echocardiography Pulmonary Artery ................................................. 141

A.4.12 Pulmonary Venous Structure DCID (12214) Echocardiography Pulmonary Veins ................................ 142

A.4.13 Lateral Mitral Annulus (12235) Echocardiography Cardiac Shunt ........................................................ 142

A.4.14 Left Ventricle Outflow Tract DCID (12223) Echocardiography Stroke Volume Origin ............................ 143

A.4.15 Right Ventricle Outflow Tract DCID (12223) Echocardiography Stroke Volume Origin ......................... 144

A.4.16 Medial Mitral Annulus DCID (12235) Mitral Valve Anatomic Sites ........................................................ 144

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 7 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

3 INTRODUCTION

3.1 AUDIENCE
This document is intended for hospital staff, health care system integrators, software designers or implementers. It is
assumed that the reader has a working understanding of DICOM.

3.2 REMARKS
DICOM, by itself, does not guarantee interoperability. However, the Conformance Statement facilitates a first- level
validation for interoperability between different applications supporting the same DICOM functionality.

This Conformance Statement is not intended to replace validation with other DICOM equipment to ensure proper
exchange of information intended.

The scope of this Conformance Statement is to facilitate communication between Philips Medical Systems and other
vendors’ Medical equipment. The Conformance Statement should be read and understood in conjunction with the DICOM
Standard [DICOM]. However, by itself it is not guaranteed to ensure the desired interoperability and successful
interconnectivity.

3.3 TERMS AND DEFINITIONS


Terms and definitions are listed here. Informal definitions are provided for the following terms used in this Conformance
Statement. The DICOM Standard is the authoritative source for formal definitions of these terms.

Abstract Syntax – the information agreed to be exchanged between applications, generally equivalent to a
Service/Object Pair (SOP) Class. Examples:Verification SOP Class, Modality Worklist Information Model Find
SOP Class, Computed Radiography Image Storage SOP Class.

Application Entity (AE) – an end point of a DICOM information exchange, including the DICOM network or media
interface software; i.e., the software that sends or receives DICOM information objects or messages. A single
device may have multiple Application Entities.

Application Entity Title – the externally known name of an Application Entity, used to identify a DICOM application
to other DICOM applications on the network.

Application Context – the specification of the type of communication used between Application Entities. Example:
DICOM network protocol.

Association – a network communication channel set up between Application Entities.

Attribute – – a unit of information in an object definition; a data element identified by a tag. The information may be
a complex data structure (Sequence), itself composed of lower level data elements. Examples: Patient ID
(0010,0020), Accession Number (0008,0050), Photometric Interpretation (0028,0004), Procedure Code Sequence
(0008,1032).

Information Object Definition (IOD) – the specified set of Attributes that comprise a type of data object; does not
represent a specific instance of the data object, but rather a class of similar data objects that have the same
properties. The Attributes may be specified as Mandatory (Type 1), Required but possibly unknown (Type 2), or
Optional (Type 3), and there may be conditions associated with the use of an Attribute (Types 1C and 2C).
Examples: US Image IOD, CT Image IOD, Print Job IOD.

Module – a set of Attributes within an Information Object Definition that are logically related to each other. Example:
Patient Module includes Patient Name, Patient ID, Patient Birth Date, and Patient Sex.
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 8 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Negotiation – first phase of Association establishment that allows Application Entities to agree on the types of data
to be exchanged and how that data will be encoded.

Presentation Context – the set of DICOM network services used over an Association, as negotiated between
Application Entities; includes Abstract Syntaxes and Transfer Syntaxes.

Protocol Data Unit (PDU) – a packet (piece) of a DICOM message sent across the network. Devices must
specify the maximum size packet they can receive for DICOM messages.

Service Class Provider (SCP) – role of an Application Entity that provides a DICOM network service; typically, a
server that performs operations requested by another Application Entity (Service Class User). Examples: Picture
Archiving and Communication System (image storage SCP, and image query/retrieve SCP), Radiology Information
System (modality worklist SCP).

Service Class User (SCU) – role of an Application Entity that uses a DICOM network service; typically, a client.
Examples: imaging modality (image storage SCU, and modality worklist SCU), imaging workstation (image
query/retrieve SCU)

Service/Object Pair (SOP) Class – the specification of the network or media transfer (service) of a particular type
of data (object); the fundamental unit of DICOM interoperability specification. Examples: Ultrasound Image Storage
Service, Basic Grayscale Print Management.

Service/Object Pair (SOP) Instance – an information object; a specific occurrence of information exchanged in a
SOP Class. Examples: a specific x-ray image.

Tag – a 32-bit identifier for a data element, represented as a pair of four digit hexadecimal numbers, the “group” and
the “element”. If the “group” number is odd, the tag is for a private (manufacturer-specific) data element. Examples:
(0010,0020) [Patient ID], (07FE,0010) [Pixel Data], (0019,0210) [private data element]

Transfer Syntax – the encoding used for exchange of DICOM information objects and messages. Examples:
JPEG compressed (images), little endian explicit value representation.

Unique Identifier (UID) – a globally unique “dotted decimal” string that identifies a specific object or a class of
objects; an ISO-8824 Object Identifier. Examples: Study Instance UID, SOP Class UID, SOP Instance UID.

Value Representation (VR) – the format type of an individual DICOM data element, such as text, an integer, a
person’s name, or a code. DICOM information objects can be transmitted with either explicit identification of the
type of each data element (Explicit VR), or without explicit identification (Implicit VR); with Implicit VR, the receiving
application must use a DICOM data dictionary to look up the format of each data element.

3.4 ABBREVIATIONS

Abbreviations of the terms that are used have been listed here.

AE Application Entity
AET Application Entity Title
CD-R Compact Disk Recordable
CSE Customer Service Engineer
DICOM Digital Imaging and Communications in Medicine
FSC File-Set Creator
FSU File-Set Updater
FSR File-Set Reader
HIS Hospital Information System
IOD Information Object Definition
IPv4 Internet Protocol version 4
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 9 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

IPv6 Internet Protocol version 6


ISO International Organization for Standards
JPEG Joint Photographic Experts Group
LOINC Logical Observation Identifiers Names and Codes
LUT Look-up Table
MPPS Modality Performed Procedure Step
MSPS Modality Scheduled Procedure Step
MWL Modality Worklist
O Optional (Key Attribute)
OSI Open Systems Interconnection
PDU Protocol Data Unit
R Required (Key Attribute)
RIS Radiology Information System
SC Secondary Capture
SCP Service Class Provider
SCU Service Class User
SOP Service-Object Pair
TCP/IP Transmission Control Protocol/Internet Protocol
U Unique (Key Attribute)
UL Upper Layer
US Ultrasound
VL Visible Light
VR Value Representation

3.5 REFERENCES
Referenced documents are listed below.

 NEMA PS3 Digital Imaging and Communications in Medicine (DICOM) Standard, available free at
http://medical.nema.org/
 Basic operator’s manual of the product
 Service Manual of the product

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 10 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

4 NETWORKING

4.1 IMPLEMENTATION MODEL


4.1.1 Application Data Flow

Remote Application
Send Storage
Entity Receives
US Images & Application
Images & SR, R e tu rn
SR, Commits Entity
s C o m m itm e n ts

Remote
Application
Update Entity Provides
Worklist Worklist Items

Workflow
Application
Entity
Acquire
Remote Application
US Images
Entity Receives
MPPS
Create/Update

Remote
Film Hardcopy Application Entity
US Images Application Prints Film Sheets
Entity

DICOM Standard Interface

Figure 4.1-1
APPLICATION DATA FLOW DIAGRAM

— The Storage Application Entity sends images to a remote AE. It is associated with the local real-world activity “Send
US Images & SR”. “Send US Images & SR” is performed upon user request for each study completed or for specific
images selected. When activated by user’s settings (auto-send), each marked set of images can be immediately
stored to a preferred destination whenever a Patient/Study is closed by the user. If the remote AE is configured as
an archive device the Storage AE will request Storage Commitment and if a commitment is successfully obtained will
record this information in the local database.
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 11 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

— The Workflow Application Entity receives Worklist information from and sends MPPS information to a remote AE. It is
associated with the local real-world activities “Update Worklist” and “Acquire Images”. When the “Update Worklist”
local real-world activity is performed the Workflow Application Entity queries a remote AE for worklist items and
provides the set of worklist items matching the query request. ”Update Worklist” is performed as a result of an
operator request or can be performed automatically at specific time intervals. When the “Acquire Images” local
real-world activity is performed the Workflow Application Entity creates and updates Modality Performed Procedure
Step instances managed by a remote AE. Acquisition of images will result in automated creation of an MPPS
Instance. Completion of the MPPS is performed as the result of an operator action.

— The Hardcopy Application Entity prints images on a remote AE (Printer). It is associated with the local real-world
activity “Film US Images”. “Film US Images” creates a print-job within the print queue containing one or more virtual
film sheets composed from images selected by the user.

4.1.2 Functional Definition of AE’s


4.1.2.1 Functional Definition of Storage Application Entity
The existence of a send-job queue entry with associated network destination will activate the Storage AE. An association
request is sent to the destination AE and upon successful negotiation of a Presentation Context the image transfer is started.
If the association cannot be opened, the related send-job is set to an error state and can be restarted by the user via job
control interface. By default, the Storage AE will not try to initiate another association for this send-job automatically.
However, an automatic retry (retry-timer, retrycount) can be configured by a CSE.

4.1.2.2 Functional Definition of Workflow Application Entity


Worklist Update attempts to download a Worklist from a remote node. If the Workflow AE establishes an Association to a
remote AE, it will transfer all worklist items via the open Association. During receiving the worklist response items are
counted and the query processing is canceled if the configurable limit of items is reached. The results will be displayed in a
separate list, which will be cleared with the next Worklist Update.

The Workflow AE performs the creation of a MPPS Instance automatically whenever images are acquired. Further
updates on the MPPS data can be performed interactively from the related MPPS user interface. The MPPS “Complete” or
“Discontinued” states can only be set from the user interface.

4.1.2.3 Functional Definition of Hardcopy Application Entity


The existence of a print-job in the print queue will activate the Hardcopy AE. An association is established with the printer
and the printer’s status determined. If the printer is operating normally, the film sheets described within the print-job will be
printed. Changes in printer status will be detected (e.g. out of film) and reported to the user. If the printer is not operating
normally, the print-job will set to an error state and can be restarted by the user via the job control interface.

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 12 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

4.1.3 Sequencing of Real-World Activities

Storage Hardcopy Workflow Department Printer Image


Scheduler Manager
Manager

1. Query Worklist

2. Receive Worklist

3. Select Workitem (MSPS)

4. Start Acquisition (Create MPPS)

5. Acquire US Images

6. Complete Acquisition (Finalize MPPS)

7. Print Acquired Images

8. Store Acquired US Images & SR

9. Commit Acquired US Images & SR

Figure 4.1-2
SEQUENCING CONSTRAINTS

Under normal scheduled workflow conditions the sequencing constraints illustrated in Figure 4.1-2 apply:

1. Query Worklist

2. Receive Worklist of Modality Scheduled Procedure Steps (MSPS)

3. Select Workitem (MSPS) from Worklist

4. Start acquisition and create MPPS

5. Acquire US Images

6. Complete acquisition and finalize MPPS

7. Print acquired images (optional step)

8. Store acquired US images and SR.

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 13 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

9. If the Image Manager is configured as an archive device the Storage AE will request Storage Commitment for the
images and SR.

Other workflow situations (e.g. unscheduled procedure steps) will have other sequencing constraints. Printing could
equally take place after the acquired images have been stored. Printing could be omitted completely if no printer is
connected or hardcopies are not required.

4.2 AE SPECIFICATIONS
4.2.1 Storage Application Entity Specification
4.2.1.1 SOP Classes
The system provides Standard Conformance to the following SOP Classes:

Table 4.2-1
SOP CLASSES FOR AE STORAGE
SOP Class Name SOP Class UID SCU SCP
Ultrasound Image 1.2.840.10008.5.1.4.1.1.6.1 Yes No
Storage
Ultrasound 1.2.840.10008.5.1.4.1.1.3.1 Yes No
Multi-frame Image
Storage
Secondary Capture 1.2.840.10008.5.1.4.1.1.7 Yes No
Image Storage
Grayscale Softcopy 1.2.840.10008.5.1.4.1.1.88.33 Yes No
Presentation State
Storage
Storage 1.2.840.10008.1.20.1 Yes No
Commitment Push
Model
Verification 1.2.840.10008.1.1 Yes Yes

4.2.1.2 Association Policies


4.2.1.2.1 General
The DICOM standard application context name for DICOM 3.0 is always proposed:

Table 4.2-2
DICOM APPLICATION CONTEXT FOR AE STORAGE
Application Context Name 1.2.840.10008.3.1.1.1

4.2.1.2.2 Number of Associations


The system initiates one Association at a time for each destination to which a transfer request is being processed in the
active job queue list. Only one job will be active at a time, the other remains pending until the active job is completed or
failed.

Table 4.2-3
NUMBER OF ASSOCIATIONS INITIATED FOR AE STORAGE
Maximum number of simultaneous Associations 1
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 14 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

The system accepts Associations to receive N-EVENT-REPORT notifications for the Storage Commitment Push Model
SOP Class.

Table 4.2-4
NUMBER OF ASSOCIATIONS ACCEPTED FOR AE STORAGE
Maximum number of simultaneous Associations 1
4.2.1.2.3 Asynchronous Nature
The system does not support asynchronous communication (multiple outstanding transactions over a single Association).

Table 4.2-5
ASYNCHRONOUS NATURE AS A SCU FOR AE STORAGE
Maximum number of outstanding asynchronous 1
transactions

4.2.1.2.4 Implementation Identifying Information


The implementation information for this Application Entity is:

Table 4.2-6
DICOM IMPLEMENTATION CLASS AND VERSION FOR AE STORAGE
Implementation Class UID 1.2.156.112601
Implementation Version Name DCMTK 3.6.0

4.2.1.3 Association Initiation Policy


4.2.1.3.1 Activity – Send Images
4.2.1.3.1.1 Description and Sequencing of Activities
A user can select images and presentation states and request them to be sent to multiple destinations (up to 3). Each
request is forwarded to the job queue and processed individually. When the “Auto-send” option is active, each marked
instance or marked set of instances stored in database will be forwarded to the network job queue for a pre-configured
auto-send target destination. It can be configured which instances will be automatically marked and the destination where
the instances are automatically sent to. The “Auto-send” is triggered by the Close Patient user application.

The Storage AE is invoked by the job control interface that is responsible for processing network archival tasks. The job
consists of data describing the instances marked for storage and the destination. An internal daemon process triggered by
a job for a specific network destination initiates a C-STORE request to store images. If the process successfully
establishes an Association to a remote Application Entity, it will transfer each marked instance one after another via the
open Association. Status of the transfer is reported through the job control interface. Only one job will be active at a time. If
the C-STORE Response from the remote Application contains a status other than Success or Warning, the Association is
aborted and the related Job is switched to a failed state. It can be restarted any time by user interaction or, if configured, by
automated retry.

The Storage AE attempts to initiate a new Association in order to issue a C-STORE request. If the job contains multiple
images then multiple C-STORE requests will be issued over the same Association.

If the Remote AE is configured as an archive device the Storage AE will, after all images and presentation states have been
sent, transmit a single Storage Commitment request (N-ACTION) over the same Association. Upon receiving the
N-ACTION response the Storage AE will delay releasing the Association for a configurable amount of time. If no
N-EVENT-REPORT is received within this time period the Association will be immediately released (i.e. notification of
Storage Commitment success or failure will be received over a separate association). However, the Storage AE is capable

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 15 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

of receiving an N-EVENT-REPORT request at any time during an association provided a Presentation Context for the
Storage Commitment Push Model has been successfully negotiated (i.e. the N-ACTION is sent at the end of one
association and the N-EVENT-REPORT is received during an association initiated for a subsequent send job or during an
association initiated by the Remote AE for the specific purpose of sending the N-EVENT-REPORT).

Storage Image
AE Manager
Manager

1. Open Association

2. C-STORE (US Image)

3. C-STORE (SR)

4. C-STORE (US Image)

5. C-STORE (SR)

6. N-ACTION (Storage Commitment Request for Images & SR)

7. N-EVENT-REPORT (Storage Commitment Response)

8. Close Association

Figure 4.2-1
SEQUENCING OF ACTIVITY – SEND IMAGES

A possible sequence of interactions between the Storage AE and an Image Manager (e.g. a storage or archive device
supporting the Storage and Storage Commitment SOP Classes as an SCP) is illustrated in Figure 4.2-1:

1. The Storage AE opens an association with the Image Manager

2. An acquired US image is transmitted to the Image Manager using a C-STORE request and the Image Manager
replies with a C-STORE response (status success).

3. A SR is transmitted to the Image Manager using a C-STORE request and the Image Manager replies with a
C-STORE response (status success).

4. Another acquired US image is transmitted to the Image Manager using a C-STORE request and the Image
Manager replies with a C-STORE response (status success).

5. Another SR is transmitted to the Image Manager using a C-STORE request and the Image Manager replies with a
C-STORE response (status success).

6. An N-ACTION request is transmitted to the Image Manager to obtain storage commitment of previously transmitted
US images and SR. The Image Manager replies with a N-ACTION response indicating the request has been
received and is being processed.

7. The Image Manager immediately transmits an N-EVENT-REPORT request notifying the Storage AE of the status of
the Storage Commitment Request (sent in step 6 using the N-ACTION message). The Storage AE replies with a
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 16 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

N-EVENT-REPORT response confirming receipt. The Image Manager could send this message at any time or
omit it entirely in favor of transmitting the N-EVENT-REPORT over a separate dedicated association (see note).

8. The Storage AE closes the association with the Image Manager.

NOTE: Many other message sequences are possible depending on the number of images and SR instances to be stored, support for
Storage Commitment and when the SCP sends the N-EVENT-REPORT. The N-EVENT-REPORT can also be sent over a
separate association initiated by the Image Manager (see Section 4.2.1.4.1 on Activity – Receive Storage Commitment
Response).

4.2.1.3.1.2 Proposed Presentation Contexts


The system is capable of proposing the Presentation Contexts shown in the following table:

Table 4.2-7
PROPOSED PRESENTATION CONTEXTS FOR ACTIVITY SEND IMAGES
Presentation Context Table
Abstract Syntax Transfer Syntax
Ext.
Name UID Name List UID List
Role Neg.
Ultrasound Image 1.2.840.10008.5.1. Implicit VR Little Endian 1.2.840.10008.1.2
Storage 4.1.1.6.1 Explicit VR Little Endian 1.2.840.10008.1.2.1 SCU None
Explicit VR Big Endian 1.2.840.10008.1.2.2
Ultrasound 1.2.840.10008.5.1. Implicit VR Little Endian 1.2.840.10008.1.2
Multi-frame Image 4.1.1.3.1 Explicit VR Little Endian 1.2.840.10008.1.2.1 SCU None
Storage Explicit VR Big Endian 1.2.840.10008.1.2.2
Secondary Capture 1.2.840.10008.5.1. Implicit VR Little Endian 1.2.840.10008.1.2
Image Storage 4.1.1.7 Explicit VR Little Endian 1.2.840.10008.1.2.1 SCU None
Explicit VR Big Endian 1.2.840.10008.1.2.2
Comprehensive 1.2.840.10008.5.1.
Implicit VR Little Endian 1.2.840.10008.1.2
Structured 4.1.1.88.33 SCU None
Report Storage Explicit VR Little Endian 1.2.840.10008.1.2.1

Presentation Contexts for Ultrasound Image Storage will only be proposed if the Send Job contains instances for these
SOP Classes.

A Presentation Context for the Storage Commitment Push Model will only be proposed if the Remote AE is configured as an
archive device.

4.2.1.3.1.3 SOP Specific Conformance Image Storage SOP Classes


All Image Storage SOP Classes supported by the Storage AE exhibit the same behavior, except where stated, and are
described together in this section.

If Ultrasound Image Storage SOP Instances are included in the Send Job and a corresponding Presentation Context is not
accepted then the Association is aborted using AP-ABORT and the send job is marked as failed. The job failure is logged
and reported to the user via the job control application. The behavior of Storage AE when encountering status codes in a
C-STORE response is summarized in the Table below:

Table 4.2-8
STORAGE C-STORE RESPONSE STATUS HANDLING BEHAVIOR

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 17 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Service
Further Meaning Error Code Behavior
Status
Success Success 0000 The SCP has successfully stored the SOP Instance. If all SOP
Instances in a send job have status success then the job is marked
as complete.
Refused Out of Resources A700-A7FF The Association is aborted using A-ABORT and the send job is
marked as failed. The status meaning is logged and the job failure
is reported to the user via the job control application. This is a
transient failure.
Error Data Set does not A900-A9FF The Association is aborted using A-ABORT and the send job is
match SOP Class marked as failed. The status meaning is logged and the job failure
is reported to the user via the job control application.
Error Cannot Understand C000-CFFF The Association is aborted using A-ABORT and the send job is
marked as failed. The status meaning is logged and the job failure
is reported to the user via the job control application.
Warning Coercion of Data B000 Image transmission is considered successful but the status meaning
Elements is logged.
Warning Data Set does not B007 Image transmission is considered successful but the status meaning
match SOP Class is logged.
Warning Elements Discarded B006 Image transmission is considered successful but the status meaning
is logged.
* * Any other The Association is aborted using A-ABORT and the send job is
status code. marked as failed. The status code is logged and the job failure is
reported to the user via the job control application.

The behavior of Storage AE during communication failure is summarized in the Table below:

Table 4.2-9
STORAGE COMMUNICATION FAILURE BEHAVIOR
Exception Behavior
Timeout The Association is aborted using A-ABORT and the send job
is marked as failed. The reason is logged and the job failure
is reported to the user via the job control application.
Association aborted by the SCP or network layers The send job is marked as failed. The reason is logged and
the job failure is reported to the user via the job control
application.

A failed send job can be restarted by user interaction. The system can be configured to automatically resend failed jobs if
a transient status code is received. The delay between resending failed jobs and the number of retries is also configurable.

The contents of Ultrasound Image Storage SOP Instances created by the system conform to the DICOM Ultrasound Image
IOD definition and are described in section 8.1.

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 18 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

4.2.1.3.1.4 SOP Specific Conformance for Storage Commitment SOP Class


4.2.1.3.1.4.1 Storage Commitment Operations (N-ACTION)
The Storage AE will request storage commitment for instances of the Ultrasound Image Storage SOP Class and SR
Storage SOP Class if the Remote AE is configured as an archive device and a presentation context for the Storage
Commitment Push Model has been accepted.

The Storage AE will consider Storage Commitment failed if no N-EVENT-REPORT is received for a Transaction UID within
a configurable time period after receiving a successful N-ACTION response (duration of applicability for a Transaction UID).

The Storage AE does not send the optional Storage Media FileSet ID & UID Attributes or the Referenced Study Component
Sequence Attribute in the N-ACTION.

The behavior of Storage AE when encountering status codes in a N-ACTION response is summarized in the Table below:

Table 4.2-10
STORAGE COMMITMENT N-ACTION RESPONSE STATUS HANDLING BEHAVIOR
Service Further
Error Code Behavior
Status Meaning
Success Success 0000 The request for storage comment is considered successfully sent. A timer is
started which will expire if no N-EVENT-REPORT for the Transaction UID is
received within a configurable timeout period.
* * Any other The Association is aborted using A-ABORT and the request for storage comment is
status marked as failed. The status meaning is logged and reported to the user.
code.

The behavior of Storage AE during communication failure is summarized in the Table below:

Table 4.2-11
STORAGE COMMITMENT COMMUNICATION FAILURE BEHAVIOR
Exception Behavior
Timeout The Association is aborted using A-ABORT and the send job
is marked as failed. The reason is logged and the job failure
is reported to the user via the job control application.
Association aborted by the SCP or network layers The send job is marked as failed. The reason is logged and
the job failure is reported to the user via the job control
application.

4.2.1.3.1.4.2 Storage Commitment Notifications (N-EVENT-REPORT)


The Storage AE is capable of receiving an N-EVENT-REPORT notification if it has successfully negotiated a Presentation
Context for the Storage Commitment Push Model (i.e. only associations established with archive devices).

Upon receipt of a N-EVENT-REPORT the timer associated with the Transaction UID will be canceled.

The behavior of Storage AE when receiving Event Types within the N-EVENT-REPORT is summarized in the Table below.

Table 4.2-12
STORAGE COMMITMENT N-EVENT-REPORT BEHAVIOUR

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 19 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Event Type Name Event Behavior


Type
ID
Storage Commitment 1 The Referenced SOP Instances under Referenced SOP Sequence (0008,1199) are
Request Successful marked within the database as “Stored & Committed (SC)” to the value of Retrieve AE
Title (0008,0054). Successfully committed SOP Instances are candidates for
automatic deletion from the local database if local resources become scarce. The
conditions under which automatic deletion is initiated and the amount of space freed are
site configurable. SOP Instances will not be deleted if they are marked with a lock flag.
The least recently accessed SOP Instances are deleted first.
Storage Commitment 2 The Referenced SOP Instances under Referenced SOP Sequence (0008,1199) are
Request Complete – treated in the same way as in the success case (Event Type 1). The Referenced SOP
Failures Exist Instances under Failed SOP Sequence (0008,1198) are marked within the database as
“Store & Commit Failed (Sf)”. The Failure Reasons are logged and the job failure is
reported to the user via the job control application. A send job that failed storage
commitment will not be automatically restarted but can be restarted by user interaction.

The reasons for returning specific status codes in a N-EVENT-REPORT response are summarized in the Table below.

Table 4.2-13
STORAGE COMMITMENT N-EVENT-REPORT RESPONSE STATUS REASONS
Service Further Error
Reasons
Status Meaning Code
Success Success 0000 The storage commitment result has been successfully received.
Failure Unrecognized 0211H The Transaction UID in the N-EVENT-REPORT request is not recognized (was
Operation never issued within an N-ACTION request).
Failure Resource 0213H The Transaction UID in the N-EVENT-REPORT request has expired (no
Limitation N-EVENT-REPORT was received within a configurable time limit).
Failure No Such Event 0113H An invalid Event Type ID was supplied in the N-EVENT-REPORT request.
Type
Failure Processing 0110H An internal error occurred during processing of the N-EVENT-REPORT. A short
Failure description of the error will be returned in Error Comment (0000,0902).
Failure Invalid 0115H One or more SOP Instance UIDs with the Referenced SOP Sequence (0008,1199)
Argument or Failed SOP Sequence (0008,1198) was not included in the Storage Commitment
Value Request associated with this Transaction UID. The unrecognized SOP Instance
UIDs will be returned within the Event Information of the N-EVENT-REPORT
response.

4.2.1.4 Association Acceptance Policy


4.2.1.4.1 Activity – Receive Storage Commitment Response
4.2.1.4.1.1 Description and Sequencing of Activities
The Storage AE will accept associations in order to receive responses to a Storage Commitment Request.

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 20 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Storage Image
AE Manager
Manager

1. Open Association

2. N-EVENT-REPORT (Storage Commitment Response)

3. Close Association

Figure 4.2-2
SEQUENCING OF ACTIVITY - RECEIVE STORAGE COMMITMENT RESPONSE

A possible sequence of interactions between the Storage AE and an Image Manager (e.g. a storage or archive device
supporting Storage Commitment SOP Classes as an SCP) is illustrated in the Figure above:

1. The Image Manager opens a new association with the Storage AE.

2. The Image Manager sends an N-EVENT-REPORT request notifying the Storage AE of the status of a previous
Storage Commitment Request. The Storage AE replies with a N-EVENT-REPORT response confirming receipt.

3. The Image Manager closes the association with the Storage AE.

The Storage AE may reject association attempts as shown in the Table below. The Result, Source and Reason/Diag
columns represent the values returned in the appropriate fields of an ASSOCIATE-RJ PDU (see PS 3.8, Section 9.3.4).
The contents of the Source column is abbreviated to save space and the meaning of the abbreviations are:

a) 1 – DICOM UL service-user

b) 2 – DICOM UL service-provider (ASCE related function)

c) 3 – DICOM UL service-provider (Presentation related function)

Table 4.2-14
ASSOCIATION REJECTION REASONS
Result Source Reason/Diag Explanation
2– c 2 – local-limit-exceeded The (configurable) maximum number of
rejected-transient simultaneous associations has been reached.
An association request with the same
parameters may succeed at a later time.
2– c 1 – temporary-congestion No associations can be accepted at this time
rejected-transient due to the real-time requirements of higher
priority activities (e.g. during image acquisition
no associations will be accepted) or because
insufficient resources are available (e.g.
memory, processes, threads). An association
request with the same parameters may
succeed at a later time.
1– a 2– The association request contained an
rejected-permanent application-context-name-not-supported unsupported Application Context Name. An
association request with the same parameters

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 21 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

will not succeed at a later time.


1– a 7 – called-AE-title-not-recognized The association request contained an
rejected-permanent unrecognized Called AE Title. An association
request with the same parameters will not
succeed at a later time unless configuration
changes are made. This rejection reason
normally occurs when the association initiator
is incorrectly configured and attempts to
address the association acceptor using the
wrong AE Title.
1– a 3 – calling-AE-title-not-recognized The association request contained an
rejected-permanent unrecognized Calling AE Title. An
association request with the same parameters
will not succeed at a later time unless
configuration changes are made. This rejection
reason normally occurs when the association
acceptor has not been configured to recognize
the AE Title of the association initiator.
1– b 1 – no-reason-given The association request could not be parsed.
rejected-permanent An association request with the same format
will not succeed at a later time.

4.2.1.4.1.2 Accepted Presentation Contexts


The Storage AE will accept Presentation Contexts as shown in the Table below.

Table 4.2-15
ACCEPTABLE PRESENTATION CONTEXTS FOR
ACTIVITY RECEIVE STORAGE COMMITMENT RESPONSE
Presentation Context Table
Abstract Syntax Transfer Syntax
Ext.
Name UID Name List UID List
Role Neg.
Storage
1.2.840.10008.1.20 Implicit VR Little Endian 1.2.840.10008.1.2
Commitment Push SCU None
.1 Explicit VR Little Endian 1.2.840.10008.1.2.1
Model
1.2.840.10008.1.1 Implicit VR Little Endian 1.2.840.10008.1.2
Verification SCP None
Explicit VR Little Endian 1.2.840.10008.1.2.1

The Storage AE will prefer to select the Explicit VR Little Endian Transfer Syntax if multiple transfer syntaxes are offered.
The Storage AE will only accept the SCU role (which must be proposed via SCP/SCU Role Selection Negotiation) within a
Presentation Context for the Storage Commitment Push Model SOP Class.

4.2.1.4.1.3 SOP Specific Conformance for Storage Commitment SOP Class


4.2.1.4.1.3.1 Storage Commitment Notifications (N-EVENT-REPORT)
Upon receipt of a N-EVENT-REPORT the timer associated with the Transaction UID will be canceled.

The behavior of Storage AE when receiving Event Types within the N-EVENT-REPORT is summarized in Table 4.2-12.

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 22 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

The reasons for returning specific status codes in a N-EVENT-REPORT response are summarized in Table 4.2-13.

4.2.1.4.1.4 SOP Specific Conformance for Verification SOP Class


The Storage AE provides standard conformance to the Verification SOP Class as an SCP. If the C-ECHO request was
successfully received, a 0000 (Success) status code will be returned in the C-ECHO response. Otherwise, a C000 (Error
– Cannot Understand) status code will be returned in the C-ECHO response.

4.2.2 Workflow Application Entity Specification


4.2.2.1 SOP Classes
The system provides Standard Conformance to the following SOP Classes:

Table 4.2-16
SOP CLASSES FOR AE WORKFLOW
SOP Class Name SOP Class UID SCU SCP

Modality Worklist Information Model – FIND 1.2.840.10008.5.1.4.31 Yes No

Modality Performed Procedure Step 1.2.840.10008.3.1.2.3.3 Yes No

4.2.2.2 Association Policies


4.2.2.2.1 General
The DICOM standard application context name for DICOM 3.0 is always proposed:

Table 4.2-17
DICOM APPLICATION CONTEXT FOR AE WORKFLOW

Application Context Name 1.2.840.10008.3.1.1.1

4.2.2.2.2 Number of Associations


The system initiates one Association at a time for a Worklist request.

Table 4.2-18
NUMBER OF ASSOCIATIONS INITIATED FOR AE WORKFLOW
Maximum number of simultaneous
1
Associations

4.2.2.2.3 Asynchronous Nature


The system does not support asynchronous communication (multiple outstanding transactions over a single Association).

Table 4.2-19
ASYNCHRONOUS NATURE AS A SCU FOR AE WORKFLOW
Maximum number of outstanding asynchronous
1
transactions

4.2.2.2.4 Implementation Identifying Information


The implementation information for this Application Entity is:

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 23 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Table 4.2-20
DICOM IMPLEMENTATION CLASS AND VERSION FOR AE WORKFLOW
1.2.156.112601
Implementation Class UID

Implementation Version Name DCMTK 3.6.0

4.2.2.3 Association Initiation Policy


4.2.2.3.1 Activity – Worklist Update
4.2.2.3.1.1 Description and Sequencing of Activities
The request for a Worklist Update is initiated by user interaction, i.e. pressing the buttons “Worklist Update”/”Patient
Worklist Query” or automatically at specific time intervals, configurable by the user. With “Worklist Update” the automated
query mechanism is performed immediately on request, while with “Patient Worklist Query” a dialog to enter search criteria
is opened and an interactive query can be performed.

The interactive Patient Worklist Query will display a dialog for entering data as search criteria. When the Query is started on
user request, only the data from the dialog will be inserted as matching keys into the query.

With automated worklist queries (including “Worklist Update”) the system always requests all items for a Scheduled
Procedure Step Start Date (actual date), Modality (US) and Scheduled Station AE Title. Query for the Scheduled Station AE
Title is configurable by a Service Engineer.

Upon initiation of the request, The system will build an Identifier for the C-FIND request, will initiate an Association to send
the request and will wait for Worklist responses. After retrieval of all responses, the system will access the local database to
add or update patient demographic data. To protect the system from overflow, the system will limit the number of processed
worklist responses to a configurable maximum. During receiving the worklist response items are counted and the query
processing is canceled by issuing a C-FIND-CANCEL if the configurable limit of items is reached. The results will be
displayed in a separate list, which will be cleared with the next worklist update.

The system will initiate an Association in order to issue a C-FIND request according to the Modality Worklist Information
Model.

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 24 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Workflow Department
AE Scheduler

1. Open Association

2. C-FIND Request (Worklist Query)

3. C-FIND Response (Worklist Item) – Status=Pending

4. C-FIND Response (Worklist Item) – Status=Pending

5. C-FIND Response – Status=Success

6. Close Association

7. Select Worklist Item

Figure 4.2-3
SEQUENCING OF ACTIVITY – WORKLIST UPDATE

A possible sequence of interactions between the Workflow AE and a Departmental Scheduler (e.g. a device such as a RIS
or HIS which supports the Modality Worklist SOP Class as an SCP) is illustrated in the Figure above:

1. The Worklist AE opens an association with the Departmental Scheduler

2. The Worklist AE sends a C-FIND request to the Departmental Scheduler containing the Worklist Query attributes.

3. The Departmental Scheduler returns a C-FIND response containing the requested attributes of the first matching
Worklist Item.

4. The Departmental Scheduler returns another C-FIND response containing the requested attributes of the second
matching Worklist Item.

5. The Departmental Scheduler returns another C-FIND response with status Success indicating that no further
matching Worklist Items exist. This example assumes that only 2 Worklist items match the Worklist Query.

