0% found this document useful (0 votes)
19 views3 pages

CT Midterm

Uploaded by

mrgrttcorpuz
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
19 views3 pages

CT Midterm

Uploaded by

mrgrttcorpuz
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
You are on page 1/ 3

CT scan

Constants:

High frequency generator - most powerful Single phase - 1


voltage, expensive
3 phase, 6 pulse - 1.35
3 phase, 12 pulse - most common generator
3 phase, 12 pulse - 1.41
Voltage ripple - (ripple means fluctuation,
HFG - 1.45
changes) insability of volatage

Formula: HU= mA x kVp x constant


Single pulse- 100% (High Voltage Generator)
EX:
3 phase, 6 pulse - 86-87 % (HVG)
GIVEN:
3 phase, 12 pulse - 96-97%(HVG), voltage =3-4
percent constant) 200 mA
High frequency generator - 1% or <1%(less than 0.05 s
1 percent), voltage = almost constant
80kVp

3, 12 pulse (1.41)

Large focal spot (1mm) - for high power ratings


(up to 60 kW) = 200x80x1.41 = 22,560HU

Small focal spot (0.6mm) for low power rating


<25kw Housing - 1-2 MHU

Anode 350,000 MHU


Heat capacity of xray tube in general
radiography
Filters - copper or aluminum filter are used
Max. - 1-2 mega heat units (1-2 MHU)
- to reduce xray
Heat units (HU) - unit of heat capacity of the
xray tube (special unit) - 70 kVp = thickness of aluminum filtration
should be 2.5 mm (total filtration )
Formula: HU= mA x kVp x constant
- 1.5 for added filtration
Joule (J)- Unit for measuring heat ( for energy or
heat)

Calorie - most commonly use for measurinv Section thickness - pre patient collimator (1-
heat 10mm)

Electron volts (eV) - sepcial unit, use to refer to 5mm - most common section thickness
teb energy of the IR
1 mm - for very high resolution 3,500

Thinner the slice increases the scan time - standard revolution per minute (RPM) of
anode
HU = mA x kVp x constant
- most common RPMS of the rotor

10,000

- maximum RPM
anode - 350,000 HU (350 kHU)
- why use high RPMS: the rotor helps to
1200 mA - maximum mA applicable for x-ray
disperse heat
15 mins - waiting time to cool down
better resolution - small focal spot (longer
tube housing exposure time)

- 1-2 MHU

- 1-2 hrs 1HU = 0.74J

- 60-90 mins

3 most common causes of x-ray tube failure: purpose of collimator

1. long exposure time - it will lead to damage to 1. reduce scatter radiation


bearings
2. to control the amount of x-rays reaching the
2. single excessive exposure - relieve to arcing IR
of the anode
2 types of collimators
3. tungsten vaporization - most common cause
1. pre patient
arcing
- controls the slice thickness and px dose
- ditting or cracking
2. post patient
3,410 deg. C
- does not control px dose
- melting of point tungsten
- also called pre-detector collimator
pyrex glass
2 types of detector
- common material used in housing
1. gas detector
negative - charge of filament
- xenon: commonly used
positive - charge of anode
2. scintillation detector
space charge
- cadmium tungstate: commonly used
- cloud of electron
- sodium iodide-thallium activated crystals
thermionic - emission of (NaTl): first scintillation crystal used
a. cesium iodide

b. bismuth germanate

c. calcium fluoride

d. cadmium tungstate

gas detectors are naturally occuring

- not commonly used because of hygroscopic

- hygroscopic: ability to absorb moisture

solid state needs to formulate

- most commonly used today

aluminum, copper - material of container

fan beam (1-10m)

- image slice

photomultiplier tube (PMT)

- multiplies the photons

3 main portions of PMT

1. photocathode

- purpose: converts x-rays to light

- light to e-

2. dynode

- multiply the electrons

- up to x13

3. collecting anode

- collect the electrons

analog-to-digital converter (ADC)

- converts electrical signal into image

scintillant

- other term of scintillation crystals

for gas detector:

tube —> px —> gas —> ionization —> signal

You might also like