HT 50 Operating Manual
HT 50 Operating Manual
HT 50 Operating Manual
Operating Manual
OPRHT50NA Rev. B
September 2008
OPRHT50NA B0908
TABLE OF CONTENTS
Section 2 . . . . . . . . . . SPECIFICATIONS
• Intended Use
• Symbols / Labeling Table
• Controls / Alarms / Monitors
• Hardware Requirements
• Miscellaneous Specifications
• Humidifier Specifications
• Air / Oxygen Entrainment Mixer Specifications
• Oxygen Blending Bag Kit Specifications
OPRHT50NA B0908
Section 5 . . . . . . . . . . VENTILATION SET UP & USE
• Introduction
• Assembling the Ventilator
• Ventilator Set Up Procedure with
Exhalation Valve Calibration
• Patient Set Up Procedure
• Built-in Humidifier (model HT50-H, HT50-HB)
• Air / Oxygen Entrainment Mixer (Optional Accessory)
• Oxygen Blending Bag Kit (Optional Accessory)
Section 7 . . . . . . . . . . TROUBLESHOOTING
• Introduction
• Table 7-1 Alarms
• Table 7-2 Built-in Humidifier
• Table 7-3 General / Clinical
• Table 7-4 Setting Limitation Messages
ii OPRHT50NA B0908
1. OPERATOR’S RESPONSIBILITY
OPRHT50NA B0908
SECTION 1
When the HT50 is used in home care and sub acute environments it
is important that the primary caregiver has received training and has
demonstrated competency in all equipment functions. A specific
written care plan must be established by the attending physician.
LIMITATION OF LIABILITY
The liability of Newport Medical Instruments, Inc. (NEWPORT) is
subject to and limited to the exclusive terms and conditions as set
forth herein. Said liability is limited whether arising out of, or related
to, the manufacture and sale of goods, their installation,
demonstration, sales representation, use, performance, or
otherwise. Any liability based upon product warranty, whether
breach of warranty or otherwise, is limited regardless of any fault
attributable to NEWPORT and the nature of the action (including
breach of warranty, negligence, and strict liability).
WARRANTY
The Newport HT50 Ventilator is guaranteed to be free of defects
for a period of two (2) years from date of delivery. The following are
exceptions to this warranty:
DEFINITIONS
WARNING Possibility of personal injury, to patient or others, if
disregarded.
TYPING CONVENTIONS
Within the text of this manual, controls, alarms, and indicators are
designated by the labeling name as they appear on the ventilator,
e.g.: P support (pressure support), P trig (pressure trigger), and
SPONT (spontaneous mode).
General Warnings
External power connection: To maintain grounding integrity when
using A.C. power, only connect to hospital grade receptacles.
Always disconnect the external power supply prior to servicing.
Always use the A.C. power cord supplied with the HT50. Make
certain the power cord ferrite is always attached to the A.C. power
cord to ensure that the HT50 meets EMC requirements.
Ensure that the oxygen source is not empty before and during the
use of the optional Air/Oxygen Entrainment Mixer or Oxygen
Blending Bag Kit.
When the HT50 is used for transport applications, ensure that the
internal battery system is fully charged prior to use.
Always ensure that the green Ext. Power LED lights after
connecting the HT50 to an external AC or DC power source (it can
take up to two minutes to light). If the LED does not light, check all
power connections and resolve any problems.
Always plug the HT50 into an external power supply source when
OPRHT50NA B0908 1-5
OPERATOR’S RESPONSIBILITY
The flow resistance of the air inlet filter, located on the right side of
ventilator, is likely to increase with repeated use. Ensure that the
filter is changed regularly.
This equipment has been tested and found to comply with the
EMC limits for the Medical Device Directive 93/42/EEC (EN55011
Class A and EN 60601-1-2). These limits are designed to provide
reasonable protection against harmful interference in a typical
medical installation. The equipment generates, uses and can
radiate radio frequency energy and, if not installed and used in
accordance with these instructions, may cause harmful
interference to other devices in the vicinity. However, there is no
guarantee that interference will not occur in a particular installation.
If this equipment does cause harmful interference with other
devices, which can be determined by turning the equipment off
and on, the user is encouraged to try to correct the interference by
one or more of the following measures:
Cautions
Only use medical grade oxygen with the Air/Oxygen Entrainment
Mixer or Oxygen Blending Bag Kit.
Contact Information
OPRHT50NA B0908
SECTION 2
INTENDED USE
This device is intended to provide continuous or intermittent mechanical
ventilator support for the care of individuals who require mechanical
ventilation. The ventilator is a restricted medical device intended for use
by qualified, trained personnel under the direction of a physician.
Specifically, the HT50 is applicable for adult and pediatric (i.e. infant, child
and adolescent) patients, greater than or equal to 10 kg or 22 lbs., who
require the following general types of ventilatory support, as prescribed by
an attending physician: positive pressure ventilation with assist/control,
SIMV and SPONT modes of ventilation. The HT50 is suitable for use in
hospital, sub-acute, emergency room, home care environments as well as
for transport and emergency response applications.
|
Main Power On Alarm Setting
▲ ▲
Equipotentiality High Alarm Set High Alarm
▼ ▼
Refer to Operating Manual Low Alarm Set Low Alarm
SYMBOLS/LABELING TABLE
Controls/Alarms/Monitors Range/Selection
1. MODE (Pressure or A/CMV
Volume Control) SIMV
SPONT
6. ƒ (Frequency) 1 to 99 b/min
15. Alarm Silence/Reset Button Pressing button silences an audible alarm violation for 60
& Indicator seconds and resets a latched alarm indicator. LED lights to
indicate that Silence is active.
16. ALARMS Indicators Indicators for violated alarms blink red. When the alarm is no
longer violated, the indicator latches (stays lit). Cancel a latched
indicator by pressing the Silence/Reset button.
17. Int. Battery Button & Indicator Pressing button displays the internal battery charge level in the
airway pressure meter (Paw) window. Use only when operating
on the internal battery system for accurate reading. LED lights
to indicate internal battery system operation and alarms.
Controls/Alarms/Monitors Range/Selection
18. FIO2 0.21 to 1.00
(with optional accessories)
19. On / Standby Button Press once to put in Setting condition. (On-Setting/LED off)
Press again to begin ventilating (On-Ventilating/LED on).
When the HT50 is ventilating, press two times to put ventilator
into Standby/Off condition (LED off).
20. Push To Unlock Buttons & Pressing button unlocks front panel buttons if locked by
Indicator automatic panel lockfeature. Auto lock is enabled/disabled in User
Set Up. LED lights to indicate panel is locked.
21. Alarms
▲Paw (High Pressure) 4 to 99 cmH2O / 4 to 99 mbar, must be 1< Low Paw
▼Paw (Low Pressure) 3 to 98 cmH2O / 3 to 98 mbar, limited by ≥ PEEP + 3 and High
Paw -1
Low Baseline Pressure Paw < PEEP - 3 cmH2O/mbar for 3 sec during exhalation
High Baseline Pressure Paw > PEEP + 8 cmH2O/mbar at onset of a breath or 3 sec
after the start of exhalation
Occlusion Paw > PEEP + 15 cmH2O/mbar at onset of a breath or 3 sec
after start of expiration
Apnea 30 sec ± 3 sec
PCV Not Reached Paw P < 50% of PCV setting
▲VI Insp. Min. Volume
•
1.1 to 50.0 L/min
▼VI Insp. Min. Volume
•
0.1 to 49.0 L/min
Check Prox Line Prox Paw does not match machine Paw during inspiration
Humidifier (5 messages) Humidifier malfunction/disconnection
Power Switchover External power to internal battery switchover alert
Battery Low Minimum of 30 minutes battery time remains until shutdown
Battery Empty Minimum of 15 minutes battery time remains until shutdown
NOTE: The time between the Battery Low Alarm violation and the Battery Empty Alarm violation
will vary depending on the ventilator load. At high volumes and pressures, the Battery Empty
Alarm will occur much sooner after the Battery Low Alarm, than it will at lower volumes and
pressures. In all cases, the stated minimum times for each alarm will be met, even if the two
alarms occur almost simultaneously.
Device Alert (5 messages) Ventilator malfunction: FAULT BAT SYS, OCCLUSION, 10V
SHUTDOWN, SYSTEM ERROR or MOTOR FAULT
Controls
Alarms
▲Paw (High Pressure) Red LED indicates high peak airway pressure, high baseline
pressure, or occlusion alarm violation.
