Cone Beam CT Guided Radiotherapy: Jan-Jakob Sonke

Download as pdf or txt
Download as pdf or txt
You are on page 1of 64

Cone Beam CT guided Radiotherapy

Jan-Jakob Sonke

Acknowledgements
NKI-AVL: Marcel van Herk, Jose Belderbos, Suzanne van Beek, Anja Betgen, Josien de Bois, ~Rianne de Jong, Michel Frenay, Danny Minkema, Tonnis Nuver, Jasper Nijkamp, Floris Pos, Monique Smitsmans, Simon van Kramen, Jochem Wolthaus, Lambert Zijp, Peter Remeijer, Coen Rasch and Joos Lebesque PMH, Toronto David Jaffray Doug Moseley Jeffrey Siewerdsen Beaumont Hospital Di Yan Alvaro Martinez
Elekta Synergy Research Group

Contents
Introduction CBCT Acquisition & Reconstruction CBCT image quality Clinical Implementation & Protocols

CT Acquisition
Conventional CT - Fan beam - 1D detector - 1 rotation = 1 slice Cone-beam CT - Cone beam - 2D detector - 1 rotation = volume (many slices)

kV image guidance: not a new idea !

First isocentric Co-60 machine in Netherlands at NKI (1960)

First Prototype CBCT Guided Linac

6.5 cm

D. Jaffray et al. Int J. Radiat. Oncol. Biol. Phys. 2002

Available Cone Beam Systems

Elekta Synergy

Siemens Artiste

Varian Trilogy C-arms

Bench Top

Acquisition and Reconstruction

Elekta Synergy Research system at NKI

Frame Rate: 5.4 fps; Acquisition Time: 1 - 2 min

Cone beam reconstruction

V ( x, y, z ) = W2 ( (W1 p( , a ( x, y, z , ), b( x, y, z , ) )) g (a ) ) d
0

dimx * dimy * dimz * Nproj computations: 5 * 109 for 2563

Imaging Field of View

Field of View
180 + fan-angle

Central detector position

Field of View: Offset Detector

Partially displaced detector position

Offset Detector

Partially displaced detector position

Offset Detector

Panel Position Central Partially displaced Fully displaced

FOV 25 cm 40 cm 50 cm

Partially displaced detector position

Geometric Calibration and QA

Geometry: Flex calibration


0.5 0.4 0.3 0.2 0.1 0 -0.1 -0.2 -200 XG (AB) YG (GT) -150 -100 -50 0 50 angle [ ] 100 150 200

Calculate center of ball bearing for all gantry angles Generate Lookup table for U & V displacements Lookup table includes Set-up error BB

displacement [cm]

Geometry: kV to MV Isoc Calibration

Gantry& Collimator Angle:

-180, -90, 0, 90, 180

Determine COG field edge & BB Calculate mean setup error

QA Phantom

QA Geometrical Accuracy
Planning CT CBCT CT

Match 3D

same ?

Match 2x 2D

DRR

EPI

3D Imaging Performance and Artifacts

Sources of cupping and streaks


Missing data (truncation)
Detector field of view 25 cm

Scattered radiation
Extra signal not from local anatomy Adds noise !

Beam hardening
Attenuation of patient smaller than expected

Ghosting
High exposure signal gives residual extra signal later

Scatter & Imaging Geometry

Cone Beam CT

Fan Beam CT

Scatter-to-primary ratio (SPR) in excess of 300% occur in lateral pelvic projection data occur for CBCT geometry

Strategies for Scatter Management


Select Minimize FOVCC to minimize SPR Optimize Air gap 0.5 0.6 m Compensators (e.g. BowTie filters) Reject Anti-scatter grid Siewerdsen et al. Med.Phys. Dec2004 Correct Scatter correction algorithm

Courtesy Jaffrey Siewerdsen

Shading

Scatter correction algorithm

Without correction
Assumption: scatter uniform and proportional to average image intensity where there is patient in the beam

With correction

Boellaard et al. Two-dimensional exit dosimetry using a liquidfilled electronic portal imaging device and a convolution model Radiother. Oncol. 44 149-157, 1997

CBCT versus Fan Beam CT


FBCT CBCT CBCT

FBCT

Motion

Moving Gas

Smitsmans et al. Med Phys. 2005

Image quality
Diet, given by a dietician based on the patients own insight, starting 7 days before treatment

Mild laxatives: Magnesium-oxide tablets (1 gram) 2 nights before CT scan and during treatments No scans/treatments before 10 am

CBCT

Moving structures are blurred over their trajectory

4D CBCT

Retrospective sorting of the projections before reconstruction yields 4D data

3D versus 4D CBCT

4D Data set 8 x 84 projections

3D Data set 670 projections

The Amsterdam Shroud (Lambert Zijp)

CC position

X-ray image #
Breathing Signal automatically extracted from projection data

Clinical Implementation

Clinical Implementation CBCT @ NKI-AvL


First clinical images on July 9th, 2003 Special team of 4 radiotherapy technicians Normal patient program during the morning Patients with extra CBCT in the afternoon Close cooperation with the physicists

