Digital Mammography Systems
Digital Mammography Systems
Company Name Fujifilm GE Healthcare GE Healthcare Hologic, Inc. Hologic, Inc. Hologic, Inc. IMS Philips Planmed Planmed Siemens
Product Name Aspire Cristalle Senographe Pristina Senographe Essential with SenoClaire Selenia Dimensions 3D Selenia Dimensions 2D Selenia Dimensions Avia Package Giotto Image 3D/3DL Microdose SI Planmed Nuance Excel Planmed Nuance Inspiration
FDA cleared, year 2014 510k - November 2016; PMA - March 2017 2014 2011 2008 2012 Yes, 2011 2013 2011 2011 Yes, 2011
Generator type Fujifilm Hi Frequency High frequency single-phase power supply High frequency single-phase power supply Constant potential, high-frequency inverter type; 7kW max.; fully Constant potential, high-frequency inverter type; 7kW max.; fully Constant potential, high-frequency inverter type; 7kW max.; fully High power (8kW) Microprocessor-controlled high frequency Microprocessor-controlled high-frequency X-ray generator Planmed; constant potential high frequency Planmed; constant potential high frequency Microproccessor -controlled
integrated into gantry integrated into gantry integrated into gantry generator high frequency generator
kV range 22 to 49 in 1 kV increments 22 to 49 kV 22 to 49 kV 20-49 kV in 0.5 kV increments 20-49 kV in 0.5 kV increments (limited to 39 kV without 20-49 kV in 0.5 kV increments (limited to 39 kV without 22 kV to 40 kV (adjustable in 0.5 kV increments) 20-40 20-35 kV, +/- 2 kV 20-35 kV, +/- 2 kV 23-35 kv
tomosynthesis option) tomosynthesis option)
mAs Range 2-600 mAs 4 to 600 mAs (depending on track, filter and kV) 4 to 500 mAs (depending on track, filter and kV) 3-500 3-500 3-500 2-600 with increments of 1 mAs 4000 LF (0,3 mm): 5 - 550 mAs (Toshiba)/5 - 600 mAs LF (0,3 mm): 5 - 550 mAs (Toshiba)/5 - 600 mAs Mo anode: 2mAs to 500 mAs
(Varian Mo&W); SF (0.1mm): 5 - 310 mAs (Varian Mo&W); SF (0.1mm): 5 - 310 mAs mode/5 mAs - 500 in AEC mode,
(Toshiba)/5 - 400 mAs (Varian Mo&W) (Toshiba)/5 - 400 mAs (Varian Mo&W) W anode: 2mAs - 630mAs
mode/5 mAs to 630 mAs in AEC
mode
mA range 20 - 200 mA Molybdenum target: 100 mA from 25 to 30kV on large focal spot, 40 mA from 25 to 30 kV on Molybdenum target: 100 mA from 25 to 30kV on large focal spot, 40 mA from 25 to 30 kV 10-200 microns, large focal spot; 10-50 microns, small focal 10-200 microns, large focal spot; 10-50 microns, small focal 10-200 microns, large focal spot; 10-50 microns, small focal spot Up to 150 mA for Large Focus, up to 45 mA for small focus 50-210 LF (0,3 mm): 80-110mA (Mo, Toshiba)/91-120 mA LF (0,3 mm): 80-110mA (Mo, Toshiba)/91-120 mA N/S
small focal spot; Rhodium target: 62 mA from 25 to 30 kV on large focal spot, 35 mA from 25 on small focal spot; Rhodium target: 62 mA from 25 to 30 kV on large focal spot, 35 mA spot spot (Varian, Mo&W) SF (0,1 mm): 20-35mA (Varian, Mo&W) SF (0,1 mm): 20-35mA
to 30 kV on small focal spot from 25 to 30 kV on small focal spot (Toshiba, Mo)/30-42 mA (Varian Mo&W) (Toshiba, Mo)/30-42 mA (Varian Mo&W)
Time range, sec. 0.01 to 10 sec. (61 steps) N/S N/S 0.1-2.5 large focal spot; 10-50 small focal spot 0.1-2.5 large focal spot; 10-50 small focal spot 0.1-2.5 large focal spot; 10-50 small focal spot 20 ms to 5 s 4-16, depending on breast thickness and glandularity LF (0.3 mm): 0.1 - 5 s SF (0.1 mm): 0.1 - 9.9 s LF (0.3 mm): 0.1 - 5 s SF (0.1 mm): 0.1 - 9.9 s exposure times, automatic: 10
ms to 4 seconds (large focus);
60 ms to 6 s (small focus)
AEC detector Auto, semi, manual Two exposure modes: automatic optimization of parameters (AOP) and manual mode Two exposure modes: automatic optimization of parameters (AOP) and manual mode 2-cell, automatic or manual 7-position photocell placement 2-cell, automatic or manual 7-position photocell placement 2-cell, automatic or manual 7-position photocell placement Anrad LMAMII detector; detector has an AEC multiple SmartAEC exposure mode continuously adjusts exposure during 48 automatically selected detector areas 48 automatically selected detector areas N/S
available available esposimete; operation: a) very short pre-exposure to detect image scan, according to feedback from exposure during image