6. The Worklist AE closes the association with the Departmental Scheduler.

7. The user selects a Worklist Item from the Worklist and prepares to acquire new images.

4.2.2.3.1.2 Proposed Presentation Contexts


The system will propose Presentation Contexts as shown in the following table:

Table 4.2-21
PROPOSED PRESENTATION CONTEXTS FOR ACTIVITY WORKLIST UPDATE

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 25 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Presentation Context Table


Abstract Syntax Transfer Syntax
Ext.
Name UID Name List UID List
Role Neg.
Modality Worklist Implicit VR Little Endian 1.2.840.10008.1.2
1.2.840.10008.5.1.
Information Model – Explicit VR Little Endian 1.2.840.10008.1.2.1 SCU None
4.31
FIND Explicit VR Big Endian 1.2.840.10008.1.2.2

4.2.2.3.1.3 SOP Specific Conformance for Modality Worklist


The behavior of the system when encountering status codes in a Modality Worklist C-FIND response is summarized in the
Table below. If any other SCP response status than "Success" or "Pending" is received by the system, a message “query
failed” will appear on the user interface.

Table 4.2-22
MODALITY WORKLIST C-FIND RESPONSE STATUS HANDLING BEHAVIOR
Service
Further Meaning Error Code Behavior
Status
Success Matching is 0000 The SCP has completed the matches. Worklist items are available for
complete display or further processing.
Refused Out of Resources A700 The Association is aborted using A-ABORT and the worklist query is
marked as failed. The status meaning is logged and reported to the user if
an interactive query. Any additional error information in the Response will
be logged.
Failed Identifier does not A900 The Association is aborted using A-ABORT and the worklist query is
match SOP Class marked as failed. The status meaning is logged and reported to the user if
an interactive query. Any additional error information in the Response will be
logged.
Failed Unable to Process C000 – The Association is aborted using A-ABORT and the worklist query is
CFFF marked as failed. The status meaning is logged and reported to the user if
an interactive query. Any additional error information in the Response will be
logged.
Cancel Matching FE00 If the query was cancelled due to too may worklist items then the SCP has
terminated due to completed the matches. Worklist items are available for display or further
Cancel request processing. Otherwise, the Association is aborted using A-ABORT and the
worklist query is marked as failed. The status meaning is logged and
reported to the user if an interactive query.
Pending Matches are FF00 The worklist item contained in the Identifier is collected for later display or
continuing further processing.
Pending Matches are FF01 The worklist item contained in the Identifier is collected for later display or
continuing – further processing. The status meaning is logged only once for each
Warning that one C-FIND operation.
or more Optional
Keys were not
supported
* * Any other The Association is aborted using A-ABORT and the worklist is marked as
status code. failed. The status meaning is logged and reported to the user if an
interactive query. Any additional error information in the Response will be
logged.
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 26 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

The behavior of the system during communication failure is summarized in the Table below.

Table 4.2-23
MODALITY WORKLIST COMMUNICATION FAILURE BEHAVIOR
Exception Behavior
Timeout The Association is aborted using A-ABORT and the worklist query marked as failed.
The reason is logged and reported to the user if an interactive query.
Association aborted by the SCP The worklist query is marked as failed. The reason is logged and reported to the user if
or network layers an interactive query.

Acquired images will always use the Study Instance UID specified for the Scheduled Procedure Step (if available). If an
acquisition is unscheduled, a Study Instance UID will be generated locally.

The Table below provides a description of the system Worklist Request Identifier and specifies the attributes that are copied
into the images. Unexpected attributes returned in a C-FIND response are ignored.

Requested return attributes not supported by the SCP are set to have no value. Non-matching responses returned by the
SCP due to unsupported optional matching keys are ignored. No attempt is made it filter out possible duplicate entries.

Table 4.2-24
WORKLIST REQUEST IDENTIFIER
Module Name Tag VR M R Q D IOD
Attribute Name
SOP Common
Specific Character Set (0008,0005) CS x
Scheduled Procedure Step
Scheduled Procedure Step Sequence (0040,0100) SQ x
> Scheduled Station AET (0040,0001) AE (S) x x
> Scheduled Procedure Step Start Date (0040,0002) DA S x x
> Scheduled Procedure Step Start Time (0040,0003) TM x x
> Modality (0008,0060) CS S x
> Scheduled Performing Physician’s Name (0040,0006) PN x x x x
> Scheduled Procedure Step Description (0040,0007) LO x x x
> Scheduled Station Name (0040,0010) SH S x
> Scheduled Procedure Step Location (0040,0011) SH S x
> Scheduled Protocol Code Sequence (0040,0008) SQ x x
> Pre-Medication (0040,0012) LO x x
> Scheduled Procedure Step ID (0040,0009) SH x x x
> Requested Contrast Agent (0032,1070) LO x x
Requested Procedure
Requested Procedure ID (0040,1001) SH S x x x x
Requested Procedure Description (0032,1060) LO x x x
Study Instance UID (0020,000D) UI x x
Requested Procedure Priority (0040,1003) SH x
Patient Transport Arrangements (0040,1004) LO x
Referenced Study Sequence (0008,1110) SQ x x
Requested Procedure Code Sequence (0032,1064) SQ x x

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 27 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Imaging Service Request


Accession Number (0008,0050) SH S x x x x
Requesting Physician (0032,1032) PN x x x
Referring Physician's Name (0008,0090) PN x x x x
Visit Identification
Admission ID (0038,0010) LO x
Visit Status
Current Patient Location (0038,0300) LO x x
Visit Admission
Admitting Diagnosis Description (0008,1080) LO x x
Patient Identification
Patient Name (0010,0010) PN R x x x x
Patient ID (0010,0020) LO R x x x x
Patient Demographic
Patient’s Birth Date (0010,0030) DA x x x x
Patient’s Sex (0010,0040) CS x x x x
Patient’s Weight (0010,1030) DS x x x
Patient's Size (0010,1020) DS x x x
Confidentiality constraint on patient data (0040,3001) LO x x
Patient Medical
Patient State (0038,0500) LO x x
Pregnancy Status (0010,21C0) US x x
Medical Alerts (0010,2000) LO x x
Allergies (0010,2110) LO x x
Special Needs (0038,0050) LO x x

The above table should be read as follows:

Module Name: The name of the associated module for supported worklist attributes.
Attribute Name: Attributes supported to build an Worklist Request Identifier.
Tag: DICOM tag for this attribute.
VR: DICOM VR for this attribute.
M: Matching keys for (automatic) Worklist Update. A "S" will indicate that the system will supply an attribute
value for Single Value Matching, a “R” will indicate Range Matching and a “*” will denote wildcard
matching. It can be configured if “Scheduled Station AE Title” is additionally supplied “(S)” and if Modality
is set to US or SC.
R: Return keys. An "x" will indicate that the system will supply this attribute as Return Key with zero length
for Universal Matching. The system will support retired date format (yyyy.mm.dd) for "Patient’s Birth
Date" and "Scheduled Procedure Step Start Date" in the response identifiers. For "Scheduled Procedure
Step Start Time" also retired time format as well as unspecified time components are supported.
Q: Interactive Query Key. An “x” " will indicate that the system will supply this attribute as matching key, if
entered in the Query Patient Worklist dialog. For example, the Patient Name can be entered thereby
restricting Worklist responses to Procedure Steps scheduled for the patient.
D: Displayed keys. An “x” indicates that this worklist attribute is displayed to the user during a patient
registration dialog. For example, Patient Name will be displayed when registering the patient prior to an
examination.
IOD: An "x" indicates that this Worklist attribute is included into all Object Instances created during
performance of the related Procedure Step.

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 28 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

The default Query Configuration is set to "Modality" (US) and "Date" (date of today). Optionally, additional matching for the
own AET is configurable.

4.2.2.3.2 Activity – Acquire Images


4.2.2.3.2.1 Description and Sequencing of Activities
After Patient registration, the system is awaiting the 1st application of Ultrasound Dose to the patient. The trigger to create
a MPPS SOP Instance is derived from this event. An Association to the configured MPPS SCP system is established
immediately and the related MPPS SOP Instance will be created.

A manual update can be performed with the MPPS user interface where is it possible to set the final state of the MPPS to
“COMPLETED” or “DISCONTINUED”. In the “Discontinued” case the user can also select the discontinuation reason from a
list corresponding to Context Group 9300. A MPPS Instance that has been sent with a state of “COMPLETED” or
“DISCONTINUED” can no longer be updated.

The system will support creation of “unscheduled cases” by allowing MPPS Instances to be communicated for locally
registered Patients.

The system only supports a 0-to-1 relationship between Scheduled and Performed Procedure Steps.

The system will initiate an Association to issue an:

— N-CREATE request according to the CREATE Modality Performed Procedure Step SOP Instance operation or a

— N-SET request to update the contents and state of the MPPS according to the SET Modality Performed Procedure Step
Information operation.

Workflow Department
AE Scheduler

1. Open Association

2. N-CREATE (MPPS) – IN PROGRESS

3. Close Association

4. Acquire Images

5. Open Association

6. N-SET (MPPS) – COMPLETED

7. Close Association

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 29 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Figure 4.2-4
SEQUENCING OF ACTIVITY – ACQUIRE IMAGES

A possible sequence of interactions between the Workflow AE and a Departmental Scheduler (e.g. a device such as a RIS
or HIS which supports the MPPS SOP Class as an SCP) is illustrated in Figure 4.2-4:

1. The Worklist AE opens an association with the Departmental Scheduler

2. The Worklist AE sends an N-CREATE request to the Departmental Scheduler to create an MPPS instance with
status of “IN PROGRESS” and create all necessary attributes. The Departmental Scheduler acknowledges the
MPPS creation with an N-CREATE response (status success).

3. The Worklist AE closes the association with the Departmental Scheduler.

4. All images are acquired and stored in the local database.

5. The Worklist AE opens an association with the Departmental Scheduler.

6. The Worklist AE sends an N-SET request to the Departmental Scheduler to update the MPPS instance with status
of “COMPLETED” and set all necessary attributes. The Departmental Scheduler acknowledges the MPPS update
with an N-SET response (status success).

7. The Worklist AE closes the association with the Departmental Scheduler.

4.2.2.3.2.2 Proposed Presentation Contexts


The system will propose Presentation Contexts as shown in the following table:

Table 4.2-25
PROPOSED PRESENTATION CONTEXTS FOR REAL-WORLD ACTIVITY ACQUIRE IMAGES
Presentation Context Table
Abstract Syntax Transfer Syntax
Ext.
Name UID Name List UID List
Role Neg.
Implicit VR Little Endian 1.2.840.10008.1.2
Modality Performed 1.2.840.10008.3.1.
Explicit VR Little Endian 1.2.840.10008.1.2.1 SCU None
Procedure Step 2.3.3
Explicit VR Big Endian 1.2.840.10008.1.2.2

4.2.2.3.2.3 SOP Specific Conformance for MPPS


The behavior of the system when encountering status codes in an MPPS N-CREATE or N–SET response is summarized in
Table 4.2-26. If any other SCP response status than "Success" or “Warning” is received by the system, a message “MPPS
update failed” will appear on the user interface.

Table 4.2-26
MPPS N-CREATE / N-SET RESPONSE STATUS HANDLING BEHAVIOR
Service Error
Further Meaning Behavior
Status Code
Success Success 0000 The SCP has completed the operation successfully.
Failure Processing Failure – 0110 The Association is aborted using A-ABORT and the MPPS is
Performed Procedure Step marked as failed. The status meaning is logged and reported to

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 30 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Object may no longer be the user. Additional information in the Response will be logged
updated (i.e. Error Comment and Error ID).
Warning Attribute Value Out of Range 0116H The MPPS operation is considered successful but the status
meaning is logged. Additional information in the Response
identifying the attributes out of range will be logged (i.e. Elements in
the Modification List/Attribute List)
* * Any other The Association is aborted using A-ABORT and the MPPS is
status marked as failed. The status meaning is logged and reported to
code. the user.

The behavior of the system during communication failure is summarized in the Table below:

Table 4.2-27
MPPS COMMUNICATION FAILURE BEHAVIOR
Exception Behavior
Timeout The Association is aborted using A-ABORT and MPPS marked as failed. The
reason is logged and reported to the user.
Association aborted by the SCP or The MPPS is marked as failed. The reason is logged and reported to the user.
network layers

Table 4.2-28 provides a description of the MPPS N-CREATE and N-SET request identifiers sent by the system. Empty
cells in the N-CREATE and N-SET columns indicate that the attribute is not sent. An “x” indicates that an appropriate value
will be sent. A “Zero length” attribute will be sent with zero length.

Table 4.2-28
MPPS N-CREATE / N-SET REQUEST IDENTIFIER
Attribute Name Tag VR N-CREATE N-SET
Specific Character Set (0008,0005) CS “ISO_IR 100” or “ISO_IR 144”
Modality (0008,0060) CS US
Referenced Patient
(0008,1120) SQ Zero length
Sequence
From Modality Worklist or user
input (all 5 components). The
Patient’s Name (0010,0010) PN
user can modify values
provided via Modality Worklist.
From Modality Worklist or user
input. The user can modify
Patient ID (0010,0020) LO
values provided via Modality
Worklist.
From Modality Worklist or user
input. The user can modify
Patient’s Birth Date (0010,0030) DA
values provided via Modality
Worklist.
From Modality Worklist or user
input. The user can modify
Patient’s Sex (0010,0040) CS
values provided via Modality
Worklist.

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 31 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Distance Source to Zero length


(0018,1110) DS x
Detector (SID)
Image Area Dose Zero length
(0018,115E) DS x
Product
From Modality Worklist or user
input. The user can modify
Study ID (0020,0010) SH
values provided via Modality
Worklist.
Performed Station AE
(0040,0241) AE MPPS AE Title
Title
Performed Station Name (0040,0242) SH From configuration
Performed Location (0040,0243) SH From configuration
Performed Procedure
(0040,0244) DA Actual start date
Step Start Date
Performed Procedure
(0040,0245) TM Actual start time
Step Start Time
Performed Procedure
(0040,0250) DA Zero length Actual end date
Step End Date
Performed Procedure
(0040,0251) TM Zero length Actual end time
Step End Time
Performed Procedure
(0040,0252) CS IN PROGRESS DISCONTINUED or COMPLETED
Step Status
If Performed Procedure Step Status
Performed Procedure (0040,0252) is “DISCONTINUED” then a
Step Discontinuation (0040,0281) SQ Zero length single item will be present containing a
Reason Code Sequence user-selected entry drawn from Context
Group 9300.
Performed Procedure Automatically created but can
(0040,0253) SH
Step ID be modified by the user.
From Modality Worklist or user
Performed Procedure input. The user can modify
(0040,0254) LO
Step Description the description provided via
Modality Worklist.
Performed Procedure Zero length
(0040,0255) LO
Type Description
Performed Protocol
(0040,0260) SQ Zero length Zero or more items
Code Sequence
Scheduled Step If 1st dose applied results in an
(0040,0270) SQ
Attributes Sequence Instance
From Modality Worklist or user
input. The user can modify
> Accession Number (0008,0050) SH
values provided via Modality
Worklist.
> Referenced Study
(0008,1110) SQ From Modality Worklist
Sequence
>> Referenced SOP
(0008,1150) UI From Modality Worklist
Class UID
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 32 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

>> Referenced SOP


(0008,1155) UI From Modality Worklist
Instance UID
> Study Instance UID (0020,000D) UI From Modality Worklist
> Requested Procedure
(0032,1060) LO From Modality Worklist
Description
> Scheduled Procedure
(0040,0007) LO From Modality Worklist
Step Description
> Scheduled Protocol
(0040,0008) SQ From Modality Worklist
Code Sequence
> Scheduled Procedure
(0040,0009) SH From Modality Worklist
Step ID
> Requested Procedure
(0040,1001) SH From Modality Worklist
ID
Performed Series if 1st dose applied results in an
(0040,0340) SQ One or more items
Sequence instance
> Retrieve AE Title (0008,0054) AE x x
> Series Description (0008,103E) LO x x
> Performing Physician’s
(0008,1050) PN x x
Name
> Operator’s Name (0008,1070) PN x x
> Referenced Image
(0008,1140) SQ One or more items One or more items
Sequence
>> Referenced SOP
(0008,1150) UI
Class UID
>> Referenced SOP
(0008,1155) UI
Instance UID
> Protocol Name (0018,1030) LO
> Series Instance UID (0020,000E) UI
> Referenced No
Zero length (SOP classes not
Imag-Composite SOP (0040,0220) SQ Zero length (SOP classes not supported)
supported)
Instance

4.2.2.4 Association Acceptance Policy


The Workflow Application Entity does not accept Associations.

4.2.3 Hardcopy Application Entity Specification


4.2.3.1 SOP Classes
The system provides Standard Conformance to the following SOP Classes:

Table 4.2-29
SOP CLASSES FOR AE HARDCOPY
SOP Class Name SOP Class UID SCU SCP

Basic Grayscale Print Management Meta 1.2.840.10008.5.1.1.9 Yes No

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 33 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Basic Color Print Management Meta 1.2.840.10008.5.1.1.18 Yes No

Basic Annotation Box SOP Class 1.2.840.10008.5.1.1.15 Yes No

Print Job SOP Class 1.2.840.10008.5.1.1.14 Yes No

Presentation LUT 1.2.840.10008.5.1.1.23 Yes No

4.2.3.2 Association Policies


4.2.3.2.1 General
The DICOM standard application context name for DICOM 3.0 is always proposed:

Table 4.2-30
DICOM APPLICATION CONTEXT FOR AE HARDCOPY

Application Context Name 1.2.840.10008.3.1.1.1

4.2.3.2.2 Number of Associations


The system initiates one Association at a time for each configured hardcopy device. Multiple hardcopy devices can be
configured.

Table 4.2-31
NUMBER OF ASSOCIATIONS INITIATED FOR AE HARDCOPY

Maximum number of simultaneous Associations 1

4.2.3.2.3 Asynchronous Nature


The system does not support asynchronous communication (multiple outstanding transactions over a single Association).

Table 4.2-32
ASYNCHRONOUS NATURE AS A SCU FOR AE HARDCOPY

Maximum number of outstanding asynchronous transactions 1

4.2.3.2.4 Implementation Identifying Information


The implementation information for this Application Entity is:

Table 4.2-33
DICOM IMPLEMENTATION CLASS AND VERSION FOR AE HARDCOPY
1.2.156.112601
Implementation Class UID

Implementation Version Name DCMTK 3.6.0

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 34 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

4.2.3.3 Association Initiation Policy


4.2.3.3.1 Activity – Film Images
4.2.3.3.1.1 Description and Sequencing of Activities
A user composes images onto film sheets and requests them to be sent to a specific hardcopy device. The user can select
the desired film format and number of copies. Each print-job is forwarded to the job queue and processed individually.

The Hardcopy AE is invoked by the job control interface that is responsible for processing network tasks. The job consists of
data describing the images and graphics to be printed as well as the requested layout and other parameters. The film
sheet is internally processed, converted to a STANDARD/1,1 page and then the page image is sent. If no association to
the printer can be established, the print-job is switched to a failed state and the user informed.

Hardcopy Printer
AE

1. Open Association

2. N-GET (Printer)

3. N-CREATE (Film Session)

4. N-CREATE (Presentation LUT)

5. N-CREATE (Film Box)

6. N-SET (Image Box)

7. N-ACTION (Film Box)

8. Print film sheets


9. N-EVENT-REPORT (Printer)

10. N-DELETE (Film Session)

11. Close Association

Figure 4.2-5
SEQUENCING OF ACTIVITY – FILM IMAGES

A typical sequence of DIMSE messages sent over an association between Hardcopy AE and a Printer is illustrated in Figure
4.2-5:

1. Hardcopy AE opens an association with the Printer

2. N-GET on the Printer SOP Class is used to obtain current printer status information. If the Printer reports a status
of FAILURE, the print-job is switched to a failed state and the user informed.

3. N-CREATE on the Film Session SOP Class creates a Film Session.

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 35 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

4. N-CREATE on the Presentation LUT SOP Class creates a Presentation LUT (if supported by the printer).

5. N-CREATE on the Film Box SOP Class creates a Film Box linked to the Film Session. A single Image Box will be
created as the result of this operation (Hardcopy AE only uses the format STANDARD\1,1)

6. N-SET on the Image Box SOP Class transfers the contents of the film sheet to the printer. If the printer does not
support the Presentation LUT SOP Class, the image data will be passed through a printer-specific correction LUT
before being sent.

7. N-ACTION on the Film Box SOP Class instructs the printer to print the Film Box

8. The printer prints the requested number of film sheets

9. The Printer asynchronously reports its status via N-EVENT-REPORT notification (Printer SOP Class). The printer
can send this message at any time. Hardcopy AE does not require the N-EVENT-REPORT to be sent. Hardcopy
AE is capable of receiving an N-EVENT-REPORT notification at any time during an association. If the Printer
reports a status of FAILURE, the print-job is switched to a failed state and the user informed.

10. N-DELETE on the Film Session SOP Class deletes the complete Film Session SOP Instance hierarchy.

11. Hardcopy AE closes the association with the Printer

Status of the print-job is reported through the job control interface. Only one job will be active at a time for each separate
hardcopy device. If any Response from the remote Application contains a status other than Success or Warning, the
Association is aborted and the related Job is switched to a failed state. It can be restarted any time by user interaction or, if
configured, by automated retry.

4.2.3.3.1.2 Proposed Presentation Contexts


The system is capable of proposing the Presentation Contexts shown in the Table below:

Table 4.2-34
PROPOSED PRESENTATION CONTEXTS FOR ACTIVITY FILM IMAGES
Presentation Context Table
Abstract Syntax Transfer Syntax
Ext.
Name UID Name List UID List
Role Neg.
Basic Grayscale Implicit VR Little Endian 1.2.840.10008.1.2
1.2.840.10008.5.1.
Print Management Explicit VR Little Endian 1.2.840.10008.1.2.1 SCU None
1.9
Meta Explicit VR Big Endian 1.2.840.10008.1.2.2
Implicit VR Little Endian 1.2.840.10008.1.2
Basic Color Print 1.2.840.10008.5.1.
Explicit VR Little Endian 1.2.840.10008.1.2.1 SCU None
Management Meta 1.18
Explicit VR Big Endian 1.2.840.10008.1.2.2
Implicit VR Little Endian 1.2.840.10008.1.2
Basic Annotation 1.2.840.10008.5.1.
Explicit VR Little Endian 1.2.840.10008.1.2.1 SCU None
Box SOP Class 1.15
Explicit VR Big Endian 1.2.840.10008.1.2.2
Implicit VR Little Endian 1.2.840.10008.1.2
1.2.840.10008.5.1.
Print Job SOP Class Explicit VR Little Endian 1.2.840.10008.1.2.1 SCU None
1.14
Explicit VR Big Endian 1.2.840.10008.1.2.2
Implicit VR Little Endian 1.2.840.10008.1.2
1.2.840.10008.5.1.
Presentation LUT Explicit VR Little Endian 1.2.840.10008.1.2.1 SCU None
1.23
Explicit VR Big Endian 1.2.840.10008.1.2.2
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 36 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

4.2.3.3.1.3 Common SOP Specific Conformance for all Print SOP Classes
The general behavior of Hardcopy AE during communication failure is summarized in the Table below. This behavior is
common for all SOP Classes supported by Hardcopy AE.

Table 4.2-35
HARDCOPY COMMUNICATION FAILURE BEHAVIOR
Exception Behavior
Timeout The Association is aborted using A-ABORT and the print-job
is marked as failed. The reason is logged and the job failure
is reported to the user via the job control application.
Association aborted by the SCP or network layers The print-job is marked as failed. The reason is logged and
the job failure is reported to the user via the job control
application.

4.2.3.3.1.4 SOP Specific Conformance for the Printer SOP Class


Hardcopy AE supports the following DIMSE operations and notifications for the Printer SOP Class:

— N-GET

— N-EVENT-REPORT

Details of the supported attributes and status handling behaviour are described in the following subsections.

4.2.3.3.1.4.1 Printer SOP Class Operations (N-GET)


Hardcopy AE uses the Printer SOP Class N-GET operation to obtain information about the current printer status. The
attributes obtained via N-GET are listed in the Table below:

Table 4.2-36
PRINTER SOP CLASS N-GET REQUEST ATTRIBUTES
Presence of
Attribute Name Tag VR Value Source
Value
Printer Status (2110,0010) CS Provided by Printer ALWAYS Printer
Printer Status Info (2110,0020) CS Provided by Printer ALWAYS Printer

The Printer Status information is evaluated as follows:

1. If Printer status (2110,0010) is NORMAL, the print-job continues to be printed.

2. If Printer status (2110,0010) is FAILURE, the print-job is marked as failed. The contents of Printer Status Info
(2110,0020) is logged and reported to the user via the job control application.

3. If Printer status (2110,0010) is WARNING, the print-job continues to be printed. The contents of Printer Status Info
(2110,0020) is logged and reported to the user via the job control application.

The behavior of Hardcopy AE when encountering status codes in a N-GET response is summarized in the Table below:

Table 4.2-37
PRINTER SOP CLASS N-GET RESPONSE STATUS HANDLING BEHAVIOR
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 37 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Service Further
Error Code Behavior
Status Meaning
Success Success 0000 The request to get printer status information was success.
* * Any other The Association is aborted using A-ABORT and the print-job is marked as
status code. failed. The status meaning is logged and reported to the user.

4.2.3.3.1.4.2 Printer SOP Class Notifications (N-EVENT-REPORT)


Hardcopy AE is capable of receiving an N-EVENT-REPORT request at any time during an association.

The behavior of Hardcopy AE when receiving Event Types within the N-EVENT-REPORT is summarized in the Table
below:

Table 4.2-38
PRINTER SOP CLASS N-EVENT-REPORT BEHAVIOUR
Event Type Event Behavior
Name Type ID
Normal 1 The print-job continues to be printed.
Warning 2 The print-job continues to be printed. The contents of Printer Status Info (2110,0020) is
logged and reported to the user via the job-control application.
Failure 3 The print-job is marked as failed. The contents of Printer Status Info (2110,0020) is logged
and reported to the user via the job-control application.
* * An invalid Event Type ID will cause a status code of 0113H to be returned in a
N-EVENT-REPORT response.

The reasons for returning specific status codes in a N-EVENT-REPORT response are summarized in the Table below:

Table 4.2-39
PRINTER SOP CLASS N-EVENT-REPORT RESPONSE STATUS REASONS
Service Further Error
Reasons
Status Meaning Code
Success Success 0000 The notification event has been successfully received.
Failure No Such Event 0113H An invalid Event Type ID was supplied in the N-EVENT-REPORT request.
Type
Failure Processing 0110H An internal error occurred during processing of the N-EVENT-REPORT. A short
Failure description of the error will be returned in Error Comment (0000,0902).

4.2.3.3.1.5 SOP Specific Conformance for the Film Session SOP Class
Hardcopy AE supports the following DIMSE operations for the Film Session SOP Class:

— N-CREATE

— N-DELETE

Details of the supported attributes and status handling behavior are described in the following subsections.

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 38 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

4.2.3.3.1.5.1 Film Session SOP Class Operations (N-CREATE)


The attributes supplied in an N-CREATE Request are listed in the Table below:

Table B4.2-40
FILM SESSION SOP CLASS N-CREATE REQUEST ATTRIBUTES
Presence of
Attribute Name Tag VR Value Source
Value
Number of
(2000,0010) IS 1 .. 10 ALWAYS User
Copies
Print
(2000,0020) CS LOW, MED or HIGH ALWAYS User
Priority
Medium Type (2000,0030) CS BLUE FILM, CLEAR FILM or PAPER ALWAYS User
Print
(2000,0020) CS LOW, MED or HIGH ALWAYS User
Priority
Film Destination (2000,0040) CS MAGAZINE or PROCESSOR ALWAYS User
Film Session
(2000,0050) LO User
Label

The behavior of Hardcopy AE when encountering status codes in a N-CREATE response is summarized in the Table below:

Table 4.2-41
FILM SESSION SOP CLASS N-CREATE RESPONSE STATUS HANDLING BEHAVIOR
Service Further Error
Behavior
Status Meaning Code
Success Success 0000 The SCP has completed the operation successfully.
Warning Attribute Value 0116H The N-CREATE operation is considered successful but the status meaning is
Out of Range logged. Additional information in the Response identifying the attributes out of
range will be logged (i.e. Elements in the Modification List/Attribute List)
Warning Attribute List 0107H The N-CREATE operation is considered successful but the status meaning is
Error logged. Additional information in the Response identifying the attributes will be
logged (i.e. Elements in the Attribute Identifier List)
* * Any other The Association is aborted using A-ABORT and the print-job is marked as failed.
status The status meaning is logged and reported to the user.
code.

4.2.3.3.1.5.2 Film Session SOP Class Operations (N-DELETE)


The behavior of Hardcopy AE when encountering status codes in a N-DELETE response is summarized in the Table below:

Table 4.2-42
PRINTER SOP CLASS N-DELETE RESPONSE STATUS HANDLING BEHAVIOR
Service Further
Error Code Behavior
Status Meaning
Success Success 0000 The SCP has completed the operation successfully.
* * Any other The Association is aborted using A-ABORT and the print-job is marked as
status code. failed. The status meaning is logged and reported to the user.

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 39 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

4.2.3.3.1.6 SOP Specific Conformance for the Presentation LUT SOP Class
Hardcopy AE supports the following DIMSE operations for the Presentation LUT SOP Class:

— N-CREATE

Details of the supported attributes and status handling behavior are described in the following subsections.

4.2.3.3.1.6.1 Presentation LUT SOP Class Operations (N-CREATE)


The attributes supplied in an N-CREATE Request are listed in the Table below:

Table 4.2-43
PRESENTATION LUT SOP CLASS N-CREATE REQUEST ATTRIBUTES
Presence of
Attribute Name Tag VR Value Source
Value
Presentation LUT
(2050,0020) CS IDENTITY ALWAYS Auto
Shape

The behavior of Hardcopy AE when encountering status codes in a N-CREATE response is summarized in the Table below:

Table 4.2-44
PRESENTATION LUT SOP CLASS N-CREATE RESPONSE STATUS HANDLING BEHAVIOR
Service
Further Meaning Error Code Behavior
Status
Success Success 0000 The SCP has completed the operation successfully.
Warning Requested Min Density or Max B605H The N-CREATE operation is considered successful but the
Density outside of printer’s status meaning is logged.
operating range
* * Any other The Association is aborted using A-ABORT and the
status code. print-job is marked as failed. The status meaning is
logged and reported to the user.
Service Further Meaning Error Code Behavior
Status
Success Success 0000 The SCP has completed the operation successfully.

4.2.3.3.1.7 SOP Specific Conformance for the Film Box SOP Class
Hardcopy AE supports the following DIMSE operations for the Presentation LUT SOP Class:

— N-CREATE

— N-ACTION

Details of the supported attributes and status handling behavior are described in the following subsections.

4.2.3.3.1.7.1 Film Box SOP Class Operations (N-CREATE)


The attributes supplied in an N-CREATE Request are listed in the Table below:

Table 4.2-45
FILM BOX SOP CLASS N-CREATE REQUEST ATTRIBUTES

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 40 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Presence of
Attribute Name Tag VR Value Source
Value
STANDARD\1,1 depending on printer. Default is
Image Display
(2010,0010) CS displayed, and user editable. Edit only when a ALWAYS Auto/User
Format
vaild substitute value is known.
Referenced Film
Session (2010,0500) SQ ALWAYS Auto
Sequence
>Referenced
(0008,1150) UI 1.2.840.10008.5.1.1.1 ALWAYS Auto
SOP Class UID
>Referenced
SOP Instance (0008,1155) UI From created Film Session SOP Instance ALWAYS Auto
UID
Film Orientation (2010,0040) CS PORTRAIT or LANDSCAPE ALWAYS User
8INX10IN, 8_5INX11IN, 10INX12IN,
10INX14IN, 11INX14IN, 11INX17IN;
Film Size ID (2010,0050) CS ALWAYS User
14INX14IN, 14INX17IN, 24CMX24CM,
24CMX30CM, A4, A3
Magnification
(2010,0060) CS REPLICATE, BILINEAR, CUBIC or NONE ALWAYS User
Type
Border Density (2010,0100) CS BLACK or WHITE ALWAYS User
Max Density (2010,0130) US 0 .. 310 Default is displayed, and user editable. ALWAYS Auto
Min Density (2010,0120) US 0 .. 50 Default is displayed, and user editable. ALWAYS Auto
Smoothing Type (2010,0080) US 0 .. 100 Default is displayed, and user editable. ALWAYS User
Empty Image
(2010,0110) CS BLACK or WHITE ALWAYS User
Density
Trim (2010,0140) CS YES or NO ALWAYS User
Configuration Default is displayed, and user editable. Edit only
(2010,0150) ST ALWAYS User
Information when a vaild substitute value is known.
Referenced
Only sent if Presentation LUT SOP Class has
Presentation LUT (2050,0500) SQ ANAP Auto
been negotiated.
Sequence
>Referenced
(0008,1150) UI 1.2.840.10008.5.1.1.23 ALWAYS Auto
SOP Class UID
>Referenced
SOP Instance (0008,1155) UI From created Presentation LUT SOP Instance ALWAYS Auto
UID

The behavior of Hardcopy AE when encountering status codes in a N-CREATE response is summarized in the Table below:

Table 4.46
FILM BOX SOP CLASS N-CREATE RESPONSE STATUS HANDLING BEHAVIOR
Service
Further Meaning Error Code Behavior
Status
Success Success 0000 The SCP has completed the operation successfully.

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 41 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Warning Requested Min Density or Max B605H The N-CREATE operation is considered successful but the
Density outside of printer’s status meaning is logged.
operating range
* * Any other The Association is aborted using A-ABORT and the
status code. print-job is marked as failed. The status meaning is
logged and reported to the user.

4.2.3.3.1.7.2 Film Box SOP Class Operations (N-ACTION)


An N-ACTION Request is issued to instruct the Print SCP to print the contents of the Film Box. The Action Reply argument
in an N-ACTION response is not evaluated.

The behavior of Hardcopy AE when encountering status codes in a N-ACTION response is summarized in the Table below:

Table 4.2-47
FILM BOX SOP CLASS N-ACTION RESPONSE STATUS HANDLING BEHAVIOR
Service Error
Further Meaning Behavior
Status Code
Success Success 0000 The SCP has completed the operation
successfully. The film has been accepted for
printing.
Warning Film Box SOP Instance hierarchy does not B603H The Association is aborted using A-ABORT and
contain Image Box SOP Instances (empty the print-job is marked as failed. The status
page) meaning is logged and reported to the user.
Warning Image size is larger than Image Box size. The B604H The N-ACTION operation is considered
image has been demagnified. successful but the status meaning is logged.
Warning Image size is larger than Image Box size. The B609H The N-ACTION operation is considered
image has been cropped to fit. successful but the status meaning is logged.
Warning Image size or Combined Print Image Size is B60AH The N-ACTION operation is considered
larger than Image Box size. The image or successful but the status meaning is logged.
combined Print Image has been decimated to
fit.
Failure Unable to create Print Job SOP Instance; print C602 The Association is aborted using A-ABORT and
queue is full. the print-job is marked as failed. The status
meaning is logged and reported to the user.
Failure Image size is larger than Image Box size. C603 The Association is aborted using A-ABORT and
the print-job is marked as failed. The status
meaning is logged and reported to the user.
Failure Combined Print Image Size is larger than Image C613 The Association is aborted using A-ABORT and
Box size. the print-job is marked as failed. The status
meaning is logged and reported to the user.
* * Any other The Association is aborted using A-ABORT and
status the print-job is marked as failed. The status
code. meaning is logged and reported to the user.

4.2.3.3.1.8 SOP Specific Conformance for the Image Box SOP Class
Hardcopy AE supports the following DIMSE operations for the Image Box SOP Class:

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 42 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

— N-SET

Details of the supported attributes and status handling behavior are described in the following subsections.