Clinical Implementation CBCT @ NKI-AvL


8 months of validation and improvement of image quality (waiting for CE marking for intervention):
Over 150 scans made to compare with EPID: prostate, head & neck, lung, bladder, sarcoma, stomach and breast patient Different scan protocols were tested Position of the detector Variation in kV and mA Variation in number of frames, by reducing gantry rotation speed

Current situation @ NKI - AvL


Patient set-up is monitored with CBCT for most of our patient groups, using a decision protocol based on bony anatomy match Radiotherapy technicians perform the acquisition, registration and evaluation (bony anatomy) Soft-tissue registrations performed by dedicated radiotherapy technicians in close cooperation with physicists and physicians

Current situation (AvL)


June 2006 We have acquired:
> 6500 CBCT scans On 3 Synergy systems > 700 patients

Archiving

Scenario I
Online Protocol 30 scans per day per machine Storing projections at high resolution (1024^2) 650 * 2 MB per image Storing high resolution scans (0.5 mm voxel size) 256 625 MB per scan ~225 GB per machine per week

Scenario II
Offline Protocol 10 scans per day per machine Storing projections at medium resolution (1024^2) 650 * 0.5 MB per image Storing medium resolution scans (1 mm voxel size) 32 MB per scan ~17 GB per machine per week

Scenario III
Offline Protocol 10 scans per day per machine Storing no projections Storing medium resolution scans (1 mm voxel size) 32 MB per scan ~1.5 GB per machine per week

Set-up Error Bony Anatomy Registration

Image analysis: comparison with reference image


reference localization

Reference image (planning CT)

Localization image (cone beam CT)

Mixed image (not matched)

Automatic matching on region of interest built-in in Synergy system


reference localization reference localization

Tumor in top of neck


Required table shift: (-3.2, -1.5, -0.6) mm

Tumor in lower part of neck


Required table shift: (+1.5, -3.2, -6.1) mm

By zooming in on a region of interest, any target can be accurately localized even if the anatomy changes shape

Matching cone beam to planning CT on bone is highly accurate example for lung treatment series - 10 days matched

Vertebrae are perfectly still

Estimated match accuracy << 1 mm SD, much better as EPID for lung

Can cone beam CT replace EPID ?


As CBCT acquisition is slower but alignment is faster Cone beam CT is matched more accurately Imaging dose is similar or lower

Cone beam CT can safely replace EPID for bony anatomy setup corrections We replaced EPID with cone beam CT The collected data is used to develop soft tissue protocols

Adaptive Radiation Therapy (ART)

Principle
Adaptive Radiation Therapy (ART *) uses imaging information of the first few treatment fractions to re-optimize the treatment plan reduction systematic error reduction treatment margins reduction dose to the rectal wall reduction of rectal toxicity **
* Yan et al., IJROBP 50 (2001) ** Peeters et al., IJROBP jan. (2006)

ART treatment scheme

Conventional plan, 10 mm

Average prostate & rectum adaptive plan, 7 mm**

CBCT first 6 days

weekly monitoring treatment

** unpublished data: Tonnis Nuver (NKI/AVL)

Average prostate
Grey-value registration
TAP / TCC / TLR / RAP / RCC / RLR

** Smitsmans et al., IJROBP 60 (2004)

Automatic prostate localization in CBCT (30 s)

Cone beam CT 10 CBCT scans: automatic bone match

Planning CT contours placed automatically

10 CBCT scans: automatic prostate match help line (GTV+3.6 mm)


Smitsmans et al., IJROBP 2004, 2005

Monitoring the treatment


Visual assessment if the prostate + SV were inside average prostate + 7 mm (PTV volume ART plan)

Variability of 4D CT Patient Models

Repeat 4D cone beam CT

Shows respiration, tumor shrinkage and baseline position variation

Base line shifts

Tumor motion is very similar but occurs at very different places. Verification is essential for accurate treatment

Local Rigid Body Registration

Visual Validation

Multiple Targets

Misalignment of the primary target

Multiple Targets

Correcting alignment of the primary target

Multiple Targets

Correct alignment of primary target might misalign the nodes

Conclusions
Organ motion limits accuracy of radiotherapy Cone-beam CT provides soft tissue contrast, is efficient and does not require moving or touching the patient (4D) CBCT provides a wealth of information (and a huge amount of data!) Dose needed for CBCT scan is considerably smaller than for standard EPID localization fields Image quality sufficient for image guidance

Conclusions
Several soft-tissue and bony anatomy based protocols in routine clinical use Substantial investment and support of vendors required to enable advanced image guided protocols Image Guidance is potentially dangerous. Do not underestimate the residual uncertainties!

Delineation variation: CT versus CT + PET

CT (T2N2) SD 7.5 mm Steenbakkers et al Radiother Oncol. 2005

CT + PET (T2N1) SD 3.5 mm

The beams will be pointed to the target the physician draws !

You might also like