CU1000 Operating Manual
CU1000 Operating Manual
CU1000
Operating Unit
Catalog No.
560-320
jina54en1-09-(2404)
From software version
3.16 (LDS3000) / 3.14 (CU1000)
INFICON GmbH
Bonner Strasse 498
50968 Cologne, Germany
INFICON Table of Contents
Table of Contents
1 About these instructions ............................................................................................................................... 5
1.1 Target groups........................................................................................................................................... 5
1.2 Other associated documents ................................................................................................................... 5
1.3 Warnings .................................................................................................................................................. 5
2 Safety............................................................................................................................................................... 6
2.1 Intended use ............................................................................................................................................ 6
2.2 Duties of the Operator.............................................................................................................................. 6
2.3 Owner requirements................................................................................................................................. 6
2.4 Dangers.................................................................................................................................................... 7
4 Description...................................................................................................................................................... 9
4.1 Device setup ............................................................................................................................................ 9
4.2 Function ................................................................................................................................................. 10
4.3 Technical data........................................................................................................................................ 10
5 Installation .................................................................................................................................................... 11
5.1 Connecting the control unit .................................................................................................................... 11
5.2 Installing the control unit ........................................................................................................................ 11
CU1000-Operating-Instructions-jina54en1-09-(2404) iii
Table of Contents INFICON
8 Gas library..................................................................................................................................................... 36
iv CU1000-Operating-Instructions-jina54en1-09-(2404)
INFICON About these instructions | 1
1.3 Warnings
DANGER
Imminent hazard resulting in death or serious injuries
WARNING
Hazardous situation resulting in potential death or serious injuries
CAUTION
Hazardous situation resulting in minor injuries
NOTICE
Hazardous situation resulting in damage to property or the environment
CU1000-Operating-Instructions-jina54en1-09-(2404) 5 / 46
2 | Safety INFICON
2 Safety
6 / 46 CU1000-Operating-Instructions-jina54en1-09-(2404)
INFICON Safety | 2
2.4 Dangers
• Only use the device away from areas with a risk of explosions.
CU1000-Operating-Instructions-jina54en1-09-(2404) 7 / 46
3 | Shipment, Transport, Storage INFICON
► Please check the scope of delivery of the product for completeness after receipt.
Transport
NOTICE
Damage due to unsuitable packaging material
Transport in unsuitable packaging material can damage the device.
► Transport the device only in the original packaging.
► Keep the original packaging.
Storage ► Always store the device in compliance with the technical data, see "Technical
Data".
8 / 46 CU1000-Operating-Instructions-jina54en1-09-(2404)
INFICON Description | 4
4 Description
Status LED
Status LED illuminated Control unit operates normally
Status LED flashing Display is set to power saving mode
CU1000-Operating-Instructions-jina54en1-09-(2404) 9 / 46
4 | Description INFICON
4.2 Function
You can use the control unit to configure the mass spectrometer module LDS3000. It
also lets you output the data stored in the MSB box.
Electrical data
560-320
Memory capacity for measured data 16 MB
Ambient conditions
560-320
Max. altitude above sea level 2000 m
Max. relative humidity above 40 °C 50%
Max. relative humidity from 31 °C to 40 °C 80% to 50% (linear abfallend)
Max. relative humidity to 40 °C 80%
Permissible ambient temperature (during 10°C - 45°C
operation)
Storage temperature -20°C - 60°C
Pollution degree 2
10 / 46 CU1000-Operating-Instructions-jina54en1-09-(2404)
INFICON Installation | 5
5 Installation
DANGER
Hearing damage from loud volume setting
Loud volume setting can damage hearing.
► Do no set volume of headphones too loud.
CU1000-Operating-Instructions-jina54en1-09-(2404) 11 / 46
5 | Installation INFICON
12 / 46 CU1000-Operating-Instructions-jina54en1-09-(2404)
INFICON Operation CU1000 | 6
6 Operation CU1000
NOTICE
Damage to touch screen from incorrect operation.
The touch screen can be damaged with a hard or pointed item.
► Operate touch screen with fingers only.
CU1000-Operating-Instructions-jina54en1-09-(2404) 13 / 46
6 | Operation CU1000 INFICON
14 / 46 CU1000-Operating-Instructions-jina54en1-09-(2404)
INFICON Operation CU1000 | 6
1 - Keyboard lock
The control unit is locked or unlocked by pressing and holding the icon for the
keyboard lock.
4 - Ser
The registered operator is shown abbreviated.
Display Meaning
Ope Operator
Sup Supervisor
Int Integrator
Ser SERVICE
5 - Zero
Background suppression is active.
6 - Caution icon
Active warnings are stored in the unit.
The active warnings can be displayed via the menu "Info > History > Active warnings".