4.2.3.3.1.8.1 Image Box SOP Class Operations (N-SET)


The attributes supplied in an N-SET Request are listed in the Table below:

Table 4.2-48-1
BASIC GRAYSCALE IMAGE BOX SOP CLASS N-SET REQUEST ATTRIBUTES
Presence of
Attribute Name Tag VR Value Source
Value
Image Position (2020,0010) US 1 ALWAYS Auto
Basic Grayscale
(2020,0110) SQ ALWAYS Auto
Image Sequence
>Samples Per Pixel (0028,0002) US 1 ALWAYS Auto
>Photometric
(0028,0004) CS MONOCHROME2 ALWAYS Auto
Interpretation
>Rows (0028,0010) US Depends on film size ALWAYS Auto
>Columns (0028,0011) US Depends on film size ALWAYS Auto
>Pixel Aspect Ratio (0028,0034) IS 1\1 ALWAYS Auto
>Bits Allocated (0028,0100) US 8 ALWAYS Auto
>Bits Stored (0028,0101) US 8 ALWAYS Auto
>High Bit (0028,0102) US 7 ALWAYS Auto
>Pixel
(0028,0103) US 0 ALWAYS Auto
Representation
>Pixel Data (7FE0,0010) OB Pixels of rendered film sheet ALWAYS Auto

Table 4.2-48
BASIC GRAYSCALE OR BASIC COLOR IMAGE BOX SOP CLASS N-SET REQUEST ATTRIBUTES
Presence of
Attribute Name Tag VR Value Source
Value
Image Position (2020,0010) US 1 ALWAYS Auto
Basic Grayscale
(2020,0110) SQ Used for BW (MONOCHROME2) print ALWAYS Auto
Image Sequence
Basic Color Image
(2020,0111) SQ Used for color (RGB) print ALWAYS Auto
Sequence
1 for MONOCHROME2
>Samples Per Pixel (0028,0002) US ALWAYS Auto
3 for RGB
>Photometric
(0028,0004) CS MONOCHROME2 or RGB ALWAYS Auto
Interpretation
Planar
(0028,0006) US Always “1”,only used for RGB print ANAP Auto
Configuration
>Rows (0028,0010) US Depends on film size ALWAYS Auto
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 43 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

>Columns (0028,0011) US Depends on film size ALWAYS Auto


>Pixel Aspect Ratio (0028,0034) IS 1\1 ALWAYS Auto
>Bits Allocated (0028,0100) US 8 ALWAYS Auto
>Bits Stored (0028,0101) US 8 ALWAYS Auto
>High Bit (0028,0102) US 7 ALWAYS Auto
>Pixel
(0028,0103) US 0 ALWAYS Auto
Representation
>Pixel Data (7FE0,0010) OB Pixels of rendered film sheet ALWAYS Auto

The behavior of Hardcopy AE when encountering status codes in a N-SET response is summarized in the Table below:

Table 4.2-49
IMAGE BOX SOP CLASS N-SET RESPONSE STATUS HANDLING BEHAVIOR
Service Error
Further Meaning Behavior
Status Code
Success Success 0000 The SCP has completed the operation
successfully. Image successfully stored in
Image Box.
Warning Image size is larger than Image Box size. The B604H The N-SET operation is considered successful
image has been demagnified. but the status meaning is logged.
Warning Requested Min Density or Max Density outside B605H The N-SET operation is considered successful
of printer’s operating range. but the status meaning is logged.
Warning Image size is larger than Image Box size. The B609H The N-SET operation is considered successful
image has been cropped to fit. but the status meaning is logged.
Warning Image size or Combined Print Image Size is B60AH The N-SET operation is considered successful
larger than Image Box size. The image or but the status meaning is logged.
combined Print Image has been decimated to
fit.
Failure Image size is larger than Image Box size. C603 The Association is aborted using A-ABORT and
the print-job is marked as failed. The status
meaning is logged and reported to the user.
Failure Insufficient memory in printer to store the C605 The Association is aborted using A-ABORT and
image. the print-job is marked as failed. The status
meaning is logged and reported to the user.
Failure Combined Print Image Size is larger than Image C613 The Association is aborted using A-ABORT and
Box size. the print-job is marked as failed. The status
meaning is logged and reported to the user.
* * Any other The Association is aborted using A-ABORT and
status the print-job is marked as failed. The status
code. meaning is logged and reported to the user.

4.2.3.4 Association Acceptance Policy


The Hardcopy Application Entity does not accept Associations.

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 44 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

4.2.4 Verification Application Entity specification


4.2.4.1 SOP Class
The system provide Standard Conformance to the following SOP Class:

Table 4.2-50
SOP CLASSES FOR AE
VERIFICATION
SOP Class Name SOP Class UID SCU SCP

Verification 1.2.840.10008.1.1 Yes Yes

4.2.4.2 Association Policies


4.2.4.2.1 General
The DICOM standard application context name for DICOM 3.0 is always proposed:

Table 4.2-51
DICOM APPLICATION CONTEXT FOR AE
VERIFICATION

Application Context Name 1.2.840.10008.3.1.1.1


4.2.4.2.2 Number of Associations
The system initiate one Association at a time for a Verification request

Table 4.2-52
NUMBER OF ASSOCIATIONS INITIATED FOR AE VERIFICATION

Maximum number of Associations Initiated 1

Table 4.2-53
NUMBER OF ASSOCIATIONS ACCEPTED FOR AE VERIFICATION

Maximum number of simultaneous Associations 1

4.2.4.2.3 Asynchronous Nature


The system do not support asynchronous communication (multiple outstanding transactions over a single
Association).

Table 4.2-54
ASYNCHRONOUS NATURE AS A SCU FOR AE VERIFICATION

Maximum number of outstanding asynchronous transactions 1

4.2.4.2.4 Implementation Identifying Information


The implementation information for this Application Entity is:

Table 4.2-55
DICOM IMPLEMENTATION CLASS AND VERSION FOR AE VERIFICATION
1.2.156.112601
Implementation Class UID

Implementation Version Name DCMTK 3.6.0

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 45 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

4.2.4.3 Association Initiation Policy


4.2.4.3.1 Activity – Verify as SCU and SCP
4.2.4.3.2 Description and Sequencing of Activities
SCU: The user selecting the “Echo” button on the DICOM configuration page initiates the verification request to the
device whose data has just been configured. This tool allows the user to ensure all data (AE Title, Port and IP
Address) was correctly entered and the remote device may be contacted. It uses C-Echo and verifies the remote
device supports all configured SOP Classes. Any SOP Classes requested that are not supported will report, “failed”.
Operations may continue, but objects of the type that are not supported will not be exported. See note in 4.2.4.3.5.1
Verification SOP Class Notifications.

SCP: The system listens on the port configured on the System Settings Configuration screen for Verification requests
initiated by other remote devices. The calling device AE must already be configured as a remote device in the
system or the association is rejected.

The system initiate an Association in order to issue:

— C-ECHO request according to the Verification SOP Class.

Verification Verification
SCU SCP
Manager
1. Open Association

2. C-ECHO Request

3. C-ECHO Response

4. Close Association

Figure 4.2-6
SEQUENCING OF ACTIVITY – ISSUE VERIFY

Verification Verification
SCP SCU
Manager
1. Open Association

2. C-ECHO Request

3. C-ECHO Response

4. Close Association

Figure 4.2-7
SEQUENCING OF ACTIVITY – RECEIVE VERIFY
4.2.4.3.3 Proposed Presentation Contexts
The system will propose Presentation Contexts as shown in the following table:

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 46 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Table 4.2-56
PROPOSED PRESENTATION CONTEXTS FOR ACTIVITY VERIFICATION
Presentation Context Table
Abstract Syntax Transfer Syntax
Ext.
Name UID Name List UID List
Role Neg.
Implicit VR Little Endian 1.2.840.10008.1.2 SCU
Verification 1.2.840.10008.1.1 None
Explicit VR Little Endian 1.2.840.10008.1.2.1 /SCP

4.2.4.3.4 SOP Specific Conformance for Verification


Table 4.2-57 summarizes the behavior of the system when receiving status codes in a C-ECHO response.

A message will appear on the user interface if the system receives any other SCP response status than “Success.”

Table 4.2-57
VERIFICATION C-ECHO RESPONSE STATUS HANDLING BEHAVIOR
Service
Further Meaning Error Code Behavior
Status
Success 0000 Device Status is set to: Verified
Refused Out of Resources A700 Device Status is set to: Not Verified
Failed Unable to Process C000 – CFFF Same as “Refused” above.
* * Any other status code. Same as “Refused” above.
4.2.4.3.4 Association Acceptance Policy
4.2.4.3.5.1 Verification SOP Class Notifications
Possible Responses when “Echo” is used in Global Device configuration:

Device Echo Succeeded


Device Echo Failed
(with messages)
{SOP Class(es)}: Verified
{SOP Class(es)}: Not Verified

Note: A given “Archive” server may not support all of the SOP Classes requested in the Verification request.
Receiving failures (“Not Verified”) responses for SOP Classes outside the scope or capability of the server will not
result in a communications failure.

For example, if the correct Image Store SOP Classes are supported and Structured Report is not, then Image Storage
will work successfully. If multiframe is not supported and loops are acquired, the transfer will fail. In this case however,
the single frame images will transfer.

Association Negotiation Request message contents for each DICOM device:

Device Type SOP Classes Requested Additional Notes


Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 47 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Storage Commitment requires


configuration of a commit server
US Image Storage even if verified with the Archive
US Image Storage (Retired) device.
US Multiframe Storage
US Multiframe Storage (Retired)
DICOM Archive Server
Storage Commitment
Comprehensive Structured Report Storage

This device must be configured even


Storage Commitment
DICOM Commit Server if the Archive device indicates
Verification
commit is supported.
Modality Performed Procedure Step
DICOM PPS Server
Verification
MWL query settings are located in
the Global System configuration
Modality Worklist page, the Modality Worklist Device
DICOM Worklist Server
Verification Selection Page and in the Patient
Search window of the Patient Data
Entry screen.
All bw printers configure this entry.
Basic Grayscale META Print If the printer supports both BW and
DICOM BW Printer
Verification Color, then this must be configured
to allow BW on that printer.
Basic Color META Print May be the same printer if color is
DICOM Color Printer
Verification also supported.

4.3 NETWORK INTERFACES


The system provides DICOM 3.0 TCP/IP Network Communication Support as defined in PS3.8 (Part 8) of the DICOM 3.0
Standard.

4.3.1 TCP/IP Stack Supported


Each process inherits its TCP/IP stack from the system’s operating systems TCP/IP stack. The local AE Port number is
always set to 104.

4.3.2 Physical Network Interface


The system supports a single network interface. The connection speed is depending on the following physical network
interfaces. One of the following physical network interfaces will be available depending on installed hardware options:

Table 4.3-1
SUPPORTED PHYSICAL NETWORK INTERFACES
Ethernet 10/100/1000*BaseT, RJ-45, UTP, STP; AutoDetect Speed, Full or Half Duplex
4.3.3 IPv4 and IPv6 Support
This product only supports IPv4 connections.

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 48 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

4.4 CONFIGURATION
DICOM and networking parameters can be configured for both the local, the system device and remote DICOM service
class providers through the System Settings DICOM Menu.
4.4.1 Local
The system local network parameters are configurable. The following network parameters can be configured for a device:
• Host Name
• IP address
• Subnet IP mask
• Default Gateway
• DICOM Application Entity Title
4.4.2 Remote
Multiple DICOM service class providers can be configured through the system presets.
4.4.2.1 DICOM Store Configuration
Configuration of DICOM Modality Worklist remote devices is enabled in the system.
• DICOM Device Application Entity Title
• IP address
• Port Number
• Max retries, Retry interval, Timeout
• Send After Every Image Stored or Send At End Of Exam
4.4.2.2 DICOM Storage Commitment Configuration
Configuration of DICOM Storage Commitment remote devices must be performed separately from DICOM Store
Configuration. The system supports Storage Commitment to the same remote device as Store or to a different device.
• DICOM Device Application Entity Title
• IP address
• Port Number
• Max retries, Retry interval, Timeout
4.4.2.3 DICOM Modality Worklist Configuration
Configuration of DICOM Modality Worklist remote devices is enabled in the system.
• DICOM Device Application Entity Title
• IP address
• Port Number
• Max retries, Retry interval, Timeout
 • Maximum number of downloaded entries

4.4.2.4 DICOM Modality Performed Procedure Step Configuration


Configuration of DICOM Modality Performed Procedure Step remote devices must be performed separately from DICOM
Modality Worklist Configuration. The ACUSON X300 supports MPPS to the same remote device as Modality Worklist or to
a different device.
• DICOM Device Application Entity Title
• IP address
• Port Number
• Max retries, Retry interval, Timeout
4.4.2.5 DICOM Print Configuration
For each DICOM Print server, the following data is configurable by the user using the System Settings DICOM Print User
Interface. The effect of changing parameters of the DICOM Print server will be seen at the next created film sheet. The
current film sheet is not affected by changing these parameters.
• DICOM Device Application Entity Title
• IP address
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 49 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

• Port Number
• Max retries, Retry interval, Timeout
• Configuration for each job according to attribute description in Section 4.2.3.3.1.5 of this document

5 MEDIA INTERCHANGE

5.1 IMPLEMENTATION MODEL


5.1.1 Application Data Flow

Export
to Media
Media Media
Application Storage
Entity Medium
Update
to Media

Figure 5.1-1
APPLICATION DATA FLOW DIAGRAM FOR MEDIA STORAGE

— The Media Application Entity exports images and Structured Reports to a disk Storage medium. It is associated with
the local real-world activity “Export to Media” and “Update to Media”. “Export to Media” and “Update to Media” is
performed upon user request for selected patients, studies, series or instances (images or Structured Reports).

5.1.2 Functional Definition of AE’s


5.1.2.1 Functional Definition of Media Application Entity
Activation of the “Export to Media” and “Update to Media” icon or menu entry will pass the currently selected patients,
studies, series or instances (images or Structured Reports) to the Media Application Entity. The SOP Instances
associated with the selection will be collected into one or more export jobs. The contents of each export job will be written
to the installed media. If the capacity of a disk is exceeded, the user is provided a dialog, stating capacity exceeded and to
insert another disk.

5.1.3 Sequencing of Real-World Activities


At least one image or Structured Report must exist and be selected before the Media Application Entity can be invoked.
The operator can insert a new media at any time before or after invokation of the Media Application Entity. The Media
Application Entity will wait indefinitely for a media to be inserted before starting to write to the device. If no media is
available the export job can be canceled from the job queue.

5.1.4 File Meta Information Options


The implementation information written to the File Meta Header in each file is:

Table 5.1-1
DICOM IMPLEMENTATION CLASS AND VERSION FOR MEDIA STORAGE
1.2.156.112601
Implementation Class UID

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 50 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Implementation Version Name DCMTK 3.6.0

5.2 AE SPECIFICATIONS
5.2.1 Media Application Entity Specification
The Media Application Entity provides standard conformance to the Media Storage Service Class. The Application
Profiles and roles are listed below:

Table 5.2-1
APPLICATION PROFILES, ACTIVITIES AND ROLES FOR MEDIA
Application Profiles Supported Real World Activity Role
STD-US-ID-SF&MF-CDR Export to Media FSC
STD-US-ID-SF&MF-DVD Update to Media FSU
STD-US-SC-SF&MF-CDR
STD-US-SC-SF&MF-DVD
STD-GEN-USB

Update functionality requires DVD+RW or CDR+RW

5.2.1.1 File Meta Information for the Application Entity


The Source Application Entity Title included in the File Meta Header is configurable (see section 5.4).

5.2.1.2 Real-World Activities


5.2.1.2.1 Activity – Export to Media
The Media Application Entity acts as an FSC when requested to export SOP Instances from the local database to media.

A dialogue will be presented allowing the user to modify the suggested media label and provides control over the available
media capacity. If the contents of the current selection do not fit on a single media an automatic separation into multiple
export jobs will be suggested which can be adapted by the user.

The user will be prompted to insert an empty media for each export job. The contents of the export job will be written
together with a corresponding DICOMDIR to media. Writing in multi-session mode is not supported. The user can cancel an
export job in the job queue.

5.2.1.2.2 Activity –Update to Media


The Media Application Entity acts as an FSU using the interchange option when requested to export SOP Instances from
the local database to media upon which DICOM data already resides.

The system user selects exams from the system’s directory for transfer to media that already contains data. The
DICOMDIR is updated allowing access to original and new data.

DVD+RW or CDR+RW media may be erased at any time, removing all previously recorded data.

5.2.1.2.1.1 Media Storage Application Profiles


See Table 5.2-1 for supported Application Profiles.

5.2.1.2.1.1.1 Options
The Media Application Entity supports the SOP Classes and Transfer Syntaxes listed in the Table below:

Table 5.2-2
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 51 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

IODS, SOP CLASSES AND TRANSFER SYNTAXES FOR OFFLINEMEDIA

Information Object Definition SOP Class UID Transfer Syntax Transfer Syntax UID

Explicit VR Little
Media Storage Directory Storage 1.2.840.10008.1.3.10 1.2.840.10008.1.2.1
Endian

Explicit VR Little
Ultrasound Image Storage 1.2.840.10008.5.1.4.1.1.6.1 1.2.840.10008.1.2.1
Endian

Explicit VR Little 1.2.840.10008.1.2.1


UID_Ultrasound Multi-frame Image Endian
1.2.840.10008.5.1.4.1.1.3.1
Storage RLE Lossless 1.2.840.10008.1.2.5
JPEG Lossy Baseline 1.2.840.10008.1.2.4.50

Comprehensive Structured Report Explicit VR Little


1.2.840.10008.5.1.4.1.1.88.33 1.2.840.10008.1.2.1
Storage Endian

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 52 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

6 SUPPORT OF CHARACTER SETS

All the system DICOM applications support the

ISO_IR 100 (ISO 8859-1:1987 Latin Alphabet No. 1 supplementary set)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 53 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

7 SECURTIY

The system does not support any specific security measures.

It is assumed that the system is used within a secured environment. It is assumed that a secured environment includes at
a minimum:

a. Firewall or router protections to ensure that only approved external hosts have network access to the system.
b. Firewall or router protections to ensure that the system only has network access to approved external hosts and
services.
c. Any communication with external hosts and services outside the locally secured environment use appropriate secure
network channels (e.g. such as a Virtual Private Network (VPN))
Other network security procedures such as automated intrusion detection may be appropriate in some environments.
Additional security features may be established by the local security policy and are beyond the scope of this conformance
statement.

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 54 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

8 ANNEXES

8.1 IOD CONTENTS


8.1.1 Created SOP Instances
Table 8.1-1 specifies the attributes of an Ultrasound Image transmitted by the system storage application.

Table 8.1-17 specifies the attributes of a Comprehensive Structured Report transmitted by the system storage application.

The following tables use a number of abbreviations. The abbreviations used in the “Presence of …” column are:

VNAP Value Not Always Present (attribute sent zero length if no value is present)
ANAP Attribute Not Always Present
ALWAYS Always Present
EMPTY Attribute is sent without a value

The abbreviations used in the “Source” column:

MWL the attribute value source Modality Worklist


USER the attribute value source is from User input
AUTO the attribute value is generated automatically
MPPS the attribute value is the same as that use for Modality Performed Procedure Step
CONFIG the attribute value source is a configurable parameter

NOTE: All dates and times are encoded in the local configured calendar and time. Date, Time and Time zone are configured using
the Service/Installation Tool.

8.1.1.1 US or US Multiframe Image IOD


Table 8.1-1
IOD OF CREATED RF SOP INSTANCES
IE Module Reference Presence of Module
Patient Patient Table 8.1-2 ALWAYS
Study General Study Table 8.1-3 ALWAYS
Patient Study Table 8.1-4 ALWAYS
Series General Series Table 8.1-5 ALWAYS
Frame of Frame of Reference Not Used
Reference
US Frame of Reference Not Used
Equipment General Equipment Table 8.1-6 ALWAYS
Image General Image Table 8.1-7 ALWAYS

Image Pixel Table 8.1-8 ALWAYS

Contrast/bolus Table 8.1-9 Only if Multi-frame

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 55 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Cine Table 8.1-10

Multi-frame Table 8.1-11

Palette Color Lookup Table Table 8.1-12 Only if Multi-frame

US Region Calibration Table 8.1-13 ALWAYS

US Image Table 8.1-14 ALWAYS

Overlay Plane Not used

VOI LUT Table 8.1-15 ALWAYS

SOP Common Table 8.1-16 ALWAYS

Table 8.1-2
PATIENT MODULE
Presence of
Attribute Name Tag VR Value Source
Value
From Modality Worklist or user input.
Values supplied via Modality Worklist
will be entered as received. Values MWL/
Patient’s Name (0010,0010) PN VNAP
supplied via user input will contain all 5 USER
components (some possibly empty). .
Maximum 64 characters.
From Modality Worklist or user input. MWL/
Patient ID (0010,0020) LO VNAP
Maximum 64 characters. USER
MWL/
Patient’s Birth Date (0010,0030) DA From Modality Worklist or user input VNAP
USER
MWL/
Patient’s Sex (0010,0040) CS From Modality Worklist or user input VNAP
USER
Referenced Patient (0008,1120) Not Used
SQ
Sequence
>Referenced SOP Class (0008,1150) Not Used
UI
UID
>Referenced SOP (0008,1155) Not Used
UI
Instance UID
Patient's Birth Time (0010,0032) TM Not Used
Other Patient IDs (0010,1000) LO Not Used
Other Patient Names (0010,1001) PN Not Used
Ethnic Group (0010,2160) SH Not Used
From User Input. As entered in the
Comments field on the New Study
Patient Comments (0010,4000) LT VNAP USER
screen but not sent if zero length.
Maximum 1024 characters.

Table 8.1-3
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 56 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

GENERAL STUDY MODULE


Presence of
Attribute Name Tag VR Value Source
Value
From Modality Worklist or generated by MWL/
Study Instance UID (0020,000D) UI ALWAYS
device AUTO
Study Date (0008,0020) DA <yyyymmdd> ALWAYS AUTO
Study Time (0008,0030) TM <hhmmss> ALWAYS AUTO
Referring Physician’s
(0008,0090) PN From Modality Worklist VNAP MWL
Name
Requested Procedure ID from Worklist MWL/
Study ID (0020,0010) SH VNAP
or User Input USER
MWL/
Accession Number (0008,0050) SH From Modality Worklist or user input VNAP
USER
Comment text box in study list.
Study Description (0008,1030) LO VNAP USER
Maximum 1024 characters.
Physician(s) of Record (0008,1048) Not Used
Referenced Study
(0008,1110) SQ From Modality Worklist VNAP MWL
Sequence
>Referenced SOP Class
(0008,1150) UI From Modality Worklist VNAP MWL
UID
>Referenced SOP
(0008,1155) UI From Modality Worklist VNAP MWL
Instance UID
Procedure Code (0008,1032) MWL Requested Procedure Code
Sequence SQ Sequence (0032,1064) Absent if ANAP MWL
unscheduled.
>Code Value (0008,0100) SH Same value as MWL attribute VNAP MWL
>Coding Scheme
(0008,0102) SH Same value as MWL attribute VNAP MWL
Designator
>Coding Scheme Version (0008,0103) SH Same value as MWL attribute VNAP MWL
>Code Meaning (0008,0104) LO Same value as MWL attribute VNAP MWL

Table 8.1-4
PATIENT STUDY MODULE
Presence
Attribute Name Tag VR Value Source
of Value
Admitting
Diagnosis (0008,1080) LO From Modality Worklist VNAP MWL
Description
Calculated from DOB input on
Patient’s Age (0010,1010) AS ALWAYS AUTO
base of actual Date
From Modality Worklist or user MWL/
Patient’s Weight (0010,1030) DS VNAP
input USER
Additional Patient’s From Modality Worklist or user
(0010,21B0) LT VNAP MWL
History input

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 57 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Table 8.1-5
GENERAL SERIES MODULE
Presence of
Attribute Name Tag VR Value Source
Value
Modality (0008,0060) CS US ALWAYS AUTO
Series Instance UID (0020,000E) UI Generated by device ALWAYS AUTO
Series Number (0020,0011) IS Generated by device ALWAYS AUTO
Series Date (0008,0021) DA <yyyymmdd> ALWAYS AUTO
Series Time (0008,0031) TM <hhmmss> ALWAYS AUTO
Performing Physician’s Physician field in Study list. Maximum 64
(0008,1050) PN VNAP USER
Name characters.
Protocol Name (0018,1030) LO Exam Type ALWAYS AUTO
Organ from Study list. Maximum 512
Series Description (0008,103E) LO VNAP USER
characters.
Operator field in Study list. Maximum 64
Operator’s Name (0008,1070) PN VNAP USER
characters.
Referenced Study (0008,1111) SQ Not Used
Component Sequence
>Referenced SOP Class (0008,1150) UI Not Used
UID
>Referenced SOP Instance (0008,1155) UI Not Used
UID
Body Part Examined (0018,0015) CS Not Used
Patient Position (0018,5100) CS Not Used
Smallest Pixel Value in (0028,0108) US Not Used
Series or
SS
Largest Pixel Value in (0028,0109) US Not Used
Series or
SS
Referenced Performed Identifies the MPPS SOP Instance to
(0008,1111) SQ ALWAYS MPPS
Procedure Step Sequence which this image is related
>Referenced SOP Class
(0008,1150) UI MPPS SOP Class UID ALWAYS MPPS
UID
>Referenced SOP Instance
(0008,1155) UI MPPS SOP Instance UID ALWAYS MPPS
UID
Request Attributes
(0040,0275) SQ Zero or 1 item will be present ALWAYS AUTO
Sequence
>Requested Procedure ID (0040,1001) SH From Modality Worklist VNAP MWL
>Scheduled Procedure
(0040,0009) SH From Modality Worklist VNAP MWL
Step ID
>Scheduled Procedure
(0040,0007) LO From Modality Worklist VNAP MWL
Step Description
>Scheduled Protocol Code (0040,0008) SQ From Modality Worklist VNAP MWL

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 58 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Sequence
Performed Procedure Step
(0040,0253) SH Same as MPPS. ALWAYS MPPS
ID
Performed Procedure Step
(0040,0244) DA Same as MPPS ALWAYS MPPS
Start Date
Performed Procedure Step
(0040,0245) TM Same as MPPS ALWAYS MPPS
Start Time
Performed Procedure Step Same as MPPS. From user input.
(0040,0254) LO VNAP MPPS
Description Maximum 64 characters.
Performed Protocol Code
(0040,0260) SQ Same as MPPS ALWAYS MPPS
Sequence
Comments on the Same as MPPS. From user input.
(0040,0280) LO VNAP MPPS
Performed Procedure Step Maximum 64 characters.

Table 8.1-6
GENERAL EQUIPMENT MODULE
Presence
Attribute Name Tag VR Value Source
of Value
Manufacturer (0008,0070) LO SONOSCAPE ALWAYS AUTO
Concatenated Hospital Name
and Department from system
Institution Name (0008,0080) LO VNAP CONFIG
configuration (see section).
Always sent even if zero length.
Station Name (0008,1010) SH From Configuration ALWAYS CONFIG
Manufacturer’s S40, S40Exp, S30, S12, S22,
(0008,1090) LO ALWAYS AUTO
Model Name S9, S9Pro, S8Exp
Device Serial
(0018,1000) LO From Configuration ALWAYS CONFIG
Number
Software Version (0018,1020) LO From Configuration ALWAYS CONFIG
Spatial (0018,1050) Not Used
DS
Resolution
Date of Last (0018,1200) Not Used
DA
Calibration
Time of Last (0018,1201) Not Used
TM
Calibration
Pixel Padding (0028,0120) US Not Used
Value or
SS

Table 8.1-7
GENERAL IMAGE MODULE
Attribute Presence of
Tag VR Value Source
Name Value
Instance (0020,0013) IS Generated by device ALWAYS AUTO

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 59 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Number
Patient
(0020,0020) CS Zero length EMPTY AUTO
Orientation
<yyyymmdd>
Content Date (0008,0023) DA ALWAYS AUTO
Date when image was acquired
<hhmmss>
Content Time (0008,0033) TM ALWAYS AUTO
Time when image was acquired
Image Type (0008,0008) CS ORIGINAL/PRIMARY/<Analysis Type*> for
uncompressed,
DERIVED/PRIMARY/ < Analysis Type *> if
compressed
Acquisition
(0020,0012) IS Not Used ALWAYS AUTO
Number
Acquisition (0008,0022) Not Used
DA
Date
Acquisition (0008,0032) Not Used
TM
Time
Image Not Used
(0020,4000) LT VNAP USER
Comments
Referenced (0008,1140) SQ Not Used
Image
Sequence
>Referenced (0008,1150) UI Not Used
SOP Class
UID
>Referenced (0008,1155) UI Not Used
SOP Instance
UID
>Referenced (0008,1160) IS Not Used
Frame Number
Derivation (0008,2111) ST Not Used
Description
Source Image (0008,2112) SQ Not Used
Sequence
>Referenced (0008,1150) UI Not Used
SOP Class
UID
>Referenced (0008,1155) UI Not Used
SOP Instance
UID
>Referenced (0008,1160) IS Not Used
Frame Number
Images in (0020,1002) IS Not Used
Acquisition
Image (0020,4000) LT Not Used
Comments

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 60 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Quality Control (0028,0300) CS Not Used


Image
Burned In (0028,0301) CS Not Used
Annotation
Lossy Image (0028,2110) CS Present “01” if image is lossy compressed, “00” if
Compression not.
Lossy Image (0028,2112) DS Used if lossy compressed.
Compression
Ratio

Table 8.1-8
IMAGE PIXEL MODULE
Presence of
Attribute Name Tag VR Value Source
Value
Samples per Pixel (0028,0002) 1 A value of 1 for “MONOCHROME2” or a
ALWAYS AUTO
value of 3 for “RGB”
Photometric (0028,0004) 1 A value of “MONOCHROME2” or “RGB”
ALWAYS AUTO
Interpretation
Rows (0028,0010) 1 Image height in pixels:640,1024,1180 ALWAYS
Columns (0028,0011) 1 Image width in pixels:480,768,768 ALWAYS
Bits Allocated (0028,0100) 1 Always 0008H ALWAYS AUTO
Bits Stored (0028,0101) 1 Always 0008H ALWAYS AUTO
High Bit (0028,0102) 1 Always 0007H ALWAYS AUTO
Pixel (0028,0103) 1 Always 0000H
ALWAYS AUTO
Representation
Pixel Data (7FE0,0010) OW /
ALWAYS AUTO
OB
Planar (0028,0006) 1C Always 0000H
Configuration
Pixel Aspect Ratio (0028,0034) 1C Not Used
Smallest Image (0028,0106) 3 Not Used
Pixel Value
Largest Image Pixel (0028,0107) 3 Not Used
Value
Red Palette Color (0028,1101) 1C Not Used
Lookup Table
Descriptor
Green Palette Color (0028,1102) 1C Not Used
Lookup Table
Descriptor
Blue Palette Color (0028,1103) 1C Not Used
Lookup Table
Descriptor
Red Palette Color (0028,1201) 1C Not Used
Lookup Table Data
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 61 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Green Palette Color (0028,1202) 1C Not Used


Lookup Table Data
Blue Palette Color (0028,1203) 1C Not Used
Lookup Table Data

Table 8.1-9
CINE MODULE
Attribute Presence
Tag VR Value Source
Name of Value
Frame Time (0018,1063) DS Only if multi-frame. ANAP AUTO
Recommende
d Display (0008,2144) IS Only if multi-frame Same as Cine Rate ANAP AUTO
Frame Rate
Cine Rate (0018,0040) IS Only if multi-frame ANAP AUTO

Table 8.1-10
MULTI-FRAME MODULE
Attribute Presence
Tag VR Value Source
Name of Value
Number of
(0028,0008) IS Only if multi-frame ANAP AUTO
Frames

Table 8.1-11
CONTRAST/BOLUS MODULE
Presence of
Attribute Name Tag VR Value Source
Value
Contrast/Bolus
(0018,0010) US Only if multi-frame ANAP AUTO
Agent
Contrast/Bolus
(0018,0012) SQ Not used
Agent Sequence
Contrast/Bolus
(0018,1040) Not used
Route
Contrast/Bolus
Administration (0018,0014) SQ Not used
Route Sequence
Contrast/Bolus
(0018,1041) Not used
Volume
Contrast/Bolus
(0018,1042) Not used
Start Time
Contrast/Bolus
(0018,1043) Not used
Stop Time
Contrast/Bolus
(0018,1044) Not used
Total Dose

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 62 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Contrast Flow
(0018,1046) Not used
Rate(s)
Contrast Flow
(0018,1047) Not used
Duration(s)
Contrast/Bolus
(0018,1048) Not used
Ingredient
Contrast/Bolus
Ingredient (0018,1049) Not used
Concentration

Table 8.1-12
PALETTE COLOR LOOKUP TABLE MODULE
Presence of
Attribute Name Tag VR Value Source
Value
Red Palette Color Only used when reading
(0028,1101) US or SS
Lookup Table Descriptor Palette images.
Only used when reading
Green Palette Color
(0028,1102) US or SS Palette images.
Lookup Table Descriptor
Only used when reading
Blue Palette Color
(0028,1102) US or SS Palette images.
Lookup Table Descriptor
Palette Color Lookup
(0028,1199) UI Not used
Table UID
Only used when reading
Red Palette Color
(0028,1201) OW Palette images.
Lookup Table Data
Only used when reading
Green Palette Color
(0028,1202) OW Palette images.
Lookup Table Data
Only used when reading
Blue Palette Color
(0028,1203) OW Palette images.
Lookup Table Data
Segmented Red Palette
Not used
Color Lookup Table Data (0028,1221) OW

Segmented Green
Palette Color Lookup Not used
(0028,1222) OW
Table Data

Segmented Red Palette


(0028,1223) OW Not used
Color Lookup Table Data

Table 8.1-13
US REGION CALIBRATION MODULE
Presence of
Attribute Name Tag VR Value Source
Value
Sequence of Ultrasound (0018,6011) A sequence is present for
Regions SQ each region on the system
display, except for ECG

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 63 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

regions. Only when set for


“Display Area”. No scaling for
“Full Screen” images
>Region Location Min x0 (0018,6018) UL Top Left position of region.
>Region Location Min y0 (0018,601A) UL Top Left position of region
Bottom Right position of
>Region Location Max x1 (0018,601C) UL region
Bottom Right position of
>Region Location Max y1 (0018,601E) UL region
Enumerated Value.
>Physical Units X 2D Image = 0003H = CM
(0018,6024) US
Direction M mode / Doppler = 0004H =
SEC
Enumerated Value.
>Physical Units Y 2D Image = 0003H = CM
(0018,6026) US
Direction M mode = 0003H = CM
Doppler = 0007H = CM / SEC
The physical value per pixel
>Physical Delta X (0018,602C) FD
increment
The physical value per pixel
>Physical Delta Y (0018,602E) FD
increment
>Reference Pixel X0 (0018,6020) SL The X pixel value of baseline
>Reference Pixel Y0 (0018,6022) SL The Y pixel value of baseline
A bit mask 0-5 if: none, 2d, M
>Region Spatial Format (0018,6012) US mode, spectral Doppler. See
DICOM PS3.3 C.8.5.5.1.1
Enumerated Value. See
>Region Data Type (0018,6014) US
DICOM PS3.3 C.8.5.5.1.2
Bit mask. See DICOM PS3.3
>Region Flags (0018,6016) UL
C.8.5.5.1.3

Table 8.1-14
US IMAGE MODULE

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 64 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Presence of
Attribute Name Tag VR Value Source
Value
Samples Per Pixel (0028,0002) A sequence is present for each
region on the system display, except
SQ for ECG regions. Only when set for
“Display Area”. No scaling for “Full
Screen” images

Photometric (0028,0004) UL Top Left position of region.