the breast density and b) the thickness difference within scan, according to feedback from exposure sensor
compressed and non compressed breast to determine kV,
filter and starting grid speed; fully microprocessor controlled
Parameters controlled Track, filter, kVp, mAs, AOP or manual Track, filter, kVp, mAs, AOP or manual Manul, auto-time, auto-kV, auto-filter Manul, auto-time, auto-kV, auto-filter Manual, auto-time, auto-kV, auto-filter kV, mAs, Ag/Rh filter N/S kV, mAs and filtration kV, mAs and filtration N/S
Breast Density and Dose
Reporting
Density reporting Optional; vendors: 1. Volpara and 2. VuComp Optional; Volpara Density 3D - specifically designed and cleared fo 3D/ Tomo mammography Optional; Volpara Density 3D - specifically designed and cleared fo 3D/ Tomo Optional, Hologic, Inc. Optional, Hologic, Inc. Optional, Hologic, Inc. N/S Philips Spectral Breast Density Measurement tool (optional) VolparaDensity (optional) VolparaDensity (optional) N/S
software (list if it is systems mammography systems
optional or integrated
and who is the software
vendor)
Vendor 1: generates patient density scorecards for each study; Breast density score compliant with BI-RADS fifth edition Breast density score compliant with BI-RADS fifth edition Volumetric breast density, volume of fibroglandular tissue, Volumetric breast density, volume of fibroglandular tissue, Volumetric breast density, volume of fibroglandular tissue, volume N/S The Philips Spectral Breast Density Measurement uses the N/A N/A N/S
volumetric numbers include volume of fibroglandular tissue, volume of breast, area breast density, BI-RADS density volume of breast, area breast density, BI-RADS density of breast, area breast density, BI-RADS density assessment, etc. spectral data from the photon-counting detector to make a direct
breast volume and volumetric breast density; vendor 2: assessment, etc. assessment, etc. measurement of the fibro-glandular content of the breast
identifies and records the radiographically dense areas of a through spectral material decomposition; the outputs from this
mammogram, and then maps this finding to a calibrated measurement are listed below; it is also possible to configure the
VuCOMP density category system to also save the complete spectral data set
Vendor 1: volumetric numbers include volume of Breast density score compliant with BI-RADS fifth edition Breast density score compliant with BI-RADS fifth edition Configurable Configurable Configurable N/S Volumetric percentage of glandular tissue; volume of glandular Patient information, breast density for each breast Patient information, breast density for each breast N/S
fibroglandular tissue, breast volume and volumetric breast tissue; total breast volume; MicroDose Density Score (correlated and average density for both breasts, VDG/ACR and average density for both breasts, VDG/ACR
density, along with a Volpara Density Grade (VDG) to the BI-RADS breast composition score) breast density category, volume of fibroglandular breast density category, volume of fibroglandular
corresponding to the BI-RADS breast density categories; tissue, volume of breast, average radiation dose/ tissue, volume of breast, average radiation dose/
vendor 2: breast area, dense area, percent breast density, image, average pressure applied using compression image, average pressure applied using compression
density catergory for each case force/contact area force/contact area
Dose software record Vendor 1; yes; vendor 2: no DoseWatch provides dose monitoring and analytics that enables you to drive quality care DoseWatch provides dose monitoring and analytics that enables you to drive quality care Optional, Hologic, Inc. Optional, Hologic, Inc. Optional, Hologic, Inc. N/S Yes (part of device software, not part of breast density N/A N/A N/S
dose improvement so you can deliver accurate diagnoses in a safe manner for each patient improvement so you can deliver accurate diagnoses in a safe manner for each patient measurement tool)