▼Paw (Low Pressure) / Apnea Red LED indicates low peak airway pressure, low baseline
pressure, apnea, or PCV (50% of PCV setting not achieved)
alarm violation.
Device Alert Red LED indicates ventilator malfunction alert.
▲VI (High Insp. Min. Volume)
•
Red LED indicates high inspiratory minute volume alarm limit is
violated.
▼VI (Low Insp. Min. Volume)
•
Red LED indicates low inspiratory minute volume alarm limit
(Back-Up Vent) is violated.
Misc. Indicators
Silence / Reset Yellow LED indicates that the audible alarm is silenced for 60
seconds.
Auto Lock On Green LED indicates that the panel is currently locked.
On / Standby Green LED indicates that the HT50 is ventilating.
P• trig Green LED blinks on to indicate patient breathing effort.
•
V (Flow) Green LED indicates that Flow is displayed in the V / I:E Ratio
numeric window display. •
I:E Green LED indicates that the I:E Ratio is displayed in the V / I:E
Ratio numeric window display. Blinking LED indicates a breath
with an inverse I:E Ratio.
Ext. Power / Green LED indicates external power is on and the internal
Charging Int. Battery battery is being charged. Red LED indicates power switchover
to internal battery.
Int. Battery (Push to Test) Yellow LED indicates internal battery is in use. LED blinks
yellow to indicate Battery Low alarm condition or blinks red to
indicate Battery Empty alarm condition.
Humidifier On Green LED indicates humidifier is active. LED blinks yellow to
indicate humidifier alarm condition.
Hardware Requirements
24. Electrical Applied parts type BF
27. RS-232C Interface /Remote 8 pin SEMCONN connector. Operates at 19,200 baud. Allows
Alarm Output put for interfacing with central alarms systems.
Miscellaneous Description
33. Regulatory and Agency Complies with the following international standards & requirements:
Standards/Requirements IEC 60601-1:1988 (+A1:1991 +A2:1995; EN 60601-1:1990
+A1:1993 +A2:1995 +A3:1996) Medical Electrical Equipment –
Part 1: General Requirements for Safety
CEI/IEC 60529:2001 Degrees of Protection Provided by
OPRHT50NA B0908 2-5
SPECIFICATIONS
40. Patient Range Adult - Pediatric (i.e. infant, child & adolescent) with
body weight >10 kg
41. Factory Default Parameters
Patient Settings:
Power Save On
Pressure Units cmH2O
Auto Panel Lock Disabled
Set Up User
42. Patient Circuit Reusable 22 mm I.D. adult/pediatric circuit with 3/16 inch
(4.8 mm) I.D. proximal pressure sensing line, 1/8 inch (3.2 mm)
I.D. exhalation valve control drive line, and exhalation valve.
Usable Volume of
Water Bottle: 265 mL
Compliance at Minimum
Water Level (Refill Line): 0.5 mL/cmH2O / mbar @ 23ºC
Compliance at Maximum
Water Level (Full Line): 0.33 mL/cmH2O / mbar @ 23ºC
Maximum Operating
Airway Pressure: 100 cmH2O / 98 mbar
Maximum Temperature at
the Patient Wye That
Triggers an Alarm: 41ºC
Pneumatic Requirements:
Oxygen 35 to 90 psig (2.4 to 6.2 Bar) full operating range
40 to 70 psig (2.7 to 4.8 Bar) accuracy ± .08
Pneumatic Requirements:
Oxygen 0-10 L/min (calibrated)
Air Atmospheric pressure
OPRHT50NA B0908
SECTION 3
Setting Condition Press the A/CMV, SIMV or SPONT button. The LED on the
selected Mode will light green to confirm the selection.
On Condition Press the A/CMV, SIMV or SPONT button. The LED on the
selected Mode will blink green and the Message Display Window
will read “PRESS AGAIN.” Press the button again within 5
seconds to confirm the mode change, or the previously selected
mode will continue.
Adjust the blinking target value for the selected control with the
▲Up / ▼Down buttons.
To Accept the new control and target value you must press the
desired control button (Volume or Pressure) a second time.
OPRHT50NA B0908 3-1
DESCRIPTION OF CONTROLS, INDICATORS, ALARMS & CONNECTIONS
On Condition: SPONT
or
Setting Condition: A/CMV, SIMV, or SPONT
Select—Adjust (▲Up / ▼Down)—Accept
Adjust the blinking target value for the selected control with the
▲Up / ▼Down buttons.
Accept the new control and target value by either pressing the
selected button again; or by pressing another button to select a
new parameter for adjustment, or by waiting 5 seconds without
making a change.
Auto Panel Lock can be enabled or disabled via User Set Up (see
pg 3-27). When the Auto Panel Lock is enabled, the Panel will
lock 30 seconds after the last button is pushed and the LED
lights green. All touch buttons (except Silence/Reset and Internal
Battery Test) are locked, preventing accidental parameter
changes.
Figure 3-1
HT50 Ventilator Front Panel
(model HT50-H1, HT50-H1B)
Figure 3-2
Newport HT50-H, HT50-HB (with built-in humidifier)
On / Standby
During Setting condition, all adjustable LEDs are lit. This allows
the operator to preset and adjust controls prior to ventilation. The
On/Standby indicator is not lit. The Message Display Window
shows “Press ON to Vent,” suggesting that the On/Standby
button needs to be pressed if you want the HT50 to start
ventilation.
MODE Control
The MODE control buttons enable the user to switch between
the following operational modes:
• A/CMV
• SIMV
• SPONT
A/CMV
(Assist / Control Mandatory Ventilation)
In A/CMV, the user may choose to pressure or volume control
mandatory breaths. In either case, all breaths delivered to the patient,
whether time (ventilator initiated) or patient-triggered, are the same.
SIMV
(Synchronized Intermittent Mandatory Ventilation)
In SIMV, the user may choose to pressure or volume control
mandatory breaths. In either case, all mandatory breaths
delivered to the patient, whether time (ventilator initiated) or
patient-triggered, are the same. In addition, the user may choose
to pressure support the spontaneous breaths in between
mandatory breaths.
SPONT
(Spontaneous Ventilation)
ƒ
(frequency)
Range: 1 to 99 b/min
NOTE: In SPONT mode, the tI setting is not utilized but the value
can be preset.
Volume Control
(tidal volume)
If the Volume Control setting causes the flow rate to reach the
maximum or minimum level of the flow specification, adjustment
of Volume Control ceases, a beep sounds, and a setting
limitation message appears in the Message Display Window.
Press the Volume Control button. The set tidal volume is displayed
in the adjacent window if the HT50 is ventilating. A “PRESS
AGAIN” message appears in the Message Display Window.
Pressure Control
(target pressure)
Range: 5 to 60 cmH2O / mbar
Press the Pressure Control button. The set target airway pressure
value is displayed in the adjacent window if the HT50 is
ventilating. A “PRESS AGAIN” message appears in the Message
Display Window.
P trig (sensitivity)
Range: 0 to 30 cmH2O/mbar
P support
(pressure support)
Range: 0 to 60 cmH2O/mbar
Manual Inflation
Range: 0 to 3.0 sec
Humidifier On Button
(HT50-H, HT50-HB only)
Range: 19ºC to 39ºC
I:E Ratio
(inspiratory time to expiratory time)
Test the HT50 dual internal battery periodically to verify that the
charge level is in the blue area. The numbers on the Paw meter do
not reflect the percent of charge.
The Int. Battery Charge Level meter is located beneath the Paw
meter. If the needle is in the red when the test button is presed the
battery charge is low. You should use an external power source.
The blue area indicates medium to full battery charge. Each
battery use time is different based on your conditions. The
numbers on the Paw meter do not reflect the percent of charge.
Press and hold the Push to Unlock button for at least one second
to unlock the panel and enable the activation of all touch buttons
for adjustment. An audible beep sounds and the LED is
extinguished. When 30 seconds have elapsed without pressing
any buttons, the Auto Lock is automatically activated again.
Silence / Reset
NOTE: The Battery Empty Alarm and the Device Alert Alarm can
not be silenced permanantly. These alarms indicate that an
alternate source of ventilation must be utilized. See pgs 3-20 and 3-
21 for more details.
Paw Meter
(airway pressure meter)
The ▲ Paw button allows the selection of the high (peak) airway
pressure alarm setting.
The ▼ Paw button allows the selection of the low peak airway
pressure alarm setting.