Interpretation
Bits Allocated (0028,0100) UL Top Left position of region

Bits Stored (0028,0101) UL Bottom Right position of region

High Bit (0028,0102) UL Bottom Right position of region

Planar Configuration (0028,0006) Enumerated Value.


US 2D Image = 0003H = CM

M mode / Doppler = 0004H = SEC


Pixel Representation (0028,0103) Enumerated Value.

2D Image = 0003H = CM
US
M mode = 0003H = CM

Doppler = 0007H = CM / SEC

Frame Increment (0028,0009) The physical value per pixel


AT
Pointer increment

Image Type (0008,0008) The physical value per pixel


CS
increment

Lossy Image (0028,2110) CS The X pixel value of baseline


Compression
Number of Stages (0008,2124) IS The Y pixel value of baseline

Number of Views in (0008,212A) A bit mask 0-5 if: none, 2d, M mode,
Stage IS spectral Doppler. See DICOM PS3.3
C.8.5.5.1.1
R Wave Time Vector (0018,6060) US Enumerated Value. See DICOM
PS3.3 C.8.5.5.1.2
Ultrasound Color Data (0028,0014) US Bit mask. See DICOM PS3.3
Present C.8.5.5.1.3
Stage Name (0008,2120) SH

Stage Code Sequence (0040,000A)

>Include “Code
Sequence Macro”
Table 8.8-1.
Stage Number (0008,2122)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 65 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

View Name (0008,2127)

View Number (0008,2128)

Number of Event (0008,2129)


Timers
Event Elapsed Time(s) (0008,2130)

Event Timer Name(s) (0008,2132)

Include 'General
Anatomy Optional
Macro' Table 10-7
View Code Sequence (0054,0220)

>Include ‘Code
Sequence Macro’
Table 8.8-1
>View Modifier Code (0054,0222)
Sequence
>>Include ‘Code
Sequence Macro’
Table 8.8-1
Acquisition DateTime (0008,002A)

Trigger Time (0018,1060)

Nominal Interval (0018,1062)

Beat Rejection Flag (0018,1080)

Low R-R Value (0018,1081)

High R-R Value (0018,1082)

Heart Rate (0018,1088)

IVUS Acquisition (0018,3100)

IVUS Pullback Rate (0018,3101)

IVUS Gated Rate (0018,3102)

IVUS Pullback Start (0018,3103)


Frame Number
IVUS Pullback Stop (0018,3104)
Frame Number
Lesion Number (0018,3105)

Output Power (0018,5000)

Transducer Data (0018,5010)

Transducer Type (0018,6031)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 66 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Focus Depth (0018,5012)

Processing Function (0018,5020)

Mechanical Index (0018,5022)

Bone Thermal Index (0018,5024)

Cranial Thermal Index (0018,5026)

Soft Tissue Thermal (0018,5027)


Index

Soft Tissue-focus (0018,5028)


Thermal Index

Soft Tissue-surface (0018,5029)


Thermal Index

Depth of Scan Field (0018,5050)

Overlay Subtype (60xx,0045)

Table 8.1-15
MODALITY LUT MODULE
Attribute Presence of
Tag VR Value Source
Name Value
Modality LUT
(0028,3000) SQ present if (0028,1040) = LOG ANAP AUTO
Sequence
> LUT
(0028,3002) US <1024,0,16> ANAP AUTO
Descriptor
> Modality
(0028,3004) LO US ANAP AUTO
LUT Type
> LUT Data (0028,3006) US LUT ANAP AUTO
Window
(0028,1050) DS 0…1023 ALWAYS AUTO
Center
Window Width (0028,1051) DS 1…1024 ALWAYS AUTO

Table 8.1-16
SOP COMMON MODULE
Attribute Presence of
Tag VR Value Source
Name Value
Specific
(0008,0005) CS “ISO_IR 100” ALWAYS CONFIG
Character Set
SOP Class 1.2.840.10008.5.1.4.1.1.6.1
(0008,0016) UI ALWAYS AUTO
UID 1.2.840.10008.5.1.4.1.1.6
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 67 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

1.2.840.10008.5.1.4.1.1.7
SOP Instance
(0008,0018) UI Generated by device ALWAYS AUTO
UID
Instance (0008,0012) <yyyymmdd>
Creation Date DA ALWAYS AUTO
Date when image was acquired
Instance (0008,0013) <hhmmss>
Creation TM ALWAYS AUTO
Time Time when image was acquired

Instance (0008,0014) Not used


UI
Creator UID
Instance
(0020,0013)
Number
8.1.1.2 Comprehensive Structured Report IOD
Table 8.1-17
IOD OF CREATED COMPREHENSIVE STRUCTURED REPORT SOP INSTANCES
IE Module Reference Presence of Module
Patient Patient Table 8.1-18 ALWAYS
Study General Study Table 8.1-19 ALWAYS
Patient Study Table 8.1-20 ALWAYS
Series General Series Table 8.1-21 ALWAYS
Equipment General Equipment Table 8.1-22 ALWAYS
Document SR Document General Table 8.1-23 ALWAYS

SR Document Content Table 8.1-24 ALWAYS

SOP Common Table 8.1-25 ALWAYS

Table 8.1-18
PATIENT MODULE
Presence of
Attribute Name Tag VR Value Source
Value
From Modality Worklist or user input.
Values supplied via Modality Worklist
will be entered as received. Values MWL/
Patient’s Name (0010,0010) PN VNAP
supplied via user input will contain all 5 USER
components (some possibly empty). .
Maximum 64 characters.
From Modality Worklist or user input. MWL/
Patient ID (0010,0020) LO VNAP
Maximum 64 characters. USER
MWL/
Patient’s Birth Date (0010,0030) DA From Modality Worklist or user input VNAP
USER
MWL/
Patient’s Sex (0010,0040) CS From Modality Worklist or user input VNAP
USER
Referenced Patient (0008,1120) SQ Not Used
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 68 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Sequence
>Referenced SOP Class (0008,1150) Not Used
UI
UID
>Referenced SOP (0008,1155) Not Used
UI
Instance UID
Patient's Birth Time (0010,0032) TM Not Used
Other Patient IDs (0010,1000) LO Not Used
Other Patient Names (0010,1001) PN Not Used
Ethnic Group (0010,2160) SH Not Used
From User Input. As entered in the
Comments field on the New Study
Patient Comments (0010,4000) LT VNAP USER
screen but not sent if zero length.
Maximum 1024 characters.

Table 8.1-19
GENERAL STUDY MODULE
Presence of
Attribute Name Tag VR Value Source
Value
From Modality Worklist or generated by MWL/
Study Instance UID (0020,000D) UI ALWAYS
device AUTO
Study Date (0008,0020) DA <yyyymmdd> ALWAYS AUTO
Study Time (0008,0030) TM <hhmmss> ALWAYS AUTO
Referring Physician’s
(0008,0090) PN From Modality Worklist VNAP MWL
Name
Requested Procedure ID from Worklist MWL/
Study ID (0020,0010) SH VNAP
or User Input USER
MWL/
Accession Number (0008,0050) SH From Modality Worklist or user input VNAP
USER
Comment text box in study list.
Study Description (0008,1030) LO VNAP USER
Maximum 1024 characters.
Physician(s) of Record (0008,1048) Not Used
Referenced Study
(0008,1110) SQ From Modality Worklist VNAP MWL
Sequence
>Referenced SOP Class
(0008,1150) UI From Modality Worklist VNAP MWL
UID
>Referenced SOP
(0008,1155) UI From Modality Worklist VNAP MWL
Instance UID
Procedure Code (0008,1032) MWL Requested Procedure Code
Sequence SQ Sequence (0032,1064) Absent if ANAP MWL
unscheduled.
>Code Value (0008,0100) SH Same value as MWL attribute VNAP MWL
>Coding Scheme
(0008,0102) SH Same value as MWL attribute VNAP MWL
Designator
>Coding Scheme Version (0008,0103) SH Same value as MWL attribute VNAP MWL

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 69 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

>Code Meaning (0008,0104) LO Same value as MWL attribute VNAP MWL

Table 8.1-20
PATIENT STUDY MODULE
Presence
Attribute Name Tag VR Value Source
of Value
Admitting
Diagnosis (0008,1080) LO From Modality Worklist VNAP MWL
Description
Calculated from DOB input on
Patient’s Age (0010,1010) AS ALWAYS AUTO
base of actual Date
From Modality Worklist or user MWL/
Patient’s Weight (0010,1030) DS VNAP
input USER
Additional Patient’s From Modality Worklist or user
(0010,21B0) LT VNAP MWL
History input

Table 8.1-21
GENERAL SERIES MODULE
Presence of
Attribute Name Tag VR Value Source
Value
Modality (0008,0060) CS US ALWAYS AUTO
Series Instance UID (0020,000E) UI Generated by device ALWAYS AUTO
Series Number (0020,0011) IS Generated by device ALWAYS AUTO
Series Date (0008,0021) DA <yyyymmdd> ALWAYS AUTO
Series Time (0008,0031) TM <hhmmss> ALWAYS AUTO
Performing Physician’s Physician field in Study list. Maximum 64
(0008,1050) PN VNAP USER
Name characters.
Protocol Name (0018,1030) LO Exam Type ALWAYS AUTO
Organ from Study list. Maximum 512
Series Description (0008,103E) LO VNAP USER
characters.
Operator field in Study list. Maximum 64
Operator’s Name (0008,1070) PN VNAP USER
characters.
Referenced Study (0008,1111) SQ Not Used
Component Sequence
>Referenced SOP Class (0008,1150) UI Not Used
UID
>Referenced SOP Instance (0008,1155) UI Not Used
UID
Body Part Examined (0018,0015) CS Not Used
Patient Position (0018,5100) CS Not Used
Smallest Pixel Value in (0028,0108) US Not Used
Series or
SS

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 70 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Largest Pixel Value in (0028,0109) US Not Used


Series or
SS
Referenced Performed Identifies the MPPS SOP Instance to
(0008,1111) SQ ALWAYS MPPS
Procedure Step Sequence which this image is related
>Referenced SOP Class
(0008,1150) UI MPPS SOP Class UID ALWAYS MPPS
UID
>Referenced SOP Instance
(0008,1155) UI MPPS SOP Instance UID ALWAYS MPPS
UID
Request Attributes
(0040,0275) SQ Zero or 1 item will be present ALWAYS AUTO
Sequence
>Requested Procedure ID (0040,1001) SH From Modality Worklist VNAP MWL
>Scheduled Procedure
(0040,0009) SH From Modality Worklist VNAP MWL
Step ID
>Scheduled Procedure
(0040,0007) LO From Modality Worklist VNAP MWL
Step Description
>Scheduled Protocol Code
(0040,0008) SQ From Modality Worklist VNAP MWL
Sequence
Performed Procedure Step
(0040,0253) SH Same as MPPS. ALWAYS MPPS
ID
Performed Procedure Step
(0040,0244) DA Same as MPPS ALWAYS MPPS
Start Date
Performed Procedure Step
(0040,0245) TM Same as MPPS ALWAYS MPPS
Start Time
Performed Procedure Step Same as MPPS. From user input.
(0040,0254) LO VNAP MPPS
Description Maximum 64 characters.
Performed Protocol Code
(0040,0260) SQ Same as MPPS ALWAYS MPPS
Sequence
Comments on the Same as MPPS. From user input.
(0040,0280) LO VNAP MPPS
Performed Procedure Step Maximum 64 characters.

Table 8.1-22
GENERAL EQUIPMENT MODULE
Presence
Attribute Name Tag VR Value Source
of Value
Manufacturer (0008,0070) LO SONOSCAPE ALWAYS AUTO
Concatenated Hospital Name
and Department from system
Institution Name (0008,0080) LO VNAP CONFIG
configuration (see section).
Always sent even if zero length.
Station Name (0008,1010) SH From Configuration ALWAYS CONFIG
Manufacturer’s
(0008,1090) LO pj0905 ALWAYS AUTO
Model Name
Device Serial
(0018,1000) LO From Configuration ALWAYS CONFIG
Number

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 71 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Software Version (0018,1020) LO From Configuration ALWAYS CONFIG


Spatial (0018,1050) Not Used
DS
Resolution
Date of Last (0018,1200) Not Used
DA
Calibration
Time of Last (0018,1201) Not Used
TM
Calibration
Pixel Padding (0028,0120) US Not Used
Value or
SS

Table 8.1-23
SR DOCUMENT GENERAL MODULE
Presence
Attribute Name Tag VR Value Source
of Value
Instance (0020,0013) IS Unique number ALWAYS AUTO
Number
Completion Flag (0040,A491) CS PARTIAL VNAP AUTO
Verification Flag (0040,A493) CS UNVERIFIED ALWAYS AUTO
Content Date (0008,0023) DA Date content created. ALWAYS AUTO
Content Time (0008,0033) TM Time content created. ALWAYS AUTO
Used when Send of
Predecessor Demand is used and Send
(0040,A360) SQ
Documents Structured Reports with
Sequence Send on Demand is ALWAYS AUTO
selected. Will indicate the
preceding SR sent in the
study.
>Study Instance (0020,000D) UI Study’s UID
ALWAYS AUTO
UID
>Referenced
(0008,1115) SQ Identifies the Series containing ALWAYS AUTO
Series
the referenced SR
Sequence
>>Reference
(0008,1199) SQ SOP Instance UID for SR ALWAYS AUTO
d SOP
Series in the study
Sequence
>>> Comprehensive SR SOP Class
(0008,1150) UI ALWAYS AUTO
Referenced 1.2.840.10008.5.1.4.1.1.88.33
SOP Class
>>>
(0008,1155) UI SOP Instance UID of the ALWAYS AUTO
Referenced
preceding SR in the study
SOP Instance
UID
Identifies Requested
Referenced ALWAYS AUTO
(0040,A370) SQ Procedures being
Request
fulfilled (completely or partially)
Sequence
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 72 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

by creation of this
Document.
>Study Instance (0020,000D) UI Same value as in MWL or auto ALWAYS AUTO
UID generated
>Referenced
(0008,1110) SQ 1 item per item in MWL, zero ALWAYS AUTO
Study
length if unscheduled
Sequence
>>Reference
(0008,1150) UI Identifies the Referenced SOP ALWAYS AUTO
d SOP Class
Class
UID
>>Reference
(0008,1155) UI Instance UID ALWAYS AUTO
d SOP
Instance UID

Current (0040,A375) SQ ALWAYS AUTO


Requested
Procedure
Evidence
Sequence
>Study Instance (0020,000D) UI Same value as in MWL or auto ALWAYS AUTO
UID
generated
>Referenced (0008,1115) SQ ALWAYS AUTO
Series Sequence
>>Series 020,000E) UI Generated by device ALWAYS AUTO
Instance UID

Table 8.1-24
SR DOCUMENT CONTENT MODULE
Presence
Attribute Name Tag VR Value Source
of Value
Content (0040,A504) SQ
Template ALWAYS AUTO
Sequence
The Root Content Item
>Template (0040,DB00) CS identifies TID 5000 (OB- ALWAYS AUTO
Identifier GYN), 5100 (Vascular).
5200 (Echo) or SONO1000
(Other).
>Mapping (0008,0105) CS DCMR ALWAYS AUTO
Resource
Content (0040,A730) SQ ALWAYS AUTO
Sequence
See Template ID
5000 for
ALWAYS AUTO
>Relationship (0040,A010) CS OB-GYN,
Type Template ID
5100 for
Vascular
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 73 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Template ID
5200 for Adult
Echo and
Template ID
SONO1000 for
Other
See Template ID ANAP AUTO
Document 5000 for
Relationship OB-GYN,
Macro Table Template ID
5100 for
Vascular
Template ID
5200 for Adult
Echo and
Template ID
SONO1000 for
Other
See Template ID
Document 5000 for
ALWAYS AUTO
Content OB-GYN,
Macro Template ID
5100 for
Vascular
Template ID
5200 for Adult
Echo and
Template ID
SONO1000 for
Other
Value Type (0040,A040) CS CONTAINER, ALWAYS AUTO
always first tag
of SR
Additional
values used:
TEXT, NUM, DATE, CODE,
IMAGE, SCOORD
Concept Name
(0040,A043) SQ ALWAYS AUTO
Code
Sequence
>Code Value (0008,0100) 125000, 125100 125200 or ALWAYS AUTO
SONO1000
>Coding
(0008,0102) DCM ALWAYS AUTO
Scheme
Designator
“OB-GYN Ultrasound
Procedure Report”,
>Code Meaning (0008,0104) ALWAYS AUTO
“Vascular Ultrasound
Procedure Report”,
“Adult
Echocardiography
Procedure Report”
Or Other Procedure Report
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 74 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Continuity of (0040,A050) CS SEPARATE ALWAYS AUTO


Content
See Template ID
Numeric 5000 for
Measurement OB-GYN, ALWAYS AUTO
Macro Template ID
5100 for
Vascular
Template ID
5200 for Adult
Echo and
Template ID
SONO1000 for
Other
See Template ID
5000 for
Code Macro OB-GYN, ALWAYS AUTO
Template ID
5100 for
Vascular
Template ID
5200 for Adult
Echo and
Template ID
SONO1000 for
Other

Table 8.1-25
SOP COMMON MODULE
Presence
Attribute Name Tag VR Value Source
of Value
SOP Class UID (0008,0016) UI 1.2.840.10008.5.1.4.1.1.88.33 ALWAYS AUTO
SOP Instance (0008,0018) UI Generated by device
ALWAYS AUTO
UID
Specific
(0008,0005) CS ISO_IR 100
Character ALWAYS CONFIG
Set

8.1.2 Used Fields in received IOD by application


The system storage application does not receive SOP Instances. The usage of attributes received via Modality Worklist is
described in section 4.2.2.3.1.3.

8.1.3 Attribute mapping


The relationships between attributes received via Modality Worklist, stored in acquired images and communicated via
MPPS are summarized in Table 8.1-17. The format and conventions used in Table 99Table 99 are the same as the
corresponding table in DICOM Part 4, Annex M.6 [DICOM].

Table 8.1-31
ATTRIBUTE MAPPING BETWEEN MODALITY WORKLIST, IMAGE AND MPPS
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 75 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Modality Worklist Image IOD MPPS IOD


Patient Name Patient Name Patient Name
Patient ID Patient ID Patient ID
Patient’s Birth Date Patient’s Birth Date Patient’s Birth Date
Patient’s Sex Patient’s Sex Patient’s Sex
Patient’s Weight Patient’s Weight
Referring Physician’s Name Referring Physician’s Name
---- ---- Scheduled Step Attributes Sequence
Study Instance UID Study Instance UID >Study Instance UID
Referenced Study Sequence Referenced Study Sequence >Referenced Study Sequence
Accession Number Accession Number >Accession Number
---- Request Attributes Sequence ----
Requested Procedure ID >Requested Procedure ID >Requested Procedure ID
Requested Procedure Description >Requested Procedure Description
Scheduled Procedure Step ID >Scheduled Procedure Step ID >Scheduled Procedure Step ID
Scheduled Procedure Step >Scheduled Procedure Step Description >Scheduled Procedure Step Description
Description
Scheduled Protocol Code Sequence >Scheduled Protocol Code Sequence ----
---- Performed Protocol Code Sequence Performed Protocol Code Sequence
---- Study ID Study ID
---- Performed Procedure Step ID Performed Procedure Step ID
---- Performed Procedure Step Start Date Performed Procedure Step Start Date
---- Performed Procedure Step Start Time Performed Procedure Step Start Time
---- Performed Procedure Step Description Performed Procedure Step Description
---- Comments on the Performed Procedure Comments on the Performed Procedure
Step Step
---- ---- Performed Series Sequence
Scheduled Performing Physician’s Performing Physician’s Name >Performing Physician’s Name
Name
Requested Procedure Code ---- Procedure Code Sequence
Sequence
---- Referenced Study Component ----
Sequence
---- >Referenced SOP Class UID SOP Class UID
---- >Referenced SOP Instance UID SOP Instance UID
---- Protocol Name Protocol Name

8.1.4 Coerced/Modified Fields


The Modality Worklist AE will truncate attribute values received in the response to a Modality Worklist Query if the value
length is longer than the maximum length permitted by the attribute’s VR.

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 76 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

8.2 CODED TERMINOLOGY AND TEMPLATES


The Workflow AE is capable of supporting arbitrary coding schemes for Procedure and Protocol Codes. The contents of
Requested Procedure Code Sequence (0032,1064) and Scheduled Protocol Code Sequence (0040,0008) supplied in
Worklist Items will be mapped to Image IOD and MPPS attributes as described in Table 8.1-31. During installation, a
service technician will establish a mapping between the site-specific codes and the Protocol Names used internally to
identify acquisition protocols.

The contents of Performed Procedure Step Discontinuation Reason Code Sequence (0040,0281) for a discontinued MPPS
will be filled with a code selected by the user from a fixed list corresponding to Context Group 9300.

8.4 GRAYSCALE IMAGE CONSISTENCY


The high resolution display monitor attached to the system can be calibrated according to the Grayscale Standard Display
Function (GSDF). The Service/Installation Tool is used together with a luminance meter to measure the Characteristic
Curve of the display system and the current ambient light. See the system Service Manual for details on the calibration
procedure and supported calibration hardware. The result of the calibration procedure is a Monitor Correction LUT that will
be active within the display subsystem after a system reboot.

8.5 STANDARD EXTENSDED / SPECIALIZED / PRIVATE SOP CLASSES


No Specialized or Private SOP Classes are supported.

8.5.1 Ultrasound Image Storage SOP Class


The Ultrasound Image Storage SOP Class is extended to create a Standard Extended SOP Class by addition of standard
and private attributes to the created SOP Instances as documented in section 8.1.

8.6 PRIVATE TRANSFER SYNTAXES


No Private Transfer Syntaxes are supported.

APPENDIX A – Structured Reports

A.1 STRUCTURED REPORTS


A.1.1 Introduction
The system implement Structured Report Templates TID 5000 (OB-GYN), 5100 (Vascular) and 5200 (Echo) from
DICOM Part 16 and SONO1000 (Others) .This Appendix describes the manner that the system measurements appear
in DICOM reports.

This Appendix contains tables of the measurement and calculations displayed on the system calculation screens and the
data dictionary of all code values, coding scheme designators and code meanings associated with those measurements
and labels. There is no attempt to provide all tags that will be sent in the SR object. For those specifics, please refer to
DICOM PS3-3, 3-16 and 3-17 for complete discussion on Comprehensive SOP Class support and the Templates
described here.
Part 17 of the DICOM Standard includes tree diagrams showing graphic examples of the structure of each template.
Briefly, an SR document will contain only the measurements, calculations and observations made during the exam. Its
exact structure is therefore determined by two main components, the measurements that are available within the context
of the template and its referenced Templates and Context groups, and those measurements and calculations that are
implemented on the system creating the report. Supplementing these constraints are private and user defined
measurements and calculations, which may be added if the Root Container Template is extensible.

In the system, the calculation packages contain a combination of template and user defined measurements and
calculations. This Appendix contains a description of the calculation packages on each system and the related templates
that support them. See the table in the Clinical Scope section below.
Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 77 of 145
SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

In each template section, there is a brief description of the mapping of measurement and calculation labels as they
appear on the system when the Analysis feature is enabled.

A.2 TID 5000 OB-GYN ULTRASOUND PROCEDURE REPORT


A.2. 1 TID 5001 OB-GYN Patient Characteristics
Label Code Meaning (Coding Scheme Designator, Code
Value)

Patient Characteristics Container: Patient Characteristics (DCM,121118)

Height Patient Height (LN, 8302-2)

Weight Patient Weight (LN, 29463-7)

Gravida Gravida (LN, 11996-6)

Para Para (LN, 11977-6)

Aborta Aborta (LN, 11612-9)

Ectopic Ectopic Pregnancies (LN, 33065-4)

Comments Comment (DCM, 121106)

A.2. 2 TID 5002 OB-GYN Procedure Summary Section


Label Code Meaning (Coding Scheme Designator, Code
Value)

OB-GYN Summary Container: Summary (DCM, 121111)

LMP LMP (LN, 11955-2)

NumberOfFetuses Number of Fetuses(LN, 11878-6)

EDD EDD (LN, 11778-8)

Comments Comment (DCM, 121106)

A.2. 3 TID 5003 OB-GYN Fetus Summary


Label ItemID Code Meaning (Coding Scheme
Designator, Code Value)

Fetus Summary Container: Fetus Summary


(125008, DCM)

Estimated Fetal Weight Estimated Weight (LN, 11727-5)

HADLOCK1(AC,FL) OB/AC/FL(Hadlock1)/EFW EFW by AC, FL, Hadlock 1984 (LN,


11750-7)

HADLOCK2(BPD,AC,FL) OB/BPD/AC/FL(Hadlock2)/EFW EFW by AC, BPD, FL, Hadlock 1985


(LN, 11735-8)

HADLOCK3(HC,AC,FL) OB/HC/AC/FL(Hadlock3)/EFW EFW by AC, FL, HC, Hadlock 1985

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 78 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

(LN, 11746-5)

Hadlock4(BPD,HC,AC,FL) OB/BPD/HC/AC/FL(Hadlock4)/EFW EFW by AC, BPD, FL, HC, Hadlock


1985 (LN, 11732-5)

Shepard(AC,BPD) OB/AC/BPD(Shepard)/EFW EFW by AC, BPD, Shepard 1982


(LN, 11739-0)

Campbell(AC) OB/AC(Campbell)/EFW EFW by AC, Campbell


1975(LN,11756-4)

Hansmann(BPD,TTD) OB/BPD/TTD(Hansmann)/EFW EFW by BPD, TTD, Hansmann


1986 (LN, 33139-7)

AUA OB/GA(AUA) Composite Ultrasound Age (LN,


11888-5)

CUA OB/GA(CUA) Composite Ultrasound Age (LN,


11888-5)

OB/OB_MHR/HR Fetal Heart Rate (LN, 11948-7)

Heart Rate OB/D_FHR/FHR Fetal Heart Rate (LN, 11948-7)

OB/M_FHR/FHR Fetal Heart Rate (LN, 11948-7)

A.2. 4 TID 5004 Fetal Biometry Ratio Section


Label Code Meaning (Coding Scheme Designator, Code
Value)

Fetal Biometry Ratios Container: Fetal Biometry Ratios (DCM, 125001)

A.2. 5 TID 5005 Fetal Biometry Section


Label Item ID Code Meaning (Coding Scheme
Designator, Code Value)

Fetal Biometry Container: Fetal Biometry (DCM, 125002)


Measurements

AC OB/AC/AC Abdominal Circumference (LN, 11979-2)

OB/ASUM/AC Abdominal Circumference (LN, 11979-2)


ASUM
OB/ASUM/AC-GA AC, ASUM 2000(LN, 33072-0)

OB/Merz/AC Abdominal Circumference (LN, 11979-2)


Merz
OB/Merz/AC-GA AC, Merz 1988(LN, 33075-3)

CFEF OB/CFEF/AC Abdominal Circumference (LN, 11979-2)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 79 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

OB/CFEF/AC-GA AC, CFEF(SONO, SONO-5005-1)

OB/Campbell /AC Abdominal Circumference (LN, 11979-2)


Campbell
OB/Campbell /AC-GA AC,Campbell 1975(LN,11889-3)

OB/Hadlock/AC Abdominal Circumference (LN, 11979-2)


Hadlock
OB/Hadlock/AC-GA AC, Hadlock 1984(LN, 11892-7)

OB/Hansmann/AC Abdominal Circumference (LN, 11979-2)


Hansmann
OB/Hansmann/AC-GA AC, Hansmann 1985(LN,33073-8)

OB/Tokyo/AC Abdominal Circumference (LN, 11979-2)


Tokyo
OB/Tokyo/AC-GA AC, Tokyo (SONO, SONO-5005-2)

APTD OB/APTD /APTD Anterior-Posterior Trunk Diameter


(LN,11819-0)

OB/Hansmann/APTD Anterior-Posterior Trunk Diameter


(LN,11819-0)
Hansmann
OB/Hansmann/APTD-GA APTD, Hansmann (SONO, SONO-5005-4)

BPD OB/BPD Biparietal Diameter (LN, 11820-8)

OB/ASUM/BPD Biparietal Diameter (LN, 11820-8)


ASUM
OB/ASUM/BPD-GA BPD, ASUM 1989(LN, 33079-5)

OB/Hadlock/BPD Biparietal Diameter (LN, 11820-8)


Hadlock
OB/Hadlock/BPD-GA BPD, Hadlock 1984(LN, 11902-4)

OB/Hansmann/BPD Biparietal Diameter (LN, 11820-8)


Hansmann
OB/Hansmann/BPD-GA BPD, Hansmann 1986(LN, 33538-0)

OB/CFEF/BPD Biparietal Diameter (LN, 11820-8)


CFEF
OB/CFEF/BPD-GA BPD, CFEF(SONO, SONO-5005-8)

OB/Campbell/BPD Biparietal Diameter (LN, 11820-8)


Campbell
OB/Campbell/BPD-GA BPD, Campbell (SONO, SONO-5005-7)

OB/Johnsen/BPD Biparietal Diameter (LN, 11820-8)


Johnsen
OB/Johnsen/BPD-GA BPD, Johnsen(SONO, SONO-5005-9)

Kurtz OB/Kurtz/BPD Biparietal Diameter (LN, 11820-8)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 80 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

OB/Kurtz/BPD-GA BPD, Kurtz 1980(LN, 11906-5)

OB/Jeanty/BPD Biparietal Diameter (LN, 11820-8)


Jeanty
OB/Jeanty/BPD-GA BPD, Jeanty 1984(LN, 11905-7)

OB/Sabbagha /BPD Biparietal Diameter (LN, 11820-8)


Sabbagha
OB/Sabbagha /BPD-GA BPD, Sabbagha 1978(LN, 11907-3)

OB/Merz/BPD Biparietal Diameter (LN, 11820-8)


Merz
OB/Merz/BPD-GA BPD,Merz 1988(LN, 33081-1)

OB/Tokyo/BPD Biparietal Diameter (LN, 11820-8)


Tokyo
OB/Tokyo/BPD-GA BPD, Tokyo 1986(LN, 33085-2)

OB/Osaka/BPD Biparietal Diameter (LN, 11820-8)


Osaka
OB/Osaka/BPD-GA BPD, Osaka 1989(LN, 33082-9)

OB/Rempen/BPD Biparietal Diameter (LN, 11820-8)


Rempen
OB/Rempen/BPD-GA BPD, Rempen 1991(LN, 33083-7)

OB/HC/HC Head Circumference (LN, 11984-2)


HC
OB/HC/OFD(HC) Occipital-Frontal Diameter(LN, 11851-3)

OB/Hadlock/HC Head Circumference (LN, 11984-2)

Hadlock OB/Hadlock/OFD(HC) Occipital-Frontal Diameter(LN, 11851-3)

OB/Hadlock/HC-GA HC by GA, Hadlock 1984(LN, 33173-6)

OB/ASUM/HC Head Circumference (LN, 11984-2)

ASUM OB/ASUM/OFD(HC) Occipital-Frontal Diameter(LN, 11851-3)

OB/ASUM/HC-GA HC, ASUM(SONO, SONO-5005-23)

OB/CFEF/HC Head Circumference (LN, 11984-2)

CFEF OB/CFEF/OFD(HC) Occipital-Frontal Diameter(LN, 11851-3)

OB/CFEF/HC-GA HC, CFEF(SONO, SONO-5005-24)

OB/Chitty /HC Head Circumference (LN, 11984-2)

Chitty OB/Chitty / OFD(HC) Occipital-Frontal Diameter(LN, 11851-3)

OB/ Chitty /HC-GA HC derived by GA, Chitty 1994 (LN, 33174-4)

Hansmann OB/Hansmann/HC Head Circumference (LN, 11984-2)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 81 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

OB/Hansmann/OFD(HC) Occipital-Frontal Diameter(LN, 11851-3)

OB/Hansmann/HC-GA HC, Hansmann 1986(LN, 33543-0)

OB/Merz/HC Head Circumference (LN, 11984-2)

Merz OB/Merz/OFD(HC) Occipital-Frontal Diameter(LN, 11851-3)

OB/Merz/HC-GA HC Merz, 1988(LN, 33115-7)

OB/Jeanty/HC Head Circumference (LN, 11984-2)

Jeanty OB/Jeanty/OFD(HC) Occipital-Frontal Diameter(LN, 11851-3)

OB/Jeanty/HC-GA HC, Jeanty 1982 (LN, 33113-2)

OB/Johnsen/HC Head Circumference (LN, 11984-2)

Johnsen OB/Johnsen/OFD(HC) Occipital-Frontal Diameter(LN, 11851-3)

OB/Johnsen/HC-GA HC, Jeanty 1982 (LN, 33113-2)

OFD OB/OFD/OFD Occipital-Frontal Diameter (LN, 11851-3)

OB/Hansmann/OFD Occipital-Frontal Diameter (LN, 11851-3)


Hansmann
OB/Hansmann/OFD-GA OFD, Hansmann 1985 (LN, 33544-8)

OB/ASUM/OFD Occipital-Frontal Diameter (LN, 11851-3)


ASUM
OB/ASUM/OFD-GA OFD, ASUM 2000(LN, 33119-9)

OB/Jeanty/OFD Occipital-Frontal Diameter (LN, 11851-3)


Jeanty
OB/Jeanty/OFD-GA OFD, Jeanty(SONO, SONO-5005-26)

OB/TAD/TAD Transverse Abdominal Diameter(LN,


TAD
11862-0)

OB/CFEF/TAD Transverse Abdominal Diameter(LN,


CFEF 11862-0)

OB/CFEF/TAD-GA TAD, CFEF(SONO, SONO-5005-29)

OB/Merz/TAD Transverse Abdominal Diameter(LN,


Merz 11862-0)

OB/Merz/TAD-GA TAD, Merz(SONO, SONO-5005-30)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 82 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Cereb OB/Cereb/Cereb Trans Cerebellar Diameter (LN, 11863-8)

OB/Hill/Cereb Trans Cerebellar Diameter (LN, 11863-8)


Hill
OB/Hill/Cereb-GA Cerebellum, Hill(SONO, SONO-5005-12)

OB/Chitty/Cereb Trans Cerebellar Diameter (LN, 11863-8)


Chitty
OB/Chitty/Cereb-GA Cerebellum, Chitty(SONO, SONO-5005-11)

OB/TTD/TTD Transverse Thoracic Diameter (LN,


TTD
11864-6)

OB/Hansmann/TTD Transverse Thoracic Diameter (LN, 11864-6)


Hansmann
OB/Hansmann/TTD-GA Transverse Thoracic Diameter,Hansmann
1985(LN, 33136-3)

Foot OB/Foot/Foot Foot Length (LN, 11965-1)

OB/Chitty/Foot Foot Length (LN, 11965-1)


Chitty
OB/Chitty/Foot-GA Foot Length, chitty(SONO, SONO-5005-20)

OB/ APAD /APAD Anterior/Posterior Thoracic Diameter(LN,


APAD
11818-2)

OB/Merz/APAD Anterior/Posterior Thoracic Diameter(LN,


11818-2)
Merz
OB/Merz/APAD-GA APAD, Merz(SONO, SONO-5005-3)

BOD OB/BOD/BOD Binocular Distance(SONO, SONO-5005-5)

OB/Jeanty/BOD Binocular Distance(SONO, SONO-5005-5)


Jeanty
OB/Jeanty/BOD-GA BOD, Jeanty(SONO, SONO-5005-6)

OB/CSP/c.s.p Cavum Septum Pellicidum(SONO,


CSP
SONO-5005-10)

CM OB/CM/CM Cisterna Magna length(LN, 11860-4)

OB/Nicolaides/CM Cisterna Magna length(LN, 11860-4)


Nicolaides
OB/Nicolaides/CM-GA CM by GA, Nicolaides(SONO,

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 83 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