How is density Vendor 1: DICOM secondary capture image; vendor 2: DICOM Structured report and secondary capture Structured report and secondary capture Radiation dose SR, MPPS, image headers Radiation dose SR, MPPS, image headers Radiation dose SR, MPPS, image headers N/S The Spectral Breast Density Measurement parameters are stored N/A N/A N/S
information stored or mammography structured report or secondary capture image as part of the DICOM header of the images and also can be
transferred storage displayed in the DICOM structured report; this enables display on
the review station together with the acquired images and
archiving
Digital Detector V
Type (CCD, aSi, aSe) aSe exclusive hexagonal close pattern pixels Patented needle structure CsI scintillator, single piece construction Patented needle structure CsI scintillator, single piece construction aSe aSe aSe Amorphous selenium (aSe) Crystalline silicon with no dead pixels aSe aSe Amporphous selenium (ase)
Spatial resolution 10 lp/mm 2-D: 7.1 lp/mm; 3-D: 3.5 lp/mm 2-D: 7.1 lp/mm; 3-D: 3.5 lp/mm 2-D: 7.1 lp/mm; 3-D: 3.5 lp/mm 6 lp/mm 10 lpmm 5.88 lp/mm 5.88 lp/mm N/S
Matrix, pixels N/S 2,850 x 2,394 pixels; approx 13 MB per image 3,062 x 2,394 pixels; approx 14 MB per image Yes Yes Yes 2,816 x 3,584 (24 x 30 cm), 2,016 x 2,816 (18 x 24 cm) Large: 24 x 26 cm: 4,915 x 5,355 pixel; small: 24 x 12.5 cm: 4,915 N/S N/S Pixel: 85 um, image matrix: 24 x
x 2,600 pixel; the numbers are a bit larger than expected from 50 30 cm: 2,816 x 3,584, 18 cmx 24
micron pixel size and physical image size because the pixel size cm: 2,016 x 2,816
is defined 20 mm below the patient support; the effective pixel
size on the patient support is 49 microns
Standard size 2,850 x 2,394 3,062 x 2,394 2,560 x 3,328 and 3,328 x 4,096 at 70 microns 2,560 x 3,328 and 3,328 x 4,096 at 70 microns 2,560 x 3,328 and 3,328 x 4,096 at 70 microns N/S ~50 Mb uncompressed, standard configuration 85 x 85 um 85 x 85 um 24 x 30 cm
High-resolution size 28 micron at 1.8 x mag 2,850 x 2,394 3,062 x 2,394 2,560 x 3,328 and 3,328 x 4,096 at 70 microns 2,560 x 3,328 and 3,328 x 4,096 at 70 microns 2,560 x 3,328 and 3,328 x 4,096 at 70 microns N/S N/S N/S N/S N/S
Imaging area N/S N/S N/S 18 x 24 cm and 24 x 29 cm 18 x 24 cm and 24 x 29 cm 18 x 24 cm and 24 x 29 cm N/S N/S N/S N/S N/S
Standard size, cm 18 x 24 cm, 24 x 30 cm 24 x 29 cm 24 x 30.7 cm 18 x 24 cm and 24 x 29 cm 18 x 24 cm and 24 x 29 cm 18 x 24 cm and 24 x 29 cm 17, 14 x 23, 94 (18 x 24) or 23, 94 x 30, 46 (24 x 30) 240 x 260 mm; studies show that this size is sufficient for 99% of 23.94 x 30.46 cm 171 x 239 cm Small: 18 x 24 cm, large: 24 x 30
all breasts, with significant improvements in positioning the cm
patient compared to larger detectors; 240 x 126 mm
(configurable only with spot compression)
High-resolution size, cm 18 x 24 cm, 24 x 30 cm 24 x 29 cm 24 x 30.7 cm 18 x 24 cm and 24 x 29 cm 18 x 24 cm and 24 x 29 cm 18 x 24 cm and 24 x 29 cm 17, 14 x 23, 94 (18 x 24) or 23, 94 x 30, 46 (24 x 30) N/S N/A N/A 16 x 20 cm, mag spot: 9 x 9 cm
Additional Features Includes QC, 1-shot phantom kit, patented comfort DQE at IEC 62220-2-3 equivalent spectrum, at 75_Gy: 70% (+/-3) at 0.5lp/mm and 64% (+/-3) at Integrated, exclusive high transmission cellular (HTC) grid; Integrated, exclusive high transmission cellular (HTC) grid; Integrated, exclusive high transmission cellular (HTC) grid; smart N/S N/S N/S N/S N/S
compression paddles 2 lp/mm; Bucky front cover thickness: 49 mm; optimized room for positioning due to the Bucky smart paddle system; Fast paddles; automatic collimation smart paddle system; Fast paddles; automatic collimation paddle system; Fast paddles; automatic collimation
depth: 470 mm
Acquisition Workstation
Workstation HP Z420 Workstation Quad core Intel i5 workstation Dual core Kontron workstation Three: AWS 8000, AWS 8000 Mobile and AWS 5000 Three: AWS 8000, AWS 8000 Mobile and AWS 5000 Three: AWS 8000, AWS 8000 Mobile and AWS 5000 MultiCore CPU, 24 GB RAM, Windows 7 embedded 64 bit OS, Intel Core i5 or better HP Z420 workstation HP Z420 workstation N/S
1 HDD of 1 Tbyte