To adjust either alarm, press the desired button once. The value
in the adjacent display window will blink. Use the ▲Up/▼Down
controls to adjust the displayed alarm setting value. The new
setting can be retained by (1) pressing the selected button again
to accept the alarm setting, (2) selecting another parameter for
adjustment, or (3) allowing five seconds to elapse without
adjustment.
NOTE: The Low ▼Paw Alarm does not function in SPONT mode.
The ▼Paw alarm limit does not apply to spontaneous breaths in
SIMV mode.
This window displays the inspiratory minute volume (in liters) and
•
is automatically updated every 10 seconds. The VI window always
displays the delivered minute volume, except when the user is in
•
the process of setting either the High or Low VI alarm limit.
button. The value in the adjacent display window will blink. Use
the ▲Up/▼Down controls to adjust the displayed alarm limit
value. The new limit can be retained by (1) pressing the selected
button again to accept the alarm setting, (2) selecting another
parameter for adjustment, or (3) allowing five seconds to elapse
without adjustment.
(user adjustable)
Back-up Ventilation
(user adjustable)
setting.
Back-up Ventilation
Back-up Ventilation is an alarmed function that activates when the
delivered inspiratory minute volume (VI) falls below the Low ▼VI
• •
•
alarm indicator blinks, an audible alarm sounds, and “LOW Vi
(BUV)” is displayed in the Message Display Window. The ventilation
settings employed by Back-up Ventilation are then displayed on the
front panel. Back-up Ventilation ceases when VI = ▼VI + 10%, at
• •
Apnea Alarm
(automatic)
Audible Alarm: Intermittent beep
Visual Alarm: ▼Paw/Apnea indicator blinks red
Message Window: APNEA
Humidifier Alarm
(automatic) HT50-H only
Audible Alarm: Intermittent 3-pulse caution beep
Visual Alarm: Humidifier indicator blinks yellow
The status of the internal battery system and the charging system is
continuously monitored and any deficiency will result in a Battery
Fault Alarm. The HT50 will continue to operate properly from an
external power source but it can not be powered by the internal
battery system.
NOTE: If the “Fault Bat Sys” device alert alarm occurs, keep the
HT50 plugged into an external popwer source. Contact Newport
Medical Technical Service Dept.
NOTE: See Pg. 3-17, Occlusion Alarm, Device, and Pg. 3-21,
Fault Battery System Alarm, for a detailed description of these
recoverable Device Alert Alarms.
There are three possible messages that will be displayed when a
non-recoverable Device Alert Alarm occurs; (1) MOTOR FAULT (2)
10V SHUTDOWN and (3) SYSTEM ERROR. For MOTOR FAULT
and10V SHUTDOWN the full text message will be displayed for 5
seconds, followed by an abbreviated form of the message, eg.
MTR Fail, or 10V Fail. A timing message will also appear,
documenting how long the condition has been present. The
format for the time is “H:MM:SS.” The SYSTEM ERROR
message will always be displayed with full text.
NOTE: If the cause of the SYSTEM ERROR does not allow the
HT50 to display the alarm message and the Device Alert indicator
to light, the ventilator will shut down and the Shut Down Alert
Alarm will activate. Upon the next power up of the unit, the
SYSTEM ERROR message will be displayed. If the cause of the
alarm has been corrected, the message can be cancelled by
pressing the Silence/Reset button.
The Shut Down Alert Alarm occurs when the ventilator is powered
Off. An intermittent audible alert indicates the ventilator is no longer
operating. The intermittent beeps will continue for at least 10
minutes or until it is silenced by pressing the Silence/Reset button.
GAS Supplies gas flow to the patient. Connect the patient breathing
OUTPUT circuit to this outlet.
EXH. VALVE
Controls the exhalation valve. Connect the exhalation valve
control drive line to this outlet.
RS-232C Connector
An 8 pin SEMCONN connector operating at 19,200 baud which
allows the ventilator system to interface with central alarm
systems and remote alarms.
A.C. connections
The HT50 uses a single inlet for both A.C. and D.C. power sources.
The inlet power connector automatically recognizes A.C. voltage
ranges from 100 to 240 and D.C. ranges from 12 to 30. A Newport
approved external battery can be attached to this connector.
Equipotential Connector
Used for electric potential equalization.
Use of the Power Cord Ferrite ensures that the HT50 meets EMC
requirements. Anytime the HT50 is operating on A.C. power, the
ferrite should be attached to the power cord. Operating from
D.C. power does not require the use of the ferrite.
NOTE: Always use the A.C. power cord supplied with the HT50.
OPTIONAL ACCESSORIES
Air/Oxygen Entrainment Mixer
Pneumatic Requirements
Oxygen 35-90 psig (2.4 to 6.2 Bar)
Pneumatic Requirements
Oxygen 0-10 L/min
USER SET UP
The User Set Up allows the operator to select a variety of
functional parameters. User Set Up parameters must be
established before the HT50 is used for ventilation.
The user may scroll through the rest of the set up parameters by
pressing the Int. Battery button. Use the ▲Up/▼Down control
buttons to change the parameter setting. To exit User Set Up, press
the Silence/Reset button once. The HT50 is now ready for use.
Pressure Units
Set Up
This setting allows the user to enable (turn on) or disable (turn
off) the Auto Panel Lock feature. For a complete description of
Auto Panel Lock see pg 3-13. The Auto Panel Lock feature is
factory set at Disabled or off. Use the ▲Up/▼Down buttons to
toggle between Enabled and Disabled.
OPRHT50NA B0908
SECTION 4
The user sets variable alarm settings for High Pressure, Low
Pressure, High Inspiratory Minute Volume and Low Inspiratory
Minute Volume, with Back Up Ventilation provided in all modes in
response to a Low Minute Volume Alarm. There are also built in
alarms for High Baseline, Low Baseline, (Circuit) Occlusion,
Apnea, PCV Not Reached, Check Prox Line, Power Switchover
(from external to internal), Low Battery and Empty Battery and
Device Alerts. The HT50-H model also provides humidifier related
alarms.
The HT50 is very easy to set up and use with clear indications of
all ventilation and alarm settings and alarm violations.
A/CMV MODE
(Assist/Control Mandatory Ventilation)
In A/CMV mode, time activated (mandatory) breaths are delivered
in accordance with the ƒ setting. Patients can trigger mandatory
breaths in addition to, or in place of, time activated (mandatory)
breaths if the effort they generate causes airway pressure to
meet the P trig setting. Every such patient effort results in a
mandatory breath. The breath can be volume or pressure
controlled. PEEP/CPAP may be added. Tidal volume is
determined by the target pressure, tI, and patient respiratory
mechanics in Pressure Control and by the tidal volume setting in
Volume Control.
SPONTANEOUS BREATH
PRESSURE A B C C A A
MANDATORY
100% BREATH
LOCKOUT
INTERVAL
Figure 4-1
Synchronized Intermittent Mandatory Ventilation (SIMV)
SIMV MODE
(Synchronized Intermittent Mandatory Ventilation)
In SIMV mode, patients receive a fixed number of volume or
pressure controlled mandatory breaths (time or patient activated)
and may breathe spontaneously between mandatory breaths,
with or without pressure support (Psupport). See Figure 4-1 for
schematic illustration. PEEP/CPAP may be added.
SPONT MODE
(Spontaneous Ventilation)
P SUPPORT
(Pressure Support)
VOLUME CONTROL
(Volume Control Ventilation)
During Volume Control ventilation, tidal volume can be set for
mandatory breaths. If a volume setting is changed while the
ventilator is operating, the change takes place in increments over
a series of breaths.
The frequency (ƒ) will only increase up to a rate that produces a 1:1
I:E ratio even if the calculated Back-up Ventilation rate is higher.
User Cancelled
Patient Cancelled
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Assembling the Ventilator . . . . . . . . . . . . . . . . . . . 5-1
Ventilator Set Up Procedure with
Exhalation Valve Calibration . . . . . . . . . . . . . . . . . 5-2
Patient Set Up Procedure . . . . . . . . . . . . . . . . . . . 5-8
Built-in Humidifier . . . . . . . . . . . . . . . . . . . . . . . . 5-10
Air/Oxygen Entrainment Mixer
(Optional Accessory) . . . . . . . . . . . . . . . . . . . . 5-13
Oxygen Blending Bag Kit
(Optional Accessory) . . . . . . . . . . . . . . . . . . . . 5-14
OPRHT50NA B0908
SECTION 5
INTRODUCTION
Familiarize yourself with the instructions in this section prior to
ventilating patients with the HT50. Following all of the listed steps
is essential to ensuring the safest possible operation of the
ventilator. Use the information in this section in conjunction with
established hospital protocols and homecare dealer instructions.