SONO-5005-42)

FTA OB/FTA/FTA Fetal Trunk Area(SONO, SONO-5005-21)

OB/Osaka/FTA Fetal Trunk Area(SONO, SONO-5005-21)


Osaka
OB/Osaka/FTA-GA Fetal Trunk Area(SONO, SONO-5005-22)

NBL OB/NBL/NBL nasal bone length(SONO, SONO-5005-32)

OB/Bunduki/NBL nasal bone length, Bunduki (SONO,


Bunduki
SONO-5005-33)

OB/Medvedev/NBL nasal bone length, Medvedev (SONO,


Medvedev
SONO-5005-34)

OB/Sonek/NBL nasal bone length, Sonek (SONO,


Sonek
SONO-5005-35)

FL OB/FL/FL Femur Length (LN, 11963-6)

Jeanty OB/Jeanty/FL Femur Length (LN, 11963-6)

OB/Jeanty/FL-GA FL, Jeanty (SONO, SONO-5005-15)

ASUM OB/ASUM/FL Femur Length (LN, 11963-6)

OB/ASUM/FL-GA FL, ASUM 2000(SONO, SONO-5005-13)

CFEF OB/CFEF/FL Femur Length (LN, 11963-6)

OB/CFEF/FL-GA FL, CFEF(SONO, SONO-5005-14)

Hadlock OB/Hadlock/FL Femur Length (LN, 11963-6)

OB/Hadlock/FL-GA FL, Hadlock 1984(LN, 11920-6)

Hohler OB/Hohler/FL Femur Length (LN, 11963-6)

OB/Hohler/FL-GA FL,Hohler 1982(LN, 11922-2)

Johnsen OB/Johnsen/FL Femur Length (LN, 11963-6)

OB/Johnsen/FL-GA FL, Johnsen(SONO, SONO-5005-16)

Merz OB/Merz/FL Femur Length (LN, 11963-6)

OB/Merz/FL-GA FL, Merz (SONO, SONO-5005-17)

Chitty OB/Chitty /FL Femur Length (LN, 11963-6)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 84 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

OB/Chitty /FL-GA FL,Chitty 1997(LN, 33098-5

Tokyo OB/Tokyo/FL Femur Length (LN, 11963-6)

OB/Tokyo/FL-GA FL, Tokyo (SONO, SONO-5005-19)

Osaka OB/Osaka/FL Femur Length (LN, 11963-6)

OB/Osaka/FL-GA FL, Osaka (SONO, SONO-5005-18)

Hansmann OB/Hansmann/FL Femur Length (LN, 11963-6)

OB/Hansmann/FL-GA FL, Hansmann 1985(LN, 11921-4)

Lt-Kid-H OB/OB_KIDNEY/Lt-Kid-H Left Kidney Thickness(LN, 11853-9)

Rt-Kid-H OB/OB_KIDNEY/Rt-Kid-H Right Kidney thickness (LN, 11855-4)

Lt-Kid-L OB/OB_KIDNEY/Lt-Kid-L Left Kidney Length(LN, 11834-9)

Rt-Kid-L OB/OB_KIDNEY/Rt-Kid-L Right Kidney length(LN, 11836-4)

Lt-Kid-W OB/OB_KIDNEY/Lt-Kid-W Left Kidney Width(LN, 11825-7)

Rt-Kid-W OB/OB_KIDNEY/Rt-Kid-W Right Kidney width(LN, 11827-3)

A.2. 6 TID 5006 Fetal Long Bones Section


Label Item ID Code Meaning (Coding Scheme
Designator, Code Value)

Fetal Long Bones Container: Fetal Long Bones (DCM,


125003)

TIB OB/TIB/TIBI Tibia Length (LN, 11968-5)

OB/Merz/TIB Tibia Length (LN, 11968-5)


Merz
OB/Merz/TIB-GA Tibia, Merz 1987(LN ,11942-0)

OB/Jeanty/TIB Tibia Length (LN, 11968-5)


Jeanty
OB/Jeanty/TIB-GA Tibia, Jeanty 1984(LN, 11941-2)

Ulna OB/Ulna/Ulna Ulna Length (LN, 11969-3)

OB/Jeanty/Ulna Ulna Length (LN, 11969-3)


Jeanty
OB/Jeanty/Ulna-GA Ulna, Jeanty 1984(LN, 11944-6)

Merz OB/Merz/Ulna Ulna Length (LN, 11969-3)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 85 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

OB/Merz/Ulna-GA Ulna, Merz(SONO, SONO-5005-31)

FIB OB/FIB/FIB Fibula length (LN, 11964-4)

OB/Jeanty/FIB Fibula length (LN, 11964-4)


Jeanty
OB/Jeanty/FIB-GA Fibula, Jeanty 1983(LN, 33097-7)

RAD OB/RAD/RAD Radius length(LN,11967-7)

OB/Merz/RAD Radius length(LN,11967-7)


Merz
OB/Merz/RAD-GA Radius, Merz 1987(LN, 11939-6)

OB/Jeanty/RAD Radius length(LN,11967-7)


Jeanty
OB/Jeanty/RAD-GA Radius, Jeanty(SONO, SONO-5005-28)

CLAV OB/CLAV/Clav. Clavicle length(LN, 11962-8)

OB/Yarkoni/Clav. Clavicle length(LN, 11962-8)


Yarkoni
OB/Yarkoni/Clav.-GA Clavical length,Yarkoni 1985(LN, 33088-6)

HL OB/HL/HL Humerus length(LN, 11966-9)

OB/ASUM/HL Humerus length(LN, 11966-9)


ASUM
OB/ASUM/HL-GA Humerus Length, ASUM 2000(LN, 33116-5)

OB/Jeanty/HL Humerus length(LN, 11966-9)


Jeanty
OB/Jeanty/HL-GA Humerus, Jeanty 1984(LN, 11936-2)

OB/Merz/HL Humerus length(LN, 11966-9)


Merz
OB/Merz/HL-GA Humerus, Merz 1987(LN, 11937-0)

OB/Osaka/HL Humerus length(LN, 11966-9)


Osaka
OB/Osaka/HL-GA Humerus Length, Osaka 1989(LN, 33117-3)

A.2. 7 TID 5007 Fetal Cranium Section


Label Item ID Code Meaning (Coding Scheme
Designator, Code Value)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 86 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Fetal Cranium Container: Fetal Cranium Section (DCM,


Section 125004)

OOD OB/OOD/OOD Outer Orbital Diameter(LN, 11629-3)

Jeanty OB/Jeanty/OOD Outer Orbital Diameter(LN, 11629-3)

OB/Jeanty/OOD-GA Outer Orbital Diameter,Jeanty (SONO,


SONO-5005-27)

HEM OB/HEM/HEM Width of Hemisphere(LN, 12170-7)

IOD OB/IOD/IOD Inner Orbital Diameter(LN, 33070-4)

NF OB/NF/NF Nuchal Fold thickness(LN, 12146-7)

VA OB/VA/Va Anterior Horn Lateral ventricular width(LN,


33197-5)

VP OB/VP/Vp Posterior Horn Lateral ventricular


width(LN, 33196-7)

A.2. 9 TID 5010 Amniotic Sac Section


Label Item ID Code Meaning (Coding Scheme Designator,
Code Value)

Amniotic Sac Container: Amniotic Sac (SRT, T-F1300)

Q1 OB/AFI/Q1 First Quadrant Diameter (LN, 11624-4)

Q2 OB/AFI/Q2 Second Quadrant Diameter (LN, 11626-9)

Q3 OB/AFI/Q3 Third Quadrant Diameter (LN, 11625-1)

Q4 OB/AFI/Q4 Fourth Quadrant Diameter (LN, 11623-6)

AFI OB/AFI/AFI Amniotic Fluid Index (LN, 11627-7)

A.2. 10 TID 5011 Early Gestation Section


Label Item ID Code Meaning (Coding Scheme Designator,
Code Value)

Fetal Biometry Container: Fetal Biometry (DCM, 125002)


Measurements

CRL OB/CRL/CRL Crown Rump Length (LN, 11957-8)

OB/ASUM/CRL Crown Rump Length (LN, 11957-8)


ASUM
OB/ASUM/CRL-GA CRL, ASUM 2000(LN, 33090-2)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 87 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

OB/Hadlock/CRL Crown Rump Length (LN, 11957-8)


Hadlock
OB/Hadlock/CRL-GA CRL, Hadlock 1992(LN, 11910-7)

OB/Hansmann/CRL Crown Rump Length (LN, 11957-8)


Hansmann
OB/Hansmann/CRL-GA CRL, Hansmann 1986(LN, 33540-6)

OB/Nelson/CRL Crown Rump Length (LN, 11957-8)


Nelson
OB/Nelson/CRL-GA CRL, Nelson 1981(LN,11913-1)

OB/Rempen/CRL Crown Rump Length (LN, 11957-8)


Rempen
OB/Rempen/CRL-GA CRL, Rempen 1991(LN, 33094-4)

OB/Robinson/CRL Crown Rump Length (LN, 11957-8)


Robinson
OB/Robinson/CRL-GA CRL, Robinson 1975(LN, 11914-9)

OB/Tokyo/CRL Crown Rump Length (LN, 11957-8)


Tokyo
OB/Tokyo/CRL-GA CRL, Tokyo 1986(LN, 33096-9)

OB/Osaka/CRL Crown Rump Length (LN, 11957-8)


Osaka
OB/Osaka/CRL-GA CRL, Osaka 1989(LN, 33093-6)

GS OB/GS/GS Gestational Sac Diameter (LN, 11850-5)

OB/Hansmann/GS Gestational Sac Diameter (LN, 11850-5)


Hansmann
OB/Hansmann/GS-GA GS, Hansmann 1979(LN, 33105-8)

OB/Hellman/GS Gestational Sac Diameter (LN, 11850-5)


Hellman
OB/Hellman/GS-GA GS, Hellman 1969(LN, 11928-9)

OB/Rempen/GS Gestational Sac Diameter (LN, 11850-5)


Rempen
OB/Rempen/GS-GA GS, Rempen 1991(LN, 11929-7)

OB/Tokyo/GS Gestational Sac Diameter (LN, 11850-5)


Tokyo
OB/Tokyo/GS-GA GS, Tokyo 1986(LN, 33108-2)

NT OB/NT/NT Nuchal Translucency (LN, 33069-6)

YS OB/YS/YS Yolk Sac length(LN, 11816-6)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 88 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

A.2. 11 TID 5012 Ovaries Section


Label Item ID Code Meaning (Coding Scheme Designator,
Code Value)

Ovaries Finding Site: Ovary (DCM, T-87000)

Left Ovary, Length OB/Ovary/Lt-Ov-L Left Ovary Length (LN, 11840-6)

Left Ovary, Height OB/Ovary/Lt-Ov-H Left Ovary Height (LN, 11857-0)

Left Ovary, Width OB/Ovary/Lt-Ov-W Left Ovary Width (LN, 11829-9)

Left Ovary, Volume OB/Ovary/Lt-Ov-Vol. Left Ovary Volume (LN, 12164-0)

Right Ovary, OB/Ovary/Rt-Ov-L Right Ovary Length (LN, 11841-4)


Length

Right Ovary, OB/Ovary/Rt-Ov-H Right Ovary Height (LN, 11858-8)


Height

Right Ovary, Width OB/Ovary/Rt-Ov-W Right Ovary Width (LN, 11830-7)

Right Ovary, OB/Ovary/Rt-Ov-Vol. Right Ovary Volume (LN, 12165-7)


Volume

Left Ovary, Length GYN/GYN_OVARY/Lt-Ov-L Left Ovary Length (LN, 11840-6)

Left Ovary, Height GYN/GYN_OVARY/Lt-Ov-H Left Ovary Height (LN, 11857-0)

Left Ovary, Width GYN/GYN_OVARY/Lt-Ov-W Left Ovary Width (LN, 11829-9)

Left Ovary, Volume GYN/GYN_OVARY/Lt-Ov-Vol. Left Ovary Volume (LN, 12164-0)

Right Ovary, GYN/GYN_OVARY/Rt-Ov-L Right Ovary Length (LN, 11841-4)


Length

Right Ovary, GYN/GYN_OVARY/Rt-Ov-H Right Ovary Height (LN, 11858-8)


Height

Right Ovary, Width GYN/GYN_OVARY/Rt-Ov-W Right Ovary Width (LN, 11830-7)

Right Ovary, GYN/GYN_OVARY/Rt-Ov-Vol. Right Ovary Volume (LN, 12165-7)


Volume

A.2. 13 TID 5014 Follicle Measurement Section


Code Meaning (Coding Scheme Designator, Code
Label Item ID
Value)

Follicles Finding Site: Ovarian Follicle(DCM, T-87600)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 89 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

GYN/GYN_FOLLICLE1/Lt-Fol1-D1

GYN/GYN_FOLLICLE1/Lt-Fol1-D2

GYN/GYN_FOLLICLE1/Lt-Fol1-D3

GYN/GYN_FOLLICLE1/Rt-Fol1-D1

GYN/GYN_FOLLICLE1/Rt-Fol1-D2
GYN/GYN_FOLLICLE1/Rt-Fol1-D3


Follicles All Follicle Diameter(LN, 11793-7)
GYN/GYN_FOLLICLE20/Lt-Fol20-D1

GYN/GYN_FOLLICLE20/Lt-Fol20-D2

GYN/GYN_FOLLICLE20/Lt-Fol20-D3

GYN/GYN_FOLLICLE20/Rt-Fol20-D1

GYN/GYN_FOLLICLE20/Rt-Fol20-D2

GYN/GYN_FOLLICLE20/Rt-Fol20-D3

GYN/GYN_FOLLICLE1/Lt-Fol1-Avg

GYN/GYN_FOLLICLE1/Rt-Fol1-Avg

Follicles … Follicle Diameter Mean(SONO, SONO-5013-1)

GYN/GYN_FOLLICLE20/Rt-Fol20-Avg

GYN/GYN_FOLLICLE20/Lt-Fol20-Avg

GYN/GYN_FOLLICLE1/Lt-Fol1- Vol

GYN/GYN_FOLLICLE1/Rt-Fol1- Vol

Follicles … Follicle Volume(SONO, SONO-5013-2)

GYN/GYN_FOLLICLE20/Rt-Fol20- Vol

GYN/GYN_FOLLICLE20/Lt-Fol20- Vol

GYN/GYN_FIBROID1/Fib1-D1

GYN/GYN_FIBROID1/Fib1-D2

GYN/GYN_FIBROID1/Fib1-D3

Fibroids All … Fibroid Diameter(SONO, SONO-5013-3)

GYN/GYN_FIBROID20/Fib20-D1

GYN/GYN_FIBROID20/Fib20-D2

GYN/GYN_FIBROID20/Fib20-D3

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 90 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

GYN/GYN_FIBROID1/Fib1-Avg

Fibroids … Fibroid Diameter Mean(SONO, SONO-5013-4)

GYN/GYN_FIBROID20/Fib2-Avg

GYN/GYN_FIBROID1/Fib1-Vol

Fibroids … Fibroid Volume(SONO, SONO-5013-5)

GYN/GYN_FIBROID20/Fib20-Vol

A.2. 14 TID 5015 Pelvis and Uterus Section


Code Meaning (Coding Scheme Designator, Code
Label Item ID
Value)

Pelvis and
Finding Site: Pelvis and Uterus (DCM, 125011)
Uterus

Length GYN/GYN_UT_LENGTH/CervixL Cervix Length (LN, 11961-0)

Height GYN/Cervix/CX-H Cervix Height(SONO, SONO-5015-1)

Width GYN/Cervix/CX-W Cervix Width(SONO, SONO-5015-2)

Uterus Container: Uterus (SRT, T-83000)

Length GYN/GYN_UTERUS/Ut-L Uterus Length (LN, 11842-2)

Width GYN/GYN_UTERUS/Ut-W Uterus Width (LN, 11865-3)

Depth GYN/GYN_UTERUS/Ut-H Uterus Height (LN, 11859-6)

Vol GYN/GYN_UTERUS/Ut-Vol Uterus Volume(LN, 33192-6)

Endometrium
GYN/GYN_UT_THICKN/Ut-Endom.Th Endometrium Thickness (LN, 12145-9)
Thickness

A.2. 15 TID 5025 OB-GYN Fetal Vascular Ultrasound Measurement Group


Code Meaning (Coding Scheme
Label Item ID
Designator, Code Value)

OB/DUCTUSART/D_DA_PS/DA-PS

OB/DUCTUSART/D_DA_RI/DA-PS
Ductus Art PS Peak Systolic Velocity(LN, 11726-7)
OB/DUCTUSART/D_DA_PI/DA-PS

OB/DUCTUSART/D_DA_PERS/DA-PS

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 91 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

OB/DUCTUSART/D_DA_AT/DA-PS

OB/DUCTUSART/D_DA_MT/DA-PS

OB/AO/D_AO_PS/Ao-PS/

OB/AO/D_AO_RI/Ao-PS

OB/AO/D_AO_PI/Ao-PS
Ao PS
OB/AO/D_AO_PERS/Ao-PS

OB/AO/D_AO_AT/Ao-PS

OB/AO/D_AO_MT/Ao-PS

OB/MCA/D_MCA_RI/Lt-MCA-PS

OB/MCA/D_MCA_RI/Rt-MCA-PS

OB/MCA/D_MCA_PI/Lt-MCA-PS

OB/MCA/D_MCA_PI/Rt-MCA-PS

OB/MCA/D_MCA_PERS/Lt-MCA-PS
Lt MCA(Rt MCA) PS
OB/MCA/D_MCA_PERS/Rt-MCA-PS

OB/MCA/D_MCA_AT/Lt-MCA-PS

OB/MCA/D_MCA_AT/Rt-MCA-PS

OB/MCA/D_MCA_MT/Lt-MCA-PS

OB/MCA/D_MCA_MT/Rt-MCA-PS

OB/SMA/D_SMA_RI/SMA-PS

OB/SMA/D_SMA_PI/SMA-PS

SMA PS OB/SMA/D_SMA_PERS/SMA-PS

OB/SMA/D_SMA_AT/SMA-PS

OB/SMA/D_SMA_MT/SMA-PS

OB/CELIACA/D_CELA_RI/Celiac A.-PS

OB/CELIACA/D_CELA_PI/Celiac A.-PS

Celiac.A. PS OB/CELIACA/D_CELA_PERS/Celiac A.-PS

OB/CELIACA/D_CELA_AT/Celiac A.-PS

OB/CELIACA/D_CELA_MT/Celiac A.-PS

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 92 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

OB/CAROTID/D_CAR_RI/Lt-Car-PS

OB/CAROTID/D_CAR_RI/Rt-Car-PS

OB/CAROTID/D_CAR_PI/Lt-Car-PS

OB/CAROTID/D_CAR_PI/Rt-Car-PS

OB/CAROTID/D_CAR_PERS/Lt-Car-PS
Carotid PS
OB/CAROTID/D_CAR_PERS/Rt-Car-PS

OB/CAROTID/D_CAR_AT/Lt-Car-PS

OB/CAROTID/D_CAR_AT/Rt-Car-PS

OB/CAROTID/D_CAR_MT/Lt-Car-PS

OB/CAROTID/D_CAR_MT/Rt-Car-PS

OB/DUCTUSART/D_DA_ED/DA-ED

OB/DUCTUSART/D_DA_RI/DA-ED

OB/DUCTUSART/D_DA_PI/DA-ED
Ductus Art ED
OB/DUCTUSART/D_DA_PERS/DA-ED

OB/DUCTUSART/D_DA_AT/DA-ED

OB/DUCTUSART/D_DA_MT/DA-ED

OB/AO/D_AO_ED/Ao-ED

OB/AO/D_AO_RI/Ao-ED

OB/AO/D_AO_PI/Ao-ED
Ao ED End Diastolic Velocity(LN, 11653-3)
OB/AO/D_AO_PERS/Ao-ED

OB/AO/D_AO_AT/Ao-ED

OB/AO/D_AO_MT/Ao-ED

OB/MCA/D_MCA_ED/Lt-MCA-ED

OB/MCA/D_MCA_ED/Rt-MCA-ED

OB/MCA/D_MCA_RI/Lt-MCA-ED
MCA ED
OB/MCA/D_MCA_RI/Rt-MCA-ED

OB/MCA/D_MCA_PI/Lt-MCA-ED

OB/MCA/D_MCA_PI/Rt-MCA-ED

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 93 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

OB/MCA/D_MCA_PERS/Lt-MCA-ED

OB/MCA/D_MCA_PERS/Rt-MCA-ED

OB/MCA/D_MCA_AT/Lt-MCA-ED

OB/MCA/D_MCA_AT/Rt-MCA-ED

OB/MCA/D_MCA_MT/Lt-MCA-ED

OB/MCA/D_MCA_MT/Rt-MCA-ED

OB/SMA/D_SMA_ED/SMA-ED

OB/SMA/D_SMA_RI/SMA-ED

OB/SMA/D_SMA_PI/SMA-ED
SMA ED
OB/SMA/D_SMA_PERS/SMA-ED

OB/SMA/D_SMA_AT/SMA-ED

OB/SMA/D_SMA_MT/SMA-ED

OB/CELIACA/D_CELA_ED/Celiac A.-ED

OB/CELIACA/D_CELA_RI/Celiac A.-ED

OB/CELIACA/D_CELA_PI/Celiac A.-ED
Celiac A. ED
OB/CELIACA/D_CELA_PERS/Celiac A.-ED

OB/CELIACA/D_CELA_AT/Celiac A.-ED

OB/CELIACA/D_CELA_MT/Celiac A.-ED

OB/CAROTID/D_CAR_ED/Lt-Car-ED

OB/CAROTID/D_CAR_ED/Rt-Car-ED

OB/CAROTID/D_CAR_RI/Lt-Car-ED

OB/CAROTID/D_CAR_RI/Rt-Car-ED

OB/CAROTID/D_CAR_PI/Lt-Car-ED
Carotid ED
OB/CAROTID/D_CAR_PI/Rt-Car-ED

OB/CAROTID/D_CAR_PERS/Lt-Car-ED

OB/CAROTID/D_CAR_PERS/Rt-Car-ED

OB/CAROTID/D_CAR_AT/Lt-Car-ED

OB/CAROTID/D_CAR_AT/Rt-Car-ED

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 94 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

OB/CAROTID/D_CAR_MT/Lt-Car-ED

OB/CAROTID/D_CAR_MT/Rt-Car-ED

OB/DUCTUSART/D_DA_RI/DA-RI

OB/DUCTUSART/D_DA_PERS/DA-RI
Ductus Art RI
OB/DUCTUSART/D_DA_AT/DA-RI

OB/DUCTUSART/D_DA_MT/DA-RI

OB/AO/D_AO_RI/Ao-RI

OB/AO/D_AO_PERS/Ao-RI
Ao RI
OB/AO/D_AO_AT/Ao-RI

OB/AO/D_AO_MT/Ao-RI

OB/MCA/D_MCA_RI/Lt-MCA-RI

OB/MCA/D_MCA_RI/Rt-MCA-RI

OB/MCA/D_MCA_PERS/Lt-MCA-RI

OB/MCA/D_MCA_PERS/Rt-MCA-RI
MCA RI
OB/MCA/D_MCA_AT/Lt-MCA-RI
Resistivity Index(LN, 12023-8)
OB/MCA/D_MCA_AT/Rt-MCA-RI

OB/MCA/D_MCA_MT/Lt-MCA-RI

OB/MCA/D_MCA_MT/Rt-MCA-RI

OB/SMA/D_SMA_RI/SMA-RI

OB/SMA/D_SMA_PERS/SMA-RI
SMA RI
OB/SMA/D_SMA_AT/SMA-RI

OB/SMA/D_SMA_MT/SMA-RI

OB/CELIACA/D_CELA_RI/Celiac A.-RI

OB/CELIACA/D_CELA_PERS/Celiac A.-RI
Celiac A. RI
OB/CELIACA/D_CELA_AT/Celiac A.-RI

OB/CELIACA/D_CELA_MT/Celiac A.-RI

OB/CAROTID/D_CAR_RI/Lt-Car-RI
Carotid RI
OB/CAROTID/D_CAR_RI/Rt-Car-RI

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 95 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

OB/CAROTID/D_CAR_PERS/Lt-Car-RI

OB/CAROTID/D_CAR_PERS/Rt-Car-RI

OB/CAROTID/D_CAR_AT/Lt-Car-RI

OB/CAROTID/D_CAR_AT/Rt-Car-RI

OB/CAROTID/D_CAR_MT/Lt-Car-RI

OB/CAROTID/D_CAR_MT/Rt-Car-RI

OB/DUCTUSART/D_DA_RI/DA-PI

OB/DUCTUSART/D_DA_PI/DA-PI
Ductus Art PI
OB/DUCTUSART/D_DA_AT/DA-PI

OB/DUCTUSART/D_DA_MT/DA-PI

OB/AO/D_AO_RI/Ao-PI

OB/AO/D_AO_PI/Ao-PI
Ao PI
OB/AO/D_AO_AT/Ao-PI

OB/AO/D_AO_MT/Ao-PI

OB/MCA/D_MCA_RI/Lt-MCA-PI

OB/MCA/D_MCA_RI/Rt-MCA-PI

OB/MCA/D_MCA_PI/Lt-MCA-PI

OB/MCA/D_MCA_PI/Rt-MCA-PI Pulsatility Index(LN, 12008-9)


MCA PI
OB/MCA/D_MCA_AT/Lt-MCA-PI

OB/MCA/D_MCA_AT/Rt-MCA-PI

OB/MCA/D_MCA_MT/Lt-MCA-PI

OB/MCA/D_MCA_MT/Rt-MCA-PI

OB/SMA/D_SMA_RI/SMA-PI

OB/SMA/D_SMA_PI/SMA-PI
SMA PI
OB/SMA/D_SMA_AT/SMA-PI

OB/SMA/D_SMA_MT/SMA-PI

OB/CELIACA/D_CELA_RI/Celiac A.-PI

Celiac A. PI OB/CELIACA/D_CELA_AT/Celiac A.-PI

OB/CELIACA/D_CELA_MT/Celiac A.-PI

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 96 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

OB/CAROTID/D_CAR_RI/Lt-Car-PI

OB/CAROTID/D_CAR_RI/Rt-Car-PI

OB/CAROTID/D_CAR_PI/Lt-Car-PI

OB/CAROTID/D_CAR_PI/Rt-Car-PI
Carotid PI
OB/CAROTID/D_CAR_AT/Lt-Car-PI

OB/CAROTID/D_CAR_AT/Rt-Car-PI

OB/CAROTID/D_CAR_MT/Lt-Car-PI

OB/CAROTID/D_CAR_MT/Rt-Car-PI

OB/DUCTUSART/D_DA_PERS/DA-S/D

Ductus Art-S/D OB/DUCTUSART/D_DA_AT/DA-S/D

OB/DUCTUSART/D_DA_MT/DA-S/D

OB/AO/D_AO_PERS/Ao-S/D

Ao-S/D OB/AO/D_AO_AT/Ao-S/D

OB/AO/D_AO_MT/Ao-S/D

OB/MCA/D_MCA_PERS/Lt-MCA-S/D

OB/MCA/D_MCA_PERS/Rt-MCA-S/D

OB/MCA/D_MCA_AT/Lt-MCA-S/D
MCA-S/D
OB/MCA/D_MCA_AT/Rt-MCA-S/D
Systolic to Diastolic Velocity Ratio(LN,
12144-2)
OB/MCA/D_MCA_MT/Lt-MCA-S/D

OB/MCA/D_MCA_MT/Rt-MCA-S/D

OB/SMA/D_SMA_PERS/SMA-S/D

SMA-S/D OB/SMA/D_SMA_AT/SMA-S/D

OB/SMA/D_SMA_MT/SMA-S/D

OB/CELIACA/D_CELA_PERS/Celiac A.-S/D

Celiac A.-S/D OB/CELIACA/D_CELA_AT/Celiac A.-S/D

OB/CELIACA/D_CELA_MT/Celiac A.-S/D

OB/CAROTID/D_CAR_PERS/Lt-Car-S/D
Carotid-S/D
OB/CAROTID/D_CAR_PERS/Rt-Car-S/D

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 97 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

OB/CAROTID/D_CAR_AT/Lt-Car-S/D

OB/CAROTID/D_CAR_AT/Rt-Car-S/D

OB/CAROTID/D_CAR_MT/Lt-Car-S/D

OB/CAROTID/D_CAR_MT/Rt-Car-S/D

OB/DUCTUSART/D_DA_AT/DA-TAmean
Ductus Art-TAmean
OB/DUCTUSART/D_DA_MT/DA-TAmean

OB/AO/D_AO_AT/Ao-TAmean
Ao-TAmean
OB/AO/D_AO_MT/Ao-TAmean

OB/MCA/D_MCA_PI/Lt-MCA-TAmean

OB/MCA/D_MCA_PI/Rt-MCA-TAmean

OB/MCA/D_MCA_AT/Lt-MCA-TAmean

OB/MCA/D_MCA_AT/Rt-MCA-TAmean
MCA-TAmean
OB/MCA/D_MCA_MT/Lt-MCA-TAmean

OB/MCA/D_MCA_MT/Rt-MCA-TAmean

OB/MCA/D_MCA_VOL/Lt-MCA-TAmean
Time averaged mean velocity(LN,
OB/MCA/D_MCA_VOL/Rt-MCA-TAmean 20352-1)

OB/SMA/D_SMA_AT/SMA-TAmean
SMA-TAmean
OB/SMA/D_SMA_MT/SMA-TAmean

OB/CELIACA/D_CELA_AT/Celiac A.-TAmean
Celiac A.-TAmean
OB/CELIACA/D_CELA_MT/Celiac A.-TAmean

OB/CAROTID/D_CAR_AT/Lt-Car-TAmean

OB/CAROTID/D_CAR_AT/Rt-Car-TAmean

OB/CAROTID/D_CAR_MT/Lt-Car-TAmean
Carotid-TAmean
OB/CAROTID/D_CAR_MT/Rt-Car-TAmean

OB/CAROTID/D_CAR_VOL/Lt-Car-TAmean

OB/CAROTID/D_CAR_VOL/Rt-Car-TAmean

OB/D_IVC_S/IVC-S
Peak Velocity During Ventricular
IVC OB/D_IVC_SAPLI/IVC-S
Systole(SONO, SONO-5025-11)
OB/D_IVC_PVIV/IVC-S

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 98 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

OB/D_IVC_S/IVC-D Peak Velocity During Ventricular


Diastole(SONO, SONO-5025-12)
OB/D_IVC_PVIV/IVC-D

OB/D_IVC_SAPLI/IVC-a Lowest Velocity During Atrial


OB/D_IVC_PVIV/IVC-a Contraction(SONO, SONO-5025-13)

Peak Velocity During Ventricular Systole


To Lowest Velocity During Atrial
OB/D_IVC_SAPLI/IVC-S/a
Contraction Velocity Ratio(SONO,
SONO-5025-14)

OB/D_IVC_SAPLI/IVC-PLI Preload Index(SONO, SONO-5025-15)

Peak Velocity Index Vein(SONO,


OB/D_IVC_PVIV/IVC-PVIV
SONO-5025-16)

Ductus Art-TAmax OB/DUCTUSART/D_DA_MT/DA-TAmax

OB/AO/D_AO_AT/Ao-TAmax
Ao-TAmax
OB/AO/D_AO_MT/Ao-TAmax

OB/MCA/D_MCA_PI/Lt-MCA-TAmax

OB/MCA/D_MCA_PI/Rt-MCA-TAmax

OB/MCA/D_MCA_AT/Lt-MCA-TAmax
MCA-TAmax
OB/MCA/D_MCA_AT/Rt-MCA-TAmax

OB/MCA/D_MCA_MT/Lt-MCA-TAmax

OB/MCA/D_MCA_MT/Rt-MCA-TAmax

OB/SMA/D_SMA_AT/SMA-TAmax
Time averaged peak velocity(LN,
SMA-TAmax
11692-1)
OB/SMA/D_SMA_MT/SMA-TAmax

OB/CELIACA/D_CELA_PI/Celiac A.-TAmax

Celiac A.-TAmax OB/CELIACA/D_CELA_AT/Celiac A.-TAmax

OB/CELIACA/D_CELA_MT/Celiac A.-TAmax

OB/CAROTID/D_CAR_PI/Lt-Car-TAmax

OB/CAROTID/D_CAR_PI/Rt-Car-TAmax

OB/CAROTID/D_CAR_AT/Lt-Car-TAmax
Carotid-TAmax
OB/CAROTID/D_CAR_AT/Rt-Car-TAmax

OB/CAROTID/D_CAR_MT/Lt-Car-TAmax

OB/CAROTID/D_CAR_MT/Rt-Car-TAmax

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 99 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

OB/DUCTUSART/D_DA_AT/DA-AT
Ductus Art-AT
OB/DUCTUSART/D_DA_MT/DA-AT

OB/AO/D_AO_AT/Ao-AT
Ao-AT
OB/AO/D_AO_MT/Ao-AT

OB/MCA/D_MCA_AT/Lt-MCA-AT

OB/MCA/D_MCA_AT/Rt-MCA-AT
MCA-AT
OB/MCA/D_MCA_MT/Lt-MCA-AT

OB/MCA/D_MCA_MT/Rt-MCA-AT
Acceleration Time(LN, 20168-1)
OB/SMA/D_SMA_AT/SMA-AT
SMA-AT
OB/SMA/D_SMA_MT/SMA-AT

OB/CELIACA/D_CELA_AT/Celiac A.-AT
Celiac A.-AT
OB/CELIACA/D_CELA_MT/Celiac A.-AT

OB/CAROTID/D_CAR_AT/Lt-Car-AT

OB/CAROTID/D_CAR_AT/Rt-Car-AT
Carotid-AT
OB/CAROTID/D_CAR_MT/Lt-Car-AT

OB/CAROTID/D_CAR_MT/Rt-Car-AT

OB/DUCTUSART/D_DA_AT/DA-VTI
Ductus Art -VTI
OB/DUCTUSART/D_DA_MT/DA-VTI

OB/AO/D_AO_AT/Ao-VTI
Ao-VTI
OB/AO/D_AO_MT/Ao-VTI

OB/MCA/D_MCA_AT/Lt-MCA-VTI

OB/MCA/D_MCA_AT/Rt-MCA-VTI
MCA-VTI Velocity Time Integral(LN, 20354-7)
OB/MCA/D_MCA_MT/Lt-MCA-VTI

OB/MCA/D_MCA_MT/Rt-MCA-VTI

OB/SMA/D_SMA_AT/SMA-VTI
SMA-VTI
OB/SMA/D_SMA_MT/SMA-VTI

OB/CELIACA/D_CELA_AT/Celiac A.-VTI
Celiac A.-VTI
OB/CELIACA/D_CELA_MT/Celiac A.-VTI

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 100 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

OB/CAROTID/D_CAR_AT/Lt-Car-VTI

OB/CAROTID/D_CAR_AT/Rt-Car-VTI
Carotid-VTI
OB/CAROTID/D_CAR_MT/Lt-Car-VTI

OB/CAROTID/D_CAR_MT/Rt-Car-VTI

OB/DUCTUSART/D_DA_AT/DA-HR

Ductus Art-HR OB/DUCTUSART/D_DA_MT/DA-HR

OB/DUCTUSART/D_DA_HR/DA-HR

OB/AO/D_AO_AT/Ao-HR

Ao-HR OB/AO/D_AO_MT/Ao-HR

OB/AO/D_AO_HR/Ao-HR

OB/MCA/D_MCA_AT/Lt-MCA-HR

OB/MCA/D_MCA_AT/Rt-MCA-HR

OB/MCA/D_MCA_MT/Lt-MCA-HR
MCA-HR
OB/MCA/D_MCA_MT/Rt-MCA-HR

OB/MCA/D_MCA_HR/Lt-MCA-HR

OB/MCA/D_MCA_HR/Rt-MCA-HR Heart Rate(LN, 8867-4)