Monitor size, inches 21.3 in. 2 types of display available: 1 MP LCD monitor; 48 cm (19-in.) medical grade; 1,280 x 1,024 2 types of display available: 1 MP LCD monitor; 48 cm (19-in.) medical grade; 1,280 x 1,024 Control monitor: 17 in. color, touch screen on some models; Control monitor: 17 in. color, touch screen on some models; Control monitor: 17 in. color, touch screen on some models; image 21.3 in. Barco TFT LED 2 Mpixel - ; - very high brightness Standard monitor 1,280 x 1,024 DICOM monitor 1,280 x 1,024; 3 21 in. 21 in. 48 cm
pixels (landscape); high luminance - up to 300 Cd/m2; contrast ratio: 2,000:1; viewing angle: pixels (landscape); high luminance - up to 300 Cd/m2; contrast ratio: 2,000:1; viewing image monitor: three: 21 in. color 2 MP, 21 in. gray 2 MP, 21 in. image monitor: three: 21 in. color 2 MP, 21 in. gray 2 MP, 21 in. monitor: three: 21 in. color 2 MP, 21 in. gray 2 MP, 21 in. gray 3 MP (250cd/m2) and resolution for medical use; or 21.3 in. Barco MP 2,048 x 1,536
170 degrees; mounted on a rotating arm for in-room access; 3 MP monitor display: high angle: 170 degrees; mounted on a rotating arm for in-room access; 3 MP monitor display: gray 3 MP gray 3 MP TFT LED diagnostic color 3 Mpixel very high brightness (400
performance color IPS 3 MP monitor; 54 cm (21.2-in.); 2,048 x 1,536 pixels (landscape); high performance color IPS 3 MP monitor; 54 cm (21.2-in.); 2,048 x 1,536 pixels cd/m2)
brightness: 1,000 Cd/m2; contrast ratio: 1,400:1; viewing angle: 170 degrees; mounted on a (landscape); brightness: 1,000 Cd/m2; contrast ratio: 1,400:1; viewing angle: 170 degrees;
rotating arm for in-room access mounted on a rotating arm for in-room access
Hard disk capacity Disk 1: 330 GB for 2-D images 1 internal 250 GB disk for the system, 1TB for image storage 1 internal 500 GB, 7,200 RPM At least 1 TB At least 1 TB At least 1 TB 70,000 images on HDD Up to 10,000 image acquisitions 1 TB 1 TB 24,7000 images on hard disk
(500 GB)
Memory 8 GB 32 GB 4 GB At least 6 GB At least 6 GB At least 6 GB 1 TB N/S 8 GB 8 GB 500 GB
Networking capabilities Printer, DMS network 4 Gigabit Ethernet port; DVI display and port connector; DICOM 3.0 platform: query/retrieve 2 Gigabit Ethernet port; DVI display; DICOM 3.0 platform: modality worklist user; modality 10, 100, 1,000 base-T ethernet 10, 100, 1,000 base-T ethernet 10, 100, 1,000 base-T ethernet Storage of MG, DX, CR, MR, US, CT, NM, DBT, SC as SCU/SCP, Gigabit copper network category 5 cable; 100 Mbps copper Intel CT x1 PCIe NIC Intel CT x1 PCIe NIC Yes
modality worklist user; modality performed procedure steps; storage provider and user; performed procedure steps; storage provider; storage commitment user; query/retrieve storage commitment/SCU, query retrieve/SCU/SCP, basic network RJ45 RJ45 connectors no network connection support
storage commitment user; query/retrieve user; basic grayscale print user; verification provider; user; basic grayscale print user; verification provider; DICOM-compliant CD-RW data print/SCU, modality worklist/SCU, MPPS/SCU verification/ for external disk with specification: USB 2.0 interface, hard disk
DICOM-compliant CD, DVD-R/-RW and USB data interchange interchange SCU/SCP, DICOM-compliant DVD/CD (multisession) based, no separate power supply, minimum storage capacity 160
GB
Gray levels 1,024 gray levels Acquisition dynamic range 14 bits Acquisition dynamic range 14 bits Acquire 14 bits, transmit 12 bits, display 8 bits Acquire 14 bits, transmit 12 bits, display 8 bits Acquire 14 bits, transmit 12 bits, display 8 bits 14 bit N/S 1,024 grey levels 1,024 grey levels 14
DICOM 3.0 compliant Yes N/S N/S Yes Yes Yes Yes Yes Yes Yes Yes
Additional features Intelligent AEC (iAEC) factors breast density and implant Connectivity features: customizable autopush to multiple DICOM databases, autoprint, Image preview time <4 sec; can preview 2-D, 3-D projections, Image preview time <4 sec; can preview 2-D, 3-D projections, Image preview time <4 sec; can preview 2-D, 3-D projections, 3-D N/S N/S N/S N/S N/S
recognition to calculate optimal dose for each exam; ISC autodelete based on storage commitment, connectivity to GE service for remote diagnostic 3-D slices, generated 2-D, contrast 2-D images at AWS 3-D slices, generated 2-D, contrast 2-D images at AWS slices, generated 2-D, contrast 2-D images at AWS
