Elisa Theoretical
Elisa Theoretical
In a typical assay designed to detect an antigen in a complex protein mixture, the antigen is immobilized either by direct
adsorption or via an antibody adsorbed to the wells of a microplate. The plate is blocked and the antigen is probed with a
specific detection antibody. The detection antibody may be directly labeled with a signal-generating enzyme or fluorophore
or it may be secondarily probed with an enzyme- or fluor-labeled secondary antibody (or avidin-biotin chemistry, see below).
For enzymatic detection, the appropriate enzyme substrate is added. The signal observed is proportional to the amount of
antigen in the sample. Washing between steps ensures that only specific (high-affinity) binding events are maintained to
cause signal at the final step.
2
Variations between ELISA protocols
A. Antigen Immobilization
Antigen immobilization varies between two principle techniques. In a traditional (direct coating) ELISA, antigens are directly
attached to the plate by passive adsorption, usually using a carbonate/bicarbonate buffer at pH >9. Most but not all proteins
bind tightly to the polystyrene surface of microplates in alkaline conditions. However, if antigen is present at low levels or
does not adhere well to the plastic, then the alternative sandwich or capture ELISA may be used. In capture (indirect coating)
ELISA, antigen-specific antibody is adsorbed onto the plastic, which in turn binds and immobilizes the antigen upon
incubation with the antigen sample. Attachment of the antibody is typically achieved using the same carbonate/bicarbonate
buffer at pH >9, or in rare instances pre-activated plates are used for a more directed attachment approach.
Sandwich ELISAs have become very popular when using complex protein samples because only the specific antigen
becomes immobilized rather than the entire sample of proteins. The more antigen that is immobilized, the higher is the
potential sensitivity of the assay. Sandwich ELISAs require two different antibodies that bind specifically to the antigen (each
reacting with a different epitope). The first antibody (bound to the plate) is called the capture antibody or coating antibody,
whereas the second antibody detects the immobilized antigen and is called the detection antibody. Such antibodies are known
as “matched pairs”; they must be validated to work in combination, as they must not compete for binding to the antigen for
accurate results to be possible. Combinations of monoclonal and polyclonal antibodies can be used; it is more common to use
the monoclonal as the coating antibody and the polyclonal as the detection antibody. Sandwich ELISAs sometimes require
more optimization than traditional ELISAs but usually the signal-to-noise (S:N) ratios are higher.
Signal amplification occurs through two mechanisms with these approaches. First, biotinylation (biotin-labeling) typically
results in multiple biotin tags per antibody molecule, thus allowing more than one streptavidin molecule to bind to each
antibody. Binding is aided by the fact that avidin-type proteins are tetrameric and have four biotin-binding sites per molecule.
Second, the process of either labeling the streptavidin molecule with enzymes or using a pre-incubated mixture of
streptavidin plus biotinylated enzyme results in conjugates having more than one enzyme. The combined effect of this
3
multiple labeling is to increase the number of enzyme molecules in the final immune complex. This increases the catalysis of
appropriate substrate and gives a stronger signal compared to a conventional enzyme-labeled secondary antibody.
D. Fluorescence
The relatively recent expansion in the number of stable fluorophores in the visible and infrared ranges has made fluorescent
signal detection an attractive option for ELISA applications. This kind of approach is common when performing multiplex
arrays, as more than one antigen can be detected simultaneously with antibodies conjugated to different fluorophores. In
fluorescence assays, the detection antibody is either labeled directly or the secondary antibody (or occasionally avidin) is
labeled for indirect detection. When multiplexing using labeled secondary antibodies it is essential to use detection antibodies
from different species in order to distinguish the separate signals. The detection limit is typically around 100pg/well which is
less sensitive than colorimetric or chemiluminescent detection. imp
Diagram of a multiplex array ELISA made made possible by using fluorescence. In this case, twelve different capture antibodies are
coated as an array of printed spots on a glass slide. Each antibody captures a different analyte and is detected by its matched detection
antibody, which is biotinylated. Finally, all spots are detected through a fluor-labeled streptavidin conjugate (in this case, the Thermo Scientific
DyLight™ 649 Fluorophore).
E. Enzymatic Detection
Two enzymes are commonly used in ELISA applications. Alkaline Phosphatase (AP) is a large enzyme used in a minority of
assays. Its size (140 KDa) makes it difficult to conjugate more than one or two molecules of the enzyme to each molecule of
an antibody or avidin, and this limits the amount of signal that can be generated. AP is also prone to stability issues unless
stored and handled correctly.