Includes:
1 ea. OPRHT50NA Operating Manual
1 ea. V24-00900-60 A.C. Power Cord with Ferrite
1 ea. HT600039 Exhalation Valve
1 ea. HT460300 Air Intake Filter Disposable (pk. of 5 filters)
1 ea. HT6004701 Prox. Inline Filter, Disposable (pk. of 5 filters)
4. Verify that the A.C. power cord supplied with the HT50 has
the Power Cord Ferrite attached between the adapter box
and the ventilator. If utilizing external power, plug the
ventilator’s electric cord into a properly grounded outlet. If
using internal battery, ensure that battery is fully charged.
Power Cord Ferrite
WARNING Always use the A.C. power cord supplied with
the HT50 that has the Power Cord Ferrite attached between
the adapter box and the ventilator to ensure that the HT50
meets EMC requirements.
Exhalation Valve Calibration 6. Calibrate the exhalation valve prior to use as follows:
d. The HT50 will start the EZ Cal and the ventilator will
automatically test the exhalation valve. If it passes the
test, the messages "Cal Completed", then "Press ON to
Vent" will be displayed.
Exhalation Valve
DETAIL A
Gas Output Connection Exhalation Valve
Proximal Line
Connector
DETAIL B
Left Side Panel Connections
Exhalation Drive Line
Figure -1
HT50-H1, HT50-H1B –Set Up Assembly
Humidifier Circuit
Humidifier
Temperature Temperature Probe Port
Probe Electrical on Exhalation Valve
Connector
Figure 5-2
HT50-H, HT50-HB (with built-in humidifier) – Set Up Assembly
Figure 5-3
HT50-H, HT50-HB with
Prox Line Tee’d into
Dry Side of Humidifier
Humidifier Inlet
Humidifier Outlet
Prox Line
Tee Adapter
Figure 5-4
HT50-H1, HT50-H1B with Third-Party Humidifier Set Up and
Prox Line Tee’d into Dry Side
Prox Line
Tee Adapter
Exh Valve Drive Line
Figure 5-5
7. Set Ptrig.
12. Check all alarm limit and control settings to ensure they are
appropriate for the patient to be ventilated.
16. Reassess ▲ High and ▼ Low Paw alarm settings and adjust
to appropriate levels.
17. Verify that the Ptrig indicator blinks each time the patient
initiates a spontaneous inspiratory effort. Readjust Ptrig as
necessary.
18. Reassess ▲ High and ▼ Low VI alarm settings and adjust
•
to appropriate levels.
•
NOTE: Newport strongly recommends that you •
set the High VI
alarm no more than 1 L above and the• Low VI alarm no more
than 1 L below the patient’s average VI in order to ensure the
quickest response to changes in patient and/or breathing circuit
conditions.
BUILT-IN HUMIDIFIER
(HT50-H, HT50-HB only)
NOTE: You will need to use an alternate source of humidification
with HT50 model HT50-H1, HT50-H1B.
General Description
Upper
2. Open the humidifier clamp knob located on the upper case of
Bottle the ventilator, then remove the humidifier bottle from the
ventilator. Open the humidifier bottle, remove the heat sink
and place absorbent paper between the heat sink ribs.
Paper
Position the heat sink inside the lower half of the bottle.
Heat Sink
3. Fill the lower half with sterile, distilled water, up to the FULL
line. Attach the upper half of the bottle to the lower half.
Lower
Bottle
4. Re-attach the humidifier bottle to the ventilator. Secure the
clamp knob.
OXYGEN ACCESSORIES
WARNING Continuous oxygen monitoring is required for
patient safety. The HT50 does not have a built-in alarm
system to notify user of a failure or disconnection of the
oxygen source.
Installation Instructions
1. Unscrew the three thumb screws on the Filter Cover from the
ventilator housing.
Inspect Filter 4. With the oxygen hose facing toward the front of the ventilator,
press the 30 mm O.D. outlet of the Mixer into the Attachment
OPRHT50NA B0908 5-13
VENTILATOR SET UP AND USE
6. Open the supply pressure valve slowly and listen to verify that
there is no hiss, indicative of a leak. Do not use the oxygen
mixer with a leak in the system.
NOTE: Before attaching the Oxygen Blending Bag, make sure that
the three hold-down screws on the Filter Cover are tight. If the
screws are not tight, ambient air may enter the HT50 from around
the inlet cover. This may change the oxygen enrichment level
delivered to the patient when the Oxygen Blending Bag is in use.
50 psig
0 - 10 L/min Regulator
Oxygen Flow Meter
Oxygen Blending
Bag Kit
Oxygen Barb
Adapter
Oxygen
Filter Cover
Oxygen Tubing
Figure 5-6
Oxygen Blending Bag Kit Assembly Diagram
The Oxygen Blending Bag is not a calibrated mixing device. The
level of oxygen enrichment achieved is affected by these
variables: minute volume, oxygen supply flow and the presence
or absence of PEEP.
Installation Instructions
The Oxygen Blending Bag Kit attaches into the Fresh Gas Intake
port on the Filter Cover, located on the right side of the HT50.
3. Close the Filter Cover, ensuring that the three thumb screws
are tight.
Table 5-1
Oxygen Supply Flow for Desired % of Oxygen
Enrichment – Without PEEP
NOTE: When PEEP is added it changes the mixing of oxygen with air. Use
Table 5-1 when the patient is ventilated without PEEP and Table 5-2 in the
presence of PEEP. Data in Table 5-2 are taken at an I:E ratio of 1:2. Different I:E
ratios may slightly affect the Desired % of Oxygen Enrichment when PEEP is in
use.
Table 5-2
Oxygen Supply Flow for Desired % of Oxygen
Enrichment – With PEEP
OPRHT50NA B0908
SECTION 6
HT50 Ventilator Wipe clean between patients and once per week while in use.
The exterior of the ventilator and all parts not in direct contact
with patients should be wiped clean with a cloth dampened with
a medical detergent or alcohol based cleaning solution.
HT50 Accessories All accessories should be thoroughly cleaned, rinsed and air
dried prior to disinfecting/sterilizing.
Oxygen Blending Bag Kit Clean and disinfect between patients and as needed while in
use.
Disassemble: Remove the Oxygen Blending Bag Kit from the HT50 air inlet.
Disconnect the oxygen tubing and slide the rubber bag off its
fitting.
Clean: Wash the outside of the rubber bag (do not immerse the bag or
get the inside of the bag wet) and the plastic parts (without
taking them apart) using a soft brush with water and a mild
detergent. Rinse thoroughly with sterile, distilled water. Shake off
excess water and place all parts on a clean towel to air dry. (Do
not heat or blow dry.)
NOTE: The rubber bag is a single patient use item. Discard the
rubber bag between patients and replace with a new bag.
Soak plastic and metal parts in a solution of one part white vinegar
to two parts of water thirty (30) minutes; rinse thoroughly with
sterile, distilled water; air dry.
Sterilize: The following sterilizing method is suitable for the plastic and
metal parts of the Oxygen Blending Bag Kit:
Humidifier Assembly Clean and disinfect once per week while in use.
(HT50-H, HT50-HB only) Clean and sterilize between patients.
Upper Bottle
Absorbent Paper
Heat Sink
Lower Bottle
Disassemble: Remove the humidifier bottle from the HT50. Open the humidifier
bottle and remove the heat sink. Remove the absorbent paper
from the heat sink and discard. Wash off any residual paper that
remains on the heat sink with running water.
Clean: Wash both halves of the bottle and the heat sink in a warm
detergent solution using a soft brush. Rinse thoroughly with
sterile water.
Shake off excess water and place all parts on a clean towel to air
dry. (Do not heat or blow dry.)
Soak in one part white vinegar to three parts of water and rinse
thoroughly with sterile, distilled water; air dry.
Sterilize: The following sterilizing methods are suitable for the bottle and
heat sink.
Humidifier Temperature Clean and disinfect twice per week while in use.
Probe Clean and sterilize between patients.
Clean: Probe cable: Wipe down with a soft cloth dampened in a mild
detergent.
Probe tips: Wash the probe tips in a mild detergent with a brush.
Rinse thoroughly with sterile, distilled water.
Shake off excess water and place all parts on a clean towel to air
dry. (Do not heat or blow dry.)
