USB COMMUNICATIONS INTERFACE

LOOP LINK 5909

DK Side 1 Contents
Warnings............................................................................. 16
UK Page 15 Safety instructions............................................................... 17
EC Declaration of Conformity............................................. 19
FR Page 29 Application.......................................................................... 20
Technical characteristics..................................................... 20
Installation........................................................................... 20
DE Seite 43 System requirements.......................................................... 20
Installation of PReset.......................................................... 20
Installation of USB drivers for Windows® 98SE, ME and XP. 20
Special instructions for Windows® 98 SE and ME......... 21
Special instructions for Windows® XP............................ 22
Installation of USB drivers for Windows® Vista.................. 23
5909 Configuration of USB communication in PReset................ 27
Connection of PR modules using EZ-Mini-Hooks.............. 28
Connection of PR modules using a modular connector..... 28
Loop Link
USB Communications
Interface UK / Europe Office
No. 5909V101-IN (1007) Tel: +44 (0)845 9000 601
F r o m s e r. n o . 0 4 0 6 5 4 0 0 1 Fax: +44 (0)845 9000 602
[email protected]
www.omniinstruments.co.uk

Australia / Asia Pacific Office USA / Canada Office


Tel +61 (0)282 442 363 Tel +1-866-849-3441
Fax +61 (0)294 751 278 Fax +1-866-628-8055
[email protected] [email protected]
www.omniinstruments.com.au www.omniinstruments.net

SIGNALS THE BEST 15


WARNING SAFETY INSTRUCTIONS
When connected to SYSTEM 5000 by the communication cable,. Definitions:
this module can get connected to hazardous electric voltages. Hazardous voltages have been defined as the ranges: 75...1500 Volt DC, and
Ignoring this warning can result in severe personal injury or 50...1000 Volt AC.
mechanical damage. Technicians are qualified persons educated or trained to mount, operate, and
GENERAL To avoid the risk of electric shock and fire, the safety instructions also trouble-shoot technically correct and in accordance with safety regulations.
of this manual must be observed and the guidelines followed. . Operators, being familiar with the contents of this manual, adjust and operate
The module must only be applied as described in the following. . the knobs or potentiometers during normal operation.
Prior to the commissioning of the module, this manual must be
examined carefully. Receipt and unpacking:
Only qualified personnel (technicians) should use this module. Unpack the module without damaging it and make sure that the manual always
If the module is used in a manner not specified by the follows the module and is always available. The packing should always follow
manufacturer, the protection provided by the module may be the module until this has been permanently mounted.
impaired. Check at the receipt of the module whether the type corresponds to the one
ordered.

WARNING Environment:
The following operations must only be carried out on a module Avoid direct sunlight, dust, high temperatures, mechanical vibrations and shock,
that is not connected to SYSTEM 5000 and a PC:. as well as rain and heavy moisture. If necessary, heating in excess of the stated
Troubleshooting the module.. limits for ambient temperatures should be avoided by way of ventilation.
HAZARD­
All modules fall under Installation Category II, Pollution Degree 1, and Insulation
OUS
Repair of the module must be done by PR electronics A/S only. Class II.
VOLTAGE
Mounting:
WARNING Only technicians who are familiar with the technical terms, warnings, and
instructions in the manual and who are able to follow these should connect the
The female communication connector of SYSTEM 5000 is
module.
connected to the input terminals on which dangerous voltages
Should there be any doubt as to the correct handling of the module, please
INSTAL- can occur. It must only be connected to the programming unit
contact your local distributor or, alternatively,
LATION 5909 by way of the attached communication cable.

PR electronics A/S, Lerbakken 10, DK-8410 Rønde tlf: +45 86 37 26 77


SYMBOL IDENTIFICATION:
Triangle with an exclamation mark: Warning/demand.
Potentially lethal situations..
.
The CE mark proves the compliance of the module with the
requirements of the EU directives..
.
The double insulation symbol shows that the module is
protected by double or reinforced insulation.

16 17
Programming: EC DECLARATION OF CONFORMITY
During programming, the measuring and connection of external voltages must
be carried out according to the specifications of this manual and the manual for As manufacturer
the unit to be programmed. PR electronics A/S
The technician must use tools and instruments that are safe to use. Lerbakken 10
Normal operation: DK-8410 Rønde
Operators are only allowed to adjust and operate modules that are safely fixed
hereby declares that the following product:
in panels, etc., thus avoiding the danger of personal injury and damage. This
means there is no electrical shock hazard, and the module is easily accessible. Type: 5909
Name: Loop Link
Cleaning:
When disconnected, the module may be cleaned with a cloth moistened with is in conformity with the following directives and standards:
distilled water. The EMC Directive 2004/108//EC and later amendments
Liability: EN 61326-1
To the extent that the instructions in this manual are not strictly observed, the For specification of the acceptable EMC performance level, refer to the.
customer cannot advance a demand against PR electronics A/S that would electrical specifications for the module.
otherwise exist according to the concluded sales agreement.
The Low Voltage Directive 2006/95/EC and later amendments
EN 61010-1

Rønde, 19 February 2010 Kim Rasmussen.