7 - Tracer gas
Set tracer gas and tracer gas concentration percentage.
Display Meaning
He Helium (4He)
H2 Hydrogen
M3 E.g. H-D, 3He or H3
CU1000-Operating-Instructions-jina54en1-09-(2404) 15 / 46
6 | Operation CU1000 INFICON
8 - Operation mode
Configured operation mode
9 - Leak rate
Current measurement for the leak rate.
10 - Graph
Graphic display of the leak rate Q(t).
11 - Time axis
Time axis of the leak rate Q(t).
16 / 46 CU1000-Operating-Instructions-jina54en1-09-(2404)
INFICON Operation CU1000 | 6
16 - Value axis
Value axis of the leak rate Q(t).
17 - Device of measurement
Device of measurement of the value axis.
WARNING
Number: 500
Displayed
Pressure sensor not connected
message
with details (Value = 4.23E-2)
To the display
Further errors or warnings
further active
501 502 errors or
warnings
(if present)
Detailed display
further active
errors or Information to
To confirm
warnings cause or remedy
all messages
(if present) to the message shown
above
You will find an overview of possible errors and warnings in the operating instructions
of the LDS3000 (mass spectrometer module), chapter "Warning and error messages".
CU1000-Operating-Instructions-jina54en1-09-(2404) 17 / 46
6 | Operation CU1000 INFICON
• the user is not authorized to change the values, see also “Operator types and
authorizations [} 21]“.
• an older version of the software run by mass spectrometer module LDS3000 does
not support this parameter.
Autoscale
Disabled: You can change the display by pressing on the intersection of the
coordinate axes and then swiping your finger along the desired axis and releasing it,
or if you press on the end of the desired coordinate axis and swipe along in the
direction of the intersection of the axes and release.
On: The display is automatically adjusted depending on the leak rate.
Control unit Display > Time axis > Time axis scale
18 / 46 CU1000-Operating-Instructions-jina54en1-09-(2404)
INFICON Operation CU1000 | 6
Diagram
Bar graph
Control unit Display > Measurement view > Measurement view mode
Trigger display on the Selection of the trigger (leak rate threshold) displayed on the touch screen.
touch screen
1
2
3
4
Assigning favorite The favorite buttons offer direct access to individual functions. They can be assigned
buttons with access control "Supervisor" or higher by the user.
Favorite 1: Middle button (see the figure in “Measurement display elements [} 14]“).
Favorite 2: Right button
Favorite 3: Button on the bottom right of the main menu.
Volume Flow switching
Display settings Check CAL
Start/Stop At AQ additionally: AQ Wizard
Measurement view Gas equivalent
(at AQ instead of : AQ, at instead of : ) - - - (= without function)
CAL
Displaying warnings The display of warnings on the touchscreen can be allowed or suppressed.
on the touchscreen
Off
On
CU1000-Operating-Instructions-jina54en1-09-(2404) 19 / 46
6 | Operation CU1000 INFICON
Control unit Settings > Set up > Control unit > Messages > Show
warnings
Show calibration note Suppress or allow the calibration note with the following content:
• Leak rate of the applied calibration leak
• No calibration should take place during the first 20 mins
OFF (suppressed)
ON (allowed)
Control unit Settings > Set up > Control unit > Messages > Show
calibration notes
Show calibration The display of the calibration request can be permitted or suppressed.
request To activate or deactivate the calibration request as such, see "Activating the
calibration request".
OFF (suppressed)
ON (allowed)
Control unit Settings > Set up > Control unit > Messages > Show
calibration request
Setting the audio Output of an acoustic signal depending on the leak rate
alarm
--- No sound
Proportional: The frequency of the audible signal is proportional to the bar graph
display or diagram height. The frequency range is 300Hz to 3300Hz.
Setpoint: The pitch is proportional to the leak rate. The signal sounds if the leak rate
exceeds the selected trigger value.
Pinpoint: The sound of the acoustic signal changes its frequency within a specific
range of leak rates. Range: A decade below the selected trigger threshold up to one
decade above. The sound keeps at a constant low and a constant high frequency
below and above this range, respectively.
Trigger: If the selected trigger threshold is exceeded, a two-pitch signal sounds.
Control unit Settings > Set up > Control unit > Audio > Audio alarm mode
Behavior with warnings or error messages: If the touch screen shows a warning or
an error, then a two-pitch signal sounds simultaneously.
Automatic switch off The touch screen can be switched off automatically after a specific time without any
of the touch screen operation to save energy.