OB/SMA/D_SMA_AT/SMA-HR

MCA-HR OB/SMA/D_SMA_MT/SMA-HR

OB/SMA/D_SMA_HR/SMA-HR

OB/CELIACA/D_CELA_AT/Celiac A.-HR

Celiac A.-HR OB/CELIACA/D_CELA_MT/Celiac A.-HR

OB/CELIACA/D_CELA_HR/Celiac A.-HR

OB/CAROTID/D_CAR_AT/Lt-Car-HR

OB/CAROTID/D_CAR_AT/Rt-Car-HR

OB/CAROTID/D_CAR_MT/Lt-Car-HR
Carotid-HR
OB/CAROTID/D_CAR_MT/Rt-Car-HR

OB/CAROTID/D_CAR_HR/Lt-Car-HR

OB/CAROTID/D_CAR_HR/Rt-Car-HR

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 101 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Ductus Art-T OB/DUCTUSART/D_DA_TIME/DA-T

Ao-T OB/AO/D_AO_TIME/Ao-T

OB/MCA/D_MCA_TIME/Lt-MCA-T
MCA-T
OB/MCA/D_MCA_TIME/Rt-MCA-T
Acceleration Time (LN, 20168-1)
SMA-T OB/SMA/D_SMA_TIME/SMA-T

Celiac A.-T OB/CELIACA/D_CELA_TIME/Celiac A.-T

OB/CAROTID/D_CAR_TIME/Lt-Car-T
Carotid-T
OB/CAROTID/D_CAR_TIME/Rt-Car-T

MCA-Vol.Flow OB/MCA/D_MCA_VOL/Lt-MCA-Vol. Flow

OB/MCA/D_MCA_VOL/Rt-MCA-Vol. Flow
Volume flow(LN, 33878-0)
Carotid-Vol.Flow OB/CAROTID/D_CAR_VOL/Lt-Car-Vol. Flow

OB/CAROTID/D_CAR_VOL/Rt-Car-Vol. Flow

A.2. 16 TID 5026 OB-GYN Pelvic Vasculatrue Section Measurement Group


Code Meaning (Coding
Label Item ID Scheme Designator,
Code Value)

Finding Site: Pelvic


Pelvic Vascular
Vascular Structure
Structure
(SRT, T-D6007)

UmbV-Dia OB/UMBILICALART/D_UMBA
Umbilical Vein (SRT,
Umbilical Vein T-F1820)
OB/UMBVD/UmbV-Diam
Diam

OB/UT_ART/Lt-UtA-Diam Vessel lumen


Uterine Art.Diam
OB/UT_ART/Rt-UtA-Diam diameter(SRT, G-0364)

OB/UMBILICALART/D_UMBA_PS/UmbA-PS

OB/UMBILICALART/D_UMBA_RI/UmbA-PS

OB/UMBILICALART/D_UMBA_PI/UmbA-PS
Peak Systolic
Umbilical Art.-PS
Velocity(LN, 11726-7)
OB/UMBILICALART/D_UMBA_PERS/UmbA-PS

OB/UMBILICALART/D_UMBA_AT/UmbA-PS

OB/UMBILICALART/D_UMBA_MT/UmbA-PS

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 102 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

OB/UTERINEART/D_UTA_PS/Lt-UtA-PS

OB/UTERINEART/D_UTA_PS/Rt-UtA-PS

OB/UTERINEART/D_UTA_RI/Lt-UtA-PS

OB/UTERINEART/D_UTA_RI/Rt-UtA-PS

OB/UTERINEART/D_UTA_PI/Lt-UtA-PS

OB/UTERINEART/D_UTA_PI/Rt-UtA-PS
Uterine Art.-PS
OB/UTERINEART/D_UTA_PERS/Lt-UtA-PS

OB/UTERINEART/D_UTA_PERS/Rt-UtA-PS

OB/UTERINEART/D_UTA_AT/Lt-UtA-PS

OB/UTERINEART/D_UTA_AT/Rt-UtA-PS

OB/UTERINEART/D_UTA_MT/Lt-UtA-PS

OB/UTERINEART/D_UTA_MT/Rt-UtA-PS

OB/UMBILICALARTLT/D_UMBALT_PS/Lt-UmA-PS

OB/UMBILICALARTLT/D_UMBALT_PS/Rt-UmA-PS

OB/UMBILICALARTLT/D_UMBALT_RI/Lt-UmA-PS

OB/UMBILICALARTLT/D_UMBALT_RI/Rt-UmA-PS

OB/UMBILICALARTLT/D_UMBALT_PI/Lt-UmA-PS

OB/UMBILICALARTLT/D_UMBALT_PI/Rt-UmA-PS
UmA-PS
OB/UMBILICALARTLT/D_UMBALT_PERS/Lt-UmA-PS

OB/UMBILICALARTLT/D_UMBALT_PERS/Rt-UmA-PS

OB/UMBILICALARTLT/D_UMBALT_AT/Lt-UmA-PS

OB/UMBILICALARTLT/D_UMBALT_AT/Rt-UmA-PS

OB/UMBILICALARTLT/D_UMBALT_MT/Lt-UmA-PS

OB/UMBILICALARTLT/D_UMBALT_MT/Rt-UmA-PS

OB/UMBILICALART/D_UMBA_ED/UmbA-ED

OB/UMBILICALART/D_UMBA_RI/UmbA-ED
End Diastolic
Umbilical Art.-ED OB/UMBILICALART/D_UMBA_PI/UmbA-ED
Velocity(LN, 11653-3)
OB/UMBILICALART/D_UMBA_PERS/UmbA-ED

OB/UMBILICALART/D_UMBA_AT/UmbA-ED

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 103 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

OB/UMBILICALART/D_UMBA_MT/UmbA-ED

OB/UTERINEART/D_UTA_ED/Lt-UtA-ED

OB/UTERINEART/D_UTA_ED/Rt-UtA-ED

OB/UTERINEART/D_UTA_RI/Lt-UtA-ED

OB/UTERINEART/D_UTA_RI/Rt-UtA-ED

OB/UTERINEART/D_UTA_PI/Lt-UtA-ED

OB/UTERINEART/D_UTA_PI/Rt-UtA-ED
Uterine Art.-ED
OB/UTERINEART/D_UTA_PERS/Lt-UtA-ED

OB/UTERINEART/D_UTA_PERS/Rt-UtA-ED

OB/UTERINEART/D_UTA_AT/Lt-UtA-ED

OB/UTERINEART/D_UTA_AT/Rt-UtA-ED

OB/UTERINEART/D_UTA_MT/Lt-UtA-ED

OB/UTERINEART/D_UTA_MT/Rt-UtA-ED

OB/UMBILICALARTLT/D_UMBALT_ED/Lt-UmA-ED

OB/UMBILICALARTLT/D_UMBALT_ED/Rt-UmA-ED

OB/UMBILICALARTLT/D_UMBALT_RI/Lt-UmA-ED

OB/UMBILICALARTLT/D_UMBALT_RI/Rt-UmA-ED

OB/UMBILICALARTLT/D_UMBALT_PI/Lt-UmA-ED

OB/UMBILICALARTLT/D_UMBALT_PI/Rt-UmA-ED
UmA-ED
OB/UMBILICALARTLT/D_UMBALT_PERS/Lt-UmA-ED

OB/UMBILICALARTLT/D_UMBALT_PERS/Rt-UmA-ED

OB/UMBILICALARTLT/D_UMBALT_AT/Lt-UmA-ED

OB/UMBILICALARTLT/D_UMBALT_AT/Rt-UmA-ED

OB/UMBILICALARTLT/D_UMBALT_MT/Lt-UmA-ED

OB/UMBILICALARTLT/D_UMBALT_MT/Rt-UmA-ED

OB/UMBILICALART/D_UMBA_RI/UmbA-RI

OB/UMBILICALART/D_UMBA_PERS/UmbA-RI
Resisitivity Index(LN,
Umbilical Art.-RI
12023-8)
OB/UMBILICALART/D_UMBA_AT/UmbA-RI

OB/UMBILICALART/D_UMBA_MT/UmbA-RI

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 104 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

OB/UTERINEART/D_UTA_RI/Lt-UtA-RI

OB/UTERINEART/D_UTA_RI/Rt-UtA-RI

OB/UTERINEART/D_UTA_PERS/Lt-UtA-RI

OB/UTERINEART/D_UTA_PERS/Rt-UtA-RI
Uterine Art.-RI
OB/UTERINEART/D_UTA_AT/Lt-UtA-RI

OB/UTERINEART/D_UTA_AT/Rt-UtA-RI

OB/UTERINEART/D_UTA_MT/Lt-UtA-RI

OB/UTERINEART/D_UTA_MT/Rt-UtA-RI

OB/UMBILICALARTLT/D_UMBALT_RI/Lt-UmA-RI

OB/UMBILICALARTLT/D_UMBALT_RI/Rt-UmA-RI

OB/UMBILICALARTLT/D_UMBALT_PERS/Lt-UmA-RI

OB/UMBILICALARTLT/D_UMBALT_PERS/Rt-UmA-RI
UmA-RI
OB/UMBILICALARTLT/D_UMBALT_AT/Lt-UmA-RI

OB/UMBILICALARTLT/D_UMBALT_AT/Rt-UmA-RI

OB/UMBILICALARTLT/D_UMBALT_MT/Lt-UmA-RI

OB/UMBILICALARTLT/D_UMBALT_MT/Rt-UmA-RI

OB/UMBILICALART/D_UMBA_RI/UmbA-PI

OB/UMBILICALART/D_UMBA_PI/UmbA-PI
Umbilical Art.-PI
OB/UMBILICALART/D_UMBA_AT/UmbA-PI

OB/UMBILICALART/D_UMBA_MT/UmbA-PI

OB/UTERINEART/D_UTA_RI/Lt-UtA-PI

OB/UTERINEART/D_UTA_RI/Rt-UtA-PI Pulsatility Index(LN,


12008-9)
OB/UTERINEART/D_UTA_PI/Lt-UtA-PI

OB/UTERINEART/D_UTA_PI/Rt-UtA-PI
Uterine Art.-PI
OB/UTERINEART/D_UTA_AT/Lt-UtA-PI

OB/UTERINEART/D_UTA_AT/Rt-UtA-PI

OB/UTERINEART/D_UTA_MT/Lt-UtA-PI

OB/UTERINEART/D_UTA_MT/Rt-UtA-PI

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 105 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

OB/UMBILICALARTLT/D_UMBALT_RI/Lt-UmA-PI

OB/UMBILICALARTLT/D_UMBALT_RI/Rt-UmA-PI

OB/UMBILICALARTLT/D_UMBALT_PI/Lt-UmA-PI

OB/UMBILICALARTLT/D_UMBALT_PI/Rt-UmA-PI
UmA-PI
OB/UMBILICALARTLT/D_UMBALT_AT/Lt-UmA-PI

OB/UMBILICALARTLT/D_UMBALT_AT/Rt-UmA-PI

OB/UMBILICALARTLT/D_UMBALT_MT/Lt-UmA-PI

OB/UMBILICALARTLT/D_UMBALT_MT/Rt-UmA-PI

OB/D_UMBV_TAMAX/UmbV-TAmax

OB/UMBILICALART/D_UMBA_PI/UmbA-TAmax
Umbilical
Art.-TAmax OB/UMBILICALART/D_UMBA_AT/UmbA-TAmax

OB/UMBILICALART/D_UMBA_MT/UmbA-TAmax

OB/UTERINEART/D_UTA_PI/Lt-UtA-Tamax

OB/UTERINEART/D_UTA_PI/Rt-UtA-TAmax

Uterine OB/UTERINEART/D_UTA_AT/Lt-UtA-TAmax
Art.-TAmax
OB/UTERINEART/D_UTA_AT/Rt-UtA-TAmax Time averaged peak
velocity(LN, 11692-1)
OB/UTERINEART/D_UTA_MT/Lt-UtA-TAmax

OB/UTERINEART/D_UTA_MT/Rt-UtA-TAmax

OB/UMBILICALARTLT/D_UMBALT_PI/Lt-UmA-TAmax

OB/UMBILICALARTLT/D_UMBALT_PI/Rt-UmA-TAmax

OB/UMBILICALARTLT/D_UMBALT_AT/Lt-UmA-TAmax
UmA-TAmax
OB/UMBILICALARTLT/D_UMBALT_AT/Rt-UmA-TAmax

OB/UMBILICALARTLT/D_UMBALT_MT/Lt-UmA-TAmax

OB/UMBILICALARTLT/D_UMBALT_MT/Rt-UmA-TAmax

OB/UMBILICALART/D_UMBA_AT/UmbA-TAmean
Umbilical
Art.-TAmean
OB/UMBILICALART/D_UMBA_MT/UmbA-TAmean Time averaged mean
velocity(LN, 20352-1)
Uterine OB/UTERINEART/D_UTA_AT/Lt-UtA-TAmean
Art.-TAmean
OB/UTERINEART/D_UTA_AT/Rt-UtA-TAmean

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 106 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

OB/UTERINEART/D_UTA_MT/Lt-UtA-TAmean

OB/UTERINEART/D_UTA_MT/Rt-UtA-TAmean

OB/UTERINEART/D_UTA_VOL/Lt-UtA-TAmean

OB/UTERINEART/D_UTA_VOL/Rt-UtA-TAmean

OB/UMBILICALARTLT/D_UMBALT_AT/Lt-UmA-TAmean

OB/UMBILICALARTLT/D_UMBALT_AT/Rt-UmA-TAmean

OB/UMBILICALARTLT/D_UMBALT_MT/Lt-UmA-TAmean
UmA-TAmean
OB/UMBILICALARTLT/D_UMBALT_MT/Rt-UmA-TAmean

OB/UMBILICALARTLT/D_UMBALT_VOL/Lt-UmA-TAmean

OB/UMBILICALARTLT/D_UMBALT_VOL/Rt-UmA-TAmean

OB/UMBILICALART/D_UMBA_PERS/UmbA-S/D

Umbilical Art.-S/D OB/UMBILICALART/D_UMBA_AT/UmbA-S/D

OB/UMBILICALART/D_UMBA_MT/UmbA-S/D

OB/UTERINEART/D_UTA_PERS/Lt-UtA-S/D

OB/UTERINEART/D_UTA_PERS/Rt-UtA-S/D

OB/UTERINEART/D_UTA_AT/Lt-UtA-S/D
Uterine Art.-S/D
OB/UTERINEART/D_UTA_AT/Rt-UtA-S/D
Systolic to Diastolic
OB/UTERINEART/D_UTA_MT/Lt-UtA-S/D Velocity Ratio(LN,
12144-2)
OB/UTERINEART/D_UTA_MT/Rt-UtA-S/D

OB/UMBILICALARTLT/D_UMBALT_PERS/Lt-UmA-S/D

OB/UMBILICALARTLT/D_UMBALT_PERS/Rt-UmA-S/D

OB/UMBILICALARTLT/D_UMBALT_AT/Lt-UmA-S/D
UmA-S/D
OB/UMBILICALARTLT/D_UMBALT_AT/Rt-UmA-S/D

OB/UMBILICALARTLT/D_UMBALT_MT/Lt-UmA-S/D

OB/UMBILICALARTLT/D_UMBALT_MT/Rt-UmA-S/D

OB/UMBILICALART/D_UMBA_AT/UmbA-AT
Umbilical Art.-AT
OB/UMBILICALART/D_UMBA_MT/UmbA-AT Acceleration Time(LN,
20168-1)
OB/UTERINEART/D_UTA_AT/Lt-UtA-AT
Uterine Art.-AT
OB/UTERINEART/D_UTA_AT/Rt-UtA-AT

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 107 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

OB/UTERINEART/D_UTA_MT/Lt-UtA-AT

OB/UTERINEART/D_UTA_MT/Rt-UtA-AT

OB/UMBILICALARTLT/D_UMBALT_AT/Lt-UmA-AT

OB/UMBILICALARTLT/D_UMBALT_AT/Rt-UmA-AT
UmA-AT
OB/UMBILICALARTLT/D_UMBALT_MT/Lt-UmA-AT

OB/UMBILICALARTLT/D_UMBALT_MT/Rt-UmA-AT

OB/UMBILICALART/D_UMBA_AT/UmbA-VTI
Umbilical Art.-VTI
OB/UMBILICALART/D_UMBA_MT/UmbA-VTI

OB/UTERINEART/D_UTA_AT/Lt-UtA-VTI

OB/UTERINEART/D_UTA_AT/Rt-UtA-VTI
Uterine Art.-VTI
OB/UTERINEART/D_UTA_MT/Lt-UtA-VTI Velocity Time
Integral(LN, 20354-7)
OB/UTERINEART/D_UTA_MT/Rt-UtA-VTI

OB/UMBILICALARTLT/D_UMBALT_AT/Lt-UmA-VTI

OB/UMBILICALARTLT/D_UMBALT_AT/Rt-UmA-VTI
UmA-VTI
OB/UMBILICALARTLT/D_UMBALT_MT/Lt-UmA-VTI

OB/UMBILICALARTLT/D_UMBALT_MT/Rt-UmA-VTI

OB/UMBILICALART/D_UMBA_AT/UmbA-HR

Umbilical Art.-HR OB/UMBILICALART/D_UMBA_MT/UmbA-HR

OB/UMBILICALART/D_UMBA_HR/UmbA-HR

OB/UTERINEART/D_UTA_AT/Lt-UtA-HR

OB/UTERINEART/D_UTA_AT/Rt-UtA-HR

OB/UTERINEART/D_UTA_MT/Lt-UtA-HR
Uterine Art.-HR
OB/UTERINEART/D_UTA_MT/Rt-UtA-HR Heart Rate(LN, 8867-4)

OB/UTERINEART/D_UTA_HR/Lt-UtA-HR

OB/UTERINEART/D_UTA_HR/Rt-UtA-HR

OB/UMBILICALARTLT/D_UMBALT_AT/Lt-UmA-HR

OB/UMBILICALARTLT/D_UMBALT_AT/Rt-UmA-HR
UmA-HR
OB/UMBILICALARTLT/D_UMBALT_MT/Lt-UmA-HR

OB/UMBILICALARTLT/D_UMBALT_MT/Rt-UmA-HR

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 108 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

OB/UMBILICALARTLT/D_UMBALT_HR/Lt-UmA-HR

OB/UMBILICALARTLT/D_UMBALT_HR/Rt-UmA-HR

Umbilical Art.-T OB/UMBILICALART/D_UMBA_TIME/UmbA-T

OB/UTERINEART/D_UTA_TIME/Lt-UtA-T
Uterine Art.-T
Acceleration Time(LN,
OB/UTERINEART/D_UTA_TIME/Rt-UtA-T
20168-1)
OB/UMBILICALARTLT/D_UMBALT_TIME/Lt-UmA-T
UmA-T
OB/UMBILICALARTLT/D_UMBALT_TIME/Rt-UmA-T

Uterine OB/UTERINEART/D_UTA_VOL/Lt-UtA-Vol. Flow


Art-Vol.Flow
OB/UTERINEART/D_UTA_VOL/Rt-UtA-Vol. Flow

OB/UMBILICALARTLT/D_UMBALT_VOL/Lt-UmA-Vol. Flow
Volume flow(LN,
UmA-Vol.Flow 33878-0)
OB/UMBILICALARTLT/D_UMBALT_VOL/Rt-UmA-Vol. Flow

Umbilical
OB/D_UMBV_TAMAX/UmbV-VenFlow
Vein-Vol.Flow

Atrial Fetal Heart


OB/M_ATFHR/Atrial FHR
Rate

GYN/GYN_M_OVARIANL/GYN_M_OVARIAN_VESSEL/Rt-OvA-D

GYN/GYN_M_OVARIANL/GYN_M_OVARIAN_VESSEL/Lt-OvA-D M-Mode vessel lumen


M-Mode vessel
diameter(SONO,
lumen diameter GYN/GYN_M_UTERINE_ART/GYN_M_UTERINE_VESSEL/Rt-UtA-D SONO-5105-8)

GYN/GYN_M_UTERINE_ART/GYN_M_UTERINE_VESSEL/Lt-UtA-D

GYN/GYN_M_OVARIANL/GYN_M_OVARIAN_STENOSIS/Lt-OvA-D1

Vessel lumen GYN/GYN_M_OVARIANL/GYN_M_OVARIAN_STENOSIS/Rt-OvA-D1 Vessel lumen


diameter diameter(SRT, G-0364)
GYN/GYN_M_UTERINE_ART/GYN_M_UTERINE_STENOSIS/Lt-UtA-D1

GYN/GYN_M_UTERINE_ART/GYN_M_UTERINE_STENOSIS/Rt-UtA-D1

GYN/GYN_M_OVARIANL/GYN_M_OVARIAN_STENOSIS/Lt-OvA-D2

Vessel outside GYN/GYN_M_OVARIANL/GYN_M_OVARIAN_STENOSIS/Rt-OvA-D2 Vessel outside


diameter GYN/GYN_M_UTERINE_ART/GYN_M_UTERINE_STENOSIS/Lt-UtA-D2 diameter(SRT, G-0365)

GYN/GYN_M_UTERINE_ART/GYN_M_UTERINE_STENOSIS/Rt-UtA-D2

Diameter GYN/GYN_M_OVARIANL/GYN_M_OVARIAN_STENOSIS/Lt-%Sten Diameter


Reduction Reduction(SRT, G-0372)
GYN/GYN_M_OVARIANL/GYN_M_OVARIAN_STENOSIS/Rt-%Sten

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 109 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

GYN/GYN_M_UTERINE_ART/GYN_M_UTERINE_STENOSIS/Lt-%Sten

GYN/GYN_M_UTERINE_ART/GYN_M_UTERINE_STENOSIS/Rt-%Sten

GYN/GYN_M_OVARIANL/GYN_M_OVARIAN_TIME/Lt-OvA-T

GYN/GYN_M_OVARIANL/GYN_M_OVARIAN_TIME/Rt-OvA-T M-Mode Cycle


M-Mode Cycle
Time(SONO,
Time
GYN/GYN_M_UTERINE_ART/GYN_M_UTERINE_TIME/Lt-UtA-T SONO-5105-9)

GYN/GYN_M_UTERINE_ART/GYN_M_UTERINE_TIME/Rt-UtA-T

GYN/GYN_M_OVARIANL/GYN_M_OVARIAN_HR/Lt-OvA-HR

GYN/GYN_M_OVARIANL/GYN_M_OVARIAN_HR/Rt-OvA-HR
Heart Rate Heart Rate(LN, 8867-4)
GYN/GYN_M_UTERINE_ART/GYN_M_UTERINE_HR/Lt-UtA-HR

GYN/GYN_M_UTERINE_ART/GYN_M_UTERINE_HR/Rt-UtA-HR

GYN/OVARIANART/GYN_D_OVARIAN_PS/Lt-OvA-PS

GYN/OVARIANART/GYN_D_OVARIAN_PS/Rt-OvA-PS

GYN/OVARIANART/GYN_D_OVARIAN_RI/Lt-OvA-PS

GYN/OVARIANART/GYN_D_OVARIAN_RI/Rt-OvA-PS

GYN/OVARIANART/GYN_D_OVARIAN_PI/Lt-OvA-PS

GYN/OVARIANART/GYN_D_OVARIAN_PI/Rt-OvA-PS Peak Systolic


Ovarian Art.-PS
Velocity(LN, 11726-7)
GYN/OVARIANART/GYN_D_OVARIAN_GROUP/Lt-OvA-PS

GYN/OVARIANART/GYN_D_OVARIAN_GROUP/Rt-OvA-PS

GYN/OVARIANART/GYN_D_OVARIAN_AT/Lt-OvA-PS

GYN/OVARIANART/GYN_D_OVARIAN_AT/Rt-OvA-PS

GYN/OVARIANART/GYN_D_OVARIAN_MT/Lt-OvA-PS

GYN/OVARIANART/GYN_D_OVARIAN_MT/Rt-OvA-PS

GYN/OVARIANART/GYN_D_OVARIAN_ED/Lt-OvA-ED

GYN/OVARIANART/GYN_D_OVARIAN_ED/Rt-OvA-ED

GYN/OVARIANART/GYN_D_OVARIAN_RI/Lt-OvA-ED
End Diastolic
Ovarian Art.-ED GYN/OVARIANART/GYN_D_OVARIAN_RI/Rt-OvA-ED
Velocity(LN, 11653-3)
GYN/OVARIANART/GYN_D_OVARIAN_PI/Lt-OvA-ED

GYN/OVARIANART/GYN_D_OVARIAN_PI/Rt-OvA-ED

GYN/OVARIANART/GYN_D_OVARIAN_GROUP/Lt-OvA-ED

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 110 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

GYN/OVARIANART/GYN_D_OVARIAN_GROUP/Rt-OvA-ED

GYN/OVARIANART/GYN_D_OVARIAN_AT/Lt-OvA-ED

GYN/OVARIANART/GYN_D_OVARIAN_AT/Rt-OvA-ED

GYN/OVARIANART/GYN_D_OVARIAN_MT/Lt-OvA-ED

GYN/OVARIANART/GYN_D_OVARIAN_MT/Rt-OvA-ED

GYN/OVARIANART/GYN_D_OVARIAN_RI/Lt-OvA-RI

GYN/OVARIANART/GYN_D_OVARIAN_RI/Rt-OvA-RI

GYN/OVARIANART/GYN_D_OVARIAN_GROUP/Lt-OvA-RI

GYN/OVARIANART/GYN_D_OVARIAN_GROUP/Rt-OvA-RI Resisitivity Index(LN,


Ovarian Art.-RI
GYN/OVARIANART/GYN_D_OVARIAN_AT/Lt-OvA-RI 12023-8)

GYN/OVARIANART/GYN_D_OVARIAN_AT/Rt-OvA-RI

GYN/OVARIANART/GYN_D_OVARIAN_MT/Lt-OvA-RI

GYN/OVARIANART/GYN_D_OVARIAN_MT/Rt-OvA-RI

GYN/OVARIANART/GYN_D_OVARIAN_MT/Lt-OvA-TAmax

GYN/OVARIANART/GYN_D_OVARIAN_MT/Rt-OvA-TAmax

Ovarian GYN/OVARIANART/GYN_D_OVARIAN_PI/Lt-OvA-TAmax Time averaged peak


Art.-TAmax velocity(LN, 11692-1)
GYN/OVARIANART/GYN_D_OVARIAN_PI/Rt-OvA-TAmax

GYN/OVARIANART/GYN_D_OVARIAN_AT/Lt-OvA-TAmax

GYN/OVARIANART/GYN_D_OVARIAN_AT/Rt-OvA-TAmax

GYN/OVARIANART/GYN_D_OVARIAN_MT/Lt-OvA-TAmean

Ovarian GYN/OVARIANART/GYN_D_OVARIAN_MT/Rt-OvA-TAmean Time averaged mean


Art.-TAmean velocity(LN, 20352-1)
GYN/OVARIANART/GYN_D_OVARIAN_AT/Lt-OvA-TAmean

GYN/OVARIANART/GYN_D_OVARIAN_AT/Rt-OvA-TAmean

GYN/OVARIANART/GYN_D_OVARIAN_RI/Lt-OvA-PI

GYN/OVARIANART/GYN_D_OVARIAN_RI/Rt-OvA-PI

GYN/OVARIANART/GYN_D_OVARIAN_PI/Lt-OvA-PI Pulsatility Index(LN,


Ovarian Art.-PI
12008-9)
GYN/OVARIANART/GYN_D_OVARIAN_PI/Rt-OvA-PI

GYN/OVARIANART/GYN_D_OVARIAN_AT/Lt-OvA-PI

GYN/OVARIANART/GYN_D_OVARIAN_AT/Rt-OvA-PI

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 111 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

GYN/OVARIANART/GYN_D_OVARIAN_MT/Lt-OvA-PI

GYN/OVARIANART/GYN_D_OVARIAN_MT/Rt-OvA-PI

GYN/OVARIANART/GYN_D_OVARIAN_GROUP/Lt-OvA-S/D

GYN/OVARIANART/GYN_D_OVARIAN_GROUP/Rt-OvA-S/D

GYN/OVARIANART/GYN_D_OVARIAN_AT/Lt-OvA-S/D Systolic to Diastolic


Ovarian Art.-S/D Velocity Ratio(LN,
GYN/OVARIANART/GYN_D_OVARIAN_AT/Rt-OvA-S/D 12144-2)

GYN/OVARIANART/GYN_D_OVARIAN_MT/Lt-OvA-S/D

GYN/OVARIANART/GYN_D_OVARIAN_MT/Rt-OvA-S/D

GYN/OVARIANART/GYN_D_OVARIAN_AT/Lt-OvA-VTI

GYN/OVARIANART/GYN_D_OVARIAN_AT/Rt-OvA-VTI Velocity Time


Ovarian Art.-VTI
GYN/OVARIANART/GYN_D_OVARIAN_MT/Lt-OvA-VTI Integral(LN, 20354-7)

GYN/OVARIANART/GYN_D_OVARIAN_MT/Rt-OvA-VTI

GYN/OVARIANART/GYN_D_OVARIAN_AT/Lt-OvA-AT

GYN/OVARIANART/GYN_D_OVARIAN_AT/Rt-OvA-AT Acceleration Time(LN,


Ovarian Art.-AT
20168-1)
GYN/OVARIANART/GYN_D_OVARIAN_MT/Lt-OvA-AT

GYN/OVARIANART/GYN_D_OVARIAN_MT/Rt-OvA-AT

GYN/OVARIANART/GYN_D_OVARIAN_AT/Lt-OvA-HR

GYN/OVARIANART/GYN_D_OVARIAN_AT/Rt-OvA-HR

GYN/OVARIANART/GYN_D_OVARIAN_MT/Lt-OvA-HR
Ovarian Art.-HR Heart Rate(LN, 8867-4)
GYN/OVARIANART/GYN_D_OVARIAN_MT/Rt-OvA-HR

GYN/OVARIANART/GYN_D_OVARIAN_HR/Lt-OvA-HR

GYN/OVARIANART/GYN_D_OVARIAN_HR/Rt-OvA-HR

GYN/D_UTERINEART_A/GYN_D_UTERINE_PS/Lt-UtA-PS

GYN/D_UTERINEART_A/GYN_D_UTERINE_PS/Rt-UtA-PS

GYN/D_UTERINEART_A/GYN_D_UTERINE_RI/Lt-UtA-PS
Peak Systolic
Uterine Art.-PS GYN/D_UTERINEART_A/GYN_D_UTERINE_RI/Rt-UtA-PS
Velocity(LN, 11726-7)
GYN/D_UTERINEART_A/GYN_D_UTERINE_PI/Lt-UtA-PS

GYN/D_UTERINEART_A/GYN_D_UTERINE_PI/Rt-UtA-PS

GYN/D_UTERINEART_A/GYN_D_UTERINE_GROUP/Lt-UtA-PS

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 112 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

GYN/D_UTERINEART_A/GYN_D_UTERINE_GROUP/Rt-UtA-PS

GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Lt-UtA-PS

GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Rt-UtA-PS

GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Lt-UtA-PS

GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Rt-UtA-PS

GYN/D_UTERINEART_A/GYN_D_UTERINE_ED/Lt-UtA-ED

GYN/D_UTERINEART_A/GYN_D_UTERINE_ED/Rt-UtA-ED

GYN/D_UTERINEART_A/GYN_D_UTERINE_RI/Lt-UtA-ED

GYN/D_UTERINEART_A/GYN_D_UTERINE_RI/Rt-UtA-ED

GYN/D_UTERINEART_A/GYN_D_UTERINE_PI/Lt-UtA-ED

GYN/D_UTERINEART_A/GYN_D_UTERINE_PI/Rt-UtA-ED End Diastolic


Uterine Art.-ED
Velocity(LN, 11653-3)
GYN/D_UTERINEART_A/GYN_D_UTERINE_GROUP/Lt-UtA-ED

GYN/D_UTERINEART_A/GYN_D_UTERINE_GROUP/Rt-UtA-ED

GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Lt-UtA-ED

GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Rt-UtA-ED

GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Lt-UtA-ED

GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Rt-UtA-ED

GYN/D_UTERINEART_A/GYN_D_UTERINE_RI/Lt-UtA-RI

GYN/D_UTERINEART_A/GYN_D_UTERINE_RI/Rt-UtA-RI

GYN/D_UTERINEART_A/GYN_D_UTERINE_GROUP/Lt-UtA-RI

GYN/D_UTERINEART_A/GYN_D_UTERINE_GROUP/Rt-UtA-RI Resistivity Index(LN,


Uterine Art.-RI
GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Lt-UtA-RI 12023-8)

GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Rt-UtA-RI

GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Lt-UtA-RI

GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Rt-UtA-RI

GYN/D_UTERINEART_A/GYN_D_UTERINE_PI/Lt-UtA-TAmax

Uterine GYN/D_UTERINEART_A/GYN_D_UTERINE_PI/Rt-UtA-TAmax Time averaged peak


Art.-TAmax velocity(LN, 11692-1)
GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Lt-UtA-TAmax

GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Rt-UtA-TAmax

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 113 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Lt-UtA-TAmax

GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Rt-UtA-TAmax

GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Lt-UtA-TAmean

GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Rt-UtA-TAmean

Uterine GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Lt-UtA-TAmean Time averaged mean


Art.-TAmean GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Rt-UtA-TAmean velocity(LN, 20352-1)

GYN/D_UTERINEART_A/GYN_D_UTERINE_VOL/Lt-UtA-TAmean

GYN/D_UTERINEART_A/GYN_D_UTERINE_VOL/Rt-UtA-TAmean

GYN/D_UTERINEART_A/GYN_D_UTERINE_RI/Lt-UtA-PI

GYN/D_UTERINEART_A/GYN_D_UTERINE_RI/Rt-UtA-PI

GYN/D_UTERINEART_A/GYN_D_UTERINE_PI/Lt-UtA-PI

GYN/D_UTERINEART_A/GYN_D_UTERINE_PI/Rt-UtA-PI Pulsatility Index(LN,


Uterine Art.-PI
12008-9)
GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Lt-UtA-PI

GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Rt-UtA-PI

GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Lt-UtA-PI

GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Rt-UtA-PI

GYN/D_UTERINEART_A/GYN_D_UTERINE_GROUP/Lt-UtA-S/D

GYN/D_UTERINEART_A/GYN_D_UTERINE_GROUP/Rt-UtA-S/D

GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Lt-UtA-S/D Systolic to Diastolic


Uterine Art.-S/D Velocity Ratio(LN,
GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Rt-UtA-S/D 12144-2)

GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Lt-UtA-S/D

GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Rt-UtA-S/D

GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Lt-UtA-AT

GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Rt-UtA-AT Acceleration Time(LN,


Uterine Art.-AT
GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Lt-UtA-AT 20168-1)

GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Rt-UtA-AT

GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Lt-UtA-VTI
Velocity Time
Uterine Art.-VTI GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Rt-UtA-VTI
Integral(LN, 20354-7)
GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Lt-UtA-VTI

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 114 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Rt-UtA-VTI

GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Lt-UtA-HR

GYN/D_UTERINEART_A/GYN_D_UTERINE_AT/Rt-UtA-HR

GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Lt-UtA-HR
Uterine Art.-HR Heart rate(LN, 8867-4)
GYN/D_UTERINEART_A/GYN_D_UTERINE_MT/Rt-UtA-HR

GYN/D_UTERINEART_A/GYN_D_UTERINE_HR/Lt-UtA-HR

GYN/D_UTERINEART_A/GYN_D_UTERINE_HR/Rt-UtA-HR

Uterine Art. -Vol. GYN/D_UTERINEART_A/GYN_D_UTERINE_VOL/Lt-UtA-Vol. Flow Volume flow(LN,