Manufacturer’s signature

18 19
Application: 1. Connect Loop Link 5909 to the USB port of the PC.
Loop Link 5909 is a USB communications interface for configuration 2. After a short while, the guide ”New hardware found” will be shown.
and monitoring of PR electronics’ PC-programmable modules. The USB This guide will help you through the installation of the first USB
communication is supported by the Windows® versions 98SE, ME, 2000, XP driver.
and Vista. PR modules available in the configuration program PReset ver. 5.0 3. When pressing [Execute] the first USB driver will be installed.
or higher, can be programmed by way of Loop Link 5909. 4. The guide starts again and helps you through the installation of the
second USB driver.
Technical characteristics:
5. When pressing [Execute] the USB driver installation will be
The PC program CD PReset ver. 5.0 or higher contains the necessary
complete.
(unsigned) USB drivers. Loop Link 5909 is supplied from the USB port of the
PC. If you want to conserve power on a laptop, the USB connector of the Special instructions for Windows® 98SE and ME:
5909 can be connected to an externally supplied hub. The communications 1. When Windows® 98SE and ME are searching for the drivers, the
connector on the PR modules is galvanically isolated from the USB connection following message will probably appear:
on the PC.
Installation:
System requirements:
Windows® 98SE, ME, 2000, XP and Vista with the following recommendations:
Memory: 16 MB.
Display resolution: 800 x 600.
Hard disk space: 12 MB 2. After clicking OK, you will see the following dialogue where the path
”CD-ROM drive:\drivers” must be specified. Either use [Browse] or
Installation of PReset: type the path.
The USB communications interface Loop Link 5909 must not be connected to
the USB port of the PC until after installation of the PReset program..
PReset will be installed on your hard disk and the registry will be updated with
all necessary information on the program. A PReset icon will automatically be
placed on the desk top.
1. Insert the CD-ROM in the CD-ROM drive and the dialogue
”InstallShield Wizard PReset - INSTALL” will start automatically.
2. When the installation program has been started, on-screen
guidelines will guide you through the rest of the installation.
3. If AutoRun is disabled in Windows® 98SE, ME, 2000, XP and Vista,
click the ”Start” button, select [Run] and start the installation
program ”CD drive:\PResetWin\setup.exe”.

3. The installation will then continue according to the guidelines


Installation of USB drivers for Windows® 98SE, ME and XP specified in ”Installation of USB drivers for Windows® 98SE, ME and
Install the USB drivers while the PReset CD is still in the drive, i.e. immediately XP”.
after the PReset installation. Please note that you need to install two USB
drivers.

20 21
Special instructions for Windows® XP: Installation of USB drivers for Windows® Vista:
1. You must say [No, not this time] to the following dialogue: Install the USB drivers while the PReset CD is still in the drive, i.e. immediately
after the PReset installation.
1. Connect Loop Link 5909 to the USB port of the PC.
2. After a short while, the ”Found New Hardware Wizard” will be
shown. This guide will help you through the installation of the USB
driver.
3. Click [Cancel] in the following dialogue:

2. You must click [Continue anyway] in the following dialogue:

4. Click [Cancel] in the following dialogue:

3. The installation will then continue according to the guidelines


specified in ”Installation of USB drivers for Windows® 98SE, ME and
XP”.

22 23
5. Click [Close] in the following dialogue: 7. Click [Browse my computer for driver software] in the following
dialogue:

6. Open the Device Manager and right click [Unknown device]. Then
left click [Update Driver Software]. 8. You will see the following dialogue where the path ”CD-ROM drive:\
Drivers” must be specified. Either use [Browse] or type the path.
Then click [Next]:

24 25
9. Click [Install this driver software anyway] in the following dialogue: Configuration of USB communication in PReset:
1. Start the PReset program and choose [Functions] [Set COM port],
and the ”Options” window is shown on screen.
2. Select ”Next 5909”. PReset now selects a COM port which will be
linked with the USB port to which Loop Link is connected.
3. If you want to configure Loop Link for other USB ports, you must
connect Loop Link to these ports and install the USB drivers for
each port. The other USB ports will not be linked with a COM port in
PReset until ”Next 5909” has been selected.

10. Click [Close] in the following dialogue:

26 27
Connection of PR modules using EZ-Mini-Hooks: Displays Programmable displays with a wide
selection of inputs and outputs for display of temperature,
1 volume and weight, etc. Feature linearisation, scaling, and
Receiving
2 Disconnect
Equipment difference measurement functions for programming via
+Vsupply PReset software.
Red * Yellow Input
PR 5331, 5333, 5334, 5335 Ex interfaces Interfaces for analogue and digital signals
Black * Green
11 (21) as well as HART® signals between sensors / I/P converters /
13 (23) frequency signals and control systems in Ex zone 0, 1 & 2
and for some modules in zone 20, 21 & 22.
Connector * Connected only for
on-line programming
File Product Input
Output Communicatio
PRetop 5331
n Language Option
08:30:00
Isolation Galvanic isolators for analogue and digital
signals as well as HART® signals. A wide product range
2004-8-10
Date:
Serial no: 043201594
Tag no: PRelectronics

Analog output
Analog input 4 - 20mA

with both loop-powered and universal isolators featuring


Output type:
Pt100 DIN/IEC
Input type: Sensor error: Upscale
0.00 - 50.00 C
Input range:
3-wire
Connection:
------

PR 6331, 6333, 6334, 6335 Loop


Cold junction comp:
1.00 sec
Response time:

Link linearisation, inversion, and scaling of output signals.

Temperature A wide selection of transmitters for DIN


form B mounting and DIN rail modules with analogue
and digital bus communication ranging from application-
USB cable
specific to universal transmitters.

Connection of PR modules using a modular connector: Universal PC or front programmable modules with
universal options for input, output and supply. This range
offers a number of advanced features such as process
calibration, linearisation and auto-diagnostics.
Programming cable
with modular connector

PR 5114, 5115, 5116, 5131


08:30:00
n Language Option
Output Communicatio
File Product Input PRetop 5331
2004-8-10
Date:
Serial no: 043201594
Tag no: PRelectronics

Analog output
Analog input Output type: 4 - 20mA
Pt100 DIN/IEC
Input type: Sensor error: Upscale
0.00 - 50.00 C
Input range:
3-wire
Connection:
------
Cold junction comp:
1.00 sec
Response time:

Loop
Link

USB cable