Horseradish Peroxidase (HRP) is a more commonly used enzyme. Its small size (40KDa) allows more molecules to be
coupled to antibodies or avidin, and this can boost signal generation. HRP is the enzyme of choice for most researchers
performing ELISAs and can be used with a variety of substrates (see below), most of which are more sensitive than AP
equivalents.
F. Substrates
Enzymatic signal generation requires the catalysis of a substrate to produce a colored or fluorescent compound or
chemiluminescence (visible light). The signal is measured using a spectrophotometric plate reader, a fluorometer with the
appropriate filters or a luminometer set to read total light output. Each type of substrate is discussed below; more information
about specific products can be found at our website.
Colorimetric substrates form a soluble, colored product that accumulates over time relative to the amount of enzyme present
imp in each well. When the desired color intensity is reached, the product absorbance is either measured directly or in some cases
a stop solution is added to provide a fixed end point for the assay. Colorimetric substrates are available for both horseradish
peroxidase (TMB, OPD, ABTS) and alkaline phosphatase (PNPP).
4
Chemifluorescent detection is also enzyme-based but the generated product is fluorescent rather than colorimetric. The signal
is measured using a fluorometer with the appropriate excitation and emission filters. Chemifluorescence reactions are either
measured over time in kinetic assays or halted using a stop solution for direct measurement. Examples of chemifluorescent
substrates for HRP are Thermo Scientific QuantaRed™ and QuantaBlu™ Substrates.
Chemiluminescence is a chemical reaction that generates energy released in the form of light. Most chemiluminescent
substrates are HRP-dependent although some AP equivalents are available. The most common approach is to use luminol in
the presence of HRP and a peroxide buffer. The luminol is oxidized and forms an excited state product that emits light as it
decays to the ground state. Light emission occurs only during the enzyme-substrate reaction, therefore when the substrate
becomes exhausted the signal ceases. Chemiluminescent detection is generally considered to be more sensitive than
colorimeteric detection. Chemiluminescent substrates for HRP include Thermo Scientific SuperSignal™ ELISA Pico and
ELISA Femto Substrates.
5
positively charged plate. Other buffered solutions such as Tris-buffered saline (TBS) or phosphate-buffered saline (PBS) at
physiological pH are sometimes used but coupling is generally not quite as efficient.
In some instances researchers require a more directed approach for attachment of the coating antibody or protein sample to
the plate, and several pre-activated plates exist for this purpose. For specific, orientated binding of the coating antibody,
plates that are pre-coated with Protein A or Protein G are available (not advised for sandwich ELISAs because of potential
cross-reactivity with detection and/or secondary antibodies). For biotinylated samples or coating antibodies, plates that are
pre-coated with streptavidin are ideal. Some assays require direct immobilization of a histidine-tagged protein, in which case
nickel- or copper-coated plates are suitable (these plates can also be used to bind and orientate capture antibodies as IgG
molecules have a histidine rich sequence in their Fc domain). For group-specific attachment of molecules via amines or thiols
to form a covalent bond, maleic-anhydride or maleimide-activated plates can be used. (These are especially useful for direct
attachment of peptide antigens, which do not coat well by passive adsorption because of their small size.) Please visit our
website for a complete listing of our many Thermo Scientific Pierce™ Coated Microplates.
B. Antibodies
Not all antibodies can be used successfully in ELISA applications, and individual antibodies must be evaluated. During the
adsorption process, the three-dimensional structure of an antigen might be altered and such that it can no longer bind its
target epitope. Some antibodies are raised against peptides; if the peptide represents an internal sequence of the antigen, then
the antibody might not bind if the whole antigen is immobilized on the plate. Furthermore, for an antibody to work
successfully in ELISA, it should react specifically with the antigen but not cross-react with a component of the blocking
buffer. For sandwich assays where two different antibodies are required, it is essential that the two antibodies react with
different epitopes on the antigen or an epitope that appears several times on the antigen. For example, if the antigen is
immobilized on the plate through the capture antibody, then the detection antibody must be able to interact with its own
epitope without steric hindrance from the first antibody or the plate. In ELISA applications where a secondary antibody is
used as part of the detection complex it is also essential that the capture and detection antibodies be raised in different animal
species so that the secondary antibody does not react with the coating antibody. Antibodies that work in combination with
each other are generally known as “matched pairs”.