Flow GYN/D_UTERINEART_A/GYN_D_UTERINE_VOL/Rt-UtA-Vol. Flow 33878-0)

GYN/D_UTERINEART_A/GYN_D_UTERINE_TIME/Lt-UtA-T Acceleration Time(LN,


Uterine Art.-T
20168-1)
GYN/D_UTERINEART_A/GYN_D_UTERINE_TIME/Rt-UtA-T

GYN/VESSEL_D/GYN_D_VESSEL_PS/Ves-PS

GYN/VESSEL_D/GYN_D_VESSEL_RI/Ves-PS

GYN/VESSEL_D/GYN_D_VESSEL_PI/Ves-PS Peak Systolic


Vessel-PS
Velocity(LN, 11726-7)
GYN/VESSEL_D/GYN_D_VESSEL_GROUP/Ves-PS

GYN/VESSEL_D/GYN_D_VESSEL_AT/Ves-PS

GYN/VESSEL_D/GYN_D_VESSEL_MT/Ves-PS

GYN/VESSEL_D/GYN_D_VESSEL_ED/Ves-ED

GYN/VESSEL_D/GYN_D_VESSEL_RI/Ves-ED

GYN/VESSEL_D/GYN_D_VESSEL_PI/Ves-ED End Diastolic


Vessel-ED
Velocity(LN, 11653-3)
GYN/VESSEL_D/GYN_D_VESSEL_GROUP/Ves-ED

GYN/VESSEL_D/GYN_D_VESSEL_AT/Ves-ED

GYN/VESSEL_D/GYN_D_VESSEL_MT/Ves-ED

GYN/VESSEL_D/GYN_D_VESSEL_RI/Ves-RI

GYN/VESSEL_D/GYN_D_VESSEL_GROUP/Ves-RI Resistivity Index(LN,


Vessel-RI
12023-8)
GYN/VESSEL_D/GYN_D_VESSEL_AT/Ves-RI

GYN/VESSEL_D/GYN_D_VESSEL_MT/Ves-RI

GYN/VESSEL_D/GYN_D_VESSEL_RI/Ves-PI Pulsatility Index(LN,


Vessel-PI
GYN/VESSEL_D/GYN_D_VESSEL_PI/Ves-PI 12008-9)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 115 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

GYN/VESSEL_D/GYN_D_VESSEL_AT/Ves-PI

GYN/VESSEL_D/GYN_D_VESSEL_MT/Ves-PI

GYN/VESSEL_D/GYN_D_VESSEL_PI/Ves-TAmax
Time averaged peak
Vessel-TAmax GYN/VESSEL_D/GYN_D_VESSEL_AT/Ves-TAmax
velocity(LN, 11692-1)
GYN/VESSEL_D/GYN_D_VESSEL_MT/Ves-TAmax

GYN/VESSEL_D/GYN_D_VESSEL_AT/Ves-TAmean Time averaged mean


Vessel-TAmean
velocity(LN, 20352-1)
GYN/VESSEL_D/GYN_D_VESSEL_MT/Ves-TAmean

GYN/VESSEL_D/GYN_D_VESSEL_AT/Ves-AT
Acceleration Time(LN,
Vessel-AT GYN/VESSEL_D/GYN_D_VESSEL_MT/Ves-AT
20168-1)
GYN/VESSEL_D/GYN_D_VESSEL_TIME/Ves-T

GYN/VESSEL_D/GYN_D_VESSEL_AT/Ves-VTI Velocity Time


Vessel-VTI
GYN/VESSEL_D/GYN_D_VESSEL_MT/Ves-VTI Integral(LN, 20354-7)

GYN/VESSEL_D/GYN_D_VESSEL_GROUP/Ves-S/D
Systolic to Diastolic
Vessel-S/D GYN/VESSEL_D/GYN_D_VESSEL_AT/Ves-S/D Velocity Ratio(LN,
12144-2)
GYN/VESSEL_D/GYN_D_VESSEL_MT/Ves-S/D

GYN/VESSEL_D/GYN_D_VESSEL_AT/Ves-HR

Vessel-HR GYN/VESSEL_D/GYN_D_VESSEL_MT/Ves-HR Heart rate(LN, 8867-4)

GYN/VESSEL_D/GYN_D_VESSEL_HR/Ves-HR

A.3 TID 5100 Vascular Ultrasound Report

A.3 .1 TID 5101 Vascular Patient Characteristics


Label Code Meaning (Coding Scheme Designator, Code
Value)

Patient Characteristics Container: Patient Characteristics (DCM,121118)

Age Subject Age (DCM, 121033)

Sex Subject Sex (DCM, 121032)

BP Systolic Blood Pressure (SRT, F-008EC)

Diastolic Blood Pressure (SRT, F-008ED)

Heart Rate Heart rate (LN, 8867-4)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 116 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

A.3 .2 TID 5102 Vascular Procedure Summary Section


Label Code Meaning (Coding Scheme Designator, Code
Value)

Summary Container: Summary (DCM, 121111)

Comments Comment (DCM, 121106)

A.3 .3 Anatomy Measure Item


Coding Code Value Code Meaning
Scheme (0008,0100) (0008,0104)
Designator
(0008,0102)

LN 11653-3 End Diastolic Velocity

LN 11726-7 Peak Systolic Velocity

LN 20352-1 Time averaged mean velocity

LN 11692-1 Time averaged peak velocity

LN 12008-9 Pulsatility Index

LN 12023-8 Resistivity Index

LN 12144-2 Systolic to Diastolic Velocity Ratio

LN 8867-4 Heart Rate

SRT G-0371 % Area Reduction

SRT G-0372 % Diameter Reduction

SRT G-0365 Vessel outside diameter

SRT G-0364 Vessel lumen diameter

SRT R-1025C Vessel Intimal Diameter

SRT R-1025D Vessel Intimal Cross-Sectional Area

SRT G-0366 Vessel lumen cross-sectional area

A.3 .4 Blood Vessel of Head DCID (12105) Intracranial Cerebral Vessels DCID (12106) Unpaired Intracranial
Cerebral Vessels (unilateral)
Laterality Label Anatomy Item ID Code Meaning (Coding Scheme
Designator, Code Value)

Blood Vessel of Head Finding Site: Blood Vessel of Head


(SRT, T-40501)

Left ICA Internal Internal Carotid Artery (SRT,

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 117 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

(SRT ,G-A Carotid Artery T-45300)


101)
CS Carotid Carotid Siphon(SRT, T-45308)
Right Siphon

(SRT,G-A PCA Posterior Posterior Communicating


100) Communicati Artery(SRT, T-45320)
ng Artery

PCA Posterior Posterior Cerebral Artery(SRT,


Cerebral T-45900)
Artery

ACA Anterior Anterior Communicating


Communicati Artery(SRT, T-45530)
ng Artery

ACA Anterior Anterior Cerebral Artery(SRT,


Cerebral T-45540)
Artery

MCA Meddle Meddle Cerebral Artery(SRT,


Cerebral T-45600)
Artery

BA Basilar Artery Basilar Artery(SRT, T-45800)

A.3.5 Artery of Neck DCID (12104) Extracranial Arteries


Laterality Label Anatomy Item ID Code Meaning (Coding Scheme
Designator, Code Value)

Artery of neck Finding Site: Artery of neck (SRT,


T-45005)

Left ICA Internal Internal Carotid Artery (SRT,


(SRT ,G-A Carotid Artery T-45300)
101)
ECA External External Carotid Artery (SRT,
Right(SRT Carotid Artery T-45200)
,G-A100)
CCA Common Common Carotid Artery (SRT,
Carotid Artery T-45100)

VA Vertebral Vertebral Artery(SRT, T-45700)


Artery

CB Carotid Bulb Carotid Bulb(SRT, T-45170)

SA Subclavian Subclavian Artery(SRT, T-46100)


Artery

SV Subclavian Subclavian Vein(SRT, T-48330)


Vein

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 118 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

A.3.6 Artery of Lower Extremity DCID (12109) Lower Extremity Arteries


Laterality Label Anatomy Item ID Code Meaning (Coding Scheme
Designator, Code Value)

Artery of Lower Extremity Finding Site: Artery of Lower


Extremity (SRT, T- 47040)

Left INT_IL Internal Iliac Internal Iliac Artery (SRT, T-46740)


(SRT ,G-A Artery
101)
EXT_IL External Iliac External Iliac Artery (SRT, T-46910)
Right(SRT Artery
,G-A100)
IL_IAC Common Iliac Common Iliac Artery (SRT,
Artery T-46710)

CFA Common Common Femoral Artery (SRT,


Femoral T-47400)
Artery

PROFUN Profunda Profunda Femoris Artery (SRT,


Femoris T-47440)
Artery

SFA Superficial Superficial Femoral Artery (SRT,


Femoral T-47403)
Artery

ATA Anterior Anterior Tibial Artery (SRT,


Tibial Artery T-47700)

PTA Posterior Posterior Tibial Artery (SRT,


Tibial Artery T-47600)

PERON Peroneal Peroneal Artery (SRT, T-47630)


Artery

A.3. 7 Vein of Lower Extremity DCID (12110) Lower Extremity Veins


Laterality Label Anatomy Item ID Code Meaning (Coding Scheme
Designator, Code Value)

Vein of Lower Extremity Finding Site: Vein of Lower


Extremity (SRT, T-49403)

POP Popliteal Vein Popliteal Vein (SRT, T-49640)


Left
(SRT ,G-A ClV Common Iliac Common Iliac Vein(SRT, T-48920)
101) Vein
Right(SRT
ElV External Iliac External Iliac Vein(SRT, T-48930)
,G-A100)
Vein

llV Internal Iliac Internal Iliac Vein(SRT, T-48940)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 119 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Vein

CFV Common Common Femoral Vein(SRT,


Femoral Vein G-035B)

SFV Superficial Superficial Femoral Vein(SRT,


Femoral Vein G-035A)

PFV Profunda Profunda Femoris Vein(SRT,


Femoris Vein T-49660)

PV Popliteal Vein Popliteal Vein(SRT, T-49640)

ATV Anterior Tibial Anterior Tibial Vein(SRT, T-49630)


Vein

PTV Posterior Posterior Tibial Vein(SRT, T-49620)


Tibial Vein

PV Peroneal Peroneal Vein(SRT, T-49650)


Vein

GSVoT Great Great Saphenous Vein of


Saphenous Thigh(SRT, R-10259)
Vein of Thigh

GSVoC Great Great Saphenous Vein of Calf(SRT,


Saphenous R-1025A)
Vein of Calf

LSV Lesser Lesser Saphenous Vein(SRT,


Saphenous T-49550)
Vein

A.3. 8 Artery of Upper Extremity DCID (12107) Upper Extremity Arteries


Laterality Label Anatomy Item ID Code Meaning (Coding Scheme
Designator, Code Value)

Artery Of Upper Extremity Finding Site: Artery Of Upper


Extremity (SRT, T-47020)

IA Innominate Innominate Artery(SRT, T-46010)


Artery
Left
AA Axillary Artery Axillary Artery (SRT, T-47100)
(SRT ,G-A
101) BA Brachial Brachial Artery (SRT, T-47160)
Artery
Right(SRT
,G-A100)
RA Radial Artery Radial Artery (SRT, T-47300)

UA Ulnar Artery Ulnar Artery (SRT, T-47200)

SPA Superficial Superficial Palmar Arch(SRT,

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 120 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Palmar Arch T-47240)

DPAoRA Deep Palmar Deep Palmar Arch of Radial Artery


Arch of (SRT, T-47340)
Radial Artery

A.3. 9 Vein of Upper Extremity DCID (12108) Upper Extremity Veins


Laterality Label Anatomy Item ID Code Meaning (Coding Scheme
Designator, Code Value)

Vein Of Upper Extremity Finding Site: Vein Of Upper


Extremity (SRT, T-49103)

IV Innominate Innominate Vein(SRT, T-48620)


Vein

IJV Internal Internal Jugular Vein (SRT,


Jugular Vein T-48170)

AV Axillary Vein Axillary Vein (SRT, T-49110)


Left
(SRT ,G-A CV Cephalic Vein Cephalic Vein (SRT, T-49240)
101)
BV Basilic Vein Basilic Vein (SRT, T-48052)
Right(SRT
,G-A100) BV Brachial Vein Brachial Vein (SRT, T-49350)

MCV Median Median Cubital Vein (SRT, T-49250)


Cubital Vein

RV Radial Vein Radial Vein (SRT, T-49340)

UV Ulnar Vein Ulnar Vein (SRT, T-49330)

A.3. 10 Vascular Structure of Kidney DCID (12115) Renal Vessels


Laterality Label Anatomy Item ID Code Meaning (Coding Scheme
Designator, Code Value)

Vascular Structure Of Kidney Finding Site: Vascular Structure Of


Kidney (SRT, T-71019)

Left
(SRT ,G-A
101) RA Renal Artery Renal Artery (SRT, T-46640)
Right(SRT
,G-A100)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 121 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

A.3. 11 Artery of Abdomen DCID (12111) Abdominal Arteries (lateral) DCID (12112) Abdominal Arteries (unilateral)
Laterality Label Anatomy Item ID Code Meaning (Coding Scheme
Designator, Code Value)

Artery of Abdomen Finding Site: Artery of Abdomen


(SRT, T-46002)

Aorta Aorta Aorta(SRT, T-42000)

CA Celiac Axis Celiac Axis(SRT, T-46400)

SMA Superior Superior Mesenteric Artery (SRT,


Mesenteric T-46510)
Artery
Left
(SRT ,G-A IMA Inferior Inferior Mesenteric Artery (SRT,
101) Mesenteric T-46520)
Artery
Right(SRT
,G-A100) ClA Common Iliac Common Iliac Artery (SRT,
Artery T-46710)

GA Gastroduode Gastroduodenal Artery (SRT,


nal Artery T-46440)

PHA Proper Proper Hepatic Artery (SRT,


Hepatic Artery T-46422 )

A.3. 12 Vein of Abdomen DCID (12113) Abdominal Veins (lateral) DCID (12114) Abdominal Veins (unilateral)
Laterality Label Anatomy Item ID Code Meaning (Coding Scheme
Designator, Code Value)

Vein of Abdomen Finding Site: Vein of Abdomen


(SRT, T- 487A0)

Left DPA Dorsal Pedal Dorsal Pedal Artery (SRT, T-47741)


(SRT ,G-A Artery
101)
lVC Inferior Vena Inferior Vena Cava (SRT, T-48710)
Right(SRT Cava
,G-A100)

iVC Inferior Vena Inferior Vena Cava(SRT, T-48710)


Cava

HV Hepatic Vein Hepatic Vein(SRT, T-48720)


Unilateral

(SRT,G-A MHV Middle Middle Hepatic Vein(SRT, T-48726)


103) Hepatic Vein

PV Portal Vein Portal Vein(SRT, T-48810)

SV Splenic Vein Splenic Vein(SRT, T-48890)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 122 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

SMV Superior Superior Mesenteric Vein (SRT,


Mesenteric T-48840)
Vein

TlPS Transjugular Transjugular Intrahepatic


Intrahepatic Portosystemic Shunt (SRT,
Portosystemi G-036C)
c Shunt

A.4 TID 5200 Echocardiography Procedure Report


A.4.1 TID 5201 Echocardiography Patient Characteristics
Label Code Meaning (Coding Scheme Designator, Code
Value)

Patient Characteristics Container: Patient Characteristics (DCM,121118)

Age Subject Age (DCM, 121033)

Sex Subject Sex (DCM, 121032)

BP Systolic Blood Pressure (SRT, F-008EC)

DBP Diastolic Blood Pressure (SRT, F-008ED)

Heart Rate Heart rate (LN, 8867-4)

BSA Body Surface Area(LN, 8277-6)

A.4.2 Left Ventricle DCID (12200) Echocardiography Left Ventricle


Label Item ID Code Meaning (Coding Scheme Designator,
Code Value)

Left Ventricle Finding Site: Left Ventricle (SRT, T-32600)

SPEllipse SPEllipse Area-Length Single Plane(DCM,125205)

2D mode(SRT,G-03A2)

Card/CARDIAC_LV_MASS_AL/C/LV
LVMI Left Ventricle Mass Index(SONO, SONO-5202-28)
MI(A-L)

Area-Length Area-Length Biplane Area-Length Biplane(DCM,125204)


Biplane
2D mode(SRT,G-03A2)

Card/CARDIAC_LV_AL/B/LVd Left Ventricle Internal End Diastolic Dimension (LN,


LVd Length
Length 29436-3)

LVd Area Card/CARDIAC_LV_AL/B/LVd Area Left Ventricular Diastolic Area(LN, 29436-3)

Card/CARDIAC_LV_AL/B/LVs Left Ventricle Internal Systolic Dimension(LN,


LVs Length
Length 29438-9)

LVs Area Card/CARDIAC_LV_AL/B/LVs Area Left Ventricular Systolic Area(LN, 29438-9)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 123 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Card/CARDIAC_LV_BULLET/B/LVA
LVAd Left Ventricular Diastolic Area(LN, G-0375)
d

Card/CARDIAC_LV_BULLET/B/LVA
LVAs Left Ventricular Systolic Area(LN, G-0374)
s

EDV Card/CARDIAC_LV_AL/B/EDV(A-L) Left Ventricular End Diastolic Volume (LN, 18026-5)

ESV Card/CARDIAC_LV_AL/B/ESV(A-L) Left Ventricular End Systolic Volume (LN, 18148-7)

SV Card/CARDIAC_LV_AL/B/SV(A-L) Stroke Volume (SRT, F-32120)

CO Card/CARDIAC_LV_AL/B/CO(A-L) Cardiac Output (SRT, F-32100)

EF Card/CARDIAC_LV_AL/B/EF(A-L) Left Ventricular Ejection Fraction (LN, 18043-0)

SI Card/CARDIAC_LV_AL/B/SI(A-L) Stroke Index (SRT, F-00078)

CI Card/CARDIAC_LV_AL/B/CI(A-L) Cardiac Index (SRT, F-32110)

Method of Disks, Biplane(DCM,125207)


Simpson Simpson
2D mode(SRT,G-03A2)

Card/CARDIAC_LV_SIMP/B/EDV(A4
C)

Card/CARDIAC_LV_SIMP/B/EDV(A2
EDV Left Ventricular End Diastolic Volume (LN, 18026-5)
C)

Card/CARDIAC_LV_SIMP/B/EDV(B
P)

Card/CARDIAC_LV_SIMP/B/ESV(A4
C)

Card/CARDIAC_LV_SIMP/B/ESV(A2
ESV Left Ventricular End Systolic Volume (LN, 18148-7)
C)

Card/CARDIAC_LV_SIMP/B/ESV(BP
)

Card/CARDIAC_LV_SIMP/B/SV(A4C
)

SV Card/CARDIAC_LV_SIMP/B/SV(A2C Stroke Volume (SRT, F-32120)


)

Card/CARDIAC_LV_SIMP/B/SV(BP)

Card/CARDIAC_LV_SIMP/B/CO(A4
C)

CO Card/CARDIAC_LV_SIMP/B/CO(A2 Cardiac Output (SRT, F-32100)


C)

Card/CARDIAC_LV_SIMP/B/CO(BP)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 124 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Card/CARDIAC_LV_SIMP/B/EF(A4C
)

EF Card/CARDIAC_LV_SIMP/B/EF(A2C Left Ventricular Ejection Fraction (LN, 18043-0)


)

Card/CARDIAC_LV_SIMP/B/EF(BP)

Card/CARDIAC_LV_SIMP/B/SI(A4C)

SI Card/CARDIAC_LV_SIMP/B/SI(A2C) Stroke Index (SRT, F-00078)

Card/CARDIAC_LV_SIMP/B/SI(BP)

Card/CARDIAC_LV_SIMP/B/CI(A4C)

CI Card/CARDIAC_LV_SIMP/B/CI(A2C) Cardiac Index (SRT, F-32110)

Card/CARDIAC_LV_SIMP/B/CI(BP)

Cube Method(DCM,125206)
CUBE CUBE
2D mode(SRT,G-03A2)

Left Ventricle Internal End Diastolic Dimension (LN,


LVIDd Card/CARDIAC_LV_CUBE/B/LVIDd
29436-3)

Card/CARDIAC_LV_CUBE/B/LVPW Left Ventricle Posterior Wall Diastolic Thickness (LN,


LVPWd
d 18152-9)

Interventricular Septum Diastolic Thickness(LN,


IVSd Card/CARDIAC_LV_CUBE/B/IVSd
18154-5)

Card/CARDIAC_LV_MASS_CUBE
LVM Left Ventricle Mass(LN, 18087-0)
/B/LVM(Cube)

Card/CARDIAC_LV_MASS_CUBE/B
LVMI Left Ventricle Mass Index(SONO, SONO-5202-30)
/LVMI(Cube)

Teichholz(DCM,125209)
TEICHHOLZ TEICHHOLZ
2D mode(SRT,G-03A2)

Card/CARDIAC_DIMENSION1/B/LVI Left Ventricle Internal End Diastolic Dimension (LN,


LVIDd
Dd 29436-3)

Card/CARDIAC_DIMENSION1/B/LVI Left Ventricle Internal Systolic Dimension (LN,


LVIDs
Ds 29438-9)

Card/CARDIAC_DIMENSION1/B/ED
EDV Left Ventricular End Diastolic Volume (LN, 18026-5)
V(Teich)

Card/CARDIAC_DIMENSION1/B/ES
ESV Left Ventricular End Systolic Volume (LN, 18148-7)
V(Teich)

SV Card/CARDIAC_DIMENSION1/B/SV( Stroke Volume (SRT, F-32120)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 125 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Teich)

Card/CARDIAC_DIMENSION1/B/CO
CO Cardiac Output (SRT, F-32100)
(Teich)

Card/CARDIAC_DIMENSION1/B/EF(
EF Left Ventricular Ejection Fraction (LN, 18043-0)
Teich)

Card/CARDIAC_DIMENSION1/B/SI(
SI Stroke Index (SRT, F-00078)
Teich)

Card/CARDIAC_DIMENSION1/B/CI(
CI Cardiac Index (SRT, F-32110)
Teich)

Card/CARDIAC_DIMENSION1/B/IVS Interventricular Septum Diastolic Thickness(LN,


IVSd
d 18154-5)

Card/CARDIAC_DIMENSION1/B/LV Left Ventricle Posterior Wall Diastolic Thickness(LN,


LVPWd
PWd 18152-9)

Card/CARDIAC_DIMENSION1/B/FS(
FS Left Ventricular Fractional Shortening(LN, 18051-3)
Teich)

Card/CARDIAC_DIMENSION1/B/IVS
IVSs Interventricular Septum Systolic Thickness(18158-6)
s

Card/CARDIAC_DIMENSION1/B/LV Left Ventricle Posterior Wall Systolic Thickness(LN,


LVPWs
PWs 18156-0)

Card/CARDIAC_DIMENSION1/B/ Interventricular Septum % Thickening(LN,


IVS%
IVS% 18054-7)

Card/CARDIAC_DIMENSION1/B/ Left Ventricle Posterior Wall % Thickening(LN,


LVPW%
LVPW% 18083-9)

Card/CARDIAC_DIMENSION1/B/ Interventricular Septum to Posterior Wall


IVS/LVPW
IVS/LVPW Thickness Ratio(LN, 18155-2)

Card/CARDIAC_DIMENSION1/B/
LVM(Cube) Left Ventricle Mass(LN, 18087-7)
LVM(Cube)

Card/CARDIAC_DIMENSION1/B/ Left Ventricle Mass Index(SONO,


LVMI(Cube)
LVMI(Cube) SONO-5202-30)

Cube Method(DCM,125206)
CUBE CUBE
M mode(SRT,G-0394)

Card/CARDIAC_M_VOLUME_CUBE Interventricular Septum Diastolic Thickness (LN,


IVSd
/M/IVSd 18154-5)

Card/CARDIAC_M_VOLUME_CUBE Left Ventricle Internal End Diastolic Dimension (LN,


LVIDd
/M/LVIDd 29436-3)

LVPWd Card/CARDIAC_M_VOLUME_CUBE Left Ventricle Posterior Wall Diastolic Thickness (LN,

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 126 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

/M/LVPWd 18152-9)

Card/CARDIAC_M_VOLUME_CUBE Interventricular Septum Systolic Thickness (LN,


IVSs
/M/IVSs 18158-6)

Card/CARDIAC_M_VOLUME_CUBE Left Ventricle Internal Systolic Dimension (LN,


LVIDs
/M/LVIDs 29438-9)

Card/CARDIAC_M_VOLUME_CUBE Left Ventricle Posterior Wall Systolic Thickness (LN,


LVPWs
/M/LVPWs 18156-0)

Card/CARDIAC_M_VOLUME_CUBE
EDV Left Ventricular End Diastolic Volume (LN, 18026-5)
/M/EDV(Cube)

Card/CARDIAC_M_VOLUME_CUBE
ESV Left Ventricular End Systolic Volume (LN, 18148-7)
/M/ESV(Cube)

Card/CARDIAC_M_VOLUME_CUBE
SV Stroke Volume (SRT, F-32120)
/M/SV(Cube)

Card/CARDIAC_M_VOLUME_CUBE
CO Cardiac Output (SRT, F-32100)
/M/CO(Cube)

Card/CARDIAC_M_VOLUME_CUBE
EF Left Ventricular Ejection Fraction (LN, 18043-0)
/M/EF(Cube)

Card/CARDIAC_M_VOLUME_CUBE
FS Left Ventricular Fractional Shortening (LN, 18051-3)
/M/FS(Cube)

Mean Velocity of Circumferential Fiber


MVCF Card/MCUBE/M/MVCF
Shortening(SONO,5202-1)

Card/CARDIAC_M_VOLUME_CUBE
SI Stroke Index (SRT, F-00078)
/M/SI(Cube)

Card/CARDIAC_M_VOLUME_CUBE
CI Cardiac Index (SRT, F-32110)
/M/CI(Cube)

Card/CARDIAC_M_VOLUME_CUBE Interventricular Septum to Posterior Wall Thickness


LVPW
/M/IVS/LVPW Ratio(LN, 18155-2)

Card/CARDIAC_M_VOLUME_CUB Interventricular Septum % Thickening(LN,


IVS%
E/M/IVS% 18054-7)

Card/CARDIAC_M_VOLUME_CUBE Left Ventricle Posterior Wall % Thickening(LN,


LVPW%
/M/LVPW% 18053-9)

Teichholz(DCM,125209)
TEICHHOLZ TEICHHOLZ
M mode(SRT,G-0394)

Card/CARDIAC_M_DIMENSION1/M/ Left Ventricle Internal End Diastolic Dimension (LN,


LVIDd
LVIDd 29436-3)

Card/CARDIAC_M_DIMENSION1/M/ Left Ventricle Internal Systolic Dimension (LN,


LVIDs
LVIDs 29438-9)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 127 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Card/CARDIAC_M_DIMENSION1/M/
EDV Left Ventricular End Diastolic Volume (LN, 18026-5)
EDV(Teich)

Card/CARDIAC_M_DIMENSION1/M/
ESV Left Ventricular End Systolic Volume (LN, 18148-7)
ESV(Teich)

Card/CARDIAC_M_DIMENSION1/M/
SV Stroke Volume (SRT, F-32120)
SV(Teich)

Card/CARDIAC_M_DIMENSION1/M/
CO Cardiac Output (SRT, F-32100)
CO(Teich)

Card/CARDIAC_M_DIMENSION1/M/
EF Left Ventricular Ejection Fraction (LN, 18043-0)
EF(Teich)

Card/CARDIAC_M_DIMENSION1/M/
FS Left Ventricular Fractional Shortening (LN, 18051-3)
FS(Teich)

Card/CARDIAC_M_DIMENSION1/M/
SI Stroke Index (SRT, F-00078)
SI(Teich)

Card/CARDIAC_M_DIMENSION1/M/
CI Cardiac Index (SRT, F-32110)
CI(Teich)

Card/CARDIAC_M_DIMENSION1/M/ Interventricular Septum Diastolic Thickness(LN,


IVSd
IVSd 18154-5)

Card/CARDIAC_M_DIMENSION1/M/ Left Ventricle Posterior Wall Diastolic Thickness(LN,


LVPWd
LVPWd 18152-9)

Card/CARDIAC_M_DIMENSION1/M/ Interventricular Septum Systolic Thickness(LN,


IVSs
IVSs 18158-6)

Card/CARDIAC_M_DIMENSION1/M/ Interventricular Septum to Posterior Wall Thickness


LVPW
IVS/LVPW Ratio(LN, 18155-2)

Card/CARDIAC_M_DIMENSION1/M/ Left Ventricle Posterior Wall Systolic Thickness(LN,


LVPWs
LVPWs 18156-0)

Method of Disks, Biplane(DCM,125207)


Simpson Simpson
Doppler Color Flow (SRT, R-409E2)

Card/CARDIAC_LV_SIMP/C/EDV(B
EDV Left Ventricular End Diastolic Volume(LN, 18026-5)
P)

Card/CARDIAC_LV_SIMP/C/ESV(B
ESV Left Ventricular End Systolic Volume(LN, 18148-7)
P)

Card/CARDIAC_LV_SIMP/C/EF(A4C
)

EF Card/CARDIAC_LV_SIMP/C/EF(A2C Left Ventricular Ejection Fraction(LN, 18043-0)


)

Card/CARDIAC_LV_SIMP/C/EF(BP)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 128 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Card/CARDIAC_LV_SIMP/C/SV(A4C
)

SV Card/CARDIAC_LV_SIMP/C/SV(A2C Stroke Volume(SRT, F-32120)


)

Card/CARDIAC_LV_SIMP/C/SV(BP)

Card/CARDIAC_LV_SIMP/C/CO(A4
C)

CO Card/CARDIAC_LV_SIMP/C/CO(A2 Cardiac Output(SRT, F-32100)


C)

Card/CARDIAC_LV_SIMP/C/CO(BP)

Card/CARDIAC_LV_SIMP/C/CI(A4C)

CI Card/CARDIAC_LV_SIMP/C/CI(A2C) Cardiac Index(SRT, F-32110)

Card/CARDIAC_LV_SIMP/C/CI(BP)

Card/CARDIAC_LV_SIMP/C/SI(A4C)

SI Card/CARDIAC_LV_SIMP/C/SI(A2C) Stroke Index(SRT, F-00078)

Card/CARDIAC_LV_SIMP/C/SI(BP)

A.4.3 Right Ventricle DCID (12204) Echocardiography Right Ventricle


Label Item ID Code Meaning (Coding Scheme Designator,
Code Value)

Finding Site: Right Ventricle(SRT, T-32500)

2D mode(SRT,G-03A2)

Card/CARDIAC_DIMENSION_RVA Right Ventricular Anterior Wall Diastolic


RVAWd
WD/B/RVAWd Thickness(LN, 18153-7)

Card/CARDIAC_DIMENSION_RVID Right Ventricular Internal Diastolic Dimension(LN,


RVIDd
D/B/RVIDd 20304-2)

Finding Site: Right Ventricle(SRT, T-32500)

M mode(SRT,G-0394)

Card/CARDIAC_M_RV_TEI/M/TV Tricuspid Valve Closure to Opening Time(LN,


Dur(a)
C-O Dur(a) G-0389)

Card/CARDIAC_M_RV_TEI/M/RVET Pulmonic Valve Ejection Time(LN, 18042-2)


RVET(b)
(b)

Card/CARDIAC_M_RV_TEI/M/RV Tei Index(SONO, SONO-5201-7)


TEI
TEI

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 129 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Card/CARDIAC_M_DIMENSION_RV Right Ventricular Anterior Wall Diastolic


RVAWd
AWD/M/RVAWd Thickness(LN, 18153-7)

Card/CARDIAC_M_DIMENSION_RV Right Ventricular Internal Diastolic Dimension(LN,


RVIDd
IDD/M/RVIDd 20304-2)

Card/CARDIAC_M_TSLOPE_RVPE Right Ventricle Pre-Ejection Period(LN, 20301-8)


RVPEP
P/M/RVPEP

Finding Site: Right Ventricle(SRT, T-32500)

Doppler Color Flow (SRT, R-409E2)

Card/CARDIAC_DIMENSION_RVA Right Ventricular Anterior Wall Diastolic


RVAWd
WD/C/RVAWd Thickness(LN, 18153-7)

Card/CARDIAC_DIMENSION_RVID Right Ventricular Internal Diastolic Dimension(LN,


RVIDd
D/C/RVIDd 20304-2)

Finding Site: Right Ventricle(SRT, T-32500)

Doppler Pulsed (SRT, R-409E4)

Card/CARDIAC_D_TV_RV_TEI/D/R
RVET(b) Pulmonic Valve Ejection Time(LN, 18042-2)
VET(b)

Card/CARDIAC_D_TV_RV_TEI/D/R
RV TEI Tei Index(Tei Index(SONO, SONO-5201-7))
V TEI

Card/CARDIAC_D_RVSP_TRVMAX/ Right Ventricular Peak Systolic Pressure(SRT,


RVSP
D/RVSP G-0380)

Card/CARDIAC_D_PV_RVET/D/RV
RVET Pulmonic Valve Ejection Time(LN, 18042-2)
ET

Card/CARDIAC_D_PV_RVPEP/D/R
RVPEP Pre-Ejection Period(LN, 18068-7)
VPEP

A.4.4 Left Atrium DCID (12205) Echocardiography Left Atrium


Label Item ID Code Meaning (Coding Scheme Designator,
Code Value)

Finding Site: Left Atrium (SRT, T-32300)

2D mode(SRT,G-03A2)

Card/CARDIAC_DIMENSION3/B/LA/
AO Left Atrium to Aortic Root Ratio(LN, 17985-3)
AO

Left Atrium Antero-posterior Systolic Dimension(LN,


LA Card/CARDIAC_DIMENSION3/B/LA
29469-4)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 130 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Card/CARDIAC_LA_SIMP/B/LA ESV
(A4C)

Card/CARDIAC_LA_SIMP/B/LA ESV
ESV Left Atrial ES Volume(DCM, 122408)
(BP)

Card/CARDIAC_LA_SIMP/B/LA ESV
(A2C)

Doppler Color Flow (SRT, R-409E2)

Card/CARDIAC_DIMENSION3/C/LA/
AO Left Atrium to Aortic Root Ratio(LN, 17985-3)
AO

Left Atrium Antero-posterior Systolic Dimension(LN,


LA Card/CARDIAC_DIMENSION3/C/LA
29469-4)

Card/CARDIAC_LA_SIMP/C/LA ESV
(A4C)

Card/CARDIAC_LA_SIMP/C/LA ESV
ESV Left Atrial ES Volume(DCM, 122408)
(BP)

Card/CARDIAC_LA_SIMP/C/LA ESV
(A2C)

M mode(SRT,G-0394)

Card/CARDIAC_M_DIMENSION3/M/L
AO Left Atrium to Aortic Root Ratio(LN, 17985-3)
A/AO

Card/CARDIAC_M_DIMENSION3/M/L Left Atrium Antero-posterior Systolic Dimension(LN,


LA
A 29469-4)

A.4.5 Right Atrium DCID (12206) Echocardiography Right Atrium


Label Item ID Code Meaning (Coding Scheme Designator,
Code Value)

Finding Site: Right Atrium (SRT, T-32200)

2D mode(SRT,G-03A2)

Card/CARDIAC_RA_SIMP/B/RA Right Atrium Systolic Volume(SONO,


ESV
ESV(A4C) SONO-5202-17)

Doppler Color Flow (SRT, R-409E2)

Card/CARDIAC_RA_SIMP/C/RA Right Atrium Systolic Volume(SONO,


ESV
ESV(A4C) SONO-5202-17)

Doppler Pulsed(SRT,R-409E4)

Card/CARDIAC_D_RVSP_TRVMAX/
RAP Right Atrium Systolic Pressure(LN, 18070-3)
D/RAP