Disinfect: Use one of the following disinfecting methods for disinfecting the
probe tips:
Soak the probe tips in one part white vinegar to one part of water
for 2 hours and rinse thoroughly with sterile, distilled water; air dry.
Sterilize: The following method is suitable for sterilizing the entire probe
assembly:
Reusable Breathing Circuits Clean once per week while in use. Always use a clean exhalation
valve (and humidifier/probe assembly if appropriate) when the
breathing circuit is reassembled for patient use.
The breathing circuit includes the tubing, water trap (if used),
proximal pressure line and exhalation valve drive line.
Disassemble: Remove the entire circuit from the ventilator. Remove the
exhalation valve, proximal inline filter (if attached) and
temperature probe (if used) from the circuit. Disassemble the
circuit to expose all surfaces for cleaning.
Clean: Use a low flow of running water or low flow of air to clear tubings
and passages of organic matter. Wash all components of the
breathing circuit with a soft brush. Rinse thoroughly with sterile,
distilled water. Shake off excess water and place all parts on a
clean towel to air dry. (Do not heat or blow dry.)
Reusable Exhalation Valve Clean and sterilize between patients and twice per week
(p/n HT600039) while in use.
Rubber Cap
Proximal Line Fitting
Temperature Probe Port
Assembled Valve: Remove the exhalation valve from the patient breathing circuit.
Remove the rubber cap from the Temperature Probe Port of the
exhalation valve and place it on the Drive Line Fitting Port of the
exhalation valve. See Figure 6-2. You will need to replace the cap
to its original location after cleaning and disinfecting is complete.
Disassembled Valve: To disassemble the current revision exhalation valve, rotate the
top cap counter-clockwise and lift it off. Lift out the valve drive
line fitting and separate it from the diaphragm. See Figure 6-3.
Disinfect: Any of the following disinfecting methods are suitable for the
NEWPORT exhalation valve (p/n HT600039):
Soak in one part white vinegar to one part of water for 2 hours
and rinse thoroughly with sterile, distilled water; air dry.
Place the exhalation valve in warm distilled water, boil the water
and allow to cool, Drain and air dry.
Sterilize: Use the following sterilizing method for the Newport exhalation
valve.
Diaphragm
1. Attach the diaphragm to the white plastic drive line fitting. The
diaphragm must be carefully seated so that it lays flat on the
white plastic drive line fitting and snaps on around the edge
completely.
NOTE: Do not try to turn the drive line fitting after securing the
cap. This may cause the diaphragm to become wrinkled or
unseated and affect ventilator performance. Perform a successful
exhalation valve calibration to verify proper valve assembly.
The air inlet particle filter, located on the right side of the
ventilator behind the Filter Cover, keeps dirt and particles out of
the ventilator’s piston system. As the filter becomes dirty it can
reduce the volume of air drawn into the ventilator.
Check the inlet filter weekly. Replace with a new filter when the
majority of the filter surface area has changed from a clean white
to a dirty brown color. Inlet filters are not reusable.
NOTE: After replacing the filter, make sure that the three hold-
down screws on the Filter Cover are tight. If the screws are not
tight, ambient air may enter the HT50 from around the inlet cover.
The Proximal Inline Filter has a very important function in the HT50.
Located at the Airway Pressure Connector, it protects the internal
pressure transducers from moisture or other contaminants that may
be present in the proximal airway tubing. Since there is no purge
flow coming from the HT50, it is important to always use a Proximal
Inline Filter (p/n HT6004701 or equivalent).
Check the Prox Inline Filter weekly. Discard it and replace with a
new filter if it appears to have gotten wet or come in contact with
a contaminant. Inline filters are not reusable.
In the event that the filter does become occluded, the primary
indication for this would be a Low Pressure Alarm indication with the
message “CHECK PROX LINE.” Should this happen, replace the
filter.
MAINTENANCE
Preventive Maintenance
The Air Inlet Filter (located behind the Filter Cover) must be
checked weekly. Replace when the majority of the filter surface
area has changed from a clean white to a dirty brown color. Air
Inlet Filters are not reusable.
Check the Prox Inline Filter weekly. Replace with a new filter if it
appears to have gotten wet or come in contact with a
contaminant. Inline filters are not reusable.
GENERAL WARNINGS
Preventive maintenance work, repairs, and service may only be
performed by NEWPORT trained or factory-authorized personnel.
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Table 7-1 Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Table 7-2 Built-in Humidifier . . . . . . . . . . . . . . . . . 7-6
Table 7-3 General / Clinical . . . . . . . . . . . . . . . . . . 7-6
Table 7-4 Setting Limitation Messages. . . . . . . . 7-12
OPRHT50NA B0908
SECTION 7
INTRODUCTION
The NEWPORT HT50 Ventilator is used in life-support situations. As
such, it is essential that all individuals using the NEWPORT HT50,
including clinicians and support staff, have a thorough
understanding of its operation. This should include a working
knowledge of the ventilator’s pneumatic and electronic systems.
Or write to:
Apnea Alarm Patient did not trigger a breath for 30 Re-evaluate patient and ventilator settings
▼ Paw/Apnea LED blinks seconds. and provide increased ventilatory support
red, audible alarm sounds as needed.
Patient efforts not detected. Trigger Use Ptrig to adjust trigger level closer to
level set improperly. baseline pressure (0 cmH2O/mbar) so that
patient efforts are detected (indicated by
the Ptrig LED illuminating green).
Battery Empty Alarm Internal battery system charge is Immediately connect HT50 to external A.C.
Int. Battery LED blinks red depleted and ventilator shutdown will or D.C. power.
continuous audible alarm occur in approximately 15 minutes.
sounds, unsilencable
Battery Low Alarm A minimum of 30 minutes of battery Plug power cord into an external power
Int. Battery LED blinks operation remains. source to charge.
yellow
intermittent audible alarm
Check Prox Line Alarm Humidity in proximal line. Change proximal line filter and re route
▼ Paw/Apnea LED blinks proximal line to a connector on the inlet
red, audible alarm sounds (dry) side of the humidifier.
Fault, Battery System Battery system is not functioning Keep HT50 plugged into external power
Device Alert LED red, properly source for proper operation. Call NEWPORT
intermittent audible alarm Technical Service Dept. immediately. See
page 1-6 and 1-7 for contact information.
High Paw Alarm Increased patient resistance or Evaluate patient. Patient may need
▲ Paw LED blinks red, decreased patient compliance. suctioning, aerosol therapy, etc.
audible alarm sounds
Increased breathing circuit resistance. Check for obstructions
(kinked tubes, water in tubing, occluded
filters, etc.)
High Paw alarm set incorrectly. Re-adjust High Paw alarm if appropriate.
Notify physician as necessary.
High Pbase Alarm Airway pressure remains above the Unblock occluded area.
▲ Paw LED blinks red, Low Paw alarm setting at beginning of
audible alarm sounds inspiration. Indicates an occlusion in Re-calibrate exhalation valve.
the circuit/exhalation valve or that the
proximal pressure line or exhalation
drive line is pinched.
High breath rate (insufficient time to Evaluate patient and make necessary
exhale). adjustments to ventilation parameters.
Ventilator autotriggering from leak or Fix leak and re-adjust Ptrig as needed.
improper Ptrig setting.
• •
High VI Alarm Increased spontaneous patient Evaluate patient. Adjust the High VI alarm
•
▲ VI LED blinks red, breathing. setting if needed.
audible alarm sounds
Increase in trach/airway leak. Evaluate leak, look for normal wake-sleep
trends and set alarms appropriately.
Increased minute volume due to
ventilator auto triggering from:
a. Leak. a. Check circuit for leak and correct. Perform
leak check (exhalation valve calibration) on
breathing circuit.
c. Loose expiratory drive line or c. Clip off ends of expiratory drive line and
proximal tubing connections loose. proximal line to tighten connection.
Low Paw Alarm Decreased patient resistance or Evaluate patient. Adjust the ventilation
A/CMV or SIMV modes, increased patient compliance. settings and/or Low Paw alarm as needed.
▼ Paw LED blinks red,
audible alarm sounds Leak or disconnect in the breathing Verify that connections are tight and leak
circuit and/or humidifier assembly. free.
Low Paw alarm set incorrectly. Readjust Low Paw alarm if appropriate.
Notify physician as necessary.