20 / 46 CU1000-Operating-Instructions-jina54en1-09-(2404)
INFICON Operation CU1000 | 6
30 s 10 min
1 min 30 min
2 min 1h
5 min ∞ (=never)
Control unit Settings > Set up > Control unit > Energy > Display off after
For the types "Integrator", "Supervisor" and "Operator", a four-digit PIN must be
assigned during registration (0000 ... 9999). "0000" is assigned to all operators ex
works.
If an operator keeps the pin "0000", this operator will always be registered is during
the start up of the system (without PIN query).
A key-operated switch can be used in addition to a PIN if an I/O module is connected.
The key-operated switch is connected to the I/O module via three digital inputs (see
operating instructions of the LDS3000).
The following table shows the authorizations of individual operator types.
CU1000-Operating-Instructions-jina54en1-09-(2404) 21 / 46
6 | Operation CU1000 INFICON
Load parameters The saved/backed-up parameters of control unit CU1000 and of the mass
spectrometer module can be loaded from a USB stick.
Control unit Function > Data > Parameters > Load
Save parameters The parameters of control unit CU1000 and of the mass spectrometer module can be
saved to a USB flash drive.
Control unit Function > Data > Parameters > Save
Display error The type of error information can be set differently for each operator type. The
information Integrator always receives the complete information.
Number: Message number
text: Brief description
Info: Expanded message information
• Only numbers
• Number and text
• Number, text and info
Control unit Function > Data > Parameter > Error info
Viewer (Operator, Supervisor)
Parameter list display Parameters can be displayed as an alphabetical list with names and current value s.
and change Each list entry is a button which, when pressed, will open the parameter's set-up
dialog box.
Control unit List > Parameters list or:
Functions > Data > Parameters > List
Display list of Parameters can be displayed as an alphabetical list with names and current change
parameter change authorizations. Each list entry is a button which, when pressed, will change access
authorizations control. Changes are possible in accordance with the hierarchy of the operator.
Control unit Functions > Data > Parameters >
Parameter Access
Access controls The authorization for changing parameters can be reset to factory setting.
Control unit Functions > Data > Parameters > Reset >
Param. Access control
22 / 46 CU1000-Operating-Instructions-jina54en1-09-(2404)
INFICON Operation CU1000 | 6
Control unit The control unit settings can be reset to factory settings.
Control unit Functions > Data > Parameters > Reset >
Control unit settings
Memory location The data stored in the control unit can be saved to a USB stick. The memory in the
control unit is limited to the recording of a 24-hour measurement. Each time after one
hour has elapsed, the file is closed and recording continues in the next file.
• USB flash drive
• Control unit
Control unit Functions > Data > Recorder > Settings >
Storage location
Copy data Copy data from the internal memory of the control unit to a connected USB stick.
Control unit Functions > Data > Recorder > Copy >
Copy files
Delete data Delete data in the internal memory of the control unit
CU1000-Operating-Instructions-jina54en1-09-(2404) 23 / 46
6 | Operation CU1000 INFICON
Control unit Functions > Data > Recorder > Delete >
Delete files
24 / 46 CU1000-Operating-Instructions-jina54en1-09-(2404)
INFICON Operation CU1000 | 6
2
1
CU1000-Operating-Instructions-jina54en1-09-(2404) 25 / 46
6 | Operation CU1000 INFICON
Scope
The explanations on the equivalence rate only refer to sniffing operation.
If you measure with the test gases helium or hydrogen, but want to display another
gas with its leak rate, use a correction factor for the test gas used.
26 / 46 CU1000-Operating-Instructions-jina54en1-09-(2404)
INFICON Operation CU1000 | 6
2
Fig. 7: Measurement screen with displayed equivalence leak rate and configured favorites key
CU1000-Operating-Instructions-jina54en1-09-(2404) 27 / 46
6 | Operation CU1000 INFICON
The options "Switch off" and "Gas equivalent no. 1...4" overwrite parameters, see "Set
equivalence factor and molar mass [} 30]".
If you select the option "User-defined", parameters then have to be configured, see
"Set equivalence factor and molar mass [} 30]".
28 / 46 CU1000-Operating-Instructions-jina54en1-09-(2404)
INFICON Operation CU1000 | 6
ð This is the gas proportion of the tracer gas in percent, e.g. for forming gas
(95/5) it would be 5%.
7 Window "Absolute pressure measuring gas".
ð Corresponds to the absolute pressure of the tracer gas in the test object in
bar.
Example
An air conditioning system is to be checked for leaks. The system is first filled with 2
bar (absolute) pure helium and checked for leaks. Later the plant will be filled with
R134a. The operating pressure is 15 bar (absolute).