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 131 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Card/CARDIAC_D_PV_PAEDP/D/RA
RAP Right Atrium Systolic Pressure(LN, 18070-3)
P

A.4.6 Aortic Valve DCID(12211) Echocardiography Aortic Valve


Label Item ID Code Meaning (Coding Scheme Designator, Code
Value)

Finding Site: Aortic Valve (SRT, T-35400)

2D mode(SRT,G-03A2)

Card/CARDIAC_DIMENSION_ASC/B Aortic Valve Cusp Separation(LN, 17996-0)


ACS
/ACS

Card/CARDIAC_DIMENSION_AVA/B Cardiovascular Orifice Area(LN, G-038E)


AVA
/AVA

M mode(SRT,G-0394)

Card/CARDIAC_M_DIMENSION_AS
ACS Aortic Valve Cusp Separation(LN, 17996-0)
C/M/ACS

Doppler Color Flow (SRT, R-409E2)

Card/CARDIAC_DIMENSION_ASC/C
ACS Aortic Valve Cusp Separation(LN, 17996-0)
/ACS

Card/CARDIAC_DIMENSION_AVA/C
AVA Cardiovascular Orifice Area(LN, G-038E)
/AVA

VTI Card/CARDIAC_PISA_MR/C/MR VTI Velocity Time Integral(LN, 20354-7)

Card/CARDIAC_PISA_AR_RAD/C/R
Rad Flow Radius(SONO, SONO-5202-21)
ad

PISA Card/CARDIAC_PISA_AR/C/PISA Flow Area(SONO, SONO-5202-28)

Doppler Pulsed(SRT,R-409E4)

Card/CARDIAC_D_AV_ARVMAX/D/A
AR Vmax Maximum Velocity(LN, 20351-3)
R Vmax

Card/CARDIAC_D_PISA_MS/D/AV
AV Vmax Peak Systolic Velocity(LN, 11726-7)
Vmax

Card/CARDIAC_D_PISA_MS/D/AV
AV VTI Peak Systolic Velocity(LN, 11726-7)
VTI

Card/CARDIAC_D_PISA_MS/D/AR
AR VTI Peak Systolic Velocity(LN, 11726-7)
VTI

HR Card/CARDIAC_D_AV_HR/D/HR Heart rate(LN, 8867-4)

AR DecT Card/CARDIAC_D_AV_ARDECT/D/A Deceleration Time(LN, 20217-6)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 132 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

R DecT

Card/CARDIAC_D_AV_ARPHT/D/AR
AR PHT Pressure Half-Time(LN, 20280-4)
PHT

Card/CARDIAC_D_AV_AVVTI/D/AV
AV Vmax Peak Systolic Velocity(LN, 11726-7)
Vmax

Card/CARDIAC_D_AV_AVVTI/D/AV
AV PGmax Peak Gradient(LN, 20247-3)
PGmax

Card/CARDIAC_D_AV_AVVTI/D/AV
AV Vmean Mean Velocity(LN, 20352-1)
Vmean

Card/CARDIAC_D_AV_AVVTI/D/AV
AV PGmean Mean Gradient(LN, 20256-4)
PGmean

Card/CARDIAC_D_AV_AVVTI/D/AV
AV VTI Velocity Time Integral(LN, 20354-7)
VTI

Card/CARDIAC_D_AVA_VTI/D/AV
AV Vmax Maximum Velocity(LN, 11726-7)
Vmax

Card/CARDIAC_D_AVA_VTI/D/AV
AV PGmax Peak Gradient(LN, 20247-3)
PGmax

Card/CARDIAC_D_AVA_VTI/D/AV
AV Vmean Mean Velocity(LN, 20352-1)
Vmean

Card/CARDIAC_D_AVA_VTI/D/AV
AV PGmean Mean Gradient(LN, 20256-4)
PGmean

Card/CARDIAC_D_AVA_VTI/D/AVA(
AVA(VTI) Cardiovascular Orifice Area(SRT, G-038E)
VTI)

Card/CARDIAC_D_AVA_VTI/D/AV
AV VTI Velocity Time Integral(LN, 20354-7)
VTI

Card/CARDIAC_D_AV_AVVMAX/D/A
AV Vmax Maximum Velocity(LN, 11726-7)
V Vmax

Card/CARDIAC_D_AV_AVVMAX/D/A
AV PGmax Peak Gradient(LN, 20247-3)
V PGmax

Card/CARDIAC_D_AV_AR/D/AR
AR Vmax Maximum Velocity(LN, 11726-7)
Vmax

Card/CARDIAC_D_AV_AR/D/AR
AR PGmax Peak Gradient(LN, 20247-3)
PGmax

Card/CARDIAC_D_AV_AR/D/AR
AR Vmean Mean Velocity(LN, 20352-1)
Vmean

Card/CARDIAC_D_AV_AR/D/AR
AR PGmean Mean Gradient(LN, 20256-4)
PGmean

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 133 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

AR VTI Card/CARDIAC_D_AV_AR/D/AR VTI Velocity Time Integral(LN, 20354-7)

AR Card/CARDIAC_D_AV_AR/D/AR
Flow Radius(SONO, SONO-5202-21)
Radius(2D) Radius(2D)

AR Flow Card/CARDIAC_D_AV_AR/D/AR
Peak Instantaneous Flow Rate(LN, 34141-2)
Rate Flow Rate

Card/CARDIAC_D_AV_AR/D/AR
AR ERO Cardiovascular Orifice Area(SRT, G-038E)
ERO

Card/CARDIAC_D_AV_AR/D/AR
AR Volume Volume Flow(LN, 33878-0)
Volume

Card/CARDIAC_D_AV_AR/D/AR
AR Fraction Regurgitant Fraction(SRT, G-0390)
Fraction

A.4.7 Mitral Valve DCID(12207) Echocardiography Mitral Valve


Label Item ID Code Meaning (Coding Scheme Designator, Code Value)

Mitral Valve Finding Site: Mitral Valve (SRT, T-35300)

2D mode 2D mode(SRT,G-03A2, SRT)

Card/CARDIAC_DIMENSION_EPSS/B
EPSS Mitral Valve EPSS, E wave (LN, 18036-4)
/EPSS

Card/CARDIAC_DIMENSION_MCS/B/
MCS Mitral Valve Cusp Separation(SONO, SONO-5202-16)
MCS

Card/CARDIAC_DIMENSION_MVDIA
MV Diam Cardiovascular Orifice Diameter(SRT, G-038F)
M/B/MV Diam

Card/CARDIAC_DIMENSION_MVA/B/
MVA Cardiovascular Orifice Area(SRT, G-038E)
MVA

M mode M mode(SRT,G-0394)

EPSS Card/MITRALVALVE/M/EPSS Mitral Valve EPSS, E wave (LN, 18036-4)

Card/CARDIAC_M_DIMENSION_MCS
MCS Mitral Valve Cusp Separation(SONO, SONO-5202-22)
/M/MCS

Card/CARDIAC_M_TSLOPE_MVDE/M Amplitude of the Mitral Valve DE wave(SONO,


MV DE
/MV DE SONO-5202-23)

MV E-F Card/CARDIAC_M_TSLOPE_MVEF/M
Mitral Valve E-F Slope by M-Mode(LN, 18040-6)
Slope /MV E-F Slope

MV E Card/CARDIAC_M_TSLOPE_MVEAM
Amplitude E Wave Mmode(SONO, SONO-5202-24)
Amp P/M/MV E Amp

MV A Card/CARDIAC_M_TSLOPE_MVAAM
Amplitude A Wave Mmode(SONO, SONO-5202-25)
Amp P/M/MV A Amp

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 134 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Doppler Color Flow (SRT, R-409E2)

Card/CARDIAC_DIMENSION_EPSS/C
EPSS Mitral Valve EPSS, E wave (LN, 18036-4)
/EPSS

Card/CARDIAC_DIMENSION_MCS/C/
MCS Mitral Valve Cusp Separation (SONO, SONO-5202-16)
MCS

Card/CARDIAC_DIMENSION_MVDIA
MV Diam Cardiovascular Orifice Diameter(SRT, G-038F)
M/C/MV Diam

Card/CARDIAC_DIMENSION_MVA/C/
MVA Cardiovascular Orifice Area(SRT, G-038E)
MVA

Card/CARDIAC_PISA_MR_RAD/C/Ra
Rad Dome Radius(SONO, SONO-5202-20)
d

PISA Card/CARDIAC_PISA_MR/C/PISA Cardiovascular Orifice Area(SRT, G-038E)

MR Flow Card/CARDIAC_PISA_MR/C/MR Flow


Peak Instantaneous Flow Rate(LN, 34141-2)
Rate Rate

AR VTI Card/CARDIAC_PISA_AR/C/AR VTI Velocity Time Integral(LN, 20354-7)

AR Flow Card/CARDIAC_PISA_AR/C/AR Flow


Peak Instantaneous Flow Rate(LN, 34141-2)
Rate Rate

Card/CARDIAC_PISA_MS_RAD/C/Ra
Rad Dome Radius(SONO, SONO-5202-20)
d

PISA Card/CARDIAC_PISA_MS/C/PISA Flow Area(SONO, SONO-5202-28)

MS VTI Card/CARDIAC_PISA_MS/C/MS VTI Velocity Time Integral(LN, 20354-7)

MS Flow Card/CARDIAC_PISA_MS/C/MS Flow


Peak Instantaneous Flow Rate(LN, 34141-2)
Rate Rate

Doppler Pulsed(SRT,R-409E4)

E Vel Card/CARDIAC_D_MV/D/MV E Vel Mitral Valve E-Wave Peak Velocity (LN, 18037-2)

E PG Card/CARDIAC_D_MV/D/MV E PG Peak Gradient(LN, 11726-7)

A Vel Card/CARDIAC_D_MV/D/MV A Vel Mitral Valve A-Wave Peak Velocity (LN, 17978-8)

A PG Card/CARDIAC_D_MV/D/MV A PG Peak Gradient(LN, 11726-7)

Card/CARDIAC_D_MV_APHT/D/MV
PHT Pressure Half-Time(LN, 20280-4)
PHT

MVA(PH Card/CARDIAC_D_MV_APHT/D/MVA(
Cardiovascular Orifice Area(SRT, G-038E)
T) PHT)

Card/CARDIAC_D_MV_EDUR/D/MV E
E Dur Mitral Valve E-Wave Duration (SONO, SONO-5202-4)
Dur

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 135 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Card/CARDIAC_D_MV_ADUR/D/MV A
A Dur Mitral Valve A-Wave Duration (SRT, G-0385)
Dur

Card/CARDIAC_D_MVA_VTI/D/MV
MV VTI Velocity Time Integral(LN , 20354-7)
VTI

Card/CARDIAC_D_MVA_VTI/D/MV
MV Vmax Maximum Velocity(LN, 11726-7)
Vmax

MV Card/CARDIAC_D_MVA_VTI/D/MV
Peak Gradient(LN, 11726-7)
PGmax PGmax

MV Card/CARDIAC_D_MVA_VTI/D/MV
Mean Velocity(LN, 11726-7)
Vmean Vmean

MV Card/CARDIAC_D_MVA_VTI/D/MV
Mean Gradient(LN, 11726-7)
PGmean PGmean

MVA(VTI Card/CARDIAC_D_MVA_VTI/D/MVA(
Cardiovascular Orifice Area(SRT, G-038E)
) VTI)

MR Flow Card/CARDIAC_D_MVA_VTI/D/MR
Peak Instantaneous Flow Rate(LN, 34141-2)
Rate Flow Rate

Card/CARDIAC_D_MVA_VTI/D/MR
MR VTI Peak Systolic Velocity(LN, 11726-7)
VTI

Card/CARDIAC_D_MVA_VTI/D/MR
MR ERO Cardiovascular Orifice Area(SRT, G-038E)
ERO

MR Card/CARDIAC_D_MVA_VTI/D/MR
Volume Flow(LN, 33878-0)
Volume Volume

MR Card/CARDIAC_D_MVA_VTI/D/MR
Regurgitant Fraction(SRT, G-0390)
Fraction Fraction

Card/CARDIAC_D_MV_DECT/D/MV
MV DecT Deceleration Time(LN, 20217-6)
DecT

MR
Card/CARDIAC_D_MV/D/MR Vmax Peak Systolic Velocity(LN, 11726-7)
Vmax

MR VTI Card/CARDIAC_D_MV/D/MR VTI Peak Systolic Velocity(LN, 11726-7)

MR dt Card/CARDIAC_D_MV_DPDT/D/MR dt Deceleration Time(LN, 20217-6)

Card/CARDIAC_D_MV_DPDT/D/MR Mitral Regurgitation dP/dt derived from Mitral Reg


MR dP/dt
dP/dt velocity(LN, 18035-6)

Card/CARDIAC_D_MVA(VTI)/D/MV
MV VTI Maximum Velocity(LN, 20351-3)
VTI

MVA(PH Card/CARDIAC_D_MV_APHT/D/MVA(
Cardiovascular Orifice Area(SRT, G-038E)
T) PHT)

MVA(PH Card/CARDIAC_D_MV_APHT/D/MVA( Cardiovascular Orifice Area(SRT, G-038E)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 136 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

T) PHT)

MV C-O Card/CARDIAC_D_LV_TEI/D/MV C-O


Mitral Valve Closure to Opening Time(SRT, G-0387)
Dur(a) Dur(a)

MR Card/CARDIAC_D_MV_VMAX/D/MR
Maximum Velocity(LN, 11726-7)
Vmax Vmax

MR Card/CARDIAC_D_MV_VMAX/D/MR
Peak Gradient(LN, 11726-7)
PGmax PGmax

MR Card/CARDIAC_D_PISA_MR/D/MR
Maximum Velocity(LN, 11726-7)
Vmax Vmax

MR Card/CARDIAC_D_PISA_MR/D/MR
Peak Gradient(LN, 11726-7)
PGmax PGmax

MR Card/CARDIAC_D_PISA_MR/D/MR
Mean Velocity(LN, 11726-7)
Vmean Vmean

MR Card/CARDIAC_D_PISA_MR/D/MR
Mean Gradient(LN, 11726-7)
PGmean PGmean

Card/CARDIAC_D_PISA_MR/D/MR
MR VTI Velocity Time Integral(LN, 11726-7)
VTI

Card/CARDIAC_D_PISA_MR/D/MR
MR Rad Flow Radius(SONO, SONO-5202-21)
Rad

MR Flow Card/CARDIAC_D_PISA_MR/D/MR
Peak Instantaneous Flow Rate(LN, 34141-2)
Rate Flow Rate

Card/CARDIAC_D_PISA_MR/D/MR
MR ERO Cardiovascular Orifice Area(SRT, G-038E)
ERO

MR Card/CARDIAC_D_PISA_MR/D/MR
Volume Flow(LN, 33878-0)
Volume Volume

MR Card/CARDIAC_D_PISA_MR/D/MR
Regurgitant Fraction(SRT, G-0390)
Fraction Fraction

Card/CARDIAC_D_PISA_MS/D/MS
MS Rad Flow Radius(SONO, SONO-5202-21)
Rad

Card/CARDIAC_D_PISA_MS/D/MS
MS Vmax Maximum Velocity(LN, 20247-3)
Vmax

MS Flow Card/CARDIAC_D_PISA_MS/D/MS
Peak Instantaneous Flow Rate(LN, 34141-2)
Rate Flow Rate

Card/CARDIAC_D_PISA_MS/D/MS
MS ERO Cardiovascular Orifice Area(SRT, G-038E)
ERO

MS Card/CARDIAC_D_PISA_MS/D/MS
Volume Flow(LN, 33878-0)
Volume Volume

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 137 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Card/CARDIAC_D_PISA_MS/D/MS
MS VTI Velocity Time Integral(LN, 20351-3)
VTI

A.4.8 Pulmonic Valve DCID(12209) Echocardiography Pulmonic Valve


Label Item ID Code Meaning (Coding Scheme Designator, Code
Value)

Pulmonic Valve Finding Site: Pulmonic Valve (SRT, T-35200)

Doppler Color Flow (SRT, R-409E2)

Rad Card/CARDIAC_PISA_PR_RAD/C/Rad Flow Radius(SONO, SONO-5202-21)

PR VTI Card/CARDIAC_PISA_PR/C/PR VTI Velocity Time Integral(LN, 20354-7)

Doppler Pulsed(SRT,R-409E4)

Card/CARDIAC_D_PV_PRVMAX/D/PR
PR Vmax Maximum Velocity(LN, 20351-3)
Vmax

PV HR Card/CARDIAC_D_PV_HR/D/PV HR Heart rate(LN, 8867-4)

Card/CARDIAC_D_PV_PVACCT/D/PV
PV AccT Acceleration Time(LN, 20168-1)
AccT

Card/CARDIAC_D_PV_PAEDP/D/PR
PR Ved End Diastolic Velocity(LN, 11653-3)
Ved

Card/CARDIAC_D_PV_PAEDP/D/PR
PR PGed Peak Gradient(LN, 20247-3)
PGed

PVA(Vma Card/CARDIAC_D_PV_PVAVMAX/D/P
Cardiovascular Orifice Area(SRT, G-038E)
x) VA(Vmax)

Card/CARDIAC_D_PV_PVAVMAX/D/P
PV Vmax Maximum Velocity(LN, 11726-7)
V Vmax

Card/CARDIAC_D_PV_PVAVTI/D/PV
PV Vmax Maximum Velocity(LN, 11726-7)
Vmax

PV Card/CARDIAC_D_PV_PVAVTI/D/PV
Peak Gradient(LN, 20247-3)
PGmax PGmax

Card/CARDIAC_D_PV_PVAVTI/D/PV
PV Vmean Mean Velocity(LN, 20352-1)
Vmean

PV Card/CARDIAC_D_PV_PVAVTI/D/PV
Mean Gradient(LN, 20256-4)
PGmean PGmean

Card/CARDIAC_D_PV_PVAVTI/D/PV
PV VTI Velocity Time Integral(LN, 20354-7)
VTI

Card/CARDIAC_D_PV_PVAVTI/D/PVA
PVA(VTI) Cardiovascular Orifice Area(SRT, G-038E)
(VTI)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 138 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

PR Vmax Card/CARDIAC_D_PR/D/PR Vmax Maximum Velocity(LN, 20351-3)

PR
Card/CARDIAC_D_PR/D/PR PGmax Peak Gradient(LN, 20247-3)
PGmax

PR Vmean Card/CARDIAC_D_PR/D/PR Vmean Mean Velocity(LN, 20352-1)

PR
Card/CARDIAC_D_PR/D/PR PGmean Mean Gradient(LN, 20256-4)
PGmean

PR VTI Card/CARDIAC_D_PR/D/PR VTI Velocity Time Integral(LN, 20354-7)

PR Rad Card/CARDIAC_D_PR/D/PR Rad Flow Radius(SONO, SONO-5202-21)

PR Flow
Card/CARDIAC_D_PR/D/PR Flow Rate Peak Instantaneous Flow Rate(LN, 34141-2)
Rate

PR ERO Card/CARDIAC_D_PR/D/PR ERO Cardiovascular Orifice Area(SRT, G-038E)

PR
Card/CARDIAC_D_PR/D/PR Volume Volume Flow(LN, 33878-0)
Volume

PR
Card/CARDIAC_D_PR/D/PR Fraction Regurgitant Fraction(SRT, G-0390)
Fraction

A.4.9 Tricuspid Valve DCID(12208) Echocardiography Tricuspid Valve


Label Item ID Code Meaning (Coding Scheme Designator, Code Value)

Tricuspid Valve Finding Site: Tricuspid Valve (SRT, T-35100)

2D mode(SRT,G-03A2, SRT)

Card/CARDIAC_DIMENSION_TVDIA
TV Diam Cardiovascular Orifice Diameter(SRT, G-038F)
M/B/TV Diam

Doppler Color Flow (SRT, R-409E2)

Card/CARDIAC_DIMENSION_TVDIA
TV Diam Cardiovascular Orifice Diameter(SRT, G-038F)
M/C/TV Diam

Rad Card/CARDIAC_PISA_TR_RAD/C/Rad Flow Radius(SONO, SONO-5202-21)

TR Flow Card/CARDIAC_PISA_TR/C/TR Flow


Peak Instantaneous Flow Rate(LN, 34141-2)
Rate Rate

TR VTI Card/CARDIAC_PISA_TR/C/TR VTI Velocity Time Integral(LN, 20354-7)

Doppler Pulsed(SRT,R-409E4)

A Card/CARDIAC_D_MV/D/E/A Tricuspid Valve E to A Ratio(LN, 18039-8)

TV Vmax Card/CARDIAC_D_TV_VMAX/D/TV
Maximum Velocity(LN, 11726-7)
Vmax

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 139 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

TV Card/CARDIAC_D_TV_VMAX/D/TV
Peak Gradient(LN, 20247-3)
PGmax PGmax

TV Vmax Card/CARDIAC_D_TV_VTI/D/TV
Maximum Velocity(LN, 11726-7)
Vmax

TV Card/CARDIAC_D_TV_VTI/D/TV
Peak Gradient(LN, 20247-3)
PGmax PGmax

TV Card/CARDIAC_D_TV_VTI/D/TV
Mean Velocity(LN, 20352-1)
Vmean Vmean

TV Card/CARDIAC_D_TV_VTI/D/TV
Mean Gradient(LN, 20256-4)
PGmean PGmean

TV VTI Card/CARDIAC_D_TV_VTI/D/TV VTI Velocity Time Integral(LN, 20354-7)

TV C-O Card/CARDIAC_D_TV_RV_TEI/D/TV
Tricuspid Valve Closure to Opening Time(SRT, G-0389)
Dur(a) C-O Dur(a)

TV E Vel Card/CARDIAC_D_EA/D/TV E Vel Tricuspid Valve E Wave Peak Velocity(LN, 18031-5)

TV E PG Card/CARDIAC_D_EA/D/TV E PG Peak Gradient(LN, 20247-3)

TV A Vel Card/CARDIAC_D_EA/D/TV A Vel Tricuspid Valve A Wave Peak Velocity(LN, 18030-7)

TV A PG Card/CARDIAC_D_EA/D/TV A PG Peak Gradient(LN, 20247-3)

TV E/A Card/CARDIAC_D_EA/D/TV E/A Tricuspid Valve E to A Ratio(LN, 18039-8)

Card/CARDIAC_D_RVSP_TRVMAX/D
TR Vmax Maximum Velocity(LN, 11726-7)
/TR Vmax

TR Card/CARDIAC_D_RVSP_TRVMAX/D
Peak Gradient(LN, 20247-3)
PGmax /TR PGmax

TR Vmax Card/CARDIAC_D_TR/D/TR Vmax Maximum Velocity(LN, 11726-7)

TR
Card/CARDIAC_D_TR/D/TR PGmax Peak Gradient(LN, 20247-3)
PGmax

TR
Card/CARDIAC_D_TR/D/TR Vmean Mean Velocity(LN, 20352-1)
Vmean

TR
Card/CARDIAC_D_TR/D/TR PGmean Mean Gradient(LN, 20256-4)
PGmean

TR VTI Card/CARDIAC_D_TR/D/TR VTI Velocity Time Integral(LN, 20354-7)

TR
Card/CARDIAC_D_TR/D/TR Rad(2D) Flow Radius(SONO, SONO-5202-21)
Rad(2D)

TR Flow
Card/CARDIAC_D_TR/D/TR Flow Rate Peak Instantaneous Flow Rate(LN, 34141-2)
Rate

TR ERO Card/CARDIAC_D_TR/D/TR ERO Cardiovascular Orifice Area(SRT, G-038E)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 140 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

TR
Card/CARDIAC_D_TR/D/TR Volume Volume Flow(LN, 33878-0)
Volume

TR
Card/CARDIAC_D_TR/D/TR Fraction Regurgitant Fraction(SRT, G-0390)
Fraction

A.4.10 Aorta DCID (12212) Echocardiography Aorta


Label Item ID Code Meaning (Coding Scheme Designator, Code
Value)

Finding Site: Aorta (SRT, T-42000)

B mode 2D mode(SRT,G-03A2, SRT)

AO Card/CARDIAC_DIMENSION3/B/AO Aortic Root Diameter(LN, 18015-8)

C mode Doppler Color Flow (SRT, R-409E2)

AO Card/CARDIAC_DIMENSION3/C/AO Aortic Root Diameter(LN, 18015-8)

M mode M mode(SRT,G-0394)

AO Card/CARDIAC_M_DIMENSION3/M/AO Aortic Root Diameter(LN, 18015-8)

A.4.11 Pulmonary artery DCID (12210) Echocardiography Pulmonary Artery


Label Item ID Code Meaning (Coding Scheme Designator, Code Value)

Finding Site: Pulmonary artery (SRT, T-44000)

B mode 2D mode(SRT,G-03A2, SRT)

Card/CARDIAC_DIMENSION_M
MPA Main Pulmonary Artery Diameter(LN, 18020-8)
PA/B/MPA

Card/CARDIAC_DIMENSION_P
PV Diam Cardiovascular Orifice Diameter(LN, G-038F)
VDIAM/B/PV Diam

C mode Doppler Color Flow (SRT, R-409E2)

Card/CARDIAC_DIMENSION_M
MPA Main Pulmonary Artery Diameter(LN, 18020-8)
PA/C/MPA

Card/CARDIAC_DIMENSION_P
PV Diam Cardiovascular Orifice Diameter(LN, G-038F)
VDIAM/C/PV Diam

D mode Doppler Pulsed(SRT,R-409E4)

MPA Card/CARDIAC_D_PV_MPAVM
Main Pulmonary Artery Peak Velocity(SRT, G-038A)
Vmax AX/D/MPA Vmax

LPA Card/CARDIAC_D_PV_LPAVMA
Left Pulmonary Artery Peak Velocity(SONO, SONO-5202-28)
Vmax X/D/LPA Vmax

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 141 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Card/CARDIAC_D_PV_PAEDP/
PAEDP Pulmonary Artery Pressure(SRT, F0212C)
D/PAEDP

A.4.12 Pulmonary Venous Structure DCID (12214) Echocardiography Pulmonary Veins


Label Item ID Code Meaning (Coding Scheme Designator, Code Value)

Finding Site: Pulmonary Venous Structure (SRT, T-48581)

Card/CARDIAC_D_PLUM_SVEL/D/
Pulm S Vel Pulmonary Vein Systolic Peak Velocity(LN, 29450-4)
Pulm S Vel

Card/CARDIAC_D_PLUM_SVTI/D/P
Pulm S VTI Pulmonary Vein S-Wave Velocity Time Integral(SRT, G-038C)
ulm S VTI

Card/CARDIAC_D_PLUM_AVEL/D/ Pulmonary Vein Atrial Contraction Reversal Peak Velocity(LN,


Pulm A Vel
Pulm A Vel 29453-8)

Card/CARDIAC_D_PLUM_DVEL/D/
Pulm D Vel Pulmonary Vein Diastolic Peak Velocity(LN, 29451-2)
Pulm D Vel

Card/CARDIAC_D_PLUM_DVTI/D/P
Pulm D VTI Pulmonary Vein D-Wave Velocity Time Integral(SRT, G-038D)
ulm D VTI

Card/CARDIAC_D_PLUM_ADUR/D/
Pulm A Dur Pulmonary Vein A-Wave Duration(SRT, G-038B)
Pulm A Dur

Card/CARDIAC_D_PLUM_DECT/D/
Pulm DecT Deceleration Time(LN, 20217-6)
Pulm DecT

A.4.13 Lateral Mitral Annulus (12235) Echocardiography Cardiac Shunt


Label Item ID Code Meaning (Coding Scheme Designator, Code Value)

Finding Site: Lateral Mitral Annulus (SRT, G-0392)

Card/CARDIAC_D_MV/D/E/E'(Late
ral)

Card/CARDIAC_D_TDI_LATERAL/ Left Ventricular Peak Early Diastolic Tissue Velocity(SRT,


Lateral
D/Ea Lateral G-037A)

Card/CARDIAC_D_TDI_LATERAL/
D/E/E'(Lateral)

Card/CARDIAC_D_TDI_LATERAL/ LV Peak Diastolic Tissue Velocity During Atrial Systole(SRT,


Aa Lateral
D/Aa Lateral G-037C)

Card/CARDIAC_D_TDI_LATERAL/
Aa(Lateral) Mitral Valve DTI Ea to Aa Ratio(SONO, SONO-5201-32)
D/Ea/Aa(Lateral)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 142 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

A.4.14 Left Ventricle Outflow Tract DCID (12223) Echocardiography Stroke Volume Origin
Label Item ID Code Meaning (Coding Scheme Designator, Code Value)

Finding Site: Left Ventricle Outflow Tract (SRT, T-32650)

Card/CARDIAC_D_MVA_VTI/D/LV
LVOT Vmax Maximum Velocity(LN, 11726-7)
OT Vmax

Card/CARDIAC_D_MVA_VTI/D/LV
LVOT PGmax Peak Gradient(LN, 11726-7)
OT PGmax

Card/CARDIAC_D_MVA_VTI/D/LV
LVOT Vmean Mean Velocity(LN, 11726-7)
OT Vmean

LVOT Card/CARDIAC_D_MVA_VTI/D/LV
Mean Gradient(LN, 11726-7)
PGmean OT PGmean

Card/CARDIAC_D_MVA_VTI/D/LV
LVOT VTI Velocity Time Integral(LN, 11726-7)
OT VTI

Card/CARDIAC_D_MVA(VTI)/D/LV
LVOT VTI Peak Systolic Velocity(LN, 11726-7)
OT VTI

Card/CARDIAC_D_AVA_VTI/D/LV
LVOT Vmax Maximum Velocity(LN, 11726-7)
OT Vmax

Card/CARDIAC_D_AVA_VTI/D/LV
LVOT PGmax Peak Gradient(LN, 20247-3)
OT PGmax

Card/CARDIAC_D_AVA_VTI/D/LV
LVOT Vmean Mean Velocity(LN, 20352-1)
OT Vmean

LVOT Card/CARDIAC_D_AVA_VTI/D/LV
Mean Gradient(LN, 20256-4)
PGmean OT PGmean

Card/CARDIAC_D_AVA_VTI/D/LV
LVOT VTI Velocity Time Integral(LN, 20354-7)
OT VTI

Card/CARDIAC_D_AVA_VTI/D/LV
LVOT(2D) Cardiovascular Orifice Diameter(SRT, G-038F)
OT(2D)

Card/CARDIAC_D_AVA_VTI/D/SV(
SV(LVOT) Stroke Volume(SRT, F-32120)
LVOT)

Card/CARDIAC_D_AVA_VTI/D/CO
CO(LVOT) Cardiac Output(SRT, F-32100)
(LVOT)

Card/CARDIAC_D_AVA_VMAX/D/
LVOT(2D) Cardiovascular Orifice Diameter(SRT, G-038F)
LVOT(2D)

Card/CARDIAC_D_AVA_VMAX/D/
AVA(Vmax) Cardiovascular Orifice Area Area(SRT, G-038E)
AVA(Vmax)

Card/CARDIAC_D_AVA_VMAX/D/
LVOT Vmax Peak Systolic Velocity(LN, 11726-7)
LVOT Vmax

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 143 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

A.4.15 Right Ventricle Outflow Tract DCID (12223) Echocardiography Stroke Volume Origin
Label Item ID Code Meaning (Coding Scheme Designator, Code Value)

Finding Site: Right Ventricle Outflow Tract (SRT, T-32550)

2D mode(SRT,G-03A2, SRT)

Card/CARDIAC_DIMENSION_RVO
RVOT Cardiovascular Orifice Diameter(SRT, G-038F)
T/B/RVOT

Doppler Color Flow (SRT, R-409E2)

Card/CARDIAC_DIMENSION_RVO
RVOT Cardiac Output(SRT, F-32100)
T/C/ RVOT

M mode(SRT,G-0394)

Card/CARDIAC_M_DIMENSION_R
RVOT Cardiovascular Orifice Diameter(SRT, G-038F)
VOT/M/RVOT

Doppler Pulsed(SRT,R-409E4)

Card/CARDIAC_D_PV_PVAVMAX/
RVOT(2D) Cardiovascular Orifice Diameter(SRT, G-038F)
D/RVOT(2D)

Card/CARDIAC_D_PV_PVAVMAX/
RVOT VMax Maximum Velocity(LN, 11726-7)
D/RVOT VMax

Card/CARDIAC_D_PV_PVAVTI/D/
RVOT(2D) Cardiovascular Orifice Diameter(SRT, G-038F)
RVOT(2D)

Card/CARDIAC_D_PV_PVAVTI/D/
RVOT Vmax Maximum Velocity(LN, 11726-7)
RVOT Vmax

Card/CARDIAC_D_PV_PVAVTI/D/
RVOT PGmax Peak Gradient(LN, 20247-3)
RVOT PGmax

Card/CARDIAC_D_PV_PVAVTI/D/
RVOT Vmean Mean Velocity(LN, 20352-1)
RVOT Vmean

RVOT Card/CARDIAC_D_PV_PVAVTI/D/
Mean Gradient(LN, 20256-4)
PGmean RVOT PGmean

Card/CARDIAC_D_PV_PVAVTI/D/
RVOT VTI Velocity Time Integral(LN, 20354-7)
RVOT VTI

A.4.16 Medial Mitral Annulus DCID (12235) Mitral Valve Anatomic Sites
Label Item ID Code Meaning (Coding Scheme Designator, Code Value)

Finding Site: Medial Mitral Annulus Contraction (SRT, G-0392)

Card/CARDIAC_D_MV/D/E/E'( Ratio of MV Peak Velocity to LV Peak Tissue Velocity E-Wave(SRT,


E'(Medial)
Medial) G-037B)

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 144 of 145


SONOSCAPE MEDICAL CORP DICOM 3.0 CONFORMANCE STATEMENT

Card/CARDIAC_D_TDI_MEDIA
Sa Medial Left Ventricular Peak Systolic Tissue Velocity(SRT, G-037D)
L/D/Sa Medial

Card/CARDIAC_D_TDI_MEDIA
Ea Medial Left Ventricular Peak Early Diastolic Tissue Velocity(SRT, G-037A)
L/D/Ea Medial

Card/CARDIAC_D_TDI_MEDIA Ratio of MV Peak Velocity to LV Peak Tissue Velocity E-Wave(SRT,


E'(medial)
L/D/E/E'(medial) G-037B)

Card/CARDIAC_D_TDI_MEDIA LV Peak Diastolic Tissue Velocity During Atrial Systole(SRT,


Aa Medial
L/D/Aa Medial G-037C)

Card/CARDIAC_D_TDI_MEDIA
Aa(Medial) Mitral Valve DTI Ea to Aa Ratio(SONO, SONO-5201-31)
L/D/Ea/Aa(Medial)

Card/CARDIAC_D_TDI_LATER
Sa Lateral Left Ventricular Peak Systolic Tissue Velocity(SRT, G-037D)
AL/D/Sa Lateral

A.4.17 Hepatic Veins DCID (12216) Echocardiography Hepatic Veins


Label Item ID Code Meaning (Coding Scheme Designator, Code Value)

Card/CARDIAC_D_HEP_SVEL/
S Vel Hepatic Vein Systolic Peak Velocity(LN, 29471-0)
D/Hep S Vel

Card/CARDIAC_D_HEP_DVEL
D Vel Hepatic Vein Diastolic Peak Velocity(LN, 29472-8)
/D/Hep D Vel

Card/CARDIAC_D_HEP_AVEL/
A Vel Hepatic Vein Atrial Contraction Reversal Peak Velocity(LN, 29474-4)
D/Hep A Vel

Card/CARDIAC_D_HEP_ADUR
A Dur Hepatic Vein A-Wave Duration(SONO, SONO-5202-27)
/D/Hep A Dur

Version 3.0.8.X / 3.1.X.X / 3.0.X.X Page 145 of 145

You might also like