Another important factor to consider when setting up an ELISA is the concentration of the antibodies. Each will require
optimization, the optimal range being partially determined by the form and origin and also by the substrate used for signal
generation. Detection antibody and enzyme conjugate working solutions should ideally be prepared in blocking solution to
reduce non-specific interactions. For recommended antibody concentrations see the Appendix at the end of this document.
C. Blocking Buffer
Blocking buffers usually consist of formulations of proteins designed to prevent non-specific binding of proteins to the plate.
An optimal blocking buffer maximizes the signal-to-noise ratio and does not react with the antibodies or target protein. If
cross-reactivity is observed then a different blocker should be tested and if repeated cross-reactivity is observed it may be
advisable to switch to a non-mammalian protein blocker such as salmon serum or a protein free blocking solution. Visit our
website for more information on our wide selection of ready-to-use blocking buffers.
Some systems may benefit from the addition of a surfactant such as Tween®-20 (a gentle non-ionic detergent) to the blocking
solution. Surfactants can help to minimize hydrophobic interactions between the blocking protein(s) and the antigen or the
antibodies. Typically a final concentration of 0.05% (v/v) Tween-20 is used. In addition, blocking buffers should be used in
sufficient volumes to completely coat the wells; for example 300l should be used for each well of a typical 96-well plate.
D. Target Antigen
The target antigen should be present in a buffer or matrix that allows it to interact with a pre-coated capture antibody or be
coupled to the plate directly. Direct coupling may require that the antigen be exchanged into a suitable coupling buffer. In
rare instances the three-dimensional structure of an antigen may be altered during the adsorption process such that it no
longer binds its target epitope. In such cases, the use of a plate pre-coated with a binding protein (such as a capture antibody)
may eliminate this problem. If the antigen is present in the form of a biological sample, then the effects of the matrix (i.e.
serum, plasma components) should be controlled by performing spike-and-recovery and linearity-of-dilution experiments.
For more information on this topic, see the related Tech Tip #58: Linearity and spike-and-recovery experiments for ELISA.
If performing a quantitative ELISA it is essential to have an equivalent standard protein (generally a purified recombinant)
where the amount of specific protein is known in advance. The standard is used to prepare a series of solutions of known
6
concentration by serial dilution of a protein stock solution followed by construction of a standard curve plotting concentration
versus absorbance. Absorbance values for samples of unknown concentration are extrapolated from this curve to determine
the actual amount of specific protein in the sample.
E. Enzyme Conjugate
The concentration of the enzyme conjugate is one of the most crucial parameters in the optimization process. The amount of
enzyme that binds directly influences the amount of signal that is generated. Too little enzyme and the signal may be very
weak with a poor signal-to-noise ratio. Too much and the background may be too high, again resulting in a poor signal-to-
noise ratio and little distinction between standards of different concentrations. For recommended enzyme conjugate
concentrations see the Appendix at the end of this document.
7
detection system is altered (for example a different antibody or substrate), then some of the other components may also need
to be re-optimized.
8
General Capture/Sandwich ELISA Protocol
This protocol represents an example of a standard capture or sandwich ELISA using a biotinylated detection antibody and
streptavidin-HRP indirect detection system with commonly used reagents and TMB (tetramethyl benzene) substrate. For
other ELISA variations, see the section entitled Modifications to the ELISA Protocol at the end of this procedure.
2:0000
1. Add 50-100 l coating 2. Cover plate, incubate for 3. Wash plate three 4. Add 300 l
antibody to each well 2 hours at room temp to times, 5 minutes each blocking buffer to
overnight at 4C each well
1:0000 1:0000
5. Cover plate and incubate 6. Remove blocker, add sample 7. Cover plate and incubate 8. Wash plate three times,
at room temp for 1 hour to or standards to each well in at room temp for 1 hour 5 minutes each
overnight at 4C. duplicate or triplicate
1:0000
9. Add biotinylated 10. Cover plate and 11. Wash plate three times, 12. Add enzyme conjugate
detection antibody to incubate at room 5 minutes each to each well
each well temperature for 1 hour
1:0000 0:3000
13. Cover plate and 14. Wash plate six times, 15. Add substrate solution 16. Develop at room temp.
incubate at room temp for 5 minutes each to each well for 30 minutes, stop reaction
1 hour if necessary
9
A. Materials Required
Hardware
Clear 96 well plate
Multi-channel precision pipettor with disposable plastic tips
Plate reader or luminometer equipped to detect the substrate
Reagents
Coating buffer: 0.2 M sodium carbonate/bicarbonate, pH 9.4
Capture antibody: Diluted in Coating Buffer (see Appendix for appropriate concentration range).