image processing optimizes contrast and IQ; 50 micron image capability
output display; automatic system start up and calibration for
efficient workflow and extended system perfomance
Review Workstation
Workstation Dell 1,920 x 1,200 wide-screen monitor SenoIris Diagnose SenoIris Diagnose SecurView DX SecurView DX SecurView DX Raffaello IntelliSpace Breast Solution HP Z420 workstation HP Z420 workstation N/S
Monitor size, inches 2 x Barco 5 MP (21.3 in. diagonal) Dual 5 MP 21.3-in. LCD display; single 12 MP 33.6-in. color LCD display; 19-in. TFT color Dual 5 MP 21.3-in. LCD display; single 12 MP 33.6-in. color LCD display; 19-in. TFT color 21.3 in. (5 MP) or 30 in.(10 MP) 21.3 in. (5 MP) or 30 in.(10 MP) 21.3 in. (5 MP) or 30 in.(10 MP) 2 x 5 MPixel Barco Coronis monitors; 21.3 in. self-calibrating, 546.4 x 185.4 x 365 493.9 mm (21.51 x 7.30 x 14.37 19.44) 2 x 21 in. 2 x 21 in. N/S
monitor monitor mono-chrome monitors portrait format with very high
resolution and brightness dedicated to digital mammography;
Barco QAWEB for calibration and quality control
Hard disk capacity 256 GB 1 x 256 GB, 2 x 1 TB SSD (additional 1 TB optional) 1 x 256 GB, 2 x 1 TB SSD (additional 1 TB optional) At least 1.7 TB, raid 5 At least 1.7 TB, raid 5 At least 1.7 TB, raid 5 1 x 1 TB (raid configuration) N/S 1 TB 1 TB N/S
Memory 16 GB 32 GB RAM 32 GB RAM At least 6 GB At least 6 GB At least 6 GB 16 GB RAM 8 GB (2 x 4 GB), 1,333 Mhz, RDIMM, low volt, dual rank, x8 8 GB 8 GB N/S
Networking capabilities 10/100/1,000 ethernet 10/100/1,000 base T Ethernet 10/100/1,000 base T Ethernet 10, 100, 1,000 base-T ethernet 10, 100, 1,000 base-T ethernet 10, 100, 1,000 base-T ethernet 100/1,000 Mbits/s N/S Intel CT x1 PCIe NIC Intel CT x1 PCIe NIC N/S
Archive/storage N/S Enterprise archive Enterprise archive SecurXchange archive SecurXchange archive SecurXchange archive N/S N/S N/S N/S N/S
Type Images stored in PACS Departmental PACS Departmental PACS Departmental PACS Departmental PACS Departmental PACS CPU with processor Intel Core i5-2500 3.30 GHz N/S N/A N/A N/S
Capacity N/A N/S N/S At least 3 TB At least 3 TB At least 3 TB 2 x 1 TB (raid configuration) N/S N/A N/A N/S
Gray levels 1,024 gray levels N/S N/S 12 bits 12 bits 12 bits N/S N/S 1,024 gray levels 1,024 gray levels N/S
DICOM 3.0 compliant Yes N/S N/S Yes Yes Yes Yes Yes Yes Yes N/S
Additional features N/S Meetings requirements of IHE radiology mammography imaging integration profile-imagine Meetings requirements of IHE radiology mammography imaging integration profile-imagine Custom workflow keypad; application synchronization; Custom workflow keypad; application synchronization; Custom workflow keypad; application synchronization; N/S Multi-modality, vendor neutral DigiPad DigiPad N/S
display actor; capable of accepting DICOM breast tomosynthesis objects; explore DBT images display actor; capable of accepting DICOM breast tomosynthesis objects; explore DBT multimodality viewing; barcode scanner; workstation cluster multimodality viewing; barcode scanner; workstation cluster multimodality viewing; barcode scanner; workstation cluster
quickly and easily with the V-Preview navigation tools (Navigate, Inspector, Bookmark) images quickly and easily with the V-Preview navigation tools (Navigate, Inspector, manager manager manager
Bookmark)
X-Ray Tube
Anode type Rotating Dual track anode: molybenum (Mo) enriched with vanadium and rhodium (Rh) Dual track anode: molybenum (Mo) enriched with vanadium and rhodium (Rh) Rotating tungsten Rotating tungsten Tungsten (W) doped with rhenium (Re)/dual anode angle Tungsten Rotating bi-angular anode Rotating bi-angular anode Dual
Heat capacity, HU 300,000 HU Anode heat storage capacity: 250 kJ (340 kHU) Anode heat storage capacity: 250 kJ (340 kHU) 300,000 (222 kJ) 300,000 (222 kJ) 300000 445kJ (600 kHU) Tube housing 700 kHU Tube housing 700 kHU 1 800 000j 430 000 HU
Heat dissipation rate, 60,000 HU/min 500 W (40 kHU/min) N/S 60000 60000 60000 0.7 kW (maximum workflow) Anode: 56 kHU/min Anode: 56 kHU/min N/S
HU/min.