Low Paw Alarm-cont’d. Ptrig setting too low (most common Re-evaluate / re-adjust Ptrig setting (especially
A/CMV or SIMV modes, with single patient use exhalation after circuit change)
▼ Paw LED blinks red, valve).
audible alarm sounds
Loose expiratory drive line or proximal Clip off ends of expiratory drive line and
tubing connections loose. proximal line to tighten connection.
Circuit disconnected for airway care or Reconnect circuit securely. Push Alarm Silence
by accident. when reconnecting after airway care (to allow
one minute for stabilization).
Low Pbase Alarm Baseline pressure is below set Pbase Verify all circuit connections are secure and
▼ Paw/Apnea LED due to airway or circuit leak or fluid leak free and that all fluid is cleared from
blinks red, audible alarm pooled in tubing. tubing.
sounds
•
Low VI Alarm / BUV Patient efforts not detected. Trigger Perform leak check on breathing circuit
•
▼ VI LED blinks red, level (Ptrig) set improperly. (exhalation valve calibration) , secure circuit
audible alarm sounds connections, evaluate Ptrig setting.
Detected patient efforts are indicated by
the Ptrig LED illuminating green.
• •
The set Low VI alarm is set above Readjust Low VI alarm setting level.
delivered mandatory minute volume.
Occlusion Alarm Exhalation valve blocked or line Check exhalation valve line. Replace
▲ Paw LED blinks red kinked. exhalation valve assembly.
at the high priority rate,
audible alarm sounds Re-calibrate exhalation valve.
Occlusion Alarm Internal defect. Primary exhalation Use an alternate means of ventilation and
▲ Paw & Device Alert LED valve solenoid is malfunctioning. call NEWPORT Technical Service Dept. for
blink red at the high priority repair procedures. See page 1-6 and 1-7
rate, audible alarm sounds for contact information.
PCV Not Reached Alarm Gross leak in the patient circuit and/or Check all patient breathing circuit
Pressure Control modes humidifier assembly. connections.
▼ Paw/Apnea LED
blinks red, audible alarm Target pressure setting requires a flow Re-evaluate the ventilator settings and
sounds rate that is beyond the HT50’s strategy.
maximal flow capability.
Power Switchover Alarm External power cord is disconnected. Reinsert power cord.
Ext. Power LED red
Int. Battery LED External power source failure. Use the dual internal battery.
blinks yellow Recharge dual internal battery when A.C. is
intermittent audible alarm available.
System Error Unrecoverable internal system failure Ventilate the patient with an alternate
e.g., memory error. means of ventilation. Make note of the
message in the display window.
System Failure Alarm Unrecoverable internal system failure. Ventilate the patient with an alternate
Device Alert LED red, means of ventilation. Make note of the
continuous audible alarm message in the display window.
Check Humidifier Humidifier bottle is loose or missing. Tighten humidifier clamp. Install humidifier
Message bottle. Press Humidifier On to restart.
Check Temp Probe Alarm Temperature probe is disconnected. Insert temperature probe into port. Press
Red LED, humidifier Humidifier On to restart.
function terminated
Faulty temperature probe. Replace temperature probe. Press
Humidifier On to restart.
Humidifier Fail Alarm Humidifier bottle is loose or missing. Tighten humidifier clamp. Install humidifier
Red LED, humidifier bottle. Press Humidifier On to restart.
function terminated
Failure of the humidifier control. Contact NEWPORT Technical Service Dept.
See page 1-6 and 1-7 for contact
information.
High Temp Core Alarm Excessive temperature in the heating If occurs repeatedly, contact NEWPORT
Red LED, humidifier element. Technical Service Dept. See page 1-6 and
function terminated 1-7 for contact information.
High Prox Temp Alarm Set humidifier temperature is lower Increase set humidifier temperature. Press
Red LED, humidifier than ambient temperature. Humidifier On to restart.
function terminated
Temperature of inhaled or exhaled gas Increase humidifier temperature setting or
coming in contact with the airway move temperature probe farther away from
temperature probe is higher than the airway. Restart humidification. Press
set temperature. Humidifier On to restart.
Set temperature was recently lowered Wait for water to cool in humidifier bottle.
by a significant amount. Restart humidification. Press Humidifier On
to restart.
Alarm volume Unintended setting. To toggle between loud and quiet, clear all
too loud or too quiet alarms, then push and hold the Alarm Silence
button for 3 seconds until you hear a beep.
Battery depleted too fast Battery is not fully charged. Charge battery to full charge level. Battery
not lasting up to 10 hours charges in 5-8 hrs from A.C. or external
D.C. Check charge level by pressing Int.
Battery (Push to Test) button and view level
Note: New battery use in Paw monitor window.
time is up to 10 hours at
nominal ventilator settings Extend battery use time by plugging into
with Power Save turned A.C. or D.C. (car adapter) when available.
ON (see Specifications on Suggestion: Optional accessory,
pg. 2-4). The first Battery Automobile 12V power cord p/n HT460085,
Low alarm will occur when can be used to plug ventilator into
a mimimum of 30 minutes automobile cigarette lighter.
of operating time remains.
Battery Empty alarm will Ensure that the green Ext. Power LED is lit
occur when a mimimum of when connected to an AC or DC power
15 minutes of battery time source (it can take up to two minutes). If
is left. LED is not lit, check connections and
resolve any problems.
Power Save is OFF. This will decrease Enter User Setup and turn Power Save ON.
battery use time by 20% to 30%.
Ventilator can not enter Power Save Press Alarm Silence/Reset to clear latched
because latched alarms indicators alarm indicators.
have not been cleared.
At your ventilator settings, note the battery
Ventilation settings used create higher use time until the first Battery Low alarm
peak and mean pressures than sounds. A minimum of 30 minutes of
nominal settings as outlined in operating time remains until shutdown. This
operating manual. will be your expected battery use time at
your settings.
CO2 rises Too much deadspace (rebreathing) in On small patients, deadspace tubing
Child’s CO2 rises the breathing circuit. (On a single-limb should be as small (15 mm ID) and short as
dramatically when put on circuit, the tubing on the patient side possible.
ventilator of the exhalation valve is deadspace.
On a "J" circuit the tubing between Please Note: The rest of the breathing
the wye connector and the patient is circuit tubing should always be 22 mm ID
deadspace.) even on small children.
Circuit disconnect/ no Proximal line is not appropriately Place proximal line at closest possible
alarm sounds placed in circuit. position to patient without creating
Breathing circuit is problems with humidity in prox line or
disconnected from patient, excessive deadspace for patient.
but there is no alarm.
If a heat moisture exhanger (HME) is in use
in place of a heated humidifier, connect the
proximal line to the breathing circuit on the
patient side of the HME to improve alarm
responsiveness.
Low Paw alarm is not appropriately Set Low Paw alarm to ensure that it
set. sounds when patient circuit is
disconnected. After setting up patient and
stabilizing ventilation, remove circuit from
patient at airway and observe peak airway
pressure that develops with the next
breath. Reconnect patient and set Low
Paw alarm above this pressure.
• •
High / Low Minute Volume VI alarm Set High / Low VI alarms to bracket
limits are not appropriately set. patient minute volume.
Exhalation Valve Cal A leak in the system. a. Check all circuit connections.
Fails (Cal Failed) b. Check that the test lung is leak-free and
that it is ≤ 1 L in size.
Reusable or single patient c. Check that the exhalation valve drive line
use exhalation valve and prox line are not reversed.
d. Use your thumb (covered with a clean
gauze pad or equivalent) instead of a
test lung to occlude circuit during
calibration.
e. If using a reusable exhalation valve,
ensure that the diaphragm is seated
properly.
f. Try a different exhalation valve.
Oxygen is being titrated directly into Make sure that oxygen is not titrated
the circuit rather than using the directly into the breathing circuit. Use the
Oxygen Blending Bag Kit or Oxygen Blending Bag Kit or Air/Oxygen
Air/Oxygen Entainment Mixer. Entrainment Mixer to provide oxygen
enrichment.
Exhalation Valve Honks Low compliance/high resistance of Make sure the breathing circuit is 22 mm ID
Exhalation valve makes circuit system. (regardless of patient size) and at least 8 ft
honking noise long for small patients.
The single patient use exhalation valve Use an exhalation valve that is approved
in use is not compatible with ventilator. for use with the HT50 Ventilator.
External Power Not Power cord is not plugged far enough Check that power cord is pushed in all the
Working into the ventilator outlet. way on both ends.
After plugging into external
Check for power to A.C. outlet or use
A.C or D.C. outlet, Ext. A.C. outlet has no power. another A.C. outlet with power.