Wash buffer: 0.1 M phosphate, 0.15 M sodium chloride, pH 7.2 containing 0.05% Tween 20
Note: 0.1M Phosphate can be replaced by 25 mM Tris, pH 7.2
Blocking buffer: 2% (w/v) Bovine Serum Albumin (BSA) in Wash Buffer
Note: alternative buffers are listed in the product list at the end of this document
Standard diluent: 2% (w/v) BSA in Wash Buffer. Note: Ideally the standard diluent composition would be as close as
possible to the sample matrix. For example if measuring the concentration of an antigen in culture
supernatant, culture medium should be used as the standard diluent. However biological sample
matrices such as serum are impossible to replicate, therefore BSA is commonly used in these
instances. Often the blocking buffer is also used as the standard diluent
Samples/standards: See preparation sections below.
Detection antibody Diluted in 1/5 strength standard diluent (see Appendix for appropriate concentration range)
(biotinylated):
Enzyme conjugate: Streptavidin-HRP diluted in 1/5 strength standard diluent (see Appendix for appropriate
concentration)
Substrate: TMB substrate (see product list for a list of alternative substrates)
Stop solution: 2M sulfuric acid
B. Preparation of Standards
Typically a standard curve may span concentrations from 0 to 1000 pg/ml but may go as high as 3000 pg/ml depending on
the predicted amount of antigen in the sample and the amount of standard protein available. Typically two-fold or three-fold
dilutions of the standard protein are prepared from the stock solution using the standard diluent. When preparing serial
dilutions of a protein standard, use fresh tips after each dilution.
C. Preparation of Samples
If the concentration of antigen in the sample potentially exceeds the highest point of the standard curve (i.e. > 1,000 pg/ml)
prepare one or more dilutions of the sample using the standard diluent.
D. Procedure
Important: Do not allow the plate to dry at any point.
1. Dilute the capture antibody to the appropriate concentration allowing sufficient volume for 50-100 l per well.
2. Add the diluted capture antibody to the plate, cover and incubate for 2 hours at room temperature (RT).
3. Remove the solution and wash the plate with 200 l per well wash buffer for 3 x 5 minutes on a shaking platform.
4. Add 300 l blocking buffer per well, cover the plate and incubate for 1 hour at room temperature. Alternatively block
overnight at 4C.
5. Prepare the samples and standards. The volume per well should be the same as the capture antibody used in step 1.
6. Remove the blocking buffer and add the samples and standards. Cover the plate and incubate for 1 hour at RT.
10
7. Remove the solution and wash the plate with 200 l per well wash buffer for 3 x 5 minutes on a shaking platform.
8. Dilute the biotinylated detection antibody to the appropriate concentration. The volume per well should be the same as
the capture antibody used in step 1.
9. Add the diluted detection antibody to the plate, cover and incubate for 1 hour at RT.
10. Remove the solution and wash the plate with 200 l per well wash buffer for 3 x 5 minutes on a shaking platform.
11. Dilute the enzyme conjugate to the appropriate concentration. The volume per well should be the same as the capture
antibody used in step 1.
12. Add the diluted enzyme conjugate to the plate, cover and incubate for 1 hour at RT.
13. Remove the solution and wash the plate with 200 l per well wash buffer for 6 x 5 minutes on a shaking platform
14. Add substrate solution to the plate. The volume per well should be the same as the capture antibody used in step 1.
15. Incubate the plate at RT until the desired color intensity is reached. Ideally a clear gradient will result for the standards.
16. Stop the reaction by adding an equal amount of stop solution.
17. If using TMB, measure the absorbance at 450 nm. For other substrates use the appropriate detection technique.
11
G. Competition ELISAs
In a competition ELISA, the assay is based on competition between the antigen in a standard/sample and an enzyme
conjugated version of the same antigen for a limited amount of antibody bound to a pre-coated plate. Both are mixed together
in the same well. As the concentration of antigen in the sample increases, the amount of labeled antigen captured by the
coating antibody decreases. Therefore there is an inverse relationship between optical density (OD) and the amount of analyte
in the sample. Amounts of labeled and non-labeled antigen to use in the assay should be determined empirically.