Target/filter W/RH Mo/Mo and Rh/Ag N/S 2-D: W Rh, W Ag; 3-D: W Al 2-D: W Rh, W Ag; 3-D: W Al W/Rh; W/Ag 0.40 mm (0.016-in.) Al Mo/Mo or RH; W/RH or Ag Mo/Mo or RH; W/RH or Ag MO anode: 2mAs to 500 mAs
combinations mode/5 mAs - 500 in AEC mode,
W anode: 2 mAs -630 mAs
mode/5 mAs to 630 mAs in AEC
mode
Focal spot size, mm 0.1 and 0.3 0.1 and 0.3 IEC on each target 0.1 and 0.3 IEC on each target 0.1 and 0.3 0.1 and 0.3 0.1/0.3 0.3 mm (0.012-in.) according to IEC 60,336 0.3 mm/0.1 mm 0.3 mm/0.1 mm Molybdeum: 0.1/0.3; tungsten:
0.1 /0.3
Positioning Assembly
Collimation N/S Filters: molybenum 0.030 mm; silver 0.030 mm Filters: molybenum 0.030 mm; rhodium 0.025 mm Automatic or user-selectable Automatic or user-selectable Automatic/manual X-ray field is fixed N/S
18 x 24 cm Yes Yes Yes Yes Yes Automatic N/S Yes Yes Yes
24 x 30 cm Yes Yes Yes Yes (24 x 29 cm) Yes (24 x 29 cm) Automatic N/S Yes No Yes
Movement locks Electromagnetic Gantry locked when compression force applied Gantry locked when compression force applied Fail-safe electromagnetic locks Fail-safe electromagnetic locks Yes All movements are motorized No, motorized movements No, motorized movements N/S
Assembly movement N/S N/S N/S Yes Yes Yes The gantry movement allow both face-to-face patient N/S N/S
positioning and conventional (technologist at the side of the
patient); the 3-D movement is useful in case of disabled
patient
Rotation, degrees negative 190 to positive 190 degrees + 180 to - 180 degrees + 185 to - 165 degrees 2-D: 195 to -155; biopsy and 3-D: 180 to -140 2-D: 195 to -155; biopsy and 3-D: 180 to -140 2-D: 195 to -155; biopsy and 3-D: 180 to -140 180 degrees to -180 degrees - isocentric programmable -100 to + 180, isocentric Motorized, -135 degrees to +180 degrees Motorized, -135 degrees to +180 degrees Plus 180 to minus 180
Vertical, cm (inches) 69 to 150 (27.2 to 59.1) Floor to image receptor distance: from 65 cm to 150 cm Floor to image receptor distance: from 65 cm to 150 cm 71-142 (27.8-56) 71-142 (27.8-56) 75 cm range (65 cm-140 cm) 76-138 Motorized, 800 mm to 1,400 mm Motorized, 800 mm to 1,400 mm N/S
SID, cm 65 66 cm 66 cm 70 70 65 cm (25.6 in.) 66 650 mm 650 mm 65 cm
Scale guide Yes, newtons (N) for force and millimeters (mm) for thickness Yes Yes Yes, two digital displays Yes, two digital displays Yes, two digital displays Yes N/A N/A N/A N/S
Handswitch Yes, also dual foot switches Dual foot-pedals for column height and compression adjustments Dual foot-pedals for column height and compression adjustments Yes, on both sides of the device Yes, on both sides of the device Yes N/S No No N/S
Radiation Output
mR/sec. at 28 kVp 540 mR/sec with W/Rh tube using 50 micron filter N/S N/S >230 >230 >230 21uGy/mAs: 3.05uG/s at 30 kV N/S 2,100 mR/s 2,100 mR/s N/S
Radiation Shield
L x W, cm (inches) 108 x 60.4cm (42.5 x 24 in.) Choice between two radiation shields: integrated to the control console (height: 190 cm or 220 Choice between two radiation shields: integrated to the control console (height: 190 cm or 203 x 86 (80 x 34) 203 x 86 (80 x 34) 60 x 193 cm 732 x 28 74 x 30.5 in. 74 x 30.5 in. N/S
cm); standalone 220 cm); standalone
Thickness 0.5 mm Pb equivalent; designed to meet the relevant N/S N/S 6 mm tempered glass, 0.5 mm leaded glass equivalent 6 mm tempered glass, 0.5 mm leaded glass equivalent 6 mm tempered glass, 0.5 mm leaded glass equivalent 0.5 mm Pb 0.5 mm 0.5 mmPb led equivalent 0.5 mmPb led equivalent N/S
requirements of 21 CFR 1,020 and IEC60601-2-45:2001