This results in the following values for the above-specified parameters:
Absolute pressure equivalence gas = 15.0
Measuring mass = 4
Percentage of measuring gas = 100.0
Absolute pressure measuring gas = 2.0
CU1000-Operating-Instructions-jina54en1-09-(2404) 29 / 46
6 | Operation CU1000 INFICON
5 Control unit: Settings > Set up > Operation modes > Equivalence rate
6 “Molar mass" button
ð (LD protocol: Command “470”)
7 Select "Mass 2", "Mass 3" or "Mass 4" to match your test gas as described
above.
ð If the test gas is set to helium, the window "Molar mass equivalent gas He"
opens.
8 Set your molar mass. In the example for 102:
ð If the equivalence factor is not equal to 1 or the molar mass is not set to factory
settings, the equivalence factor is displayed both on the calibration result and on
the measurement screen.
30 / 46 CU1000-Operating-Instructions-jina54en1-09-(2404)
INFICON Operation CU1000 | 6
Fig. 8: Top left: Display of Molar Mass (102) and Equivalence Factor (127)
NOTICE
Aborted connection
Loss of data due to disconnection
► Do not switch off the device and do not remove the USB flash drive while the
software is being updated!
► Switch the device off and back on after a software update has taken place.
CU1000-Operating-Instructions-jina54en1-09-(2404) 31 / 46
6 | Operation CU1000 INFICON
6.3.7.2 Checking and updating the software version of the MSB box
The current software is available from the Inficon support.
The functions of the XL Sniffer adapter set are taken into consideration in system
software version 2.11 or higher.
1 Copy the file with the file name extension ".bin" into the root directory of a USB
stick.
2 Connect the USB flash drive to the USB port on the device.
3 Select: "Functions > Data > Update > MSB".
ð The display shows information on the current and the new software version
as well as on the boot loader.
4 Check the version information.
ð Select the "Start" button to start the update.
ð Do not switch off the device and do not remove the USB flash drive while the
software is being updated! Do not switch off the device and do not remove
the USB flash drive while the software is being updated!
5 Follow the instructions on the touchscreen and wait until the update is
complete.
6 If the system displays warning 104 or 106, confirm with "C".
32 / 46 CU1000-Operating-Instructions-jina54en1-09-(2404)
INFICON Operation CU1000 | 6
2 Connect the USB flash drive to the USB port on the device.
3 Select: "Functions > Data > Update > I/O module"
ð The display shows information on the current and the new software as well as
on the current boot loader.
4 Check the version information.
5 Select the "Start" button to start the update.
ð Do not switch off the device and do not remove the USB flash drive while the
software is being updated!
6 Follow the instructions on the touchscreen and wait until the update is
complete.
ð The following tips are shown after selecting the "Start" button on the
touchscreen:
• Connect and switch on the IO1000.
• Activate boot mode (switch DIP S2.3 on and off once).
• When the STATUS LED flashes green, press OK.
CU1000-Operating-Instructions-jina54en1-09-(2404) 33 / 46
7 | Decommissioning the measuring instrument INFICON
WARNING
Danger due to harmful substances
Contaminated devices could endanger health. The contamination declaration serves
to protect all persons who come into contact with the device.
► Fill in the declaration of contamination completely.
34 / 46 CU1000-Operating-Instructions-jina54en1-09-(2404)
INFICON Decommissioning the measuring instrument | 7
CU1000-Operating-Instructions-jina54en1-09-(2404) 35 / 46
8 | Gas library INFICON
8 Gas library
The operating software of the device contains a list of approx. 100 gases which could
be relevant in the refrigeration industry.
The list is stored in the nonvolatile flash memory of the operating unit and can be
updated. The user can access this list while pre-defining the equivalence gases, see
"Configure gas list [} 28]". The user can then select from the pre-defined gases during
gas equivalent selection, see "Gas equivalent selection [} 27]".
The library of the device has the following factory-defined content:
36 / 46 CU1000-Operating-Instructions-jina54en1-09-(2404)
INFICON Gas library | 8
CU1000-Operating-Instructions-jina54en1-09-(2404) 37 / 46
8 | Gas library INFICON
38 / 46 CU1000-Operating-Instructions-jina54en1-09-(2404)
INFICON Gas library | 8
CU1000-Operating-Instructions-jina54en1-09-(2404) 39 / 46
8 | Gas library INFICON
40 / 46 CU1000-Operating-Instructions-jina54en1-09-(2404)
INFICON Gas library | 8
CU1000-Operating-Instructions-jina54en1-09-(2404) 41 / 46
8 | Gas library INFICON
42 / 46 CU1000-Operating-Instructions-jina54en1-09-(2404)
INFICON Gas library | 8
CU1000-Operating-Instructions-jina54en1-09-(2404) 43 / 46
9 | CE Declaration of Conformity INFICON
9 CE Declaration of Conformity
44 / 46 CU1000-Operating-Instructions-jina54en1-09-(2404)