ELISA Optimization
This section describes optimization steps for each component of the assay starting at the capture antibody (for sandwich
ELISAs) right through the enzyme conjugate and choice of substrate. Although each component is described separately, in
many instances it is possible to optimize two components simultaneously by performing a checkerboard titration as shown in
the layout below.
12
D. Optimizing Sample Concentration
1. Prepare different concentrations of the sample using standard diluent. Test a wide range of sample concentrations,
keeping in mind the detection limit of the substrate being used.
2. Apply an equal volume of each concentration to the plate and proceed from step 6 of the protocol.
3. Check for strong signal versus low background.
4. To confirm that the biological sample matrix is not masking or enhancing the signal, spike-and-recovery and linearity-of-
dilution experiments should be performed. For more information on this topic, see the related Tech Tip #58: Linearity
and spike-and-recovery experiments for ELISA.
E. Optimizing the Detection Antibody Concentration
1. Prepare different concentrations of the detection antibody in standard diluent (see ranges described in the Appendix).
2. Apply an equal volume of each concentration to the plate and proceed from step 9 of the protocol.
3. Check for strong signal versus low background.
F. Optimizing the Enzyme Conjugate Concentration
1. Prepare different concentrations of the enzyme conjugate in standard diluent according to the range described in the
Appendix. Ensure the concentration is in accordance with the range described for the substrate.
2. Apply an equal volume of each concentration to the plate and proceed from step 12 of the protocol.
3. Check for strong signal versus low background.
G. Optimizing Signal Detection
1. Select substrate(s) based on likely amount of antigen in sample and ability to detect with appropriate instrument.
2. Apply the working solution to the plate and proceed from step 14 of the protocol.
3. If antigen can clearly be detected over a dynamic range then the substrate is appropriate. If the antigen is below the
threshold for detection then select a more sensitive substrate.
Appendix
The following tables provide recommended ranges for different ELISA components. Concentrations are guidelines only; for
best results optimize each component individually.
Recommended starting concentration ranges for coating and detection antibodies for ELISA optimization. The
use of non-purified antibodies will work but may result in higher background. It is generally recommended to use affinity
purified antibodies for optimal signal:noise ratio.
Source Coating Antibody Detection Antibody
Polyclonal serum 5-15 g/ml* 1-10 g/ml*
Crude ascites 5-15 g/ml* 1-10 g/ml*
Affinity purified polyclonal 1-12 g/ml 0.5-5 g/ml
Affinity purified monoclonal 1-12 g/ml 0.5-5 g/ml
Recommended secondary antibody concentrations for ELISA in different systems. Check the instructions
for the substrate as they may recommend a more defined concentration range for the enzyme conjugate.
Enzyme System Concentration
HRP Colorimetric system 20-200 ng/ml
Chemifluorescent system 25-50 ng/ml
Chemiluminescent system 10-100 ng/ml
AP Colorimetric system 100-200 ng/ml
Chemiluminescent system 40-200 ng/ml
13
References
Engvall E and Perlmann P (1971). Enzyme linked immunosorbent assay (ELISA) quantitative assay of immunoglobulin G. Immunochemistry, v8 p871-875.
Kohler C and Milstein C (1975). Continuous cultures of fused cells secreting antibody of predefined specificity. Nature, v256 p495-497.
Avrameas S, Uriel J. !966). Méthode de marquage d’antigènes et d’anticorps avec des enzymes et son application en immunodiffusion. C R Acad Sci Hebd
Seances Acad Sci D. v262 p2543-2545.
Avrameas S. (1969). Coupling of enzymes to proteins with glutaraldehyde. Immunochemistry v6 p43-52.
Nakane PK and Pierce GB (1967). Enzyme-labeled antibodies for the light and electron microscopic localization of tissue antigens. J Cell Biol v33 p307-
318.
SuperSignal® Technology is protected by U.S. Patent # 6,432,662.
Tween® is a registered trademark of ICI Americas.
© 2010 Thermo Fisher Scientific Inc. All rights reserved. Unless otherwise indicated, all trademarks are property of Thermo Fisher Scientific Inc. and its
subsidiaries. Printed in the USA.
14