Compression system Auto compression: 0N to 200 N; manual compression: 0N to Motorized/manual Motorized/manual Yes, with lightweight, flexing paddles, automatically shifting Yes, with lightweight, flexing paddles, automatically shifting Yes, with lightweight, flexing paddles, automatically shifting paddle SCS-sensitive compression system-double power driven Motorized compression Motorized Motorized 3kg to 20 kg automatic and
200 N paddle positions and auto-sensing paddle identification paddle positions and auto-sensing paddle identification positions and auto-sensing paddle identification speed: fast up to the breast and very slow when compression manual adjustments
starts; foot-pedal control
Force, Newtons 200 N Motorized up to 20 daN; manual up to 27 daN Motorized up to 20 daN; manual up to 30 daN 178 maximum, motorized; 300 maximum, manual 178 maximum, motorized; 300 maximum, manual 178 maximum, motorized; 300 maximum, manual Up to 200 N 200 maximum Adjustable maximum: 20 - 25 daN Adjustable maximum: 20 - 25 daN Neutons
Grid ratio 6 to 1 Grid ratio: 11:1 in both 2-D and DBT acquisitions 2-D ratio 5:1 frequency 36l/cm - 3-D grid for DBT acquisition 4:1; exclusive HTC grid 4:1; exclusive HTC grid 4:1 grid 31 lp/mm-low absorption N/A r5 r5 0.209027778
Bucky N/A 24 x 29 cm 24 x 31 cm Sealed; auto-retracting HTC grid Sealed; auto-retracting HTC grid Yes N/A Linear grid, motorized Linear grid, motorized No Bucky, FFDM unit
Magnification device Yes, 1.8x 1.5 and 1.8 (2-D) 1.5 and 1.8 (2-D) 2-D: carbon fiber platform for 1.5x or 1.8x 2-D: carbon fiber platform for 1.5x or 1.8x 2-D: carbon fiber platform for 1.5x or 1.8x with fiagnostic imaging 1.8 Diagnostic scan Mag factors 1.6, 1.8 Mag factors 1.6, 1.8, 2.0 Yes
kit option
Stereotactic device Available N/S Yes (2D) Optional stereotactic biopsy guidance system with Optional stereotactic biopsy guidance system with Optional stereotactic biopsy guidance system with tomosynthesis Yes using same digital detector area N/A DigiGuide DigiGuide Yes
tomosynthesis guidance option tomosynthesis guidance option guidance option
Film ID system N/A Digital detector, no film Digital detector, no film Included in the print function Included in the print function N/S N/A N/A N/A DICOM header
Label printer N/A N/S N/S N/A N/A N/S N/A N/A N/A No
Power requirements 200 V AC/208V AC/220 V AC (single pPhase) Input frequency: 50 Hz/60 Hz; input voltage: single-phase 200-240 V~; EATON UPS 5P650 650 Input frequency: 50Hz/60Hz; input voltage: single phase 200/208/220/240 V~, APC smart AWS: 100 120 200 220 230 240 VAC; gantry: 200 209 220 230 AWS: 100 120 200 220 230 240 VAC; gantry: 200 209 220 230 AWS: 100 120 200 220 230 240 VAC; gantry: 200 209 220 230 240 8 kW output-240/220/208/200 VAC None 208 VAC 50/60Hz - 240 VAC 208 VAC 50/60Hz - 240 VAC 208 V, 230 V, 227 V, +/-10%
VA UPS 750 VA 240 VAC nominal 10%, 5060 Hz 5% permanently wired; single- 240 VAC nominal 10%, 5060 Hz 5% permanently wired; single- VAC nominal 10%, 5060 Hz 5% permanently wired; single-phase single phase, 208 V, 230 V 240V
phase gantry phase gantry gantry +/-10% two phase, 50/50 Hz+Hz
H x W x D, cm (inches) Gantry: 24.5 x 50 x 90 (62.4 x 127 x 228.5) Gantry: H 202 cm (80 in.) x W 72 cm (28 in.) x D 131 cm (52 in.); acquisition station: H 188 cm Gantry: H 193 cm (76in.) x W 70 cm (28 in.) x D 127 cm (50 in.); acquisition station: H 156 Gantry: 223 (max.) x 66 x 138 (87.8 x 26 x 54.25); acquisition Gantry: 223 (max.) x 66 x 138 (87.8 x 26 x 54.25); acquisition Gantry: 223 (max.) x 66 x 138 (87.8 x 26 x 54.25); acquisition 144 x 93.5 x 215 73.2 x 28 x 05 40.4 x 39.1 x 29.7 in minimum 40.4 x 39.1 x 29.7 in minimum N/S