Power indicator on panel
does not light after two D.C. Auto lighter outlet is not active Make sure auto lighter outlet is active with
minutes with engine off. engine off or turn engine on.
HT50 cannot detect external power. Switch to alternate ventilator and contact
Newport Medical Technical Service Dept.
Frequency is 1.5 Times Ventilator is in Back up Ventilation in Back up Ventilation will stop and the
Set Value response to the Low Minute Volume respiratory frequency will return to normal
Ventilator is alarming and Alarm being violated. when the monitored minute volume is 10%
respiratory frequency is 1.5 above the Low Minute Volume Alarm limit.
times the original set value
Manual Inflation Button High Paw alarm setting reached during If a higher inflation pressure is needed,
Breath terminates and Manual inflation. increase the High Paw alarm limit setting to
High Paw alarm is violated a safe but appropriate level. Otherwise,
decrease flow rate or manual inflation time.
Manual Inflation Button Mandatory flow is set too low. Evaluate ventilation settings. If appropriate,
Can’t generate adequate decrease inspiratory time to increase flow.
rise in pressure
Gross leak in patient breathing circuit. Check/secure all patient breathing circuit
connections.
PEEP/CPAP Control Leak in breathing circuit, humidifier, Find and correct leak.
Monitored Pbase is less endotracheal tube cuff, patient
than set Pbase. interface. etc.
Pressure Meter Breathing circuit resistance caused by Temporarily disconnect the patient circuit
Pressure does not return an occluded filter/exhalation valve, from the ventilator GAS OUTPUT outlet. If
to zero when PEEP is set pooled water or lodged secretions the pressure meter returns to zero, the
to zero which prevent the free exit of patient cause of the elevated baseline pressure is
exhalation. circuit resistance.
Bounce/rebound from test lung Use a test lung with more physiologic
performance.
Pressure Not Rising Massive leak in the breathing circuit. Locate leak and fix it.
Ventilator sounds like it is
delivering breaths but Exhalation valve diaphragm has come Replace exhalation valve/breathing circuit.
pressure is not rising unseated.
during the breath
Settings Change When Set up: Default selected in User Set Patient must be off ventilator.
Vent Turned Off and On Up parameters. Enter User Set Up and select Set up: User.
Ventilator changes settings
when turned off and on
Trigger Problem Inappropriate Ptrig setting. Adjust the Ptrig towards "0" until the
Patient cannot trigger the ventilator auto-triggers, then slowly
ventilator increase the Ptrig setting until the auto-
triggering stops.
Pressure Support is cycling off late. Manage airway leak / fix circuit leak.
•
Trigger Problem In volume control, flow is set Check flow setting in V display window on
Patient double-triggers the inappropriately low. face panel. If it is too low for patient need,
ventilator (patient has decrease inspiratory time (tI) setting until
trouble exhaling because the flow is set appropriately.
second breath is being
delivered) Pressure support is set too low for Re-evaluate pressure support setting.
patient need.
Ventilator Makes Noise Cylinder is turned off or empty. Check that cylinder is turned on and that it
When Air/Oxygen Mixer is not empty.
Is Connected
HT50 makes a loud noise
when using the Air Oxygen
Entrainment Mixer
connected to a gas cylinder
Ventilator Pistons Move The HT50 generates a 7.5 L/min of Ventilator is operating correctly.
Between Breaths continuous flow in between breaths
Ventilator sounds like the when PEEP is > 0 cmH2O.
dual micro pistons
continue to move between
breaths
Water in Breathing Room temperature is cooler than the a. Place water trap inline with the
Circuit Tubing heated, humidified breathing gas in breathing circuit and empty it regularly.
the circuit. When the gas in the circuit b. Wrap circuit tubing with pipe insulation
cools, water precipitates out. or plastic sandwich wrap.
c. Use a heated wire circuit.
• •
Reached Max V Maximum V (flow) setting has been Increase inspiratory time (tI) or decrease
reached. tidal volume to resolve.
• •
Reached Min V Minimum V (flow) setting has been Decrease inspiratory time (tI) or increase
reached. tidal volume to resolve.
Inverse I:E Inverse I:E Ratio has been reached. Decrease inspiratory time (tI) or decrease
frequency (ƒ) to resolve.
Reached Max I:E Inverse I:E Ratio has reached 3:1 Decrease inspiratory time (tI) or decrease
frequency (ƒ) to resolve.
•
• V (flow) display is not available in
V Unavailable •
Pressure Control because V is variable
and regulated breath by breath.
PEEP + PS Too High Set PEEP + Psupport is higher than 60. Adjust PEEP or Psupport to resolve.
PC – PEEP Too Low Pressure Control value minus PEEP is Adjust PEEP or Pressure Control to
less than 5 cmH2O/mbar. resolve.
앖 – PEEP Too Low High Pressure alarm limit minus PEEP Adjust PEEP or High Paw alarm setting to
is less than 5 cmH2O/mbar. resolve.
Mixer makes a pronounced Oxygen source gas pressure is low. Check that oxygen source gas is 50 psig.
clicking sound during
normal operation Oxygen source regulator is oscillating. Check oxygen source regulator. If noise
continues contact Newport Technical
Service Dept.
Oxygen leaks out of Mixer diaphragm is leaking. Contact Newport Technical Service Dept.
Mixer when connected
to 50 psig oxygen gas
source.
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Set Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Standard Test Settings . . . . . . . . . . . . . . . . . . . . . A-2
Quick Check Procedure . . . . . . . . . . . . . . . . . . . . A-2
Pass / Fail Check Off Sheet . . . . . . . . . . . . . . . . . A-5
Abbreviated Check Procedure . . . . . . . . . . . . . . . A-6
OPRHT50NA B0908
APPENDIX A
INTRODUCTION
This procedure is intended to assist qualified operators to
establish a routine program for verifying proper HT50 operation.
Perform this Quick Check procedure each time the ventilator is
prepared for clinical use.
NOTE: When ventilator changes are made allow time for the new
settings to stabilize. In certain conditions (e.g., VT adjustment),
the HT50 may take five or six breaths to make the adjustment.
Pre-test Inspection
1. Remove the three screws from the Filter Cover. Inspect the
filter. Replace the filter if it is dirty. Re-install screws.
Set Up
Control Setting
MODE A/CMV
Volume Control 500 mL
tI 1.0 sec
ƒ 15 b/min
P trig (–)1 cmH2O / mbar
▼Paw 3 cmH2O / mbar
▲Paw 99 cmH2O / 97 mbar
▼VI alarm
•
0.1 L (minimum setting)
▲ VI alarm
•
50 L (maximum setting)
PEEP/CPAP 0 cmH2O / mbar
P support 0 cmH2O / mbar
Humidifier Off
• Reconnect the A.C. power cord. Verify that within two minutes
the “Int. Battery” indicator turns off and the Ext.
Power/Charging indicator turns green. Press the Silence/Reset
button to clear the Message Display Window.
• Disconnect the test lung from the breathing circuit and verify
that after two breaths an audible and visual Low Paw alarm
occurs.
• Attach the test lung to the breathing circuit and verify that the
audible alarm ceases and the ▼Paw LED latches. Press
Silence / Reset button to clear the alarm and the message in
the Message Display Window.
• Press the ▲Up control until “VT/ V i/f” is displayed. Verify that
•
•
VT= 450-550, V i = 6-9 and ƒ = 13-17 is displayed.
• Unplug the HT50 from AC power, clear the alarm with the
Alarm/Silence button and then press and hold the Int. Battery
(Push to Test) button while powered on internal battery. Verify
that the Int. Battery charge level needle on the Paw (airway
pressure) meter is in the blue zone. This indicates the dual
internal battery is charged.
The ventilator is ready for operation when all tests have been completed
successfully.
Note any comments on inspection of unit, corrective action taken, or recommendations for
further action.
1. Remove the three screws from the inlet filter cover. Inspect
the filter. Replace if dirty. Re-install screws.
4. The HT50 will start the EZ Cal and the ventilator will
automatically test the exhalation valve. If it passes the test,
the messages "Cal Completed", then "Press ON to Vent" will
be displayed.
4a. If the test fails, the message "Cal Failed" will be displayed.
4c. If using a test lung during the EZ Cal, remove the test lung
and occlude the patient connection instead.
4d. Check the integrity of the circuit and connections, then press
the Manual Inflation button twice to initiate calibration again.
