PH MV White Paper
PH MV White Paper
PH MV White Paper
The millivolt reading from a theoretically perfect pH sensor in a pH 7 solution is zero. In nominal
pH 4 and pH 10 buffers at 25 °C, millivolt readings are 177.48 and -177.48, respectively. In
practice, millivolt readings will differ in an amount referred to as the asymmetry potential of the
electrode. As the electrodes age and degrade with use, mV readings during calibration will
change – decreasing in acidic solutions and increasing in basic solutions.
Changes in mV readings observed during calibration procedures over time can be used as an
indicator of electrode condition and the quality of the calibration procedure. Small deviations
from theoretical mV readings or historic readings are not a concern because they are corrected
during the calibration process and are generally stable between calibration events. Additional
confidence in the calibration can be built by performing a linearity check. This is done by
placing the sensor in a third reference standard (different from the two used in calibration) and
comparing the reading to the expected reading.
Reading mV directly could allow the user to derive a unique mathematical equation relating
voltage to pH that is different from the one used in the pH meter. Deriving a unique equation
has little purpose for most water monitoring plans.
Follow best practices for pH calibration procedures, considering the goals and
constraints of a water quality monitoring program
Record the mV reading during calibrations in a calibration log or laboratory notebook
Watch for sudden changes from prior calibrations to indicate there was an error or
change in the calibration procedure, incomplete maintenance, or sensor damage
Use large deviations that develop with time along with slow response times or erratic
readings as an indicator that measuring electrode or reference electrode needs to be
regenerated or replaced.
Published guidelines on acceptable mV readings or slope percents vary. Rather than relying on
absolute values of mV or slope to determine sensor quality, it is best to incorporate the data to a
comprehensive QA/QC program where it can be used to assess overall sensor, calibration, and
data quality.
References