(74 in.) x W 77 cm (30 in.) x D 37 cm (15 in.) cm (61 in.) x W 71 cm (28 in.) x D 40 cm (16 in.); generator H 133 cm (52 in.) x W 64 cm station: 207 x 92.7 x 58.5 (81.5 x 36.5 x 23) station: 207 x 92.7 x 58.5 (81.5 x 36.5 x 23) station: 207 x 92.7 x 58.5 (81.5 x 36.5 x 23)
(25 in.) x D 44 cm (17 in.)
Weight, kg (lb) 370 Gantry: 420 kg (926 lb); control station without monitors: 160 kg (353 lb) Acquisition station: 217 kg (478 lb); gantry: 420 kg (926 lb); generator: 160kg (353 lb) Gantry: 400 (882); AWS: 154 (340) Gantry: 400 (882); AWS: 154 (340) Gantry: 400 (882); AWS: 154 (340) 330 65 lbs 421 lbs 421 lbs Examinination stand and
generator: 770 lbs; control desk
with radiation sheild: 275 lbs
Optional accessories Biopsy paddle, rectangle spot compression paddle, flex Additional 24 x 29 cm paddle; additional 19 x 23 cm sliding paddle; 24 x 29 cm flexible Options: 24 x 31 cm ergonomic paddle that conforms to the breast, 19 x 23 cm ergonomic Affirm biopsy guidance system; tomo guided biopsy software; Affirm biopsy guidance system; tomo guided biopsy software; Diagnostic imaging kit; Affirm biopsy guidance system; tomo Stereotactic biopsy systems-Giotto Image 3D is pre-arranged Accessory shelf N/S N/S N/S
paddles, stereotactice biopsy system compression paddle; 19 x 23 cm flexible and sliding compression paddle; 10 x 23 sliding sliding paddle that conforms to the breast, 1.5 and 1.8 magnification stands with C-view generated 2-D image software; I-View contrast enhanced C-view generated 2-D image software; I-View contrast enhanced guided biopsy software; C-view generated 2-D image software; I- for full field digital biopsy both in upright and prone position;
implant/small breast compression paddle; square spot sliding compression paddle; round spot dedicated paddles (19 x 23 cm, round spot, square spot), square spot sliding compression 2-D software; SecurView DX diagnostic workstation; SecurView 2-D software; SecurView DX diagnostic workstation; SecurView View contrast enhanced 2-D software; SecurView DX diagnostic the system can be transformed into a stereotactic prone
sliding paddle; 2-D localization 19 x 23 Swiss Cheese sliding compression paddle; 2-D paddle, round spot sliding compression paddle, 2-D localizer with standard and round spot RT workstation; workstation cluster manager; advanced RT workstation; workstation cluster manager; advanced workstation; SecurView RT workstation; workstation cluster biopsy system using same detector of the mammography unit
localization 19 x 23 sliding standard compression paddle; 2-D crosshair device; X-ray biopsy paddles, bar code reader, mobile mounting kit, stereotaxy add-on (2-D), senoIris workflow manager; extra paddles; needle localization kit; CAD workflow manager; extra paddles; needle localization kit; CAD manager; advanced workflow manager; extra paddles; needle and adding the MammoBed table
protective shield; bar code reader; senoIris review station providing technologist easy access Review station providing technologist easy access to prior images and reports in the exam and Quantra Breast Density Analysis software and Quantra Breast Density Analysis software localization kit; CAD and Quantra breast density analysis software
to prior images and reports in the exam room room
Warranty 1 year 1 year system warranty standard 1 year system warranty standard 1 year parts and labor; 2 year prorated manufacturers warranty 1 year parts and labor; 2 year prorated manufacturers warranty on N/S 1 year 1 year 1 year N/S
on X-ray tube X-ray tube