Operation Verification
OPRHT50NA B0908
APPENDIX B
HUMIDIFIER ACCESSORIES
1. HT460910 Humidifier Bottle – upper 1
2. HT460920 Humidifier Bottle – lower 1
3. HT460930 Humidifier – heat sink 1
4. HT460940 Humidifier – airway temperature probe 1
5. HT460950 Humidifier – hose connection (12 in) 1
6. V12-34001-62 Humidifier – absorbent paper 10/ea
AIR/OXYGEN MIXER
7. HT460700 Air/oxygen mixer – DISS 1240 1
8. HT600044 Oxygen blending bag kit (1L) 1
MOUNTING ACCESSORIES
21. HT460231 Compact stand assembly 1
with support arm and breathing circuit hanger
22. HT460200 Rail / bed bracket 1
continued
Contact NEWPORT MEDICAL or your local NEWPORT MEDICAL representative for pricing and
availability on all accessories and parts listed.
OPRHT50NA B0908 B-1
HT50 ACCESSORIES
MOUNTING ACCESSORIES
23. HTBKT01 Bracket, humidifier pole mount (for stand) 1
24. HTECH01 Bracket, e-cylinder holder (for stand) 1
25. HTECH02 Bracket, dual e-cylinder holder (for stand) 1
MISCELLANEOUS ACCESSORIES
26. HT460300 Air inlet particle filter (disposable) 5/ea
27. LNG500A Adult test lung (500 mL) with R20 restrictor 1
28. V24-00900-60 Electrical power cord / A.C. with Ferrite 1
29. HT460400 Power cord ferrite 1
30. HT460085 Automobile 12 V accessory power cord 1
31. OPRHT50NA Operating manual for HT50 1
32. SERHT50NA Service manual for HT50 1
33. HT6004701 Prox inline filters (disposable) 5/ea
34. BAT3300A External Batttery Pack Assy 1
35. CHG3313P Battery Charger for BAT3300A 1
Contact NEWPORT MEDICAL or your local NEWPORT MEDICAL representative for pricing and
availability on all accessories and parts listed.
OPRHT50NA B0908
APPENDIX C
2. Why do I need to place a filter inline with the proximal pressure tubing?
6. Is it possible to use a heat moisture exchanger (HME) with the HT50’s single-limb breathing
circuit? Is the continuous flow generated when PEEP is added a problem?
Exhalation Valve I
Calibration 5-3
Cleaning 6-6 I:E Ratio 3-12
Exhalation Valve Connector (Exh. Valve) 3-25 Inlet Filter. See Air Inlet Particle Filter
External Power Connector (A.C./D.C.) 3-26 Inspiratory Minute Volume 3-15, 3-23
Inspiratory Time. See tI
Intended Use 2-1
F Internal Battery, Dual
Factory Default Parameters 2-6 Alarms 3-20 - 3-21
Filters. See Air Inlet Filter, Prox. Inline Filter Charge Level Meter 3-13
Flow Rate. See V
•
General Warnings 1-5
ƒ (frequency) 3-8, 3-23 Maintenance 6-10
Fresh Gas Intake 3-25 Recharge Time 3-20
Front Panel Specifications 2-4 – 2-5
Alarms 3-14 Test Button 3-12
Controls & Indicators 3-5 Inverse I:E Message 3-24
Photo 3-3, 3-4 Inverse I:E Ratio 3-12
Message Display Window 3-23
Overview 3-1 L
Left Side Connectors 3-24
G Low Baseline Pressure Alarm 3-18
Gas Output Connector 3-24 Low Insp. Minute Volume Alarm
Alarm 3-16
Back-up Ventilation. See Back-up Ventilation
H Control & Display 3-15
Hardware Requirements 2-5 Low Pressure Alarm
High Baseline Pressure Alarm 3-17 Alarm 3-15
High Insp. Minute Volume Alarm Control & Display 3-14
Alarm 3-16
Control & Display 3-15 M
High Pressure Alarm
Alarm 3-15 Maintenance 6-9
•
Control & Display 3-14 Mandatory Flow. See V
High Prox Temp Alarm 3-19 Manual Inflation 3-11
High Temp Core Alarm 3-19 Mean Paw 3-23
Home Caregiver Notes & Warnings 3-25, 3-26, Message Display Window 3-23
3-27, 3-29, 5-1, 5-4, 5-8, 5-9, 5-10, 6-5, 6-8, Alarm Messages 3-24
6-9, 7-1, A-1, A-6 Monitoring Messages 3-22
Humidifier (HT50-H, HT50-HB) 5-10 Setting Limitations Messages 3-24
Alarms 3-19 Mixer. See Air/Oxygen Entrainment Mixer
Cleaning 6-3 Modes
General Description 5-10 A/CMV 3-6, 4-2
Prep For Use 5-11 How to change 3-1
Set Up And Operation 5-12 SIMV 3-6, 4-3
Specifications 2-7 SPONT 3-7, 4-3
Warnings & Precautions 5-11 Modes Control 3-5
Humidifier Fail Alarm 3-19 Motor Fault 3-21
Humidifier On Button 3-11
O Q
Occlusion Alarm 3-17 Quick Check Procedure A-1 to A-4
Circuit 3-17 Quick Check Procedure Check-Off Sheet A-5
Device 3-17
On/Standby Button 3-5 R
Operator’s Responsibility 1-1
Oxygen Accessories. See Air/Oxygen Reached Max I:E Message 3-24
•
Entrainment Mixer, Oxygen Blending Bag Reached Max V Message 3-24
•
Oxygen Blending Bag 3-27, 5-14 Reached Min V Message 3-24
Assembly Diagram 5-15 Respiratory Frequency. See ƒ
Cleaning 6-2 Right Side Connectors 3-25
Disassembly 6-2 RS-232 Interface (Comm. Port) 3-25
Installation 5-15
Oxygen Supply Flow Graphs 5-16 S
Specifications 2-8
Warnings & Precautions 5-14 Sensitivity. See Ptrig
Oxygen Monitoring 5-12 Set Up & Use 5-1 to 5-17
Air/Oxygen Entrainment Mixer 5-13
P Humidifier 5-10
Oxygen Blending Bag 5-14
Panel Lock. See Auto Panel Lock Patient 5-8
Parameter Settings 2-6 Ventilator 5-1
Parts & Accessories List B-1 Setting Limitation Messages 3-24
Patient Range 2-6 Shut Down Alert 3-22
Patient Set Up Procedure. See Set Up & Use Shut Down, How To 3-1
Paw Meter 3-14, 3-28 Silence/Reset 3-13
Paw, Base 3-23 SIMV (Synchronized Intermittent Mandatory
Paw, Mean 3-23 Ventilation) 3-6, 4-3
Paw, Peak 3-23 Specifications 2-2
PC – Peep Too Low 3-24 Air/Oxygen Mixer 2-8
PCV Not Reached Alarm 3-19 Controls/Alarms/Monitors 2-2
Peep + PS Too High 3-24 Hardware 2-5
PEEP/CPAP 3-10 Humidifier 2-7
Power Cord Ferrite 3-26, 5-2 Miscellaneous 2-5
Power Save 3-28 Oxygen Blending Bag 2-8
Power Supply Requirement 2-5, 3-26 SPONT Mode 3-7, 4-3
Power Switchover Alarm 3-21 Standard Test Settings A-2
Pressure Control 3-9, 4-4 Start Up Parameter Setting. See User Set Up
Switch To Volume Control 3-1, 3-8 Symbols/Labeling Table 2-1
Pressure Support. See Psupport System Error 3-21
Pressure Units 3-28 System Overview 4-1
Preventive Maintenance 6-9
Proximal Inline Filter 5-5, 6-9 T
Proximal Line. See Assembly Diagram
Psupport (Pressure Support) 3-10, 4-4 Target Pressure. See Pressure Control
Ptrig (Sensitivity) 3-10 Tech. Set Up. See User Set Up
Push to Unlock Button 3-13 Temperature Probe Connector (Temp. Probe) 3-
25
Theory of Operation 4-1 to 4-6
tI (inspiratory time) 3-8
Tidal Volume. See Volume Control
U
Unlock Panel. See Auto Panel Lock
Up and Down Control 3-7
User Set Up 3-27
Exiting 3-29
How to access 3-27
Parameters 3-28
V
•
V (Mandatory Flow) 3-12
•
V Unavailable Message 3-24
Ventilator Set Up. See Set Up & Use
Volume Control 3-8, 4-5
Switch To Pressure Control 3-1, 3-9
W
General Warnings 1-4
Warranty 1-2
Water Trap. See Breathing Circuits