MUCLecture 2021 11245963
MUCLecture 2021 11245963
Lecture Four
Sarah Mohammed
Second stage
ــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ
Theoretical Lecture
KVp adjustment & mA control
KVp Adjustment
Kilovoltage peak (kVp) the kilovoltage applied to the x-ray tube, which
accelerates electrons from the cathode to the anode. Tube voltage,
determines the quantity (intensity) and quality (energy) of the photons
generated.
The energy of the photon emitted from the X-ray tube depends on the
energy of the electrons that bombard the target. The energy of the
electron is, in turn, determined by the peak kilovoltage used.
1
Almustaqbal University College
Lecture Four
Sarah Mohammed
Second stage
ــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ
Image quality
2
Almustaqbal University College
Lecture Four
Sarah Mohammed
Second stage
ــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ
A higher kVp will make X-ray beam more penetrating, this will
result in less difference in attenuation between the different parts of
the image, leading to lower contrast.
The tube current affects only the intensity (quantity) but not the quality of
the X-rays (figure 3). As the tube current increases the intensity also
increases.
3
Almustaqbal University College
Lecture Four
Sarah Mohammed
Second stage
ــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ
Milliampere-seconds (mAs)
For example:
mA x t = mAs
600 mA x 0.1 sec. = 60 mAs
4
Almustaqbal University College
Lecture Four
Sarah Mohammed
Second stage
ــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ
The exposure timers control the length of an X-ray exposure. There are
two basic types of exposure timers, namely, electronic timers, and
automatic exposure control (AEC).
1- Electronic Timers
In electronic timer, the length of the X-ray exposure is determined by the
time required to charge a capacitor through a resistance. When the
exposure button starts exposure, it also starts charging a capacitor. The
exposure time is terminated, when the capacitor is charged to a specified
value, necessary to turn on associated electronic circuit. This time can be
varied by varying the value of the resistance in the charging circuit.
Modern generators use electronic timers, in order to obtain very accurate
exposure duration. These timers can accurately control exposures from <
1 ms to > 1 s.
electronic device that converts visible light energy into electrical energy.
A photodiode is a solid-state device that performs the same function.
Phototimer AEC devices are considered exit-type devices because the
detectors are positioned behind the image receptor (Figure 2) so that
radiation must exit the image receptor before it is measured by the
detectors.
6
Almustaqbal University College
Lecture Four
Sarah Mohammed
Second stage
ــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ
paddles have been exposed. Photo timers have largely been replaced with
ionization chamber systems.
7
Almustaqbal University College
Lecture Four
Sarah Mohammed
Second stage
ــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ
Checking a timer:
If the exposure time set on the diagnostic x-ray unit is not optimal, the
radiograph can be under exposed or over exposed. This may lead to
repeat examinations. Hence, there is a need to test the timer of the x-ray
unit periodically.
Spinning top test tool (figure 5) is used to check the exposure time.
It consist of a rotating circular brass plate with small rectangular
portion cut (hole) at its periphery.
The tool is placed on the cassette, loaded with film. For a set time,
the unit is energized, while the top is rotating.
The pulses passing through the hole of the circular plate, produces
equally spaced rectangular density patterns, on the film.
8
Almustaqbal University College
Lecture Four
Sarah Mohammed
Second stage
ــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ
Questions:
1. Define KVp?
2. What are the consideration that should be taken to adjust the KVp?
3. How the KVp effect the contrast image?
4. What does the mAs mean?
5. What are the KVp and mA effect on?
6. How the electronic timers work?
7. What are the types of automatic exposure control (AEC)? Explain
on of them?
8. What is the entrance type device of AEC?
9. How can you checking the exposure time?